🇵🇱 Okręgowy Urząd Miar w Krakowie PL

PCA AP 082 • Kraków, Poland • 117 calibration items

Metrology & calibration

Okręgowy Urząd Miar w Krakowie is a calibration laboratory accredited by PCA under number AP 082 (Kraków, Poland). Its accreditation scope covers 117 calibration items in the following areas: Metrology & calibration. Calibrated items include: Multimeters, Load weights, Mass standards of accuracy class F2, Non-automatic weighing instruments. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; Direct method; At permanent location (S).

At a glance

  • Accreditation number: PCA AP 082
  • Calibration items in scope: 117
  • Types of calibrated items: 87
  • Calibration principles used: 12
  • City: Kraków, Poland
  • Valid until: 18-01-2030
Accreditation number
AP 082
Accreditation body
PCA
Address
ul. Krupnicza 11, 31-123 Kraków
City
Kraków, Poland
Division
Calibration Laboratories Group
Valid until
18-01-2030

What this laboratory calibrates (87)

Calibration principles (12)

Certificate information

Issue date
13.01.2026
Issue no.
24
Certificate date
20.12.2019
Accreditation cycle
13.01.2026 — 18.01.2030
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (117)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. The value of the uncertainty covered by the CMC expressed as a percentage is the relative measurement uncertainty and refers to the percentage share of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Force gauges
Transducers
Calibration — Force
0,1 N to 100 N · CMC 0,05 % 100 N to 5000 N · CMC 0,05 % 1 kN to 2 kN · CMC 0,07 % 3 kN to 50 kN · CMC 0,05 %
Force
Internal procedure; At permanent location (S) PO-7.2/42 based on PN EN ISO 376:2011
2 Force gauges
Transducers
Calibration — Force
50 N to 500 kN · CMC 0,12 %
Force
Internal procedure; Outside of permanent location (P) PO-7.2/42 based on PN EN ISO 376:2011
3 Static testing machines
Technological devices
Calibration — Force
for compressive and tensile forces — 0,1 N to 100 N · CMC 0,09 % for compressive and tensile forces — 10 N to 500 kN · CMC 0,12 % for compressive forces — 500 kN to 3000 kN · CMC 0,24 %
Force
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/97 based on PN-EN ISO 7500-1:2018-05 PN EN ISO 12390-4:2025-08 cl. 4.4 PO-7.2/104 based on ASTM E4-24
4 Static testing machines
Technological devices
- crosshead speed
Calibration — Force
0 mm/min to 2500 mm/min · CMC 0,1 %
Force
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/127 based on ASTM E2658-15 (2023)
5 Extensometers Calibration — Force
0 mm to 1300 mm · CMC 0,2 % (not less than 0,67 µm)
Force
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/129 based on PN-EN ISO 9513:2013-06
6 Charpy pendulum impact testing machines Calibration — Impact strength
1 J to 750 J · CMC 0,55 % KN
Impact strength
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/71 based on PN-EN ISO 148-2:2017-02
7 Rockwell hardness testers Calibration — Hardness
hardness — 20 HRA to 88 HRA · CMC 0,6 HRA hardness — 20 HRBW to 100 HRBW · CMC 0,7 HRBW hardness — 20 HRC to 70 HRC · CMC 0,6 HRC hardness — 70 HREW to 100 HREW · CMC 0,6 HREW hardness — 70 HR15N to 91 HR15N · CMC 0,6 HR15N hardness — 42 HR30N to 80 HR30N · CMC 0,6 HR30N hardness — 67 HR15TW to 93 HR15TW · CMC 0,9 HR15TW hardness — 29 HR30TW 82 HR30TW · CMC 0,9 HR30TW force — 29,42 N to 1471 N · CMC 0,24 % length — 20 µm to 240 µm · CMC 0,42 µm cycle time — 1 s to 15 s · CMC 0,05 s
Hardness
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/99 based on PN EN ISO 6508-2:2024-06
8 Rockwell hardness testers Calibration — Hardness
hardness — 20 HRA to 88 HRA · CMC 0,6 HRA hardness — 40 HRBW to 100 HRBW · CMC 0,7 HRBW hardness — 20 HRC to 65 HRC · CMC 0,6 HRC hardness — 70 HREW to 100 HREW · CMC 0,6 HREW hardness — 70 HR15N to 92 HR15N · CMC 0,6 HR15N hardness — 42 HR30N to 82 HR30N · CMC 0,6 HR30N hardness — 67 HR15TW to 93 HR15TW · CMC 0,9 HR15TW hardness — 29 HR30TW to 83 HR30TW · CMC 0,9 HR30TW force — 29,42 N to 1471 N · CMC 0,24 % length — 20 µm to 240 µm · CMC 0,42 µm cycle time — 1 s to 15 s · CMC 0,05 s
