🇵🇱 Centralne Laboratorium Ochrony Radiologicznej PL

PCA AP 057 • Warszawa, Poland • 12 calibration items

Metrology & calibration

Centralne Laboratorium Ochrony Radiologicznej is a calibration laboratory accredited by PCA under number AP 057 (Warszawa, Poland). Its accreditation scope covers 12 calibration items in the following areas: Metrology & calibration. Calibrated items include: Ionising radiation dosimeters, TLD dosimeters, Charcoal detectors, Dosimetric instruments for beta radiation measurement. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Beta radiation measurement; At permanent locati…, Internal procedure; Gamma radiation measurement; At permanent locat….

At a glance

  • Accreditation number: PCA AP 057
  • Calibration items in scope: 12
  • Types of calibrated items: 12
  • Calibration principles used: 6
  • City: Warszawa, Poland
  • Valid until: 06-11-2027
Accreditation number
AP 057
Accreditation body
PCA
Address
ul. Konwaliowa 7, 03-194 Warszawa
City
Warszawa, Poland
Division
Calibration Laboratory for Dosimetric and Radon Instruments
Valid until
06-11-2027

What this laboratory calibrates (12)

Calibration principles (6)

Certificate information

Issue date
11.09.2026
Issue no.
22
Certificate date
10.10.2019
Accreditation cycle
16.10.2023 — 06.11.2027
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (12)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 %. A value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Dosimetric instruments for gamma radiation measurement Calibration — Dosimetric quantities
- air kerma rate; Cs-137, Co-60 — 1 µGy/h to 10 µGy/h · CMC 8,0 % - exposure rate; Cs-137, Co-60 — 0,1 mR/h to 1 mR/h · CMC 8,0 % - absorbed dose rate in air; Cs-137, Co-60 — 1 µGy/h to 10 µGy/h · CMC 8,0 % - dose equivalent rate; Cs-137, Co-60 — 1 µSv/h to 10 µSv/h · CMC 8,0 % - photon dose equivalent rate; Cs-137, Co-60 — 1 µSv/h to 10 µSv/h · CMC 8,0 % - ambient dose equivalent rate; Cs-137, Co-60 — 1 µSv/h to 10 µSv/h · CMC 8,0 % - personal dose equivalent rate; Cs-137, Co-60 — 1 µSv/h to 10 µSv/h · CMC 8,0 % - air kerma rate; Cs-137 — 0,01 mGy/h to 80 mGy/h · CMC 4,0 % - air kerma rate; Co-60 — 0,01 mGy/h to 4 mGy/h · CMC 5,0 % - air kerma rate; Am-241 — 15 µGy/h to 20 µGy/h · CMC 5,0 % - exposure rate; Cs-137 — 1 mR/h to 9000 mR/h · CMC 4,0 % - exposure rate; Co-60 — 1 mR/h to 500 mR/h · CMC 5,0 % - exposure rate; Am-241 — 2 mR/h to 2 mR/h · CMC 5,0 % - absorbed dose rate in air; Cs-137 — 0,01 mGy/h to 60 mGy/h · CMC 4,0 % - absorbed dose rate in air; Co-60 — 0,01 mGy/h to 5 mGy/h · CMC 5,0 % - absorbed dose rate in air; Am-241 — 15 µGy/h to 20 µGy/h · CMC 5,0 % - dose equivalent rate; Cs-137 — 0,01 mSv/h to 80 mSv/h · CMC 4,0 % - dose equivalent rate; Co-60 — 0,01 mSv/h to 5 mSv/h · CMC 5,0 % - dose equivalent rate; Am-241 — 15 µSv/h to 20 µSv/h · CMC 5,0 % - ambient dose equivalent rate; Cs-137 — 0,01 mSv/h to 95 mSv/h · CMC 4,0 % - ambient dose equivalent rate; Co-60 — 0,01 mSv/h to 5 mSv/h · CMC 5,0 % - ambient dose equivalent rate; Am-241 — 30 µSv/h to 40 µSv/h · CMC 5,0 % - personal dose equivalent rate; Cs-137 — 0,01 mSv/h to 95 mSv/h · CMC 4,0 % - personal dose equivalent rate; Co-60 — 0,01 mSv/h to 5 mSv/h · CMC 5,0 % - personal dose equivalent rate; Am-241 — 30 µSv/h to 40 µSv/h · CMC 5,0 % - air kerma; Cs-137 — 0,0003 mGy to 460 mGy · CMC 4,0 % - air kerma; Co-60 — 0,0002 mGy to 32 mGy · CMC 5,0 % - air kerma; Am-241 — 0,0004 mGy to 180 mGy · CMC 5,0 % - exposure; Cs-137 — 0,003 mR to 50000 mR · CMC 4,0 % - exposure; Co-60 — 0,002 mR to 4000 