🇨🇿 AVL Moravia s.r.o. PL

ČIA 2385 • Brno, Czech Republic • 31 calibration items

Metrology & calibration

AVL Moravia s.r.o. is a calibration laboratory accredited by ČIA under number 2385 (Brno, Czech Republic). Its accreditation scope covers 31 calibration items in the following areas: Metrology & calibration. Calibrated items include: AC current / Multimeters, ammeters and inspection devices, current…, Active electrical power / Wattmeters, network analyzers and inspect…, Apparent electrical power / Wattmeters, network analyzers and inspe…, DC current / Ammeters, multimeters, clamp meters and current sensor…. Most frequently used calibration principles: Generation by a standard calibrator, Direct measurement by a standard multimeter, Comparison using reference weights and calibration arms.

At a glance

  • Accreditation number: ČIA 2385
  • Calibration items in scope: 31
  • Types of calibrated items: 24
  • Calibration principles used: 18
  • City: Brno, Czech Republic
  • Valid until: 24-10-2027
Accreditation number
2385
Accreditation body
ČIA
Address
Olomoucká 3419/7, Židenice, 618 00 Brno
City
Brno, Czech Republic
Division
Laboratorium wzorcujące
Valid until
24-10-2027

What this laboratory calibrates (24)

Calibration principles (18)

Laboratory sections (2)