Hardness
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/102 based on ASTM E18-25
9 Vickers hardness testers Calibration — Hardness
hardness — 100 HV 0,05 to 1000 HV 0,05 · CMC 5% hardness — 100 HV 0,1 to 1000 HV 0,1 · CMC 5% hardness — 100 HV 0,2 to 1000 HV 0,2 · CMC 5% hardness — 100 HV 0,3 to 1000 HV 0,3 · CMC 5% hardness — 100 HV 0,5 to 1000 HV 0,5 · CMC 5% hardness — 100 HV 1 to 1000 HV 1 · CMC 3% hardness — 100 HV 2 to 1000 HV 2 · CMC 3% hardness — 100 HV 5 to 1000 HV 5 · CMC 3% hardness — 100 HV 10 to 1000 HV 10 · CMC 3% hardness — 100 HV 30 to 1000 HV 30 · CMC 3% force — 0,4903 N to 294,2 N · CMC 0,24 % length — 0 mm to 0,6 mm · CMC 0,5 µm cycle time — 1 s to 20 s · CMC 0,05 s
Hardness
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/100 based on PN EN ISO 6507-2:2018-05
10 Vickers hardness testers Calibration — Hardness
hardness — 100 HV 0,05 to 1000 HV 0,05 · CMC 5% hardness — 100 HV 0,1 to 1000 HV 0,1 · CMC 5% hardness — 100 HV 0,2 to 1000 HV 0,2 · CMC 5% hardness — 100 HV 0,3 to 1000 HV 0,3 · CMC 5% hardness — 100 HV 0,5 to 1000 HV 0,5 · CMC 5% hardness — 100 HV 1 to 1000 HV 1 · CMC 3% hardness — 100 HV 2 to 1000 HV 2 · CMC 3% hardness — 100 HV 5 to 1000 HV 5 · CMC 3% hardness — 100 HV 10 to 1000 HV 10 · CMC 3% hardness — 100 HV 30 to 1000 HV 30 · CMC 3% force — 0,4903 N to 294,2 N · CMC 0,24 % length — 0 mm to 0,6 mm · CMC 0,5 µm cycle time — 1 s to 20 s · CMC 0,05 s
Hardness
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/126 based on ASTM E92-23 ASTM E384-22
11 Brinell hardness testers Calibration — Hardness
hardness — 100 HBW to 400 HBW 10/3000 · CMC 2,2 % hardness — 100 HBW to 400 HBW 10/1000 · CMC 2,2 % hardness — 100 HBW to 400 HBW 5/750 · CMC 2,2 % hardness — 100 HBW to 200 HBW 10/500 · CMC 2,2 % hardness — 100 HBW to 200 HBW 5/250 · CMC 2,2 % hardness — 100 HBW to 400 HBW 2,5/187,5 · CMC 2,2 % hardness — 100 HBW to 200 HBW 2,5/62,5 · CMC 2,2 % hardness — 100 HBW to 200 HBW 2,5/31,25 · CMC 2,2 % hardness — 100 HBW to 600 HBW 1/30 · CMC 2,2 % force — 294,2 N to 29,420 kN · CMC 0,24 % length — 0 mm to 1 mm · CMC 0,5 µm length — 0 mm to 10 mm · CMC 1 µm cycle time — 1 s to 20 s · CMC 0,05 s
Hardness
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/98 based on PN EN ISO 6506-2:2019-10
12 Brinell hardness testers Calibration — Hardness
hardness — 100 HBW to 400 HBW 10/3000 · CMC 2,2 % hardness — 100 HBW to 400 HBW 10/1000 · CMC 2,2 % hardness — 100 HBW to 400 HBW 5/750 · CMC 2,2 % hardness — 100 HBW to 200 HBW 10/500 · CMC 2,2 % hardness — 100 HBW to 250 HBW 5/250 · CMC 2,2 % hardness — 100 HBW to 400 HBW 2,5/187,5 · CMC 2,2 % hardness — 100 HBW to 250 HBW 2,5/62,5 · CMC 2,2 % hardness — 100 HBW to 200 HBW 2,5/31,25 · CMC 2,2 % hardness — 100 HBW to 600 HBW 1/30 · CMC 2,2 % force — 294,2 N to 29,420 kN · CMC 0,24 % length — 0 mm to 1 mm · CMC 0,5 µm length — 0 mm to 10 mm · CMC 1 µm cycle time — 1 s to 20 s · CMC 0,05 s
Hardness
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/103 based on ASTM E10-23
13 Knoop hardness testers Calibration — Hardness
hardness — HK 0,1 · CMC 8% hardness — HK 0,2 · CMC 8% hardness — HK 0,5 · CMC 5% force — 0,9807 N to 4,903 N · CMC 0,24 % length — 0 mm to 0,3 mm · CMC 0,5 µm cycle time — 1 s to 20 s · CMC 0,05 s
Hardness
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/130 based on PN EN ISO 4545-2:2018-04
14 Knoop hardness testers Calibration — Hardness
hardness — HK 0,1 · CMC 8% hardness — HK 0,2 · CMC 8% hardness — HK 0,5 · CMC 5% force — 0,9807 N to 4,903 N · CMC 0,24 % length — 0 mm to 0,3 mm · CMC 0,5 µm cycle time — 1 s to 20 s · CMC 0,05 s
Hardness
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/126 based on ASTM E92-23 ASTM E384-22
15 Non-automatic weighing instruments Calibration — Mass (weighing instruments)
up to 300 g · CMC 8 · 10⁻⁵ % 0,3 kg do 50 kg · CMC 1 · 10⁻⁴ %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/15 based on EURAMET cg-18 v.4.0
16 Mass standards of accuracy class F1
Weights of accuracy class F1
