🇵🇱 CZAH-POMIAR Sp. z o.o. PL
Metrology & calibration
CZAH-POMIAR Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 062 (Katowice, Poland). Its accreditation scope covers 16 calibration items in the following areas: Metrology & calibration. Calibrated items include: Noble metal thermocouples (type S, R, B), Electrical thermometers (including electronic), Base metal thermocouples (type J, T, K, E, N), Industrial resistance thermometer sensors. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; At ITS-90 fixed points: freezing point of zinc,….
At a glance
- Accreditation number: PCA AP 062
- Calibration items in scope: 16
- Types of calibrated items: 17
- Calibration principles used: 7
- City: Katowice, Poland
- Valid until: 25-01-2028
- Accreditation number
- AP 062
- Accreditation body
- PCA
- Address
- ul. Porcelanowa 25, 40-241 Katowice
- City
- Katowice, Poland
- Division
- Temperature Measurement Laboratory
- Valid until
- 25-01-2028
What this laboratory calibrates (17)
- Noble metal thermocouples (type S, R, B) — 3 calibration items
- Electrical thermometers (including electronic) — 2 calibration items
- Base metal thermocouples (type J, T, K, E, N) — 1 calibration item
- Industrial resistance thermometer sensors — 1 calibration item
- Line scanners — 1 calibration item
- Long standard platinum resistance thermometer (SPRT) sensors — 1 calibration item
- Noble metal thermocouples — 1 calibration item
- Photoelectric pyrometers — 1 calibration item
- Pyrometers — 1 calibration item
- Radiation pyrometers — 1 calibration item
- Temperature indicators (meters) compatible with resistance sensors — 1 calibration item
- Temperature indicators (meters) compatible with thermoelectric sensors — 1 calibration item
- Temperature simulators compatible with resistance sensors — 1 calibration item
- Temperature simulators compatible with thermoelectric sensors — 1 calibration item
- Temperature transmitters compatible with resistance sensors — 1 calibration item
- and 2 more
Calibration principles (7)
- Internal procedure; At permanent location (S) — 9 calibration items
- Internal procedure; At permanent location (S) and outside of it (P) — 2 calibration items
- Internal procedure; At ITS-90 fixed points: freezing point of zinc, freezing point of aluminium, freezing point of copper; At permanent location (S) — 1 calibration item
- Internal procedure; At ITS-90 fixed points: triple point of mercury, triple point of water, melting point of gallium, freezing point of tin, freezing point of zinc, freezing point of aluminium; At permanent location (S) — 1 calibration item
- Internal procedure; Outside of permanent location (P) — 1 calibration item
- Internal procedure; Wire method At the melting point of gold; At permanent location (S) — 1 calibration item
- Internal procedure; Wire method At the melting point of palladium; At permanent location (S) — 1 calibration item
Certificate information
- Issue date
- 26.11.2025
- Issue no.
- 20
- Certificate date
- 15.01.2020
- Accreditation cycle
- 05.01.2024 — 25.01.2028
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (16)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Temperature indicators (meters) compatible with thermoelectric sensors | Calibration — Electric quantity simulation -270 °C to 1820 °C · CMC 0,7 °C Electric quantity simulation |
Internal procedure; At permanent location (S) and outside of it (P) | PT-19.01/5 PT-19.01/6 based on EURAMET cg-11 v.2.0 | |
| 2 | Temperature indicators (meters) compatible with resistance sensors | Calibration — Electric quantity simulation -200 °C to 850 °C · CMC 0,7 °C Electric quantity simulation |
Internal procedure; At permanent location (S) and outside of it (P) | PT-19.01/5 PT-19.01/6 based on EURAMET cg-11 v.2.0 | |
| 3 | Temperature simulators compatible with thermoelectric sensors | Calibration — Electric quantity simulation -270 °C to 1820 °C · CMC 0,7 °C Electric quantity simulation |
Internal procedure; At permanent location (S) | PT-19.01/5 based on EURAMET cg-11 v.2.0 | |
| 4 | Temperature simulators compatible with resistance sensors | Calibration — Electric quantity simulation -200 °C to 850 °C · CMC 0,7 °C Electric quantity simulation |
Internal procedure; At permanent location (S) | PT-19.01/5 based on EURAMET cg-11 v.2.0 | |
