🇵🇱 Okręgowy Urząd Miar w Gdańsku PL
Metrology & calibration
Okręgowy Urząd Miar w Gdańsku is a calibration laboratory accredited by PCA under number AP 086 (Gdańsk, Poland). Its accreditation scope covers 149 calibration items in the following areas: Metrology & calibration. Calibrated items include: Power network parameter meters, Clamp meters, Multimeters, Calibrators. Most frequently used calibration principles: Internal procedure; Direct method; At permanent location (S), Internal procedure; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P).
At a glance
- Accreditation number: PCA AP 086
- Calibration items in scope: 149
- Types of calibrated items: 123
- Calibration principles used: 42
- City: Gdańsk, Poland
- Valid until: 18-01-2030
- Accreditation number
- AP 086
- Accreditation body
- PCA
- Address
- ul. Polanki 124c, 80-308 Gdańsk
- City
- Gdańsk, Poland
- Division
- Calibration Laboratories Group
- Valid until
- 18-01-2030
What this laboratory calibrates (123)
- Power network parameter meters — 17 calibration items
- Clamp meters — 8 calibration items
- Multimeters — 6 calibration items
- Calibrators — 4 calibration items
- Fixed resistors — 4 calibration items
- Power network analysers — 4 calibration items
- Residual current device testers — 4 calibration items
- Shape testers — 4 calibration items
- Breakdown testers — 3 calibration items
- Mass standards, weights of accuracy class M1 — 3 calibration items
- Mass standards, weights of accuracy class M2 — 3 calibration items
- Oscilloscope calibrators — 3 calibration items
- Oscilloscopes — 3 calibration items
- Resistance standards — 3 calibration items
- – for compressive forces — 2 calibration items
- and 108 more
Calibration principles (42)
- Internal procedure; Direct method; At permanent location (S) — 36 calibration items
- Internal procedure; At permanent location (S) — 33 calibration items
- Internal procedure; At permanent location (S) and outside of it (P) — 14 calibration items
- Internal procedure; Calibration using an optical coordinate measuring machine with an imaging probe; At permanent location (S) — 7 calibration items
- Internal procedure; Calibration using gauge blocks; At permanent location (S) — 5 calibration items
- At permanent location (S) — 3 calibration items
- Internal procedure; Calibration using a dial indicator instrument; At permanent location (S) — 3 calibration items
- Internal procedure; Calibration using a goniometer; At permanent location (S) — 3 calibration items
- Internal procedure; Direct method; At permanent location (S) and outside of it (P) — 3 calibration items
- Internal procedure; Calibration using a dividing head; At permanent location (S) — 2 calibration items
- Internal procedure; Calibration using a micrometer head; At permanent location (S) — 2 calibration items
- Internal procedure; Calibration using gauge blocks; At permanent location (S) and outside of it (P) — 2 calibration items
- Internal procedure; Comparison method Calibration in a climatic chamber; At permanent location (S) — 2 calibration items
- Internal procedure; Direct / indirect method; At permanent location (S) — 2 calibration items
- Internal procedure; Direct method Substitution method for 2,2 A to 20 A; At permanent location (S) — 2 calibration items
- and 27 more
Certificate information
- Issue date
- 24.06.2026
- Issue no.
- 30
- Certificate date
- 26.11.2020
- Accreditation cycle
- 19.12.2025 — 18.01.2030
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (149)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Analogue dial indicators | Calibration — Length scale interval 0,1 mm — 0 mm to 25 mm · CMC 20 µm scale interval 0,01 mm — 0 mm to 25 mm · CMC 5 µm Length |
Internal procedure; Calibration using a micrometer head; At permanent location (S) | PP/6W1/S20/05 | |
| 2 | Analogue dial indicators | Calibration — Length scale interval 0,01 mm — 0 mm to 30 mm · CMC 3 µm Length |
Internal procedure; Calibration using gauge blocks; At permanent location (S) and outside of it (P) | PP/6W1/S20/11 | |
| 3 | Digital indicators | Calibration — Length resolution 0,01 mm — 0 mm to 25 mm · CMC 10 µm Length |
Internal procedure; Calibration using a micrometer head; At permanent location (S) | PP/6W1/S20/05 | |
| 4 | Digital indicators | Calibration — Length resolution 0,0005 mm; resolution 0,001 mm — 0 mm to 30 mm · CMC 2 µm Length |
Internal procedure; Calibration using gauge blocks; At permanent location (S) and outside of it (P) | PP/6W1/S20/11 | |
| 5 | Laser distance meters | Calibration — Length 0 m to 20 m · CMC 2 mm 0 mm do 40 m · CMC 3 mm Length |
Internal procedure; Calibration using a reference laser distance meter; At permanent location (S) | PP/6W1/S16/03 | |
| 6 | Reference foils | Calibration — Length 0 mm to 4 mm · CMC Q[1,4; 1,2 · Lₙ] µm Length |
Internal procedure; Calibration using a dial indicator instrument; At permanent location (S) | PP/6W1/S23/02 | |
| 7 | Micrometer depth gauges | Calibration — Length 0 mm to 25 mm · CMC 2 µm 25 mm to 50 mm · CMC 2 µm 50 mm to 75 mm · CMC 2 µm 75 mm to 100 mm · CMC 2 µm 100 mm to 125 mm · CMC 2 µm 125 mm to 150 mm · CMC 3 µm 150 mm to 175 mm · CMC 3 µm 175 mm to 200 mm · CMC 3 µm 200 mm do 225 mm · CMC 4 µm 225 mm do 250 mm · CMC 4 µm 250 mm do 275 mm · CMC 5 µm 275 mm do 300 mm · CMC 5 µm Length |
Internal procedure; Calibration using gauge blocks; At permanent location (S) | PP/6W1/S20/06 | |
| 8 | Caliper depth gauges | Calibration — Length 0 mm to 300 mm · CMC 0,02 mm Length |
At permanent location (S) | PN-EN ISO 13385-2:2020 | |
| 9 | Micrometer heads | Calibration — Length 0 mm to 30 mm · CMC 3 µm Length |
Internal procedure; Calibration using a dial indicator instrument; At permanent location (S) | PP/6W1/S23/03 | |
| 10 | Dial thickness gauges | Calibration — Length 0 mm to 30 mm · CMC 5 µm Length |
Internal procedure; Calibration using gauge blocks; At permanent location (S) | PP/6W1/S20/10 | |
| 11 | Ultrasonic thickness gauges | Calibration — Length 0 mm to 40 mm · CMC Q[0,032; 0,0058 · lₙ] µm Length |
Internal procedure; At permanent location (S) | PP/6W1/S20/15 | |
| 12 | Penetration needles | Calibration — Length diameter of the cylindrical part — 0,9 mm to 1,1 mm · CMC 2 µm axis deviation of the conical and cylindrical sections — 0 mm to 0,3 mm · CMC 8 µm tip diameter — 0,1 mm to 0,2 mm · CMC 2 µm Length |
Internal procedure; Calibration using an optical coordinate measuring machine with an imaging probe; At permanent location (S) | PP/6W1/S24/01 | |
| 13 | Coating thickness gauges | Calibration — Length 0 mm to 250 µm · CMC 2,2 µm 0 mm to 600 µm · CMC 2,4 µm 0 mm to 1000 µm · CMC 2,7 µm Length |
Internal procedure; Calibration using reference foils; At permanent location (S) | PP/6W1/S20/14 | |
| 14 | External micrometers | Calibration — Length 0 mm to 25 mm · CMC 2 µm 25 mm to 50 mm · CMC 2 µm 50 mm to 75 mm · CMC 2 µm 75 mm to 100 mm · CMC 2 µm 100 mm to 125 mm · CMC 3 µm 125 mm to 150 mm · CMC 3 µm 150 mm to 175 mm · CMC 3 µm 175 mm to 200 mm · CMC 4 µm 200 mm do 225 mm · CMC 4 µm 225 mm do 250 mm · CMC 4 µm 250 mm do 275 mm · CMC 5 µm 275 mm do 300 mm · CMC 5 µm 300 mm do 325 mm · CMC 5 µm 325 mm do 350 mm · CMC 6 µm 350 mm do 375 mm · CMC 6 µm 375 mm do 400 mm · CMC 6 µm 400 mm do 425 mm · CMC 7 µm 425 mm do 450 mm · CMC 7 µm 450 mm do 475 mm · CMC 7 µm 475 mm do 500 mm · CMC 8 µm Length |
Internal procedure; Calibration using gauge blocks; At permanent location (S) | PP/6W1/S20/06 | |
| 15 | Internal micrometers | Calibration — Length 5 mm do 30 mm · CMC 2 µm 30 mm do 55 mm · CMC 2 µm 50 mm to 75 mm · CMC 2 µm 75 mm to 100 mm · CMC 2 µm 100 mm to 125 mm · CMC 3 µm Length |
Internal procedure; Calibration using gauge blocks; At permanent location (S) | PP/6W1/S20/06 | |
| 16 | Toolmaker's measuring microscopes | Calibration — Length 0 mm to 50 mm · CMC 2 µm Length |
Internal procedure; Calibration using a line standard; At permanent location (S) and outside of it (P) | PP/6W1/S15/02 | |
| 17 | Optical parallels | Calibration — Length up to 100 mm · CMC Q[0,59; 0,021 · lₙ] µm Length |
Internal procedure; Measurement of the deviation of the central length from the nominal length Calibration using an optical dial comparator; At permanent location (S) | PP/6W1/S19/01 | |
| 18 | Gauge blocks (classes 0, 1, 2) | Calibration — Length 0,5 mm to 100 mm · CMC Q[71; 1,4 · lₙ] nm Length |
Internal procedure; Calibration using a two-probe comparator; At permanent location (S) | PP/6W1/S21/01 | |
| 19 | Shape testers | Calibration — Length 0 mm to 1000 mm · CMC 4 µm Length |
Internal procedure; Direct length measurement using a caliper, micrometer, rule, dial indicator or coordinate measuring machine with an imaging probe; At permanent location (S) | PP/6W1/S20/17 | |
| 20 | Shape testers | Calibration — Length 0 mm to 100 mm · CMC 2 µm Length |
Internal procedure; Comparative measurement using a dial indicator and gauge blocks; At permanent location (S) | PP/6W1/S23/05 | |
| 21 | Semi-rigid, rigid and tape measures Tape measures |
Calibration — Length semi-rigid, rigid and tape measures — 0 m to 5 m · CMC Q[0,12; 0,015 · L] mm tape measures — 0 m to 20 m · CMC Q[0,12; 0,016 · L] mm tape measures — 0 m to 40 m · CMC Q[0,36; 0,016(L - 20)] mm tape measures — 0 m to 60 m · CMC Q[0,49; 0,016(L - 40)] mm tape measures — 0 m to 80 m · CMC Q[0,60; 0,016(L - 60)] mm tape measures — 0 m to 100 m · CMC Q[0,69; 0,16 (L - 80)] mm Length |
Internal procedure; Calibration using a reference rule; At permanent location (S) | PP/6W1/S16/01 | |
| 22 | Calipers | Calibration — Length external dimensions — 0 mm to 400 mm · CMC 0,02 mm external dimensions — 0 mm to 600 mm · CMC 0,03 mm external dimensions — 0 mm to 1000 mm · CMC 0,04 mm internal dimensions — 0 mm to 380 mm · CMC 0,02 mm Length |
At permanent location (S) | PN-EN ISO 13385-1:2019 | |
| 23 | Feeler gauges | Calibration — Length 0,03 mm to 2 mm · CMC 4 µm Length |
Internal procedure; Calibration using a dial indicator instrument; At permanent location (S) | PP/6W1/S23/01 | |
| 24 | Taper feeler gauges | Calibration — Length 0 mm to 60 mm · CMC 0,2 mm Length |
Internal procedure; Calibration using an optical coordinate measuring machine with an imaging probe; At permanent location (S) | PP/6W1/S24/08 | |
| 25 | Dial bore gauges | Calibration — Length 4 mm to 9,5 mm · CMC 3 µm 10 mm to 18 mm · CMC 3 µm 18 mm to 30 mm · CMC 3 µm 30 mm to 50 mm · CMC 3 µm 50 mm to 150 mm · CMC 4 µm 150 mm to 380 mm · CMC 8 µm Length |
Internal procedure; Calibration using gauge blocks and a micrometer head; At permanent location (S) | PP/6W1/S20/07 | |
| 26 | Two-point internal micrometers | Calibration — Length 50 mm to 75 mm · CMC 2 µm 75 mm to 100 mm · CMC 2 µm 100 mm to 125 mm · CMC 3 µm 125 mm to 150 mm · CMC 3 µm 150 mm to 175 mm · CMC 3 µm 175 mm to 200 mm · CMC 4 µm 200 mm do 225 mm · CMC 4 µm 225 mm do 250 mm · CMC 4 µm 250 mm do 275 mm · CMC 5 µm 275 mm do 300 mm · CMC 5 µm 300 mm do 325 mm · CMC 5 µm 325 mm do 350 mm · CMC 6 µm 350 mm do 375 mm · CMC 6 µm Length |
Internal procedure; Calibration using gauge blocks; At permanent location (S) | PP/6W1/S20/06 | |
