🇨🇿 Institut pro testování a certifikaci, a.s. PL
Metrology & calibration
Institut pro testování a certifikaci, a.s. is a calibration laboratory accredited by ČIA under number 2222 (Uherské Hradiště, Czech Republic). Its accreditation scope covers 151 calibration items in the following areas: Metrology & calibration. Calibrated items include: Direct-indicating thermometers, Thermocouples, Controllers and indicators, Alternating current / current sources. Most frequently used calibration principles: Direct measurement by a length gauge, Direct measurement by a standard microscope, Direct measurement using parallel gauge blocks.
At a glance
- Accreditation number: ČIA 2222
- Calibration items in scope: 151
- Types of calibrated items: 112
- Calibration principles used: 81
- City: Uherské Hradiště, Czech Republic
- Valid until: 07-01-2027
- Accreditation number
- 2222
- Accreditation body
- ČIA
- Address
- Sokolovská 573, 686 01 Uherské Hradiště
- City
- Uherské Hradiště, Czech Republic
- Division
- Calibration laboratory
- Valid until
- 07-01-2027
What this laboratory calibrates (112)
- Direct-indicating thermometers — 8 calibration items
- Thermocouples — 6 calibration items
- Controllers and indicators — 4 calibration items
- Alternating current / current sources — 3 calibration items
- Time, time interval / counters, stopwatches, generators, timers, synthesizers, converters — 3 calibration items
- Absorbance (transmittance) in the UV-VIS-NIR spectrum / photometers and spectrometers, “Elisa readers”, solid and liquid filters for checking absorbance, stray light and spectral resolution — 2 calibration items
- Alternating current / digital, analogue multimeters and ammeters, testing and inspection instruments and their parts, current probes — 2 calibration items
- Alternating current, peak and peak-to-peak value / sources of non-harmonic and pulse signals, pulse testers — 2 calibration items
- Alternating-current voltage / analogue multimeters, voltmeters, oscilloscopes, testing and inspection instruments and their parts — 2 calibration items
- Alternating-current voltage / voltage sources and calibrators, multifunction calibrators, testing and inspection instruments and their parts, high voltage testers — 2 calibration items
- Clamp ammeters and clamp current probes — 2 calibration items
- DC resistance / resistors, calibrators, simulators, testing and inspection instruments and their parts — 2 calibration items
- DC resistance / standard resistors, shunts — 2 calibration items
- Direct current / clamp ammeters and multimeters, clamp current probes — 2 calibration items
- Direct current / current sources, testing and inspection instruments and their parts — 2 calibration items
- and 97 more
Calibration principles (81)
- Direct measurement by a length gauge — 7 calibration items
- Direct measurement by a standard microscope — 7 calibration items
- Direct measurement using parallel gauge blocks — 7 calibration items
- Comparison with a standard resistance sensor in liquid bath — 6 calibration items
- Direct measurement on a calibrator of inspection instruments — 6 calibration items
- Comparison with a standard resistance sensor in dry block — 5 calibration items
- Direct measurement by a standard multimeter — 5 calibration items
- Direct measurement of force on a calibrator and measurement of tip geometry using an image analyzer — 5 calibration items
- Comparison with a standard thermoelectric sensor in a horizontal oven — 4 calibration items
- Generation by a reference power calibrator — 4 calibration items
- Measurement by a vector network analyzer — 4 calibration items
- Comparison with a standard digital manometer — 3 calibration items
- Current generation by a standard calibrator — 3 calibration items
- Direct measurement of standard filters — 3 calibration items
- Direct measurement of standard measuring rings — 3 calibration items
- and 66 more
Laboratory sections (4)
- Uherské Hradiště — 130 calibration items
- Brno — 14 calibration items
- Zlín — 6 calibration items
- Prague — 1 calibration item
Laboratory locations
- Institut pro testování a certifikaci, a.s. (main site) — Sokolovská 573, 686 01 Uherské Hradiště
- Uherské Hradiště — Sokolovská 573, 686 01 Uherské Hradiště
- Zlín — třída Tomáše Bati 5264, 760 01 Zlín
- Prague — Křelovická 970, 104 00 Praha
- Brno — Křižíkova 70, 612 00 Brno
Certificate information
- Certificate date
- 19.12.2025
Accreditation scope — calibration items (151)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Uherské Hradiště | |||||
| 1 | Slide gauges | Calibration – length from 0 mm to 100 mm · CMC 12 µm from 100 mm to 500 mm · CMC 14 µm CMC for the field of measured quantity: Length |
Direct measurement using parallel gauge blocks | KP 1/D | |
| 2 | Depth gauges | Calibration – length from 0 mm to 200 mm · CMC 13 µm CMC for the field of measured quantity: Length |
Direct measurement using parallel gauge blocks | KP 1/D | |
| 3 | Height gauges | Calibration – length from 0 mm to 500 mm · CMC 14 µm CMC for the field of measured quantity: Length |
Direct measurement using parallel gauge blocks | KP 1/D | |
| 4 | Micrometers | Calibration – length from 0 mm to 50 mm · CMC 1,6 µm from 50 mm to 100 mm · CMC 2,0 µm od 100 mm do 200 mm · CMC 3,2 µm CMC for the field of measured quantity: Length |
Direct measurement using parallel gauge blocks | KP 2/D | |
| 5 | Micrometer depth gauges | Calibration – length from 0 mm to 100 mm · CMC 2 µm CMC for the field of measured quantity: Length |
Direct measurement using parallel gauge blocks | KP 2/D | |
| 6 | Three-contact internal gauge with a micrometer screw | Calibration – length from 0 mm to 50 mm · CMC 1,6 µm od 50 mm do 75 mm · CMC 1,8 µm od 75 mm do 100 mm · CMC 2,0 µm CMC for the field of measured quantity: Length |
Direct measurement of standard measuring rings | KP 2/D | |
| 7 | Three-contact internal gauge with a lever mechanism | Calibration – length from 0 mm to 50 mm · CMC 2,1 µm od 50 mm do 75 mm · CMC 2,3 µm od 75 mm do 100 mm · CMC 2,5 µm CMC for the field of measured quantity: Length |
Direct measurement of standard measuring rings | KP 2/D | |
| 8 | Two-contact internal gauge | Calibration – length 0 mm to 35 mm · CMC 2,5 µm 35 mm to 65 mm · CMC 2,6 µm 65 mm to 75 mm · CMC 2,7 µm 75 mm to 100 mm · CMC 2,9 µm CMC for the field of measured quantity: Length |
