🇵🇱 "Energopomiar" Sp. z o.o. PL

PCA AP 131 • Gliwice, Poland • 25 calibration items

Metrology & calibration

"Energopomiar" Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 131 (Gliwice, Poland). Its accreditation scope covers 25 calibration items in the following areas: Metrology & calibration. Calibrated items include: Calibrators, Multimeters, Digital ammeters, Digital voltmeters. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Direct method; At permanent location (S), Internal procedure; Direct method; At permanent location (S) and ou….

At a glance

  • Accreditation number: PCA AP 131
  • Calibration items in scope: 25
  • Types of calibrated items: 26
  • Calibration principles used: 4
  • City: Gliwice, Poland
  • Valid until: 22-12-2029
Accreditation number
AP 131
Accreditation body
PCA
Address
ul. Józefa Sowińskiego 3, 44-121 Gliwice
City
Gliwice, Poland
Division
Metrology Laboratory
Valid until
22-12-2029

What this laboratory calibrates (26)

Calibration principles (4)

Certificate information

Issue date
26.11.2025
Issue no.
18
Certificate date
19.02.2025
Accreditation cycle
26.11.2025 — 22.12.2029
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (25)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. A CMC uncertainty value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Multimeters
Digital voltmeters
Digital ammeters
Calibration — DC voltage
2 mV to 200 mV · CMC 3,5 · 10⁻⁵ · U + 0,0058 mV 0,2 V to 2 V · CMC 3,5 · 10⁻⁵ · U + 0,000012 V 2 V to 20 V · CMC 3,5 · 10⁻⁵ · U + 0,00012 V 20 V to 200 V · CMC 3,5 · 10⁻⁵ · U + 0,0013 V 200 V to 1000 V · CMC 3,7 · 10⁻⁵ · U + 0,0053 V
DC voltage
Internal procedure; Direct method; At permanent location (S) and outside of it (P) I_AM_04 based on EURAMET cg-15 v.3.0
2 Calibrators Calibration — DC voltage
2 mV to 20 mV · CMC 0,03 % 20 mV to 2 V · CMC 0,008 % 2 V to 1000 V · CMC 0,01 %
DC voltage
Internal procedure; Direct method; At permanent location (S) I_AM_05 based on EURAMET cg-15 v.3.0
3 Multimeters
Digital voltmeters
Digital ammeters
Calibration — DC current
0,01 mA to 0,2 mA · CMC 5,6 · 10⁻⁵ · I + 0,000058 mA 0,2 mA to 2 mA · CMC 1,2 · 10⁻⁴ · I + 0,00013 mA 2 mA to 20 mA · CMC 1,2 · 10⁻⁴ · I + 0,0012 mA 20 mA to 200 mA · CMC 1,4 · 10⁻⁴ · I + 0,011 mA 0,2 A to 2 A · CMC 2,3 · 10⁻⁴ · I + 0,00012 A 2 A to 20 A · CMC 6 · 10⁻⁴ · I + 0,0045 A
DC current
Internal procedure; Direct method; At permanent location (S) and outside of it (P) I_AM_04 based on EURAMET cg-15 v.3.0
4 Calibrators Calibration — DC current
0,2 mA to 2 mA · CMC 0,2 % 2 mA to 4 mA · CMC 0,07 % 4 mA to 10 mA · CMC 0,04 % 10 mA to 100 mA · CMC 0,02 % 100 mA to 1 A · CMC 0,1 % 1 A to 10 A · CMC 0,16 %
DC current
Internal procedure; Direct method; At permanent location (S) I_AM_05 based on EURAMET cg-15 v.3.0
5 Multimeters
Digital voltmeters
Digital ammeters
Calibration — AC voltage
f = 20 Hz to 20 kHz — 2 mV to 20 mV · CMC 3,7 · 10⁻⁴ · U + 0,014 mV f = 20 Hz to 20 kHz — 20 mV to 200 mV · CMC 4 · 10⁻⁴ · U + 0,022 mV f = 20 Hz to 20 kHz — 0,2 V to 2 V · CMC 3,6 · 10⁻⁴ · U + 0,00012 V f = 20 Hz to 20 kHz — 2 V to 20 V · CMC 3,6 · 10⁻⁴ · U + 0,0012 V f = 40 Hz to 20 kHz — 20 V to 200 V · CMC 3,6 · 10⁻⁴ · U + 0,012 V f = 40 Hz to 5 kHz — 200 V to 750 V · CMC 4,4 · 10⁻⁴ · U + 0,03 V
AC voltage
Internal procedure; Direct method; At permanent location (S) and outside of it (P) I_AM_04 based on EURAMET cg-15 v.3.0
6 Calibrators Calibration — AC voltage
f = 20 Hz to 10 kHz — 2 mV to 20 mV · CMC 0,4 % f = 20 Hz to 10 kHz — 20 mV to 20 V · CMC 0,3 % f = 20 Hz to 10 kHz — 20 V to 750 V · CMC 0,17 %