| 27 | Caliper height gauges | Calibration — Length 0 mm do 300 mm · CMC 0,02 mm 0 mm do 750 mm · CMC 0,03 mm 0 mm to 1000 mm · CMC 0,04 mm Length |
At permanent location (S) | PN-EN ISO 13225:2012 | |
| 28 | Line standards of measuring magnifiers | Calibration — Length 10 mm · CMC 4,3 µm 20 mm · CMC 4,6 µm 30 mm · CMC 5,1 µm Length |
Internal procedure; Calibration using an optical coordinate measuring machine with an imaging probe; At permanent location (S) | PP/6W1/S24/06 | |
| 29 | Step wedges for ultrasonic thickness gauges | Calibration — Length 0 mm to 40 mm · CMC 0,03 mm Length |
Internal procedure; At permanent location (S) | PP/6W1/S23/04 | |
| 30 | Penetration needles | Calibration — Angle cone angle — 8° to 10° · CMC 5’ squareness deviation of the tip relative to the axis — 0° to 3° · CMC 15’ Angle |
Internal procedure; Calibration using an optical coordinate measuring machine with an imaging probe; At permanent location (S) | PP/6W1/S24/01 | |
| 31 | Universal bevel protractors | Calibration — Angle 4° × 90° · CMC 5’ Angle |
Internal procedure; Calibration using angle gauge blocks; At permanent location (S) | PP/6W1/S20/13 | |
| 32 | 90° try squares | Calibration — Angle length of the longer arm — up to 300 mm · CMC 7,0 µm Angle |
Internal procedure; Calibration using an optical coordinate measuring machine with an imaging probe; At permanent location (S) | PP/6W1/S24/07 | |
| 33 | Toolmaker's measuring microscopes – goniometric head |
Calibration — Angle 0° to 360° · CMC 3’ Angle |
Internal procedure; Calibration using angle gauge blocks; At permanent location (S) and outside of it (P) | PP/6W1/S15/02 | |
| 34 | Johansson angle gauge blocks | Calibration — Angle 0° to 180° · CMC 1,9” Angle |
Internal procedure; Calibration using a goniometer; At permanent location (S) | PP/6W1/S17/01 | |
| 35 | Kusznikow angle gauge blocks | Calibration — Angle 0° to 180° · CMC 1,9” Angle |
Internal procedure; Calibration using a goniometer; At permanent location (S) | PP/6W1/S17/01 | |
| 36 | Wringing angle gauge blocks | Calibration — Angle 0° to 180° · CMC 1,9” Angle |
Internal procedure; Calibration using a goniometer; At permanent location (S) | PP/6W1/S17/01 | |
| 37 | Angle gauge blocks with matt measuring surfaces | Calibration — Angle 0° to 30° · CMC 20’’ Angle |
Internal procedure; Calibration using an optical coordinate measuring machine with an imaging probe; At permanent location (S) | PP/6W1/S24/02 | |
| 38 | Builders' spirit levels | Calibration — Angle 0 m to 2,5 m · CMC 0,1° Angle |
Internal procedure; Calibration using a dividing head; At permanent location (S) | PP/6W1/S20/16 | |
| 39 | Digital levels | Calibration — Angle 4° × 90° · CMC 0,1° Angle |
Internal procedure; Calibration using a dividing head; At permanent location (S) | PP/6W1/S20/16 | |
| 40 | Shape testers | Calibration — Angle 0° to 180° · CMC 1,5’ Angle |
Internal procedure; Angle measurement using a measuring microscope; At permanent location (S) | PP/6W1/S24/05 | |
| 41 | Line standards of measuring magnifiers | Calibration — Angle 0° to 180° · CMC 0,1° Angle |
Internal procedure; Calibration using an optical coordinate measuring machine with an imaging probe; At permanent location (S) | PP/6W1/S24/06 | |
| 42 | Optical flats | Calibration — Length (surface geometry) up to 80 mm · CMC 0,04 µm Length (surface geometry) |
Internal procedure; Calibration using optical flats; At permanent location (S) | PP/6W1/S18/01 | |
| 43 | Optical parallels | Calibration — Length (surface geometry) flatness deviation — up to 100 mm · CMC 0,08 µm parallelism deviation — up to 100 mm · CMC 0,6 µm Length (surface geometry) |
Internal procedure; Calibration using an optical flat and an optical dial instrument; At permanent location (S) | PP/6W1/S19/01 | |
| 44 | Surface plates | Calibration — Length (surface geometry) up to 1600 mm x 1200 mm · CMC 3 µm Length (surface geometry) |
At permanent location (S) and outside of it (P) | PN-ISO 8512-2:1999 | |
| 45 | Optical coordinate measuring machines with an imaging probe | Calibration — Length (coordinate measurements) 1 mm to 300 mm · CMC Q[1,4; 0,0049 · L] µm Length (coordinate measurements) |
Internal procedure; Outside of permanent location (P) | PP/6W1/S15/05 based on PN-EN ISO 10360-7:2011 | |
| 46 | Shape testers | Calibration — Length (coordinate measurements) 0,1 mm do 300 mm · CMC Q[3,5;0,01 · L] µm 0° to 360° · CMC 20” Length (coordinate measurements) |
Internal procedure; Length and angle measurement using an optical coordinate measuring machine with an imaging probe; At permanent location (S) | PP/6W1/S24/10 | |
| 47 | Static testing machines – for tensile forces – for compressive forces Technological devices – for tensile forces – for compressive forces |
Calibration — Force 0,1 N to 10 kN; 0,1 N to 3 MN; 0,1 N to 10 kN; 0,1 N to 2 MN · CMC 0,13 % for proving devices of class 0,5 (tension) 0,1 N to 10 kN; 0,1 N to 3 MN; 0,1 N to 10 kN; 0,1 N to 2 MN · CMC 0,25 % for proving devices of class 1 (tension) 0,1 N to 10 kN; 0,1 N to 3 MN; 0,1 N to 10 kN; 0,1 N to 2 MN · CMC 0,48% for proving devices of class 2 (tension) 0,1 N to 10 kN; 0,1 N to 3 MN; 0,1 N to 10 kN; 0,1 N to 2 MN · CMC 0,13 % for proving devices of class 0,5 (compression) 0,1 N to 10 kN; 0,1 N to 3 MN; 0,1 N to 10 kN; 0,1 N to 2 MN · CMC 0,25 % for proving devices of class 1 (compression) 0,1 N to 10 kN; 0,1 N to 3 MN; 0,1 N to 10 kN; 0,1 N to 2 MN · CMC 0,48 % for proving devices of class 2 (compression) Force |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W1/S25/01 based on PN-EN ISO 7500-1:2018 | |
| 48 | Force gauges | Calibration — Force 0,1 N to 5 kN · CMC 0,03 % compression/tension 5 kN to 600 kN · CMC 0,05 % compression/tension Force |
Internal procedure; At permanent location (S) | PP/6W1/S28/01 | |
| 49 | Brinell hardness testers | Calibration — Hardness hardness — 100 HBW2,5/187,5 to 450 HBW2,5/187,5 · CMC 3,0% hardness — 100 HBW10/1000 to 300 HBW10/1000 · CMC 3,0% hardness — 100 HBW10/3000 to 350 HBW10/3000 · CMC 3,0% force — 1839 N to 29420 N · CMC 0,26 % length — 1 mm to 7 mm · CMC 0,15 % · L (mm), not less than 0,5 µm Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W1/S26/03 based on PN-EN ISO 6506-2 | |
| 50 | Rockwell hardness testers | Calibration — Hardness hardness — 20 HRA to 95 HRA · CMC 0,6 HRA hardness — 10 HRBW to 100 HRBW · CMC 0,6 HRB hardness — 10 HRC to 70 HRC · CMC 0,6 HRC force — 98 N to 1471 N · CMC 0,26 % Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W1/S26/01 based on PN-EN ISO 6508-2 | |
| 51 | Vickers hardness testers | Calibration — Hardness hardness — 100 HV30 to 600 HV30 · CMC 3,0% hardness — 100 HV10 to 800 HV10 · CMC 3,0% hardness — 100 HV5 to 800 HV5 · CMC 3,0% force — 49,03 N to 294,2 N · CMC 0,26 % length — 0 mm to 1 mm · CMC 0,15 % · L (mm), not less than 0,5 µm Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W1/S26/02 based on PN-EN ISO 6507-2 | |
| 52 | Automatic belt weighers | Calibration — Mass (weighing instruments) up to 30 000 kg · CMC 0,15 % Mass (weighing instruments) |
Internal procedure; The measuring range refers to the largest mass of a material portion weighed at once on the non-automatic control instrument; Outside of permanent location (P) | PP/6W1/S04/02 based on OIML R50-1:2014 (p. 7) OIML R50-2:2014 (p. 10) | |
| 53 | Non-automatic electronic weighing instruments | Calibration — Mass (weighing instruments) 0 g to 300 g · CMC 6,8 · 10⁻⁷ · m 300 g to 1200 g · CMC 8,2 · 10⁻⁶ · m 1200 g to 35 kg · CMC 2,0 · 10⁻⁵ · m 35 kg to 600 kg · CMC 1,7 · 10⁻⁴ · m 600 kg to 4000 kg · CMC 4,1 · 10⁻⁴ · m 4000 kg to 120000 kg · CMC 7,5 · 10⁻⁴ · m Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W1/S02/02 based on EURAMET cg-18 v.4.0 | |
| 54 | Mass standards, weights of accuracy class E2 | Calibration — Mass (weights and mass standards) 1 mg · CMC 0,002 mg 2 mg · CMC 0,002 mg 5 mg · CMC 0,002 mg 10 mg · CMC 0,002 mg 20 mg · CMC 0,003 mg 50 mg · CMC 0,004 mg 100 mg · CMC 0,005 mg 200 mg · CMC 0,006 mg 500 mg · CMC 0,008 mg 1g · CMC 0,010 mg 2g · CMC 0,012 mg 5g · CMC 0,015 mg 10 g · CMC 0,020 mg 20 g · CMC 0,025 mg 50 g · CMC 0,03 mg 100 g · CMC 0,05 mg 200 g · CMC 0,10 mg 500 g · CMC 0,25 mg 1 kg · CMC 0,50 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/03 based on OIML R-111-1:2004 Annex C | |
| 55 | Mass standards, weights of accuracy class F1 | Calibration — Mass (weights and mass standards) 1 mg, 2 mg, 5 mg · CMC 0,006 mg 10 mg · CMC 0,008 mg 20 mg · CMC 0,010 mg 50 mg · CMC 0,013 mg 100 mg · CMC 0,016 mg 200 mg · CMC 0,020 mg 500 mg · CMC 0,026 mg 1g · CMC 0,030 mg 2g · CMC 0,040 mg 5g · CMC 0,050 mg 10 g · CMC 0,060 mg 20 g · CMC 0,080 mg 50 g · CMC 0,10 mg 100 g · CMC 0,16 mg 200 g · CMC 0,30 mg 500 g · CMC 0,80 mg 1 kg · CMC 1,6 mg 2 kg · CMC 3 mg 5 kg · CMC 8 mg 10 kg · CMC 16 mg 20 kg · CMC 33 mg 50 kg · CMC 83 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 56 | Mass standards, weights of accuracy class F2 | Calibration — Mass (weights and mass standards) 1 mg, 2 mg, 5 mg · CMC 0,020 mg 10 mg · CMC 0,026 mg 20 mg · CMC 0,033 mg 50 mg · CMC 0,040 mg 100 mg · CMC 0,053 mg 200 mg · CMC 0,060 mg 500 mg · CMC 0,083 mg 1g · CMC 0,10 mg 2g · CMC 0,13 mg 5g · CMC 0,16 mg 10 g · CMC 0,20 mg 20 g · CMC 0,26 mg 50 g · CMC 0,33 mg 100 g · CMC 0,53 mg 200 g · CMC 1,0 mg 500 g · CMC 2,6 mg 1 kg · CMC 5,0 mg 2 kg · CMC 10 mg 5 kg · CMC 26 mg 10 kg · CMC 53 mg 20 kg · CMC 100 mg 50 kg · CMC 260 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 57 | Mass standards, weights of accuracy class M1 | Calibration — Mass (weights and mass standards) 1 mg, 2 mg, 5 mg · CMC 0,06 mg 10 mg · CMC 0,08 mg 20 mg · CMC 0,10 mg 50 mg · CMC 0,13 mg 100 mg · CMC 0,16 mg 200 mg · CMC 0,20 mg 500 mg · CMC 0,26 mg 1g · CMC 0,33 mg 2g · CMC 0,40 mg 5g · CMC 0,53 mg 10 g · CMC 0,66 mg 20 g · CMC 0,83 mg 50 g · CMC 1,0 mg 100 g · CMC 1,6 mg 200 g · CMC 3,3 mg 500 g · CMC 8,3 mg 1 kg · CMC 16 mg 2 kg · CMC 33 mg 5 kg · CMC 83 mg 10 kg · CMC 160 mg 20 kg · CMC 330 mg 50 kg · CMC 830 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 58 | Mass standards, weights of accuracy class M1 | Calibration — Mass (weights and mass standards) 100 kg · CMC 1,6 g 200 kg · CMC 3g 500 kg · CMC 8g 1000 kg · CMC 16 g Mass (weights and mass standards) |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W1/S03/02 based on OIML R-111-1:2004 Annex C | |
| 59 | Mass standards, weights of accuracy class M2 | Calibration — Mass (weights and mass standards) 100 mg · CMC 0,5 mg 200 mg · CMC 0,6 mg 500 mg · CMC 0,8 mg 1g · CMC 1,0 mg 2g · CMC 1,3 mg 5g · CMC 1,6 mg 10 g · CMC 2,0 mg 20 g · CMC 2,6 mg 50 g · CMC 3,3 mg 100 g · CMC 5 mg 200 g · CMC 10 mg 500 g · CMC 26 mg 1 kg · CMC 53 mg 2 kg · CMC 100 mg 5 kg · CMC 260 mg 10 kg · CMC 530 mg 20 kg · CMC 1000 mg 50 kg · CMC 2600 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 60 | Mass standards, weights of accuracy class M2 | Calibration — Mass (weights and mass standards) 100 kg · CMC 5g 200 kg · CMC 10 g 500 kg · CMC 25 g 1000 kg · CMC 50 g Mass (weights and mass standards) |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W1/S03/02 based on OIML R-111-1:2004 Annex C | |