mR · CMC 5,0 % - exposure; Am-241 — 0,005 mR to 20 mR · CMC 5,0 % - absorbed dose in air; Cs-137 — 0,0003 mGy to 460 mGy · CMC 4,0 % - absorbed dose in air; Co-60 — 0,0002 mGy to 40 mGy · CMC 5,0 % - absorbed dose in air; Am-241 — 0,0004 mGy to 180 mGy · CMC 5,0 % - dose equivalent; Cs-137 — 0,0003 mSv to 460 mSv · CMC 4,0 % - dose equivalent; Co-60 — 0,0002 mSv to 40 mSv · CMC 5,0 % - dose equivalent; Am-241 — 0,0004 mSv to 180 mSv · CMC 5,0 % - ambient dose equivalent; Cs-137 — 0,0003 mSv to 550 mSv · CMC 4,0 % - ambient dose equivalent; Co-60 — 0,0002 mSv to 40 mSv · CMC 5,0 % - ambient dose equivalent; Am-241 — 0,7 µSv to 300 µSv · CMC 5,0 % - personal dose equivalent; Cs-137 — 0,0003 mSv to 550 mSv · CMC 4,0 % - personal dose equivalent; Co-60 — 0,0002 mSv to 40 mSv · CMC 5,0 % - personal dose equivalent; Am-241 — 0,8 µSv to 340 µSv · CMC 5,0 %
Dosimetric quantities
Internal procedure; At permanent location (S) PW 1G
2 TLD dosimeters
Ionising radiation dosimeters
Calibration — Dosimetric quantities
- air kerma; Cs-137 — 0,0003 mGy to 460 mGy · CMC 4,0 % - air kerma; Co-60 — 0,0002 mGy to 32 mGy · CMC 5,0 % - air kerma; Am-241 — 0,0004 mGy to 180 mGy · CMC 5,0 % - exposure; Cs-137 — 0,003 mR to 50000 mR · CMC 4,0 % - exposure; Co-60 — 0,002 mR to 4000 mR · CMC 5,0 % - exposure; Am-241 — 0,005 mR to 20 mR · CMC 5,0 % - absorbed dose in air; Cs-137 — 0,0003 mGy to 460 mGy · CMC 4,0 % - absorbed dose in air; Co-60 — 0,0002 mGy to 40 mGy · CMC 5,0 % - absorbed dose in air; Am-241 — 0,0004 mGy to 180 mGy · CMC 5,0 % - dose equivalent; Cs-137 — 0,0003 mSv to 460 mSv · CMC 4,0 % - dose equivalent; Co-60 — 0,0002 mSv to 40 mSv · CMC 5,0 % - dose equivalent; Am-241 — 0,0004 mSv to 180 mSv · CMC 5,0 % - ambient dose equivalent; Cs-137 — 0,0003 mSv to 550 mSv · CMC 4,0 % - ambient dose equivalent; Co-60 — 0,0002 mSv to 40 mSv · CMC 5,0 % - ambient dose equivalent; Am-241 — 0,7 µSv to 300 µSv · CMC 5,0 % - personal dose equivalent; Cs-137 — 0,0003 mSv to 550 mSv · CMC 4,0 % - personal dose equivalent; Co-60 — 0,0002 mSv to 40 mSv · CMC 5,0 % - personal dose equivalent; Am-241 — 0,8 µSv to 340 µSv · CMC 5,0 %
Dosimetric quantities
Internal procedure; Gamma radiation measurement; At permanent location (S) PW 1G
3 Dosimetric instruments for X-radiation measurement Calibration — Dosimetric quantities
- air kerma rate — 0,003 mGy/h to 114 mGy/h · CMC 4,5 % - exposure rate — 0,3 mR/h to 11000 mR/h · CMC 4,5 % - absorbed dose rate in air — 0,003 mGy/h to 114 mGy/h · CMC 4,5 % - dose equivalent rate — 0,003 mSv/h to 114 mSv/h · CMC 4,5 % - ambient dose equivalent rate — 0,004 mSv/h to 180 mSv/h · CMC 4,5 % - personal dose equivalent rate — 0,004 mSv/h to 200 mSv/h · CMC 4,5 % - air kerma — 0,0001 mGy to 1000 mGy · CMC 4,5 % - exposure — 0,001 mR to 91000 mR · CMC 4,5 % - absorbed dose in air — 0,0001 mGy to 1000 mGy · CMC 4,5 % - dose equivalent — 0,0001 mSv to 1000 mSv · CMC 4,5 % - ambient dose equivalent — 0,0001 mSv to 1400 mSv · CMC 4,5 % - personal dose equivalent — 0,0001 mSv to 1600 mSv · CMC 4,5 %
Dosimetric quantities
Internal procedure; X-radiation for spectra N-40 to N-300; At permanent location (S) PW 2X
4 TLD dosimeters
Ionising radiation dosimeters
Calibration — Dosimetric quantities
- air kerma — 0,0001 mGy to 1000 mGy · CMC 4,5 % - exposure — 0,001 mR to 91000 mR · CMC 4,5 % - absorbed dose in air — 0,0001 mGy to 1000 mGy · CMC 4,5 % - dose equivalent — 0,0001 mSv to 1000 mSv · CMC 4,5 % - ambient dose equivalent — 0,0001 mSv to 1400 mSv · CMC 4,5 % - personal dose equivalent — 0,0001 mSv to 1600 mSv · CMC 4,5 %