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,006 mg 2 mg · CMC 0,006 mg 5 mg · CMC 0,006 mg 10 mg · CMC 0,008 mg 20 mg · CMC 0,010 mg 50 mg · CMC 0,012 mg 100 mg · CMC 0,016 mg 200 mg · CMC 0,020 mg 500 mg · CMC 0,025 mg 1g · CMC 0,03 mg 2g · CMC 0,04 mg 5g · CMC 0,05 mg 10 g · CMC 0,06 mg 20 g · CMC 0,08 mg 50 g · CMC 0,10 mg 100 g · CMC 0,16 mg 200 g · CMC 0.3 mg 500 g · CMC 0,8 mg 1 kg · CMC 1,6 mg 2 kg · CMC 3,0 mg 5 kg · CMC 8,0 mg 10 kg · CMC 16 mg 20 kg · CMC 30 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
17 Mass standards of accuracy class F2
Weights of accuracy class F2
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0.75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
18 Load weights Calibration — Mass (weights and mass standards)
1 g to 40000 g · CMC 1,2 · 10⁻³ %
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
19 pH meters Calibration — pH
pH — 1 to 14 · CMC 0,004 DC voltage — -1400 mV to 1400 mV · CMC 0,3 mV
pH
Internal procedure; Electrical method; At permanent location (S) PO-7.2/37
20 pH electrodes Calibration — pH
45 mV to 70 mV · CMC 0,3
pH
Internal procedure; Slope of the electrode characteristic; At permanent location (S) PO-7.2/93
21 Conductivity meters Calibration — Specific electrical conductivity (conductometry)
0,250 µS/cm to 1,999 µS/cm · CMC 0,007 µS/cm 2,000 µS/cm to 19,999 µS/cm · CMC 0,3 % 20,00 µS/cm to 199,99 µS/cm · CMC 0,10 % 0,200 mS/cm to 1,999 mS/cm · CMC 0,18 % 2,000 mS/cm to 19,999 mS/cm · CMC 0,22 % 20,00 mS/cm to 99,99 mS/cm · CMC 0,24 % 100,00 mS/cm to 150,00 mS/cm · CMC 1,5 %
Specific electrical conductivity (conductometry)
Internal procedure; Electrical method; At permanent location (S) PO-7.2/38
22 Spectrophotometers Calibration — Spectral transmittance
zakres widmowy 400 nm to 850 nm — 0,1 to 0,3 · CMC 0,0031 zakres widmowy 400 nm to 850 nm — 0,3 to 0,5 · CMC 0,0039 zakres widmowy 400 nm to 850 nm — 0,5 to 1,0 · CMC 0,0052
Spectral transmittance
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/87
23 Spectrophotometers Calibration — Optical density of spectral transmittance
zakres widmowy 400 nm to 850 nm — 0,0 to 0,3 · CMC 0,0038 zakres widmowy 400 nm to 850 nm — 0,3 to 0,5 · CMC 0,0042 zakres widmowy 400 nm to 850 nm — 0,5 to 1,0 · CMC 0,0049 wavelength — 400 nm to 800 nm · CMC 0,20 nm
Optical density of spectral transmittance
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/87
24 Electronic pressure gauges
Bourdon tube pressure gauges
- absolute pressure (gas)
Calibration — Pressure
750 hPa to 1150 hPa · CMC 0,2 hPa
Pressure
Internal procedure; At permanent location (S) PO-7.2/17
25 Electronic pressure gauges
Bourdon tube pressure gauges
Electronic and mechanical blood pressure manometers
- gauge pressure (gas)
Calibration — Pressure
-90 kPa to 0 kPa · CMC 0,03 kPa 0 kPa to 100 kPa · CMC 0,05 kPa 100 kPa to 1000 kPa · CMC 0,2 kPa 1000 kPa do 2000 kPa · CMC 0,4 kPa 2000 kPa do 3000 kPa · CMC 2 kPa
Pressure
Internal procedure; At permanent location (S) PO-7.2/17
26 Electronic pressure gauges
Bourdon tube pressure gauges
- gauge pressure (liquid)
Calibration — Pressure
0,00 MPa do 0,6 MPa · CMC 0,0003 MPa 0,6 MPa do 1 MPa · CMC 0,0005 MPa 1 MPa do 6 MPa · CMC 0,001 MPa 6 MPa to 25 MPa · CMC 0,01 MPa 25 MPa do 40 MPa · CMC 0,02 MPa 40 MPa do 70 MPa · CMC 0,03 MPa
Pressure
Internal procedure; At permanent location (S) PO-7.2/17
27 Electrical thermometers (including electronic) Calibration — Temperature (electrical thermometry)
-50 °C to -40 °C · CMC 0,07 °C -40 °C to 0 °C · CMC 0,04 °C 0 °C · CMC 0,02 °C 0 °C to 260 °C · CMC 0,04 °C 260 °C to 1100 °C · CMC 0,7 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; At permanent location (S) PO-7.2/101
28 Electrical thermometers (including electronic) Calibration — Temperature (electrical thermometry)
-20 °C to 150 °C · CMC 0,13 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; Outside of permanent location (P) PO-7.2/101
29 Noble metal thermocouples Calibration — Temperature (electrical thermometry)
100 °C to 1100 °C · CMC 0,5 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PO-7.2/46 based on EURAMET cg-8 v.3.1