Laboratory locations

Certificate information

Certificate date
14.04.2026

Accreditation scope — calibration items (31)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Workplace Hranice
1 Length / Roller diameter of dynamometers for vehicle testing Calibration – length
Diameter of roller determined from circumference length — 1,2 m to 1,5 m · CMC 0,09 m
CMC for the field of measured quantity: Length
Direct measurement by diametrical tape AW-02-1019
2 Velocity on the roller surface / Dynamometers for vehicle testing Calibration – mechanical motion
891,2 Hz to 267,3 kHz — 1 km·h⁻¹ to 300 km·h⁻¹ · CMC 0,021 % + 0,0039 km·h⁻¹
CMC for the field of measured quantity: Mechanical motion
Calculation based on the specified reference speed and roller diameter AW-02-1016
3 Speed sensors with a high number of pulses per revolution / Dynamometers for vehicle testing Calibration – mechanical motion
0,0 kHz to 102,4 kHz — 0 min⁻¹ to 500 min⁻¹ · CMC 0,000036 % + 0,0058 min⁻¹ 102,4 kHz to 204,8 kHz — 500 min⁻¹ to 1000 min⁻¹ · CMC 0,00010 % + 0,0055 min⁻¹ 204,8 kHz to 307,2 kHz — 1000 min⁻¹ to 1500 min⁻¹ · CMC 0,00016 % + 0,0050 min⁻¹ 307,2 kHz to 409,6 kHz — 1500 min⁻¹ to 2000 min⁻¹ · CMC 0,00019 % + 0,0044 min⁻¹ 409,6 kHz to 512,0 kHz — 2000 min⁻¹ to 2500 min⁻¹ · CMC 0,00022 % + 0,0039 min⁻¹
CMC for the field of measured quantity: Mechanical motion
Comparison with a standard counter AW-02-1007
4 Air flow velocity / Fans for motor vehicle tests Calibration – mechanical motion
10 km·h⁻¹ to 140 km·h⁻¹ · CMC 2,44 % + 0,82 km·h⁻¹
CMC for the field of measured quantity: Mechanical motion
Direct measurement by a standard anemometer AW-02-1017
5 Force on the roller surface / Dynamometers for vehicle testing Calibration – force
500 N to 25000 N · CMC 0,023 % + 0,026 N
CMC for the field of measured quantity: Force
Calculation based on the specified reference torque and roller diameter AW-02-1015
6 Torque / Dynamometers for vehicle testing Calibration – force
300 Nm to 15000 Nm · CMC 0,031 % + 0,014 Nm
CMC for the field of measured quantity: Force
Comparison using reference weights and calibration arms AW-02-1011
7 Time Interval / Time base of dynamometers for vehicle testing Calibration – time and frequency quantities
1000 s · CMC 0,0015 s 2000 s · CMC 0,0034 s
CMC for the field of measured quantity: Time and frequency quantities
Comparison with a standard time interval generator AW-02-1018
Workplace Brno
8 Digital speed meters, speed sensors, stroboscopes Calibration – mechanical motion
optically; 1 pulse per revolution — 1 min⁻¹ to 100000 min⁻¹ · CMC 0,006 % pulsed (IRC sensor); 0,001 kHz to 100 kHz — 10 min⁻¹ to 30000 min⁻¹ · CMC 0,006 %
CMC for the field of measured quantity: Mechanical motion
Comparison with a standard speed meter AW-02-1007
9 Torque / Dynamometers for rotating machines tests Calibration – force
0,01 Nm to 10 Nm · CMC 0,2 % 10 Nm to 20 Nm · CMC 0,1 % 20 Nm to 100 Nm · CMC 0,075 % 100 Nm to 10000 Nm · CMC 0,05 %
CMC for the field of measured quantity: Force
Comparison using reference weights and calibration arms AW-02-1011
10 Resistance and thermoelectric temperature sensors Calibration – temperature
Sensors with a diameter up to 9.5 mm and a length up to 102 mm — 0 °C · CMC 0,3 °C
CMC for the field of measured quantity: Temperature
Generation by crushed ice AW-02-1010
