PCA AP 083 • Wrocław, Poland • 56 calibration items

Metrology & calibration

Okręgowy Urząd Miar we Wrocławiu is a calibration laboratory accredited by PCA under number AP 083 (Wrocław, Poland). Its accreditation scope covers 56 calibration items in the following areas: Metrology & calibration. Calibrated items include: Calibrators, Multimeters, Sound level meters, Bridges. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Direct method; At permanent location (S), Internal procedure; Direct method using gauge blocks; At permanent….

At a glance

  • Accreditation number: PCA AP 083
  • Calibration items in scope: 56
  • Types of calibrated items: 59
  • Calibration principles used: 10
  • City: Wrocław, Poland
  • Valid until: 18-01-2030
Accreditation number
AP 083
Accreditation body
PCA
Address
ul. Młodych Techników 61/63, 53-647 Wrocław
City
Wrocław, Poland
Division
Calibration Laboratories Group
Valid until
18-01-2030

What this laboratory calibrates (59)

Calibration principles (10)

Certificate information

Issue date
10.11.2026
Issue no.
25
Certificate date
10.01.2020
Accreditation cycle
22.12.2025 — 18.01.2030
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (56)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. The value of the uncertainty covered by the CMC expressed as a percentage is the relative measurement uncertainty and refers to the percentage share of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Gauge blocks of class 0 Calibration — Length
0,5 mm to 100 mm · CMC Q[0,08; 1,2 · L] µm 125 mm to 500 mm · CMC Q[0,08; 1,4 · L] µm
Length
Internal procedure; Comparison method using a two-probe comparator; At permanent location (S) PW/L1/02 based on PN-EN ISO 3650:2000
2 Gauge blocks 1, 2 Calibration — Length
0,5 mm to 100 mm · CMC Q[0,10; 1,7 · L] µm 125 mm to 500 mm · CMC Q[0,10; 1,9 · L] µm
Length
Internal procedure; Comparison method using a two-probe comparator; At permanent location (S) PW/L1/02 based on PN-EN ISO 3650:2000
3 Calipers Calibration — Length
0 mm to 300 mm · CMC 0,04 mm 300 mm to 500 mm · CMC 0,05 mm
Length
Internal procedure; Direct method using gauge blocks; At permanent location (S) PW/L1/01 IW1-PW/L1/01
4 Caliper depth gauges Calibration — Length
0 mm to 500 mm · CMC 0,06 mm
Length
Internal procedure; Direct method using gauge blocks; At permanent location (S) PW/L1/01 IW1-PW/L1/01
5 Caliper height gauges Calibration — Length
0 mm to 500 mm · CMC 0,05 mm 500 mm to 800 mm · CMC 0,06 mm 800 mm to 1000 mm · CMC 0,07 mm
Length
Internal procedure; Direct method using gauge blocks; At permanent location (S) PW/L1/01 IW1-PW/L1/01
6 External micrometers Calibration — Length
0 mm do 75 mm · CMC 2 µm 75 mm to 100 mm · CMC 3 µm 100 mm do 200 mm · CMC 3 µm
Length
Internal procedure; Direct method using gauge blocks; At permanent location (S) PW/L1/01 IW2-PW/L1/01
7 Internal micrometers Calibration — Length
5 mm to 30 mm · CMC 2 µm 30 mm to 55 mm · CMC 2 µm 50 mm to 100 mm · CMC 3 µm 100 mm to 150 mm · CMC 3 µm
Length
Internal procedure; Direct method using gauge blocks; At permanent location (S) PW/L1/01 IW2-PW/L1/01
