🇵🇱 Narodowe Centrum Badań Jądrowych, Ośrodek Radioizotopów POLATOM PL
Metrology & calibration
Narodowe Centrum Badań Jądrowych, Ośrodek Radioizotopów POLATOM is a calibration laboratory accredited by PCA under number AP 120 (Otwock, Poland). Its accreditation scope covers 14 calibration items in the following areas: Metrology & calibration. Calibrated items include: Radioactive solutions, Solid single-radionuclide sources, Surface sources, - surface emission rate. Most frequently used calibration principles: Internal procedure; Measurement method using an HPGe spectrometer;…, Internal procedure; Measurement method using 4π ionisation chambers…, Internal procedure; Measurement method using a 2π proportional coun….
At a glance
- Accreditation number: PCA AP 120
- Calibration items in scope: 14
- Types of calibrated items: 7
- Calibration principles used: 9
- City: Otwock, Poland
- Valid until: 18-12-2028
- Accreditation number
- AP 120
- Accreditation body
- PCA
- Address
- ul. Andrzeja Sołtana 7, 05-400 Otwock
- City
- Otwock, Poland
- Division
- Radioactivity Standards Laboratory
- Valid until
- 18-12-2028
What this laboratory calibrates (7)
- Radioactive solutions — 6 calibration items
- Solid single-radionuclide sources — 4 calibration items
- Surface sources — 2 calibration items
- - surface emission rate — 1 calibration item
- - surface emission rate of radiation — 1 calibration item
- Radioactivity meters with 4π ionisation chambers — 1 calibration item
- Solid multigamma sources — 1 calibration item
Calibration principles (9)
- Internal procedure; Measurement method using an HPGe spectrometer; At permanent location (S) — 3 calibration items
- Internal procedure; Measurement method using 4π ionisation chambers Measurement method using the CAPINTEC CRC-15 BETA meter; At permanent location (S) — 2 calibration items
- Internal procedure; Measurement method using a 2π proportional counter; At permanent location (S) — 2 calibration items
- Internal procedure; Triple-to-double coincidence ratio (TDCR) measurement method 4π(LS)-γ measurement method TDCRγ measurement method Direct methods; At permanent location (S) — 2 calibration items
- Internal procedure; Measurement method using 4π ionisation chambers; At permanent location (S) — 1 calibration item
- Internal procedure; Measurement method using a scintillation counter with a NaI(Tl) crystal; At permanent location (S) — 1 calibration item
- Internal procedure; Measurement method using the TRICARB 2910 TR beta scintillation spectrometer Measurement method using the WALLAC 1411 beta scintillation spectrometer; At permanent location (S) — 1 calibration item
- Internal procedure; Method of calibrating activity meters by measuring a reference material; At permanent location (S) and outside of it (P) — 1 calibration item
- Internal procedure; Triple-to-double coincidence ratio (TDCR) measurement method 4π(LS)-γ measurement method X-γ coincidence measurement method TDCRγ measurement method Direct methods; At permanent location (S) — 1 calibration item
Certificate information
- Issue date
- 15.09.2026
- Issue no.
