🇵🇱 Okręgowy Urząd Miar w Poznaniu PL
Metrology & calibration
Okręgowy Urząd Miar w Poznaniu is a calibration laboratory accredited by PCA under number AP 084 (Poznań, Poland). Its accreditation scope covers 40 calibration items in the following areas: Metrology & calibration. Calibrated items include: Multimeters, Calibrators, Clamp meters, Distortion meters. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Direct method; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P).
At a glance
- Accreditation number: PCA AP 084
- Calibration items in scope: 40
- Types of calibrated items: 39
- Calibration principles used: 4
- City: Poznań, Poland
- Valid until: 18-01-2030
- Accreditation number
- AP 084
- Accreditation body
- PCA
- Address
- ul. Krakowska 19, 61-893 Poznań
- City
- Poznań, Poland
- Division
- Calibration Laboratories Group
- Valid until
- 18-01-2030
What this laboratory calibrates (39)
- Multimeters — 5 calibration items
- Calibrators — 4 calibration items
- Clamp meters — 4 calibration items
- Distortion meters — 4 calibration items
- Power quality meters — 4 calibration items
- Bourdon tube pressure gauges — 3 calibration items
- Electronic pressure gauges — 3 calibration items
- Pressure transmitters — 3 calibration items
- Spectrophotometers — 3 calibration items
- Digital ammeters — 2 calibration items
- Digital voltmeters — 2 calibration items
- Gauge blocks (classes 0, 1, 2) — 2 calibration items
- Brinell hardness testers — 1 calibration item
- Digital ohmmeters — 1 calibration item
- Electrically controlled electronic timers — 1 calibration item
- and 24 more
Calibration principles (4)
- Internal procedure; At permanent location (S) — 20 calibration items
- Internal procedure; Direct method; At permanent location (S) — 12 calibration items
- Internal procedure; At permanent location (S) and outside of it (P) — 6 calibration items
- Internal procedure; Outside of permanent location (P) — 2 calibration items
Certificate information
- Issue date
- 18.12.2025
- Issue no.
- 26
- Certificate date
- 20.01.2020
- Accreditation cycle
- 18.12.2025 — 18.01.2030
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (40)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Testing machines for static tests Technological equipment |
Calibration — Force - for tensile forces — 0,01 N to 100 N · CMC 0,1 % - for tensile forces — 100 N to 250 kN · CMC 0,12 % - for tensile forces — 250 kN to 1 MN · CMC 0,24 % - for compressive forces — 0,01 N to 100 N · CMC 0,1 % - for compressive forces — 100 N to 500 kN · CMC 0,12 % - for compressive forces — 500 kN to 3 MN · CMC 0,24 % Force |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure IW/L1/S7/01 based on PN-EN ISO 7500-1:2018-05 | |
| 2 | Torque wrenches Torque screwdrivers |
Calibration — Torque 2 Nm to 1000 Nm · CMC 0,5 % Torque |
Internal procedure; At permanent location (S) | Internal procedure IW/L1/S9/01 | |
