🇵🇱 "Energopomiar" Sp. z o.o. PL
Metrology & calibration
"Energopomiar" Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 131 (Gliwice, Poland). Its accreditation scope covers 25 calibration items in the following areas: Metrology & calibration. Calibrated items include: Calibrators, Multimeters, Digital ammeters, Digital voltmeters. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Direct method; At permanent location (S), Internal procedure; Direct method; At permanent location (S) and ou….
At a glance
- Accreditation number: PCA AP 131
- Calibration items in scope: 25
- Types of calibrated items: 26
- Calibration principles used: 4
- City: Gliwice, Poland
- Valid until: 22-12-2029
- Accreditation number
- AP 131
- Accreditation body
- PCA
- Address
- ul. Józefa Sowińskiego 3, 44-121 Gliwice
- City
- Gliwice, Poland
- Division
- Metrology Laboratory
- Valid until
- 22-12-2029
What this laboratory calibrates (26)
- Calibrators — 6 calibration items
- Multimeters — 6 calibration items
- Digital ammeters — 4 calibration items
- Digital voltmeters — 4 calibration items
- Electronic pressure gauges — 2 calibration items
- Pressure transmitters — 2 calibration items
- Temperature indicators (meters), including temperature controllers — 2 calibration items
- Base metal thermocouples (K, N) — 1 calibration item
- Bourdon tube pressure gauges — 1 calibration item
- Clamp meters — 1 calibration item
- Digital ohmmeters — 1 calibration item
- Electrical thermometers (including electronic) — 1 calibration item
- Frequency meters — 1 calibration item
- Noble metal thermocouples (S) — 1 calibration item
- Resistance thermometer sensors — 1 calibration item
- and 11 more
Calibration principles (4)
- Internal procedure; At permanent location (S) — 7 calibration items
- Internal procedure; Direct method; At permanent location (S) — 6 calibration items
- Internal procedure; Direct method; At permanent location (S) and outside of it (P) — 6 calibration items
- Internal procedure; Indirect method; At permanent location (S) and outside of it (P) — 6 calibration items
Certificate information
- Issue date
- 26.11.2025
- Issue no.
- 18
- Certificate date
- 19.02.2025
- Accreditation cycle
- 26.11.2025 — 22.12.2029
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (25)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Multimeters Digital voltmeters Digital ammeters |
Calibration — DC voltage 2 mV to 200 mV · CMC 3,5 · 10⁻⁵ · U + 0,0058 mV 0,2 V to 2 V · CMC 3,5 · 10⁻⁵ · U + 0,000012 V 2 V to 20 V · CMC 3,5 · 10⁻⁵ · U + 0,00012 V 20 V to 200 V · CMC 3,5 · 10⁻⁵ · U + 0,0013 V 200 V to 1000 V · CMC 3,7 · 10⁻⁵ · U + 0,0053 V DC voltage |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | I_AM_04 based on EURAMET cg-15 v.3.0 | |
| 2 | Calibrators | Calibration — DC voltage 2 mV to 20 mV · CMC 0,03 % 20 mV to 2 V · CMC 0,008 % 2 V to 1000 V · CMC 0,01 % DC voltage |
Internal procedure; Direct method; At permanent location (S) | I_AM_05 based on EURAMET cg-15 v.3.0 | |
| 3 | Multimeters Digital voltmeters Digital ammeters |
Calibration — DC current 0,01 mA to 0,2 mA · CMC 5,6 · 10⁻⁵ · I + 0,000058 mA 0,2 mA to 2 mA · CMC 1,2 · 10⁻⁴ · I + 0,00013 mA 2 mA to 20 mA · CMC 1,2 · 10⁻⁴ · I + 0,0012 mA 20 mA to 200 mA · CMC 1,4 · 10⁻⁴ · I + 0,011 mA 0,2 A to 2 A · CMC 2,3 · 10⁻⁴ · I + 0,00012 A 2 A to 20 A · CMC 6 · 10⁻⁴ · I + 0,0045 A DC current |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | I_AM_04 based on EURAMET cg-15 v.3.0 | |
| 4 | Calibrators | Calibration — DC current 0,2 mA to 2 mA · CMC 0,2 % 2 mA to 4 mA · CMC 0,07 % 4 mA to 10 mA · CMC 0,04 % 10 mA to 100 mA · CMC 0,02 % 100 mA to 1 A · CMC 0,1 % 1 A to 10 A · CMC 0,16 % DC current |
Internal procedure; Direct method; At permanent location (S) | I_AM_05 based on EURAMET cg-15 v.3.0 | |
| 5 | Multimeters Digital voltmeters Digital ammeters |
