🇨🇿 BD SENSORS s.r.o. PL

ČIA 2233 • Buchlovice, Czech Republic • 2 calibration items

Metrology & calibration

BD SENSORS s.r.o. is a calibration laboratory accredited by ČIA under number 2233 (Buchlovice, Czech Republic). Its accreditation scope covers 2 calibration items in the following areas: Metrology & calibration. Calibrated items include: Mechanical pressure gauges and electromechanical manometers, Piston gauges. Most frequently used calibration principles: EURAMET cg-17, EURAMET cg-3 (cross-floating method, calculation of the effective a….

At a glance

  • Accreditation number: ČIA 2233
  • Calibration items in scope: 2
  • Types of calibrated items: 2
  • Calibration principles used: 2
  • City: Buchlovice, Czech Republic
  • Valid until: 20-04-2028
Accreditation number
2233
Accreditation body
ČIA
Address
Hradišťská 817, 687 08 Buchlovice
City
Buchlovice, Czech Republic
Division
Laboratorium wzorcujące
Valid until
20-04-2028

What this laboratory calibrates (2)

Calibration principles (2)

Certificate information

Certificate date
16.12.2025

Accreditation scope — calibration items (2)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Mechanical pressure gauges and electromechanical manometers Calibration – pressure
gauge pressure, gas — -94 kPa to -45,9 kPa · CMC 2,5·10⁻⁵ · (pamb – |pe|) + 1,4 Pa gauge pressure, gas — -45,9 kPa to -10 kPa · CMC 6·10⁻⁵ · |pe| gauge pressure, gas — -10 kPa to -7,5 kPa · CMC 0,6 Pa gauge pressure, gas — -7,5 kPa to -1,35 kPa · CMC 8·10⁻⁵ · |pe| gauge pressure, gas — -1,35 kPa to 1,35 kPa · CMC 0,11 Pa gauge pressure, gas — 1,35 kPa to 5 kPa · CMC 8·10⁻⁵ · pe gauge pressure, gas — 5 kPa to 350 kPa · CMC 2,5·10⁻⁵ · pe + 0,2 Pa gauge pressure, gas — 0,35 MPa to 7 MPa · CMC 2,5·10⁻⁵ · pe + 2 Pa gauge pressure, gas — 7 MPa to 20 MPa · CMC 6·10⁻⁵ · pe absolute pressure, gas — 0,01 kPa to 5 kPa · CMC 1,5 Pa absolute pressure, gas — 5 kPa to 350 kPa · CMC 2,5·10⁻⁵ · p + 1 Pa absolute pressure, gas — 0,35 MPa to 7 MPa · CMC 2,5·10⁻⁵ · p + 2 Pa absolute pressure, gas — 7 MPa to 20 MPa · CMC 6·10⁻⁵ · p + 5 Pa gauge pressure, liquid — 0,4 MPa to 2 MPa · CMC 0,16 kPa gauge pressure, liquid — 2 MPa to 100 MPa · CMC 8·10⁻⁵ · pe gauge pressure, liquid — 100 MPa to 250 MPa · CMC 1,2·10⁻⁴ · pe absolute pressure, liquid — 0,4 MPa to 2 MPa · CMC 0,16 kPa absolute pressure, liquid — 2 MPa to 100 MPa · CMC 8·10⁻⁵ · (p – pamb) absolute pressure, liquid — 100 MPa to 250 MPa · CMC 1,2·10⁻⁴ · (p – pamb)
CMC for the field of measured quantity: Pressure
EURAMET cg-17 KP-801, KP 802
2 Piston gauges Calibration – pressure
gauge pressure, gas — -95 kPa to -10 kPa · CMC 6·10⁻⁵ · |pe| gauge pressure, gas — -10 kPa to -3 kPa · CMC 0,6 Pa gauge pressure, gas — 5 kPa to 350 kPa · CMC 2,5·10⁻⁵ · pe + 0,2 Pa gauge pressure, gas — 0,35 MPa to 7 MPa · CMC 2,5·10⁻⁵ · pe + 2 Pa gauge pressure, gas — 7 MPa to 20 MPa · CMC 6·10⁻⁵ · pe gauge pressure, liquid — 0,1 MPa to 1,5 MPa · CMC 8·10⁻⁵ · pe + 10 Pa gauge pressure, liquid — 1,5 MPa to 100 MPa · CMC 8·10⁻⁵ · pe gauge pressure, liquid — 100 MPa to 250 MPa · CMC 1,2·10⁻⁴ · pe
CMC for the field of measured quantity: Pressure
EURAMET cg-3 (cross-floating method, calculation of the effective area of the pressure balance and mass of the weights) KP-803

Reference standards (3)

KP 802
KP-801
KP-803

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2233