🇵🇱 Okręgowy Urząd Miar w Krakowie PL
Metrology & calibration
Okręgowy Urząd Miar w Krakowie is a calibration laboratory accredited by PCA under number AP 082 (Kraków, Poland). Its accreditation scope covers 117 calibration items in the following areas: Metrology & calibration. Calibrated items include: Multimeters, Load weights, Mass standards of accuracy class F2, Non-automatic weighing instruments. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; Direct method; At permanent location (S).
At a glance
- Accreditation number: PCA AP 082
- Calibration items in scope: 117
- Types of calibrated items: 87
- Calibration principles used: 12
- City: Kraków, Poland
- Valid until: 18-01-2030
- Accreditation number
- AP 082
- Accreditation body
- PCA
- Address
- ul. Krupnicza 11, 31-123 Kraków
- City
- Kraków, Poland
- Division
- Calibration Laboratories Group
- Valid until
- 18-01-2030
What this laboratory calibrates (87)
- Multimeters — 8 calibration items
- Load weights — 7 calibration items
- Mass standards of accuracy class F2 — 7 calibration items
- Non-automatic weighing instruments — 7 calibration items
- Weights of accuracy class F2 — 7 calibration items
- 25 kg mass standards — 6 calibration items
- Mass standards of accuracy class M1 — 6 calibration items
- Mass standards of accuracy class M2 — 6 calibration items
- Power quality meters — 6 calibration items
- Weights of accuracy class M1 — 6 calibration items
- Weights of accuracy class M2 — 6 calibration items
- Bourdon tube pressure gauges — 5 calibration items
- Calibrators — 4 calibration items
- Electronic pressure gauges — 3 calibration items
- Rockwell hardness testers — 3 calibration items
- and 72 more
Calibration principles (12)
- Internal procedure; At permanent location (S) — 68 calibration items
- Internal procedure; At permanent location (S) and outside of it (P) — 21 calibration items
- Internal procedure; Direct method; At permanent location (S) — 10 calibration items
- Internal procedure; Indirect method; At permanent location (S) — 7 calibration items
- Internal procedure; Direct or indirect method; At permanent location (S) — 2 calibration items
- Internal procedure; Electrical method; At permanent location (S) — 2 calibration items
- Internal procedure; Outside of permanent location (P) — 2 calibration items
- Internal procedure; Bridge method; At permanent location (S) — 1 calibration item
- Internal procedure; Bridge method; At permanent location (S) and outside of it (P) — 1 calibration item
- Internal procedure; Direct method Calibration at fixed points; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method; Outside of permanent location (P) — 1 calibration item
- Internal procedure; Slope of the electrode characteristic; At permanent location (S) — 1 calibration item
Certificate information
- Issue date
- 13.01.2026
- Issue no.
- 24
- Certificate date
- 20.12.2019
- Accreditation cycle
- 13.01.2026 — 18.01.2030
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (117)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Force gauges Transducers |
Calibration — Force 0,1 N to 100 N · CMC 0,05 % 100 N to 5000 N · CMC 0,05 % 1 kN to 2 kN · CMC 0,07 % 3 kN to 50 kN · CMC 0,05 % Force |
Internal procedure; At permanent location (S) | PO-7.2/42 based on PN EN ISO 376:2011 | |
| 2 | Force gauges Transducers |
Calibration — Force 50 N to 500 kN · CMC 0,12 % Force |
Internal procedure; Outside of permanent location (P) | PO-7.2/42 based on PN EN ISO 376:2011 | |
| 3 | Static testing machines Technological devices |
Calibration — Force for compressive and tensile forces — 0,1 N to 100 N · CMC 0,09 % for compressive and tensile forces — 10 N to 500 kN · CMC 0,12 % for compressive forces — 500 kN to 3000 kN · CMC 0,24 % Force |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/97 based on PN-EN ISO 7500-1:2018-05 PN EN ISO 12390-4:2025-08 cl. 4.4 PO-7.2/104 based on ASTM E4-24 | |
| 4 | Static testing machines Technological devices - crosshead speed |
Calibration — Force 0 mm/min to 2500 mm/min · CMC 0,1 % Force |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/127 based on ASTM E2658-15 (2023) | |
| 5 | Extensometers | Calibration — Force 0 mm to 1300 mm · CMC 0,2 % (not less than 0,67 µm) Force |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/129 based on PN-EN ISO 9513:2013-06 | |
| 6 | Charpy pendulum impact testing machines | Calibration — Impact strength 1 J to 750 J · CMC 0,55 % KN Impact strength |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/71 based on PN-EN ISO 148-2:2017-02 | |