Direct measurement of standard measuring rings | KP 2/D | |
| 9 | Pasameters | Calibration – length from 0 mm to 25 mm · CMC 0,5 µm od 25 mm do 50 mm · CMC 0,8 µm od 50 mm do 75 mm · CMC 1,2 µm od 75 mm do 100 mm · CMC 1,5 µm od 100 mm do 125 mm · CMC 1,8 µm od 125 mm do 150 mm · CMC 2,2 µm od 150 mm do 175 mm · CMC 2,5 µm od 175 mm do 200 mm · CMC 2,9 µm CMC for the field of measured quantity: Length |
Direct measurement using parallel gauge blocks | KP 2/D | |
| 10 | Micropasameters | Calibration – length from 0 mm to 25 mm · CMC 2,0 µm od 25 mm do 50 mm · CMC 2,5 µm od 50 mm do 75 mm · CMC 3,2 µm od 75 mm do 100 mm · CMC 3,8 µm od 100 mm do 125 mm · CMC 4,5 µm od 125 mm do 150 mm · CMC 5,4 µm od 150 mm do 175 mm · CMC 6,1 µm od 175 mm do 200 mm · CMC 6,9 µm CMC for the field of measured quantity: Length |
Direct measurement using parallel gauge blocks | KP 2/D | |
| 11 | Centesimal indicators | Calibration – length from 0 mm to 100 mm · CMC (2·L + 6) µm CMC for the field of measured quantity: Length |
Comparison with a standard length gauge | KP 3/D | |
| 12 | Millesimal indicators | Calibration – length from 0 mm to 50 mm · CMC (2·L + 1) µm CMC for the field of measured quantity: Length |
Comparison with a standard length gauge | KP 3/D | |
| 13 | Cylindrical gauges | Calibration – length from 0 mm to 200 mm · CMC (3·L + 0,5) µm CMC for the field of measured quantity: Length |
Direct measurement by a length gauge | KP 3/D | |
| 14 | Slot gauges, feeler gauges | Calibration – length from 0 mm to 200 mm · CMC (3·L + 0,5) µm CMC for the field of measured quantity: Length |
Direct measurement by a length gauge | KP 3/D | |
| 15 | Limit plug gauges | Calibration – length from 0 mm to 200 mm · CMC (5·L + 3) µm CMC for the field of measured quantity: Length |
Direct measurement by a length gauge | KP 3/D | |
| 16 | Precision plug gauges | Calibration – length from 0 mm to 50 mm · CMC (1·L + 0,5) µm CMC for the field of measured quantity: Length |
Direct measurement by a length gauge | KP 3/D | |
| 17 | Measuring wires | Calibration – length from 0,17 mm to 6,35 mm · CMC (1·L + 0,5) µm CMC for the field of measured quantity: Length |
Direct measurement by a length gauge | KP 3/D | |
| 18 | Setting rings | Calibration – length 10 mm to 200 mm · CMC (4·L + 1,2) µm CMC for the field of measured quantity: Length |
Direct measurement by a length gauge | KP 3/D | |
| 19 | Micrometer depth gauges | Calibration – length from 0 mm to 100 mm · CMC 2,0 µm CMC for the field of measured quantity: Length |
Direct measurement by a length gauge | KP 3/D | |
| 20 | Steel rules | Calibration – length from 0 mm to 250 mm · CMC 0,05 mm CMC for the field of measured quantity: Length |
Direct measurement by a standard microscope | KP 4/D | |
| 21 | Rules – straightness | Calibration – length from 0 mm to 250 mm · CMC 0,007 mm CMC for the field of measured quantity: Length |
Direct measurement by a standard microscope | KP 4/D | |
| 22 | Thread gauges | Calibration – length from 0 mm to 300 mm · CMC 0,01 mm CMC for the field of measured quantity: Length |
Direct measurement by a standard microscope | KP 4/D | |
| 23 | Radius gauges | Calibration – length from 0 mm to 300 mm · CMC 0,01 mm CMC for the field of measured quantity: Length |
Direct measurement by a standard microscope | KP 4/D | |
| 24 | Profile gauges | Calibration – length from 0 mm to 300 mm · CMC (1·L + 0,5) µm CMC for the field of measured quantity: Length |
Direct measurement by a standard microscope | KP 4/D | |
| 25 | Parallel gauge blocks | Calibration – length from 0,5 mm to 100 mm · CMC (2·L + 0,2) µm CMC for the field of measured quantity: Length |
Comparison with standard parallel gauge blocks | KP 21/D | |
| 26 | Angles | Calibration – plane angle from 0° to 180° · CMC 2,5′ CMC for the field of measured quantity: Plane angle |
Direct measurement by a standard microscope | KP 4/D | |
| 27 | Angle gauges | Calibration – plane angle from 0° to 360° · CMC 2,5′ CMC for the field of measured quantity: Plane angle |
Direct measurement by a standard microscope | KP 4/D | |
| 28 | Speed / analogue and digital contact and contactless revolution counters, rpm sensors, stroboscopes | Calibration – mechanical motion (speed) 1 min⁻¹ to 10 000 min⁻¹ · CMC 0,006 min⁻¹ 10 000 min⁻¹ to 999 999 min⁻¹ · CMC 0,03 min⁻¹ CMC for the field of measured quantity: Mechanical motion (speed) |
Direct measurement of optical signal or direct comparison with a standard revolution counter | KP 6/F | |
| 29 | Speed / analogue and digital contact and contactless revolution counters, rpm sensors, stroboscopes | Calibration – mechanical motion (speed) 2 min⁻¹ to 999,99 min⁻¹ · CMC 0,1 min⁻¹ 1 000 min⁻¹ to 9 999 min⁻¹ · CMC 0,9 min⁻¹ 10 000 min⁻¹ to 999 999 min⁻¹ · CMC 7,0 min⁻¹ CMC for the field of measured quantity: Mechanical motion (speed) |
Measurement by a standard revolution counter | KP 6/F | |
| 30 | Deformation pressure gauges | Calibration – pressure, mechanical stress Overpressure; gas — 0 kPa to 10 kPa · CMC 0,015 % + 0,058 kPa Overpressure; gas — 10 kPa to 400 kPa · CMC 0,046 % + 0,07 kPa Overpressure; gas — 0,4 MPa to 2 MPa · CMC 0,046 % + 0,18 kPa Overpressure; gas — 2 MPa to 6 MPa · CMC 0,015 % + 0,42 kPa Underpressure; gas — 0 kPa to 95 kPa · CMC 0,046 % + 0,07 kPa Absolute pressure; gas — 70 kPa to 130 kPa · CMC 0,015 % + 0,12 kPa Absolute pressure; gas — 130 kPa to 520 kPa · CMC 0,046 % + 0,13 kPa Absolute pressure; gas — 0,52 MPa to 2,12 MPa · CMC 0,046 % + 0,21 kPa Absolute pressure; gas — 2,12 MPa to 6 MPa · CMC 0,015 % + 0,44 kPa Overpressure, absolute pressure; gas, liquid — 6 MPa to 16 MPa · CMC 0,015 % + 1,2 kPa Overpressure, absolute pressure; liquid — 16 MPa to 60 MPa · CMC 0,046 % + 11 kPa Overpressure, absolute pressure; liquid — 60 MPa to 100 MPa · CMC 0,012 % + 8,3 kPa Barometric pressure — 70 kPa to 120 kPa · CMC 0,082 kPa CMC for the field of measured quantity: Pressure, mechanical stress |
Comparison with a standard digital manometer | KP 1/P | |