| 3 | Rockwell hardness testers | Calibration — Hardness 20 HRA to 88 HRA · CMC 0,5 HRA 20 HRBW to 45 HRBW · CMC 1,0 HRBW 45 HRBW to 100 HRBW · CMC 0,8 HRBW 20 HRC to 70 HRC · CMC 0,6 HRC 70 HR15N to 91 HR15N · CMC 0,5 HR15N 42 HR30N to 80 HR30N · CMC 0,5 HR30N 20 HR45N to 70 HR45N · CMC 0,5 HR45N - force — 29,42 N to 1471 N · CMC 0,24 % Hardness |
Internal procedure; Outside of permanent location (P) | Internal procedure IW/L1/S6/02 based on PN-EN ISO 6508-2:2015-04 | |
| 4 | Vickers hardness testers | Calibration — Hardness HV0,1 · CMC 7,0 % HV0,2 · CMC 6,0 % HV0,3 · CMC 4,5 % HV0,5 · CMC 4% HV1 · CMC 3% HV2 · CMC 2% HV5 · CMC 2% HV10 · CMC 2% HV30 · CMC 2,5 % - force — 0,9807 N to 294,2 N · CMC 0,24 % - length — 0 mm to 1 mm · CMC 0,0005 mm Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure IW/L1/S6/01 based on PN-EN ISO 6507-2:2018-05 | |
| 5 | Brinell hardness testers | Calibration — Hardness HBW2,5/187,5 · CMC 2% HBW5/250 · CMC 2% HBW5/750 · CMC 2% HBW10/3000 · CMC 2% - force — 1839 N to 29420 N · CMC 0,24 % - length — 0 mm to 1 mm · CMC 0,0005 mm - length — 1 mm to 7 mm · CMC 0,005 mm Hardness |
Internal procedure; Outside of permanent location (P) | Internal procedure IW/L1/S6/03 based on PN-EN ISO 6506-2:2019-10 | |
| 6 | Non-automatic electronic weighing instruments Non-automatic mechanical weighing instruments |
Calibration — Mass (weighing instruments) up to 500 g · CMC 2 x 10⁻⁴ % 500 g do 20 kg · CMC 10 x 10⁻⁴ % 20 kg do 60 kg · CMC 17 x 10⁻⁴ % 60 kg do 150 kg · CMC 12 x 10⁻³ % Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure IW/L1/S2/01 based on EURAMET cg-18 v.4.0 | |
| 7 | Mass standards and weights of accuracy class F1 | Calibration — Mass (weights and mass standards) 1 mg · CMC 0,005 mg 2 mg · CMC 0,005 mg 5 mg · CMC 0,005 mg 10 mg · CMC 0,006 mg 20 mg · CMC 0,01 mg 50 mg · CMC 0,01 mg 100 mg · CMC 0,01 mg 200 mg · CMC 0,01 mg 500 mg · CMC 0,02 mg 1g · CMC 0,02 mg 2g · CMC 0,03 mg 5g · CMC 0,04 mg 10 g · CMC 0,05 mg 20 g · CMC 0,06 mg 50 g · CMC 0,1 mg 100 g · CMC 0,1 mg 200 g · CMC 0,2 mg 500 g · CMC 0,6 mg 1 kg · CMC 1,2 mg 2 kg · CMC 2 mg 5 kg · CMC 6 mg 10 kg · CMC 12 mg 20 kg · CMC 25 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | Internal procedure IW/L1/S1/01 based on OIML R111-1:2004 Annex C | |
| 8 | Mass standards and weights of accuracy class F2 | Calibration — Mass (weights and mass standards) 1 mg · CMC 0,01 mg 2mg · CMC 0,01 mg 5 mg · CMC 0,01 mg 10 mg · CMC 0,02 mg 20 mg · CMC 0,02 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,1 mg 2g · CMC 0,1 mg 5g · CMC 0,1 mg 10 g · CMC 0,1 mg 20 g · CMC 0,2 mg 50 g · CMC 0,2 mg 100 g · CMC 0,4 mg 200 g · CMC 0,7 mg 500 g · CMC 2,0 mg 1 kg · CMC 4 mg 2 kg · CMC 7 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | Internal procedure IW/L1/S1/01 based on OIML R111-1:2004 Annex C | |
| 9 | Mass standards and weights of accuracy class M1 | Calibration — Mass (weights and mass standards) 1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,1 mg 50 mg · CMC 0,1 mg 100 mg · CMC 0,1 mg 200 mg · CMC 0,1 mg 500 mg · CMC 0,2 mg 1g · CMC 0,2 mg 2g · CMC 0,3 mg 5g · CMC 0,4 mg 10 g · CMC 0,5 mg 20 g · CMC 0,6 mg 50 g · CMC 0,7 mg 100 g · CMC 1,2 mg 200 g · CMC 2 mg 500 g · CMC 6 mg 1 kg · CMC 12 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | Internal procedure IW/L1/S1/01 based on OIML R111-1:2004 Annex C | |