| 61 | 25 kg mass standards | Calibration — Mass (weights and mass standards) 25 kg · CMC 400 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 62 | Load weights | Calibration — Mass (weights and mass standards) 1 g to 51 kg · CMC 1,6 · 10⁻⁵· m Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 63 | pH electrodes | Calibration — pH slope of the characteristic — 50 mV to 70 mV · CMC 0,25 mV electrode efficiency — 80 % to 105 % · CMC 0,40 % pH for E=0 — 5 to 9 · CMC 0,022 pH |
Internal procedure; At permanent location (S) | PP/6W2/S29/01 | |
| 64 | pH meters | Calibration — pH 0 to 14 · CMC 0,003 -1000 mV to 1000 mV · CMC 0,2 mV -1400 mV to - 1100 mV · CMC 0,3 mV 1100 mV to 1400 mV · CMC 0,3 mV pH |
Internal procedure; Electrical method; At permanent location (S) | PP/6W2/S28/01 | |
| 65 | Conductivity meters | Calibration — Specific electrical conductivity (conductometry) 0,1 µS/cm to 19,99 µS/cm · CMC 0,05 % 20 µS/cm to 200 mS/cm · CMC 0,07 % Specific electrical conductivity (conductometry) |
Internal procedure; Electrical method; At permanent location (S) | PP/6W2/S30/01 | |
| 66 | Breath analysers | Calibration — Mass concentration (breath analysis) 0,00 mg/l do 0,40 mg/l · CMC 0,01 mg/l 0,41 mg/l do 1,00 mg/l · CMC 0,02 mg/l 1,01 mg/l to 1,50 mg/l · CMC 0,03 mg/l 1,51 mg/l to 2,00 mg/l · CMC 0,05 mg/l Mass concentration (breath analysis) |
Internal procedure; Mass concentration of ethanol in exhaled air (mg of ethanol in 1 l of air); At permanent location (S) | PP/6W2/S34/01 based on OIML R126-2:2021, A1.1, A2 | |
| 67 | Calibrators | Calibration — DC voltage 0 mV to 200 mV · CMC 1,8 ∙ 10⁻⁵ ∙ U + 0,65 µV 0,2 V to 2 V · CMC 6,6 ∙ 10⁻⁶ ∙ U + 2,1 µV 2 V to 20 V · CMC 6,6 ∙ 10⁻⁶ ∙ U + 14 µV 20 V to 200 V · CMC 9,8 ∙ 10⁻⁶ ∙ U + 0,17 mV 200 V to 1000 V · CMC 1,1 ∙ 10⁻⁵ ∙ U + 2 mV DC voltage |
Internal procedure; Direct / indirect method; At permanent location (S) | PP/6W2/S01/02 | |
| 68 | Oscilloscope calibrators | Calibration — DC voltage 1 mV to 100 V · CMC 0,05 % DC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S14/01 | |
| 69 | Analogue voltmeters | Calibration — DC voltage 1 mV to 210 mV · CMC 2 ∙ 10⁻⁴ ∙ U + 16µV 0,21 V to 2,1 V · CMC 2,1 ∙ 10⁻⁴ ∙ U + 0,15 mV 2,1 V to 21 V · CMC 2,1 ∙ 10⁻⁴ ∙ U + 1,5 mV 21 V to 210 V · CMC 2 ∙ 10⁻⁴ ∙ U + 15 mV 210 V to 1000 V · CMC 2 ∙ 10⁻⁴ ∙ U + 0,12 V DC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S08/01 | |
| 70 | Digital voltmeters Multimeters |
Calibration — DC voltage 0 mV to 22 mV · CMC 5,0 µV 22 V to 220 mV · CMC 1,7 ∙ 10⁻⁵ ∙ U + 1,5 µV 0,22 V to 2,2 V · CMC 1,1 ∙ 10⁻⁵ ∙ U + 4,1 µV 2,2 V to 11 V · CMC 7,5 ∙ 10⁻⁶ ∙ U + 27 µV 11 V to 22 V · CMC 7,5 ∙ 10⁻⁶ ∙ U + 53 µV 22 V to 220 V · CMC 1,1 ∙ 10⁻⁵ ∙ U + 0,38 mV 220 V to 1000 V · CMC 1,5 ∙ 10⁻⁵ ∙ U + 3,8 mV DC voltage |
Internal procedure; Direct / indirect method; At permanent location (S) | PP/6W2/S01/01 based on EURAMET cg-15 v. 3.0 | |
| 71 | Clamp meters Power network parameter meters |
Calibration — DC voltage 10 mV to 320 mV · CMC 1,9 ∙ 10⁻⁵ ∙ U + 58 µV 0,32 V to 3,2 V · CMC 2 ∙ 10⁻⁵ ∙ U + 0,6 mV 3,2 V to 32 V · CMC 2 ∙ 10⁻⁵ ∙ U + 5,8 mV 32 V to 320 V · CMC 2 ∙ 10⁻⁵ ∙ U + 58 mV 320 V to 1000 V · CMC 7,5 ∙ 10⁻⁴ ∙ U - 0,17 V DC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S05/05 | |
| 72 | Power network parameter meters Breakdown testers Safety testers |
Calibration — DC voltage 50 V to 1000 V · CMC 0,58 V DC voltage |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | PP/6W2/S07/01 | |
| 73 | Oscilloscopes | Calibration — DC voltage 1 mV to 100 V · CMC 0,29 % DC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S15/01 based on EURAMET cg-7 v. 1.0 | |
| 74 | Calibrators | Calibration — DC current 0 µA to 200 µA · CMC 2,3 ∙ 10⁻⁵ ∙ I + 0,0013 µA 0,2 mA to 2 mA · CMC 2,2 ∙ 10⁻⁵ ∙ I + 0,013 µA 2 mA to 20 mA · CMC 3,2 ∙ 10⁻⁵ ∙ I + 0,13 µA 20 mA to 200 mA · CMC 1,4 ∙ 10⁻⁴ ∙ I + 2,9 µA 0,2 mA to 2 A · CMC 3,1 ∙ 10⁻⁴ ∙ I + 0,31 mA 2 mA to 20 A · CMC 5,4 ∙ 10⁻⁴ ∙ I + 4,9 mA DC current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S01/02 | |
| 75 | Clamp meters | Calibration — DC current 100 mA to 320 mA · CMC 1 ∙ 10⁻⁴ ∙ I + 0,53 µA 0,32 A to 3,2 A · CMC 8 ∙ 10⁻⁴ ∙ I + 0,4 mA 3,2 A to 10,5 A · CMC 5 ∙ 10⁻⁴ ∙ I + 0,005 A 10,5 A to 20 A · CMC 7 ∙ 10⁻⁴ ∙ I + 0,0069 A DC current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S04/01 | |
| 76 | Power network parameter meters | Calibration — DC current 100 mA to 320 mA · CMC 1 ∙ 10⁻⁴ ∙ I + 0,53 µA 0,32 A to 3,2 A · CMC 8 ∙ 10⁻⁴ ∙ I + 0,4 mA 3,2 A to 10,5 A · CMC 5 ∙ 10⁻⁴ ∙ I + 0,005 A 10,5 A to 20 A · CMC 7 ∙ 10⁻⁴ ∙ I + 0,0069 A DC current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S05/05 | |
| 77 | Analogue ammeters | Calibration — DC current 1 mA to 21 mA · CMC 1∙10⁻⁴ ∙ I + 0,013 mA 21 mA to 210 mA · CMC 1∙10⁻⁴ ∙ I + 0,13 mA 0,21 A to 2,1 A · CMC 1∙10⁻⁴ ∙ I + 0,0013 A 2,1 A to 21 A · CMC 1,8∙10⁻⁴ ∙ I + 0,008 A DC current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S08/01 | |
| 78 | Multimeters Digital ammeters |
Calibration — DC current 0 µA to 200 µA · CMC 9 ∙ 10⁻⁵ ∙ I + 0,016 µA 0,22 mA to 2,2 mA · CMC 7,5 ∙ 10⁻⁵ ∙ I + 0,028 µA 2,2 mA to 22 mA · CMC 7,5 ∙ 10⁻⁵ ∙ I + 0,22 µA 22 mA to 220 mA · CMC 9,5 ∙ 10⁻⁵ ∙ I + 3,1 µA 0,22 A to 2,2 A · CMC 1,7 ∙ 10⁻⁴ ∙ I + 55 µA 2,2 A to 20 A · CMC 5,4 ∙ 10⁻⁴ ∙ I + 4,9 mA DC current |
Internal procedure; Direct method Substitution method for 2,2 A to 20 A; At permanent location (S) | PP/6W2/S01/01 based on EURAMET cg-15 v. 3.0 | |
| 79 | Power network analysers Power network parameter meters Clamp meters Residual current device testers |
Calibration — AC voltage 50 Hz — 10 mV to 32 mV · CMC 4 ∙ 10⁻⁴ ∙ U + 0,13 mV 50 Hz — 32 mV to 320 mV · CMC 4 ∙ 10⁻⁴ ∙ U + 0,056 mV 50 Hz — 0,32 V to 3,2 V · CMC 4 ∙ 10⁻⁴ ∙ U + 0,0006 V 50 Hz — 3,2 V to 32 V · CMC 5 ∙ 10⁻⁴ ∙ U + 0,0022 V 50 Hz — 32 V to 105 V · CMC 8 ∙ 10⁻⁴ ∙ U - 0,0025 V 50 Hz — 105 V to 320 V · CMC 5 ∙ 10⁻⁴ ∙ U + 0,046 V 50 Hz — 320 V to 800 V · CMC 6 ∙ 10⁻⁴ ∙ U + 0,08 V 50 Hz — 800 V to 1000 V · CMC 4 ∙ 10⁻⁴ ∙ U + 0,5 V AC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S05/05 | |
| 80 | Generators | Calibration — AC voltage 20 Hz to 20 kHz — 50 mV to 100 mV · CMC 7 ∙ 10⁻⁴ ∙ U + 0,0462 mV 20 Hz to 20 kHz — 0,1 V to 1 V · CMC 7 ∙ 10⁻⁴ ∙ U + 0,0003 V 20 Hz to 20 kHz — 1 V to 10 V · CMC 7 ∙ 10⁻⁴ ∙ U + 0,0035 V 20 Hz to 20 kHz — 10 V to 30 V · CMC 7 ∙ 10⁻⁴ ∙ U + 0,035 V 20 kHz to 100 kHz — 1 mV to 3 V · CMC 0,5 % 100 kHz to 1 MHz — 1 mV to 3 V · CMC 0,8 % AC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S12/01 | |
| 81 | Calibrators | Calibration — AC voltage 1 Hz to 40 Hz — 0,1 mV to 120 mV · CMC 2,1 ∙ 10⁻⁴ ∙ U + 18 µV 1 Hz to 40 Hz — 0,12 V to 1,2 V · CMC 1,8 ∙ 10⁻⁴ ∙ U + 0,11 mV 1 Hz to 40 Hz — 1,2 V to 12 V · CMC 1,8 ∙ 10⁻⁴ ∙ U + 1,1 mV 1 Hz to 40 Hz — 12 V to 120 V · CMC 2,1 ∙ 10⁻⁴ ∙ U + 11 mV 40 Hz to 2 kHz — 0,1 to 120 mV · CMC 2,1 ∙ 10⁻⁴ ∙ U + 17 µV 40 Hz to 2 kHz — 0,12 V to 1,2 V · CMC 1,8 ∙ 10⁻⁴ ∙ U + 61 µV 40 Hz to 2 kHz — 1,2 V to 12 V · CMC 1,8 ∙ 10⁻⁴ ∙ U + 0,56 mV 40 Hz to 2 kHz — 12 V to 120 V · CMC 2,1 ∙ 10⁻⁴ ∙ U + 6,6 mV 40 Hz to 2 kHz — 120 V to 500 V · CMC 2,6 ∙ 10⁻⁴ ∙ U + 0,11 V 40 Hz to 2 kHz — 500 V to 1000 V · CMC 2,6 ∙ 10⁻⁴ ∙ U + 0,2 V 2 kHz to 10 kHz — 0,1 mV to 120 mV · CMC 3 ∙ 10⁻⁴ ∙ U + 17 µV 2 kHz to 10 kHz — 0,12 V to 1,2 V · CMC 3 ∙ 10⁻⁴ ∙ U + 61 µV 2 kHz to 10 kHz — 1,2 V to 12 V · CMC 3 ∙ 10⁻⁴ ∙ U + 0,56 mV 2 kHz to 10 kHz — 12 V to 120 V · CMC 2,6 ∙ 10⁻⁴ ∙ U + 6,6 mV 2 kHz to 10 kHz — 120 V to 500 V · CMC 2,6 ∙ 10⁻⁴ ∙ U + 0,13 V 2 kHz to 10 kHz — 500 V to 1000 V · CMC 2,6 ∙ 10⁻⁴ ∙ U + 0,2 V 10 kHz to 30 kHz — 0,1 mV to 120 mV · CMC 5,3 ∙ 10⁻⁴ ∙ U + 29 µV 10 kHz to 30 kHz — 0,12 V to 1,2 V · CMC 5,3 ∙ 10⁻⁴ ∙ U + 0,11 mV 10 kHz to 30 kHz — 1,2 V to 12 V · CMC 5,3 ∙ 10⁻⁴ ∙ U + 1,1 mV 10 kHz to 30 kHz — 12 V to 120 V · CMC 5,3 ∙ 10⁻⁴ ∙ U + 11 mV 10 kHz to 30 kHz — 120 V to 500 V · CMC 5,3 ∙ 10⁻⁴ ∙ U + 0,26 V 10 kHz to 30 kHz — 500 V to 1000 V · CMC 5,3 ∙ 10⁻⁴ ∙ U + 0,44 V 30 kHz to 100 kHz — 0,1 mV to 120 mV · CMC 1,2 ∙ 10⁻³ ∙ U + 72 µV 30 kHz to 100 kHz — 0,12 V to 1,2 V · CMC 1,3 ∙ 10⁻³ ∙ U + 0,26 mV 30 kHz to 100 kHz — 1,2 V to 12 V · CMC 1,3 ∙ 10⁻³ ∙ U + 2,4 mV 30 kHz to 100 kHz — 12 V to 120 V · CMC 1,4 ∙ 10⁻³ ∙ U + 28 mV 100 kHz to 300 kHz — 0,1 mV to 120 mV · CMC 4,8 ∙ 10⁻³ ∙ U + 0,84 mV 100 kHz to 300 kHz — 0,12 V to 1,2 V · CMC 4,8 ∙ 10⁻³ ∙ U + 8,2 mV 300 kHz to 1 MHz — 0,1 mV to 120 mV · CMC 2,3 ∙ 10⁻² ∙ U + 2,4 mV 300 kHz to 1 MHz — 0,12 V to 1,2 V · CMC 2,3 ∙ 10⁻² ∙ U + 25 mV AC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S01/02 | |
| 82 | Oscilloscope calibrators | Calibration — AC voltage 1 kHz — 1 mV to 100 V · CMC 0,005 % AC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S14/01 | |
| 83 | Analogue voltmeters | Calibration — AC voltage 50 Hz — 1 mV to 210 mV · CMC 6 ∙ 10⁻⁴ ∙ U + 0,087 mV 50 Hz — 0,21 V to 2,1 V · CMC 6 ∙ 10⁻⁴ ∙ U + 0,0003 V 50 Hz — 2,1 V to 21 V · CMC 5 ∙ 10⁻⁴ ∙ U + 0,0032 V 50 Hz — 21 V to 210 V · CMC 5 ∙ 10⁻⁴ ∙ U + 0,032 V 50 Hz — 210 V to 1000 V · CMC 5 ∙ 10⁻⁴ ∙ U + 0,18 V AC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S08/01 | |
| 84 | Power network parameter meters Breakdown testers Electrical safety testers |
Calibration — AC voltage 50 Hz — 100 V to 750 V · CMC 9 ∙ 10⁻⁴ ∙ U + 0,5 V 50 Hz — 750 V to 1000 V · CMC 5 ∙ 10⁻⁴ ∙ U + 0,01 kV AC voltage |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | PP/6W2/S07/01 | |
| 85 | Power network parameter meters Residual current device testers (touch voltage) |
Calibration — AC voltage 50 Hz — 1 V to 100 V · CMC 0,058 V AC voltage |
Internal procedure; Indirect method; At permanent location (S) | PP/6W2/S05/05 | |
| 86 | Digital voltmeters Multimeters |
Calibration — AC voltage 10 Hz to 20 Hz — 2,2 mV to 22 mV · CMC 5,4 ∙ 10⁻⁴ ∙ U + 13 µV 10 Hz to 20 Hz — 22 mV to 220 mV · CMC 5,4 ∙ 10⁻⁴ ∙ U + 39 µV 10 Hz to 20 Hz — 0,22 V to 2,2 V · CMC 5,4 ∙ 10⁻⁴ ∙ U + 0,12 mV 10 Hz to 20 Hz — 2,2 V to 22 V · CMC 5,4 ∙ 10⁻⁴ ∙ U + 1,2 mV 10 Hz to 20 Hz — 22 V to 220 V · CMC 5,4 ∙ 10⁻⁴ ∙ U + 13 mV 20 Hz to 40 Hz — 2,2 mV to 22 mV · CMC 2,1 ∙ 10⁻⁴ ∙ U + 13 µV 20 Hz to 40 Hz — 22 mV to 220 mV · CMC 2,1 ∙ 10⁻⁴ ∙ U + 27 µV 20 Hz to 40 Hz — 0,22 V to 2,2 V · CMC 2 ∙ 10⁻⁴ ∙ U + 60 µV 20 Hz to 40 Hz — 2,2 V to 22 V · CMC 2 ∙ 10⁻⁴ ∙ U + 0,63 mV 20 Hz to 40 Hz — 22 V to 220 V · CMC 2 ∙ 10⁻⁴ ∙ U + 6,7 mV 40 Hz to 20 kHz — 0,22 mV to 2,2 mV · CMC 1,8 ∙ 10⁻⁴ ∙ U + 12 µV 40 Hz to 20 kHz — 2,2 mV to 22 mV · CMC 1,8 ∙ 10⁻⁴ ∙ U + 13 µV 40 Hz to 20 kHz — 22 mV to 220 mV · CMC 1,3 ∙ 10⁻⁴ ∙ U + 21 µV 40 Hz to 20 kHz — 0,22 V to 2,2 V · CMC 9 ∙ 10⁻⁵ ∙ U + 43 µV 40 Hz to 20 kHz — 2,2 V to 22 V · CMC 9 ∙ 10⁻⁵ ∙ U + 0,4 mV 40 Hz to 20 kHz — 22 V to 220 V · CMC 1,2 ∙ 10⁻⁴ ∙ U + 4,5 mV 20 kHz to 50 kHz — 0,22 mV to 2,2 mV · CMC 3,5 ∙ 10⁻⁴ ∙ U + 12 µV 20 kHz to 50 kHz — 2,2 mV to 22 mV · CMC 4,5 ∙ 10⁻⁴ ∙ U + 13 µV 20 kHz to 50 kHz — 22 mV to 220 mV · CMC 2,7 ∙ 10⁻⁴ ∙ U + 21 µV 20 kHz to 50 kHz — 0,22 V to 2,2 V · CMC 1,5 ∙ 10⁻⁴ ∙ U + 47 µV 20 kHz to 50 kHz — 2,2 V to 22 V · CMC 1,5 ∙ 10⁻⁴ ∙ U + 0,5 mV 20 kHz to 50 kHz — 22 V to 220 V · CMC 1,8 ∙ 10⁻⁴ ∙ U + 5,4 mV 50 kHz to 100 kHz — 2,2 mV to 22 mV · CMC 1,1 ∙ 10⁻³ ∙ U + 20 µV 50 kHz to 100 kHz — 22 mV to 220 mV · CMC 7,1 ∙ 10⁻⁴ ∙ U + 56 µV 50 kHz to 100 kHz — 0,22 V to 2,2 V · CMC 1,9 ∙ 10⁻⁴ ∙ U + 0,15 mV 50 kHz to 100 kHz — 2,2 V to 22 V · CMC 1,9 ∙ 10⁻⁴ ∙ U + 1,3 mV 50 kHz to 100 kHz — 22 V to 220 V · CMC 3,3 ∙ 10⁻⁴ ∙ U + 16 mV 100 kHz to 300 kHz — 2,2 mV to 22 mV · CMC 2,4 ∙ 10⁻³ ∙ U + 31 µV 100 kHz to 300 kHz — 22 mV to 220 mV · CMC 1,5 ∙ 10⁻³ ∙ U + 64 µV 100 kHz to 300 kHz — 0,22 V to 2,2 V · CMC 7,7 ∙ 10⁻⁴ ∙ U + 0,26 mV 100 kHz to 300 kHz — 2,2 V to 22 V · CMC 5,6 ∙ 10⁻⁴ ∙ U + 2,3 mV 100 kHz to 300 kHz — 22 V to 220 V · CMC 2 ∙ 10⁻³ ∙ U + 46 mV 300 kHz to 500 kHz — 2,2 mV to 22 mV · CMC 3,1 ∙ 10⁻³ ∙ U + 66 µV 300 kHz to 500 kHz — 22 mV to 220 mV · CMC 3,1 ∙ 10⁻³ ∙ U + 85 µV 300 kHz to 500 kHz — 0,22 V to 2,2 V · CMC 2,2 ∙ 10⁻³ ∙ U + 0,5 mV 300 kHz to 500 kHz — 2,2 V to 22 V · CMC 2,2 ∙ 10⁻³ ∙ U + 6,5 mV 500 kHz to 1 MHz — 2,2 mV to 22 mV · CMC 6,1 ∙ 10⁻³ ∙ U + 88 µV 500 kHz to 1 MHz — 22 mV to 220 mV · CMC 6 ∙ 10⁻³ ∙ U + 0,22 mV 500 kHz to 1 MHz — 0,22 V to 2,2 V · CMC 3,7 ∙ 10⁻³ ∙ U + 1,3 mV 500 kHz to 1 MHz — 2,2 V to 22 V · CMC 3,3 ∙ 10⁻³ ∙ U + 21 mV 15 Hz to 50 Hz — 220 V to 250 V · CMC 6,6 ∙ 10⁻⁴ ∙ U + 75 mV 50 Hz to 1 kHz — 220 V to 1000 V · CMC 1,6 ∙ 10⁻⁴ ∙ U + 47 mV AC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S01/01 based on EURAMET cg-15 v. 3.0 | |