11 Resistance and thermoelectric temperature sensors Calibration – temperature
Sensors with a diameter up to 9.5 mm and a length up to 102 mm — 35 °C to 100 °C · CMC 0,5 °C Sensors with a diameter up to 9.5 mm and a length up to 102 mm — 100 °C to 200 °C · CMC 0,7 °C Sensors with a diameter up to 9.5 mm and a length up to 102 mm — 200 °C to 350 °C · CMC 1,0 °C
CMC for the field of measured quantity: Temperature
Generation by standard calibration furnace AW-02-1010
12 DC voltage / Voltmeters, multimeters and inspection devices Calibration – electrical quantities
0 mV to 200 mV · CMC 0,0044 % + 5,3µV 0,2 V to 2 V · CMC 0,0044 % + 7,5µV 2 V to 20 V · CMC 0,0036 % + 68 µV 20 V to 200 V · CMC 0,0044 % + 750 µV 200 V to 1025 V · CMC 0,0044 % + 7,5 mV
CMC for the field of measured quantity: Electrical quantities
Generation by a standard calibrator AW-02-1003
13 DC voltage / Sources and calibrators Calibration – electrical quantities
0 mV to 100 mV · CMC 0,0043 % + 4,8 µV 100 mV to 1000 mV · CMC 0,0029 % + 11 µV 1 V to 10 V · CMC 0,0028 % + 71 µV 10 V to 100 V · CMC 0,0044 % + 920 µV 100 V to 1000 V · CMC 0,0048 % + 14 mV
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard multimeter AW-02-1003
14 DC current / Ammeters, multimeters, clamp meters and current sensors, current / current converters, current / voltage converters Calibration – electrical quantities
0 µA to 200 µA · CMC 0,014 % + 0,049 µA 0,2 mA to 2 mA · CMC 0,012 % + 0,067µA 2 mA to 20 mA · CMC 0,0065 % + 0,51µA 20 mA to 200 mA · CMC 0,0081 % + 5,2 µA 0,2 A to 2 A · CMC 0,013 % + 64 µA 2 A to 30 A · CMC 0,049 % + 0,71 mA
CMC for the field of measured quantity: Electrical quantities
Generation by a standard calibrator AW-02-1004
15 DC current / Ammeters, multimeters, clamp meters and current sensors, current / current converters, current / voltage converters Calibration – electrical quantities
30 A to 1500 A · CMC 0,35 %
CMC for the field of measured quantity: Electrical quantities
Generation by a calibrator with a multi-turn coil AW-02-1004
16 DC current / Sources and calibrators Calibration – electrical quantities
0 µA to 100 µA · CMC 0,058 % + 31 nA 0,1 mA to 1 mA · CMC 0,058 % + 72 nA 1 mA to 10 mA · CMC 0,058 % + 2,4 µA 10 mA to 100 mA · CMC 0,058 % + 7,8 µA 100 mA to 400 mA · CMC 0,058 % + 50 µA 0,4 A to 1 A · CMC 0,058 % + 0,26 mA 1 A to 3 A · CMC 0,12 % + 1,8 mA 3 A to 10 A · CMC 0,18 % + 4,2 mA
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard multimeter AW-02-1004
17 AC voltage / Voltmeters, multimeters and inspection devices Calibration – electrical quantities
10 Hz to 45 Hz — 1 mV to 200 mV · CMC 0,29 % + 84 µV 45 Hz to 1000 Hz — 1 mV to 200 mV · CMC 0,049 % + 38 µV 1 kHz to 20 kHz — 1 mV to 200 mV · CMC 0,12 % + 59 µV 20 kHz to 100 kHz — 1 mV to 200 mV · CMC 0,46 % + 0,15 mV 100 kHz to 500 kHz — 1 mV to 200 mV · CMC 0,97 % + 0,52 mV 10 Hz to 45 Hz — 0,2 V to 2 V · CMC 0,29 % + 0,6 mV 45 Hz to 1000 Hz — 0,2 V to 2 V · CMC 0,049 % + 0,19 mV 1 kHz to 20 kHz — 0,2 V to 2 V · CMC 0,097 % + 0,24 mV 20 kHz to 100 kHz — 0,2 V to 2 V · CMC 0,36 % + 3,4 mV 100 kHz to 500 kHz — 0,2 V to 2 V · CMC 0,55 % + 6,5 mV 10 Hz to 45 Hz — 2 V to 20 V · CMC 0,29 % + 5,3 mV 45 Hz to 1000 Hz — 2 V to 20 V · CMC 0,046 % + 1,9 mV 1 kHz to 20 kHz — 2 V to 20 V · CMC 0,081 % + 2,4 mV 20 kHz to 100 kHz — 2 V to 20 V · CMC 0,29 % + 55 mV 30 Hz to 45 Hz — 20 V to 200 V · CMC 0,065 % + 35 mV 45 Hz to 1000 Hz — 20 V to 200 V · CMC 0,049 % + 17 mV 1 kHz to 20 kHz — 20 V to 200 V · CMC 0,12 % + 68 mV 30 Hz to 45 Hz — 200 V to 1020 V · CMC 0,065 % + 0,36 V 45 Hz to 1000 Hz — 200 V to 1020 V · CMC 0,049 % + 0,21 V 1 kHz to 20 kHz — 200 V to 1020 V · CMC 0,2 % + 0,73 V