30 Base metal thermocouples Calibration — Temperature (electrical thermometry)
100 °C to 1100 °C · CMC 1,4 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PO-7.2/68 based on EURAMET cg-8 v.3.1
31 Resistance thermometer sensors Calibration — Temperature (electrical thermometry)
0,01 (ppw) · CMC 0,007 °C -50 °C to -40 °C · CMC 0,07 °C -40 °C to 0 °C · CMC 0,04 °C 0 °C · CMC 0,02 °C 0 °C to 260 °C · CMC 0,04 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; At permanent location (S) PO-7.2/117
32 Liquid-in-glass thermometers
(with a scale interval not less than 0,1 °C)
Calibration — Temperature (non-electrical thermometry)
-50 °C to -40 °C · CMC 0,07 °C -40 °C to 0 °C · CMC 0,04 °C 0 °C · CMC 0,02 °C 0 °C to 100 °C · CMC 0,04 °C 100 °C to 260 °C · CMC 0,05 °C
Temperature (non-electrical thermometry)
Internal procedure; At permanent location (S) PO-7.2/51
33 Digital voltmeters
Multimeters
Calibrators
Power quality meters
Calibration — DC voltage
1 mV to 10 mV · CMC 0,006 % 10 mV to 1000 V · CMC 0,002 %
DC voltage
Internal procedure; Direct method; At permanent location (S) PO-7.2/105 PO-7.2/112
34 Digital ammeters
Multimeters
Calibrators
Power quality meters
Calibration — DC current
1 mA to 20 A · CMC 0,006 %
DC current
Internal procedure; Direct or indirect method; At permanent location (S) PO-7.2/105 PO-7.2/112 PO-7.2/120
35 Clamp meters Calibration — DC current
3,2 A to 1000 A · CMC 0,25 %
DC current
Internal procedure; Direct method; At permanent location (S) PO-7.2/105
36 Digital voltmeters
Multimeters
Calibrators
Power quality meters
Breakdown testers
Breakdown voltage meters
Calibration — AC voltage
40 Hz to 20 kHz — 10 mV to 1000 V · CMC 0,015 %
AC voltage
Internal procedure; Direct method; At permanent location (S) PO-7.2/65 PO-7.2/105 PO-7.2/112 PO-7.2/120
37 Digital ammeters
Calibrators
Multimeters
Power quality meters
Calibration — AC current
40 Hz to 1 kHz — 1 mA to 1 A · CMC 0,015 % 40 Hz to 1 kHz — 1 A to 20 A · CMC 0,028 % 1 kHz to 5 kHz — 1 mA to 1 A · CMC 0,028 %
AC current
Internal procedure; Direct or indirect method; At permanent location (S) PO-7.2/105 PO-7.2/112 PO-7.2/120
38 Clamp meters Calibration — AC current
50 Hz — 3,2 A to 1000 A · CMC 0,56 %
AC current
Internal procedure; Direct method; At permanent location (S) PO-7.2/105
39 Digital ohmmeters
Multimeters
Calibration — DC resistance
0,01 Ω · CMC 0,008 % 0,1 Ω · CMC 0,003 % 1 Ω · CMC 0,003 % 10 Ω · CMC 0,003 % 100 Ω · CMC 0,003 % 1 kΩ · CMC 0,003 % 10 kΩ · CMC 0,003 % 100 kΩ · CMC 0,006 % 1 MΩ · CMC 0,006 % 10 MΩ · CMC 0,006 % 100 MΩ · CMC 0,8 %
DC resistance
Internal procedure; Direct method Calibration at fixed points; At permanent location (S) PO-7.2/105 based on EURAMET cg-15 v.3.0
40 Digital ohmmeters
Multimeters
Calibration — DC resistance
0,01 Ω to 0,1 Ω · CMC 0,003 % + 0,000002 Ω 0,1 Ω to 1 Ω · CMC 0,009 % + 0,000005 Ω 1 Ω to 10 Ω · CMC 0,009 % + 0,000005 Ω 10 Ω to 100 Ω · CMC 0,009 % + 0,000002 Ω 100 Ω to 1000 Ω · CMC 0,0012 % + 0,0006 Ω 1 kΩ to 10 kΩ · CMC 0,0012 % + 0,000005 kΩ 10 kΩ to 100 kΩ · CMC 0,0012 % + 0,000005 kΩ 0,1 MΩ to 1 MΩ · CMC 0,007 % + 0,00000004 MΩ 1 MΩ to 10 MΩ · CMC 0,03 % + 0,00004 MΩ 10 MΩ to 100 MΩ · CMC 0,01 % + 0,001 MΩ
DC resistance
Internal procedure; Direct method; At permanent location (S) PO-7.2/105 based on EURAMET cg-15 v.3.0
41 Resistance calibrators Calibration — DC resistance
0 Ω to 2 Ω · CMC 0,0023 % + 0,00005 Ω 2 Ω to 20 Ω · CMC 0,0012 % + 0,00005 Ω 20 Ω to 200 Ω · CMC 0,0010 % + 0,00006 Ω 0,2 kΩ to 2 kΩ · CMC 0,0010 % + 0,00002 kΩ 2 kΩ to 20 kΩ · CMC 0,0010 % + 0,000006 kΩ 20 kΩ to 200 kΩ · CMC 0,0010 % + 0,00006 kΩ 0,2 MΩ to 2 MΩ · CMC 0,0011 % + 0,000002 MΩ 2 MΩ to 20 MΩ · CMC 0,0025 % + 0,0002 MΩ 20 MΩ to 200 MΩ · CMC 0,008 % + 0,002 MΩ 0,2 GΩ to 2 GΩ · CMC 0,025 % + 0,0002 GΩ 2 GΩ to 20 GΩ · CMC 0,2 % + 0,02 GΩ
DC resistance
Internal procedure; Direct method; At permanent location (S) PO-7.2/112 based on EURAMET cg-15 v.3.0
42 Power quality meters Calibration — DC resistance