| 10 | Bourdon tube pressure gauges Electronic pressure gauges Pressure transmitters |
Calibration — Pressure - gauge pressure – negative and positive pressure (gas) — -0,095 MPa to 0,1 MPa · CMC 6 · 10⁻⁵ MPa - gauge pressure – negative and positive pressure (gas) — 0,1 MPa to 0,6 MPa · CMC 3 · 10⁻⁴ MPa - gauge pressure – negative and positive pressure (gas) — 0,6 MPa to 6 MPa · CMC 5 · 10⁻⁴ · P Pressure |
Internal procedure; At permanent location (S) | Internal procedure IW/L1/S4/01 | |
| 11 | Bourdon tube pressure gauges Electronic pressure gauges Pressure transmitters |
Calibration — Pressure - gauge pressure – positive pressure (liquid - oil) — 0 MPa to 0,4 MPa · CMC 3,2 · 10⁻⁴ MPa - gauge pressure – positive pressure (liquid - oil) — 0,4 MPa to 0,6 MPa · CMC 8 · 10⁻⁴ · P - gauge pressure – positive pressure (liquid - oil) — 0,6 MPa to 7 MPa · CMC 2,2 · 10⁻³ MPa - gauge pressure – positive pressure (liquid - oil) — 7 MPa to 70 MPa · CMC 3,2 · 10⁻⁴ · P Pressure |
Internal procedure; At permanent location (S) | Internal procedure IW/L1/S4/01 | |
| 12 | Bourdon tube pressure gauges Electronic pressure gauges Pressure transmitters |
Calibration — Pressure - gauge pressure – positive pressure (liquid - water) — 0 MPa to 7 MPa · CMC 4,2 · 10⁻³ MPa - gauge pressure – positive pressure (liquid - water) — 7 MPa to 40 MPa · CMC 6 · 10⁻⁴ · P Pressure |
Internal procedure; At permanent location (S) | Internal procedure IW/L1/S4/01 | |
| 13 | Gauge blocks (classes 0, 1, 2) | Calibration — Length 0,5 mm to 100 mm · CMC Q[53; 0,96 · L] nm Length |
Internal procedure; At permanent location (S) | Internal procedure IW/L2/S12/01 based on PN-EN ISO 3650:2000 | |
| 14 | Gauge blocks (classes 0, 1, 2) | Calibration — Length 125 mm to 500 mm · CMC Q[117; 1,42 · L] nm Length |
Internal procedure; At permanent location (S) | Internal procedure IW/L2/S2/05 based on PN-EN ISO 3650:2000 | |
| 15 | Vernier calipers | Calibration — Length 0 mm to 300 mm · CMC 28 µm 300 mm to 800 mm · CMC 44 µm 800 mm to 1000 mm · CMC 54 µm Length |
Internal procedure; At permanent location (S) | Internal procedure IW/L2/S5/02 based on PN-EN ISO 13385-1:2019-12 | |
| 16 | Vernier depth gauges | Calibration — Length 0 mm to 300 mm · CMC 28 µm 300 mm to 630 mm · CMC 44 µm Length |
Internal procedure; At permanent location (S) | Internal procedure IW/L2/S5/02 based on PN-EN ISO 13385-2:2020-10 | |
| 17 | Vernier height gauges | Calibration — Length 0 mm to 300 mm · CMC 28 µm 300 mm to 630 mm · CMC 44 µm Length |
Internal procedure; At permanent location (S) | Internal procedure IW/L2/S5/02 based on PN-EN ISO 13225:2012 | |
| 18 | Special vernier instruments | Calibration — Length 0 mm to 300 mm · CMC 28 µm 300 mm to 800 mm · CMC 44 µm 800 mm to 1000 mm · CMC 54 µm Length |