| 31 | Pressure converters | Calibration – pressure, mechanical stress Overpressure; gas — 0 kPa to 10 kPa · CMC 0,015 % + 0,0018 kPa Overpressure; gas — 10 kPa to 400 kPa · CMC 0,048 % + 0,035 kPa Overpressure; gas — 0,4 MPa to 2 MPa · CMC 0,048 % + 0,12 kPa Overpressure; gas — 2 MPa to 6 MPa · CMC 0,015 % + 0,35 kPa Underpressure; gas — 0 kPa to 95 kPa · CMC 0,048 % + 0,035 kPa Absolute pressure; gas — 70 kPa to 130 kPa · CMC 0,015 % + 0,1 kPa Absolute pressure; gas — 130 kPa to 520 kPa · CMC 0,048 % + 0,11 kPa Absolute pressure; gas — 0,52 MPa to 2,12 MPa · CMC 0,048 % + 0,16 kPa Absolute pressure; gas — 2,12 MPa to 6 MPa · CMC 0,015 % + 0,36 kPa Overpressure, absolute pressure; gas, liquid — 6 MPa to 16 MPa · CMC 0,015 % + 0,93 kPa Overpressure, absolute pressure; liquid — 16 MPa to 60 MPa · CMC 0,048 % + 11 kPa Overpressure, absolute pressure; liquid — 60 MPa to 100 MPa · CMC 0,012 % + 8,1 kPa Barometric pressure — 70 kPa to 120 kPa · CMC 0,12 % CMC for the field of measured quantity: Pressure, mechanical stress |
Comparison with a standard digital manometer | KP 2/P | |
| 32 | Digital manometers | Calibration – pressure, mechanical stress Overpressure; gas — 0 kPa to 10 kPa · CMC 0,015 % + 0,0018 kPa Overpressure; gas — 10 kPa to 400 kPa · CMC 0,046 % + 0,039 kPa Overpressure; gas — 0,4 MPa to 2 MPa · CMC 0,046 % + 0,14 kPa Overpressure; gas — 2 MPa to 6 MPa · CMC 0,015 % + 0,36 kPa Underpressure; gas — 0 kPa to 95 kPa · CMC 0,046 % + 0,039 kPa Absolute pressure; gas — 70 kPa to 130 kPa · CMC 0,015 % + 0,1 kPa Absolute pressure; gas — 130 kPa to 520 kPa · CMC 0,046 % + 0,11 kPa Absolute pressure; gas — 0,52 MPa to 2,12 MPa · CMC 0,046 % + 0,17 kPa Absolute pressure; gas — 2,12 MPa to 6 MPa · CMC 0,015 % + 0,37 kPa Overpressure, absolute pressure; gas, liquid — 6 MPa to 16 MPa · CMC 0,015 % + 0,98 kPa Overpressure, absolute pressure; liquid — 16 MPa to 60 MPa · CMC 0,046 % + 0,11 kPa Overpressure, absolute pressure; liquid — 60 MPa to 100 MPa · CMC 0,012 % + 8,2 kPa Barometric pressure — 70 kPa to 120 kPa · CMC 0,058 kPa CMC for the field of measured quantity: Pressure, mechanical stress |
Comparison with a standard digital manometer | KP 3/P | |
| 33 | Thermocouples | Calibration – temperature from −40 °C to 0 °C · CMC 0,6 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in liquid bath | KP 1/T | |
| 34 | Thermocouples | Calibration – temperature 0 °C · CMC 0,6 °C from 0 °C to 100 °C · CMC 0,6 °C from 100 °C to 250 °C · CMC 0,7 °C od 250 °C do 600 °C · CMC 0,8 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in dry block | KP 1/T | |
| 35 | Thermocouples | Calibration – temperature from 600 °C to 800 °C · CMC 1,5 °C od 800 °C do 1 200 °C · CMC 2,2 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard thermoelectric sensor in a horizontal oven | KP 1/T | |
| 36 | Glass thermometers | Calibration – temperature from −40 °C to 0 °C · CMC 0,20 °C 0 °C · CMC 0,05 °C from 0 °C to 100 °C · CMC 0,14 °C from 100 °C to 250 °C · CMC 0,15 °C from 250 °C to 300 °C · CMC 0,30 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in liquid bath | KP 3/T | |
| 37 | Direct-indicating thermometers | Calibration – temperature −80 °C to −40 °C · CMC 0,3 °C CMC for the field of measured quantity: Temperature |
Comparison with a reference direct indicating temperature sensor in an electrical vertical furnace | KP 4/T | |
| 38 | Direct-indicating thermometers | Calibration – temperature from −40 °C to 0 °C · CMC 0,07 °C 0 °C · CMC 0,06 °C from 0 °C to 100 °C · CMC 0,09 °C from 100 °C to 250 °C · CMC 0,10 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in liquid bath | KP 4/T | |
| 39 | Direct-indicating thermometers | Calibration – temperature from 250 °C to 350 °C · CMC 0,44 °C od 350 °C do 600 °C · CMC 0,62 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in dry block | KP 4/T | |
| 40 | Direct-indicating thermometers | Calibration – temperature from 600 °C to 800 °C · CMC 1,6 °C od 800 °C do 1 200 °C · CMC 2,2 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard thermoelectric sensor in a horizontal oven | KP 4/T | |
| 41 | Controllers and indicators | Calibration – temperature Type K — −270 °C to 0 °C · CMC 0,3 °C Type K — 0 °C to 1370 °C · CMC 0,3 °C Type S — −50 °C to 0 °C · CMC 0,9 °C Type S — 0 °C to 1 780 °C · CMC 0,6 °C Type B — 200 °C to 500 °C · CMC 1,5 °C Type B — 500 °C to 1 820 °C · CMC 0,6 °C Type R — 0 °C to 1 768 °C · CMC 0,6 °C Type J — −210 °C to 0 °C · CMC 0,2 °C Type J — 0 °C to 1 200 °C · CMC 0,2 °C Type N — −270 °C to 0 °C · CMC 0,2 °C Type N — 0 °C to 1 300 °C · CMC 0,2 °C CMC for the field of measured quantity: Temperature |
Simulation of thermoelectric temperature sensors using DC voltage | KP 5/T | |
| 42 | Controllers and indicators | Calibration – temperature RTD — −200 °C to −50 °C · CMC 0,05 °C RTD — −50 °C to 850 °C · CMC 0,05 °C CMC for the field of measured quantity: Temperature |
Simulation of resistance temperature sensors using DC electrical resistors | KP 5/T | |
| 43 | Non-contact thermometers | Calibration – temperature od −30 °C do 23 °C · CMC 2,2 °C od 23 °C do 150 °C · CMC 2,1 °C from 150 °C to 250 °C · CMC 2,2 °C from 250 °C to 350 °C · CMC 2,5 °C from 350 °C to 500 °C · CMC 3,0 °C CMC for the field of measured quantity: Temperature |
Direct measurement on a black body, comparison with a standard contactless thermometer | KP 6/T | |
| 44 | Direct-indicating thermometers integrated in conditioning and thermal chambers and baths | Calibration – temperature od −80 °C do 250 °C · CMC 1,0 °C od 250 °C do 600 °C · CMC 1,1 °C from 600 °C to 700 °C · CMC 1,7 °C od 700 °C do 850 °C · CMC 1,9 °C od 850 °C do 1 200 °C · CMC 2,9 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard digital thermometer in a chamber (characterisation) | KP 22/T | |
| 45 | Direct-indicating thermometers and temperature measuring chains integrated in conditioning and thermal chambers and baths | Calibration – temperature od −20 °C do 250 °C · CMC 1,0 °C od 250 °C do 600 °C · CMC 1,1 °C from 600 °C to 700 °C · CMC 1,7 °C od 700 °C do 850 °C · CMC 1,9 °C od 850 °C do 1 000 °C · CMC 2,9 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard digital thermometer in a chamber (characterisation according to the standard AMS 2750) | KP 21/T | |
| 46 | Temperature controllers and indicators of conditioning and thermal chambers and baths | Calibration – temperature Type K — −20 °C to 1 370 °C · CMC 0,3 °C Type N — −20 °C to 1 300 °C · CMC 0,2 °C Type S — 0 °C to 1 780 °C · CMC 0,6 °C Type B — 200 °C to 500 °C · CMC 1,5 °C Type B — 500 °C to 1 800 °C · CMC 0,6 °C CMC for the field of measured quantity: Temperature |