8 Micrometers with built-in dial indicator Calibration — Length
0 mm do 75 mm · CMC 2 µm 75 mm to 100 mm · CMC 2 µm
Length
Internal procedure; Direct method using gauge blocks; At permanent location (S) PW/L1/01 IW2-PW/L1/01
9 Micrometer depth gauges Calibration — Length
0 mm to 75 mm · CMC 2 µm 75 mm to 150 mm · CMC 3 µm
Length
Internal procedure; Direct method using gauge blocks; At permanent location (S) PW/L1/01 IW2-PW/L1/01
10 Indicating snap gauges Calibration — Length
0 mm to 100 mm · CMC 2 µm
Length
Internal procedure; Direct method using gauge blocks; At permanent location (S) PW/L1/01 IW2-PW/L1/01
11 Analogue dial indicators with a scale interval of 0,01 mm Calibration — Length
0 mm to 50 mm · CMC 6 µm
Length
Internal procedure; Direct method using a micrometer head; At permanent location (S) PW/L1/01 IW3-PW/L1/01
12 Digital indicators with a resolution of 0,01 mm Calibration — Length
0 mm to 50 mm · CMC 10 µm
Length
Internal procedure; Direct method using a micrometer head; At permanent location (S) PW/L1/01 IW3-PW/L1/01
13 Digital indicators with a resolution of 0,001 mm Calibration — Length
0 mm to 50 mm · CMC 3 µm
Length
Internal procedure; Direct method using a micrometer head; At permanent location (S) PW/L1/01 IW3-PW/L1/01
14 Dial bore gauges with a scale interval of 0,01 mm Calibration — Length
4 mm to 50 mm · CMC 7 µm 50 mm to 160 mm · CMC 7 µm
Length
Internal procedure; Direct method using a micrometer head; At permanent location (S) PW/L1/01 IW3-PW/L1/01
15 Dial bore gauges with a scale interval of 0,001 mm Calibration — Length
18 mm to 35 mm · CMC 2 µm 35 mm to 60 mm · CMC 2 µm 50 mm to 150 mm · CMC 3 µm
Length
Internal procedure; Direct method using a micrometer head; At permanent location (S) PW/L1/01 IW3-PW/L1/01
16 Feeler gauges Calibration — Length
0,03 mm to 1,00 mm · CMC 0,8 µm
Length
Internal procedure; Comparison method using gauge blocks and an optimeter; At permanent location (S) PW/L1/07
17 Rigid rules
Semi-rigid rules
Calibration — Length
0 m to 1 m · CMC 0,08 mm
Length
Internal procedure; Comparison method using a rigid or semi-rigid rule; At permanent location (S) PW/L1/04
18 Tape measures
Rigid rules
Semi-rigid rules
Calibration — Length
0 m to 5 m · CMC (0,08 + 0,01 · L) mm
Length
Internal procedure; Comparison method using a rigid or semi-rigid rule; At permanent location (S) PW/L1/04
19 Static testing machines for tensile and compressive forces
Technological devices for tensile and compressive forces
Calibration — Force
0,01 N to 2000 kN · CMC 0,25 %
Force
Internal procedure; Using class 1 reference force transducers; Outside of permanent location (P) PW/L1/15 based on PN-EN ISO 7500-1:2018-05
20 Rockwell hardness testers Calibration — Hardness
hardness — 45 HRA to 88 HRA · CMC 0,6 HRA hardness — 90 HRB to 100 HRB · CMC 0,6 HRB hardness — 20 HRC to 70 HRC · CMC 0,6 HRC force — 98 N to 1471 N · CMC 0,28 %
Hardness
Internal procedure; At permanent location (S) and outside of it (P) PW/L1/16 based on PN-EN ISO 6508-2:2024-06
21 Non-automatic electronic weighing instruments Calibration — Mass (weighing instruments)
