🇨🇿 Dopravní podnik hl. m. Prahy, akciová společnost PL

ČIA 2425 • Praha 4, Czech Republic • 7 calibration items

Metrology & calibration

Dopravní podnik hl. m. Prahy, akciová společnost is a calibration laboratory accredited by ČIA under number 2425 (Praha 4, Czech Republic). Its accreditation scope covers 7 calibration items in the following areas: Metrology & calibration. Calibrated items include: AC voltage / AC voltage meters, AC voltage / AC voltage sources, Calipers, DC voltage / DC voltage meters. Most frequently used calibration principles: Direct comparison with a multimeter, Direct measurement with a multimeter, Comparative measurement with a reference torque transducer.

At a glance

  • Accreditation number: ČIA 2425
  • Calibration items in scope: 7
  • Types of calibrated items: 7
  • Calibration principles used: 5
  • City: Praha 4, Czech Republic
  • Valid until: 15-05-2027
Accreditation number
2425
Accreditation body
ČIA
Address
Sliačská 1150/1, 141 00 Praha 4
City
Praha 4, Czech Republic
Division
Laboratoria wzorcujące DPP
Valid until
15-05-2027

What this laboratory calibrates (7)

Calibration principles (5)

Laboratory sections (3)

Laboratory locations

Certificate information

Certificate date
15.05.2024

Accreditation scope — calibration items (7)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Workplace No. 1 – field of length
1 Calipers Calibration – length
0 mm to 150 mm · CMC 25 µm
CMC for the field of measured quantity: Length
Comparison with reference gauge blocks, rings and gauges KP 1-002
2 Micrometer callipers Calibration – length
0 mm to 25 mm · CMC 2 µm
CMC for the field of measured quantity: Length
Comparison with gauge blocks and gauges KP 1-001
Workplace No. 3 - field of torque
3 Torque / Torque wrenches, torque screwdrivers Calibration – torque
0,2 N·m to 1000 N·m · CMC 1 %
CMC for the field of measured quantity: Torque
Comparative measurement with a reference torque transducer KP 3-001
Workplace No. 2 - field of electrical quantities
4 DC voltage / DC voltage meters Calibration – electrical quantities
0 mV to 120 mV · CMC 0,0042 % + 5,2 µV 120 mV to 1,2 V · CMC 0,0015 % 1,2 V to 12 V · CMC 0,0013 % 12 V to 120 V · CMC 0,0017 % 120 V to 1000 V · CMC 0,0016 %
CMC for the field of measured quantity: Electrical quantities
Direct comparison with a multimeter KP 2-001, KP 2-003, KP 2-004
5 DC voltage / DC voltage sources Calibration – electrical quantities
0 mV to 120 mV · CMC 0,0042 % + 5,2 µV 120 mV to 1,2 V · CMC 0,0015 % 1,2 V to 12 V · CMC 0,0013 % 12 V to 120 V · CMC 0,0017 % 120 V to 1000 V · CMC 0,0016 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement with a multimeter KP 2-001, KP 2-003, KP 2-004
6 AC voltage / AC voltage meters Calibration – electrical quantities