AC voltage
Internal procedure; Direct method; At permanent location (S) I_AM_05 based on EURAMET cg-15 v.3.0
7 Multimeters
Digital voltmeters
Digital ammeters
Calibration — AC current
f = 20 Hz to 5 kHz — 0,01 mA to 0,2 mA · CMC 1,5 · 10⁻⁴ · I + 0,00058 mA f = 20 Hz to 5 kHz — 0,2 mA to 2 mA · CMC 6,1 · 10⁻⁴ · I + 0,00048 mA f = 20 Hz to 5 kHz — 2mA to 20 mA · CMC 7,2 · 10⁻⁴ · I + 0,0016 mA f = 20 Hz to 5 kHz — 20 mA to 200 mA · CMC 6,5 · 10⁻⁴ · I + 0,019 mA f = 20 Hz to 5 kHz — 0,2 A to 2 A · CMC 6,8 · 10⁻⁴ · I + 0,00024 A f = 40 Hz to 1 kHz — 2 A to 20 A · CMC 9,9 · 10⁻⁴ · I + 0,007 A
AC current
Internal procedure; Direct method; At permanent location (S) and outside of it (P) I_AM_04 based on EURAMET cg-15 v.3.0
8 Calibrators Calibration — AC current
f = 40 Hz to 5 kHz — 0,2 mA to 2 mA · CMC 0,6 % f = 40 Hz to 5 kHz — 2 mA to 10 A · CMC 0,3 %
AC current
Internal procedure; Direct method; At permanent location (S) I_AM_05 based on EURAMET cg-15 v.3.0
9 Multimeters
Digital ohmmeters
Calibration — DC resistance
1 Ω to 1,9 MΩ · CMC 0,012 % 1,9 MΩ to 19 MΩ · CMC 0,035 %
DC resistance
Internal procedure; Direct method; At permanent location (S) and outside of it (P) I_AM_04 based on EURAMET cg-15 v.3.0
10 Calibrators Calibration — DC resistance
1 Ω to 10 Ω · CMC 0,05 % 10 Ω to 1 MΩ · CMC 0,02 % 1 MΩ to 10 MΩ · CMC 0,05 % 10 MΩ to 100 MΩ · CMC 1 %
DC resistance
Internal procedure; Direct method; At permanent location (S) I_AM_05 based on EURAMET cg-15 v.3.0
11 Clamp meters
Single-phase digital active power meters
Three-phase digital active power meters
Calibration — AC power
f = 50 Hz; U = 10 V to 700 V; I = 0,1 A to 50 A; cos φ = 1 — 10 W to 94,5 kW · CMC 0,18 %
AC power
Internal procedure; At permanent location (S) I_AM_09 based on EURAMET cg-15 v.3.0
12 Single-phase digital apparent power meters
Three-phase digital apparent power meters
Calibration — AC power
f = 50 Hz; U = 10 V to 700 V; I = 0,1 A to 50 A — 10 VA to 94,5 kVA · CMC 0,18 %
AC power
Internal procedure; At permanent location (S) I_AM_09 based on EURAMET cg-15 v.3.0
13 Temperature transmitters (for resistance thermometer sensors) Calibration — Electrical simulation of quantities
-200 °C to 200 °C · CMC 0,17 °C 200 °C to 500 °C · CMC 0,22 °C 500 °C to 850 °C · CMC 0,30 °C
Electrical simulation of quantities
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) I_AM_03 based on EURAMET cg-11 v.2.0
14 Temperature transmitters (for thermocouples) Calibration — Electrical simulation of quantities
-200 °C to 0 °C · CMC 0,40 °C 0 °C to 200 °C · CMC 0,26 °C 200 °C to 1000 °C · CMC 0,35 °C 1000 °C to 1820 °C · CMC 0,44 °C
Electrical simulation of quantities
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) I_AM_03 based on EURAMET cg-11 v.2.0
15 Temperature simulators (for resistance thermometer sensors) Calibration — Electrical simulation of quantities
-200 °C to 200 °C · CMC 0,04 °C 200 °C to 500 °C · CMC 0,08 °C 500 °C to 850 °C · CMC 0,13 °C
Electrical simulation of quantities
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) I_AM_03 based on EURAMET cg-11 v.2.0
16 Temperature simulators (for thermocouples) Calibration — Electrical simulation of quantities
-200 °C to 0 °C · CMC 0,36 °C 0 °C to 200 °C · CMC 0,16 °C 200 °C to 1000 °C · CMC 0,20 °C 1000 °C to 1820 °C · CMC 0,29 °C
Electrical simulation of quantities
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) I_AM_03 based on EURAMET cg-11 v.2.0
17 Temperature indicators (meters), including temperature controllers
