🇨🇿 MSP - GLOBAL s.r.o. PL

ČIA 2286 • Dobrá, Czech Republic • 8 calibration items

Metrology & calibration

MSP - GLOBAL s.r.o. is a calibration laboratory accredited by ČIA under number 2286 (Dobrá, Czech Republic). Its accreditation scope covers 8 calibration items in the following areas: Metrology & calibration. Calibrated items include: – current loops, – Resistance sensors, – Type “K” thermocouples, Deformation and digital manometers, pressure transducers. Most frequently used calibration principles: Comparison with a reference simulator of el. quantities, Comparison with a reference thermometer in a dry block, Comparison with a reference thermometer in a dry block and in a hor….

At a glance

  • Accreditation number: ČIA 2286
  • Calibration items in scope: 8
  • Types of calibrated items: 11
  • Calibration principles used: 8
  • City: Dobrá, Czech Republic
  • Valid until: 28-08-2028
Accreditation number
2286
Accreditation body
ČIA
Address
Dobrá 240, 739 51 Dobrá
City
Dobrá, Czech Republic
Division
Czeska Służba Wzorcująca
Valid until
28-08-2028

What this laboratory calibrates (11)

Calibration principles (8)

Certificate information

Certificate date
28.08.2023

Accreditation scope — calibration items (8)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Slide gauges: division 0,01 mm and 0,02 mm, division 0,05 and 0,1 mm Calibration – length
0 mm to 250 mm · CMC 30 µm
CMC for the field of measured quantity: Length
Comparison with reference parallel gauge blocks and rings KBP 7.2, KBP 7.1
2 Micrometer calliper gauges division 1 µm and 10 µm Calibration – length
0 mm to 300 mm · CMC 3 µm
CMC for the field of measured quantity: Length
Comparison with reference parallel gauge blocks KBP 8
3 Electronic and mechanical scales with non-automatic function Calibration – mass
class F1 weight — 0 kg to 0,22 kg · CMC 1,4 mg class F1 weight — 0,22 kg to 1,6 kg · CMC 5,8 mg class F1 weight — 1,6 kg to 5 kg · CMC 42 mg class F1 weight — 5 kg to 10 kg · CMC 81 mg class F2 weight — 10 kg to 16 kg · CMC 0,13 g class F2 weight — 16 kg to 20 kg · CMC 1,2 g class M1 weight — 20 kg to 50 kg · CMC 3,2 g class M1 weight — 50 kg to 100 kg · CMC 6,3 g
CMC for the field of measured quantity: Mass
Direct loading with a reference weight KBP 14.1, KBP 14.2
4 Torque / torque wrenches and screwdrivers Calibration – force, mechanical tests
2 Nm to 1 100 Nm · CMC 1,4 %
CMC for the field of measured quantity: Force, mechanical tests
Comparison with a reference torque sensor KBP 25
5 Deformation and digital manometers, pressure transducers Calibration – pressure, mechanical stress
gas — 0 kPa to 35 kPa · CMC 22 Pa gas — 35 kPa to 200 kPa · CMC 61 Pa gas — 200 kPa to 2 000 kPa · CMC 0,6 kPa oil — 1,2 MPa to 12 MPa · CMC 0,03 % water — 0 MPa to 60 MPa · CMC 0,07 MPa
CMC for the field of measured quantity: Pressure, mechanical stress
gas: comparison with a reference digital pressure gauge; oil: comparison with a reference piston pressure gauge; water: comparison with a reference digital pressure gauge KBP 9, KBP 9.1, KBP 10
6 Resistance thermometers Pt100 Calibration – temperature
0 °C to 200 °C · CMC 0,32 °C 200 °C to 400 °C · CMC 0,54 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermometer in a dry block KBP 5
7 Indicating thermometers and temperature measuring chains Calibration – temperature
0 °C to 200 °C · CMC 0,32 °C 200 °C to 400 °C · CMC 0,54 °C 400 °C to 650 °C · CMC 2,1 °C 650 °C to 1 100 °C · CMC 2,8 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermometer in a dry block and in a horizontal furnace KBP 15
8 Measuring chains – Simulation of electrical output signal:
– Type “K” thermocouples
– Resistance sensors
– current loops
Calibration – temperature
type “K” thermocouples — - 200 °C to 1000 °C · CMC 0,6 °C type “K” thermocouples — 1000 °C to 1100 °C · CMC 0,7 °C resistance sensors — - 200 °C to 200 °C · CMC 0,20 °C resistance sensors — 200 °C to 600 °C · CMC 0,34 °C resistance sensors — 600 °C to 850 °C · CMC 0,50 °C current loops — 0 mA to 4 mA · CMC 3 µA current loops — 4 mA to 12 mA · CMC 6 µA current loops — 12 mA to 25 mA · CMC 8 µA
CMC for the field of measured quantity: Temperature
Comparison with a reference simulator of el. quantities KBP 4.1

Reference standards (12)

KBP 10
KBP 14.1
KBP 14.2
KBP 15
KBP 25
KBP 4.1
KBP 5
KBP 7.1
KBP 7.2
KBP 8
KBP 9
KBP 9.1

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2286