10 kΩ to 100 kΩ · CMC 0,013 % + 0,58 kΩ 100 kΩ to 1000 kΩ · CMC 0,078 % + 0,51 kΩ 1 MΩ to 5 MΩ · CMC 0,031 % + 0,0056 MΩ 5 MΩ to 10 MΩ · CMC 0,058 % + 0,0042 MΩ 10 MΩ to 100 MΩ · CMC 0,08 % + 0,052 MΩ 100 MΩ to 1000 MΩ · CMC 1,2 % + 0,15 MΩ 1 GΩ to 10 GΩ · CMC 3,5 % + 0,0006 GΩ 10 GΩ to 100 GΩ · CMC 0,6 % + 0,0004 GΩ 100 GΩ to 1000 GΩ · CMC 0,6 % + 0,004 GΩ
DC resistance
Internal procedure; Direct method; At permanent location (S) PO-7.2/105 based on EURAMET cg-15 v.3.0
43 Variable resistors Calibration — DC resistance
0,001 Ω to 0,01 Ω · CMC 0,0021 % + 0,000007 Ω 0,01 Ω to 0,1 Ω · CMC 0,0021 % + 0,000007 Ω 0,1 Ω to 1 Ω · CMC 0,0021 % + 0,000007 Ω 1 Ω to 10 Ω · CMC 0,0015 % + 0,00006 Ω 10 Ω to 100 Ω · CMC 0,0015 % + 0,00006 Ω 100 Ω to 1000 Ω · CMC 0,0015 % + 0,0006 Ω 1 kΩ to 10 kΩ · CMC 0,0015 % + 0,000006 kΩ 10 kΩ to 100 kΩ · CMC 0,0015 % + 0,00006 kΩ 100 kΩ to 1000 kΩ · CMC 0,0016 % + 0,0012 kΩ 1 MΩ to 10 MΩ · CMC 0,0026 % + 0,00012 MΩ 10 MΩ to 100 MΩ · CMC 0,0080 % 0,0012 MΩ 100 MΩ to 1000 MΩ · CMC 0,021 % + 0,12 MΩ 1 GΩ to 10 GΩ · CMC 0,18 % + 0,012 GΩ 10 GΩ to 100 GΩ · CMC 0,18 % + 0,012 GΩ
DC resistance
Internal procedure; Direct method; At permanent location (S) PO-7.2/107 based on PN-EN 60477-1:2022-12
44 Fixed resistors Calibration — DC resistance
0,0001 Ω · CMC 0,02 % 0,001 Ω to 1 GΩ · CMC 0,003 % 10 GΩ · CMC 0,06 % 20 GΩ · CMC 0,25 %
DC resistance
Internal procedure; Indirect method; At permanent location (S) PO-7.2/11 based on PN-EN 60477-1:2022-12
45 Capacitance meters
Multimeters
Calibration — Capacitance
1 kHz — 1 pF to 10 nF · CMC 0,1 % 1 kHz — 10 nF to 110 µF · CMC 0,15 %
Capacitance
Internal procedure; At permanent location (S) PO-7.2/105
46 Inductance meters
Multimeters
Calibration — Inductance
1 kHz — 10 µH to 100 µH · CMC 1,5 % 1 kHz — 100 µH to 1 mH · CMC 0,15 % 1 kHz — 1 mH to 1 H · CMC 0,05 %
Inductance
Internal procedure; At permanent location (S) PO-7.2/105
47 Analogue kilovoltmeters
Voltmeters
Calibration — High voltage and current
50 Hz — 1 kV to 75 kV · CMC 0,5 %
High voltage and current
Internal procedure; Indirect method; At permanent location (S) PO-7.2/65
48 Digital kilovoltmeters
Voltmeters
Calibration — High voltage and current
50 Hz — 1 kV to 75 kV · CMC 0,08 %
High voltage and current
Internal procedure; Indirect method; At permanent location (S) PO-7.2/105
49 Current transformers
- AC current ratio (50 Hz)
Calibration — High voltage and current
0,1 A to 5000 A / 5 A / 1 A · CMC 0,013 % 0,1 A to 5000 A / 5 A / 1 A · CMC 1′
High voltage and current
Internal procedure; Bridge method; At permanent location (S) and outside of it (P) PO-7.2/62
50 Voltage transformers
- AC voltage ratio (50 Hz)
Calibration — High voltage and current
3 kV to 35 kV / 0,11 kV / 0,10 kV · CMC 0,013 % 3 kV to 35 kV / 0,11 kV / 0,10 kV · CMC 1'
High voltage and current
Internal procedure; Bridge method; At permanent location (S) PO-7.2/62
51 Temperature indicators (meters), including temperature controllers Calibration — Electrical simulation of quantities
-270 °C to 1820 °C · CMC 0,03 °C
Electrical simulation of quantities
Internal procedure; Indirect method; At permanent location (S) PO-7.2/111 based on EURAMET cg-11 v.2.0
52 Temperature simulators Calibration — Electrical simulation of quantities
-270 °C to 1820 °C · CMC 0,03 °C
Electrical simulation of quantities
Internal procedure; Indirect method; At permanent location (S) PO-7.2/110 based on EURAMET cg-11 v.2.0
53 Temperature indicators (meters), including temperature controllers Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,01 °C
Electrical simulation of quantities
Internal procedure; Indirect method; At permanent location (S) PO-7.2/111 based on EURAMET cg-11 v.2.0
54 Temperature simulators Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,01 °C
Electrical simulation of quantities
Internal procedure; Indirect method; At permanent location (S) PO-7.2/110 based on EURAMET cg-11 v.2.0
55 Mechanical stopwatches Calibration — Time (time interval)
0 h to 1 h · CMC 0,06 s + 8 · 10⁻⁶ · τ
Time (time interval)
Internal procedure; At permanent location (S) PO-7.2/113
56 Electronic stopwatches Calibration — Time (time interval)