| 87 | Oscilloscopes | Calibration — AC voltage 1 kHz — 1 mV to 100 V · CMC 0,29 % AC voltage |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S15/01 based on EURAMET cg-7 v. 1.0 | |
| 88 | Calibrators | Calibration — AC current 40 Hz to 2 kHz — 10 mA to 200 µA · CMC 6,5 ∙ 10⁻⁴ ∙ I + 0,032 µA 40 Hz to 2 kHz — 0,2 mA to 2 mA · CMC 6,5 ∙ 10⁻⁴ ∙ I + 0,27 µA 40 Hz to 2 kHz — 2 mA to 20 mA · CMC 6,5 ∙ 10⁻⁴ ∙ I + 2,7 µA 40 Hz to 2 kHz — 20 mA to 200 mA · CMC 6,5 ∙ 10⁻⁴ ∙ I + 26 µA 40 Hz to 2 kHz — 0,2 A to 2 A · CMC 6,9 ∙ 10⁻⁴ ∙ I + 0,51 mA 40 Hz to 2 kHz — 2 A to 20 A · CMC 2 ∙ 10⁻³ ∙ I + 5,4 mA 2 kHz to 5 kHz — 10 µA to 200 µA · CMC 1,3 ∙ 10⁻³ ∙ I + 0,05 µA 2 kHz to 5 kHz — 0,2 mA to 2 mA · CMC 1,3 ∙ 10⁻³ ∙ I + 0,42 µA 2 kHz to 5 kHz — 2 mA to 20 mA · CMC 1,3 ∙ 10⁻³ ∙ I + 3,6 µA 2 kHz to 5 kHz — 20 mA to 200 mA · CMC 1,2 ∙ 10⁻³ ∙ I + 34 µA 2 kHz to 5 kHz — 0,2 A to 2 A · CMC 1,3 ∙ 10⁻³ ∙ I + 0,71 mA 2 kHz to 5 kHz — 2 A to 20 A · CMC 2 ∙ 10⁻³ ∙ I + 11 mA AC current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S01/02 | |
| 89 | Analogue ammeters | Calibration — AC current 50 Hz — 1 mA to 21 mA · CMC 3 ∙ 10⁻⁴ ∙ I + 0,016 mA 50 Hz — 21 mA to 210 mA · CMC 3 ∙ 10⁻⁴ ∙ I + 0,15 mA 50 Hz — 0,21 A to 2,1 A · CMC 2 ∙ 10⁻⁴ ∙ I + 0,0018 A 50 Hz — 2,1 A to 21 A · CMC 2 ∙ 10⁻⁴ ∙ I + 0,018 A AC current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S08/01 | |
| 90 | Digital ammeters Multimeters |
Calibration — AC current 40 Hz to 1 kHz — 10 µA to 220 µA · CMC 2,3 ∙ 10⁻⁴ ∙ I + 0,028 µA 40 Hz to 1 kHz — 0,22 mA to 2,2 mA · CMC 2,3 ∙ 10⁻⁴ ∙ I + 0,16 µA 40 Hz to 1 kHz — 2,2 mA to 22 mA · CMC 2,3 ∙ 10⁻⁴ ∙ I + 1,6 µA 40 Hz to 1 kHz — 22 mA to 220 mA · CMC 2,3 ∙ 10⁻⁴ ∙ I + 14 µA 40 Hz to 1 kHz — 0,22 A to 2,2 A · CMC 5,5 ∙ 10⁻⁴ ∙ I + 0,16 mA 40 Hz to 1 kHz — 2,2 A to 20 A · CMC 2 ∙ 10⁻³ ∙ I + 5,4 mA 1 kHz to 5 kHz — 10 µA to 220 µA · CMC 6,3 ∙ 10⁻⁴ ∙ I + 0,037 µA 1 kHz to 5 kHz — 0,22 mA to 2,2 mA · CMC 4,4 ∙ 10⁻⁴ ∙ I + 0,33 µA 1 kHz to 5 kHz — 2,2 mA to 22 mA · CMC 4,4 ∙ 10⁻⁴ ∙ I + 2,1 µA 1 kHz to 5 kHz — 22 mA to 220 mA · CMC 4,4 ∙ 10⁻⁴ ∙ I + 16 µA 1 kHz to 5 kHz — 0,22 A to 2,2 A · CMC 9,5 ∙ 10⁻⁴ ∙ I + 0,27 mA AC current |
Internal procedure; Direct method Substitution method for 2,2 A to 20 A; At permanent location (S) | PP/6W2/S01/01 based on EURAMET cg-15 v. 3.0 | |
| 91 | Power network parameter meters Power network analysers |
Calibration — AC current 50 Hz — 100 mA to 320 mA · CMC 1,7 ∙ 10⁻³ ∙ I - 0,019 mA 50 Hz — 0,32 A to 3,2 A · CMC 1,8 ∙ 10⁻³ ∙ I + 0,0006 A 50 Hz — 3,2 A to 10,5 A · CMC 2,3 ∙ 10⁻³ ∙ I + 0,0057 A 50 Hz — 10,5 A to 20 A · CMC 2,4 ∙ 10⁻³ ∙ I + 0,0087 A AC current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S05/05 | |
| 92 | Clamp meters | Calibration — AC current 50 Hz — 100 mA to 320 mA · CMC 1,7 ∙ 10⁻³ ∙ I - 0,019 mA 50 Hz — 0,32 A to 3,2 A · CMC 1,8 ∙ 10⁻³ ∙ I + 0,0006 A 50 Hz — 3,2 A to 10,5 A · CMC 2,3 ∙ 10⁻³ ∙ I + 0,0057 A 50 Hz — 10,5 A to 20 A · CMC 2,4 ∙ 10⁻³ ∙ I + 0,0087 A AC current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S04/01 | |
| 93 | Power network parameter meters Residual current device testers |
Calibration — AC current 50 Hz; RCD tripping current; RCD rated residual current — 5 mA to 1100 mA · CMC 1,39 ∙ 10⁻² ∙ I + 0,07 mA AC current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S05/05 | |
| 94 | Resistance calibrators | Calibration — DC resistance 0 Ω to 0,2 Ω · CMC 18 µΩ 0,2 Ω to 2 Ω · CMC 3,8 ∙ 10⁻⁵ ∙ R + 14 µΩ 2 Ω to 20 Ω · CMC 2,4 ∙ 10⁻⁵ ∙ R + 0,44 mΩ 20 Ω to 200 Ω · CMC 2,1 ∙ 10⁻⁵ ∙ R + 2 mΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S01/02 | |
| 95 | Multimeters | Calibration — DC resistance 0,2 kΩ to 2 kΩ · CMC 2,1 ∙ 10⁻⁵ ∙ R + 17 mΩ 2 kΩ to 20 kΩ · CMC 2,1 ∙ 10⁻⁵ ∙ R + 33 mΩ 20 kΩ to 200 kΩ · CMC 2,2 ∙ 10⁻⁵ ∙ R + 2 Ω 0,2 MΩ to 2 MΩ · CMC 2,5 ∙ 10⁻⁵ ∙ R + 34 Ω 2 MΩ to 20 MΩ · CMC 3,9 ∙ 10⁻⁵ ∙ R + 0,1 kΩ 20 MΩ to 200 MΩ · CMC 1,6 ∙ 10⁻⁴ ∙ R + 6,3 kΩ 0,2 GΩ to 2 GΩ · CMC 5,3 ∙ 10⁻⁴ ∙ R + 0,33 MΩ DC resistance |
Internal procedure; Substitution method; At permanent location (S) | PP/6W2/S01/01 based on EURAMET cg-15 v. 3.0 02/2015 | |
| 96 | Variable resistors Fixed resistors |
Calibration — DC resistance 2 GΩ to 20 GΩ · CMC 3,1 ∙ 10⁻³ ∙ R + 23 MΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S10/04 PP/6W2/S10/05 | |
| 97 | Power network parameter meters Clamp meters |
Calibration — DC resistance 0,01 Ω to 0,1 Ω · CMC 7,8 ∙ 10⁻³ ∙ R + 0,0005 Ω 0,1 Ω to 1 Ω · CMC 2,6 ∙ 10⁻³ ∙ R + 0,0055 Ω 1 Ω to 10 Ω · CMC 8 ∙ 10⁻⁴ ∙ R + 0,0051 Ω 10 Ω to 100 Ω · CMC 6 ∙ 10⁻⁴ ∙ R + 0,0028 Ω 100 Ω to 1000 Ω · CMC 6 ∙ 10⁻⁴ ∙ R + 0,026 Ω 1 kΩ to 10 kΩ · CMC 6 ∙ 10⁻⁴ ∙ R + 0,0003 kΩ 10 kΩ to 100 kΩ · CMC 6 ∙ 10⁻⁴ ∙ R + 0,0026 kΩ 100 kΩ to 1000 kΩ · CMC 6 ∙ 10⁻⁴ ∙ R + 0,027 kΩ 1 MΩ to 10 MΩ · CMC 7 ∙ 10⁻⁴∙ R + 0,0002 MΩ 10 MΩ to 60 MΩ · CMC 2 ∙ 10⁻³ ∙ R + 0,0022 MΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S05/05 | |
| 98 | Power network parameter meters - insulation resistance |
Calibration — DC resistance 10 kΩ to 100 kΩ · CMC 6 ∙ 10⁻⁴ ∙ R + 0,00008 kΩ 100 kΩ to 1000 kΩ · CMC 6 ∙ 10⁻⁴ ∙ R + 0,0012 kΩ 1 MΩ to 10 MΩ · CMC 7 ∙ 10⁻⁴ ∙ R + 0,0002 MΩ 10 MΩ to 100 MΩ · CMC 1,8 ∙ 10⁻³ ∙ R + 0,002 MΩ 100 MΩ to 1000 MΩ · CMC 1,1 ∙ 10⁻³ ∙ R + 0,07 MΩ 1 GΩ to 10 GΩ · CMC 1,73 ∙ 10⁻² ∙ R + 0,000005 GΩ 10 GΩ to 100 GΩ · CMC 1,73 ∙ 10⁻² ∙ R - 0,00009 GΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S05/05 | |
| 99 | Digital ohmmeters | Calibration — DC resistance 0,0001 Ω · CMC 0,06 % 0,001 Ω · CMC 0,4 % 0,002 Ω; 0,003 Ω · CMC 0,25 % 0,01 Ω · CMC 0,07 % 0,02 Ω; 0,03 Ω · CMC 0,04 % 0,1 Ω; 0,2 Ω; 0,3 Ω · CMC 0,03 % 1 Ω; 10 Ω; 100 Ω; 1000 Ω; 10 kΩ; 100 kΩ · CMC 0,01 % 1 MΩ · CMC 0,006 % 10 MΩ · CMC 0,001 % 100 MΩ · CMC 0,004 % 1 GΩ · CMC 0,04 % 10 GΩ · CMC 0,25 % 1 Ω to 10 Ω · CMC 0,12 % 10 Ω to 10 M Ω · CMC 0,06 % 10 MΩ to 100 MΩ · CMC 0,12 % 100 MΩ to 1000 MΩ · CMC 0,20 % DC resistance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S10/03 | |
| 100 | Fixed resistors Resistance standards |
Calibration — DC resistance 0,0001 Ω; 0,001 Ω; 0,002 Ω; 0,003 Ω; 0,01 Ω; 0,02 Ω; 0,03 Ω; 0,1 Ω; 0,2 Ω; 0,3 Ω · CMC 0,0000041 Ω 1Ω · CMC 0,0000047 Ω 10 Ω · CMC 0,000024 Ω 25 Ω · CMC 0,000074 Ω 100 Ω · CMC 0,00035 Ω 1 kΩ · CMC 0,0000047 kΩ 10 kΩ · CMC 0,000024 kΩ 100 kΩ · CMC 0,00023 kΩ 1 MΩ · CMC 0,000012 MΩ 10 MΩ · CMC 0,000070 MΩ 100 MΩ · CMC 0,0017 MΩ 1 GΩ · CMC 0,0023 GΩ DC resistance |
Internal procedure; Direct comparison method; At permanent location (S) | PP/6W2/S10/05 | |
| 101 | Fixed resistors Resistance standards |
Calibration — DC resistance 10 GΩ · CMC 0,025 GΩ 20 GΩ · CMC 0,05 GΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S10/05 | |
| 102 | Power network parameter meters Loop impedance meters Residual current device testers - loop resistance |
Calibration — AC resistance 50 Hz — 0,5 Ω to 1000 Ω · CMC 5,8 ∙ 10⁻³ ∙ R + 0,026 Ω AC resistance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S05/05 | |
| 103 | Power network parameter meters - earth resistance Earth resistance meters |
Calibration — AC resistance 10 Hz to 1 kHz — 0,1 Ω to 1 Ω · CMC 4,1 ∙ 10⁻³ ∙ R + 0,0006 Ω 10 Hz to 1 kHz — 1 Ω to 10 Ω · CMC 1,2 ∙ 10⁻³ ∙ R + 0,0004 Ω 10 Hz to 1 kHz — 10 Ω to 100 Ω · CMC 0,08 % 10 Hz to 1 kHz — 100 Ω to 1000 Ω · CMC 0,07 % 10 Hz to 1 kHz — 1000 Ω to 10000 Ω · CMC 3 ∙ 10⁻⁴ ∙ R + 0,37 Ω 10 Hz to 1 kHz — 1 kΩ to 10 kΩ · CMC 9 ∙ 10⁻⁵ ∙ R + 0,0057 kΩ AC resistance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S05/05 | |
| 104 | Digital ohmmeters Bridges |
Calibration — AC resistance 1 kHz — 0,1 Ω to 1 Ω · CMC 4,1 ∙ 10⁻³ ∙ R + 0,0006 Ω 1 kHz — 1 Ω to 10 Ω · CMC 1,1 ∙ 10⁻³ ∙ R + 0,0004 Ω 1 kHz — 10 Ω to 100 Ω · CMC 5 ∙ 10⁻⁴ ∙ R + 0,003 Ω 1 kHz — 100 Ω to 1000 Ω · CMC 3 ∙ 10⁻⁴ ∙ R + 0,04 Ω 1 kHz — 1 kΩ to 10 kΩ · CMC 3 ∙ 10⁻⁴ ∙ R + 0,0004 kΩ 1 kHz — 10 kΩ to 100 kΩ · CMC 3 ∙ 10⁻⁴ ∙ R + 0,004 kΩ 1 kHz — 100 kΩ to 1000 kΩ · CMC 7 ∙ 10⁻⁴ ∙ R - 0,004 kΩ AC resistance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S10/02 | |
| 105 | Fixed resistors Variable resistors Resistance standards |
Calibration — AC resistance 1kHz — 0,1 Ω to 1 Ω · CMC 7 ∙ 10⁻⁴ ∙ R + 0,0006 Ω 1kHz — 1Ω · CMC 0,15 % 1kHz — 1 Ω to 10 Ω · CMC 1,6 ∙ 10⁻³ ∙ R - 0,004 Ω 1kHz — 10 Ω · CMC 0,18 % 1kHz — 10 Ω to 100 kΩ · CMC 0,03 % 1kHz — 100 kΩ · CMC 0,01 % 1kHz — 100 kΩ to 1000 kΩ · CMC 1,9 ∙ 10⁻⁴ ∙ R - 0,013 kΩ AC resistance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S10/01 | |
| 106 | Variable standard capacitors Fixed standard capacitors |
Calibration — Capacitance 1 kHz — 1 pF to 10 pF · CMC 0,1 % 1 kHz — 10 pF to 100 pF · CMC 0,3 % Capacitance |
Internal procedure; Substitution method; At permanent location (S) | PP/6W2/S10/01 | |
| 107 | Variable standard capacitors Fixed standard capacitors |
Calibration — Capacitance 1 kHz — 100 pF to 1000 pF · CMC 0,05 % 1 kHz — 1 nF to 10 nF · CMC 0,04 % 1 kHz — 10 nF to 100 nF · CMC 0,04 % 1 kHz — 100 nF to 1000 nF · CMC 0,04 % 1 kHz — 1 µF to 10 µF · CMC 0,04 % Capacitance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S10/01 | |
| 108 | Capacitance meters Capacitance bridges |
Calibration — Capacitance 1 kHz — 1 pF to 10 pF · CMC 0,3 % 1 kHz — 10 pF to 100 pF · CMC 0,1 % 1 kHz — 100 pF to 1000 pF · CMC 1,8 ∙ 10⁻⁴ ∙ C + 0,6 1 kHz — 1 nF to 10 nF · CMC 2,6 ∙ 10⁻⁴ ∙ C + 0,0006 1 kHz — 10 nF to 100 nF · CMC 0,4 % 1 kHz — 100 nF to 1000 nF · CMC 0,4 % 1 kHz — 1 µF to 10 µF · CMC 1,2 % Capacitance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S10/02 | |
| 109 | Multimeters | Calibration — Capacitance 0,1 nF to 11 µF · CMC 1,20 % Capacitance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S01/01 | |
| 110 | Inductors Variable standard inductors Fixed standard inductors |