| 7 | Rockwell hardness testers | Calibration — Hardness hardness — 20 HRA to 88 HRA · CMC 0,6 HRA hardness — 20 HRBW to 100 HRBW · CMC 0,7 HRBW hardness — 20 HRC to 70 HRC · CMC 0,6 HRC hardness — 70 HREW to 100 HREW · CMC 0,6 HREW hardness — 70 HR15N to 91 HR15N · CMC 0,6 HR15N hardness — 42 HR30N to 80 HR30N · CMC 0,6 HR30N hardness — 67 HR15TW to 93 HR15TW · CMC 0,9 HR15TW hardness — 29 HR30TW 82 HR30TW · CMC 0,9 HR30TW force — 29,42 N to 1471 N · CMC 0,24 % length — 20 µm to 240 µm · CMC 0,42 µm cycle time — 1 s to 15 s · CMC 0,05 s Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/99 based on PN EN ISO 6508-2:2024-06 | |
| 8 | Rockwell hardness testers | Calibration — Hardness hardness — 20 HRA to 88 HRA · CMC 0,6 HRA hardness — 40 HRBW to 100 HRBW · CMC 0,7 HRBW hardness — 20 HRC to 65 HRC · CMC 0,6 HRC hardness — 70 HREW to 100 HREW · CMC 0,6 HREW hardness — 70 HR15N to 92 HR15N · CMC 0,6 HR15N hardness — 42 HR30N to 82 HR30N · CMC 0,6 HR30N hardness — 67 HR15TW to 93 HR15TW · CMC 0,9 HR15TW hardness — 29 HR30TW to 83 HR30TW · CMC 0,9 HR30TW force — 29,42 N to 1471 N · CMC 0,24 % length — 20 µm to 240 µm · CMC 0,42 µm cycle time — 1 s to 15 s · CMC 0,05 s Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/102 based on ASTM E18-25 | |
| 9 | Vickers hardness testers | Calibration — Hardness hardness — 100 HV 0,05 to 1000 HV 0,05 · CMC 5% hardness — 100 HV 0,1 to 1000 HV 0,1 · CMC 5% hardness — 100 HV 0,2 to 1000 HV 0,2 · CMC 5% hardness — 100 HV 0,3 to 1000 HV 0,3 · CMC 5% hardness — 100 HV 0,5 to 1000 HV 0,5 · CMC 5% hardness — 100 HV 1 to 1000 HV 1 · CMC 3% hardness — 100 HV 2 to 1000 HV 2 · CMC 3% hardness — 100 HV 5 to 1000 HV 5 · CMC 3% hardness — 100 HV 10 to 1000 HV 10 · CMC 3% hardness — 100 HV 30 to 1000 HV 30 · CMC 3% force — 0,4903 N to 294,2 N · CMC 0,24 % length — 0 mm to 0,6 mm · CMC 0,5 µm cycle time — 1 s to 20 s · CMC 0,05 s Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/100 based on PN EN ISO 6507-2:2018-05 | |
| 10 | Vickers hardness testers | Calibration — Hardness hardness — 100 HV 0,05 to 1000 HV 0,05 · CMC 5% hardness — 100 HV 0,1 to 1000 HV 0,1 · CMC 5% hardness — 100 HV 0,2 to 1000 HV 0,2 · CMC 5% hardness — 100 HV 0,3 to 1000 HV 0,3 · CMC 5% hardness — 100 HV 0,5 to 1000 HV 0,5 · CMC 5% hardness — 100 HV 1 to 1000 HV 1 · CMC 3% hardness — 100 HV 2 to 1000 HV 2 · CMC 3% hardness — 100 HV 5 to 1000 HV 5 · CMC 3% hardness — 100 HV 10 to 1000 HV 10 · CMC 3% hardness — 100 HV 30 to 1000 HV 30 · CMC 3% force — 0,4903 N to 294,2 N · CMC 0,24 % length — 0 mm to 0,6 mm · CMC 0,5 µm cycle time — 1 s to 20 s · CMC 0,05 s Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/126 based on ASTM E92-23 ASTM E384-22 | |
| 11 | Brinell hardness testers | Calibration — Hardness hardness — 100 HBW to 400 HBW 10/3000 · CMC 2,2 % hardness — 100 HBW to 400 HBW 10/1000 · CMC 2,2 % hardness — 100 HBW to 400 HBW 5/750 · CMC 2,2 % hardness — 100 HBW to 200 HBW 10/500 · CMC 2,2 % hardness — 100 HBW to 200 HBW 5/250 · CMC 2,2 % hardness — 100 HBW to 400 HBW 2,5/187,5 · CMC 2,2 % hardness — 100 HBW to 200 HBW 2,5/62,5 · CMC 2,2 % hardness — 100 HBW to 200 HBW 2,5/31,25 · CMC 2,2 % hardness — 100 HBW to 600 HBW 1/30 · CMC 2,2 % force — 294,2 N to 29,420 kN · CMC 0,24 % length — 0 mm to 1 mm · CMC 0,5 µm length — 0 mm to 10 mm · CMC 1 µm cycle time — 1 s to 20 s · CMC 0,05 s Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/98 based on PN EN ISO 6506-2:2019-10 | |
| 12 | Brinell hardness testers | Calibration — Hardness hardness — 100 HBW to 400 HBW 10/3000 · CMC 2,2 % hardness — 100 HBW to 400 HBW 10/1000 · CMC 2,2 % hardness — 100 HBW to 400 HBW 5/750 · CMC 2,2 % hardness — 100 HBW to 200 HBW 10/500 · CMC 2,2 % hardness — 100 HBW to 250 HBW 5/250 · CMC 2,2 % hardness — 100 HBW to 400 HBW 2,5/187,5 · CMC 2,2 % hardness — 100 HBW to 250 HBW 2,5/62,5 · CMC 2,2 % hardness — 100 HBW to 200 HBW 2,5/31,25 · CMC 2,2 % hardness — 100 HBW to 600 HBW 1/30 · CMC 2,2 % force — 294,2 N to 29,420 kN · CMC 0,24 % length — 0 mm to 1 mm · CMC 0,5 µm length — 0 mm to 10 mm · CMC 1 µm cycle time — 1 s to 20 s · CMC 0,05 s Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/103 based on ASTM E10-23 | |
| 13 | Knoop hardness testers | Calibration — Hardness hardness — HK 0,1 · CMC 8% hardness — HK 0,2 · CMC 8% hardness — HK 0,5 · CMC 5% force — 0,9807 N to 4,903 N · CMC 0,24 % length — 0 mm to 0,3 mm · CMC 0,5 µm cycle time — 1 s to 20 s · CMC 0,05 s Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/130 based on PN EN ISO 4545-2:2018-04 | |