Internal procedure; At permanent location (S) | Internal procedure IW/L2/S5/02 | |
| 19 | External micrometers | Calibration — Length 0 mm to 50 mm · CMC 2 µm 50 mm do 100 mm · CMC 3 µm 100 mm do 150 mm · CMC 4 µm 150 mm do 200 mm · CMC 5 µm Length |
Internal procedure; At permanent location (S) | Internal procedure IW/L2/S5/14 based on PN-EN ISO 3611:2023-10 | |
| 20 | Rotameters | Calibration — Volume flow rate (flow - gases) 1 dm³/h to 30 dm³/h · CMC (0,0007 x Q2 - 0,01 x Q + 0,85) % 30 dm³/h to 15000 dm³/h · CMC 0,38 % Volume flow rate (flow - gases) |
Internal procedure; At permanent location (S) | Internal procedure IW/L2/S14/03 | |
| 21 | Digital voltmeters Multimeters Clamp meters |
Calibration — DC voltage 0,02 mV to 2 mV · CMC 5,8 µV 2 mV to 10 mV · CMC 3,5 · 10⁻⁵ · U + 5,8 µV 10 mV to 200 mV · CMC 8,1 · 10⁻⁶ · U + 0,12 µV 0,2 V to 2 V · CMC 6,9 · 10⁻⁶ · U + 0,46 µV 2 V to 20 V · CMC 6,9 · 10⁻⁶ · U + 2,3 µV 20 V to 1000 V · CMC 1,2 · 10⁻⁵ · U + 0,23 mV DC voltage |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 based on Euramet cg-15 v. 3.0 02/2015 | |
| 22 | Calibrators | Calibration — DC voltage 10 mV to 200 mV · CMC 8,1 · 10⁻⁶ · U + 0,12 µV 0,2 V to 2 V · CMC 6,9 · 10⁻⁶ · U + 0,46 µV 2 V to 20 V · CMC 6,9 · 10⁻⁶ · U + 2,3 µV 20 V to 1000 V · CMC 1,2 · 10⁻⁵ · U + 0,23 mV DC voltage |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 | |
| 23 | Digital ammeters Multimeters Clamp meters |
Calibration — DC current 10 µA to 200 mA · CMC 1,2 · 10⁻⁴ · I + 1,2 µA 0,2 A to 2 A · CMC 2,3 · 10⁻⁴ · I + 23 µA 2 A to 20 A · CMC 5,8 · 10⁻⁴ · I + 4,6 mA DC current |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 based on Euramet cg-15 v. 3.0 02/2015 | |
| 24 | Calibrators | Calibration — DC current 10 µA to 200 mA · CMC 1,2 · 10⁻⁴ · I + 1,2 µA 0,2 A to 2 A · CMC 2,3 · 10⁻⁴ · I + 23 µA 2 A to 10 A · CMC 4,6 · 10⁻⁴ · I + 0,23 mA DC current |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 | |
| 25 | Digital voltmeters Power quality meters Distortion meters Multimeters Clamp meters |
Calibration — AC voltage 20 Hz to 40 Hz — 0,02 mV to 2 mV · CMC 12 µV 20 Hz to 40 Hz — 2 mV to 10 mV · CMC 3,5 · 10⁻⁴ · U + 12 µV 20 Hz to 40 Hz — 10 mV to 200 mV · CMC 2,5 · 10⁻⁴ · U + 4,6 µV 20 Hz to 40 Hz — 0,2 V to 20 V · CMC 2,0 · 10⁻⁴ · U + 0,23 mV 40 Hz to 100 Hz — 0,02 mV to 2 mV · CMC 12 µV 40 Hz to 100 Hz — 2 mV to 10 mV · CMC 3,5 · 10⁻⁴ · U + 12 µV 40 Hz to 100 Hz — 10 mV to 200 mV · CMC 2,3 · 10⁻⁴ · U + 4,6 µV 40 Hz to 100 Hz — 0,2 V to 1000 V · CMC 1,7 · 10⁻⁴ · U + 12 mV 100 Hz to 2 kHz — 0,02 mV to 2 mV · CMC 12 µV 100 Hz to 2 kHz — 2 mV to 10 mV · CMC 3,5 · 10⁻⁴ · U + 12 µV 100 Hz to 2 kHz — 10 mV to 200 mV · CMC 2,3 · 10⁻⁴ · U + 2,3 µV 100 Hz to 2 kHz — 0,2 V to 200 V · CMC 1,5 · 10⁻⁴ · U + 2,3 mV 100 Hz to 2 kHz — 200 V to 1000 V · CMC 1,7 · 10⁻⁴ · U + 12 mV 2 kHz to 5 kHz — 0,02 mV to 2 mV · CMC 12 µV 2 kHz to 5 kHz — 2 mV to 10 mV · CMC 3,5 · 10⁻⁴ · U + 12 µV 2 kHz to 5 kHz — 10 mV to 200 mV · CMC 2,3 · 10⁻⁴ · U + 4,6 µV 2 