Calibration — Inductance 1 kHz — 1 µH · CMC 6% 1 kHz — 2 µH; 3 µH; 5 µH · CMC 3% 1 kHz — 10 µH · CMC 1% 1 kHz — 20 µH; 30 µH; 50 µH · CMC 0,6 % 1 kHz — 100 µH; 200 µH; 300 µH · CMC 0,2 % 1 kHz — 500 µH; 1 mH; 2 mH; 3 mH; 5mH · CMC 0,1 % 1 kHz — 10 mH; 20 mH; 30 mH; 50 mH; 100 mH; 200 mH; 300 mH; 500 mH; 1H · CMC 0,05 % 1 kHz — 100 µH to 1000 µH · CMC 0,2 % 1 kHz — 1 mH to 10 mH · CMC 0,1 % 1 kHz — 10 mH to 100 mH · CMC 0,05 % 1 kHz — 100 mH to 1000 mH · CMC 0,05 % 1 kHz — 1 H to 10 H · CMC 0,1 % Inductance |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S10/01 | |
| 111 | Inductance meters Inductance bridges |
Calibration — Inductance 1 kHz — 1 µH · CMC 6% 1 kHz — 2 µH; 3 µH; 5 µH · CMC 3% 1 kHz — 10 µH · CMC 1% 1 kHz — 20 µH; 30 µH; 50 µH · CMC 0,6 % 1 kHz — 100 µH; 200 µH; 300 µH · CMC 0,2 % 1 kHz — 500 µH; 1 mH; 2 mH; 3 mH; 5mH · CMC 0,1 % 1 kHz — 10 mH; 20 mH; 30 mH; 50 mH; 100 mH; 200 mH; 300 mH; 500 mH; 1H · CMC 0,05 % Inductance |
Internal procedure; Direct method using fixed inductance standards; At permanent location (S) | PP/6W2/S10/02 | |
| 112 | Inductance meters Inductance bridges |
Calibration — Inductance 1 kHz — 100 µH to 1000 µH · CMC 0,5 % 1 kHz — 1 mH to 10 mH · CMC 0,5 % 1 kHz — 10 mH to 100 mH · CMC 0,3 % 1 kHz — 100 mH to 1000 mH · CMC 0,3 % 1 kHz — 1 H to 10 H · CMC 0,3 % Inductance |
Internal procedure; Direct method using variable inductance standards; At permanent location (S) | PP/6W2/S10/02 | |
| 113 | Clamp meters Power network parameter meters - DC current |
Calibration — High voltage and current 16 A to 32 A · CMC 2,4 ∙ 10⁻³ ∙ I + 0,0011 A 32 A to 105 A · CMC 2,5 ∙ 10⁻³ ∙ I + 0,0012 A 105 A to 200 A · CMC 2,3 ∙ 10⁻³ ∙ I + 0,030 A 200 A to 525 A · CMC 2,7 ∙ 10⁻³ ∙ I - 0,030 A 525 A to 1000 A · CMC 2,3 ∙ 10⁻³ ∙ I + 0,28 A High voltage and current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S04/01 | |
| 114 | Power network analysers Clamp meters Power network parameter meters - AC current |
Calibration — High voltage and current 50 Hz — 16 A to 32 A · CMC 3,3 ∙ 10⁻³ ∙ I + 0,0052 A 50 Hz — 32 A to 160 A · CMC 3,3 ∙ 10⁻³ ∙ I + 0,027 A 50 Hz — 160 A to 1000 A · CMC 3,3 ∙ 10⁻³ ∙ I + 0,45 A High voltage and current |
Internal procedure; Direct method; At permanent location (S) | PP/6W2/S04/01 | |
| 115 | Power network parameter meters Breakdown testers Electrical safety testers - DC/AC voltage |
Calibration — High voltage and current 50 Hz — 1 kV to 7 kV · CMC 1,6 ∙ 10⁻³ ∙ U + 0,011 kV 50 Hz — 7 kV to 15 kV · CMC 0,21 kV 50 Hz — 15 kV to 30 kV · CMC 0,45 kV High voltage and current |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | PP/6W2/S07/01 | |
| 116 | Temperature simulators Temperature indicators (meters) |
Calibration — Electrical simulation of quantities -200 °C to 1800 °C · CMC 0,014 °C; Matryca CMC na str. 23 - 25 CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 0 °C · CMC 0,12 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 100 °C · CMC 0,091 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 200 °C · CMC 0,079 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 300 °C · CMC 0,072 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 400 °C · CMC 0,069 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 500 °C · CMC 0,068 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 600 °C · CMC 0,067 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 700 °C · CMC 0,066 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 800 °C · CMC 0,065 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 900 °C · CMC 0,064 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 1000 °C · CMC 0,064 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 1100 °C · CMC 0,064 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 1200 °C · CMC 0,064 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 1300 °C · CMC 0,065 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 1400 °C · CMC 0,067 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 1500 °C · CMC 0,069 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 1600 °C · CMC 0,072 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type R — 1700 °C · CMC 0,076 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 0 °C · CMC 0,12 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 100 °C · CMC 0,093 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 200 °C · CMC 0,083 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 300 °C · CMC 0,077 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 400 °C · CMC 0,075 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 500 °C · CMC 0,074 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 600 °C · CMC 0,074 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 700 °C · CMC 0,073 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 800 °C · CMC 0,073 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 900 °C · CMC 0,072 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 1000 °C · CMC 0,072 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 1100 °C · CMC 0,072 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 1200 °C · CMC 0,072 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 1300 °C · CMC 0,073 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 1400 °C · CMC 0,075 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 1500 °C · CMC 0,078 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 1600 °C · CMC 0,081 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type S — 1700 °C · CMC 0,085 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 600 °C · CMC 0,12 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 700 °C · CMC 0,10 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 800 °C · CMC 0,094 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 900 °C · CMC 0,087 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 1000 °C · CMC 0,082 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 1100 °C · CMC 0,078 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 1200 °C · CMC 0,076 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 1300 °C · CMC 0,074 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 1400 °C · CMC 0,073 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 1500 °C · CMC 0,073 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 1600 °C · CMC 0,074 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 1700 °C · CMC 0,076 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type B — 1800 °C · CMC 0,079 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — -200 °C · CMC 0,064 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — -100 °C · CMC 0,056 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 0 °C · CMC 0,050 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 100 °C · CMC 0,050 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 200 °C · CMC 0,051 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 300 °C · CMC 0,051 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 400 °C · CMC 0,052 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 500 °C · CMC 0,052 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 600 °C · CMC 0,053 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 700 °C · CMC 0,053 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 800 °C · CMC 0,048 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 900 °C · CMC 0,049 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 1000 °C · CMC 0,051 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 1100 °C · CMC 0,052 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type J — 1200 °C · CMC 0,054 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type T — -200 °C · CMC 0,062 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type T — -100 °C · CMC 0,047 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type T — 0 °C · CMC 0,038 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type T — 100 °C · CMC 0,037 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type T — 200 °C · CMC 0,037 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type T — 300 °C · CMC 0,037 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type T — 400 °C · CMC 0,038 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — -200 °C · CMC 0,040 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — -100 °C · CMC 0,029 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 0 °C · CMC 0,023 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 100 °C · CMC 0,023 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 200 °C · CMC 0,023 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 300 °C · CMC 0,024 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 400 °C · CMC 0,025 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 500 °C · CMC 0,026 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 600 °C · CMC 0,027 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 700 °C · CMC 0,028 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 800 °C · CMC 0,030 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 900 °C · CMC 0,032 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type E — 1000 °C · CMC 0,033 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — -200 °C · CMC 0,065 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — -100 °C · CMC 0,049 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 0 °C · CMC 0,057 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 100 °C · CMC 0,057 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 200 °C · CMC 0,058 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 300 °C · CMC 0,058 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 400 °C · CMC 0,059 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 500 °C · CMC 0,071 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 600 °C · CMC 0,072 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 700 °C · CMC 0,073 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 800 °C · CMC 0,074 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 900 °C · CMC 0,075 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 1000 °C · CMC 0,076 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 1100 °C · CMC 0,077 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 1200 °C · CMC 0,079 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type K — 1300 °C · CMC 0,081 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — -200 °C · CMC 0,078 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — -100 °C · CMC 0,044 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 0 °C · CMC 0,039 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 100 °C · CMC 0,038 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 200 °C · CMC 0,038 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 300 °C · CMC 0,038 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 400 °C · CMC 0,038 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 500 °C · CMC 0,039 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 600 °C · CMC 0,050 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 700 °C · CMC 0,051 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 800 °C · CMC 0,052 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 900 °C · CMC 0,053 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 1000 °C · CMC 0,054 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 1100 °C · CMC 0,056 