| 14 | Knoop hardness testers | Calibration — Hardness hardness — HK 0,1 · CMC 8% hardness — HK 0,2 · CMC 8% hardness — HK 0,5 · CMC 5% force — 0,9807 N to 4,903 N · CMC 0,24 % length — 0 mm to 0,3 mm · CMC 0,5 µm cycle time — 1 s to 20 s · CMC 0,05 s Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/126 based on ASTM E92-23 ASTM E384-22 | |
| 15 | Non-automatic weighing instruments | Calibration — Mass (weighing instruments) up to 300 g · CMC 8 · 10⁻⁵ % 0,3 kg do 50 kg · CMC 1 · 10⁻⁴ % Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/15 based on EURAMET cg-18 v.4.0 | |
| 16 | Mass standards of accuracy class F1 Weights of accuracy class F1 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,006 mg 2 mg · CMC 0,006 mg 5 mg · CMC 0,006 mg 10 mg · CMC 0,008 mg 20 mg · CMC 0,010 mg 50 mg · CMC 0,012 mg 100 mg · CMC 0,016 mg 200 mg · CMC 0,020 mg 500 mg · CMC 0,025 mg 1g · CMC 0,03 mg 2g · CMC 0,04 mg 5g · CMC 0,05 mg 10 g · CMC 0,06 mg 20 g · CMC 0,08 mg 50 g · CMC 0,10 mg 100 g · CMC 0,16 mg 200 g · CMC 0.3 mg 500 g · CMC 0,8 mg 1 kg · CMC 1,6 mg 2 kg · CMC 3,0 mg 5 kg · CMC 8,0 mg 10 kg · CMC 16 mg 20 kg · CMC 30 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 17 | Mass standards of accuracy class F2 Weights of accuracy class F2 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0.75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 18 | Load weights | Calibration — Mass (weights and mass standards) 1 g to 40000 g · CMC 1,2 · 10⁻³ % Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 19 | pH meters | Calibration — pH pH — 1 to 14 · CMC 0,004 DC voltage — -1400 mV to 1400 mV · CMC 0,3 mV pH |
Internal procedure; Electrical method; At permanent location (S) | PO-7.2/37 | |
| 20 | pH electrodes | Calibration — pH 45 mV to 70 mV · CMC 0,3 pH |
Internal procedure; Slope of the electrode characteristic; At permanent location (S) | PO-7.2/93 | |
| 21 | Conductivity meters | Calibration — Specific electrical conductivity (conductometry) 0,250 µS/cm to 1,999 µS/cm · CMC 0,007 µS/cm 2,000 µS/cm to 19,999 µS/cm · CMC 0,3 % 20,00 µS/cm to 199,99 µS/cm · CMC 0,10 % 0,200 mS/cm to 1,999 mS/cm · CMC 0,18 % 2,000 mS/cm to 19,999 mS/cm · CMC 0,22 % 20,00 mS/cm to 99,99 mS/cm · CMC 0,24 % 100,00 mS/cm to 150,00 mS/cm · CMC 1,5 % Specific electrical conductivity (conductometry) |
Internal procedure; Electrical method; At permanent location (S) | PO-7.2/38 | |
| 22 | Spectrophotometers | Calibration — Spectral transmittance zakres widmowy 400 nm to 850 nm — 0,1 to 0,3 · CMC 0,0031 zakres widmowy 400 nm to 850 nm — 0,3 to 0,5 · CMC 0,0039 zakres widmowy 400 nm to 850 nm — 0,5 to 1,0 · CMC 0,0052 Spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/87 | |
| 23 | Spectrophotometers | Calibration — Optical density of spectral transmittance zakres widmowy 400 nm to 850 nm — 0,0 to 0,3 · CMC 0,0038 zakres widmowy 400 nm to 850 nm — 0,3 to 0,5 · CMC 0,0042 zakres widmowy 400 nm to 850 nm — 0,5 to 1,0 · CMC 0,0049 wavelength — 400 nm to 800 nm · CMC 0,20 nm Optical density of spectral transmittance |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/87 | |
| 24 | Electronic pressure gauges Bourdon tube pressure gauges - absolute pressure (gas) |
Calibration — Pressure 750 hPa to 1150 hPa · CMC 0,2 hPa Pressure |
Internal procedure; At permanent location (S) | PO-7.2/17 | |
| 25 | Electronic pressure gauges Bourdon tube pressure gauges Electronic and mechanical blood pressure manometers - gauge pressure (gas) |
Calibration — Pressure -90 kPa to 0 kPa · CMC 0,03 kPa 0 kPa to 100 kPa · CMC 0,05 kPa 100 kPa to 1000 kPa · CMC 0,2 kPa 1000 kPa do 2000 kPa · CMC 0,4 kPa 2000 kPa do 3000 kPa · CMC 2 kPa Pressure |
Internal procedure; At permanent location (S) | PO-7.2/17 | |
| 26 | Electronic pressure gauges Bourdon tube pressure gauges - gauge pressure (liquid) |
Calibration — Pressure 0,00 MPa do 0,6 MPa · CMC 0,0003 MPa 0,6 MPa do 1 MPa · CMC 0,0005 MPa 1 MPa do 6 MPa · CMC 0,001 MPa 6 MPa to 25 MPa · CMC 0,01 MPa 25 MPa do 40 MPa · CMC 0,02 MPa 40 MPa do 70 MPa · CMC 0,03 MPa Pressure |
Internal procedure; At permanent location (S) | PO-7.2/17 | |
| 27 | Electrical thermometers (including electronic) | Calibration — Temperature (electrical thermometry) -50 °C to -40 °C · CMC 0,07 °C -40 °C to 0 °C · CMC 0,04 °C 0 °C · CMC 0,02 °C 0 °C to 260 °C · CMC 0,04 °C 260 °C to 1100 °C · CMC 0,7 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | PO-7.2/101 | |