0 g to 220 g · CMC 2,3·10⁻⁴ % 220 g to 12 kg · CMC 8,5·10⁻⁴ % 12 kg to 300 kg · CMC 6·10⁻³ %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) PW/L1/12 based on EURAMET cg-18 v.4.0
22 Mass standards of accuracy class E2 Calibration — Mass (weights and mass standards)
1g · CMC 0,010 mg 2g · CMC 0,013 mg 5g · CMC 0,016 mg 10 g · CMC 0,020 mg 20 g · CMC 0,026 mg 50 g · CMC 0,03 mg 100 g · CMC 0,05 mg 200 g · CMC 0,10 mg 500 g · CMC 0,25 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PW/L1/14 based on OIML R-111-1:2004 Annex C
23 Mass standards of accuracy class F1
Weights of accuracy class F1
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,006 mg 2 mg · CMC 0,006 mg 5 mg · CMC 0,006 mg 10 mg · CMC 0,008 mg 20 mg · CMC 0,010 mg 50 mg · CMC 0,013 mg 100 mg · CMC 0,016 mg 200 mg · CMC 0,020 mg 500 mg · CMC 0,026 mg 1g · CMC 0,033 mg 2g · CMC 0,040 mg 5g · CMC 0,053 mg 10 g · CMC 0,06 mg 20 g · CMC 0,08 mg 50 g · CMC 0,10mg 100 g · CMC 0,16 mg 200 g · CMC 0,33 mg 500 g · CMC 0,8 mg 1 kg · CMC 1,6 mg 2 kg · CMC 3 mg 5 kg · CMC 8 mg 10 kg · CMC 16 mg 20 kg · CMC 33 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PW/L1/11 based on OIML R-111-1:2004 Annex C
24 Mass standards of accuracy class F2
Weights of accuracy class F2
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,020 mg 2 mg · CMC 0,020 mg 5 mg · CMC 0,020 mg 10 mg · CMC 0,026 mg 20 mg · CMC 0,033 mg 50 mg · CMC 0,040 mg 100 mg · CMC 0,053 mg 200 mg · CMC 0,066 mg 500 mg · CMC 0,083 mg 1g · CMC 0,100 mg 2g · CMC 0,133 mg 5g · CMC 0,166 mg 10 g · CMC 0,200 mg 20 g · CMC 0,26 mg 50 g · CMC 0,33 mg 100 g · CMC 0,53 mg 200 g · CMC 1,0 mg 500 g · CMC 2,6 mg 1 kg · CMC 5,3 mg 2 kg · CMC 10 mg 5 kg · CMC 26 mg 10 kg · CMC 53 mg 20 kg · CMC 100 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PW/L1/11 based on OIML R-111-1:2004 Annex C
25 Mass standards of accuracy class M1
Weights of accuracy class M1
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,066 mg 2 mg · CMC 0,066 mg 5 mg · CMC 0,066 mg 10 mg · CMC 0,083 mg 20 mg · CMC 0,10 mg 50 mg · CMC 0,13 mg 100 mg · CMC 0,16 mg 200 mg · CMC 0,20 mg 500 mg · CMC 0,26 mg 1g · CMC 0,33 mg 2g · CMC 0,40 mg 5g · CMC 0,53 mg 10 g · CMC 0,66 mg 20 g · CMC 0,83 mg 50 g · CMC 1,0 mg 100 g · CMC 1,6 mg 200 g · CMC 3 mg 500 g · CMC 8 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PW/L1/11 based on OIML R-111-1:2004 Annex C
26 Mass standards of accuracy class M1
Weights of accuracy class M1
Calibration — Mass (weights and mass standards)
1 kg · CMC 16 mg 2 kg · CMC 33 mg 5 kg · CMC 83 mg 10 kg · CMC 166 mg 20 kg · CMC 333 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PW/L1/11 based on OIML R-111-1:2004 Annex C
27 25 kg mass standards Calibration — Mass (weights and mass standards)
25 kg · CMC 416 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PW/L1/11 based on OIML R-111-1:2004 Annex C
28 Load weights Calibration — Mass (weights and mass standards)
1 kg to 30 kg · CMC 1·10⁻⁴ %
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PW/L1/11
29 Grain density meters 1/4 L Calibration — Density (solids)
