🇨🇿 Technický a zkušební ústav stavební Praha, s.p. PL
Metrology & calibration
Technický a zkušební ústav stavební Praha, s.p. is a calibration laboratory accredited by ČIA under number 2275 (Praha 9, Czech Republic). Its accreditation scope covers 11 calibration items in the following areas: Metrology & calibration. Calibrated items include: Flowmeters that can be calibrated with water, Force / tensile testing machines and presses, working force gauges, Deformation manometers, electromechanical manometers with a pressur…, Dial indicators. Most frequently used calibration principles: Measurement with parallel gauge blocks, Check impacts on a standard anvil, Comparison with a standard digital thermometer in the calibrated de….
At a glance
- Accreditation number: ČIA 2275
- Calibration items in scope: 11
- Types of calibrated items: 9
- Calibration principles used: 10
- City: Praha 9, Czech Republic
- Valid until: 25-03-2027
- Accreditation number
- 2275
- Accreditation body
- ČIA
- Address
- Prosecká 811/76a, Prosek, 190 00 Praha 9
- City
- Praha 9, Czech Republic
- Division
- Laboratorium wzorcujące TZÚS Praha, s.p. – oddział TIS
- Valid until
- 25-03-2027
What this laboratory calibrates (9)
- Flowmeters that can be calibrated with water — 2 calibration items
- Force / tensile testing machines and presses, working force gauges — 2 calibration items
- Deformation manometers, electromechanical manometers with a pressure transducer or with digital indication — 1 calibration item
- Dial indicators — 1 calibration item
- Direct-indicating thermometers — 1 calibration item
- Direct-indicating thermometers, thermometric devices — 1 calibration item
- Electronic length sensors — 1 calibration item
- Hardness / Schmidt type hardness testers (for concrete and other materials) — 1 calibration item
- Slide gauges — 1 calibration item
Calibration principles (10)
- Measurement with parallel gauge blocks — 2 calibration items
- Check impacts on a standard anvil — 1 calibration item
- Comparison with a standard digital thermometer in the calibrated device — 1 calibration item
- Comparison with a standard force-proving instrument — 1 calibration item
- Comparison with a standard piston manometer — 1 calibration item
- Comparison with a standard Pt100 resistance temperature sensor in a water and oil bath — 1 calibration item
- Loading with weights — 1 calibration item
- Mass method — 1 calibration item
- Measurement on a spiral microscope — 1 calibration item
- Volume method — 1 calibration item
Laboratory sections (2)
- Technický a zkušební ústav stavební Praha, s.p. – TIS Branch — 10 calibration items
- Technický a zkušební ústav stavební Praha, s.p. – Branch 0200 — 1 calibration item
Laboratory locations
- Technický a zkušební ústav stavební Praha, s.p. (main site) — Prosecká 811/76a, Prosek, 190 00 Praha 9
- Technický a zkušební ústav stavební Praha, s.p. – oddział TIS — Prosecká 811/76a, 190 00 Praha 9
- Technický a zkušební ústav stavební Praha, s.p. – oddział 0200 — Nemanická 441, 370 10 České Budějovice
Certificate information
- Certificate date
- 30.10.2024
Accreditation scope — calibration items (11)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Technický a zkušební ústav stavební Praha, s.p. – TIS Branch | |||||
| 1 | Dial indicators | Calibration – length 0 mm to 100 mm · CMC 9 µm CMC for the field of measured quantity: Length |
Measurement on a spiral microscope | IP 0960K003 | |
| 2 | Slide gauges | Calibration – length 0 mm to 250 mm · CMC 0,015 mm 250 mm to 450 mm · CMC 0,03 mm 450 mm to 1000 mm · CMC 0,05 mm CMC for the field of measured quantity: Length |
Measurement with parallel gauge blocks | IP 0960K006 | |
| 3 | Electronic length sensors | Calibration – length 0 mm to 250 mm · CMC 0,01 mm CMC for the field of measured quantity: Length |
Measurement with parallel gauge blocks | IP 0960K005 | |
| 4 | Flowmeters that can be calibrated with water | Calibration – flow rate cold water — 0,005 m³/h to 20 m³/h · CMC 0,15 % hot water — 0,005 m³/h to 20 m³/h · CMC 0,20 % cold water — 20 m³/h to 200 m³/h · CMC 0,15 % hot water — 20 m³/h to 200 m³/h · CMC 0,20 % CMC for the field of measured quantity: Flow rate |
Mass method | IP 0960K011 | |
| 5 | Flowmeters that can be calibrated with water | Calibration – flow rate cold water — 0,005 m³/h to 20 m³/h · CMC 0,20 % hot water — 0,005 m³/h to 20 m³/h · CMC 0,25 % cold water — 20 m³/h to 200 m³/h · CMC 0,20 % hot water — 20 m³/h to 200 m³/h · CMC 0,25 % CMC for the field of measured quantity: Flow rate |
Volume method | IP 0960K011 | |
| 6 | Force / tensile testing machines and presses, working force gauges | Calibration – force, mechanical tests Tension — 1 N to 30 N · CMC 0,080 % Compression — 1 N to 30 N · CMC 0,080 % CMC for the field of measured quantity: Force, mechanical tests |
Loading with weights | IP 0960K072 | |
| 7 | Force / tensile testing machines and presses, working force gauges | Calibration – force, mechanical tests Tension — 30 N to 200 N · CMC 0,040 % Tension — 200 N to 20 kN · CMC 0,025 % Tension — 20 kN to 100 kN · CMC 0,060 % Compression — 30 N to 200 N · CMC 0,040 % Compression — 200 N to 2 kN · CMC 0,025 % Compression — 2 kN to 100 kN · CMC 0,040 % Compression — 100 kN to 500 kN · CMC 0,070 % Compression — 500 kN to 1 MN · CMC 0,090 % Compression — 1 MN to 5 MN · CMC 0,025 % CMC for the field of measured quantity: Force, mechanical tests |
Comparison with a standard force-proving instrument | IP 0960K072 | |
| 8 | Hardness / Schmidt type hardness testers (for concrete and other materials) | Calibration – force, mechanical tests Type L — 73 Rₖ to 77 Rₖ · CMC 0,7 Rₖ Type N — 79 Rₖ to 83 Rₖ · CMC 0,7 Rₖ CMC for the field of measured quantity: Force, mechanical tests |
Check impacts on a standard anvil | IP 0960K001 | |
| 9 | Deformation manometers, electromechanical manometers with a pressure transducer or with digital indication | Calibration – pressure gauge pressure, liquid — 0,025 MPa to 60 MPa · CMC 0,08 % CMC for the field of measured quantity: Pressure |
Comparison with a standard piston manometer | IP 0960K018 | |
| 10 | Direct-indicating thermometers | Calibration – temperature 0 °C to 180 °C · CMC 0,2 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard Pt100 resistance temperature sensor in a water and oil bath | IP 0960K020 | |
| Technický a zkušební ústav stavební Praha, s.p. – Branch 0200 | |||||
| 11 | Direct-indicating thermometers, thermometric devices | Calibration – temperature -40 °C to 100 °C · CMC 0,3 °C 100 °C to 250 °C · CMC 0,4 °C 250 °C to 500 °C · CMC 1,4 °C 500 °C to 900 °C · CMC 1,9 °C 900 °C to 1 200 °C · CMC 2,7 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard digital thermometer in the calibrated device | IP 0960K014 | |
Reference standards (9)
IP 0960K001
IP 0960K003
IP 0960K005
IP 0960K006
IP 0960K011
IP 0960K014
IP 0960K018
IP 0960K020
IP 0960K072