Simulation of thermoelectric temperature sensors using DC voltage (characterisation according to the standard AMS 2750) | KP 21/T | |
| 47 | Relative humidity / humidity meters | Calibration – air humidity (20 to 50) °C — 5 % RH to 70 % RH · CMC 2,2 % RH (20 to 50) °C — 70 % RH to 95 % RH · CMC 3,2 % RH CMC for the field of measured quantity: Air humidity |
Direct comparison with a reference hygrometer in a climatic chamber | KP 1/V | |
| 48 | Direct-current voltage / direct-current voltage meters (digital and analogue multimeters, voltmeters, oscilloscopes, testing and inspection instruments and their parts) | Calibration – electrical quantities from 0 mV to 1 mV · CMC 1,3 µV from 1 mV to 20 mV · CMC 1,7 µV from 20 mV to 100 mV · CMC 2,1 µV from 100 mV to 200 mV · CMC 3,4 µV od 200 mV do 500 mV · CMC 0,5 µV od 500 mV do 900 mV · CMC 0,0011 % od 0,9 V do 1,5 V · CMC 0,0009 % from 1,5 V to 2 V · CMC 0,0008 % od 2 V do 4 V · CMC 0,0007 % od 4 V do 10 V · CMC 0,00055 % from 10 V to 20 V · CMC 0,00051 % od 20 V do 40 V · CMC 0,0011 % od 40 V do 100 V · CMC 0,000 94 % from 100 V to 200 V · CMC 0,000 82 % od 200 V do 400 V · CMC 0,0014 % od 400 V do 1100 V · CMC 0,0013 % od 1100 V do 6000 V · CMC 0,6 % CMC for the field of measured quantity: Electrical quantities |
Voltage generation by a standard calibrator | KP 1/E, KP 2/E, KP 3/E, KP 18/E | |
| 49 | Direct-current voltage / direct-current voltage sources and calibrators, multifunction calibrators, testing and inspection instruments and their parts, high voltage testers | Calibration – electrical quantities from 0 mV to 120 mV · CMC 0,00096 % + 0,71 µV od 120 mV do 1 200 mV · CMC 0,00072 % + 4,6 µV from 1,2 V to 12 V · CMC 0,00060 % + 4,6 µV from 12 V to 120 V · CMC 0,00088 % + 29 µV od 120 V do 1 100 V · CMC 0,00098 % + 82 µV CMC for the field of measured quantity: Electrical quantities |
Direct measurement by a standard multimeter | KP 8/E | |
| 50 | Direct-current voltage / direct-current voltage sources and calibrators, multifunction calibrators, testing and inspection instruments and their parts, high voltage testers | Calibration – electrical quantities 1100 V to 15 kV · CMC 0,6 % CMC for the field of measured quantity: Electrical quantities |
Measurement by a standard multimeter with a HV probe | KP 8/E | |
| 51 | Direct current / direct-current meters (digital and analogue multimeters, ammeters, testing and inspection instruments and their parts, current probes) | Calibration – electrical quantities od 0 µA do 2 µA · CMC 3,8 nA od 2 µA do 10 µA · CMC 4,5 nA from 10 µA to 20 µA · CMC 5,5 nA od 20 µA do 40 µA · CMC 0,027 % od 40 µA do 100 µA · CMC 0,019 % from 100 µA to 200 µA · CMC 0,015 % od 200 µA do 200 mA · CMC 0,011 % od 200 mA do 400 mA · CMC 0,027 % from 400 mA to 1 A · CMC 0,019 % from 1 A to 2 A · CMC 0,015 % from 2 A to 3 A · CMC 0,045 % CMC for the field of measured quantity: Electrical quantities |
Current generation by a standard calibrator | KP 1/E, KP 2/E, KP 3/E | |
| 52 | Direct current / direct-current meters (digital and analogue multimeters, ammeters, testing and inspection instruments and their parts, current probes) | Calibration – electrical quantities from 3 A to 4 A · CMC 0,038 % od 4 A do 30 A · CMC 0,035 % od 30 A do 50 A · CMC 0,15 % od 50 A do 100 A · CMC 0,28 % CMC for the field of measured quantity: Electrical quantities |
Comparison with current measured by a standard shunt with multimeter | KP 1/E, KP 2/E, KP 3/E | |
| 53 | Direct current / clamp ammeters and multimeters, clamp current probes | Calibration – electrical quantities 1 A to 5 A · CMC 0,2 % CMC for the field of measured quantity: Electrical quantities |
Current generation by a standard calibrator | KP 1/E, KP 2/E, KP 3/E | |
| 54 | Direct current / clamp ammeters and multimeters, clamp current probes | Calibration – electrical quantities 5 A to 750 A · CMC 0,13 % CMC for the field of measured quantity: Electrical quantities |
Simulation of reference current using calibrator and current coil | KP 1/E, KP 2/E, KP 3/E | |
| 55 | Direct current / current sources, testing and inspection instruments and their parts | Calibration – electrical quantities od 0 µA do 12 µA · CMC 0,0045 % + 6,0 nA from 12 µA to 120 µA · CMC 0,0050 % + 6,7 nA from 120 µA to 1,2 mA · CMC 0,0050 % + 4,5 nA from 1,2 mA to 12 mA · CMC 0,0044 % + 0,045 µA from 12 mA to 120 mA · CMC 0,0065 % + 0,45 µA od 120 mA do 1,05 A · CMC 0,016 % + 9,8 µA CMC for the field of measured quantity: Electrical quantities |
Direct measurement by a standard multimeter | KP 1/E, KP 2/E, KP 3/E | |
| 56 | Direct current / current sources, testing and inspection instruments and their parts | Calibration – electrical quantities 1,05 A to 12 A · CMC 0,0073 % + 25 µA 12 A to 20 A · CMC 0,0058 % + 4 µA 20 A to 50 A · CMC 0,042 % + 61 µA 50 A to 100 A · CMC 0,047 % CMC for the field of measured quantity: Electrical quantities |
Indirect measurement using a standard shunt and multimeter | KP 8/E | |
| 57 | Direct current / current sources | Calibration – electrical quantities 1 A to 40 A · CMC 3,8 % 40 A to 100 A · CMC 5 % 100 A to 250 A · CMC 4 % 250 A to 400 A · CMC 2,5 % 400 A to 1 000 A · CMC 2 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement by a standard clamp multimeter | KP 8/E | |