180 g to 220 g · CMC 0,7 g 72 kg/hL to 88 kg/hL · CMC 0,28 kg/hL
Density (solids)
Internal procedure; At permanent location (S) PW/L1/13
30 Grain density meters 1 L Calibration — Density (solids)
720 g to 880 g · CMC 2,0 g 72 kg/hL to 88 kg/hL · CMC 0,20 kg/hL
Density (solids)
Internal procedure; At permanent location (S) PW/L1/13
31 Spectrophotometers Calibration — Optical density of spectral transmittance
optical density of spectral transmittance; spectral range 400 nm to 800 nm — 0,10 to 0,20 · CMC 0,0042 optical density of spectral transmittance; spectral range 400 nm to 800 nm — 0,20 to 0,40 · CMC 0,0045 optical density of spectral transmittance; spectral range 400 nm to 800 nm — 0,40 to 0,66 · CMC 0,0054 optical density of spectral transmittance; spectral range 400 nm to 800 nm — 0,66 to 1,00 · CMC 0,0064 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC Blank 0 mg/l: 0,0365 · CMC 0,0044 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC Blank 0 mg/l: 0,0387 · CMC 0,0044 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC Blank 0 mg/l: 0,0444 · CMC 0,0045 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC Blank 0 mg/l: 0,0500 · CMC 0,0045 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 20 mg/l: 0,2482 · CMC 0,0050 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 20 mg/l: 0,1359 · CMC 0,0048 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 20 mg/l: 0,3288 · CMC 0,0052 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 20 mg/l: 0,2963 · CMC 0,0051 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 40 mg/l: 0,4555 · CMC 0,0057 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 40 mg/l: 0,2273 · CMC 0,0053 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 40 mg/l: 0,6082 · CMC 0,0060 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 40 mg/l: 0,5360 · CMC 0,0058 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 60 mg/l: 0,6710 · CMC 0,0064 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 60 mg/l: 0,3245 · CMC 0,0058 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 60 mg/l: 0,8997 · CMC 0,0069 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 60 mg/l: 0,7861 · CMC 0,0067 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 80 mg/l: 0,8852 · CMC 0,0073 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 80 mg/l: 0,4219 · CMC 0,0066 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 80 mg/l: 1,1899 · CMC 0,0079 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 80 mg/l: 1,0366 · CMC 0,0076 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 100 mg/l: 1,0970 · CMC 0,0092 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 100 mg/l: 0,5169 · CMC 0,0083 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 100 mg/l: 1,4815 · CMC 0,0098 Liquid filters; wavelengths: 235 nm, 257 nm, 313 nm, 350 nm — PDC 100 mg/l: 1,2864 · CMC 0,0095 wavelength — 400 nm to 800 nm · CMC 0,18 nm
Optical density of spectral transmittance
Internal procedure; At permanent location (S) and outside of it (P) PW/L1/10
32 Bourdon tube pressure gauges
Electronic pressure gauges
- gauge pressure - negative and positive pressure