CMC for the field of measured quantity: Electrical quantities
Generation by a standard calibrator AW-02-1003
18 AC voltage / Sources and calibrators Calibration – electrical quantities
3 Hz to 5 Hz — 1 mV to 100 mV · CMC 1,2 % + 0,059 mV 5 Hz to 10 Hz — 1 mV to 100 mV · CMC 0,41 % + 0,059 mV 10 Hz to 20 kHz — 1 mV to 100 mV · CMC 0,07 % + 0,049 mV 20 kHz to 50 kHz — 1 mV to 100 mV · CMC 0,14 % + 0,064 mV 50 kHz to 100 kHz — 1 mV to 100 mV · CMC 0,7 % + 0,12 mV 100 kHz to 300 kHz — 1 mV to 100 mV · CMC 4,7 % + 0,72 mV 3 Hz to 5 Hz — 0,1 V to 1 V · CMC 1,2 % + 0,45 mV 5 Hz to 10 Hz — 0,1 V to 1 V · CMC 0,41 % + 0,45 mV 10 Hz to 20 kHz — 0,1 V to 1 V · CMC 0,07 % + 0,35 mV 20 kHz to 20 kHz — 0,1 V to 1 V · CMC 0,14 % + 0,59 mV 50 kHz to 100 kHz — 0,1 V to 1 V · CMC 0,7 % + 0,94 mV 100 kHz to 300 kHz — 0,1 V to 1 V · CMC 4,7 % + 0,77 mV 3 Hz to 5 Hz — 1 V to 10 V · CMC 1,2 % + 4,5 mV 5 Hz to 10 Hz — 1 V to 10 V · CMC 0,41 % + 4,5 mV 10 Hz to 20 kHz — 1 V to 10 V · CMC 0,07 % + 3,5 mV 20 kHz to 50 kHz — 1 V to 10 V · CMC 0,14 % + 5,9 mV 50 kHz to 100 kHz — 1 V to 10 V · CMC 0,7 % + 9,4 mV 100 kHz to 300 kHz — 1 V to 10 V · CMC 4,7 % + 58 mV 3 Hz to 5 Hz — 10 V to 100 V · CMC 1,2 % + 45 mV 5 Hz to 10 Hz — 10 V to 100 V · CMC 0,41 % + 45 mV 10 Hz to 20 kHz — 10 V to 100 V · CMC 0,07 % + 36 mV 20 kHz to 50 kHz — 10 V to 100 V · CMC 0,14 % + 59 mV 50 kHz to 100 kHz — 10 V to 100 V · CMC 0,7 % + 95 mV 100 kHz to 300 kHz — 10 V to 100 V · CMC 4,7 % + 0,77 V 3 Hz to 5 Hz — 100 V to 1000 V · CMC 1,2 % + 0,27V 5 Hz to 10 Hz — 100 V to 1000 V · CMC 0,41 % + 0,27 V 10 Hz to 20 kHz — 100 V to 1000 V · CMC 0,07 % + 0,56 V 20 kHz to 50 kHz — 100 V to 1000 V · CMC 0,14 % + 0,71 V 50 kHz to 100 kHz — 100 V to 1000 V · CMC 0,7 % + 0,92 V
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard multimeter AW-02-1003
19 AC current / Multimeters, ammeters and inspection devices, current / current converters, current / voltage converters Calibration – electrical quantities
10 Hz to 45 Hz — 20 µA to 200 µA · CMC 0,28 % + 0,41 µA 45 Hz to 1000 Hz — 20 µA to 200 µA · CMC 0,13 % + 0,41 µA 1 kHz to 10 kHz — 20 µA to 200 µA · CMC 1,8 % + 0,41 µA 10 Hz to 45 Hz — 0,2 mA to 2 mA · CMC 0,28 % + 0,87 µA 45 Hz to 1000 Hz — 0,2 mA to 2 mA · CMC 0,12 % + 0,73 µA 1 kHz to 10 kHz — 0,2 mA to 2 mA · CMC 0,97 % + 1,2 µA 10 Hz to 45 Hz — 2 mA to 20 mA · CMC 0,28 % + 8,6 µA 45 Hz to 1000 Hz — 2 mA to 20 mA · CMC 0,12 % + 7,1 µA 1 kHz to 10 kHz — 2 mA to 20 mA · CMC 0,65 % + 12 µA 10 Hz to 45 Hz — 20 mA to 200 mA · CMC 0,28 % + 86 µA 45 Hz to 1000 Hz — 20 mA to 200 mA · CMC 0,12 % + 71 µA 1 kHz to 10 kHz — 20 mA to 200 mA · CMC 0,65 % + 0,12 mA 10 Hz to 45 Hz — 0,2 A to 2 A · CMC 0,28 % + 0,95 mA 45 Hz to 1000 Hz — 0,2 A to 2 A · CMC 0,13 % + 0,81 mA 1 kHz to 5 kHz — 0,2 A to 2 A · CMC 0,81 % + 1,3 mA 30 Hz to 45 Hz — 2 A to 30 A · CMC 0,25 % + 9,7 mA 45 Hz to 100 Hz — 2 A to 30 A · CMC 0,057 % + 6,6 mA 0,1 kHz to 1 kHz — 2 A to 30 A · CMC 0,41 % + 9,0 mA
CMC for the field of measured quantity: Electrical quantities
