🇵🇱 Okręgowy Urząd Miar w Katowicach PL
Metrology & calibration
Okręgowy Urząd Miar w Katowicach is a calibration laboratory accredited by PCA under number AP 085 (Katowice, Poland). Its accreditation scope covers 25 calibration items in the following areas: Metrology & calibration. Calibrated items include: Calibrators, Multimeters, Analogue ammeters, Analogue voltmeters. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Direct method; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P).
At a glance
- Accreditation number: PCA AP 085
- Calibration items in scope: 25
- Types of calibrated items: 27
- Calibration principles used: 5
- City: Katowice, Poland
- Valid until: 18-01-2030
- Accreditation number
- AP 085
- Accreditation body
- PCA
- Address
- ul. Rynek 9, 40-957 Katowice
- City
- Katowice, Poland
- Division
- Calibration Laboratories Group
- Valid until
- 18-01-2030
What this laboratory calibrates (27)
- Calibrators — 5 calibration items
- Multimeters — 4 calibration items
- Analogue ammeters — 2 calibration items
- Analogue voltmeters — 2 calibration items
- Digital ammeters — 2 calibration items
- Digital voltmeters — 2 calibration items
- Electronic pressure gauges — 2 calibration items
- Gauge blocks (classes 0, 1, 2) — 2 calibration items
- Spectrophotometers — 2 calibration items
- Spring pressure gauges — 2 calibration items
- Conductometers — 1 calibration item
- Digital ohmmeters — 1 calibration item
- Digital timing machines (chronocomparators) — 1 calibration item
- Electronic stopwatches — 1 calibration item
- Fixed resistors — 1 calibration item
- and 12 more
Calibration principles (5)
- Internal procedure; At permanent location (S) — 10 calibration items
- Internal procedure; Direct method; At permanent location (S) — 9 calibration items
- Internal procedure; At permanent location (S) and outside of it (P) — 2 calibration items
- Internal procedure; Comparison method in vertical arrangement; At permanent location (S) — 2 calibration items
- Internal procedure; Electrical method; At permanent location (S) — 2 calibration items
Certificate information
- Issue date
- 19.12.2025
- Issue no.
- 19
- Certificate date
- 13.08.2020
- Accreditation cycle
- 19.12.2025 — 18.01.2030
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (25)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Gauge blocks (classes 0, 1, 2) | Calibration — Length 0,46 mm to 100 mm · CMC Q[57; 1,9·L] nm Length |
Internal procedure; Comparison method in vertical arrangement; At permanent location (S) | I/W1/09/01 | |
| 2 | Gauge blocks (classes 0, 1, 2) | Calibration — Length 125 mm to 500 mm · CMC Q[95; 2,0·L] nm Length |
Internal procedure; Comparison method in vertical arrangement; At permanent location (S) | I/W1/09/01 | |
| 3 | Johansson angle gauge blocks Kusznikow angle gauge blocks Wringing angle gauge blocks |