Calibration — Pressure
-0,1 MPa to 0,1 MPa · CMC 2·10⁻⁵ MPa 0,1 MPa to 0,4 MPa · CMC 1·10⁻⁴ MPa 0,4 MPa to 1 MPa · CMC 2·10⁻⁴ MPa 1 MPa to 2,5 MPa · CMC 6·10⁻⁴ MPa 2,5 MPa to 5 MPa · CMC 1·10⁻³ MPa 5 MPa to 6 MPa · CMC 2·10⁻³ MPa 6 MPa to 10 MPa · CMC 4·10⁻³ MPa 10 MPa to 22 MPa · CMC 1·10⁻² MPa 22 MPa to 50 MPa · CMC 2·10⁻² MPa 50 MPa to 60 MPa · CMC 3·10⁻² MPa
Pressure
Internal procedure; Direct method; At permanent location (S) PW/L2/01 IW1- PW/L2/01 IW2- PW/L2/01 IW5- PW/L2/01 IW7- PW/L2/01
33 Sound calibrators Calibration — Sound pressure (sound in air)
sound pressure level; re 20 µPa; nominal frequency: 1 kHz — 90 dB to 120 dB · CMC 0,10 dB
Sound pressure (sound in air)
Internal procedure; At permanent location (S) PW/L3/13 based on PN-EN 60942:2005
34 Sound level meters Calibration — Sound pressure (sound in air)
response of the sound level meter to the sound calibrator signal; re 20 µPa — 90 dB to 130 dB · CMC 0,2 dB
Sound pressure (sound in air)
Internal procedure; At permanent location (S) PW/L3/14 PW/L3/21 based on PN-EN 60651:2002 PN-EN 60804:2002 PN-EN 61672-1:2005 PN-EN 61672-3:2007
35 Sound level meters Calibration — Sound pressure (sound in air)
response of the sound level meter to electrical test signals; re 20 µPa, frequency range 20 Hz to 20 kHz — 0 dB to 140 dB · CMC 0,2 dB
Sound pressure (sound in air)
Internal procedure; At permanent location (S) PW/L3/14 PW/L3/21 based on PN-EN 60651:2002 PN-EN 60804:2002 PN-EN 61672-1:2005 PN-EN 61672-3:2007
36 Sound level meters Calibration — Sound pressure (sound in air)
free-field frequency response of the sound level meter; frequency range — 20 Hz to 1 kHz · CMC 0,2 dB free-field frequency response of the sound level meter; frequency range — 2 kHz to 8 kHz · CMC 0,3 dB free-field frequency response of the sound level meter; frequency range — for 16 kHz · CMC 0,5 dB free-field frequency response of the sound level meter; frequency range — for 20 kHz · CMC 0,6 dB
Sound pressure (sound in air)
Internal procedure; At permanent location (S) PW/L3/14 PW/L3/21 based on PN-EN 60651:2002 PN-EN 60804:2002
37 Sound level meters Calibration — Sound pressure (sound in air)
free-field frequency response of the sound level meter; frequencies — 125 Hz, 1 kHz · CMC 0,2 dB free-field frequency response of the sound level meter; frequencies — 4 kHz, 8 kHz · CMC 0,3 dB
Sound pressure (sound in air)
Internal procedure; At permanent location (S) PW/L3/14 PW/L3/21 based on PN-EN 61672-1:2005 PN-EN 61672-3:2007
38 Vibration calibrators
- vibration acceleration
Calibration — Vibration acceleration
RMS acceleration value; nominal frequency: 15,92 Hz — 1,0 ms⁻² · CMC 1,3 % RMS or peak acceleration value; nominal frequencies: 79,6 Hz, 159,2 Hz — 3,16 ms⁻² · CMC 1,2 % RMS or peak acceleration value; nominal frequencies: 79,6 Hz, 159,2 Hz — 7,07 ms⁻² · CMC 1,2 % RMS or peak acceleration value; nominal frequencies: 79,6 Hz, 159,2 Hz — 10 ms⁻² · CMC 1,2 %
Vibration acceleration
Internal procedure; At permanent location (S) PW/L3/04
39 pH meters Calibration — PH
pH — 0 to 14 · CMC 0,003 DC voltage — -2300 mV to 2300 mV · CMC 0,2 mV