20 Hz to 40 Hz — 1,0 mV to 12 mV · CMC 0,45 % 40 Hz to 1 kHz — 1,0 mV to 12 mV · CMC 0,18 % 1 kHz to 20 kHz — 1,0 mV to 12 mV · CMC 0,19 % 20 kHz to 50 kHz — 1,0 mV to 12 mV · CMC 0,27 % 50 kHz to 100 kHz — 1,0 mV to 12 mV · CMC 0,73 % 20 Hz to 40 Hz — 12 mV to 120 mV · CMC 0,054 % 40 Hz to 1 kHz — 12 mV to 120 mV · CMC 0,031 % 1 kHz to 20 kHz — 12 mV to 120 mV · CMC 0,039 % 20 kHz to 50 kHz — 12 mV to 120 mV · CMC 0,058 % 50 kHz to 100 kHz — 12 mV to 120 mV · CMC 0,12 % 20 Hz to 40 Hz — 120 mV to 1,2 V · CMC 0,054 % 40 Hz to 1 kHz — 120 mV to 1,2 V · CMC 0,031 % 1 kHz to 20 kHz — 120 mV to 1,2 V · CMC 0,039 % 20 kHz to 50 kHz — 120 mV to 1,2 V · CMC 0,058 % 50 kHz to 100 kHz — 120 mV to 1,2 V · CMC 0,12 % 20 Hz to 40 Hz — 1,2 V to 12 V · CMC 0,054 % 40 Hz to 1 kHz — 1,2 V to 12 V · CMC 0,031 % 1 kHz to 20 kHz — 1,2 V to 12 V · CMC 0,039 % 20 kHz to 50 kHz — 1,2 V to 12 V · CMC 0,058 % 50 kHz to 100 kHz — 1,2 V to 12 V · CMC 0,12 % 20 Hz to 40 Hz — 12 V to 120 V · CMC 0,069 % 40 Hz to 1 kHz — 12 V to 120 V · CMC 0,046 % 1 kHz to 20 kHz — 12 V to 120 V · CMC 0,046 % 20 kHz to 50 kHz — 12 V to 120 V · CMC 0,064 % 50 kHz to 100 kHz — 12 V to 120 V · CMC 0,16 % 20 Hz to 40 Hz — 120 V to 1000 V · CMC 0,073 % 40 Hz to 1 kHz — 120 V to 1000 V · CMC 0,027 % 1 kHz to 20 kHz — 120 V to 1000 V · CMC 0,083 % 20 kHz to 50 kHz — 120 V to 1000 V · CMC 0,15 % 50 kHz to 100 kHz — 120 V to 1000 V · CMC 0,36 %
CMC for the field of measured quantity: Electrical quantities
Direct comparison with a multimeter KP 2-002, KP 2-004
7 AC voltage / AC voltage sources Calibration – electrical quantities
20 Hz to 40 Hz — 1,0 mV to 12 mV · CMC 0,45 % 40 Hz to 1 kHz — 1,0 mV to 12 mV · CMC 0,18 % 1 kHz to 20 kHz — 1,0 mV to 12 mV · CMC 0,19 % 20 kHz to 50 kHz — 1,0 mV to 12 mV · CMC 0,27 % 50 kHz to 100 kHz — 1,0 mV to 12 mV · CMC 0,73 % 20 Hz to 40 Hz — 12 mV to 120 mV · CMC 0,054 % 40 Hz to 1 kHz — 12 mV to 120 mV · CMC 0,031 % 1 kHz to 20 kHz — 12 mV to 120 mV · CMC 0,039 % 20 kHz to 50 kHz — 12 mV to 120 mV · CMC 0,058 % 50 kHz to 100 kHz — 12 mV to 120 mV · CMC 0,12 % 20 Hz to 40 Hz — 120 mV to 1,2 V · CMC 0,054 % 40 Hz to 1 kHz — 120 mV to 1,2 V · CMC 0,031 % 1 kHz to 20 kHz — 120 mV to 1,2 V · CMC 0,039 % 20 kHz to 50 kHz — 120 mV to 1,2 V · CMC 0,058 % 50 kHz to 100 kHz — 120 mV to 1,2 V · CMC 0,12 % 20 Hz to 40 Hz — 1,2 V to 12 V · CMC 0,054 % 40 Hz to 1 kHz — 1,2 V to 12 V · CMC 0,031 % 1 kHz to 20 kHz — 1,2 V to 12 V · CMC 0,039 % 20 kHz to 50 kHz — 1,2 V to 12 V · CMC 0,058 % 50 kHz to 100 kHz — 1,2 V to 12 V · CMC 0,12 % 20 Hz to 40 Hz — 12 V to 120 V · CMC 0,069 % 40 Hz to 1 kHz — 12 V to 120 V · CMC 0,046 % 1 kHz to 20 kHz — 12 V to 120 V · CMC 0,046 % 20 kHz to 50 kHz — 12 V to 120 V · CMC 0,064 % 50 kHz to 100 kHz — 12 V to 120 V · CMC 0,16 % 20 Hz to 40 Hz — 120 V to 1000 V · CMC 0,073 % 40 Hz to 1 kHz — 120 V to 1000 V · CMC 0,027 % 1 kHz to 20 kHz — 120 V to 1000 V · CMC 0,083 % 20 kHz to 50 kHz — 120 V to 1000 V · CMC 0,15 % 50 kHz to 100 kHz — 120 V to 1000 V · CMC 0,36 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement with a multimeter KP 2-002, KP 2-004

Reference standards (7)

KP 1-001
KP 1-002
KP 2-001
KP 2-002
KP 2-003
KP 2-004
KP 3-001

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2425