Calibration — Angle 180° · CMC 3,7” Angle |
Internal procedure; At permanent location (S) | I/W1/09/03 | |
| 4 | Electronic pressure gauges Spring pressure gauges |
Calibration — Pressure - gauge pressure: gas — - 0,003 MPa to - 0,1 MPa · CMC 5 · 10⁻⁶ MPa - gauge pressure: gas — 0,001 MPa to 0,1 MPa · CMC 5 · 10⁻⁶ MPa - gauge pressure: gas — >0,1 MPa to 3,5 MPa · CMC 6 · 10⁻⁵ MPa Pressure |
Internal procedure; At permanent location (S) | I/W1/09/12 I/W1/09/12a I/W1/09/12b I/W1/09/12c I/W1/09/12d I/W1/09/12e | |
| 5 | Electronic pressure gauges Spring pressure gauges |
Calibration — Pressure - gauge pressure: liquid — 0,02 MPa to 0,6 MPa · CMC 5 · 10⁻⁴ MPa - gauge pressure: liquid — >0,6 MPa to 6,0 MPa · CMC 1 · 10⁻³ MPa - gauge pressure: liquid — >6,0 MPa to 100,0 MPa · CMC 2 · 10⁻² MPa Pressure |
Internal procedure; At permanent location (S) | I/W1/09/12 I/W1/09/12a I/W1/09/12b I/W1/09/12c I/W1/09/12d I/W1/09/12e | |
| 6 | pH meters | Calibration — pH - pH — 0 to 14 · CMC 0,002 - DC voltage — -1900 mV to 1900 mV · CMC 0,2 mV pH |
Internal procedure; Electrical method; At permanent location (S) | I/W3/09/01 | |
| 7 | Conductometers | Calibration — Specific electrical conductivity (conductometry) 0,2 µS/cm to 2 µS/cm · CMC 0,3 % 0,002 mS/cm to 200 mS/cm · CMC 0,2 % Specific electrical conductivity (conductometry) |
Internal procedure; Electrical method; At permanent location (S) | I/W3/09/13 | |
| 8 | Calibrators | Calibration — DC voltage 10 mV to 1000 V · CMC 0,003 % DC voltage |
Internal procedure; Direct method; At permanent location (S) | I/W3/09/12 | |
| 9 | Analogue voltmeters Digital voltmeters Multimeters |
Calibration — DC voltage 0,2 mV to 2 mV · CMC 0,5 % 2 mV to 20 mV · CMC 0,1 % 20 mV to 200 mV · CMC 0,01 % 0,2 V to 1000 V · CMC 0,005 % DC voltage |
Internal procedure; Direct method; At permanent location (S) | I/W3/09/03 I/W3/09/10 | |
| 10 | Calibrators | Calibration — DC current 10 µA to 200 mA · CMC 0,007 % 200 mA to 2 A · CMC 0,03 % 2 A to 10 A · CMC 0,08 % DC current |
Internal procedure; Direct method; At permanent location (S) | I/W3/09/12 | |
| 11 | Analogue ammeters Digital ammeters Multimeters |
Calibration — DC current 0,2 mA to 200 mA · CMC 0,03 % 0,2 A to 2 A · CMC 0,07 % 2 A to 10 A · CMC 0,15 % DC current |
Internal procedure; Direct method; At permanent location (S) | I/W3/09/03 I/W3/09/10 | |
| 12 | Calibrators | Calibration — AC voltage 40 Hz to 10 kHz — 10 mV to 1000 V · CMC 0,02 % AC voltage |
Internal procedure; Direct method; At permanent location (S) | I/W3/09/12 | |
| 13 | Analogue voltmeters Digital voltmeters Multimeters |
Calibration — AC voltage 40 Hz to 20 kHz — 0,2 mV to 2 mV · CMC 1% 40 Hz to 20 kHz — 2 mV to 20 mV · CMC 0,2 % 40 Hz to 20 kHz — 20 mV to 200 mV · CMC 0,1 % 40 Hz to 20 kHz — 0,2 V to 200 V · CMC 0,07 % 40 Hz to 5 kHz — 200 V to 1000 V · CMC 0,07 % AC voltage |