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 1200 °C · CMC 0,057 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type N — 1300 °C · CMC 0,059 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 0 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 100 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 200 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 300 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 400 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 500 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 600 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 700 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 800 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 900 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 1000 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 1100 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 1200 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 1300 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 1400 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 1500 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 1600 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 1700 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 1800 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 1900 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 2000 °C · CMC 0,67 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 2100 °C · CMC 0,68 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 2200 °C · CMC 0,68 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type C — 2300 °C · CMC 0,68 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 0 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 100 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 200 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 300 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 400 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 500 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 600 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 700 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 800 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 900 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 1000 °C · CMC 0,40 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 1100 °C · CMC 0,41 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 1200 °C · CMC 0,41 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 1300 °C · CMC 0,41 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 1400 °C · CMC 0,41 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 1500 °C · CMC 0,41 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 1600 °C · CMC 0,41 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 1700 °C · CMC 0,41 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 1800 °C · CMC 0,41 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 1900 °C · CMC 0,41 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 2000 °C · CMC 0,42 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 2100 °C · CMC 0,42 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 2200 °C · CMC 0,42 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 2300 °C · CMC 0,42 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 2400 °C · CMC 0,43 °C CMC matrix: temperature meters with a resolution of 0,01 °C for thermocouple type A — 2500 °C · CMC 0,43 °C CMC matrix: temperature simulators for thermocouple type R — 0 °C · CMC 0,95 °C CMC matrix: temperature simulators for thermocouple type R — 100 °C · CMC 0,67 °C CMC matrix: temperature simulators for thermocouple type R — 200 °C · CMC 0,57 °C CMC matrix: temperature simulators for thermocouple type R — 300 °C · CMC 0,51 °C CMC matrix: temperature simulators for thermocouple type R — 400 °C · CMC 0,48 °C CMC matrix: temperature simulators for thermocouple type R — 500 °C · CMC 0,46 °C CMC matrix: temperature simulators for thermocouple type R — 600 °C · CMC 0,44 °C CMC matrix: temperature simulators for thermocouple type R — 700 °C · CMC 0,42 °C CMC matrix: temperature simulators for thermocouple type R — 800 °C · CMC 0,41 °C CMC matrix: temperature simulators for thermocouple type R — 900 °C · CMC 0,39 °C CMC matrix: temperature simulators for thermocouple type R — 1000 °C · CMC 0,38 °C CMC matrix: temperature simulators for thermocouple type R — 1100 °C · CMC 0,37 °C CMC matrix: temperature simulators for thermocouple type R — 1200 °C · CMC 0,36 °C CMC matrix: temperature simulators for thermocouple type R — 1300 °C · CMC 0,36 °C CMC matrix: temperature simulators for thermocouple type R — 1400 °C · CMC 0,35 °C CMC matrix: temperature simulators for thermocouple type R — 1500 °C · CMC 0,36 °C CMC matrix: temperature simulators for thermocouple type R — 1600 °C · CMC 0,36 °C CMC matrix: temperature simulators for thermocouple type R — 1700 °C · CMC 0,37 °C CMC matrix: temperature simulators for thermocouple type S — 0 °C · CMC 0,93 °C CMC matrix: temperature simulators for thermocouple type S — 100 °C · CMC 0,68 °C CMC matrix: temperature simulators for thermocouple type S — 200 °C · CMC 0,59 °C CMC matrix: temperature simulators for thermocouple type S — 300 °C · CMC 0,55 °C CMC matrix: temperature simulators for thermocouple type S — 400 °C · CMC 0,52 °C CMC matrix: temperature simulators for thermocouple type S — 500 °C · CMC 0,51 °C CMC matrix: temperature simulators for thermocouple type S — 600 °C · CMC 0,49 °C CMC matrix: temperature simulators for thermocouple type S — 700 °C · CMC 0,48 °C CMC matrix: temperature simulators for thermocouple type S — 800 °C · CMC 0,46 °C CMC matrix: temperature simulators for thermocouple type S — 900 °C · CMC 0,45 °C CMC matrix: temperature simulators for thermocouple type S — 1000 °C · CMC 0,43 °C CMC matrix: temperature simulators for thermocouple type S — 1100 °C · CMC 0,42 °C CMC matrix: temperature simulators for thermocouple type S — 1200 °C · CMC 0,42 °C CMC matrix: temperature simulators for thermocouple type S — 1300 °C · CMC 0,41 °C CMC matrix: temperature simulators for thermocouple type S — 1400 °C · CMC 0,41 °C CMC matrix: temperature simulators for thermocouple type S — 1500 °C · CMC 0,42 °C CMC matrix: temperature simulators for thermocouple type S — 1600 °C · CMC 0,42 °C CMC matrix: temperature simulators for thermocouple type S — 1700 °C · CMC 0,44 °C CMC matrix: temperature simulators for thermocouple type B — 600 °C · CMC 0,84 °C CMC matrix: temperature simulators for thermocouple type B — 700 °C · CMC 0,73 °C CMC matrix: temperature simulators for thermocouple type B — 800 °C · CMC 0,65 °C CMC matrix: temperature simulators for thermocouple type B — 900 °C · CMC 0,59 °C CMC matrix: temperature simulators for thermocouple type B — 1000 °C · CMC 0,55 °C CMC matrix: temperature simulators for thermocouple type B — 1100 °C · CMC 0,51 °C CMC matrix: temperature simulators for thermocouple type B — 1200 °C · CMC 0,48 °C CMC matrix: temperature simulators for thermocouple type B — 1300 °C · CMC 0,46 °C CMC matrix: temperature simulators for thermocouple type B — 1400 °C · CMC 0,44 °C CMC matrix: temperature simulators for thermocouple type B — 1500 °C · CMC 0,43 °C CMC matrix: temperature simulators for thermocouple type B — 1600 °C · CMC 0,43 °C CMC matrix: temperature simulators for thermocouple type B — 1700 °C · CMC 0,43 °C CMC matrix: temperature simulators for thermocouple type B — 1800 °C · CMC 0,44 °C CMC matrix: temperature simulators for thermocouple type J — -200 °C · CMC 0,23 °C CMC matrix: temperature simulators for thermocouple type J — -100 °C · CMC 0,13 °C CMC matrix: temperature simulators for thermocouple type J — 0 °C · CMC 0,11 °C CMC matrix: temperature simulators for thermocouple type J — 100 °C · CMC 0,10 °C CMC matrix: temperature simulators for thermocouple type J — 200 °C · CMC 0,10 °C CMC matrix: temperature simulators for thermocouple type J — 300 °C · CMC 0,10 °C CMC matrix: temperature simulators for thermocouple type J — 400 °C · CMC 0,10 °C CMC matrix: temperature simulators for thermocouple type J — 500 °C · CMC 0,059 °C CMC matrix: temperature simulators for thermocouple type J — 600 °C · CMC 0,060 °C CMC matrix: temperature simulators for thermocouple type J — 700 °C · CMC 0,059 °C CMC matrix: temperature simulators for thermocouple type J — 800 °C · CMC 0,054 °C CMC matrix: temperature simulators for thermocouple type J — 900 °C · CMC 0,056 °C CMC matrix: temperature simulators for thermocouple type J — 1000 °C · CMC 0,059 °C CMC matrix: temperature simulators for thermocouple type J — 1100 °C · CMC 0,061 °C CMC matrix: temperature simulators for thermocouple type J — 1200 °C · CMC 0,063 °C CMC matrix: temperature simulators for thermocouple type T — -200 °C · CMC 0,32 °C CMC matrix: temperature simulators for thermocouple type T — -100 °C · CMC 0,18 °C CMC matrix: temperature simulators for thermocouple type T — 0 °C · CMC 0,13 °C CMC matrix: temperature simulators for thermocouple type T — 100 °C · CMC 0,11 °C CMC matrix: temperature simulators for thermocouple type T — 200 °C · CMC 0,10 °C CMC matrix: temperature simulators for thermocouple type T — 300 °C · CMC 0,092 °C CMC matrix: temperature simulators for thermocouple type T — 400 °C · CMC 0,088 °C CMC matrix: temperature simulators for thermocouple type E — -200 °C · CMC 0,20 °C CMC matrix: temperature simulators for thermocouple type E — -100 °C · CMC 0,11 °C CMC matrix: temperature simulators for thermocouple type E — 0 °C · CMC 0,087 °C CMC matrix: temperature simulators for thermocouple type E — 100 °C · CMC 0,076 °C CMC matrix: temperature simulators for thermocouple type E — 200 °C · CMC 0,070 °C CMC matrix: temperature simulators for thermocouple type E — 300 °C · CMC 0,067 °C CMC matrix: temperature simulators for thermocouple type E — 400 °C · CMC 0,031 °C CMC matrix: temperature simulators for thermocouple type E — 500 °C · CMC 0,032 °C CMC matrix: temperature simulators for thermocouple type E — 600 °C · CMC 0,034 °C CMC matrix: temperature simulators for thermocouple type E — 700 °C · CMC 0,036 °C CMC matrix: temperature simulators for thermocouple type E — 800 °C · CMC 0,037 °C CMC matrix: temperature simulators for thermocouple type E — 900 °C · CMC 0,040 °C CMC matrix: temperature simulators for thermocouple type E — 1000 °C · CMC 0,042 °C CMC matrix: temperature simulators for thermocouple type K — -200 °C · CMC 0,33 °C CMC matrix: temperature simulators for thermocouple type K — -100 °C · CMC 0,17 °C CMC matrix: temperature simulators for thermocouple type K — 0 °C · CMC 0,14 °C CMC matrix: temperature simulators for thermocouple type K — 100 °C · CMC 0,13 °C CMC matrix: temperature simulators for thermocouple type K — 200 °C · CMC 0,14 °C CMC matrix: temperature simulators for thermocouple type K — 300 °C · CMC 0,13 °C CMC matrix: temperature simulators for thermocouple type K — 400 °C · CMC 0,13 °C CMC matrix: temperature simulators for thermocouple type K — 500 °C · CMC 0,14 °C CMC matrix: temperature simulators for thermocouple type K — 600 °C · CMC 0,081 °C CMC matrix: temperature simulators for thermocouple type K — 700 °C · CMC 0,082 °C CMC matrix: temperature simulators for thermocouple type K — 800 °C · CMC 0,084 °C CMC matrix: temperature simulators for thermocouple type K — 900 °C · CMC 0,086 °C CMC matrix: temperature simulators for thermocouple type K — 1000 °C · CMC 0,088 °C CMC matrix: temperature simulators for thermocouple type K — 1100 °C · CMC 0,090 °C CMC matrix: temperature simulators for thermocouple type K — 1200 °C · CMC 0,092 °C CMC matrix: temperature simulators for thermocouple type K — 1300 °C · CMC 0,096 °C CMC matrix: temperature simulators for thermocouple type N — -200 °C · CMC 0,50 °C CMC matrix: temperature simulators for thermocouple type N — -100 °C · CMC 0,24 °C CMC matrix: temperature simulators for thermocouple type N — 0 °C · CMC 0,20 °C CMC matrix: temperature simulators for thermocouple type N — 100 °C · CMC 0,17 °C CMC matrix: temperature simulators for thermocouple type N — 200 °C · CMC 0,15 °C CMC matrix: temperature simulators for thermocouple type N — 300 °C · CMC 0,14 °C CMC matrix: temperature simulators for thermocouple type N — 400 °C · CMC 0,14 °C CMC matrix: temperature simulators for thermocouple type N — 500 °C · CMC 0,13 °C CMC matrix: temperature simulators for thermocouple type N — 600 °C · CMC 0,13 °C CMC matrix: temperature simulators for thermocouple type N — 700 °C · CMC 0,064 °C CMC matrix: temperature simulators for thermocouple type N — 800 °C · CMC 0,066 °C CMC matrix: temperature simulators for thermocouple type N — 900 °C · CMC 0,067 °C CMC matrix: temperature simulators for thermocouple type N — 1000 °C · CMC 0,069 °C CMC matrix: temperature simulators for thermocouple type N — 1100 °C · CMC 0,071 °C CMC matrix: temperature simulators for thermocouple type N — 1200 °C · CMC 0,073 °C CMC matrix: temperature simulators for thermocouple type N — 1300 °C · CMC 0,076 °C CMC matrix: temperature simulators for thermocouple type C — 0 °C · CMC 0,76 °C CMC matrix: temperature simulators for thermocouple type C — 100 °C · CMC 0,74 °C CMC matrix: temperature simulators for thermocouple type C — 200 °C · CMC 0,73 °C CMC matrix: temperature simulators for thermocouple type C — 300 °C · CMC 0,72 °C CMC matrix: temperature simulators for thermocouple type C — 400 °C · CMC 0,72 °C CMC matrix: temperature simulators for thermocouple type C — 500 °C · CMC 0,71 °C CMC matrix: temperature simulators for thermocouple type C — 600 °C · CMC 0,71 °C CMC matrix: temperature simulators for thermocouple type C — 700 °C · CMC 0,71 °C CMC matrix: temperature simulators for thermocouple type C — 800 °C · CMC 0,72 °C CMC matrix: temperature simulators for thermocouple type C — 900 °C · CMC 0,72 °C CMC matrix: temperature simulators for thermocouple type C — 1000 °C · CMC 0,72 °C CMC matrix: temperature simulators for thermocouple type C — 1100 °C · CMC 0,72 °C CMC matrix: temperature simulators for thermocouple type C — 1200 °C · CMC 0,73 °C CMC matrix: temperature simulators for thermocouple type C — 1300 °C · CMC 0,68 °C CMC matrix: temperature simulators for thermocouple type C — 1400 °C · CMC 0,68 °C CMC matrix: temperature simulators for thermocouple type C — 1500 °C · CMC 0,68 °C CMC matrix: temperature simulators for thermocouple type C — 1600 °C · CMC 0,68 °C CMC matrix: temperature simulators for thermocouple type C — 1700 °C · CMC 0,68 °C CMC matrix: temperature simulators for thermocouple type C — 1800 °C · CMC 0,68 °C CMC matrix: temperature simulators for thermocouple type C — 1900 °C · CMC 0,69 °C CMC matrix: temperature simulators for thermocouple type C — 2000 °C · CMC 0,69 °C CMC matrix: temperature simulators for thermocouple type C — 2100 °C · CMC 0,69 °C CMC matrix: temperature simulators for thermocouple type C — 2200 °C · CMC 0,70 °C CMC matrix: temperature simulators for thermocouple type C — 2300 °C · CMC 0,70 °C CMC matrix: temperature simulators for thermocouple type A — 0 °C · CMC 0,58 °C CMC matrix: temperature simulators for thermocouple type A — 100 °C · CMC 0,53 °C CMC matrix: temperature simulators for thermocouple type A — 200 °C · CMC 0,51 °C CMC matrix: temperature simulators for thermocouple type A — 300 °C · CMC 0,50 °C CMC matrix: temperature simulators for thermocouple type A — 400 °C · CMC 0,50 °C CMC matrix: temperature simulators for thermocouple type A — 500 °C · CMC 0,50 °C CMC matrix: temperature simulators for thermocouple type A — 600 °C · CMC 0,50 °C CMC matrix: temperature simulators for thermocouple type A — 700 °C · CMC 0,50 °C CMC matrix: temperature simulators for thermocouple type A — 800 °C · CMC 0,50 °C CMC matrix: temperature simulators for thermocouple type A — 900 °C · CMC 0,51 °C CMC matrix: temperature simulators for thermocouple type A — 1000 °C · CMC 0,51 °C CMC matrix: temperature simulators for thermocouple type A — 1100 °C · CMC 0,52 °C CMC matrix: temperature simulators for thermocouple type A — 1200 °C · CMC 0,53 °C CMC matrix: temperature simulators for thermocouple type A — 1300 °C · CMC 0,53 °C CMC matrix: temperature simulators for thermocouple type A — 1400 °C · CMC 0,42 °C CMC matrix: temperature simulators for thermocouple type A — 1500 °C · CMC 0,43 °C CMC matrix: temperature simulators for thermocouple type A — 1600 °C · CMC 0,43 °C CMC matrix: temperature simulators for thermocouple type A — 1700 °C · CMC 0,43 °C CMC matrix: temperature simulators for thermocouple type A — 1800 °C · CMC 0,43 °C CMC matrix: temperature simulators for thermocouple type A — 1900 °C · CMC 0,44 °C CMC matrix: temperature simulators for thermocouple type A — 2000 °C · CMC 0,44 °C CMC matrix: temperature simulators for thermocouple type A — 2100 °C · CMC 0,45 °C CMC matrix: temperature simulators for thermocouple type A — 2200 °C · CMC 0,46 °C CMC matrix: temperature simulators for thermocouple type A — 2300 °C · CMC 0,47 °C CMC matrix: temperature simulators for thermocouple type A — 2400 °C · CMC 0,48 °C CMC matrix: temperature simulators for thermocouple type A — 2500 °C · CMC 0,48 °C Electrical simulation of quantities |
Internal procedure; Indirect electrical method; At permanent location (S) | PP/6W2/S01/03 based on EURAMET cg-11 v. 2.0, PN-EN 60584-1:2014 | |
| 117 | Temperature simulators Temperature indicators (meters) |
Calibration — Electrical simulation of quantities -200 °C to 850 °C · CMC 0,0015 °C; Matryca CMC na str. 23 - 25 CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — -200 °C · CMC 0,0021 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — -100 °C · CMC 0,015 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — 0 °C · CMC 0,018 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — 100 °C · CMC 0,021 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — 200 °C · CMC 0,024 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — 300 °C · CMC 0,069 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — 400 °C · CMC 0,074 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — 500 °C · CMC 0,078 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — 600 °C · CMC 0,083 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — 700 °C · CMC 0,088 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — 800 °C · CMC 0,094 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt10 — 850 °C · CMC 0,097 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — -200 °C · CMC 0,0052 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — -100 °C · CMC 0,0080 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — 0 °C · CMC 0,026 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — 100 °C · CMC 0,029 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — 200 °C · CMC 0,032 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — 300 °C · CMC 0,035 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — 400 °C · CMC 0,039 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — 500 °C · CMC 0,042 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — 600 °C · CMC 0,046 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — 700 °C · CMC 0,050 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — 800 °C · CMC 0,054 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt25 — 850 °C · CMC 0,056 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — -200 °C · CMC 0,0022 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — -100 °C · CMC 0,0081 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — 0 °C · CMC 0,011 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — 100 °C · CMC 0,013 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — 200 °C · CMC 0,016 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — 300 °C · CMC 0,061 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — 400 °C · CMC 0,065 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — 500 °C · CMC 0,069 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — 600 °C · CMC 0,074 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — 700 °C · CMC 0,079 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — 800 °C · CMC 0,084 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt100 — 850 °C · CMC 0,087 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — -200 °C · CMC 0,0015 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — -100 °C · CMC 0,0074 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — 0 °C · CMC 0,0098 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — 100 °C · CMC 0,013 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — 200 °C · CMC 0,015 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — 300 °C · CMC 0,022 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — 400 °C · CMC 0,025 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — 500 °C · CMC 0,028 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — 600 °C · CMC 0,031 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — 700 °C · CMC 0,035 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — 800 °C · CMC 0,038 °C CMC matrix: temperature meters (with a reading resolution of 0,001 °C) and temperature simulators, Pt1000 — 850 °C · CMC 0,040 °C Electrical simulation of quantities |
Internal procedure; Indirect electrical method; At permanent location (S) | PP/6W2/S01/03 based on EURAMET cg-11 v. 2.0, PN-EN 60751: 2022 | |
| 118 | Digital timing comparators | Calibration — Time (time interval) -120 s/d to 120 s/d · CMC 0,006 s/d Time (time interval) |
Internal procedure; At permanent location (S) | PP/6W2/S16/01 | |
| 119 | Period generators Oscilloscope calibrators |
Calibration — Time (time interval) period generators — 10 ns to 0,5 s · CMC 1·10⁻¹⁰ ·T oscilloscope calibrators — 10 ns to 0,5 s · CMC 0,0001 % Time (time interval) |
Internal procedure; At permanent location (S) | PP/6W2/S12/01 PP/6W2/S14/01 | |
| 120 | Period meters (frequency counters, timers) | Calibration — Time (time interval) 10 ns to 5 s · CMC 1·10⁻¹⁰ · T Time (time interval) |
Internal procedure; At permanent location (S) | PP/6W2/S11/01 | |
| 121 | Power network parameter meters (residual current device testers) |
Calibration — Time (time interval) 20 ms to 190 ms · CMC 1,1 ms 200 ms to 390 ms · CMC 1,2 ms 400 ms to 1 s · CMC 8,2 ms Time (time interval) |
Internal procedure; At permanent location (S) | PP/6W2/S05/05 | |