kHz to 5 kHz — 0,2 V to 200 V · CMC 1,7 · 10⁻⁴ · U + 2,3 mV 2 kHz to 5 kHz — 200 V to 1000 V · CMC 1,7 · 10⁻⁴ · U + 12 mV 5 kHz to 10 kHz — 0,02 mV to 2 mV · CMC 12 µV 5 kHz to 10 kHz — 2 mV to 10 mV · CMC 3,5 · 10⁻⁴ · U + 12 µV 5 kHz to 10 kHz — 10 mV to 200 mV · CMC 2,3 · 10⁻⁴ · U + 4,6 µV 5 kHz to 10 kHz — 0,2 V to 200 V · CMC 1,7 · 10⁻⁴ · U + 12 mV 10 kHz to 20 kHz — 0,02 mV to 2 mV · CMC 12 µV 10 kHz to 20 kHz — 2 mV to 10 mV · CMC 3,5 · 10⁻⁴ · U + 12 µV 10 kHz to 20 kHz — 10 mV to 200 mV · CMC 4,6 · 10⁻⁴ · U + 9,2 µV 10 kHz to 20 kHz — 0,2 V to 200 V · CMC 2,9 · 10⁻⁴ · U + 4,6 mV 20 kHz to 30 kHz — 0,02 mV to 2 mV · CMC 28 µV 20 kHz to 30 kHz — 2 mV to 10 mV · CMC 2,3 · 10⁻³ · U + 24 µV 20 kHz to 30 kHz — 10 mV to 200 mV · CMC 4,6 · 10⁻⁴ · U + 9,2 µV 20 kHz to 30 kHz — 0,2 V to 200 V · CMC 2,9 · 10⁻⁴ · U + 4,6 mV 30 kHz to 50 kHz — 0,02 mV to 2 mV · CMC 28 µV 30 kHz to 50 kHz — 2 mV to 10 mV · CMC 2,3 · 10⁻³ · U + 24 µV 30 kHz to 50 kHz — 10 mV to 200 mV · CMC 8,1 · 10⁻⁴ · U + 23 µV 30 kHz to 50 kHz — 0,2 V to 200 V · CMC 5,8 · 10⁻⁴ · U + 23 mV AC voltage |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 based on Euramet cg-15 v. 3.0 02/2015 | |
| 26 | Distortion meters Power quality meters |
Calibration — AC voltage 50 Hz; - total harmonic distortion (THD), up to the 25th harmonic — 0 % to 20 % · CMC 0,1 % AC voltage |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 | |
| 27 | Calibrators | Calibration — AC voltage 10 Hz to 40 Hz — 10 mV to 200 mV · CMC 2,5 · 10⁻⁴ · U + 4,6 µV 10 Hz to 40 Hz — 0,2 V to 200 V · CMC 2,0 · 10⁻⁴ · U + 2,3 mV 10 Hz to 40 Hz — 200 V to 1000 V · CMC 2,0 · 10⁻⁴ · U + 12 mV 40 Hz to 100 Hz — 10 mV to 200 mV · CMC 2,3 · 10⁻⁴ · U + 4,6 µV 40 Hz to 100 Hz — 0,2 V to 200 V · CMC 1,7 · 10⁻⁴ · U + 2,3 mV 40 Hz to 100 Hz — 200 V to 1000 V · CMC 2,0 · 10⁻⁴ · U + 12 mV 100 Hz to 2 kHz — 10 mV to 200 mV · CMC 2,3 · 10⁻⁴ · U + 2,3 µV 100 Hz to 2 kHz — 0,2 V to 200 V · CMC 1,5 · 10⁻⁴ · U + 2,3 mV 100 Hz to 2 kHz — 200 V to 1000 V · CMC 2,0 · 10⁻⁴ · U + 12 mV 2 kHz to 10 kHz — 10 mV to 200 mV · CMC 2,3 · 10⁻⁴ · U + 4,6 µV 2 kHz to 10 kHz — 0,2 V to 200 V · CMC 1,7 · 10⁻⁴ · U + 2,3 mV 2 kHz to 10 kHz — 200 V to 1000 V · CMC 2,0 · 10⁻⁴ · U + 12 mV 10 kHz to 30 kHz — 10 mV to 200 mV · CMC 4,6 · 10⁻⁴ · U + 9,2 µV 10 kHz to 30 kHz — 0,2 V to 200 V · CMC 2,9 · 10⁻⁴ · U + 4,6 mV 10 kHz to 30 kHz — 200 V to 1000 V · CMC 2,0 · 10⁻⁴ · U + 12 mV 30 kHz to 50 kHz — 10 mV to 200 mV · CMC 8,1 · 10⁻⁴ · U + 23 µV 30 kHz to 50 kHz — 0,2 V to 200 V · CMC 5,8 · 10⁻⁴ · U + 23 mV 30 kHz to 50 kHz — 200 V to 1000 V · CMC 2,0 · 10⁻⁴ · U + 12 mV AC voltage |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 | |
| 28 | Digital ammeters Power quality meters Distortion meters Multimeters Clamp meters |
