🇵🇱 CZAH-POMIAR Sp. z o.o. PL

PCA AP 062 • Katowice, Poland • 16 calibration items

Metrology & calibration

CZAH-POMIAR Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 062 (Katowice, Poland). Its accreditation scope covers 16 calibration items in the following areas: Metrology & calibration. Calibrated items include: Noble metal thermocouples (type S, R, B), Electrical thermometers (including electronic), Base metal thermocouples (type J, T, K, E, N), Industrial resistance thermometer sensors. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; At ITS-90 fixed points: freezing point of zinc,….

At a glance

  • Accreditation number: PCA AP 062
  • Calibration items in scope: 16
  • Types of calibrated items: 17
  • Calibration principles used: 7
  • City: Katowice, Poland
  • Valid until: 25-01-2028
Accreditation number
AP 062
Accreditation body
PCA
Address
ul. Porcelanowa 25, 40-241 Katowice
City
Katowice, Poland
Division
Temperature Measurement Laboratory
Valid until
25-01-2028

What this laboratory calibrates (17)

Calibration principles (7)

Certificate information

Issue date
26.11.2025
Issue no.
20
Certificate date
15.01.2020
Accreditation cycle
05.01.2024 — 25.01.2028
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (16)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Temperature indicators (meters) compatible with thermoelectric sensors Calibration — Electric quantity simulation
-270 °C to 1820 °C · CMC 0,7 °C
Electric quantity simulation
Internal procedure; At permanent location (S) and outside of it (P) PT-19.01/5 PT-19.01/6 based on EURAMET cg-11 v.2.0
2 Temperature indicators (meters) compatible with resistance sensors Calibration — Electric quantity simulation
-200 °C to 850 °C · CMC 0,7 °C
Electric quantity simulation
Internal procedure; At permanent location (S) and outside of it (P) PT-19.01/5 PT-19.01/6 based on EURAMET cg-11 v.2.0
3 Temperature simulators compatible with thermoelectric sensors Calibration — Electric quantity simulation
-270 °C to 1820 °C · CMC 0,7 °C
Electric quantity simulation
Internal procedure; At permanent location (S) PT-19.01/5 based on EURAMET cg-11 v.2.0
4 Temperature simulators compatible with resistance sensors Calibration — Electric quantity simulation
-200 °C to 850 °C · CMC 0,7 °C
Electric quantity simulation
Internal procedure; At permanent location (S) PT-19.01/5 based on EURAMET cg-11 v.2.0
5 Temperature transmitters compatible with thermoelectric sensors Calibration — Electric quantity simulation
type R, S, B — 0 °C to 1700 °C · CMC 0,7 °C type J, K, N — -100 °C to 1300 °C · CMC 0,2 °C
Electric quantity simulation
Internal procedure; At permanent location (S) PT-7.15/3 based on EURAMET cg-11 v.2.0
6 Temperature transmitters compatible with resistance sensors Calibration — Electric quantity simulation
-200 °C to 800 °C · CMC 0,2 °C
Electric quantity simulation
Internal procedure; At permanent location (S) PT-7.15/3 based on EURAMET cg-11 v.2.0
7 Base metal thermocouples (type J, T, K, E, N) Calibration — Temperature (electrical thermometry)
0 °C to 1200 °C · CMC 1,1 °C 1200 °C to 1260 °C · CMC 1,4 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PT-19.01/2 based on EURAMET cg-8 v.3.1
8 Noble metal thermocouples (type S, R, B) Calibration — Temperature (electrical thermometry)
Zn (419,527) °C · CMC 0,19 °C Al (660,323) °C · CMC 0,19 °C Cu (1084,62) °C · CMC 0,23 °C