| 28 | Electrical thermometers (including electronic) | Calibration — Temperature (electrical thermometry) -20 °C to 150 °C · CMC 0,13 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; Outside of permanent location (P) | PO-7.2/101 | |
| 29 | Noble metal thermocouples | Calibration — Temperature (electrical thermometry) 100 °C to 1100 °C · CMC 0,5 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PO-7.2/46 based on EURAMET cg-8 v.3.1 | |
| 30 | Base metal thermocouples | Calibration — Temperature (electrical thermometry) 100 °C to 1100 °C · CMC 1,4 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PO-7.2/68 based on EURAMET cg-8 v.3.1 | |
| 31 | Resistance thermometer sensors | Calibration — Temperature (electrical thermometry) 0,01 (ppw) · CMC 0,007 °C -50 °C to -40 °C · CMC 0,07 °C -40 °C to 0 °C · CMC 0,04 °C 0 °C · CMC 0,02 °C 0 °C to 260 °C · CMC 0,04 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | PO-7.2/117 | |
| 32 | Liquid-in-glass thermometers (with a scale interval not less than 0,1 °C) |
Calibration — Temperature (non-electrical thermometry) -50 °C to -40 °C · CMC 0,07 °C -40 °C to 0 °C · CMC 0,04 °C 0 °C · CMC 0,02 °C 0 °C to 100 °C · CMC 0,04 °C 100 °C to 260 °C · CMC 0,05 °C Temperature (non-electrical thermometry) |
Internal procedure; At permanent location (S) | PO-7.2/51 | |
| 33 | Digital voltmeters Multimeters Calibrators Power quality meters |
Calibration — DC voltage 1 mV to 10 mV · CMC 0,006 % 10 mV to 1000 V · CMC 0,002 % DC voltage |
Internal procedure; Direct method; At permanent location (S) | PO-7.2/105 PO-7.2/112 | |
| 34 | Digital ammeters Multimeters Calibrators Power quality meters |
Calibration — DC current 1 mA to 20 A · CMC 0,006 % DC current |
Internal procedure; Direct or indirect method; At permanent location (S) | PO-7.2/105 PO-7.2/112 PO-7.2/120 | |
| 35 | Clamp meters | Calibration — DC current 3,2 A to 1000 A · CMC 0,25 % DC current |
Internal procedure; Direct method; At permanent location (S) | PO-7.2/105 | |
| 36 | Digital voltmeters Multimeters Calibrators Power quality meters Breakdown testers Breakdown voltage meters |
Calibration — AC voltage 40 Hz to 20 kHz — 10 mV to 1000 V · CMC 0,015 % AC voltage |
Internal procedure; Direct method; At permanent location (S) | PO-7.2/65 PO-7.2/105 PO-7.2/112 PO-7.2/120 | |
| 37 | Digital ammeters Calibrators Multimeters Power quality meters |
Calibration — AC current 40 Hz to 1 kHz — 1 mA to 1 A · CMC 0,015 % 40 Hz to 1 kHz — 1 A to 20 A · CMC 0,028 % 1 kHz to 5 kHz — 1 mA to 1 A · CMC 0,028 % AC current |
Internal procedure; Direct or indirect method; At permanent location (S) | PO-7.2/105 PO-7.2/112 PO-7.2/120 | |
| 38 | Clamp meters | Calibration — AC current 50 Hz — 3,2 A to 1000 A · CMC 0,56 % AC current |
Internal procedure; Direct method; At permanent location (S) | PO-7.2/105 | |
| 39 | Digital ohmmeters Multimeters |
Calibration — DC resistance 0,01 Ω · CMC 0,008 % 0,1 Ω · CMC 0,003 % 1 Ω · CMC 0,003 % 10 Ω · CMC 0,003 % 100 Ω · CMC 0,003 % 1 kΩ · CMC 0,003 % 10 kΩ · CMC 0,003 % 100 kΩ · CMC 0,006 % 1 MΩ · CMC 0,006 % 10 MΩ · CMC 0,006 % 100 MΩ · CMC 0,8 % DC resistance |
Internal procedure; Direct method Calibration at fixed points; At permanent location (S) | PO-7.2/105 based on EURAMET cg-15 v.3.0 | |
| 40 | Digital ohmmeters Multimeters |
Calibration — DC resistance 0,01 Ω to 0,1 Ω · CMC 0,003 % + 0,000002 Ω 0,1 Ω to 1 Ω · CMC 0,009 % + 0,000005 Ω 1 Ω to 10 Ω · CMC 0,009 % + 0,000005 Ω 10 Ω to 100 Ω · CMC 0,009 % + 0,000002 Ω 100 Ω to 1000 Ω · CMC 0,0012 % + 0,0006 Ω 1 kΩ to 10 kΩ · CMC 0,0012 % + 0,000005 kΩ 10 kΩ to 100 kΩ · CMC 0,0012 % + 0,000005 kΩ 0,1 MΩ to 1 MΩ · CMC 0,007 % + 0,00000004 MΩ 1 MΩ to 10 MΩ · CMC 0,03 % + 0,00004 MΩ 10 MΩ to 100 MΩ · CMC 0,01 % + 0,001 MΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | PO-7.2/105 based on EURAMET cg-15 v.3.0 | |
| 41 | Resistance calibrators | Calibration — DC resistance 0 Ω to 2 Ω · CMC 0,0023 % + 0,00005 Ω 2 Ω to 20 Ω · CMC 0,0012 % + 0,00005 Ω 20 Ω to 200 Ω · CMC 0,0010 % + 0,00006 Ω 0,2 kΩ to 2 kΩ · CMC 0,0010 % + 0,00002 kΩ 2 kΩ to 20 kΩ · CMC 0,0010 % + 0,000006 kΩ 20 kΩ to 200 kΩ · CMC 0,0010 % + 0,00006 kΩ 0,2 MΩ to 2 MΩ · CMC 0,0011 % + 0,000002 MΩ 2 MΩ to 20 MΩ · CMC 0,0025 % + 0,0002 MΩ 20 MΩ to 200 MΩ · CMC 0,008 % + 0,002 MΩ 0,2 GΩ to 2 GΩ · CMC 0,025 % + 0,0002 GΩ 2 GΩ to 20 GΩ · CMC 0,2 % + 0,02 GΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | PO-7.2/112 based on EURAMET cg-15 v.3.0 | |