0 h to 24 h · CMC 0,02 s + 8 · 10⁻⁷ · τ
Time (time interval)
Internal procedure; At permanent location (S) PO-7.2/114
57 Power quality meters Calibration — Time (time interval)
20 ms to 400 ms · CMC 0,8 ms >400 ms to 1 s · CMC 10 ms
Time (time interval)
Internal procedure; At permanent location (S) PO-7.2/120
58 Non-automatic weighing instruments Calibration — Mass (weighing instruments)
up to 300 g · CMC 1 · 10⁻⁴ % 0,3 kg do 50 kg · CMC 1,5 · 10⁻³ % 50 kg do 6000 kg · CMC 1,4 · 10⁻² %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/15 based on EURAMET cg-18 v.4.0
59 Mass standards of accuracy class F2
Weights of accuracy class F2
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
60 Mass standards of accuracy class M1
Weights of accuracy class M1
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
61 Mass standards of accuracy class M2
Weights of accuracy class M2
Calibration — Mass (weights and mass standards)
100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
62 25 kg mass standards Calibration — Mass (weights and mass standards)
25 kg · CMC 280 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
63 Load weights Calibration — Mass (weights and mass standards)
1 g to 25000 g · CMC 1,2 · 10⁻³ %
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
64 Non-automatic weighing instruments Calibration — Mass (weighing instruments)
up to 20 kg · CMC 2,9 · 10⁻³ % 20 kg to 6000 kg · CMC 1,4 · 10⁻² %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/15 based on EURAMET cg-18 v.4.0
65 Mass standards of accuracy class F2
Weights of accuracy class F2
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
66 Mass standards of accuracy class M1
Weights of accuracy class M1
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
67 Mass standards of accuracy class M2
Weights of accuracy class M2
Calibration — Mass (weights and mass standards)
100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
68 25 kg mass standards Calibration — Mass (weights and mass standards)
25 kg · CMC 280 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
69 Load weights Calibration — Mass (weights and mass standards)
1 g to 25000 g · CMC 1,2 · 10⁻³ %
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
70 Rockwell hardness testers Calibration — Hardness
hardness — 20 HRA to 88 HRA · CMC 0,7 HRA hardness — 20 HRBW to 100 HRBW · CMC 0,8 HRBW hardness — 20 HRC to 70 HRC · CMC 0,7 HRC force — 98,07 N to 1471 N · CMC 0,26 %
Hardness
Internal procedure; Outside of permanent location (P) PO-7.2/99 based on PN EN ISO 6508-2:2024-06
71 Non-automatic weighing instruments Calibration — Mass (weighing instruments)
up to 300 g · CMC 1,4 · 10⁻⁴ % 0,3 kg do 50 kg · CMC 5 · 10⁻⁴ % 50 kg do 10000 kg · CMC 1,4 · 10⁻² %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/15 based on EURAMET cg-18 v.4.0
72 Mass standards of accuracy class F2
Weights of accuracy class F2
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
73 Mass standards of accuracy class M1
Weights of accuracy class M1
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
74 Mass standards of accuracy class M2
Weights of accuracy class M2
Calibration — Mass (weights and mass standards)
100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
75 25 kg mass standards Calibration — Mass (weights and mass standards)
25 kg · CMC 280 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
76 Load weights Calibration — Mass (weights and mass standards)
1 g to 25000 g · CMC 1,2 · 10⁻³ %
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
77 Non-automatic weighing instruments Calibration — Mass (weighing instruments)
up to 300 g · CMC 1 · 10⁻⁴ % 0,3 kg do 50 kg · CMC 1 · 10⁻³ % 50 kg do 6000 kg · CMC 1,5 · 10⁻² %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/15 based on EURAMET cg-18 v.4.0
78 Mass standards of accuracy class F2
Weights of accuracy class F2
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
79 Mass standards of accuracy class M1
Weights of accuracy class M1
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