| 58 | Alternating-current voltage / analogue multimeters, voltmeters, oscilloscopes, testing and inspection instruments and their parts | Calibration – electrical quantities 10 Hz to 100 kHz — 0,1 mV to 1 mV · CMC 8 µV 100 kHz to 300 kHz — 0,1 mV to 1 mV · CMC 18 µV 300 kHz to 1 MHz — 0,1 mV to 1 mV · CMC 34 µV 10 Hz to 100 kHz — 1 mV to 10 mV · CMC 15 µV 100 kHz to 300 kHz — 1 mV to 10 mV · CMC 30 µV 300 kHz to 1 MHz — 1 mV to 10 mV · CMC 70 µV 10 Hz to 30 kHz — 10 mV to 20 mV · CMC 13 µV 30 kHz to 100 kHz — 10 mV to 20 mV · CMC 18 µV 100 kHz to 300 kHz — 10 mV to 20 mV · CMC 45 µV 300 kHz to 1 MHz — 10 mV to 20 mV · CMC 110 µV 10 Hz to 30 kHz — 20 mV to 40 mV · CMC 17 µV 30 kHz to 100 kHz — 20 mV to 40 mV · CMC 29 µV 100 kHz to 300 kHz — 20 mV to 40 mV · CMC 75 µV 300 kHz to 1 MHz — 20 mV to 40 mV · CMC 270 µV 10 Hz to 30 kHz — 40 mV to 200 mV · CMC 0,045 % 30 kHz to 100 kHz — 40 mV to 200 mV · CMC 0,075 % 100 kHz to 300 kHz — 40 mV to 200 mV · CMC 0,2 % 300 kHz to 1 MHz — 40 mV to 200 mV · CMC 0,7 % 10 Hz to 300 Hz — 200 mV to 400 mV · CMC 0,04 % 300 Hz to 30 kHz — 200 mV to 400 mV · CMC 0,03 % 30 kHz to 100 kHz — 200 mV to 400 mV · CMC 0,04 % 100 kHz to 300 kHz — 200 mV to 400 mV · CMC 0,1 % 300 kHz to 1 MHz — 200 mV to 400 mV · CMC 0,55 % 10 Hz to 300 Hz — 0,4 V to 1 V · CMC 0,022 % 300 Hz to 30 kHz — 0,4 V to 1 V · CMC 0,016 % 30 kHz to 100 kHz — 0,4 V to 1 V · CMC 0,025 % 100 kHz to 300 kHz — 0,4 V to 1 V · CMC 0,07 % 300 kHz to 1 MHz — 0,4 V to 1 V · CMC 0,47 % 10 Hz to 300 Hz — 1 V to 2 V · CMC 0,015 % 300 Hz to 30 kHz — 1 V to 2 V · CMC 0,01 % 30 kHz to 100 kHz — 1 V to 2 V · CMC 0,02 % 100 kHz to 300 kHz — 1 V to 2 V · CMC 0,05 % 300 kHz to 1 MHz — 1 V to 2 V · CMC 0,43 % 10 Hz to 300 Hz — 2 V to 4 V · CMC 0,03 % 300 Hz to 30 kHz — 2 V to 4 V · CMC 0,02 % 30 kHz to 100 kHz — 2 V to 4 V · CMC 0,03 % 100 kHz to 300 kHz — 2 V to 4 V · CMC 0,11 % 300 kHz to 1 MHz — 2 V to 4 V · CMC 0,55 % 10 Hz to 300 Hz — 4 V to 10 V · CMC 0,02 % 300 Hz to 30 kHz — 4 V to 10 V · CMC 0,013 % 30 kHz to 100 kHz — 4 V to 10 V · CMC 0,02 % 100 kHz to 300 kHz — 4 V to 10 V · CMC 0,086 % 300 kHz to 1 MHz — 4 V to 10 V · CMC 0,47 % 10 Hz to 300 Hz — 10 V to 20 V · CMC 0,015 % 300 Hz to 30 kHz — 10 V to 20 V · CMC 0,01 % 30 kHz to 100 kHz — 10 V to 20 V · CMC 0,015 % 100 kHz to 300 kHz — 10 V to 20 V · CMC 0,075 % 300 kHz to 1 MHz — 10 V to 20 V · CMC 0,42 % 10 Hz to 300 Hz — 20 V to 40 V · CMC 0,03 % 300 Hz to 30 kHz — 20 V to 40 V · CMC 0,019 % 30 kHz to 100 kHz — 20 V to 40 V · CMC 0,035 % 100 kHz to 300 kHz — 20 V to 40 V · CMC 0,11 % 300 kHz to 1 MHz — 20 V to 40 V · CMC 0,15 % 10 Hz to 300 Hz — 40 V to 100 V · CMC 0,022 % 300 Hz to 10 kHz — 40 V to 100 V · CMC 0,01 % 10 kHz to 30 kHz — 40 V to 100 V · CMC 0,014 % 30 kHz to 100 kHz — 40 V to 100 V · CMC 0,03 % 100 kHz to 200 kHz — 40 V to 100 V · CMC 0,08 % 10 Hz to 300 Hz — 100 V to 200 V · CMC 0,02 % 300 Hz to 30 kHz — 100 V to 200 V · CMC 0,01 % 30 kHz to 100 kHz — 100 V to 200 V · CMC 0,023 % 100 kHz to 200 kHz — 100 V to 200 V · CMC 0,065 % 45 Hz to 300 Hz — 200 V to 750 V · CMC 0,03 % 300 Hz to 10 kHz — 200 V to 750 V · CMC 0,025 % 10 kHz to 30 kHz — 200 V to 750 V · CMC 0,03 % 30 kHz to 100 kHz — 200 V to 750 V · CMC 0,15 % 45 Hz to 300 Hz — 750 V to 1100 V · CMC 0,027 % 300 Hz to 10 kHz — 750 V to 1100 V · CMC 0,022 % 10 kHz to 30 kHz — 750 V to 1100 V · CMC 0,025 % CMC for the field of measured quantity: Electrical quantities |
Voltage generation by a standard calibrator | KP 1/E, KP 2/E, KP 3/E, KP 18/E | |
| 59 | Alternating-current voltage / analogue multimeters, voltmeters, oscilloscopes, testing and inspection instruments and their parts | Calibration – electrical quantities 40 Hz to 100 Hz — 1100 V to 5000 V · CMC 2 % CMC for the field of measured quantity: Electrical quantities |
Comparison with a standard multimeter with a HV probe | KP 1/E, KP 2/E, KP 3/E, KP 18/E | |
| 60 | Alternating-current voltage / voltage sources and calibrators, multifunction calibrators, testing and inspection instruments and their parts, high voltage testers | Calibration – electrical quantities 1 Hz to 40 Hz — 0,1 mV to 12 mV · CMC 0,062 % + 3,5 µV 40 Hz to 1 000 Hz — 0,1 mV to 12 mV · CMC 0,048 % + 1,4 µV 1 kHz to 20 kHz — 0,1 mV to 12 mV · CMC 0,048 % + 1,4 µV 20 kHz to 50 kHz — 0,1 mV to 12 mV · CMC 0,13 % + 1,4 µV 50 kHz to 100 kHz — 0,1 mV to 12 mV · CMC 0,6 % + 1,4 µV 100 kHz to 300 kHz — 0,1 mV to 12 mV · CMC 4,7 % + 2,4 µV 1 Hz to 40 Hz — 12 mV to 120 mV · CMC 0,036 % + 2,8 µV 40 Hz to 1 000 Hz — 12 mV to 120 mV · CMC 0,017 % + 1,7 µV 1 kHz to 20 kHz — 12 mV to 120 mV · CMC 0,017 % + 1,7 µV 20 kHz to 50 kHz — 12 mV to 120 mV · CMC 0,038 % + 1,7 µV 50 kHz to 100 kHz — 12 mV to 120 mV · CMC 0,13 % + 1,9 µV 100 kHz to 300 kHz — 12 mV to 120 mV · CMC 0,37 % + 12 µV 300 kHz to 1 MHz — 12 mV to 120 mV · CMC 1,2 % + 11 µV 1 Hz to 40 Hz — 120 mV to 1,2 V · CMC 0,028 % + 32 µV 40 Hz to 1 000 Hz — 120 mV to 1,2 V · CMC 0,01 % + 22 µV 1 kHz to 20 kHz — 120 mV to 1,2 V · CMC 0,01 % + 22 µV 20 kHz to 50 kHz — 120 mV to 1,2 V · CMC 0,035 % + 22 µV 50 kHz to 100 kHz — 120 mV to 1,2 V · CMC 0,093 % + 23 µV 100 kHz to 300 kHz — 120 mV to 1,2 V · CMC 0,35 % + 120 µV 300 kHz to 1 MHz — 120 mV to 1,2 V · CMC 1,2 % + 80 µV 1 Hz to 40 Hz — 1,2 V to 12 V · CMC 0,028 % + 320 µV 40 Hz to 1 000 Hz — 1,2 V to 12 V · CMC 0,0093 % + 220 µV 1 kHz to 20 kHz — 1,2 V to 12 V · CMC 0,0093 % + 220 µV 20 kHz to 50 kHz — 1,2 V to 12 V · CMC 0,035 % + 220 µV 50 kHz to 100 kHz — 1,2 V to 12 V · CMC 0,093 % + 230 µV 100 kHz to 300 kHz — 1,2 V to 12 V · CMC 0,35 % + 1,2 mV 300 kHz to 1 MHz — 1,2 V to 12 V · CMC 1,2 % + 0,77 mV 1 Hz to 40 Hz — 12 V to 120 V · CMC 0,036 % + 3,8 mV 40 Hz to 1 000 Hz — 12 V to 120 V · CMC 0,024 % + 2,2 mV 1 kHz to 20 kHz — 12 V to 120 V · CMC 0,024 % + 2,2 mV 20 kHz to 50 kHz — 12 V to 120 V · CMC 0,041 % + 2,2 mV 50 kHz to 100 kHz — 12 V to 120 V · CMC 0,14 % + 2,3 mV 100 kHz to 300 kHz — 12 V to 120 V · CMC 0,47 % + 12 mV 300 kHz to 1 MHz — 12 V to 120 V · CMC 1,2 % + 16 mV 1 Hz to 40 Hz — 120 V to 700 V · CMC 0,047 % + 46 mV 40 Hz to 1 000 Hz — 120 V to 700 V · CMC 0,047 % + 23 mV 1 kHz to 20 kHz — 120 V to 700 V · CMC 0,047 % + 23 mV 20 kHz to 50 kHz — 120 V to 700 V · CMC 0,014 % + 23 mV 50 kHz to 100 kHz — 120 V to 700 V · CMC 0,24 % + 23 mV CMC for the field of measured quantity: Electrical quantities |
Direct measurement by a standard multimeter | KP 8/E | |
| 61 | Alternating-current voltage / voltage sources and calibrators, multifunction calibrators, testing and inspection instruments and their parts, high voltage testers | Calibration – electrical quantities 40 Hz to 100 Hz — 1100 V to 11 kV · CMC 2 % CMC for the field of measured quantity: Electrical quantities |
Measurement by a standard multimeter with a HV probe | KP 8/E | |