Dosimetric quantities
Internal procedure; X-radiation measurement for spectra N-40 to N-300; At permanent location (S) PW 2X
5 Dosimetric instruments for beta radiation measurement Calibration — Dosimetric quantities
- absorbed dose rate in air; Sr-90 — 14 mGy/h to 300 mGy/h · CMC 8,0 % - absorbed dose rate in air; Kr-85 — 60 mGy/h · CMC 8,0 % - directional dose equivalent rate; Sr-90 — 14 mSv/h to 300 mSv/h · CMC 8,0 % - directional dose equivalent rate; Kr-85 — 57 mSv/h · CMC 8,0 % - personal dose equivalent rate; Sr-90 — 14 mSv/h to 300 mSv/h · CMC 8,0 % - personal dose equivalent rate; Kr-85 — 57 mSv/h · CMC 8,0 % - absorbed dose in air; Sr-90 — 0,01 mGy to 5000 mGy · CMC 8,0 % - absorbed dose in air; Kr-85 — 0,01 mGy to 500 mGy · CMC 8,0 % - directional dose equivalent; Sr-90 — 0,01 mSv to 2500 mSv · CMC 8,0 % - directional dose equivalent; Kr-85 — 0,01 mSv to 250 mSv · CMC 8,0 % - personal dose equivalent; Sr-90 — 0,01 mSv to 2500 mSv · CMC 8,0 % - personal dose equivalent; Kr-85 — 0,01 mSv to 250 mSv · CMC 8,0 %
Dosimetric quantities
Internal procedure; At permanent location (S) PW 6B
6 TLD dosimeters
Ionising radiation dosimeters
Calibration — Dosimetric quantities
- absorbed dose in air; Sr-90 — 0,01 mGy to 5000 mGy · CMC 8,0 % - absorbed dose in air; Kr-85 — 0,01 mGy to 500 mGy · CMC 8,0 % - directional dose equivalent; Sr-90 — 0,01 mSv to 2500 mSv · CMC 8,0 % - directional dose equivalent; Kr-85 — 0,01 mSv to 250 mSv · CMC 8,0 % - personal dose equivalent; Sr-90 — 0,01 mSv to 2500 mSv · CMC 8,0 % - personal dose equivalent; Kr-85 — 0,01 mSv to 250 mSv · CMC 8,0 %
Dosimetric quantities
Internal procedure; Beta radiation measurement; At permanent location (S) PW 6B
7 Dosimetric instruments for neutron radiation measurement Calibration — Dosimetric quantities
- ambient dose equivalent rate; ²⁴¹Am/Be — 10 µSv/h to 200 µSv/h · CMC 17,0 % - personal dose equivalent rate; ²⁴¹Am/Be — 10 µSv/h to 200 µSv/h · CMC 17,0 % - ambient dose equivalent; ²⁴¹Am/Be — 0,005 mSv to 20 mSv · CMC 17,0 % - personal dose equivalent; ²⁴¹Am/Be — 0,005 mSv to 20 mSv · CMC 17,0 %
Dosimetric quantities
Internal procedure; At permanent location (S) PW 4N
8 TLD dosimeters
Ionising radiation dosimeters
Calibration — Dosimetric quantities
- ambient dose equivalent; ²⁴¹Am/Be — 0,005 mSv to 20 mSv · CMC 17,0 % - personal dose equivalent; ²⁴¹Am/Be — 0,005 mSv to 20 mSv · CMC 17,0 %
Dosimetric quantities
Internal procedure; Neutron radiation measurement PW 4N
9 Meters and monitors for alpha/ beta surface contamination measurement Calibration — Surface emission of radiation
Am-241 — 3,59 s⁻¹cm⁻² · CMC 8,0 % C-14 — 2,38 s⁻¹cm⁻² · CMC 8,0 % Cl-36 — 4,05 s⁻¹cm⁻² · CMC 7,0 % Sr-90 — 6,74 s⁻¹cm⁻² · CMC 7,0 %
Surface emission of radiation
Internal procedure; At permanent location (S) PW 3S
10 Instruments for measuring radon activity concentration in air Calibration — Radon measurements
Rn-222 — 50 Bq/m³ to 1 000 Bq/m³ · CMC 6,5 % Rn-222 — 1 kBq/m³ to 50 kBq/m³ · CMC 6,1 %
Radon measurements
Internal procedure; At permanent location (S) PW 5R
11 Instruments for measuring potential alpha energy concentration Calibration — Radon measurements
Rn-222 — 1 µJ/m³ to 250 µJ/m³ · CMC 6,0 %
Radon measurements
Internal procedure; At permanent location (S) PW 5R
12 Track detectors
Charcoal detectors
Electrets
Calibration — Radon measurements
Rn-222 — 6 kBq·h/m³ to 6 000 kBq·h/m³ · CMC 6,1 %
Radon measurements
Internal procedure; At permanent location (S) PW 5R

Reference standards (6)

PW 1G
PW 2X
PW 3S
PW 4N
PW 5R
PW 6B