Generation by a standard calibrator AW-02-1004
20 AC current / Multimeters, ammeters and inspection devices, current / current converters, current / voltage converters Calibration – electrical quantities
50 Hz to 400 Hz — 30 A to 600 A · CMC 0,12 % 50 Hz to 60 Hz — 600 A to 1500 A · CMC 0,35 %
CMC for the field of measured quantity: Electrical quantities
Generation by a calibrator with a multi-turn coil AW-02-1004
21 AC current / Sources and calibrators Calibration – electrical quantities
3 Hz to 5 Hz — 1 µA to 100 µA · CMC 1,3 % + 0,081 µA 5 Hz to 10 Hz — 1 µA to 100 µA · CMC 0,41 % + 0,081 µA 10 Hz to 5 kHz — 1 µA to 100 µA · CMC 0,18 % + 0,08 µA 5 kHz to 10 kHz — 1 µA to 100 µA · CMC 0,41 % + 0,83 µA 3 Hz to 5 Hz — 0,1 mA to 1 mA · CMC 1,2 % + 0,55 µA 5 Hz to 10 Hz — 0,1 mA to 1 mA · CMC 0,35 % + 0,55 µA 10 Hz to 5 kHz — 0,1 mA to 1 mA · CMC 0,12 % + 0,56 µA 5 kHz to 10 kHz — 0,1 mA to 1 mA · CMC 0,23 % + 3,4 µA 3 Hz to 5 Hz — 1 mA to 10 mA · CMC 1,3 % + 7,6 µA 5 Hz to 10 Hz — 1 mA to 10 mA · CMC 0,41 % + 7,6 µA 10 Hz to 5 kHz — 1 mA to 10 mA · CMC 0,18 % + 7,4 µA 5 kHz to 10 kHz — 1 mA to 10 mA · CMC 0,41 % + 83 µA 3 Hz to 5 Hz — 10 mA to 100 mA · CMC 1,2 % + 0,055 mA 5 Hz to 10 Hz — 10 mA to 100 mA · CMC 0,35 % + 0,055 mA 10 Hz to 5 kHz — 10 mA to 100 mA · CMC 0,12 % + 0,053 mA 5 kHz to 10 kHz — 10 mA to 100 mA · CMC 0,24 % + 0,32 mA 3 Hz to 5 Hz — 100 mA to 400 mA · CMC 1,2 % + 0,49 mA 5 Hz to 10 Hz — 100 mA to 400 mA · CMC 0,35 % + 0,49 mA 10 Hz to 5 kHz — 100 mA to 400 mA · CMC 0,12 % + 0,53 mA 5 kHz to 10 kHz — 100 mA to 400 mA · CMC 0,24 % + 4,4 mA 3 Hz to 5 Hz — 400 mA to 1 A · CMC 1,2 % + 0,55 mA 5 Hz to 10 Hz — 400 mA to 1 A · CMC 0,35 % + 0,55 mA 10 Hz to 5 kHz — 400 mA to 1 A · CMC 0,12 % + 0,71 mA 5 kHz to 10 kHz — 400 mA to 1 A · CMC 0,41 % + 11 mA 3 Hz to 5 Hz — 1 A to 3 A · CMC 1,3 % + 2,7 mA 5 Hz to 10 Hz — 1 A to 3 A · CMC 0,41 % + 2,7 mA 10 Hz to 5 kHz — 1 A to 3 A · CMC 0,18 % + 3,9 mA 5 kHz to 10 kHz — 1 A to 3 A · CMC 0,41 % + 27 mA 3 Hz to 5 Hz — 3 A to 10 A · CMC 1,3 % + 8,5 mA 5 Hz to 10 Hz — 3 A to 10 A · CMC 0,41 % + 8,5 mA 10 Hz to 5 kHz — 3 A to 10 A · CMC 0,18 % + 13 mA 5 kHz to 10 kHz — 3 A to 10 A · CMC 0,41 % + 89 mA
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard multimeter AW-02-1004
22 DC resistance / Multimeters, ohmmeters, inspection devices, resistance / voltage converters Calibration – electrical quantities
100 µΩ · CMC 0,015 % 1 mΩ · CMC 0,032 % 10 mΩ · CMC 0,037 % 100 mΩ · CMC 0,0050 % 1 Ω · CMC 0,0050 % 10 Ω · CMC 0,0060 % 100 Ω · CMC 0,0051 % 1 kΩ · CMC 0,0051 % 10 kΩ · CMC 0,0050 % 100 kΩ · CMC 0,0051 % 1 MΩ · CMC 0,0080 % 10 MΩ · CMC 0,0081 % 100 MΩ · CMC 0,013 % 1 GΩ · CMC 0,040 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement of a standard resistance AW-02-1013, AW-02-1014
23 DC resistance / Decade resistance boxes, resistors and calibrators Calibration – electrical quantities
0 Ω to 10 Ω · CMC 0,012 % + 3,5 mΩ 10 Ω to 100 Ω · CMC 0,012 % + 4,8 mΩ 100 Ω to 1 kΩ · CMC 0,012 % + 15 mΩ 1 kΩ to 10 kΩ · CMC 0,012 % + 0,15 Ω 10 kΩ to 100 kΩ · CMC 0,012 % + 1,6 Ω 100 kΩ to 1 MΩ · CMC 0,012 % + 24 Ω 1 MΩ to 10 MΩ · CMC 0,047 % + 0,42 kΩ 10 MΩ to 100 MΩ · CMC 0,93 % + 16 kΩ 100 MΩ to 1 GΩ · CMC 2,4 % + 0,2 MΩ
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard multimeter AW-02-1012