| 42 | Power quality meters | Calibration — DC resistance 10 kΩ to 100 kΩ · CMC 0,013 % + 0,58 kΩ 100 kΩ to 1000 kΩ · CMC 0,078 % + 0,51 kΩ 1 MΩ to 5 MΩ · CMC 0,031 % + 0,0056 MΩ 5 MΩ to 10 MΩ · CMC 0,058 % + 0,0042 MΩ 10 MΩ to 100 MΩ · CMC 0,08 % + 0,052 MΩ 100 MΩ to 1000 MΩ · CMC 1,2 % + 0,15 MΩ 1 GΩ to 10 GΩ · CMC 3,5 % + 0,0006 GΩ 10 GΩ to 100 GΩ · CMC 0,6 % + 0,0004 GΩ 100 GΩ to 1000 GΩ · CMC 0,6 % + 0,004 GΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | PO-7.2/105 based on EURAMET cg-15 v.3.0 | |
| 43 | Variable resistors | Calibration — DC resistance 0,001 Ω to 0,01 Ω · CMC 0,0021 % + 0,000007 Ω 0,01 Ω to 0,1 Ω · CMC 0,0021 % + 0,000007 Ω 0,1 Ω to 1 Ω · CMC 0,0021 % + 0,000007 Ω 1 Ω to 10 Ω · CMC 0,0015 % + 0,00006 Ω 10 Ω to 100 Ω · CMC 0,0015 % + 0,00006 Ω 100 Ω to 1000 Ω · CMC 0,0015 % + 0,0006 Ω 1 kΩ to 10 kΩ · CMC 0,0015 % + 0,000006 kΩ 10 kΩ to 100 kΩ · CMC 0,0015 % + 0,00006 kΩ 100 kΩ to 1000 kΩ · CMC 0,0016 % + 0,0012 kΩ 1 MΩ to 10 MΩ · CMC 0,0026 % + 0,00012 MΩ 10 MΩ to 100 MΩ · CMC 0,0080 % 0,0012 MΩ 100 MΩ to 1000 MΩ · CMC 0,021 % + 0,12 MΩ 1 GΩ to 10 GΩ · CMC 0,18 % + 0,012 GΩ 10 GΩ to 100 GΩ · CMC 0,18 % + 0,012 GΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | PO-7.2/107 based on PN-EN 60477-1:2022-12 | |
| 44 | Fixed resistors | Calibration — DC resistance 0,0001 Ω · CMC 0,02 % 0,001 Ω to 1 GΩ · CMC 0,003 % 10 GΩ · CMC 0,06 % 20 GΩ · CMC 0,25 % DC resistance |
Internal procedure; Indirect method; At permanent location (S) | PO-7.2/11 based on PN-EN 60477-1:2022-12 | |
| 45 | Capacitance meters Multimeters |
Calibration — Capacitance 1 kHz — 1 pF to 10 nF · CMC 0,1 % 1 kHz — 10 nF to 110 µF · CMC 0,15 % Capacitance |
Internal procedure; At permanent location (S) | PO-7.2/105 | |
| 46 | Inductance meters Multimeters |
Calibration — Inductance 1 kHz — 10 µH to 100 µH · CMC 1,5 % 1 kHz — 100 µH to 1 mH · CMC 0,15 % 1 kHz — 1 mH to 1 H · CMC 0,05 % Inductance |
Internal procedure; At permanent location (S) | PO-7.2/105 | |
| 47 | Analogue kilovoltmeters Voltmeters |
Calibration — High voltage and current 50 Hz — 1 kV to 75 kV · CMC 0,5 % High voltage and current |
Internal procedure; Indirect method; At permanent location (S) | PO-7.2/65 | |
| 48 | Digital kilovoltmeters Voltmeters |
Calibration — High voltage and current 50 Hz — 1 kV to 75 kV · CMC 0,08 % High voltage and current |
Internal procedure; Indirect method; At permanent location (S) | PO-7.2/105 | |
| 49 | Current transformers - AC current ratio (50 Hz) |
Calibration — High voltage and current 0,1 A to 5000 A / 5 A / 1 A · CMC 0,013 % 0,1 A to 5000 A / 5 A / 1 A · CMC 1′ High voltage and current |
Internal procedure; Bridge method; At permanent location (S) and outside of it (P) | PO-7.2/62 | |
| 50 | Voltage transformers - AC voltage ratio (50 Hz) |
Calibration — High voltage and current 3 kV to 35 kV / 0,11 kV / 0,10 kV · CMC 0,013 % 3 kV to 35 kV / 0,11 kV / 0,10 kV · CMC 1' High voltage and current |
Internal procedure; Bridge method; At permanent location (S) | PO-7.2/62 | |
| 51 | Temperature indicators (meters), including temperature controllers | Calibration — Electrical simulation of quantities -270 °C to 1820 °C · CMC 0,03 °C Electrical simulation of quantities |
Internal procedure; Indirect method; At permanent location (S) | PO-7.2/111 based on EURAMET cg-11 v.2.0 | |
| 52 | Temperature simulators | Calibration — Electrical simulation of quantities -270 °C to 1820 °C · CMC 0,03 °C Electrical simulation of quantities |
Internal procedure; Indirect method; At permanent location (S) | PO-7.2/110 based on EURAMET cg-11 v.2.0 | |
| 53 | Temperature indicators (meters), including temperature controllers | Calibration — Electrical simulation of quantities -200 °C to 850 °C · CMC 0,01 °C Electrical simulation of quantities |
Internal procedure; Indirect method; At permanent location (S) | PO-7.2/111 based on EURAMET cg-11 v.2.0 | |
| 54 | Temperature simulators | Calibration — Electrical simulation of quantities -200 °C to 850 °C · CMC 0,01 °C Electrical simulation of quantities |
Internal procedure; Indirect method; At permanent location (S) | PO-7.2/110 based on EURAMET cg-11 v.2.0 | |
| 55 | Mechanical stopwatches | Calibration — Time (time interval) 0 h to 1 h · CMC 0,06 s + 8 · 10⁻⁶ · τ Time (time interval) |
Internal procedure; At permanent location (S) | PO-7.2/113 | |
| 56 | Electronic stopwatches | Calibration — Time (time interval) 0 h to 24 h · CMC 0,02 s + 8 · 10⁻⁷ · τ Time (time interval) |
Internal procedure; At permanent location (S) | PO-7.2/114 | |
| 57 | Power quality meters | Calibration — Time (time interval) 20 ms to 400 ms · CMC 0,8 ms >400 ms to 1 s · CMC 10 ms Time (time interval) |