- 16
- Certificate date
- 31.01.2020
- Accreditation cycle
- 25.11.2024 — 18.12.2028
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (14)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Surface sources - surface emission rate of radiation |
Calibration — Reference materials - measurement for radionuclides: ¹⁴C, ³⁶Cl, ⁵⁷Co, ⁶⁰Co, ⁹⁰Sr,+⁹⁰Y, ¹³⁷Cs, ¹⁴⁷Pm, ²⁰⁴Tl, ²⁴¹Am, ²³⁹Pu — 1 s⁻¹cm⁻² to 90 s⁻¹cm⁻² · CMC 1,5 % Reference materials - measurement |
Internal procedure; Measurement method using a 2π proportional counter; At permanent location (S) | BW-PP-15 | |
| 2 | Surface sources - surface emission rate |
Calibration — Reference materials - measurement for radionuclides: ¹⁴C, ³⁶Cl, ⁵⁷Co, ⁶⁰Co, ⁹⁰Sr,+⁹⁰Y, ¹³⁷Cs, ¹⁴⁷Pm, ²⁰⁴Tl, ²⁴¹Am, ²³⁹Pu — 150 s⁻¹ to 9000 s⁻¹ · CMC 1,5 % Reference materials - measurement |
Internal procedure; Measurement method using a 2π proportional counter; At permanent location (S) | BW-PP-15 | |
| 3 | Radioactive solutions | Calibration — Activity of radionuclides for radionuclides: ¹⁸F, ³²P, ³⁶Cl, ⁴⁵Ca, ⁴⁶Sc, ⁴⁷Sc, ⁵⁷Co, ⁶⁰Co, ⁸⁸Y, ⁸⁹Sr, ¹³¹I, ¹³⁴Cs, ¹⁶¹Tb, ¹⁷⁷Lu, ¹⁹²Ir, ²²³Ra, ²²⁵Ac — 5 Bq to 500 GBq · CMC 0,8 % Activity of radionuclides |
Internal procedure; Triple-to-double coincidence ratio (TDCR) measurement method 4π(LS)-γ measurement method TDCRγ measurement method Direct methods; At permanent location (S) | BW-PP-1, BW-PP-2, BW-PP-11 | |
| 4 | Radioactive solutions | Calibration — Activity of radionuclides for radionuclides: ¹⁴C, ²⁴Na ³⁵S, ⁵⁹Fe, ¹²³I, ¹²⁵I, ¹⁶⁹Yb, ¹⁸⁸Re, ¹⁸⁸W+¹⁸⁸Re, ²⁰³Hg, ⁵⁸Co — 5 Bq to 500 GBq · CMC 1,0 % Activity of radionuclides |
Internal procedure; Triple-to-double coincidence ratio (TDCR) measurement method 4π(LS)-γ measurement method X-γ coincidence measurement method TDCRγ measurement method Direct methods; At permanent location (S) | BW-PP-1, BW-PP-2, BW-PP-3, BW-PP-11 | |
| 5 | Radioactive solutions | Calibration — Activity of radionuclides for radionuclides: ⁶⁵Zn, ⁷⁵Se ⁸⁵Sr, ⁹⁰Sr+⁹⁰Y, ⁹⁰Y, ⁹⁹Mo+⁹⁹ᵐTc, ⁹⁹ᵐTc, ¹⁰⁶Ru+¹⁰⁶Rh, ¹¹⁰ᵐAg, ¹³³Ba, ¹⁹⁸Au, ²⁰⁴Tl, ²⁴¹Am — 5 Bq to 500 GBq · CMC 1,2 % for radionuclides: ⁵¹Cr, ⁶³Ni ⁶⁴Cu, ¹³⁷Cs, ¹⁵²Eu, ²²Na, ⁵⁴Mn ¹²⁴Sb, ¹³⁹Ce, ¹⁷⁰Tm — 5 Bq to 500 GBq · CMC 1,5 % for radionuclides: ³H, ⁵⁵Fe ¹⁰⁹Cd, ¹¹³Sn — 5 Bq to 500 GBq · CMC 2,0 % Activity of radionuclides |