24 DC resistance / Decade resistance boxes, resistors and calibrators Calibration – electrical quantities
0 mΩ to 0,1 mΩ · CMC 0,74 µΩ 0,1 mΩ to 1 mΩ · CMC 0,025 % + 0,74 µΩ 1 mΩ to 10 mΩ · CMC 0,052 % + 0,45 µΩ 10 mΩ to 100 mΩ · CMC 0,020 % + 3,6 µΩ 0,1 Ω to 1 Ω · CMC 0,019 % + 4,3 µΩ 1 Ω to 10 Ω · CMC 0,013 % + 61 µΩ 10 Ω to 100 Ω · CMC 0,012 % + 170 µΩ
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard calibrator and multimeter (voltamper method) AW-02-1012
25 Active electrical power / Wattmeters, network analyzers and inspection devices with ranges (1 to 1,000) V and (45 to 100) Hz Calibration – electrical quantities
(0.3 to 30) A; cos φ = 1 — 0,3 W to 30 kW · CMC 0,12 % (0.3 to 30) A; cos φ = (0.8 to 0.99) — 0,3 W to 30 kW · CMC 0,35 % (0.3 to 30) A; cos φ = (0.5 to 0.8) — 0,3 W to 30 kW · CMC 0,75 %
CMC for the field of measured quantity: Electrical quantities
Generation by a standard calibrator AW-02-1005
26 Active electrical power / Wattmeters, network analyzers and inspection devices with ranges (1 to 1,000) V and (45 to 100) Hz Calibration – electrical quantities
(30 to 600) A; cos φ = 1 — 30 kW to 600 kW · CMC 0,13 % (30 to 600) A; cos φ = (0.8 to 0.99) — 30 kW to 600 kW · CMC 0,35 % (30 to 600) A; cos φ = (0.5 to 0.8) — 30 kW to 600 kW · CMC 0,75 %
CMC for the field of measured quantity: Electrical quantities
Comparison with a standard wattmeter with a current sensor AW-02-1005
27 Apparent electrical power / Wattmeters, network analyzers and inspection devices with ranges (1 to 1,000) V Calibration – electrical quantities
(0.3 to 30) A; (45 to 400) Hz — 0,3 VA to 30 kVA · CMC 0,12 %
CMC for the field of measured quantity: Electrical quantities
Generation by a standard calibrator AW-02-1005
28 Apparent electrical power / Wattmeters, network analyzers and inspection devices with ranges (1 to 1,000) V Calibration – electrical quantities
(30 to 600) A; (45 to 150) Hz — 30 kVA to 600 kVA · CMC 0,12 %
CMC for the field of measured quantity: Electrical quantities
Comparison with a standard wattmeter with a current sensor AW-02-1005
29 Time interval / Manually controlled stopwatches, time switches Calibration – time and frequency quantities
1 s to 3600 s · CMC 0,42 s
CMC for the field of measured quantity: Time and frequency quantities
Comparison with a standard counter AW-02-1001
30 Frequency / Frequency meters and frequency standards Calibration – time and frequency quantities
1 Hz · CMC 0,0032 % 10 Hz · CMC 0,0032 % 100 Hz · CMC 0,0032 % 1 kHz · CMC 0,0032 % 10 kHz · CMC 0,0032 % 100 kHz · CMC 0,0032 % 1 MHz · CMC 0,0032 % 10 MHz · CMC 0,0032 %
CMC for the field of measured quantity: Time and frequency quantities
Generation by a standard frequency generator AW-02-1006
31 Frequency / Frequency meters and frequency standards Calibration – time and frequency quantities
3 Hz to 5 Hz · CMC 0,05 % 5 Hz to 10 Hz · CMC 0,01 % 10 Hz to 40 Hz · CMC 0,01 % 40 Hz to 1 MHz · CMC 0,001 %
CMC for the field of measured quantity: Time and frequency quantities
Direct measurement by a standard counter AW-02-1008

Reference standards (17)

AW-02-1001
AW-02-1003
AW-02-1004
AW-02-1005
AW-02-1006
AW-02-1007
AW-02-1008
AW-02-1010
AW-02-1011
AW-02-1012
AW-02-1013
AW-02-1014
AW-02-1015
AW-02-1016
AW-02-1017
AW-02-1018
AW-02-1019

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2385