| 122 | Time interval meters (frequency counters, timers) | Calibration — Time (time interval) 0,1 µs to 10000 s · CMC 1·10⁻¹⁰ · τ + 5·10⁻⁹ s Time (time interval) |
Internal procedure; At permanent location (S) | PP/6W2/S11/01 | |
| 123 | Oscilloscopes | Calibration — Time (time interval) 50 ns to 0,5 s · CMC 0,07 % Time (time interval) |
Internal procedure; At permanent location (S) | PP/6W2/S15/01 | |
| 124 | Electronic stopwatches | Calibration — Time (time interval) 0 h to 24 h · CMC 3·10⁻⁷ · τ + 0,01 s Time (time interval) |
Internal procedure; At permanent location (S) | PP/6W2/S24/02 | |
| 125 | Mechanical stopwatches | Calibration — Time (time interval) 0 h to 1 h · CMC 1·10⁻⁵ · τ + 0,06 s Time (time interval) |
Internal procedure; At permanent location (S) | PP/6W2/S24/01 | |
| 126 | Quartz and non-quartz generators | Calibration — Frequency square waveforms — 0,2 Hz to 3 GHz · CMC 1·10⁻¹⁰ · f sinusoidal waveforms — 10 kHz to 3 GHz · CMC 1·10⁻¹⁰ ·f sinusoidal waveforms — 0,2 Hz to 10 kHz · CMC 10 µHz Frequency |
Internal procedure; At permanent location (S) | PP/6W2/S12/01 | |
| 127 | Digital frequency meters | Calibration — Frequency square waveforms — 0,2 Hz to 10 MHz · CMC 1·10⁻¹⁰ · f sinusoidal waveforms — 10 kHz to 2,7 GHz · CMC 1·10⁻¹⁰ · f Frequency |
Internal procedure; At permanent location (S) | PP/6W2/S11/01 | |
| 128 | Power network analysers Power network parameter meters Clamp meters |
Calibration — Frequency 40 Hz to 1 kHz · CMC 0,006 Hz Frequency |
Internal procedure; At permanent location (S) | PP/6W2/S05/05 | |
| 129 | Thermohygrometers | Calibration — Relative humidity 10 °C to 23 °C; 50 % rh to 80 % rh — 50 % rh · CMC 1,0 % rh 10 °C to 23 °C; 50 % rh to 80 % rh — 80 % rh · CMC 1,4 % rh 23 °C to 40 °C; 20 % rh to 80 % rh — 20 % rh · CMC 0,7 % rh 23 °C to 40 °C; 20 % rh to 80 % rh — 80 % rh · CMC 1,4 % rh 40 °C; 20 % rh to 50 % rh — 20 % rh · CMC 0,7 % rh 40 °C; 20 % rh to 50 % rh — 50 % rh · CMC 1,4 % rh Relative humidity |
Internal procedure; Comparison method Calibration in a climatic chamber; At permanent location (S) | PP/6W2/S41/01 | |
| 130 | Spectrophotometers (VIS and NIR) | Calibration — Optical density of spectral transmittance zakres widmowy 400 nm to 890 nm — 0 to 0,3 · CMC 0,0045 zakres widmowy 400 nm to 890 nm — 0,3001 to 0,8 · CMC 0,0054 zakres widmowy 400 nm to 890 nm — 0,8001 to 1,9 · CMC 0,0069 Optical density of spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W2/S27/01 | |
| 131 | Spectrophotometers (UV) | Calibration — Optical density of spectral transmittance zakres widmowy 235 nm to 400 nm — 0 to 0,3 · CMC 0,0045 zakres widmowy 235 nm to 400 nm — 0,3001 to 0,8 · CMC 0,0054 zakres widmowy 235 nm to 400 nm — 0,8001 to 1,9 · CMC 0,0069 Optical density of spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W2/S27/01 | |
| 132 | Reference optical filters | Calibration — Optical density of spectral transmittance zakres widmowy 235 nm to 890 nm — 0 to 0,3 · CMC 0,0039 zakres widmowy 235 nm to 890 nm — 0,3001 to 0,8 · CMC 0,0048 zakres widmowy 235 nm to 890 nm — 0,8001 to 1,9 · CMC 0,0062 Optical density of spectral transmittance |
Internal procedure; At permanent location (S) | PP/6W2/S27/02 | |
| 133 | Spectrophotometers (VIS and NIR) | Calibration — Spectral transmittance zakres widmowy 400 nm to 890 nm — 0,01 to 0,15 · CMC 0,0031 zakres widmowy 400 nm to 890 nm — 0,151 to 0,5 · CMC 0,0051 zakres widmowy 400 nm to 890 nm — 0,501 to 1 · CMC 0,0079 Spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W2/S27/01 | |
| 134 | Spectrophotometers (UV) | Calibration — Spectral transmittance zakres widmowy 235 nm to 400 nm — 0,01 to 0,15 · CMC 0,0031 zakres widmowy 235 nm to 400 nm — 0,151 to 0,5 · CMC 0,0051 zakres widmowy 235 nm to 400 nm — 0,501 to 1 · CMC 0,0079 Spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W2/S27/01 | |
| 135 | Reference optical filters | Calibration — Spectral transmittance zakres widmowy 235 nm to 890 nm — 0,01 to 0,15 · CMC 0,0023 zakres widmowy 235 nm to 890 nm — 0,151 to 0,5 · CMC 0,0046 zakres widmowy 235 nm to 890 nm — 0,501 to 1 · CMC 0,0075 Spectral transmittance |
Internal procedure; At permanent location (S) | PP/6W2/S27/02 | |
| 136 | Spectrophotometers VIS, UV, NIR | Calibration — Spectral transmittance wavelength — 275 nm to 810 nm · CMC 0,28 nm Spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W2/S27/01 | |
| 137 | Reference optical wavelength filters | Calibration — Spectral transmittance wavelength — 275 nm to 810 nm · CMC 0,26 nm Spectral transmittance |
Internal procedure; At permanent location (S) | PP/6W2/S27/02 | |
| 138 | Electronic pressure gauges - absolute pressure |
Calibration — Pressure 850 hPa to 1100 hPa · CMC 0,4 hPa Pressure |
Internal procedure; At permanent location (S) | PP/6W2/S42/01 | |
| 139 | Electrical thermometers (including electronic) | Calibration — Temperature (electrical thermometry) -30 °C to 60 °C · CMC 0,03 °C 60 °C to 150 °C · CMC 0,06 °C 150 °C to 200 °C · CMC 1,5 °C 200 °C to 300 °C · CMC 2,0 °C 300 °C to 400 °C · CMC 3,5 °C 400 °C to 500 °C · CMC 4,5 °C Temperature (electrical thermometry) |
Internal procedure; Comparison method Calibration in a liquid bath, temperature calibrator; At permanent location (S) | PP/6W2/S40/01 | |
| 140 | Electrical thermometers (including electronic) | Calibration — Temperature (electrical thermometry) -30 °C to -10 °C · CMC 1,2 °C -10 °C to 0 °C · CMC 0,7 °C 0 °C · CMC 0,5 °C 0 °C to 25 °C · CMC 0,2 °C 25 °C to 50 °C · CMC 0,3 °C 50 °C to 80 °C · CMC 0,7 °C Temperature (electrical thermometry) |
Internal procedure; Comparison method Calibration in a climatic chamber; At permanent location (S) | PP/6W2/S41/01 | |
| 141 | Thermal imaging cameras Radiation pyrometers |
Calibration — Temperature (radiation thermometry) -15 °C do 0 °C · CMC 2,0 °C 0 °C do 100 °C · CMC 1,5 °C 100 °C to 200 °C · CMC 2,0 °C 200 °C do 350 °C · CMC 2,5 °C 350 °C do 500 °C · CMC 3,5 °C Temperature (radiation thermometry) |
Internal procedure; At permanent location (S) | PP/6W2/S43/01 based on the FLUKE manufacturer's measurement instructions and MSL guides | |
| 142 | Bourdon tube pressure gauges Electronic pressure gauges |
Calibration — Pressure gauge pressure - negative and positive pressure (gas) — -0,003 MPa to -0,09 MPa · CMC 4,9·10⁻⁵ MPa gauge pressure - negative and positive pressure (gas) — 0,02 MPa to 1 MPa · CMC 1,5·10⁻⁴ MPa gauge pressure - positive pressure (oil) — 0,02 MPa to 0,6 MPa · CMC 2,5·10⁻⁴ MPa gauge pressure - positive pressure (oil) — 0,6 MPa to 6 MPa · CMC 3,5·10⁻³ MPa gauge pressure - positive pressure (oil) — 6 MPa to 25 MPa · CMC 1,5·10⁻³ MPa gauge pressure - positive pressure (oil) — 25 MPa to 70 MPa · CMC 4,0·10⁻² MPa gauge pressure - positive pressure (oil) — 70 MPa to 260 MPa · CMC 1,5·10⁻¹ MPa Pressure |
Internal procedure; At permanent location (S) | PP/6W3/S01/01 based on EURAMET cg-17 v. 4.1 | |
| 143 | Non-automatic electronic weighing instruments | Calibration — Mass (weighing instruments) 0 g to 300 g · CMC 6,8 · 10⁻⁷ · m 300 g to 1200 g · CMC 8,2 · 10⁻⁶ · m 1200 g to 35 kg · CMC 2,0 · 10⁻⁵ · m 35 kg to 600 kg · CMC 1,7 · 10⁻⁴ · m 600 kg to 4000 kg · CMC 4,1 · 10⁻⁴ · m Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W1/S02/02 based on EURAMET cg-18 v.4.0 | |
| 144 | Mass standards, weights of accuracy class F1 | Calibration — Mass (weights and mass standards) 1g · CMC 0,030 mg 2g · CMC 0,040 mg 5g · CMC 0,050 mg 10 g · CMC 0,060 mg 20 g · CMC 0,080 mg 50 g · CMC 0,10 mg 100 g · CMC 0,16 mg 200 g · CMC 0,30 mg 500 g · CMC 0,80 mg 1 kg · CMC 1,6 mg 2 kg · CMC 3 mg 5 kg · CMC 8 mg 10 kg · CMC 16 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 145 | Mass standards, weights of accuracy class F2 | Calibration — Mass (weights and mass standards) 1 mg, 2 mg, 5 mg · CMC 0,020 mg 10 mg · CMC 0,026 mg 20 mg · CMC 0,033 mg 50 mg · CMC 0,040 mg 100 mg · CMC 0,053 mg 200 mg · CMC 0,060 mg 500 mg · CMC 0,083 mg 1g · CMC 0,10 mg 2g · CMC 0,13 mg 5g · CMC 0,16 mg 10 g · CMC 0,20 mg 20 g · CMC 0,26 mg 50 g · CMC 0,33 mg 100 g · CMC 0,53 mg 200 g · CMC 1,0 mg 500 g · CMC 2,6 mg 1 kg · CMC 5,0 mg 2 kg · CMC 10 mg 5 kg · CMC 26 mg 10 kg · CMC 53 mg 20 kg · CMC 100 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 146 | Mass standards, weights of accuracy class M1 | Calibration — Mass (weights and mass standards) 1 mg, 2 mg, 5 mg · CMC 0,06 mg 10 mg · CMC 0,08 mg 20 mg · CMC 0,10 mg 50 mg · CMC 0,13 mg 100 mg · CMC 0,16 mg 200 mg · CMC 0,20 mg 500 mg · CMC 0,26 mg 1g · CMC 0,33 mg 2g · CMC 0,40 mg 5g · CMC 0,53 mg 10 g · CMC 0,66 mg 20 g · CMC 0,83 mg 50 g · CMC 1,0 mg 100 g · CMC 1,6 mg 200 g · CMC 3,3 mg 500 g · CMC 8,3 mg 1 kg · CMC 16 mg 2 kg · CMC 33 mg 5 kg · CMC 83 mg 10 kg · CMC 160 mg 20 kg · CMC 330 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 147 | Mass standards, weights of accuracy class M2 | Calibration — Mass (weights and mass standards) 100 mg · CMC 0,5 mg 200 mg · CMC 0,6 mg 500 mg · CMC 0,8 mg 1g · CMC 1,0 mg 2g · CMC 1,3 mg 5g · CMC 1,6 mg 10 g · CMC 2,0 mg 20 g · CMC 2,6 mg 50 g · CMC 3,3 mg 100 g · CMC 5 mg 200 g · CMC 10 mg 500 g · CMC 26 mg 1 kg · CMC 53 mg 2 kg · CMC 100 mg 5 kg · CMC 260 mg 10 kg · CMC 530 mg 20 kg · CMC 1000 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 148 | 25 kg mass standards | Calibration — Mass (weights and mass standards) 25 kg · CMC 400 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) and outside of it (P) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
| 149 | Load weights | Calibration — Mass (weights and mass standards) 1 g to 25 kg · CMC 1,6 · 10⁻⁵ · m Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PP/6W1/S03/01 based on OIML R-111-1:2004 Annex C | |
Reference standards (92)
EURAMET cg-11 v. 2.0,
EURAMET cg-15 v. 3.0
EURAMET cg-15 v. 3.0 02/2015
EURAMET cg-17 v. 4.1
EURAMET cg-18 v.4.0
EURAMET cg-7 v. 1.0
OIML R-111-1:2004 Załącznik C
OIML R126-2:2021, A1.1, A2
OIML R50-1:2014 (p. 7)
OIML R50-2:2014 (p. 10)
PN-EN 60584-1:2014
PN-EN 60751: 2022
PN-EN ISO 10360-7:2011
PN-EN ISO 13225:2012
PN-EN ISO 13385-1:2019
PN-EN ISO 13385-2:2020
PN-EN ISO 6506-2
PN-EN ISO 6507-2
PN-EN ISO 6508-2
PN-EN ISO 7500-1:2018
PN-ISO 8512-2:1999
PP/6W1/S02/02
PP/6W1/S03/01
PP/6W1/S03/02
PP/6W1/S03/03
PP/6W1/S04/02
PP/6W1/S15/02
PP/6W1/S15/05
PP/6W1/S16/01
PP/6W1/S16/03
PP/6W1/S17/01
PP/6W1/S18/01
PP/6W1/S19/01
PP/6W1/S20/05
PP/6W1/S20/06
PP/6W1/S20/07
PP/6W1/S20/10
PP/6W1/S20/11
PP/6W1/S20/13
PP/6W1/S20/14
PP/6W1/S20/15
PP/6W1/S20/16
PP/6W1/S20/17
PP/6W1/S21/01
PP/6W1/S23/01
PP/6W1/S23/02
PP/6W1/S23/03
PP/6W1/S23/04
PP/6W1/S23/05
PP/6W1/S24/01
PP/6W1/S24/02
PP/6W1/S24/05
PP/6W1/S24/06
PP/6W1/S24/07
PP/6W1/S24/08
PP/6W1/S24/10
PP/6W1/S25/01
PP/6W1/S26/01
PP/6W1/S26/02
PP/6W1/S26/03
PP/6W1/S28/01
PP/6W2/S01/01
PP/6W2/S01/02
PP/6W2/S01/03
PP/6W2/S04/01
PP/6W2/S05/05
PP/6W2/S07/01
PP/6W2/S08/01
PP/6W2/S10/01
PP/6W2/S10/02
PP/6W2/S10/03
PP/6W2/S10/04
PP/6W2/S10/05
PP/6W2/S11/01
PP/6W2/S12/01
PP/6W2/S14/01
PP/6W2/S15/01
PP/6W2/S16/01
PP/6W2/S24/01
PP/6W2/S24/02
PP/6W2/S27/01
PP/6W2/S27/02
PP/6W2/S28/01
PP/6W2/S29/01
PP/6W2/S30/01
PP/6W2/S34/01
PP/6W2/S40/01
PP/6W2/S41/01
PP/6W2/S42/01
PP/6W2/S43/01
PP/6W3/S01/01