Internal procedure; At permanent location (S) | PO-7.2/120 | |
| 58 | Non-automatic weighing instruments | Calibration — Mass (weighing instruments) up to 300 g · CMC 1 · 10⁻⁴ % 0,3 kg do 50 kg · CMC 1,5 · 10⁻³ % 50 kg do 6000 kg · CMC 1,4 · 10⁻² % Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/15 based on EURAMET cg-18 v.4.0 | |
| 59 | Mass standards of accuracy class F2 Weights of accuracy class F2 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 60 | Mass standards of accuracy class M1 Weights of accuracy class M1 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 61 | Mass standards of accuracy class M2 Weights of accuracy class M2 |
Calibration — Mass (weights and mass standards) 100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 62 | 25 kg mass standards | Calibration — Mass (weights and mass standards) 25 kg · CMC 280 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 63 | Load weights | Calibration — Mass (weights and mass standards) 1 g to 25000 g · CMC 1,2 · 10⁻³ % Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 64 | Non-automatic weighing instruments | Calibration — Mass (weighing instruments) up to 20 kg · CMC 2,9 · 10⁻³ % 20 kg to 6000 kg · CMC 1,4 · 10⁻² % Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/15 based on EURAMET cg-18 v.4.0 | |
| 65 | Mass standards of accuracy class F2 Weights of accuracy class F2 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 66 | Mass standards of accuracy class M1 Weights of accuracy class M1 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 67 | Mass standards of accuracy class M2 Weights of accuracy class M2 |
Calibration — Mass (weights and mass standards) 100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 68 | 25 kg mass standards | Calibration — Mass (weights and mass standards) 25 kg · CMC 280 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 69 | Load weights | Calibration — Mass (weights and mass standards) 1 g to 25000 g · CMC 1,2 · 10⁻³ % Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 70 | Rockwell hardness testers | Calibration — Hardness hardness — 20 HRA to 88 HRA · CMC 0,7 HRA hardness — 20 HRBW to 100 HRBW · CMC 0,8 HRBW hardness — 20 HRC to 70 HRC · CMC 0,7 HRC force — 98,07 N to 1471 N · CMC 0,26 % Hardness |
Internal procedure; Outside of permanent location (P) | PO-7.2/99 based on PN EN ISO 6508-2:2024-06 | |
| 71 | Non-automatic weighing instruments | Calibration — Mass (weighing instruments) up to 300 g · CMC 1,4 · 10⁻⁴ % 0,3 kg do 50 kg · CMC 5 · 10⁻⁴ % 50 kg do 10000 kg · CMC 1,4 · 10⁻² % Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/15 based on EURAMET cg-18 v.4.0 | |
| 72 | Mass standards of accuracy class F2 Weights of accuracy class F2 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 73 | Mass standards of accuracy class M1 Weights of accuracy class M1 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 74 | Mass standards of accuracy class M2 Weights of accuracy class M2 |
Calibration — Mass (weights and mass standards) 100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 75 | 25 kg mass standards | Calibration — Mass (weights and mass standards) 25 kg · CMC 280 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 76 | Load weights | Calibration — Mass (weights and mass standards) 1 g to 25000 g · CMC 1,2 · 10⁻³ % Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 77 | Non-automatic weighing instruments | Calibration — Mass (weighing instruments) up to 300 g · CMC 1 · 10⁻⁴ % 0,3 kg do 50 kg · CMC 1 · 10⁻³ % 50 kg do 6000 kg · CMC 1,5 · 10⁻² % Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/15 based on EURAMET cg-18 v.4.0 | |
| 78 | Mass standards of accuracy class F2 Weights of accuracy class F2 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 79 | Mass standards of accuracy class M1 Weights of accuracy class M1 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 80 | Mass standards of accuracy class M2 Weights of accuracy class M2 |
Calibration — Mass (weights and mass standards) 100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 81 | 25 kg mass standards | Calibration — Mass (weights and mass standards) 25 kg · CMC 280 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 82 | Load weights | Calibration — Mass (weights and mass standards) 1 g to 25000 g · CMC 1,2 · 10⁻³ % Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 83 | Gauge blocks of accuracy classes 0, 1, 2 | Calibration — Length 0,5 mm to 100 mm · CMC Q[0,09; 1,5 · L] µm 125 mm to 500 mm · CMC Q[0,2; 1,5 · L] µm Length |
