🇵🇱 APLISENS S.A. PL

PCA AP 122 • Warszawa, Poland • 3 calibration items

Metrology & calibration

APLISENS S.A. is a calibration laboratory accredited by PCA under number AP 122 (Warszawa, Poland). Its accreditation scope covers 3 calibration items in the following areas: Metrology & calibration. Calibrated items include: Pressure transmitters, - absolute pressure (medium – gas), - gauge pressure (medium – gas), Electromagnetic flowmeters. Most frequently used calibration principles: Internal procedure; At permanent location (S).

At a glance

  • Accreditation number: PCA AP 122
  • Calibration items in scope: 3
  • Types of calibrated items: 4
  • Calibration principles used: 1
  • City: Warszawa, Poland
  • Valid until: 23-03-2029
Accreditation number
AP 122
Accreditation body
PCA
Address
ul. Morelowa 7, 03-192 Warszawa
City
Warszawa, Poland
Division
APLISENS Calibration Laboratory
Valid until
23-03-2029

What this laboratory calibrates (4)

Calibration principles (1)

Certificate information

Issue date
16.01.2026
Issue no.
14
Certificate date
10.03.2020
Accreditation cycle
24.02.2025 — 23.03.2029
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (3)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. A value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Pressure transmitters
- gauge pressure (medium – gas)
Calibration — Pressure
0 kPa to 4 kPa · CMC 0,0008 kPa 4 kPa to 10 kPa · CMC 0,0022 kPa 10 kPa to 25 kPa · CMC 0,0053 kPa 25 kPa to 40 kPa · CMC 0,0081 kPa 40 kPa to 100 kPa · CMC 0,019 kPa 100 kPa to 160 kPa · CMC 0,025 kPa 160 kPa to 200 kPa · CMC 0,026 kPa 200 kPa to 250 kPa · CMC 0,038 kPa 250 kPa to 600 kPa · CMC 0,086 kPa 600 kPa to 1000 kPa · CMC 0,16 kPa 1000 kPa to 1600 kPa · CMC 0,24 kPa 1600 kPa to 2500 kPa · CMC 0,38 kPa 2500 kPa to 6000 kPa · CMC 1 kPa 6000 kPa to 10000 kPa · CMC 1,5 kPa 10000 kPa to 16000 kPa · CMC 2,5 kPa 16000 kPa to 25000 kPa · CMC 3,5 kPa 25000 kPa to 35000 kPa · CMC 5,3 kPa
Pressure
Internal procedure; At permanent location (S) PR-01 based on EURAMET cg-17 v.4.1
2 Pressure transmitters
- absolute pressure (medium – gas)
Calibration — Pressure
12 kPa to 40 kPa · CMC 0,035 kPa 40 kPa to 100 kPa · CMC 0,038 kPa 100 kPa to 160 kPa · CMC 0,043 kPa 160 kPa to 200 kPa · CMC 0,046 kPa 200 kPa to 250 kPa · CMC 0,052 kPa 250 kPa to 300 kPa · CMC 0,056 kPa 300 kPa to 600 kPa · CMC 0,1 kPa 600 kPa to 1000) kPa · CMC 0,16 kPa 1000 kPa to 1600 kPa · CMC 0,25 kPa 1600 kPa to 2500 kPa · CMC 0,39 kPa 2500 kPa to 6000 kPa · CMC 1,1 kPa 6000 kPa to 10000 kPa · CMC 1,7 kPa 10000 kPa to 16000 kPa · CMC 2,6 kPa 16000 kPa to 25000 kPa · CMC 3,9 kPa 25000 kPa to 35000 kPa · CMC 5,5 kPa
Pressure
Internal procedure; At permanent location (S) PR-01 based on EURAMET cg-17 v.4.1
3 Electromagnetic flowmeters Calibration — Volume flow rate (flow liquids)
0,4 m³/h to 1600 m³/h · CMC 0,2 %
Volume flow rate (flow liquids)
Internal procedure; At permanent location (S) PR-06 based on PN-EN ISO 4064-2:2017-07 clause 7.4

Reference standards (4)

EURAMET cg-17 v.4.1
PN-EN ISO 4064-2:2017-07 pkt. 7.4
PR-01
PR-06

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20122