Internal procedure; Triple-to-double coincidence ratio (TDCR) measurement method 4π(LS)-γ measurement method TDCRγ measurement method Direct methods; At permanent location (S) | BW-PP-1, BW-PP-2, BW-PP-11 | |
| 6 | Radioactive solutions | Calibration — Activity of radionuclides for radionuclides: ¹⁸F, ³²P, ³⁶Cl, ⁴⁵Ca, ⁴⁶Sc,⁴⁷Sc, ⁵⁷Co, ⁶⁰Co, ⁸⁸Y, ⁸⁹Sr, ¹³¹I, ¹³⁴Cs, ¹⁶¹Tb, ¹⁷⁷Lu, ¹⁹²Ir — 5 Bq to 500 GBq · CMC 1,0 % for radionuclides: ¹⁴C, ²⁴Na, ³⁵S, ⁵⁹Fe, ¹²³I, ¹²⁵I, ¹⁶⁹Yb, ¹⁸⁸Re, ¹⁸⁸W+¹⁸⁸Re, ²⁰³Hg, ⁵⁸Co — 5 Bq to 500 GBq · CMC 1,2 % for radionuclides: ⁶⁵Zn, ⁷⁵Se ⁸⁵Sr, ⁹⁰Sr+⁹⁰Y, ⁹⁰Y, ⁹⁹Mo+⁹⁹ᵐTc ⁹⁹ᵐTc, ¹⁰⁶Ru+¹⁰⁶Rh, ¹¹⁰ᵐAg ¹³³Ba, ¹⁹⁸Au, ²⁰⁴Tl, ²⁴¹Am — 5 Bq to 500 GBq · CMC 1,4 % for radionuclides: ⁵¹Cr, ⁶³Ni ⁶⁴Cu, ¹³⁷Cs, ¹⁵²Eu, ²²Na, ⁵⁴Mn ¹²⁴Sb, ¹³⁹Ce, ¹⁷⁰Tm — 5 Bq to 500 GBq · CMC 1,6 % for radionuclides: ³H, ⁵⁵Fe ¹⁰⁹Cd, ¹¹³Sn — 5 Bq to 500 GBq · CMC 2,1 % Activity of radionuclides |
Internal procedure; Measurement method using the TRICARB 2910 TR beta scintillation spectrometer Measurement method using the WALLAC 1411 beta scintillation spectrometer; At permanent location (S) | BW-PP-7, BW-PP-8 | |
| 7 | Radioactive solutions | Calibration — Activity of radionuclides for radionuclides: ¹³¹I, ¹³⁴Cs, ⁴⁶Sc, ⁶⁰Co, ¹⁸F, ⁵⁷Co, ⁸⁸Y, ⁸⁹Sr ¹⁷⁷Lu, ¹⁹²Ir, ²⁴Na, ⁵⁹Fe, ¹²³I, ¹²⁵I, ¹⁶⁹Yb, ¹⁸⁸Re, ¹⁸⁸W+¹⁸⁸Re, ²⁰³Hg ⁵⁸Co, ⁶⁵Zn, ⁷⁵Se, ⁸⁵Sr ⁹⁹Mo+⁹⁹ᵐTc, ⁹⁹ᵐTc, ¹⁰⁶Ru+¹⁰⁶Rh ¹¹⁰ᵐAg, ¹³³Ba, ¹⁹⁸Au, ²⁰⁴Tl, ²⁴¹Am ⁵¹Cr, ⁶⁴Cu, ¹³⁷Cs, ¹⁵²Eu, ²²Na ⁵⁴Mn, ¹²⁴Sb, ¹³⁹Ce, ¹⁷⁰Tm, ¹⁰⁹Cd ¹¹³Sn — 5 Bq to 500 GBq · CMC 3,0 % Activity of radionuclides |
Internal procedure; Measurement method using an HPGe spectrometer; At permanent location (S) | BW-PP-4 | |
| 8 | Radioactive solutions | Calibration — Activity of radionuclides for radionuclides: ¹³¹I, ¹³⁴Cs ⁴⁶Sc, ⁶⁰Co, ¹⁸F, ³²P, ⁵⁷Co, ⁸⁸Y ⁸⁹Sr, ¹⁷⁷Lu, ¹⁹²Ir, ²⁴Na, ⁵⁹Fe, ¹²³I, ¹²⁵I, ¹⁶⁹Yb, ¹⁸⁸Re, ¹⁸⁸W+¹⁸⁸Re ²⁰³Hg, ⁵⁸Co, ⁶⁵Zn, ⁷⁵Se, ⁸⁵Sr, ⁹⁰Y ⁹⁹Mo+⁹⁹ᵐTc, ⁹⁹ᵐTc, ¹⁰⁶Ru+¹⁰⁶Rh ¹¹⁰ᵐAg, ¹³³Ba, ¹⁹⁸Au, ²⁰⁴Tl, ²⁴¹Am ⁵¹Cr, ⁶⁴Cu, ¹³⁷Cs, ¹⁵²Eu, ²²Na ⁵⁴Mn, ¹²⁴Sb, ¹³⁹Ce, ¹⁷⁰Tm, ¹⁰⁹Cd ¹¹³Sn — 1 MBq to 50 GBq · CMC 2,5 % Activity of radionuclides |
Internal procedure; Measurement method using 4π ionisation chambers Measurement method using the CAPINTEC CRC-15 BETA meter; At permanent location (S) | BW-PP-5, BW-PP-10 | |