| 62 | Alternating-current voltage, peak and peak-to-peak value of periodic non-harmonic signals / sources of non-harmonic and pulse signals, pulse testers | Calibration – electrical quantities Bandwidth: 1 Hz to 300 kHz; Impedance: Z = 50 Ω — 10 mV to 5 V · CMC 1,8 % Bandwidth: 300 kHz to 2 GHz; Impedance: Z = 50 Ω — 10 mV to 5 V · CMC 4,0 % Bandwidth: 1 Hz to 300 kHz; Impedance: Z = 50 Ω — 5 V to 38 V · CMC 2,9 % Bandwidth: 300 kHz to 1 GHz; Impedance: Z = 50 Ω — 5 V to 38 V · CMC 4,5 % Bandwidth: 1 Hz to 300 kHz; Impedance: Z = 1 MΩ — 10 mV to 40 V · CMC 1,8 % Bandwidth: 300 kHz to 400 MHz; Impedance: Z = 1 MΩ — 10 mV to 40 V · CMC 4,0 % Bandwidth: 1 Hz to 400 MHz; Impedance: Z = 10 MΩ — 40 V to 400 V · CMC 4,5 % Bandwidth: 1 Hz to 400 MHz; Impedance: Z = 66,7 MΩ — 40 V to 1000 V · CMC 4,5 % CMC for the field of measured quantity: Electrical quantities |
Measurement by a standard oscilloscope | KP 8/E | |
| 63 | Pulse amplitude, peak and peak-to-peak value of voltage / sources of non-harmonic and pulse signals, pulse testers | Calibration – electrical quantities Z = 1 MΩ; rise time >1,3 ns — 10 mV to 40 V · CMC 4,0 % Z = 50 kΩ; rise time >140 ns — 10 V to 5000 V · CMC 5,6 % Z = 1 kΩ; rise time >4,2 ns — 10 V to 5000 V · CMC 5,0 % Z = 50 Ω; rise time >3,5 ns — 1 V to 5000 V · CMC 4,9 % CMC for the field of measured quantity: Electrical quantities |
Measurement by a standard oscilloscope with a voltage probe | KP 8/E | |
| 64 | Alternating current, peak and peak-to-peak value / sources of non-harmonic and pulse signals, pulse testers | Calibration – electrical quantities 1 Hz to 20 kHz — 100 mA to 1,5 A · CMC 2,7 % 1 Hz to 20 kHz — 1,5 A to 15 A · CMC 2,4 % CMC for the field of measured quantity: Electrical quantities |
Measurement by a standard oscilloscope with a shunt or clamp probe | KP 8/E | |
| 65 | Alternating current, peak and peak-to-peak value / sources of non-harmonic and pulse signals, pulse testers | Calibration – electrical quantities rise time >140 ns — 1 A to 2 500 A · CMC 4,9 % CMC for the field of measured quantity: Electrical quantities |
Measurement by a standard oscilloscope with a current shunt | KP 8/E | |
| 66 | Alternating current / digital, analogue multimeters and ammeters, testing and inspection instruments and their parts, current probes | Calibration – electrical quantities 10 Hz to 1 kHz — 1 µA to 10 µA · CMC 0,015 µA 1 kHz to 5 kHz — 1 µA to 10 µA · CMC 0,02 µA 10 Hz to 1 kHz — 10 µA to 50 µA · CMC 0,025 µA 1 kHz to 5 kHz — 10 µA to 50 µA · CMC 0,035 µA 10 Hz to 1 kHz — 50 µA to 100 µA · CMC 0,035 µA 1 kHz to 5 kHz — 50 µA to 100 µA · CMC 0,06 µA 10 Hz to 1 kHz — 100 µA to 200 µA · CMC 0,035 % 1 kHz to 5 kHz — 100 µA to 200 µA · CMC 0,06 % 10 Hz to 1 kHz — 200 µA to 400 µA · CMC 0,071 % 1 kHz to 5 kHz — 200 µA to 400 µA · CMC 0,083 % 10 Hz to 1 kHz — 400 µA to 800 µA · CMC 0,045 % 1 kHz to 5 kHz — 400 µA to 800 µA · CMC 0,06 % 10 Hz to 1 kHz — 800 µA to 2 mA · CMC 0,03 % 1 kHz to 5 kHz — 800 µA to 2 mA · CMC 0,045 % 10 Hz to 1 kHz — 2 mA to 3 mA · CMC 0,075 % 1 kHz to 5 kHz — 2 mA to 3 mA · CMC 0,085 % 10 Hz to 1 kHz — 3 mA to 4 mA · CMC 0,055 % 1 kHz to 5 kHz — 3 mA to 4 mA · CMC 0,065 % 10 Hz to 1 kHz — 4 mA to 8 mA · CMC 0,045 % 1 kHz to 5 kHz — 4 mA to 8 mA · CMC 0,06 % 10 Hz to 1 kHz — 8 mA to 20 mA · CMC 0,033 % 1 kHz to 5 kHz — 8 mA to 20 mA · CMC 0,045 % 10 Hz to 1 kHz — 20 mA to 30 mA · CMC 0,075 % 1 kHz to 5 kHz — 20 mA to 30 mA · CMC 0,085 % 10 Hz to 1 kHz — 30 mA to 40 mA · CMC 0,055 % 1 kHz to 5 kHz — 30 mA to 40 mA · CMC 0,065 % 10 Hz to 1 kHz — 40 mA to 80 mA · CMC 0,045 % 1 kHz to 5 kHz — 40 mA to 80 mA · CMC 0,055 % 10 Hz to 1 kHz — 80 mA to 200 mA · CMC 0,03 % 1 kHz to 5 kHz — 80 mA to 200 mA · CMC 0,045 % 10 Hz to 1 kHz — 200 mA to 300 mA · CMC 0,15 % 1 kHz to 5 kHz — 200 mA to 300 mA · CMC 2 % 10 Hz to 1 kHz — 300 mA to 800 mA · CMC 0,1 % 1 kHz to 5 kHz — 300 mA to 800 mA · CMC 0,15 % 10 Hz to 1 kHz — 800 mA to 2 000 mA · CMC 0,06 % 1 kHz to 5 kHz — 800 mA to 2 000 mA · CMC 0,11 % 10 Hz to 1 kHz — 2 A to 4 A · CMC 0,1 % 1 kHz to 5 kHz — 2 A to 4 A · CMC 0,2 % 5 kHz to 10 kHz — 2 A to 4 A · CMC 0,5 % 10 kHz to 20 kHz — 2 A to 4 A · CMC 2 % 10 Hz to 1 kHz — 4 A to 6 A · CMC 0,07 % 1 kHz to 5 kHz — 4 A to 6 A · CMC 0,14 % 5 kHz to 10 kHz — 4 A to 6 A · CMC 0,38 % 10 kHz to 20 kHz — 4 A to 6 A · CMC 1,5 % 10 Hz to 1 kHz — 6 A to 11 A · CMC 0,07 % 1 kHz to 5 kHz — 6 A to 11 A · CMC 0,13 % 5 kHz to 10 kHz — 6 A to 11 A · CMC 0,35 % 10 kHz to 20 kHz — 6 A to 11 A · CMC 1,3 % 40 Hz to 400 Hz — 11 A to 20 A · CMC 0,35 % CMC for the field of measured quantity: Electrical quantities |
Current generation by a standard calibrator | KP 1/E, KP 2/E, KP 3/E | |
| 67 | Alternating current / digital, analogue multimeters and ammeters, testing and inspection instruments and their parts, current probes | Calibration – electrical quantities 40 Hz to 300 Hz — 20 A to 30 A · CMC 0,1 % 40 Hz to 250 Hz — 30 A to 50 A · CMC 1,2 % 40 Hz to 160 Hz — 50 A to 100 A · CMC 2 % CMC for the field of measured quantity: Electrical quantities |
Comparison with current measured by a standard shunt with multimeter | KP 1/E, KP 2/E, KP 3/E | |
| 68 | Clamp ammeters and clamp current probes | Calibration – electrical quantities 40 Hz to 300 Hz — 1 A to 5 A · CMC 0,25 % CMC for the field of measured quantity: Electrical quantities |
Direct generation of current with a calibrator | KP 1/E | |
| 69 | Clamp ammeters and clamp current probes | Calibration – electrical quantities 40 Hz to 300 Hz — 5 A to 20 A · CMC 0,16 % 40 Hz to 300 Hz — 20 A to 30 A · CMC 0,33 % 40 Hz to 300 Hz — 30 A to 100 A · CMC 0,25 % 40 Hz to 300 Hz — 100 A to 250 A · CMC 0,35 % 40 Hz to 300 Hz — 250 A to 750 A · CMC 0,2 % CMC for the field of measured quantity: Electrical quantities |
Simulation of reference current using calibrator and current coil | KP 1/E | |