Internal procedure; Direct method; At permanent location (S) | I/W3/09/03 I/W3/09/10 | |
| 14 | Calibrators | Calibration — AC current 40 Hz to 10 kHz — 10 µA to 200 µA · CMC 0,1 % 40 Hz to 10 kHz — 0,2 mA to 200 mA · CMC 0,06 % 40 Hz to 2 kHz — 0,2 A to 2 A · CMC 0,1 % 40 Hz to 2 kHz — 2 A to 10 A · CMC 0,2 % 2 kHz to 10 kHz — 2 A to 10 A · CMC 0,4 % AC current |
Internal procedure; Direct method; At permanent location (S) | I/W3/09/12 | |
| 15 | Analogue ammeters Digital ammeters Multimeters |
Calibration — AC current 40 Hz to 5 kHz — 0,2 mA to 200 mA · CMC 0,1 % 40 Hz to 5 kHz — 0,2 A to 2 A · CMC 0,15 % 40 Hz to 1 kHz — 2 A to 10 A · CMC 0,3 % AC current |
Internal procedure; Direct method; At permanent location (S) | I/W3/09/03 I/W3/09/10 | |
| 16 | Fixed resistors Variable resistors Digital ohmmeters Calibrators |
Calibration — DC resistance 0,001 Ω to 1 Ω · CMC 0,005 % 1 Ω to 100 MΩ · CMC 0,003 % 100 MΩ to 1 GΩ · CMC 0,01 % DC resistance |
Internal procedure; Direct method; At permanent location (S) | I/W3/09/03 I/W3/09/04 I/W3/09/06 I/W3/09/12 | |
| 17 | Mechanical stopwatches | Calibration — Time (time interval) 0 h to 1 h · CMC 5·10⁻⁵·τ + 0,06 s Time (time interval) |
Internal procedure; At permanent location (S) | I/W3/09/11 | |
| 18 | Electronic stopwatches | Calibration — Time (time interval) 0 h to 24 h · CMC 6·10⁻⁷·τ + 0,02 s Time (time interval) |
Internal procedure; At permanent location (S) | I/W3/09/05 | |
| 19 | Period meters Period generators |
Calibration — Time (time interval) 0,75 ns to 10 s · CMC (15 µHz ·T +1,2·10⁻⁹)·T 5 ns to 10 s · CMC (15 µHz ·T +1,2·10⁻⁹)·T Time (time interval) |
Internal procedure; At permanent location (S) | I/W3/09/09 I/W3/09/14 | |
| 20 | Time interval meters Time interval generator |
Calibration — Time (time interval) 20 ns to 2000 s · CMC 1,2·10⁻⁹·τ + 10 ns Time (time interval) |
Internal procedure; At permanent location (S) | I/W3/09/09 I/W3/09/14 | |
| 21 | Digital timing machines (chronocomparators) | Calibration — Time (time interval) -120 s/d to 120 s/d · CMC 0,7·Δ₁ s/d Time (time interval) |
Internal procedure; At permanent location (S) | I/W3/09/08 | |
| 22 | Frequency meters | Calibration — Frequency 0,1 Hz to 1,35 GHz · CMC 1,2·10⁻⁹ · f + 15 µHz Frequency |
Internal procedure; At permanent location (S) | I/W3/09/09 | |
| 23 | Quartz generators | Calibration — Frequency 0,1 Hz to 10 kHz · CMC 15 µHz + 1,2·10⁻⁹·f 10 kHz to 350 MHz · CMC 1,2 · 10⁻⁹·f Frequency |
Internal procedure; At permanent location (S) | I/W3/09/14 | |