80 Mass standards of accuracy class M2
Weights of accuracy class M2
Calibration — Mass (weights and mass standards)
100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
81 25 kg mass standards Calibration — Mass (weights and mass standards)
25 kg · CMC 280 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
82 Load weights Calibration — Mass (weights and mass standards)
1 g to 25000 g · CMC 1,2 · 10⁻³ %
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
83 Gauge blocks of accuracy classes 0, 1, 2 Calibration — Length
0,5 mm to 100 mm · CMC Q[0,09; 1,5 · L] µm 125 mm to 500 mm · CMC Q[0,2; 1,5 · L] µm
Length
Internal procedure; At permanent location (S) PO-7.2/39
84 Optical parallels Calibration — Length
length deviation — length up to 80 mm · CMC 0,93 µm
Length
Internal procedure; At permanent location (S) PO-7.2/67
85 Calipers Calibration — Length
0 mm to 1000 mm · CMC (0,04 + 0,01 · L) mm
Length
Internal procedure; At permanent location (S) PO-7.2/74
86 Caliper height gauges Calibration — Length
0 mm to 1000 mm · CMC (0,04 + 0,01 · L) mm
Length
Internal procedure; At permanent location (S) PO-7.2/75
87 Caliper depth gauges Calibration — Length
0 mm to 500 mm · CMC (0,04 + 0,01 · L) mm
Length
Internal procedure; At permanent location (S) PO-7.2/76
88 External micrometers Calibration — Length
0 mm to 175 mm · CMC (0,9 + 2,0 · L) µm
Length
Internal procedure; At permanent location (S) PO-7.2/77
89 Micrometer depth gauges Calibration — Length
0 mm to 150 mm · CMC (4,0 + 0,3 · L) µm
Length
Internal procedure; At permanent location (S) PO-7.2/78
90 Two-point internal micrometers Calibration — Length
0 mm to 200 mm · CMC (2,6 + 1,1 · L) µm
Length
Internal procedure; At permanent location (S) PO-7.2/114
91 Rigid rules
Semi-rigid rules
Folding rules
Draper's rules
Calibration — Length
0 mm to 1000 mm · CMC (0,15 + 0,001 · L) mm
Length
Internal procedure; At permanent location (S) PO-7.2/2
92 Rigid rules
Semi-rigid rules
Folding rules
Calibration — Length
0 mm to 5000 mm · CMC (0,25 + 0,01 · L) mm
Length
Internal procedure; At permanent location (S) PO-7.2/2
93 Tape measures Calibration — Length
0 mm to 5000 mm · CMC (0,29 + 0,01 · L) mm 0 mm to 10000 mm · CMC 0,68·mm 0 mm to 15000 mm · CMC 1,02·mm 0 mm to 20000 mm · CMC 1,36·mm 0 mm to 25000 mm · CMC 1,70·mm 0 mm to 30000 mm · CMC 2,04·mm
Length
Internal procedure; At permanent location (S) PO-7.2/2
94 Johansson angle gauge blocks
Kusznikow angle gauge blocks
Wringing angle gauge blocks
Calibration — Angle
0º to 360º · CMC 4,1”
Angle
Internal procedure; At permanent location (S) PO-7.2/59
95 Universal bevel protractors Calibration — Angle
0º to 360º · CMC 3,5’
Angle
Internal procedure; At permanent location (S) PO-7.2/58
96 90º cylindrical squares Calibration — Angle
squareness deviation — up to 315 mm · CMC 2,40 µm
Angle
Internal procedure; At permanent location (S) PO-7.2/81
97 90º try squares Calibration — Angle
flatness and straightness deviation — up to 600 mm longer arm · CMC 3,4 µm squareness deviation — up to 600 mm longer arm · CMC 4,0 µm parallelism deviation — up to 600 mm longer arm · CMC 3,6 µm
Angle
Internal procedure; At permanent location (S) PO-7.2/82
98 Bar spirit levels Calibration — Angle
flatness deviation — 0 mm/m to 1 mm/m · CMC 4,2 µm zero setting error — 0 mm/m to 1 mm/m · CMC 0,30 scale int. scale interval error — 0 mm/m to 1 mm/m · CMC 0,031 mm/m
Angle
Internal procedure; At permanent location (S) PO-7.2/122
99 Builders' spirit levels Calibration — Angle
zero setting error — up to 120 cm · CMC 0,28 mm
Angle
Internal procedure; At permanent location (S) PO-7.2/123
100 Optical flats Calibration — Length (surface geometry)
measuring surface — diameter up to 100 mm · CMC 0,05 µm
Length (surface geometry)
Internal procedure; At permanent location (S) PO-7.2/66
101 Optical parallels Calibration — Length (surface geometry)
flatness deviation — length up to 80 mm · CMC 0,07 µm parallelism deviation — length up to 80 mm · CMC 0,26 µm
Length (surface geometry)
Internal procedure; At permanent location (S) PO-7.2/67