| 70 | Alternating current / current sources | Calibration – electrical quantities 10 Hz to 20 Hz — 1 µA to 120 µA · CMC 0,46 % + 35 nA 20 Hz to 45 Hz — 1 µA to 120 µA · CMC 0,17 % + 35 nA 45 Hz to 100 Hz — 1 µA to 120 µA · CMC 0,07 % + 35 nA 100 Hz to 1 000 Hz — 1 µA to 120 µA · CMC 0,071 % + 35 nA 1 kHz to 5 kHz — 1 µA to 120 µA · CMC 0,061 % + 35 nA 5 kHz to 20 kHz — 1 µA to 120 µA · CMC 0,061 % + 35 nA 10 Hz to 20 Hz — 120 µA to 1,2 mA · CMC 0,46 % + 230 nA 20 Hz to 45 Hz — 120 µA to 1,2 mA · CMC 0,17 % + 230 nA 45 Hz to 100 Hz — 120 µA to 1,2 mA · CMC 0,07 % + 230 nA 100 Hz to 1 000 Hz — 120 µA to 1,2 mA · CMC 0,037 % + 230 nA 1 kHz to 5 kHz — 120 µA to 1,2 mA · CMC 0,044 % + 230 nA 5 kHz to 20 kHz — 120 µA to 1,2 mA · CMC 0,044 % + 230 nA 10 Hz to 20 Hz — 1,2 mA to 12 mA · CMC 0,46 % + 2,3 µA 20 Hz to 45 Hz — 1,2 mA to 12 mA · CMC 0,17 % + 2,3 µA 45 Hz to 100 Hz — 1,2 mA to 12 mA · CMC 0,07 % + 2,3 µA 100 Hz to 1 000 Hz — 1,2 mA to 12 mA · CMC 0,037 % + 2,3 µA 1 kHz to 5 kHz — 1,2 mA to 12 mA · CMC 0,041 % + 2,3 µA 5 kHz to 20 kHz — 1,2 mA to 12 mA · CMC 0,041 % + 2,3 µA 10 Hz to 20 Hz — 12 mA to 120 mA · CMC 0,46 % + 23 µA 20 Hz to 45 Hz — 12 mA to 120 mA · CMC 0,17 % + 23 µA 45 Hz to 100 Hz — 12 mA to 120 mA · CMC 0,07 % + 23 µA 100 Hz to 1 000 Hz — 12 mA to 120 mA · CMC 0,037 % + 23 µA 1 kHz to 5 kHz — 12 mA to 120 mA · CMC 0,04 % + 23 µA 5 kHz to 20 kHz — 12 mA to 120 mA · CMC 0,04 % + 23 µA 10 Hz to 20 Hz — 120 mA to 1 A · CMC 0,46 % + 230 µA 20 Hz to 45 Hz — 120 mA to 1 A · CMC 0,17 % + 230 µA 45 Hz to 100 Hz — 120 mA to 1 A · CMC 0,07 % + 230 µA 100 Hz to 1 000 Hz — 120 mA to 1 A · CMC 0,12 % + 230 µA 1 kHz to 5 kHz — 120 mA to 1 A · CMC 0,13 % + 230 µA 5 kHz to 20 kHz — 120 mA to 1 A · CMC 0,13 % + 230 µA CMC for the field of measured quantity: Electrical quantities |
Direct measurement by a standard multimeter | KP 8/E | |
| 71 | Alternating current / current sources | Calibration – electrical quantities 40 Hz to 1 kHz — 1,2 A to 12 A · CMC 0,14 % + 330 µA 1 kHz to 20 kHz — 1,2 A to 12 A · CMC 0,15 % + 41 µA 40 Hz to 1 kHz — 12 A to 20 A · CMC 0,14 % + 420 µA 1 kHz to 20 kHz — 12 A to 20 A · CMC 0,15 % + 490 µA 40 Hz to 250 Hz — 20 A to 50 A · CMC 1 % + 44 µA 40 Hz to 250 Hz — 50 A to 100 A · CMC 2 % + 48 µA CMC for the field of measured quantity: Electrical quantities |
Indirect measurement by a multimeter with a standard shunt | KP 8/E | |
| 72 | Alternating current / current sources | Calibration – electrical quantities 40 Hz to 300 Hz — 1 A to 40 A · CMC 1,2 % + 0,5 A 40 Hz to 300 Hz — 40 A to 400 A · CMC 0,6 % + 5 A 40 Hz to 150 Hz — 400 A to 1 500 A · CMC 0,6 % + 23 A CMC for the field of measured quantity: Electrical quantities |
Measurement by a standard clamp ammeter | KP 8/E | |
| 73 | DC resistance / ohmmeters, testing and inspection instruments and their parts | Calibration – electrical quantities 0 Ω · CMC 20 µΩ 0,1 mΩ · CMC 6 nΩ 1 mΩ · CMC 50 nΩ 0,01 Ω · CMC 0,3 µΩ 0,1 Ω · CMC 2 µΩ 1 Ω · CMC 12 µΩ 10 Ω · CMC 200 µΩ 100 Ω · CMC 1,2 mΩ 1 kΩ · CMC 14 mΩ 10 kΩ · CMC 0,14 Ω 100 kΩ · CMC 1,7 Ω 1 MΩ · CMC 30 Ω 10 MΩ · CMC 60 Ω 100 MΩ · CMC 9 kΩ 1 GΩ · CMC 0,4 MΩ 10 GΩ · CMC 0,4 GΩ 100 GΩ · CMC 5 GΩ 1 TΩ · CMC 100 GΩ CMC for the field of measured quantity: Electrical quantities |
Direct measurement of fixed resistance standards | KP 1/E, KP 2/E | |
| 74 | DC resistance / standard resistors, shunts | Calibration – electrical quantities 0,1 mΩ · CMC 10 nΩ 1 mΩ · CMC 70 nΩ 0,01 Ω · CMC 0,4 µΩ 0,1 Ω · CMC 3 µΩ 1 Ω · CMC 20 µΩ 10 Ω · CMC 250 µΩ 100 Ω · CMC 1,5 mΩ 1 kΩ · CMC 40 mΩ 10 kΩ · CMC 0,4 Ω 100 kΩ · CMC 7 Ω CMC for the field of measured quantity: Electrical quantities |
Comparison by ratio method – measurement of voltage drop at constant measuring current | KP 6/E | |
| 75 | DC resistance / standard resistors, shunts | Calibration – electrical quantities 1 MΩ · CMC 40 Ω 10 MΩ · CMC 700 Ω 100 MΩ · CMC 12 kΩ 1 GΩ · CMC 0,6 MΩ CMC for the field of measured quantity: Electrical quantities |
Comparison by ratio method – measurement of standard and calibrated resistance | KP 6/E | |
| 76 | DC resistance / resistors, calibrators, simulators, testing and inspection instruments and their parts | Calibration – electrical quantities from 0 Ω to 12 Ω · CMC 0,002 % + 78 µΩ from 12 Ω to 120 Ω · CMC 0,0016 % + 570 µΩ from 0,12 kΩ to 1,2 kΩ · CMC 0,0014 % + 540 µΩ from 1,2 kΩ to 12 kΩ · CMC 0,0014 % + 5,4 mΩ from 12 kΩ to 120 kΩ · CMC 0,0016 % + 45 mΩ from 0,12 MΩ to 1,2 MΩ · CMC 0,0026 % + 1,9 Ω from 1,2 MΩ to 12 MΩ · CMC 0,0068 % + 110 Ω from 12 MΩ to 120 MΩ · CMC 0,058 % + 1,3 kΩ od 0,12 GΩ do 1,2 GΩ · CMC 0,57 % + 0,35 MΩ CMC for the field of measured quantity: Electrical quantities |
Direct measurement by a standard multimeter | KP 8/E | |
| 77 | DC resistance / resistors, calibrators, simulators, testing and inspection instruments and their parts | Calibration – electrical quantities from 0,1 MΩ to 1 MΩ · CMC 0,004 % from 1 MΩ to 10 MΩ · CMC 0,01 % od 10 MΩ do 200 MΩ · CMC 0,02 % od 200 MΩ do 500 MΩ · CMC 0,03 % od 500 MΩ do 1000 MΩ · CMC 0,1 % od 1000 MΩ do 2000 MΩ · CMC 0,2 % od 2000 MΩ do 5000 MΩ · CMC 0,3 % od 5000 MΩ do 10000 MΩ · CMC 0,5 % od 10000 MΩ do 50000 MΩ · CMC 1 % od 50000 MΩ do 100000 MΩ · CMC 2 % CMC for the field of measured quantity: Electrical quantities |
Indirect comparison – method of resistance divider excited by constant measuring voltage | KP 6/E | |
| 78 | HF level (voltage) / HF voltmeters, measuring receivers, synthesizers, oscilloscopes, spectral and circuit analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) | Calibration – electrical quantities 0,1 MHz to 1,4 GHz — 20 dBµV (10 µV) to 47 dBµV (224 µV) · CMC 4 % 1,4 GHz to 2,08 GHz — 20 dBµV (10 µV) to 47 dBµV (224 µV) · CMC 3,5 % 10 MHz to 30 MHz — 47 dBµV (224 µV) to 67 dBµV (2,24 mV) · CMC 4 % 30 MHz to 2,08 GHz — 47 dBµV (224 µV) to 67 dBµV (2,24 mV) · CMC 3,5 % 10 MHz to 30 MHz — 67 dBµV (2,24 mV) to 87 dBµV (22,4 mV) · CMC 4 % 30 MHz to 2,08 GHz — 67 dBµV (2,24 mV) to 87 dBµV (22,4 mV) · CMC 3,5 % 100 kHz to 300 kHz — 87 dBµV (22,4 mV) to 117 dBµV (707 mV) · CMC 17 % 300 kHz to 600 kHz — 87 dBµV (22,4 mV) to 117 dBµV (707 mV) · CMC 14 % 600 kHz to 1 MHz — 87 dBµV (22,4 mV) to 117 dBµV (707 mV) · CMC 8,5 % 1 MHz to 1,3 GHz — 87 dBµV (22,4 mV) to 117 dBµV (707 mV) · CMC 2,6 % 1,3 GHz to 2,08 GHz — 87 dBµV (22,4 mV) to 117 dBµV (707 mV) · CMC 3,2 % 100 kHz to 300 kHz — 117 dBµV (707 mV) to 137 dBµV (7,07 V) · CMC 17 % 300 kHz to 600 kHz — 117 dBµV (707 mV) to 137 dBµV (7,07 V) · CMC 14 % 600 kHz to 1 MHz — 117 dBµV (707 mV) to 137 dBµV (7,07 V) · CMC 8,5 % 1 MHz to 1,3 GHz — 117 dBµV (707 mV) to 137 dBµV (7,07 V) · CMC 3,5 % 1,3 GHz to 2,08 GHz — 117 dBµV (707 mV) to 137 dBµV (7,07 V) · CMC 4 % 100 kHz to 2,08 GHz — 137 dBµV (7,07 V) to 147 dBµV (22,4 V) · CMC 2,9 % 100 kHz to 2,08 GHz — 147 dBµV (22,4 V) to 152 dBµV (39,0 V) · CMC 3,7 % CMC for the field of measured quantity: Electrical quantities |
Indirect comparison with standard HF wattmeter | KP 13/E, KP 18/E | |