| 24 | Spectrophotometers | Calibration — Optical density of spectral transmittance filter PDC Blank — 0,0352 for λ=350nm · CMC 0,0035 filter PDC Blank — 0,0366 for λ=313nm · CMC 0,0035 filter PDC Blank — 0,0411 for λ=257nm · CMC 0,0035 filter PDC Blank — 0,0455 for λ=235nm · CMC 0,0035 filter PDC 20 mg/l — 0,2489 for λ=350nm · CMC 0,0039 filter PDC 20 mg/l — 0,1352 for λ=313nm · CMC 0,0039 filter PDC 20 mg/l — 0,3286 for λ=257nm · CMC 0,0039 filter PDC 20 mg/l — 0,2946 for λ=235nm · CMC 0,0039 filter PDC 40mg/l — 0,4500 for λ=350nm · CMC 0,0044 filter PDC 40mg/l — 0,2255 for λ=313nm · CMC 0,0044 filter PDC 40mg/l — 0,6014 for λ=257nm · CMC 0,0044 filter PDC 40mg/l — 0,5284 for λ=257nm · CMC 0,0044 filter PDC 60mg/l — 0,6656 for λ=350nm · CMC 0,0050 filter PDC 60mg/l — 0,3219 for λ=313nm · CMC 0,0050 filter PDC 60mg/l — 0,8912 for λ=257nm · CMC 0,0050 filter PDC 60mg/l — 0,7774 for λ=235nm · CMC 0,0050 filter PDC 80mg/l — 0,8811 for λ=350nm · CMC 0,0058 filter PDC 80mg/l — 0,4214 for λ=313nm · CMC 0,0058 filter PDC 80mg/l — 1,1854 for λ=257nm · CMC 0,0058 filter PDC 80mg/l — 1,0314 for λ=235nm · CMC 0,0058 filter PDC 100mg/l — 1,0975 for λ=350nm · CMC 0,0079 filter PDC 100mg/l — 0,5173 for λ=313nm · CMC 0,0079 filter PDC 100mg/l — 1,4809 for λ=257nm · CMC 0,0079 filter PDC 100mg/l — 1,2844 for λ=235nm · CMC 0,0079 0 to 0,2 · CMC 0,0038 0,2 to 0,5 · CMC 0,0041 0,5 to 1 · CMC 0,0059 Optical density of spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | I/W3/09/02 | |
| 25 | Spectrophotometers | Calibration — Spectral transmittance filter PDC Blank — 0,9222 for λ=350nm · CMC 0,0064 filter PDC Blank — 0,9192 for λ=313nm · CMC 0,0064 filter PDC Blank — 0,9096 for λ=257nm · CMC 0,0064 filter PDC Blank — 0,9005 for λ=235nm · CMC 0,0064 filter PDC 20 mg/l — 0,5638 for λ=350nm · CMC 0,0055 filter PDC 20 mg/l — 0,7325 for λ=313nm · CMC 0,0055 filter PDC 20 mg/l — 0,4692 for λ=257nm · CMC 0,0055 filter PDC 20 mg/l — 0,5075 for λ=235nm · CMC 0,0055 filter PDC 40mg/l — 0,3548 for λ=350nm · CMC 0,0048 filter PDC 40mg/l — 0,5950 for λ=313nm · CMC 0,0048 filter PDC 40mg/l — 0,2504 for λ=257nm · CMC 0,0048 filter PDC 40mg/l — 0,2962 for λ=235nm · CMC 0,0048 filter PDC 60mg/l — 0,2160 for λ=350nm · CMC 0,0042 filter PDC 60mg/l — 0,4766 for λ=313nm · CMC 0,0042 filter PDC 60mg/l — 0,1285 for λ=257nm · CMC 0,0042 filter PDC 60mg/l — 0,1669 for λ=235nm · CMC 0,0042 filter PDC 80mg/l — 0,1315 for λ=350nm · CMC 0,0037 filter PDC 80mg/l — 0,3790 for λ=313nm · CMC 0,0037 filter PDC 80mg/l — 0,0653 for λ=257nm · CMC 0,0037 filter PDC 80mg/l — 0,0930 for λ=235nm · CMC 0,0037 filter PDC 100mg/l — 0,0799 for λ=350nm · CMC 0,0032 filter PDC 100mg/l — 0,3039 for λ=313nm · CMC 0,0032 filter PDC 100mg/l — 0,0330 for λ=257nm · CMC 0,0032 filter PDC 100mg/l — 0,0520 for λ=235nm · CMC 0,0032 0,1 to 0,3 · CMC 0,0030 0,3 to 0,6 · CMC 0,0040 0,6 to 1 · CMC 0,0050 - wavelength — 270 nm to 780 nm · CMC 0,2 nm Spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | I/W3/09/02 | |
Reference standards (21)
I/W1/09/01
I/W1/09/03
I/W1/09/12
I/W1/09/12a
I/W1/09/12b
I/W1/09/12c
I/W1/09/12d
I/W1/09/12e
I/W3/09/01
I/W3/09/02
I/W3/09/03
I/W3/09/04
I/W3/09/05
I/W3/09/06
I/W3/09/08
I/W3/09/09
I/W3/09/10
I/W3/09/11
I/W3/09/12
I/W3/09/13
I/W3/09/14