102 Comparison roughness specimens Calibration — Length (surface geometry)
parameter Rₐ up to 12,5 µm · CMC 30 %
Length (surface geometry)
Internal procedure; At permanent location (S) PO-7.2/80
103 Straight edges Calibration — Length (surface geometry)
up to 150 mm · CMC 1,4 µm 150 mm do 500 mm · CMC 2,6 µm
Length (surface geometry)
Internal procedure; At permanent location (S) PO-7.2/79
104 Non-automatic weighing instruments Calibration — Mass (weighing instruments)
up to 350 g · CMC 1,4 · 10⁻⁴ % 0,35 kg to 10 kg · CMC 1,2 · 10⁻³ % 10 kg to 30 kg · CMC 3,2 · 10⁻³ % 30 kg to 600 kg · CMC 2,9 · 10⁻² % 600 kg to 6000 kg · CMC 6,4 · 10⁻² %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/15 based on EURAMET cg-18 v.4.0
105 Mass standards of accuracy class F2
Weights of accuracy class F2
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
106 Mass standards of accuracy class M1
Weights of accuracy class M1
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
107 Mass standards of accuracy class M2
Weights of accuracy class M2
Calibration — Mass (weights and mass standards)
100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
108 25 kg mass standards Calibration — Mass (weights and mass standards)
25 kg · CMC 280 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
109 Load weights Calibration — Mass (weights and mass standards)
1 g to 25000 g · CMC 1,2 · 10⁻³ %
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
110 Bourdon tube pressure gauges
- gauge pressure (gas)
Calibration — Pressure
-0,1 MPa do -0,005 MPa · CMC 0,0002 MPa 0,005 MPa to 0,25 MPa · CMC 0,0002 MPa
Pressure
Internal procedure; At permanent location (S) PO-7.2/1
111 Bourdon tube pressure gauges
- gauge pressure (liquid)
Calibration — Pressure
0,02 MPa do 0,6 MPa · CMC 0,0005 MPa 0,6 MPa to 6 MPa · CMC 0,006 MPa 6 MPa do 60 MPa · CMC 0,07 MPa
Pressure
Internal procedure; At permanent location (S) PO-7.2/1
112 Non-automatic weighing instruments Calibration — Mass (weighing instruments)
up to 300 g · CMC 1,4 · 10⁻⁴ % 0,3 kg do 50 kg · CMC 5 · 10⁻⁴ % 50 kg do 6000 kg · CMC 1,4 · 10⁻² %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) PO-7.2/15 based on EURAMET cg-18 v.4.0
113 Mass standards of accuracy class F2
Weights of accuracy class F2
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
114 Mass standards of accuracy class M1
Weights of accuracy class M1
Calibration — Mass (weights and mass standards)
1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
115 Mass standards of accuracy class M2
Weights of accuracy class M2
Calibration — Mass (weights and mass standards)
100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
116 25 kg mass standards Calibration — Mass (weights and mass standards)
25 kg · CMC 280 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C
117 Load weights Calibration — Mass (weights and mass standards)
1 g to 25000 g · CMC 1,2 · 10⁻³ %
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PO-7.2/88 based on OIML R111-1:2004 Annex C

Reference standards (75)

ASTM E10-23
ASTM E18-25
ASTM E2658-15 (2023)
ASTM E384-22
ASTM E4-24
ASTM E92-23
EURAMET cg-11 v.2.0
EURAMET cg-15 v.3.0
EURAMET cg-18 v.4.0
EURAMET cg-8 v.3.1
OIML R111-1:2004 Załącznik C
PN EN ISO 12390-4:2025-08 cl. 4.4
PN EN ISO 376:2011
PN EN ISO 4545-2:2018-04
PN EN ISO 6506-2:2019-10
PN EN ISO 6507-2:2018-05
PN EN ISO 6508-2:2024-06
PN-EN 60477-1:2022-12
PN-EN ISO 148-2:2017-02
PN-EN ISO 7500-1:2018-05
PN-EN ISO 9513:2013-06
PO-7.2/1
PO-7.2/100
PO-7.2/101
PO-7.2/102
PO-7.2/103
PO-7.2/104
PO-7.2/105
PO-7.2/107
PO-7.2/11
PO-7.2/110
PO-7.2/111
PO-7.2/112
PO-7.2/113
PO-7.2/114
PO-7.2/117
PO-7.2/120
PO-7.2/122
PO-7.2/123
PO-7.2/126
PO-7.2/127
PO-7.2/129
PO-7.2/130
PO-7.2/15
PO-7.2/17
PO-7.2/2
PO-7.2/37
PO-7.2/38
PO-7.2/39
PO-7.2/42
PO-7.2/46
PO-7.2/51
PO-7.2/58
PO-7.2/59
PO-7.2/62
PO-7.2/65
PO-7.2/66
PO-7.2/67
PO-7.2/68
PO-7.2/71
PO-7.2/74
PO-7.2/75
PO-7.2/76
PO-7.2/77
PO-7.2/78
PO-7.2/79
PO-7.2/80
PO-7.2/81
PO-7.2/82
PO-7.2/87
PO-7.2/88
PO-7.2/93
PO-7.2/97
PO-7.2/98
PO-7.2/99

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20082