Temperature (electrical thermometry)
Internal procedure; At ITS-90 fixed points: freezing point of zinc, freezing point of aluminium, freezing point of copper; At permanent location (S) PT-19.01/14
9 Noble metal thermocouples (type S, R, B) Calibration — Temperature (electrical thermometry)
Au (1064,18) °C · CMC 0,45 °C
Temperature (electrical thermometry)
Internal procedure; Wire method At the melting point of gold; At permanent location (S) PT-19.01/14
10 Noble metal thermocouples (type S, R, B) Calibration — Temperature (electrical thermometry)
Pd (1553,5) °C · CMC 1,60 °C
Temperature (electrical thermometry)
Internal procedure; Wire method At the melting point of palladium; At permanent location (S) PT-19.01/14
11 Noble metal thermocouples Calibration — Temperature (electrical thermometry)
type S and R — 0 °C to 1085 °C · CMC 0,6 °C type S and R — 1085 °C to 1200 °C · CMC 1,0 °C type S and R — 1200 °C to 1300 °C · CMC 1,3 °C type S and R — 1300 °C to 1500 °C · CMC 1,8 °C type B — 400 °C to 1085 °C · CMC 1,0 °C type B — 1085 °C to 1500 °C · CMC 1,8 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PT-19.01/1 based on EURAMET cg-8 v.3.1
12 Industrial resistance thermometer sensors Calibration — Temperature (electrical thermometry)
-40 °C to 0 °C · CMC 0,03 °C 0 °C to 300 °C · CMC 0,11 °C 300 °C to 550 °C · CMC 0,22 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PT-19.01/7 based on DKD-R 5-1 (09/2018)
13 Electrical thermometers (including electronic) Calibration — Temperature (electrical thermometry)
-20 °C to 100 °C · CMC 0,2 °C 100 °C do 650 °C · CMC 0,6 °C 650 °C do 1200 °C · CMC 1,2 °C 1200 °C do 1500 °C · CMC 2,0 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PT-19.01/3 PT-19.01/4 based on EURAMET cg-8 v.3.1 and DKD-R 5-1 (09/2018)
14 Electrical thermometers (including electronic) Calibration — Temperature (electrical thermometry)
50 °C to 650 °C · CMC 0,6 °C
Temperature (electrical thermometry)
Internal procedure; Outside of permanent location (P) PT-19.01/3 PT-19.01/4 based on EURAMET cg-8 v.3.1 and DKD-R 5-1 (09/2018)
15 Long standard platinum resistance thermometer (SPRT) sensors Calibration — Temperature (electrical thermometry)
Hg (-38,8344) °C · CMC 1,2 mK H₂O (0,01) °C · CMC 1,0 mK Ga (29,7646) °C · CMC 1,5 mK Sn (231,928) °C · CMC 1,6 mK Zn (419,527) °C · CMC 1,8 mK Al (660,323) °C · CMC 4,4 mK
Temperature (electrical thermometry)
Internal procedure; At ITS-90 fixed points: triple point of mercury, triple point of water, melting point of gallium, freezing point of tin, freezing point of zinc, freezing point of aluminium; At permanent location (S) PT-19.01/15
16 Pyrometers
Radiation pyrometers
Photoelectric pyrometers
Thermal imaging cameras
Line scanners
Calibration — Temperature (radiation thermometry)
-30 °C do 100 °C · CMC 1,1 °C 100 °C to 200 °C · CMC 2,6 °C 200 °C do 300 °C · CMC 3,1 °C 300 °C do 500 °C · CMC 4,0 °C
Temperature (radiation thermometry)
Internal procedure; At permanent location (S) PT-19.03/1 based on MSL Technical Guide 22 v.3 and VDI/VDE 3511 Part 4.4

Reference standards (16)

DKD-R 5-1 (09/2018)
EURAMET cg-11 v.2.0
EURAMET cg-8 v.3.1
MSL Technical Guide 22 v.3
PT-19.01/1
PT-19.01/14
PT-19.01/15
PT-19.01/2
PT-19.01/3
PT-19.01/4
PT-19.01/5
PT-19.01/6
PT-19.01/7
PT-19.03/1
PT-7.15/3
VDI/VDE 3511 Part 4.4