🇵🇱 Toropol Sp. z o.o. PL

PCA AP 130 • Warszawa, Poland • 16 calibration items

Metrology & calibration

Toropol Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 130 (Warszawa, Poland). Its accreditation scope covers 16 calibration items in the following areas: Metrology & calibration. Calibrated items include: - for compressive forces, - for tensile forces, Bar sieves, Laboratory test sieves. Most frequently used calibration principles: Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; At permanent location (S), Internal procedure; measurement using a coordinate measuring micros….

At a glance

  • Accreditation number: PCA AP 130
  • Calibration items in scope: 16
  • Types of calibrated items: 17
  • Calibration principles used: 4
  • City: Warszawa, Poland
  • Valid until: 21-12-2029
Accreditation number
AP 130
Accreditation body
PCA
Address
ul. Czarna Droga 29, 03-620 Warszawa
City
Warszawa, Poland
Division
Toropol Calibration Laboratory
Valid until
21-12-2029

What this laboratory calibrates (17)

Calibration principles (4)

Certificate information

Issue date
26.11.2025
Issue no.
22
Certificate date
11.12.2020
Accreditation cycle
26.11.2025 — 21.12.2029
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (16)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. A CMC uncertainty value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Laboratory test sieves Calibration — Length
0,020 mm to 0,5 mm · CMC 3,0 µm 0,5 mm to 2 mm · CMC 10 µm 2 mm to 10 mm · CMC 20 µm
Length
Internal procedure; measurement using a coordinate measuring microscope; At permanent location (S) I-07.1, I-07.2, ISO 3310-1, ISO 3310-2
2 Laboratory test sieves Calibration — Length
10 mm to 30 mm · CMC 60 µm 30 mm to 125 mm · CMC 90 µm
Length
Internal procedure; measurement using a calliper; At permanent location (S) I-07.2 based on ISO 3310-2
3 Bar sieves Calibration — Length
2,5 mm to 10 mm · CMC 20 µm
Length
Internal procedure; measurement using a coordinate measuring microscope; At permanent location (S) I-07.4 based on PN-EN 933-3
4 Bar sieves Calibration — Length
10 mm to 50 mm · CMC 60 µm
Length
Internal procedure; measurement using a calliper; At permanent location (S) I-07.4 based on PN-EN 933-3
5 Wire cloth basket sieves Calibration — Length
0,040 mm to 0,5 mm · CMC 3 µm
Length
Internal procedure; measurement using a coordinate measuring microscope; At permanent location (S) I-07.3
6 Analogue dial indicators with a scale interval of 0,01 mm Calibration — Length
0 mm do 30 mm · CMC 4,5 µm 0 mm to 50 mm · CMC 4,8 µm
Length
Internal procedure; At permanent location (S) I-04
7 Digital indicators with a resolution of 0,01 mm Calibration — Length
0 mm do 25,4 mm · CMC 5,1 µm 0 mm to 50,8 mm · CMC 5,4 µm
Length
Internal procedure; At permanent location (S) I-05
8 Digital indicators with a resolution of 0,001 mm Calibration — Length
0 mm to 50,8 mm · CMC 2,5 µm
Length
Internal procedure; At permanent location (S) I-05
9 Digital displacement transducers with a resolution of 0,01 mm Calibration — Length
0 mm to 50 mm · CMC 8,0 µm
Length
Internal procedure; At permanent location (S) and outside of it (P) I-05
10 Static testing machines
- for compressive forces
Calibration — Force
0,1 N to 30 N · CMC 0,15 %1) 0,03 kN to 3000 kN · CMC 0,25 % 3000 kN to 5000 kN · CMC 0,35 %2)
Force
Internal procedure; At permanent location (S) and outside of it (P) I-01 based on PN-EN ISO 7500-1:2018-05, PN-EN 12390-4:2020-03
11 Static testing machines
- for tensile forces
Calibration — Force
0,1 N to 30 N · CMC 0,15 %1) 0,03 kN to 300 kN · CMC 0,25 %
Force
Internal procedure; At permanent location (S) and outside of it (P) I-01 based on PN-EN ISO 7500-1:2018-05, PN-EN 12390-4:2020-03
12 Technological devices
- for compressive forces
Calibration — Force
0,1 N to 30 N · CMC 0,15 %1) 0,03 kN to 3000 kN · CMC 0,25 % 3000 kN to 5000 kN · CMC 0,35 %2)
Force
Internal procedure; At permanent location (S) and outside of it (P) I-01 based on PN-EN ISO 7500-1:2018-05, PN-EN 12390-4:2020-03
13 Technological devices
- for tensile forces
Calibration — Force
0,1 N to 30 N · CMC 0,15 %1) 0,03 kN to 300 kN · CMC 0,25 %
Force
Internal procedure; At permanent location (S) and outside of it (P) I-01 based on PN-EN ISO 7500-1:2018-05, PN-EN 12390-4:2020-03
14 Concrete presses Calibration — Force
0,03 N to 3000 kN · CMC 0,25 % 3000 kN to 5000 kN · CMC 0,35 %2)
Force
Internal procedure; At permanent location (S) and outside of it (P) I-01 based on PN-EN ISO 7500-1:2018-05, PN-EN 12390-4:2020-03
15 Extensometers
(mounted in static testing machines)
Calibration — Force
0 mm to 700 mm · CMC 0,25 %3)
Force
Internal procedure; At permanent location (S) and outside of it (P) I-10 based on PN-EN 9513:2013-06
16 Electronic pressure gauges
Bourdon tube pressure gauges
- gauge pressure – overpressure
Calibration — Pressure
0 MPa to 1,6 MPa · CMC 0,001 MPa 1,6 MPa to 6,0 MPa · CMC 0,004 MPa
Pressure
Internal procedure; At permanent location (S) and outside of it (P) I-12 based on Euramet/cg-17/v.4.0

Reference standards (16)

Euramet/cg-17/v.4.0
I-01
I-04
I-05
I-07.1
I-07.2
I-07.3
I-07.4
I-10
I-12
ISO 3310-1
ISO 3310-2
PN-EN 12390-4:2020-03
PN-EN 933-3
PN-EN 9513:2013-06
PN-EN ISO 7500-1:2018-05

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20130