🇵🇱 APLISENS S.A. PL
Metrology & calibration
APLISENS S.A. is a calibration laboratory accredited by PCA under number AP 122 (Warszawa, Poland). Its accreditation scope covers 3 calibration items in the following areas: Metrology & calibration. Calibrated items include: Pressure transmitters, - absolute pressure (medium – gas), - gauge pressure (medium – gas), Electromagnetic flowmeters. Most frequently used calibration principles: Internal procedure; At permanent location (S).
At a glance
- Accreditation number: PCA AP 122
- Calibration items in scope: 3
- Types of calibrated items: 4
- Calibration principles used: 1
- City: Warszawa, Poland
- Valid until: 23-03-2029
- Accreditation number
- AP 122
- Accreditation body
- PCA
- Address
- ul. Morelowa 7, 03-192 Warszawa
- City
- Warszawa, Poland
- Division
- APLISENS Calibration Laboratory
- Valid until
- 23-03-2029
What this laboratory calibrates (4)
- Pressure transmitters — 2 calibration items
- - absolute pressure (medium – gas) — 1 calibration item
- - gauge pressure (medium – gas) — 1 calibration item
- Electromagnetic flowmeters — 1 calibration item
Calibration principles (1)
- Internal procedure; At permanent location (S) — 3 calibration items
Certificate information
- Issue date
- 16.01.2026
- Issue no.
- 14
- Certificate date
- 10.03.2020
- Accreditation cycle
- 24.02.2025 — 23.03.2029
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (3)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Pressure transmitters - gauge pressure (medium – gas) |
Calibration — Pressure 0 kPa to 4 kPa · CMC 0,0008 kPa 4 kPa to 10 kPa · CMC 0,0022 kPa 10 kPa to 25 kPa · CMC 0,0053 kPa 25 kPa to 40 kPa · CMC 0,0081 kPa 40 kPa to 100 kPa · CMC 0,019 kPa 100 kPa to 160 kPa · CMC 0,025 kPa 160 kPa to 200 kPa · CMC 0,026 kPa 200 kPa to 250 kPa · CMC 0,038 kPa 250 kPa to 600 kPa · CMC 0,086 kPa 600 kPa to 1000 kPa · CMC 0,16 kPa 1000 kPa to 1600 kPa · CMC 0,24 kPa 1600 kPa to 2500 kPa · CMC 0,38 kPa 2500 kPa to 6000 kPa · CMC 1 kPa 6000 kPa to 10000 kPa · CMC 1,5 kPa 10000 kPa to 16000 kPa · CMC 2,5 kPa 16000 kPa to 25000 kPa · CMC 3,5 kPa 25000 kPa to 35000 kPa · CMC 5,3 kPa Pressure |
Internal procedure; At permanent location (S) | PR-01 based on EURAMET cg-17 v.4.1 | |
| 2 | Pressure transmitters - absolute pressure (medium – gas) |
Calibration — Pressure 12 kPa to 40 kPa · CMC 0,035 kPa 40 kPa to 100 kPa · CMC 0,038 kPa 100 kPa to 160 kPa · CMC 0,043 kPa 160 kPa to 200 kPa · CMC 0,046 kPa 200 kPa to 250 kPa · CMC 0,052 kPa 250 kPa to 300 kPa · CMC 0,056 kPa 300 kPa to 600 kPa · CMC 0,1 kPa 600 kPa to 1000) kPa · CMC 0,16 kPa 1000 kPa to 1600 kPa · CMC 0,25 kPa 1600 kPa to 2500 kPa · CMC 0,39 kPa 2500 kPa to 6000 kPa · CMC 1,1 kPa 6000 kPa to 10000 kPa · CMC 1,7 kPa 10000 kPa to 16000 kPa · CMC 2,6 kPa 16000 kPa to 25000 kPa · CMC 3,9 kPa 25000 kPa to 35000 kPa · CMC 5,5 kPa Pressure |
Internal procedure; At permanent location (S) | PR-01 based on EURAMET cg-17 v.4.1 | |
| 3 | Electromagnetic flowmeters | Calibration — Volume flow rate (flow liquids) 0,4 m³/h to 1600 m³/h · CMC 0,2 % Volume flow rate (flow liquids) |
Internal procedure; At permanent location (S) | PR-06 based on PN-EN ISO 4064-2:2017-07 clause 7.4 | |
Reference standards (4)
EURAMET cg-17 v.4.1
PN-EN ISO 4064-2:2017-07 pkt. 7.4
PR-01
PR-06