PH
Internal procedure; Electrical method; At permanent location (S) PW/L3/10
40 Conductivity meters Calibration — Specific electrical conductivity (conductometry)
specific electrical conductivity — 0,0001 mS/cm to 2 mS/cm · CMC 0,2 % specific electrical conductivity — 2 mS/cm to 500 mS/cm · CMC 0,2 %
Specific electrical conductivity (conductometry)
Internal procedure; Electrical method; At permanent location (S) PW/L3/11
41 Multimeters
Digital voltmeters
Calibration — DC voltage
0,1 mV to 20 mV · CMC 0,033 % 20 mV to 200 mV · CMC 0,004 % 200 mV to 2 V · CMC 0,00089 % 2 V to 20 V · CMC 0,00039 % 20 V to 200 V · CMC 0,00064 % 200 V to 1000 V · CMC 0,00095 %
DC voltage
Internal procedure; Direct method; At permanent location (S) PW/L3/07
42 Calibrators Calibration — DC voltage
10 mV to 200 mV · CMC 0,00075 % 200 mV to 20 V · CMC 0,00046 % 20 V to 200 V · CMC 0,00069 % 200 V to 1000 V · CMC 0,00078 %
DC voltage
Internal procedure; Direct method; At permanent location (S) PW/L3/05
43 Multimeters
Digital ammeters
Calibration — DC current
10 µA to 200 µA · CMC 0,013 % 0,2 mA to 200 mA · CMC 0,0053 % 0,2 A to 2 A · CMC 0,013 % 2 A to 10 A · CMC 0,043 %
DC current
Internal procedure; Direct method; At permanent location (S) PW/L3/07
44 Calibrators Calibration — DC current
10 µA to 200 µA · CMC 0,0093 % 0,2 mA to 2 mA · CMC 0,034 % 2 mA to 20 mA · CMC 0,0036 % 20 mA to 200 mA · CMC 0,0066 % 0,2 A to 2 A · CMC 0,021 % 2 A to 10 A · CMC 0,057 %
DC current
Internal procedure; Direct method; At permanent location (S) PW/L3/05
45 Multimeters
Digital voltmeters
Calibration — AC voltage
10 Hz — 10 mV to 200 mV · CMC 0,039 % 10 Hz — 0,2 V to 2 V · CMC 0,032 % 10 Hz — 2 V to 20 V · CMC 0,033 % 10 Hz to 40 Hz — 10 mV to 200 mV · CMC 0,021 % 10 Hz to 40 Hz — 0,2 V to 2 V · CMC 0,015 % 10 Hz to 40 Hz — 2 V to 20 V · CMC 0,016 % 40 Hz to 300 Hz — 200 µV to 2 mV · CMC 0,43 % 40 Hz to 300 Hz — 2 mV to 20 mV · CMC 0,052 % 40 Hz to 300 Hz — 20 mV to 200 mV · CMC 0,016 % 40 Hz to 300 Hz — 200 mV to 2 V · CMC 0,009 % 40 Hz to 300 Hz — 2 V to 20 V · CMC 0,0073 % 40 Hz to 300 Hz — 20 V to 200 V · CMC 0,0085 % 40 Hz to 300 Hz — 200 V to 1000 V · CMC 0,019 % 300 Hz to 10 kHz — 200 µV to 2 mV · CMC 0,36 % 300 Hz to 10 kHz — 2 mV to 20 mV · CMC 0,045 % 300 Hz to 10 kHz — 20 mV to 200 mV · CMC 0,014 % 300 Hz to 10 kHz — 200 mV to 2 V · CMC 0,0077 % 300 Hz to 10 kHz — 2 V to 20 V · CMC 0,0055 % 300 Hz to 10 kHz — 20 V to 200 V · CMC 0,0066 % 300 Hz to 10 kHz — 200 V to 1000 V · CMC 0,015 % 10 kHz to 30 kHz — 200 µV to 2 mV · CMC 0,36 % 10 kHz to 30 kHz — 2 mV to 20 mV · CMC 0,046 % 10 kHz to 30 kHz — 20 mV to 200 mV · CMC 0,016 % 10 kHz to 30 kHz — 200 mV to 2 V · CMC 0,0077 % 10 kHz to 30 kHz — 2 V to 20 V · CMC 0,0055 % 10 kHz to 30 kHz — 20 V to 200 V · CMC 0,0084 % 10 kHz to 30 kHz — 200 V to 1000 V · CMC 0,020 % 30 kHz to 100 kHz — 200 µV to 2 mV · CMC 0,45 % 30 kHz to 100 kHz — 2 mV to 20 mV · CMC 0,072 % 30 kHz to 100 kHz — 20 mV to 200 mV · CMC 0,042 % 30 kHz to 100 kHz — 200 mV to 20 V · CMC 0,011 % 30 kHz to 100 kHz — 20 V to 200 V · CMC 0,016 % 30 kHz to 100 kHz — 200 V to 1000 V · CMC 0,026 % 100 kHz to 330 kHz — 200 µV to 2 mV · CMC 0,78 % 100 kHz to 330 kHz — 2 mV to 20 mV · CMC 0,19 % 100 kHz to 330 kHz — 20 mV to 200 mV · CMC 0,13 % 100 kHz to 330 kHz — 200 mV to 2 V · CMC 0,036 % 100 kHz to 330 kHz — 2 V to 20 V · CMC 0,035 % 100 kHz to 330 kHz — 20 V to 200 V · CMC 0,053 % 300 kHz to 1 MHz — 200 µV to 2 mV · CMC 0,0023 · U + 0,024 mV 300 kHz to 1 MHz — 2 mV to 20 mV · CMC 0,4 % 300 kHz to 1 MHz — 20 mV to 200 mV · CMC 0,3 % 300 kHz to 1 MHz — 200 mV to 20 V · CMC 0,2 %
AC voltage
Internal procedure; Direct method; At permanent location (S) PW/L3/07
46 Calibrators Calibration — AC voltage
for 10 Hz — 20 mV to 200 mV · CMC 0,038 % dla 10 Hz to 40 Hz — 20 mV to 200 mV · CMC 0,021 % 40 Hz to 100 Hz — 10 mV to 200 mV · CMC 0,018 % 40 Hz to 100 Hz — 200 mV to 200 V · CMC 0,011 % 40 Hz to 100 Hz — 200 V to 1000 V · CMC 0,015 % 100 Hz to 2000 Hz — 10 mV to 200 mV · CMC 0,017 % 100 Hz to 2000 Hz — 200 mV to 200 V · CMC 0,01 % 100 Hz to 2000 Hz — 200 V to 1000 V · CMC 0,015 % 2 kHz to 10 kHz — 10 mV to 200 mV · CMC 0,00014 · U + 0,004 mV 2 kHz to 10 kHz — 200 mV to 200 V · CMC 0,012 % 2 kHz to 10 kHz — 200 V to 1000 V · CMC 0,013 % 10 kHz to 30 kHz — 10 mV to 200 mV · CMC 0,00034 · U + 0,008 mV 10 kHz to 30 kHz — 200 mV to 200 V · CMC 0,024 % 10 kHz to 30 kHz — 200 V to 1000 V · CMC 0,026 % 3 kHz to 100 kHz — 20 mV to 200 mV · CMC 0,00077 · U + 0,02 mV 3 kHz to 100 kHz — 200 mV to 200 V · CMC 0,067 % 3 kHz to 100 kHz — 200 V to 1000 V · CMC 0,077 % 100 kHz to 300 kHz — 200 mV to 200 V · CMC 0,4 % 300 kHz to 1 MHz — 200 mV to 200 V · CMC 2,0 %
AC voltage
Internal procedure; Direct method; At permanent location (S) PW/L3/05
47 Multimeters
Digital ammeters
Calibration — AC current
10 Hz to 1 kHz — 10 µA to 200 µA · CMC 0,026 % 10 Hz to 1 kHz — 0,2 mA to 200 mA · CMC 0,019 % 10 Hz to 1 kHz — 0,2 A to 2 A · CMC 0,042 % 10 Hz to 1 kHz — 2 A to 10 A · CMC 0,061 % 1 kHz to 5 kHz — 10 µA to 200 µA · CMC 0,045 % 1 kHz to 5 kHz — 0,2 mA to 200 mA · CMC 0,030 % 1 kHz to 5 kHz — 0,2 A to 2 A · CMC 0,062 % 1 kHz to 5 kHz — 2 A to 10 A · CMC 0,12 %
AC current
Internal procedure; Direct method; At permanent location (S) PW/L3/07
48 Calibrators Calibration — AC current
50 Hz to 2 kHz — 10 µA to 200 µA · CMC 0,061 % 50 Hz to 2 kHz — 0,2 mA to 200 mA · CMC 0,041 % 50 Hz to 2 kHz — 0,2 A to 2 A · CMC 0,073 % 50 Hz to 2 kHz — 2 A to 10 A · CMC 0,1 % 2 kHz to 5 kHz — 10 µA to 200 µA · CMC 0,0003 · I + 0,02 µA 2 kHz to 5 kHz — 0,2 mA to 200 mA · CMC 0,039 % 2 kHz to 5 kHz — 0,2 A to 2 A · CMC 0,083 % 2 kHz to 5 kHz — 2 A to 10 A · CMC 0,26 %
AC current
Internal procedure; Direct method; At permanent location (S) PW/L3/05
49 Multimeters
Digital ohmmeters
Calibration — DC resistance
0,1 Ω to 1 Ω · CMC 0,014 % 1 Ω to 10 Ω · CMC 0,0030 % 10 Ω to 10 kΩ · CMC 0,0011 % 10 kΩ to 100 kΩ · CMC 0,0014 % 100 kΩ to 1 MΩ · CMC 0,0032 % 1 MΩ to 10 MΩ · CMC 0,0063 % 10 MΩ to 100 MΩ · CMC 0,022 %
DC resistance
Internal procedure; Direct method; At permanent location (S) PW/L3/07
50 Calibrators Calibration — DC resistance
