🇨🇿 INPEK spol. s r.o. PL

ČIA 2335 • Praha 8, Czech Republic • 2 calibration items

Metrology & calibration

INPEK spol. s r.o. is a calibration laboratory accredited by ČIA under number 2335 (Praha 8, Czech Republic). Its accreditation scope covers 2 calibration items in the following areas: Metrology & calibration. Calibrated items include: Concentration of gaseous components, NO2/NO converter. Most frequently used calibration principles: Comparison with the standard (CRM and dilution unit).

At a glance

  • Accreditation number: ČIA 2335
  • Calibration items in scope: 2
  • Types of calibrated items: 2
  • Calibration principles used: 1
  • City: Praha 8, Czech Republic
  • Valid until: 13-03-2030
Accreditation number
2335
Accreditation body
ČIA
Address
V Holešovičkách 94/41, Libeň, 182 00 Praha 8
City
Praha 8, Czech Republic
Division
Laboratorium wzorcujące
Valid until
13-03-2030

What this laboratory calibrates (2)

Calibration principles (1)

Certificate information

Certificate date
13.03.2025

Accreditation scope — calibration items (2)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Concentration of gaseous components Calibration – amount of substance
NO — 10 µmol/mol to 2000 µmol/mol · CMC 1,08 % NO₂ — 10 µmol/mol to 500 µmol/mol · CMC 2,22 % N₂O — 10 µmol/mol to 500 µmol/mol · CMC 2,22 % CO — 10 µmol/mol to 3000 µmol/mol · CMC 0,99 % CO — 0,003 mol/mol to 0,22 mol/mol · CMC 1,16 % SO₂ — 10 µmol/mol to 2000 µmol/mol · CMC 1,00 % CO₂ — 0,0024 mol/mol to 0,20 mol/mol · CMC 1,09 % O₂ — 0,004 mol/mol to 0,21 mol/mol · CMC 1,00 % C₃H₈ — 10 µmol/mol to 10000 µmol/mol · CMC 1,04 %
CMC for the field of measured quantity: Amount of substance
Comparison with the standard (CRM and dilution unit) PPK 001 (ČSN EN ISO 9169)
2 NO2/NO converter Calibration – amount of substance
converter efficiency 0 to 100% — 200 µmol/mol to 1000 µmol/mol · CMC 2,56 % converter efficiency 0 to 100% — 20 µmol/mol to 200 µmol/mol · CMC 3,35 %
CMC for the field of measured quantity: Amount of substance
Comparison with the standard (CRM and dilution unit) PPK 002 (ČSN EN ISO 9169; ČSN EN 14792)

Reference standards (4)

PPK 001
PPK 002
ČSN EN 14792
ČSN EN ISO 9169

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2335