| 9 | Solid single-radionuclide sources | Calibration — Activity of radionuclides for radionuclides: ⁶⁰Co, ⁸⁸Y — 1 kBq to 1 MBq · CMC 1,0 % for radionuclides: ⁵⁷Co, ⁸⁵Sr ¹³³Ba, ²⁴¹Am — 1 kBq to 1 MBq · CMC 1,5 % for radionuclides: ²²Na, ⁵¹Cr ⁵⁴Mn, ⁵⁵Fe, ⁶⁵Zn, ¹⁰⁹Cd, ¹¹³Sn ¹³⁷Cs, ¹⁵²Eu — 1 kBq to 1 MBq · CMC 2,0 % Activity of radionuclides |
Internal procedure; Measurement method using a scintillation counter with a NaI(Tl) crystal; At permanent location (S) | BW-PP-14 | |
| 10 | Solid single-radionuclide sources | Calibration — Activity of radionuclides for radionuclides: ⁶⁰Co, ⁸⁸Y ⁵⁷Co, ⁸⁵Sr, ¹³³Ba, ²⁴¹Am, ²²Na ⁵¹Cr, ⁵⁴Mn, ⁵⁵Fe, ⁶⁵Zn, ¹⁰⁹Cd ¹¹³Sn, ¹³⁷Cs, ¹⁵²Eu, ²⁰³Hg, ¹³⁹Ce ⁴⁰K — 5 Bq to 200 kBq · CMC 3,0 % Activity of radionuclides |
Internal procedure; Measurement method using an HPGe spectrometer; At permanent location (S) | BW-PP-4 | |
| 11 | Solid single-radionuclide sources | Calibration — Activity of radionuclides for radionuclides: ⁵⁷Co, ¹³⁷Cs ¹³³Ba — 1 MBq to 500 MBq · CMC 2,5 % Activity of radionuclides |
Internal procedure; Measurement method using 4π ionisation chambers Measurement method using the CAPINTEC CRC-15 BETA meter; At permanent location (S) | BW-PP-5, BW-PP-10 | |
| 12 | Solid single-radionuclide sources | Calibration — Activity of radionuclides for radionuclides: ¹³¹I — 2 MBq to 10 GBq · CMC 1,3 % Activity of radionuclides |
Internal procedure; Measurement method using 4π ionisation chambers; At permanent location (S) | BW-PP-5 | |
| 13 | Solid multigamma sources | Calibration — Activity of radionuclides for radionuclides: ²⁴¹Am, ¹⁰⁹Cd ⁵⁷Co, ⁵¹Cr, ¹¹³Sn, ⁸⁵Sr, ¹³⁷Cs ⁵⁴Mn, ⁶⁵Zn, ⁶⁰Co, ⁸⁸Y, ¹⁵²Eu ¹³³Ba, ²⁰³Hg, ¹³⁹Ce, ⁴⁰K — 5 Bq to 200 kBq · CMC 3,0 % Activity of radionuclides |
Internal procedure; Measurement method using an HPGe spectrometer; At permanent location (S) | BW-PP-4 | |
| 14 | Radioactivity meters with 4π ionisation chambers | Calibration — Activity of radionuclides 1 MBq to 50 GBq · CMC 2,5 % Activity of radionuclides |
Internal procedure; Method of calibrating activity meters by measuring a reference material; At permanent location (S) and outside of it (P) | BW-PP-9 | |
Reference standards (12)
BW-PP-1
BW-PP-10
BW-PP-11
BW-PP-14
BW-PP-15
BW-PP-2
BW-PP-3
BW-PP-4
BW-PP-5
BW-PP-7
BW-PP-8
BW-PP-9