Calibration — AC voltage f = 20 Hz to 20 kHz — 2 mV to 20 mV · CMC 3,7 · 10⁻⁴ · U + 0,014 mV f = 20 Hz to 20 kHz — 20 mV to 200 mV · CMC 4 · 10⁻⁴ · U + 0,022 mV f = 20 Hz to 20 kHz — 0,2 V to 2 V · CMC 3,6 · 10⁻⁴ · U + 0,00012 V f = 20 Hz to 20 kHz — 2 V to 20 V · CMC 3,6 · 10⁻⁴ · U + 0,0012 V f = 40 Hz to 20 kHz — 20 V to 200 V · CMC 3,6 · 10⁻⁴ · U + 0,012 V f = 40 Hz to 5 kHz — 200 V to 750 V · CMC 4,4 · 10⁻⁴ · U + 0,03 V AC voltage |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | I_AM_04 based on EURAMET cg-15 v.3.0 | |
| 6 | Calibrators | Calibration — AC voltage f = 20 Hz to 10 kHz — 2 mV to 20 mV · CMC 0,4 % f = 20 Hz to 10 kHz — 20 mV to 20 V · CMC 0,3 % f = 20 Hz to 10 kHz — 20 V to 750 V · CMC 0,17 % AC voltage |
Internal procedure; Direct method; At permanent location (S) | I_AM_05 based on EURAMET cg-15 v.3.0 | |
| 7 | Multimeters Digital voltmeters Digital ammeters |
Calibration — AC current f = 20 Hz to 5 kHz — 0,01 mA to 0,2 mA · CMC 1,5 · 10⁻⁴ · I + 0,00058 mA f = 20 Hz to 5 kHz — 0,2 mA to 2 mA · CMC 6,1 · 10⁻⁴ · I + 0,00048 mA f = 20 Hz to 5 kHz — 2mA to 20 mA · CMC 7,2 · 10⁻⁴ · I + 0,0016 mA f = 20 Hz to 5 kHz — 20 mA to 200 mA · CMC 6,5 · 10⁻⁴ · I + 0,019 mA f = 20 Hz to 5 kHz — 0,2 A to 2 A · CMC 6,8 · 10⁻⁴ · I + 0,00024 A f = 40 Hz to 1 kHz — 2 A to 20 A · CMC 9,9 · 10⁻⁴ · I + 0,007 A AC current |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | I_AM_04 based on EURAMET cg-15 v.3.0 | |
| 8 | Calibrators | Calibration — AC current f = 40 Hz to 5 kHz — 0,2 mA to 2 mA · CMC 0,6 % f = 40 Hz to 5 kHz — 2 mA to 10 A · CMC 0,3 % AC current |
Internal procedure; Direct method; At permanent location (S) | I_AM_05 based on EURAMET cg-15 v.3.0 | |
| 9 | Multimeters Digital ohmmeters |
Calibration — DC resistance 1 Ω to 1,9 MΩ · CMC 0,012 % 1,9 MΩ to 19 MΩ · CMC 0,035 % DC resistance |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | I_AM_04 based on EURAMET cg-15 v.3.0 | |
| 10 | Calibrators | Calibration — DC resistance 1 Ω to 10 Ω · CMC 0,05 % 10 Ω to 1 MΩ · CMC 0,02 % 1 MΩ to 10 MΩ · CMC 0,05 % 10 MΩ to 100 MΩ · CMC 1 % DC resistance |
Internal procedure; Direct method; At permanent location (S) | I_AM_05 based on EURAMET cg-15 v.3.0 | |
| 11 | Clamp meters Single-phase digital active power meters Three-phase digital active power meters |
Calibration — AC power f = 50 Hz; U = 10 V to 700 V; I = 0,1 A to 50 A; cos φ = 1 — 10 W to 94,5 kW · CMC 0,18 % AC power |
Internal procedure; At permanent location (S) | I_AM_09 based on EURAMET cg-15 v.3.0 | |
| 12 | Single-phase digital apparent power meters Three-phase digital apparent power meters |
Calibration — AC power f = 50 Hz; U = 10 V to 700 V; I = 0,1 A to 50 A — 10 VA to 94,5 kVA · CMC 0,18 % AC power |
Internal procedure; At permanent location (S) | I_AM_09 based on EURAMET cg-15 v.3.0 | |
| 13 | Temperature transmitters (for resistance thermometer sensors) | Calibration — Electrical simulation of quantities -200 °C to 200 °C · CMC 0,17 °C 200 °C to 500 °C · CMC 0,22 °C 500 °C to 850 °C · CMC 0,30 °C Electrical simulation of quantities |
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) | I_AM_03 based on EURAMET cg-11 v.2.0 | |
| 14 | Temperature transmitters (for thermocouples) | Calibration — Electrical simulation of quantities -200 °C to 0 °C · CMC 0,40 °C 0 °C to 200 °C · CMC 0,26 °C 200 °C to 1000 °C · CMC 0,35 °C 1000 °C to 1820 °C · CMC 0,44 °C Electrical simulation of quantities |
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) | I_AM_03 based on EURAMET cg-11 v.2.0 | |
| 15 | Temperature simulators (for resistance thermometer sensors) | Calibration — Electrical simulation of quantities -200 °C to 200 °C · CMC 0,04 °C 200 °C to 500 °C · CMC 0,08 °C 500 °C to 850 °C · CMC 0,13 °C Electrical simulation of quantities |
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) | I_AM_03 based on EURAMET cg-11 v.2.0 | |