Calibration — AC current 20 Hz to 40 Hz — 0,02 mA to 200 mA · CMC 3,5 · 10⁻⁴ · I + 23 µA 20 Hz to 40 Hz — 0,2 A to 2 A · CMC 5,8 · 10⁻⁴ · I + 0,23 mA 40 Hz to 500 Hz — 10 µA to 200 mA · CMC 3,5 · 10⁻⁴ · I + 23 µA 40 Hz to 500 Hz — 0,2 A to 2 A · CMC 5,8 · 10⁻⁴ · I + 0,23 mA 40 Hz to 500 Hz — 2 A to 150 A · CMC 1,2 · 10⁻³ · I + 69 mA 500 Hz to 1 kHz — 10 µA to 200 mA · CMC 3,5 · 10⁻⁴ · I + 23 µA 500 Hz to 1 kHz — 0,2 A to 2 A · CMC 5,8 · 10⁻⁴ · I + 0,23 mA 500 Hz to 1 kHz — 2 A to 10 A · CMC 9,2 · 10⁻⁴ · I + 3,5 mA 500 Hz to 1 kHz — 10 A to 20 A · CMC 9,2 · 10⁻⁴ · I + 6,9 mA 1 kHz to 5 kHz — 10 µA to 200 mA · CMC 3,5 · 10⁻⁴ · I + 23 µA 1 kHz to 5 kHz — 0,2 A to 2 A · CMC 5,8 · 10⁻⁴ · I + 0,23 mA AC current |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 based on Euramet cg-15 v. 3.0 02/2015 | |
| 29 | Distortion meters Power quality meters |
Calibration — AC current 50 Hz; - total harmonic distortion (THD), up to the 25th harmonic — 0 % to 20 % · CMC 0,2 % AC current |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 | |
| 30 | Calibrators | Calibration — AC current 10 Hz to 1 kHz — 10 µA to 200 mA · CMC 3,5 · 10⁻⁴ · I + 23 µA 10 Hz to 1 kHz — 0,2 A to 2 A · CMC 6,9 · 10⁻⁴ · I + 0,46 mA 10 Hz to 1 kHz — 2 A to 10 A · CMC 9,5 · 10⁻⁴ · I + 1,2 mA 1 kHz to 5 kHz — 10 µA to 200 mA · CMC 3,5 · 10⁻⁴ · I + 23 µA 1 kHz to 5 kHz — 0,2 A to 10 A · CMC 2,9 · 10⁻³ · I + 1,2 mA AC current |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 | |
| 31 | Digital ohmmeters Multimeters |
Calibration — DC resistance 0,1 mΩ · CMC 0,7 % 1 mΩ · CMC 0,6 % 0,01 Ω to 0,1 Ω · CMC 0,6 % 0,1 Ω to 1 Ω · CMC 0,058 % 1 Ω to 10 Ω · CMC 0,0041 % 10 Ω to 100 Ω · CMC 0,0020 % 0,1 kΩ to 1000 kΩ · CMC 0,0012 % 1 MΩ to 2 MΩ · CMC 0,0018 % 2 MΩ to 4 MΩ · CMC 0,029 % 4 MΩ to 20 MΩ · CMC 0,041 % 20 MΩ to 100 MΩ · CMC 0,065 % 100 MΩ to 1000 MΩ · CMC 0,058 % 1 GΩ to 10 GΩ · CMC 0,082 % 10 GΩ to 20 GΩ · CMC 0,081 % 20 GΩ to 70 GΩ · CMC 0,15 % 70 GΩ to 111 GΩ · CMC 0,20 % DC resistance |
Internal procedure; Direct method; At permanent location (S) | Internal procedure IW/L3/S11/01 based on Euramet cg-15 v. 3.0 02/2015 | |
| 32 | Resistance calibrators Fixed resistors Variable resistors |
Calibration — DC resistance 0,01 Ω to 0,1 Ω · CMC 0,33 % 0,1 Ω to 1 Ω · CMC 0,034 % 1 Ω to 19 Ω · CMC 0,0019 % 19 Ω to 190 Ω · CMC 0,0013 % 190 Ω to 190 kΩ · CMC 0,0011 % 190 kΩ to 1,9 MΩ · CMC 0,0017 % 1,9 MΩ to 19 MΩ · CMC 0,0040 % 19 MΩ to 190 MΩ · CMC 0,040 % 190 MΩ to 1,9 GΩ · CMC 0,41 % DC resistance |
Internal procedure; Direct method; At permanent location (S) | Internal procedures IW/L3/S11/01 IW/L3/S13/01 | |
| 33 | Mechanical stopwatches | Calibration — Time (time interval) 60 min · CMC (0,06 + 1,2·10⁻⁵ · τ) s1) Time (time interval) |
Internal procedure; At permanent location (S) | Internal procedure IW/L3/S14/02 | |
| 34 | Electronic stopwatches | Calibration — Time (time interval) 24 hours · CMC (0,003 + 2,8·10⁻⁷ · τ) s1) Time (time interval) |
Internal procedure; At permanent location (S) | Internal procedure IW/L3/S14/01 | |
| 35 | Electrically controlled electronic timers | Calibration — Time (time interval) 0,001 s to 10000 s · CMC (0,0006 + 5·10⁻⁷ · τ) s1) Time (time interval) |
Internal procedure; At permanent location (S) | Internal procedure IW/L3/S14/01 | |