Temperature indicators (meters) for resistance thermometer sensors
Calibration — Electrical simulation of quantities
-200 °C to 200 °C · CMC 0,08 °C 200 °C to 500 °C · CMC 0,15 °C 500 °C to 850 °C · CMC 0,20 °C
Electrical simulation of quantities
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) I_AM_03 based on EURAMET cg-11 v.2.0
18 Temperature indicators (meters), including temperature controllers
Temperature indicators (meters) for thermocouples
Calibration — Electrical simulation of quantities
-200 °C to 0 °C · CMC 0,36 °C 0 °C to 200 °C · CMC 0,16 °C 200 °C to 1000 °C · CMC 0,20 °C 1000 °C to 1820 °C · CMC 0,29 °C
Electrical simulation of quantities
Internal procedure; Indirect method; At permanent location (S) and outside of it (P) I_AM_03 based on EURAMET cg-11 v.2.0
19 Multimeters
Frequency meters
Calibration — Frequency
20 Hz to 99,999 Hz · CMC 1 · 10⁻⁵ · f + 0,0014 Hz 100 Hz to 999,99 Hz · CMC 1,1 · 10⁻⁵ · f + 0,013 Hz 1 kHz to 9,9999 kHz · CMC 1,1 · 10⁻⁵ · f + 0,00013 kHz 10 kHz to 99,999 kHz · CMC 1,1 · 10⁻⁵ · f + 0,0013 kHz
Frequency
Internal procedure; Direct method; At permanent location (S) and outside of it (P) I_AM_04
20 Calibrators Calibration — Frequency
20 Hz to 100 kHz · CMC 0,01 %
Frequency
Internal procedure; Direct method; At permanent location (S) I_AM_05
21 Bourdon tube pressure gauges
Electronic pressure gauges
Pressure transmitters
Calibration — Pressure
- gauge pressure (medium – gas) — -95 kPa to -60 kPa · CMC 0,02 % - gauge pressure (medium – gas) — -60 kPa to 60 kPa · CMC 0,012 kPa - gauge pressure (medium – gas) — 60 kPa to 3500 kPa · CMC 0,02 % - gauge pressure (medium – liquid) — 100 kPa to 600 kPa · CMC 0,12 kPa - gauge pressure (medium – liquid) — 0,6 MPa to 70 MPa · CMC 0,02 %
Pressure
Internal procedure; At permanent location (S) I_AM_02 based on EURAMET cg-17 v. 4.1
22 Electronic pressure gauges
Pressure transmitters
Calibration — Pressure
- absolute pressure (medium – gas, liquid) — 7 kPa to 250 kPa · CMC 0,07 kPa - absolute pressure (medium – gas, liquid) — 250 kPa to 700 kPa · CMC 0,2 kPa - absolute pressure (medium – gas, liquid) — 700 kPa to 70100 kPa · CMC 0,02 %
Pressure
Internal procedure; At permanent location (S) I_AM_02 based on EURAMET cg-17 v. 4.1
23 Resistance thermometer sensors Calibration — Temperature (electrical thermometry)
-10 °C to 250 °C · CMC 0,06 °C 250 °C to 400 °C · CMC 0,37 °C 400 °C to 640 °C · CMC 0,83 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) I_AM_06 based on DKD-R 5-1 (09/2018)
24 Noble metal thermocouples (S)
Base metal thermocouples (K, N)
Calibration — Temperature (electrical thermometry)
-10 °C to 160 °C · CMC 0,7 °C 160 °C to 199 °C · CMC 0,9 °C 199 °C to 400 °C · CMC 1,4 °C 400 °C to 600 °C · CMC 1,9 °C 600 °C to 900 °C · CMC 2,2 °C 900 °C to 1100 °C · CMC 2,6 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) I_AM_07 based on EURAMET cg-8 v.3.1
25 Electrical thermometers (including electronic)
Temperature transmitters (containing temperature sensors)
Calibration — Temperature (electrical thermometry)
-10 °C to 250 °C · CMC 0,06 °C 250 °C to 400 °C · CMC 0,37 °C 400 °C to 640 °C · CMC 0,83 °C 640 °C to 1100 °C · CMC 2,3 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) I_AM_08 based on EURAMET cg-8 v.3.1, DKD-R 5-1 (09/2018)

Reference standards (13)

DKD-R 5-1 (09/2018)
EURAMET cg-11 v.2.0
EURAMET cg-15 v.3.0
EURAMET cg-17 v. 4.1
EURAMET cg-8 v.3.1
I_AM_02
I_AM_03
I_AM_04
I_AM_05
I_AM_06
I_AM_07
I_AM_08
I_AM_09

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20131