| 16 | Temperature simulators (for thermocouples) | Calibration — Electrical simulation of quantities -200 °C to 0 °C · CMC 0,36 °C 0 °C to 200 °C · CMC 0,16 °C 200 °C to 1000 °C · CMC 0,20 °C 1000 °C to 1820 °C · CMC 0,29 °C Electrical simulation of quantities |
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) | I_AM_03 based on EURAMET cg-11 v.2.0 | |
| 17 | Temperature indicators (meters), including temperature controllers Temperature indicators (meters) for resistance thermometer sensors |
Calibration — Electrical simulation of quantities -200 °C to 200 °C · CMC 0,08 °C 200 °C to 500 °C · CMC 0,15 °C 500 °C to 850 °C · CMC 0,20 °C Electrical simulation of quantities |
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) | I_AM_03 based on EURAMET cg-11 v.2.0 | |
| 18 | Temperature indicators (meters), including temperature controllers Temperature indicators (meters) for thermocouples |
Calibration — Electrical simulation of quantities -200 °C to 0 °C · CMC 0,36 °C 0 °C to 200 °C · CMC 0,16 °C 200 °C to 1000 °C · CMC 0,20 °C 1000 °C to 1820 °C · CMC 0,29 °C Electrical simulation of quantities |
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) | I_AM_03 based on EURAMET cg-11 v.2.0 | |
| 19 | Multimeters Frequency meters |
Calibration — Frequency 20 Hz to 99,999 Hz · CMC 1 · 10⁻⁵ · f + 0,0014 Hz 100 Hz to 999,99 Hz · CMC 1,1 · 10⁻⁵ · f + 0,013 Hz 1 kHz to 9,9999 kHz · CMC 1,1 · 10⁻⁵ · f + 0,00013 kHz 10 kHz to 99,999 kHz · CMC 1,1 · 10⁻⁵ · f + 0,0013 kHz Frequency |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | I_AM_04 | |
| 20 | Calibrators | Calibration — Frequency 20 Hz to 100 kHz · CMC 0,01 % Frequency |
Internal procedure; Direct method; At permanent location (S) | I_AM_05 | |
| 21 | Bourdon tube pressure gauges Electronic pressure gauges Pressure transmitters |
Calibration — Pressure - gauge pressure (medium – gas) — -95 kPa to -60 kPa · CMC 0,02 % - gauge pressure (medium – gas) — -60 kPa to 60 kPa · CMC 0,012 kPa - gauge pressure (medium – gas) — 60 kPa to 3500 kPa · CMC 0,02 % - gauge pressure (medium – liquid) — 100 kPa to 600 kPa · CMC 0,12 kPa - gauge pressure (medium – liquid) — 0,6 MPa to 70 MPa · CMC 0,02 % Pressure |
Internal procedure; At permanent location (S) | I_AM_02 based on EURAMET cg-17 v. 4.1 | |
| 22 | Electronic pressure gauges Pressure transmitters |
Calibration — Pressure - absolute pressure (medium – gas, liquid) — 7 kPa to 250 kPa · CMC 0,07 kPa - absolute pressure (medium – gas, liquid) — 250 kPa to 700 kPa · CMC 0,2 kPa - absolute pressure (medium – gas, liquid) — 700 kPa to 70100 kPa · CMC 0,02 % Pressure |
Internal procedure; At permanent location (S) | I_AM_02 based on EURAMET cg-17 v. 4.1 | |
| 23 | Resistance thermometer sensors | Calibration — Temperature (electrical thermometry) -10 °C to 250 °C · CMC 0,06 °C 250 °C to 400 °C · CMC 0,37 °C 400 °C to 640 °C · CMC 0,83 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | I_AM_06 based on DKD-R 5-1 (09/2018) | |
| 24 | Noble metal thermocouples (S) Base metal thermocouples (K, N) |
Calibration — Temperature (electrical thermometry) -10 °C to 160 °C · CMC 0,7 °C 160 °C to 199 °C · CMC 0,9 °C 199 °C to 400 °C · CMC 1,4 °C 400 °C to 600 °C · CMC 1,9 °C 600 °C to 900 °C · CMC 2,2 °C 900 °C to 1100 °C · CMC 2,6 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | I_AM_07 based on EURAMET cg-8 v.3.1 | |
| 25 | Electrical thermometers (including electronic) Temperature transmitters (containing temperature sensors) |
Calibration — Temperature (electrical thermometry) -10 °C to 250 °C · CMC 0,06 °C 250 °C to 400 °C · CMC 0,37 °C 400 °C to 640 °C · CMC 0,83 °C 640 °C to 1100 °C · CMC 2,3 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | I_AM_08 based on EURAMET cg-8 v.3.1, DKD-R 5-1 (09/2018) | |
Reference standards (13)
DKD-R 5-1 (09/2018)
EURAMET cg-11 v.2.0
EURAMET cg-15 v.3.0
EURAMET cg-17 v. 4.1
EURAMET cg-8 v.3.1
I_AM_02
I_AM_03
I_AM_04
I_AM_05
I_AM_06
I_AM_07
I_AM_08
I_AM_09