| 36 | Spectrophotometers | Calibration — Optical density of spectral transmittance - optical density of spectral transmittance (absorbance); in the spectral range 400 nm to 880 nm — 0,0000 to 1,6500 · CMC calculated by formula 2): 0,0033 for D = 0,1000; 0,0041 for D = 0,5000; 0,0065 for D = 1,2000; 0,0097 for D = 1,6000 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC Blank 0 mg/l: 0,0404 · CMC 0,0035 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC Blank 0 mg/l: 0,0436 · CMC 0,0035 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC Blank 0 mg/l: 0,0525 · CMC 0,0036 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC Blank 0 mg/l: 0,0607 · CMC 0,0036 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 20 mg/l: 0,2495 · CMC 0,0041 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 20 mg/l: 0,1395 · CMC 0,0039 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 20 mg/l: 0,3359 · CMC 0,0040 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 20 mg/l: 0,3086 · CMC 0,0040 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC40 mg/l: 0,4580 · CMC 0,0048 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC40 mg/l: 0,2307 · CMC 0,0044 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC40 mg/l: 0,6148 · CMC 0,0048 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC40 mg/l: 0,5447 · CMC 0,0047 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 60 mg/l: 0,6743 · CMC 0,0056 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 60 mg/l: 0,3283 · CMC 0,0050 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 60 mg/l: 0,9058 · CMC 0,0057 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 60 mg/l: 0,7933 · CMC 0,0055 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 80 mg/l: 0,8884 · CMC 0,0066 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 80 mg/l: 0,4260 · CMC 0,0059 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 80 mg/l: 1,1970 · CMC 0,0067 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 80 mg/l: 1,0454 · CMC 0,0068 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 100 mg/l: 1,1023 · CMC 0,0085 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 100 mg/l: 0,5220 · CMC 0,0078 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 100 mg/l: 1,4901 · CMC 0,0091 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 100 mg/l: 1,2963 · CMC 0,0088 Optical density of spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure IW/L3/S3/01 | |
| 37 | Spectrophotometers | Calibration — Spectral transmittance - spectral transmittance (transmittance); in the spectral range 400 nm to 880 nm — 2,20 % to 80,00 % · CMC calculated by formula 3): 0,11 % for τ = 10,00 %; 0,36 % for τ = 50,00 %; 0,51 % for τ = 80,00 % Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC Blank 0 mg/l: 0,9112 · CMC 0,0059 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC Blank 0 mg/l: 0,9045 · CMC 0,0059 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC Blank 0 mg/l: 0,8862 · CMC 0,0059 