| 5 | Temperature transmitters compatible with thermoelectric sensors | Calibration — Electric quantity simulation type R, S, B — 0 °C to 1700 °C · CMC 0,7 °C type J, K, N — -100 °C to 1300 °C · CMC 0,2 °C Electric quantity simulation |
Internal procedure; At permanent location (S) | PT-7.15/3 based on EURAMET cg-11 v.2.0 | |
| 6 | Temperature transmitters compatible with resistance sensors | Calibration — Electric quantity simulation -200 °C to 800 °C · CMC 0,2 °C Electric quantity simulation |
Internal procedure; At permanent location (S) | PT-7.15/3 based on EURAMET cg-11 v.2.0 | |
| 7 | Base metal thermocouples (type J, T, K, E, N) | Calibration — Temperature (electrical thermometry) 0 °C to 1200 °C · CMC 1,1 °C 1200 °C to 1260 °C · CMC 1,4 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PT-19.01/2 based on EURAMET cg-8 v.3.1 | |
| 8 | Noble metal thermocouples (type S, R, B) | Calibration — Temperature (electrical thermometry) Zn (419,527) °C · CMC 0,19 °C Al (660,323) °C · CMC 0,19 °C Cu (1084,62) °C · CMC 0,23 °C Temperature (electrical thermometry) |
Internal procedure; At ITS-90 fixed points: freezing point of zinc, freezing point of aluminium, freezing point of copper; At permanent location (S) | PT-19.01/14 | |
| 9 | Noble metal thermocouples (type S, R, B) | Calibration — Temperature (electrical thermometry) Au (1064,18) °C · CMC 0,45 °C Temperature (electrical thermometry) |
Internal procedure; Wire method At the melting point of gold; At permanent location (S) | PT-19.01/14 | |
| 10 | Noble metal thermocouples (type S, R, B) | Calibration — Temperature (electrical thermometry) Pd (1553,5) °C · CMC 1,60 °C Temperature (electrical thermometry) |
Internal procedure; Wire method At the melting point of palladium; At permanent location (S) | PT-19.01/14 | |
| 11 | Noble metal thermocouples | Calibration — Temperature (electrical thermometry) type S and R — 0 °C to 1085 °C · CMC 0,6 °C type S and R — 1085 °C to 1200 °C · CMC 1,0 °C type S and R — 1200 °C to 1300 °C · CMC 1,3 °C type S and R — 1300 °C to 1500 °C · CMC 1,8 °C type B — 400 °C to 1085 °C · CMC 1,0 °C type B — 1085 °C to 1500 °C · CMC 1,8 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PT-19.01/1 based on EURAMET cg-8 v.3.1 | |
| 12 | Industrial resistance thermometer sensors | Calibration — Temperature (electrical thermometry) -40 °C to 0 °C · CMC 0,03 °C 0 °C to 300 °C · CMC 0,11 °C 300 °C to 550 °C · CMC 0,22 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PT-19.01/7 based on DKD-R 5-1 (09/2018) | |
| 13 | Electrical thermometers (including electronic) | Calibration — Temperature (electrical thermometry) -20 °C to 100 °C · CMC 0,2 °C 100 °C do 650 °C · CMC 0,6 °C 650 °C do 1200 °C · CMC 1,2 °C 1200 °C do 1500 °C · CMC 2,0 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PT-19.01/3 PT-19.01/4 based on EURAMET cg-8 v.3.1 and DKD-R 5-1 (09/2018) | |
| 14 | Electrical thermometers (including electronic) | Calibration — Temperature (electrical thermometry) 50 °C to 650 °C · CMC 0,6 °C Temperature (electrical thermometry) |
Internal procedure; Outside of permanent location (P) | PT-19.01/3 PT-19.01/4 based on EURAMET cg-8 v.3.1 and DKD-R 5-1 (09/2018) | |
| 15 | Long standard platinum resistance thermometer (SPRT) sensors | Calibration — Temperature (electrical thermometry) Hg (-38,8344) °C · CMC 1,2 mK H₂O (0,01) °C · CMC 1,0 mK Ga (29,7646) °C · CMC 1,5 mK Sn (231,928) °C · CMC 1,6 mK Zn (419,527) °C · CMC 1,8 mK Al (660,323) °C · CMC 4,4 mK Temperature (electrical thermometry) |
Internal procedure; At ITS-90 fixed points: triple point of mercury, triple point of water, melting point of gallium, freezing point of tin, freezing point of zinc, freezing point of aluminium; At permanent location (S) | PT-19.01/15 | |
| 16 | Pyrometers Radiation pyrometers Photoelectric pyrometers Thermal imaging cameras Line scanners |
Calibration — Temperature (radiation thermometry) -30 °C do 100 °C · CMC 1,1 °C 100 °C to 200 °C · CMC 2,6 °C 200 °C do 300 °C · CMC 3,1 °C 300 °C do 500 °C · CMC 4,0 °C Temperature (radiation thermometry) |
Internal procedure; At permanent location (S) | PT-19.03/1 based on MSL Technical Guide 22 v.3 and VDI/VDE 3511 Part 4.4 | |
Reference standards (16)
DKD-R 5-1 (09/2018)
EURAMET cg-11 v.2.0
EURAMET cg-8 v.3.1
MSL Technical Guide 22 v.3
PT-19.01/1
PT-19.01/14
PT-19.01/15
PT-19.01/2
PT-19.01/3
PT-19.01/4
PT-19.01/5
PT-19.01/6
PT-19.01/7
PT-19.03/1
PT-7.15/3
VDI/VDE 3511 Part 4.4