Internal procedure; At permanent location (S) | PO-7.2/39 | |
| 84 | Optical parallels | Calibration — Length length deviation — length up to 80 mm · CMC 0,93 µm Length |
Internal procedure; At permanent location (S) | PO-7.2/67 | |
| 85 | Calipers | Calibration — Length 0 mm to 1000 mm · CMC (0,04 + 0,01 · L) mm Length |
Internal procedure; At permanent location (S) | PO-7.2/74 | |
| 86 | Caliper height gauges | Calibration — Length 0 mm to 1000 mm · CMC (0,04 + 0,01 · L) mm Length |
Internal procedure; At permanent location (S) | PO-7.2/75 | |
| 87 | Caliper depth gauges | Calibration — Length 0 mm to 500 mm · CMC (0,04 + 0,01 · L) mm Length |
Internal procedure; At permanent location (S) | PO-7.2/76 | |
| 88 | External micrometers | Calibration — Length 0 mm to 175 mm · CMC (0,9 + 2,0 · L) µm Length |
Internal procedure; At permanent location (S) | PO-7.2/77 | |
| 89 | Micrometer depth gauges | Calibration — Length 0 mm to 150 mm · CMC (4,0 + 0,3 · L) µm Length |
Internal procedure; At permanent location (S) | PO-7.2/78 | |
| 90 | Two-point internal micrometers | Calibration — Length 0 mm to 200 mm · CMC (2,6 + 1,1 · L) µm Length |
Internal procedure; At permanent location (S) | PO-7.2/114 | |
| 91 | Rigid rules Semi-rigid rules Folding rules Draper's rules |
Calibration — Length 0 mm to 1000 mm · CMC (0,15 + 0,001 · L) mm Length |
Internal procedure; At permanent location (S) | PO-7.2/2 | |
| 92 | Rigid rules Semi-rigid rules Folding rules |
Calibration — Length 0 mm to 5000 mm · CMC (0,25 + 0,01 · L) mm Length |
Internal procedure; At permanent location (S) | PO-7.2/2 | |
| 93 | Tape measures | Calibration — Length 0 mm to 5000 mm · CMC (0,29 + 0,01 · L) mm 0 mm to 10000 mm · CMC 0,68·mm 0 mm to 15000 mm · CMC 1,02·mm 0 mm to 20000 mm · CMC 1,36·mm 0 mm to 25000 mm · CMC 1,70·mm 0 mm to 30000 mm · CMC 2,04·mm Length |
Internal procedure; At permanent location (S) | PO-7.2/2 | |
| 94 | Johansson angle gauge blocks Kusznikow angle gauge blocks Wringing angle gauge blocks |
Calibration — Angle 0º to 360º · CMC 4,1” Angle |
Internal procedure; At permanent location (S) | PO-7.2/59 | |
| 95 | Universal bevel protractors | Calibration — Angle 0º to 360º · CMC 3,5’ Angle |
Internal procedure; At permanent location (S) | PO-7.2/58 | |
| 96 | 90º cylindrical squares | Calibration — Angle squareness deviation — up to 315 mm · CMC 2,40 µm Angle |
Internal procedure; At permanent location (S) | PO-7.2/81 | |
| 97 | 90º try squares | Calibration — Angle flatness and straightness deviation — up to 600 mm longer arm · CMC 3,4 µm squareness deviation — up to 600 mm longer arm · CMC 4,0 µm parallelism deviation — up to 600 mm longer arm · CMC 3,6 µm Angle |
Internal procedure; At permanent location (S) | PO-7.2/82 | |
| 98 | Bar spirit levels | Calibration — Angle flatness deviation — 0 mm/m to 1 mm/m · CMC 4,2 µm zero setting error — 0 mm/m to 1 mm/m · CMC 0,30 scale int. scale interval error — 0 mm/m to 1 mm/m · CMC 0,031 mm/m Angle |
Internal procedure; At permanent location (S) | PO-7.2/122 | |
| 99 | Builders' spirit levels | Calibration — Angle zero setting error — up to 120 cm · CMC 0,28 mm Angle |
Internal procedure; At permanent location (S) | PO-7.2/123 | |
| 100 | Optical flats | Calibration — Length (surface geometry) measuring surface — diameter up to 100 mm · CMC 0,05 µm Length (surface geometry) |
Internal procedure; At permanent location (S) | PO-7.2/66 | |
| 101 | Optical parallels | Calibration — Length (surface geometry) flatness deviation — length up to 80 mm · CMC 0,07 µm parallelism deviation — length up to 80 mm · CMC 0,26 µm Length (surface geometry) |
Internal procedure; At permanent location (S) | PO-7.2/67 | |
| 102 | Comparison roughness specimens | Calibration — Length (surface geometry) parameter Rₐ up to 12,5 µm · CMC 30 % Length (surface geometry) |
Internal procedure; At permanent location (S) | PO-7.2/80 | |
| 103 | Straight edges | Calibration — Length (surface geometry) up to 150 mm · CMC 1,4 µm 150 mm do 500 mm · CMC 2,6 µm Length (surface geometry) |
Internal procedure; At permanent location (S) | PO-7.2/79 | |
| 104 | Non-automatic weighing instruments | Calibration — Mass (weighing instruments) up to 350 g · CMC 1,4 · 10⁻⁴ % 0,35 kg to 10 kg · CMC 1,2 · 10⁻³ % 10 kg to 30 kg · CMC 3,2 · 10⁻³ % 30 kg to 600 kg · CMC 2,9 · 10⁻² % 600 kg to 6000 kg · CMC 6,4 · 10⁻² % Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/15 based on EURAMET cg-18 v.4.0 | |