1 Ω to 2 Ω · CMC 0,0021 % 2 Ω to 20 Ω · CMC 0,0014 % 20 Ω to 200 kΩ · CMC 0,0011 % 200 kΩ to 2 MΩ · CMC 0,0019 % 2 MΩ to 20 MΩ · CMC 0,0047 % 20 MΩ to 200 MΩ · CMC 0,075 % 100 MΩ · CMC 0,016 % 0,2 GΩ to 1 GΩ · CMC 0,55 %
DC resistance
Internal procedure; Direct method; At permanent location (S) PW/L3/05
51 Bridges
Digital ohmmeters
Calibration — AC resistance
1 kHz — 1 Ω to 100 kΩ · CMC 0,0003 · R + 2 mΩ
AC resistance
Internal procedure; Direct method; At permanent location (S) PW/L3/01
52 Variable standard inductors
Fixed standard inductors
Bridges
Inductance meters
Calibration — Inductance
1 kHz — 1 µH · CMC 5,2 % 1 kHz — 2 µH · CMC 2,5 % 1 kHz — 3 µH · CMC 2,0 % 1 kHz — 5 µH · CMC 1,6 % 1 kHz — 10 µH · CMC 0,52 % 1 kHz — 20 µH · CMC 0,26 % 1 kHz — 30 µH · CMC 0,22 % 1 kHz — 50 µH · CMC 0,17 % 1 kHz — 100 µH · CMC 0,11 % 1 kHz — 200 µH · CMC 0,10 % 1 kHz — 300 µH · CMC 0,08 % 1 kHz — 500 µH · CMC 0,06 % 1 kHz — 1 mH · CMC 0,05 % 1 kHz — 2 mH · CMC 0,05 % 1 kHz — 3 mH · CMC 0,05 % 1 kHz — 5 mH · CMC 0,04 % 1 kHz — 10 mH · CMC 0,04 % 1 kHz — 20 mH · CMC 0,04 % 1 kHz — 30 mH · CMC 0,04 % 1 kHz — 50 mH · CMC 0,04 % 1 kHz — 100 mH · CMC 0,04 % 1 kHz — 200 mH · CMC 0,04 % 1 kHz — 300 mH · CMC 0,04 % 1 kHz — 500 mH · CMC 0,04 % 1 kHz — 1H · CMC 0,04 % 1 kHz — 2H · CMC 0,12 % 1 kHz — 5H · CMC 0,12 % 1 kHz — 10 H · CMC 0,12 %
Inductance
Internal procedure; At permanent location (S) PW/L3/01 PW/L3/06
53 Variable standard capacitors
Fixed standard capacitors
Bridges
Capacitance meters
Calibration — Capacitance
1 kHz — 10 pF · CMC 0,007 % 1 kHz — 100 pF · CMC 0,007 % 1 kHz — 500 pF · CMC 0,007 % 1 kHz — 1000 pF · CMC 0,007 % 1 kHz — 5000 pF · CMC 0,007 % 1 kHz — 10000pF · CMC 0,007 % 1 kHz — 0,1 pF to 10 pF · CMC 0,03 % 1 kHz — 10 pF to 0,1 µF · CMC 0,012 % 1 kHz — 0,1 µF to 1 µF · CMC 0,012 % 1 kHz — 1 µF to 10 µF · CMC 0,03 % 1 kHz — 10 µF to 100 µF · CMC 0,05 %
Capacitance
Internal procedure; At permanent location (S) PW/L3/01 PW/L3/06
54 Period meters (frequency counters-timers) Calibration — Time (time interval)
0,1 µs to 1 s · CMC (3·10⁻⁵ Hz·T + 1,9·10⁻⁹) T
Time (time interval)
Internal procedure; Direct method; At permanent location (S) PW/L3/18
55 Quartz generators Calibration — Frequency
10 Hz to 225 MHz · CMC 3·10⁻⁵ Hz + 1,9·10⁻⁹ f
Frequency
Internal procedure; Direct method; At permanent location (S) PW/L3/19
56 Digital frequency meters Calibration — Frequency
1 Hz to 80 MHz · CMC 3·10⁻⁵ Hz + 1,9·10⁻⁹ f
Frequency
Internal procedure; Direct method; At permanent location (S) PW/L3/18

Reference standards (41)

EURAMET cg-18 v.4.0
IW1- PW/L2/01
IW1-PW/L1/01
IW2- PW/L2/01
IW2-PW/L1/01
IW3-PW/L1/01
IW5- PW/L2/01
IW7- PW/L2/01
OIML R-111-1:2004 Załącznik C
PN-EN 60651:2002
PN-EN 60804:2002
PN-EN 60942:2005
PN-EN 61672-1:2005
PN-EN 61672-3:2007
PN-EN ISO 3650:2000
PN-EN ISO 6508-2:2024-06
PN-EN ISO 7500-1:2018-05
PW/L1/01
PW/L1/02
PW/L1/04
PW/L1/07
PW/L1/10
PW/L1/11
PW/L1/12
PW/L1/13
PW/L1/14
PW/L1/15
PW/L1/16
PW/L2/01
PW/L3/01
PW/L3/04
PW/L3/05
PW/L3/06
PW/L3/07
PW/L3/10
PW/L3/11
PW/L3/13
PW/L3/14
PW/L3/18
PW/L3/19
PW/L3/21

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20083