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC Blank 0 mg/l: 0,8696 · CMC 0,0059 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 20 mg/l: 0,5630 · CMC 0,0051 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 20 mg/l: 0,7253 · CMC 0,0050 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 20 mg/l: 0,4614 · CMC 0,0051 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 20 mg/l: 0,4914 · CMC 0,0051 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 40 mg/l: 0,3483 · CMC 0,0049 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 40 mg/l: 0,5879 · CMC 0,0045 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 40 mg/l: 0,2428 · CMC 0,0049 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 40 mg/l: 0,2853 · CMC 0,0049 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 60 mg/l: 0,2117 · CMC 0,0045 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 60 mg/l: 0,4695 · CMC 0,0040 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 60 mg/l: 0,1242 · CMC 0,0046 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 60 mg/l: 0,1610 · CMC 0,0046 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 80 mg/l: 0,1293 · CMC 0,0042 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 80 mg/l: 0,3750 · CMC 0,0042 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 80 mg/l: 0,0635 · CMC 0,0042 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 80 mg/l: 0,0901 · CMC 0,0042 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 100 mg/l: 0,0790 · CMC 0,0040 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 100 mg/l: 0,3006 · CMC 0,0039 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 100 mg/l: 0,0324 · CMC 0,0040 Liquid filters; wavelengths: 350 nm, 313 nm, 257 nm, 235 nm; values given are nominal values — PDC 100 mg/l: 0,0505 · CMC 0,0040 Spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure IW/L3/S3/01 | |
| 38 | Spectrophotometers | Calibration — Spectral transmittance - wavelength — 270,00 nm to 820,00 nm · CMC 0,17 nm - wavelength — 820,01 nm to 900,00 nm · CMC 0,51 nm Spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure IW/L3/S3/01 | |
| 39 | Luxmeters | Calibration — Illuminance 10 lx to 10000 lx · CMC 2,1 % Illuminance |
Internal procedure; At permanent location (S) | Internal procedure IW/L3/S1/01 | |
| 40 | Photometric calibrators | Calibration — Illuminance 85 lx to 115 lx · CMC 3,2 % Illuminance |
Internal procedure; At permanent location (S) | Internal procedure IW/L3/S8/01 | |
Reference standards (32)
EURAMET cg-18 v.4.0
Euramet cg-15 v. 3.0 02/2015
IW/L1/S1/01
IW/L1/S2/01
IW/L1/S4/01
IW/L1/S6/01
IW/L1/S6/02
IW/L1/S6/03
IW/L1/S7/01
IW/L1/S9/01
IW/L2/S12/01
IW/L2/S14/03
IW/L2/S2/05
IW/L2/S5/02
IW/L2/S5/14
IW/L3/S1/01
IW/L3/S11/01
IW/L3/S13/01
IW/L3/S14/01
IW/L3/S14/02
IW/L3/S3/01
IW/L3/S8/01
OIML R111-1:2004 Załącznik C
PN-EN ISO 13225:2012
PN-EN ISO 13385-1:2019-12
PN-EN ISO 13385-2:2020-10
PN-EN ISO 3611:2023-10
PN-EN ISO 3650:2000
PN-EN ISO 6506-2:2019-10
PN-EN ISO 6507-2:2018-05
PN-EN ISO 6508-2:2015-04
PN-EN ISO 7500-1:2018-05