| 105 | Mass standards of accuracy class F2 Weights of accuracy class F2 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 106 | Mass standards of accuracy class M1 Weights of accuracy class M1 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 107 | Mass standards of accuracy class M2 Weights of accuracy class M2 |
Calibration — Mass (weights and mass standards) 100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 108 | 25 kg mass standards | Calibration — Mass (weights and mass standards) 25 kg · CMC 280 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 109 | Load weights | Calibration — Mass (weights and mass standards) 1 g to 25000 g · CMC 1,2 · 10⁻³ % Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 110 | Bourdon tube pressure gauges - gauge pressure (gas) |
Calibration — Pressure -0,1 MPa do -0,005 MPa · CMC 0,0002 MPa 0,005 MPa to 0,25 MPa · CMC 0,0002 MPa Pressure |
Internal procedure; At permanent location (S) | PO-7.2/1 | |
| 111 | Bourdon tube pressure gauges - gauge pressure (liquid) |
Calibration — Pressure 0,02 MPa do 0,6 MPa · CMC 0,0005 MPa 0,6 MPa to 6 MPa · CMC 0,006 MPa 6 MPa do 60 MPa · CMC 0,07 MPa Pressure |
Internal procedure; At permanent location (S) | PO-7.2/1 | |
| 112 | Non-automatic weighing instruments | Calibration — Mass (weighing instruments) up to 300 g · CMC 1,4 · 10⁻⁴ % 0,3 kg do 50 kg · CMC 5 · 10⁻⁴ % 50 kg do 6000 kg · CMC 1,4 · 10⁻² % Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | PO-7.2/15 based on EURAMET cg-18 v.4.0 | |
| 113 | Mass standards of accuracy class F2 Weights of accuracy class F2 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,015 mg 2 mg · CMC 0,015 mg 5 mg · CMC 0,015 mg 10 mg · CMC 0,020 mg 20 mg · CMC 0,025 mg 50 mg · CMC 0,03 mg 100 mg · CMC 0,04 mg 200 mg · CMC 0,05 mg 500 mg · CMC 0,06 mg 1g · CMC 0,07 mg 2g · CMC 0,10 mg 5g · CMC 0,12 mg 10 g · CMC 0,15 mg 20 g · CMC 0,20 mg 50 g · CMC 0,25 mg 100 g · CMC 0,40 mg 200 g · CMC 0,75 mg 500 g · CMC 2,0 mg 1 kg · CMC 4,0 mg 2 kg · CMC 7,5 mg 5 kg · CMC 20 mg 10 kg · CMC 40 mg 20 kg · CMC 75 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 114 | Mass standards of accuracy class M1 Weights of accuracy class M1 |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,05 mg 2 mg · CMC 0,05 mg 5 mg · CMC 0,05 mg 10 mg · CMC 0,06 mg 20 mg · CMC 0,07 mg 50 mg · CMC 0,10 mg 100 mg · CMC 0,12 mg 200 mg · CMC 0,15 mg 500 mg · CMC 0,20 mg 1g · CMC 0,25 mg 2g · CMC 0,30 mg 5g · CMC 0,40 mg 10 g · CMC 0,50 mg 20 g · CMC 0,60 mg 50 g · CMC 0,70 mg 100 g · CMC 1,25 mg 200 g · CMC 2,5 mg 500 g · CMC 6,0 mg 1 kg · CMC 12,5 mg 2 kg · CMC 25 mg 5 kg · CMC 60 mg 10 kg · CMC 125 mg 20 kg · CMC 250 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 115 | Mass standards of accuracy class M2 Weights of accuracy class M2 |
Calibration — Mass (weights and mass standards) 100 mg · CMC 0,40 mg 200 mg · CMC 0,50 mg 500 mg · CMC 0,60 mg 1g · CMC 0,75 mg 2g · CMC 1,00 mg 5g · CMC 1,25 mg 10 g · CMC 1,5 mg 20 g · CMC 2,0 mg 50 g · CMC 2,5 mg 100 g · CMC 4,0 mg 200 g · CMC 7,5 mg 500 g · CMC 20 mg 1 kg · CMC 40 mg 2 kg · CMC 75 mg 5 kg · CMC 200 mg 10 kg · CMC 400 mg 20 kg · CMC 750 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 116 | 25 kg mass standards | Calibration — Mass (weights and mass standards) 25 kg · CMC 280 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
| 117 | Load weights | Calibration — Mass (weights and mass standards) 1 g to 25000 g · CMC 1,2 · 10⁻³ % Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | PO-7.2/88 based on OIML R111-1:2004 Annex C | |
Reference standards (75)
ASTM E10-23
ASTM E18-25
ASTM E2658-15 (2023)
ASTM E384-22
ASTM E4-24
ASTM E92-23
EURAMET cg-11 v.2.0
EURAMET cg-15 v.3.0
EURAMET cg-18 v.4.0
EURAMET cg-8 v.3.1
OIML R111-1:2004 Załącznik C
PN EN ISO 12390-4:2025-08 cl. 4.4
PN EN ISO 376:2011
PN EN ISO 4545-2:2018-04
PN EN ISO 6506-2:2019-10
PN EN ISO 6507-2:2018-05
PN EN ISO 6508-2:2024-06
PN-EN 60477-1:2022-12
PN-EN ISO 148-2:2017-02
PN-EN ISO 7500-1:2018-05
PN-EN ISO 9513:2013-06
PO-7.2/1
PO-7.2/100
PO-7.2/101
PO-7.2/102
PO-7.2/103
PO-7.2/104
PO-7.2/105
PO-7.2/107
PO-7.2/11
PO-7.2/110
PO-7.2/111
PO-7.2/112
PO-7.2/113
PO-7.2/114
PO-7.2/117
PO-7.2/120
PO-7.2/122
PO-7.2/123
PO-7.2/126
PO-7.2/127
PO-7.2/129
PO-7.2/130
PO-7.2/15
PO-7.2/17
PO-7.2/2
PO-7.2/37
PO-7.2/38
PO-7.2/39
PO-7.2/42
PO-7.2/46
PO-7.2/51
PO-7.2/58
PO-7.2/59
PO-7.2/62
PO-7.2/65
PO-7.2/66
PO-7.2/67
PO-7.2/68
PO-7.2/71
PO-7.2/74
PO-7.2/75
PO-7.2/76
PO-7.2/77
PO-7.2/78
PO-7.2/79
PO-7.2/80
PO-7.2/81
PO-7.2/82
PO-7.2/87
PO-7.2/88
PO-7.2/93
PO-7.2/97
PO-7.2/98
PO-7.2/99