🇨🇿 MIKROKOM, s.r.o. PL

ČIA 2311 • Praha 4, Czech Republic • 11 calibration items

Metrology & calibration

MIKROKOM, s.r.o. is a calibration laboratory accredited by ČIA under number 2311 (Praha 4, Czech Republic). Its accreditation scope covers 11 calibration items in the following areas: Metrology & calibration. Calibrated items include: Attenuation (1 dB) / optical reflectometers OTDR, Linearity / optical spectrum analysers for fibre optics, Linearity / power meters for fibre optics, Maximum power wavelength / light sources for fibre optics. Most frequently used calibration principles: Comparison with a standard optical power meter by substitution method, Comparison with a standard optical power meter using an optical att…, Wavelength measurement by a standard optical spectrum analyser.

At a glance

  • Accreditation number: ČIA 2311
  • Calibration items in scope: 11
  • Types of calibrated items: 11
  • Calibration principles used: 8
  • City: Praha 4, Czech Republic
  • Valid until: 22-12-2030
Accreditation number
2311
Accreditation body
ČIA
Address
Pod vinicí 622/22, Modřany, 143 00 Praha 4
City
Praha 4, Czech Republic
Division
Laboratorium wzorcujące
Valid until
22-12-2030

What this laboratory calibrates (11)

Calibration principles (8)

Certificate information

Certificate date
22.12.2025

Accreditation scope — calibration items (11)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 RF signal level / RF measuring receivers Calibration – electrical quantities
1 MHz to 2050 MHz; 75 Ω — -70 dBm to +7,4 dBm · CMC 0,64 dB 1 MHz to 2050 MHz; 50 Ω — -70 dBm to +7,4 dBm · CMC 0,58 dB
CMC for the field of measured quantity: Electrical quantities
Measurement of power generated by the standard KP06
2 Optical power level / power meters for fibre optics Calibration – optical quantities
800 nm to 1650 nm — -35 dBm to -4 dBm · CMC 2,8 %
CMC for the field of measured quantity: Optical quantities
Comparison with a standard optical power meter by substitution method KP01
3 Linearity / power meters for fibre optics Calibration – optical quantities
800 nm to 1650 nm — -60 dBm to -4 dBm · CMC 0,14 dB
CMC for the field of measured quantity: Optical quantities
Comparison with a standard optical power meter using an optical attenuator KP01
4 Optical power level / light sources for fibre optics Calibration – optical quantities
800 nm to 1650 nm — -60 dBm to +10 dBm · CMC 0,15 dB
CMC for the field of measured quantity: Optical quantities
Power measurement by a standard optical power meter KP02
5 Maximum power wavelength / light sources for fibre optics Calibration – optical quantities
600 nm to 1 650 nm · CMC 0,11 nm
CMC for the field of measured quantity: Optical quantities
Wavelength measurement by a standard optical spectrum analyser KP02
6 Wavelength / optical spectrum analysers for fibre optics Calibration – optical quantities
1 250 nm to 1 650 nm · CMC 0,11 nm 1 530 nm to 1 560 nm · CMC 0,012 nm
CMC for the field of measured quantity: Optical quantities
1 250 nm to 1 650 nm: comparison with a standard optical spectrum analyser; 1 530 nm to 1 560 nm: wavelength measurement using a standard absorption cell KP04
7 Optical power level / optical spectrum analysers for fibre optics Calibration – optical quantities
800 nm to 1650 nm — -35 dBm to -4 dBm · CMC 0,16 dB
CMC for the field of measured quantity: Optical quantities
Comparison with a standard optical power meter by substitution method KP04
8 Linearity / optical spectrum analysers for fibre optics Calibration – optical quantities
800 nm to 1650 nm — -60 dBm to -5 dBm · CMC 0,19 dB
CMC for the field of measured quantity: Optical quantities
Comparison with a standard optical power meter using an optical attenuator KP04
9 Maximum power wavelength / optical reflectometers OTDR Calibration – optical quantities
600 nm to 1650 nm · CMC 0,11 nm
CMC for the field of measured quantity: Optical quantities
Wavelength measurement by a standard optical spectrum analyser KP05
10 Optical length of SMF fibre / optical reflectometers OTDR Calibration – optical quantities
spectral bands 1310 nm, 1550 nm, 1625 nm — 20,17 km · CMC 2,3 m
CMC for the field of measured quantity: Optical quantities
Optical length measurement of an optical fibre standard KP05
11 Attenuation (1 dB) / optical reflectometers OTDR Calibration – optical quantities
800 nm to 1650 nm — -35 dBm to -4 dBm · CMC 0,021 dB
CMC for the field of measured quantity: Optical quantities
Attenuation measurement at various optical lengths of the fibre – comparison with a standard optical power meter using an optical attenuator KP05

Reference standards (5)

KP01
KP02
KP04
KP05
KP06

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2311