| 79 | HF level (voltage) / HF generators, circuit analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) | Calibration – electrical quantities 0,1 MHz to 1,4 GHz — 20 dBµV (10 µV) to 47 dBµV (224 µV) · CMC 4 % 1,4 GHz to 2,08 GHz — 20 dBµV (10 µV) to 47 dBµV (224 µV) · CMC 3,5 % 10 MHz to 30 MHz — 47 dBµV (224 µV) to 67 dBµV (2,24 mV) · CMC 4 % 30 MHz to 2,08 GHz — 47 dBµV (224 µV) to 67 dBµV (2,24 mV) · CMC 3,5 % 10 MHz to 30 MHz — 67 dBµV (2,24 µV) to 87 dBµV (22,4 mV) · CMC 4 % 30 MHz to 2,08 GHz — 67 dBµV (2,24 µV) to 87 dBµV (22,4 mV) · CMC 3,5 % 100 kHz to 300 kHz — 87 dBµV (22,4 mV) to 117 dBµV (707 mV) · CMC 17 % 300 kHz to 600 kHz — 87 dBµV (22,4 mV) to 117 dBµV (707 mV) · CMC 14 % 600 kHz to 1 MHz — 87 dBµV (22,4 mV) to 117 dBµV (707 mV) · CMC 8,5 % 1 MHz to 1,3 GHz — 87 dBµV (22,4 mV) to 117 dBµV (707 mV) · CMC 2,6 % 1,3 GHz to 2,08 GHz — 87 dBµV (22,4 mV) to 117 dBµV (707 mV) · CMC 3,2 % 100 kHz to 300 kHz — 117 dBµV (707 mV) to 137 dBµV (7,07 V) · CMC 17 % 300 kHz to 600 kHz — 117 dBµV (707 mV) to 137 dBµV (7,07 V) · CMC 14 % 600 kHz to 1 MHz — 117 dBµV (707 mV) to 137 dBµV (7,07 V) · CMC 8,5 % 1 MHz to 1,3 GHz — 117 dBµV (707 mV) to 137 dBµV (7,07 V) · CMC 3,5 % 1,3 GHz to 2,08 GHz — 117 dBµV (707 mV) to 137 dBµV (7,07 V) · CMC 4 % 100 kHz to 2,08 GHz — 137 dBµV (7,07 V) to 147 dBµV (22,4 V) · CMC 2,9 % 100 kHz to 2,08 GHz — 147 dBµV (22,4 V) to 152 dBµV (39,0 V) · CMC 3,7 % CMC for the field of measured quantity: Electrical quantities |
Indirect measurement of HF voltage - by measuring HF power at impedance 50 Ω | KP 11/E, KP 12/E | |
| 80 | HF level (power) / oscillators, circuit analyzers, testers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) | Calibration – electrical quantities 10 MHz to 20 MHz — 1 pW (−90 dBm) to 2,5 pW (−86 dBm) · CMC 8,7 % 20 MHz to 50 MHz — 1 pW (−90 dBm) to 2,5 pW (−86 dBm) · CMC 6,5 % 50 MHz to 7 GHz — 1 pW (−90 dBm) to 2,5 pW (−86 dBm) · CMC 6,2 % 10 MHz to 20 MHz — 2,5 pW (−86 dBm) to 800 pW (−61 dBm) · CMC 7,8 % 20 MHz to 50 MHz — 2,5 pW (−86 dBm) to 800 pW (−61 dBm) · CMC 5,2 % 50 MHz to 7 GHz — 2,5 pW (−86 dBm) to 800 pW (−61 dBm) · CMC 4,8 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement by a standard measuring receiver | KP 15/E | |
| 81 | HF level (power) / oscillators, circuit analyzers, testers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) | Calibration – electrical quantities 10 MHz to 20 MHz — 800 pW (−61 dBm) to 2 nW (−57 dBm) · CMC 8,5 % 20 MHz to 50 MHz — 800 pW (−61 dBm) to 2 nW (−57 dBm) · CMC 6,2 % 50 MHz to 7 GHz — 800 pW (−61 dBm) to 2 nW (−57 dBm) · CMC 5,5 % 7 GHz to 15 GHz — 800 pW (−61 dBm) to 2 nW (−57 dBm) · CMC 6,0 % 15 GHz to 18 GHz — 800 pW (−61 dBm) to 2 nW (−57 dBm) · CMC 6,6 % 10 MHz to 20 MHz — 2 nW (−57 dBm) to 10 µW (−20 dBm) · CMC 7,5 % 20 MHz to 50 MHz — 2 nW (−57 dBm) to 10 µW (−20 dBm) · CMC 4,8 % 50 MHz to 7 GHz — 2 nW (−57 dBm) to 10 µW (−20 dBm) · CMC 3,9 % 7 GHz to 15 GHz — 2 nW (−57 dBm) to 10 µW (−20 dBm) · CMC 4,5 % 15 GHz to 18 GHz — 2 nW (−57 dBm) to 10 µW (−20 dBm) · CMC 5,4 % 100 kHz to 300 kHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 7,3 % 300 kHz to 1 MHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 4,0 % 1 MHz to 50 MHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 3,6 % 50 MHz to 1 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 3,0 % 1 GHz to 2 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 3,7 % 2 GHz to 15 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 4,8 % 15 GHz to 18 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 5,1 % 18 GHz to 20 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 5,5 % 20 GHz to 26,5 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 4,6 % 26,5 GHz to 40,8 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 6,0 % 100 kHz to 300 kHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 7,7 % 300 kHz to 1 MHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 4,1 % 1 MHz to 50 MHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 3,4 % 50 MHz to 1 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 2,6 % 1 GHz to 2 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 3,7 % 2 GHz to 15 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 4,6 % 15 GHz to 18 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 4,9 % 18 GHz to 20 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 5,3 % 20 GHz to 26,5 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 4,3 % 26,5 GHz to 40,8 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 5,6 % 100 kHz to 300 kHz — 10 mW to 100 W (+20 dBm) · CMC 6,2 % 300 kHz to 1 MHz — 10 mW to 100 W (+20 dBm) · CMC 6,2 % 1 MHz to 10 MHz — 10 mW to 100 W (+20 dBm) · CMC 5,4 % 10 MHz to 50 MHz — 10 mW to 100 W (+20 dBm) · CMC 4,6 % 50 MHz to 1 GHz — 10 mW to 100 W (+20 dBm) · CMC 4,4 % 1 GHz to 15 GHz — 10 mW to 100 W (+20 dBm) · CMC 5,8 % 15 GHz to 18 GHz — 10 mW to 100 W (+20 dBm) · CMC 6,0 % 18 GHz to 20 GHz — 10 mW to 100 W (+20 dBm) · CMC 6,4 % 20 GHz to 26,5 GHz — 10 mW to 100 W (+20 dBm) · CMC 5,5 % 26,5 GHz to 40,8 GHz — 10 mW to 100 W (+20 dBm) · CMC 6,5 % 100 kHz to 300 kHz — 100 mW (+20 dBm) to 3 W (+33 dBm) · CMC 6,0 % 300 kHz to 1 MHz — 100 mW (+20 dBm) to 3 W (+33 dBm) · CMC 6,0 % 1 MHz to 10 MHz — 100 mW (+20 dBm) to 3 W (+33 dBm) · CMC 5,2 % 10 MHz to 500 MHz — 100 mW (+20 dBm) to 3 W (+33 dBm) · CMC 4,8 % 500 MHz to 1 GHz — 100 mW (+20 dBm) to 3 W (+33 dBm) · CMC 5,9 % 1 GHz to 4,2 GHz — 3 mW (+33 dBm) to 10 W (+40 dBm) · CMC 6,7 % 100 kHz to 300 kHz — 3 mW (+33 dBm) to 10 W (+40 dBm) · CMC 7,3 % 300 kHz to 1 MHz — 3 mW (+33 dBm) to 10 W (+40 dBm) · CMC 7,3 % 1 MHz to 10 MHz — 3 mW (+33 dBm) to 10 W (+40 dBm) · CMC 6,6 % 10 MHz to 500 MHz — 3 mW (+33 dBm) to 10 W (+40 dBm) · CMC 6,3 % 500 MHz to 1 GHz — 3 mW (+33 dBm) to 10 W (+40 dBm) · CMC 7,1 % 1 GHz to 4,2 GHz — 3 mW (+33 dBm) to 10 W (+40 dBm) · CMC 7,6 % 100 kHz to 300 kHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 7,4 % 300 kHz to 1 MHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 7,4 % 1 MHz to 10 MHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 6,7 % 10 MHz to 500 MHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 6,2 % 500 MHz to 1 GHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 6,9 % 1 GHz to 2 GHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 8,0 % 2 GHz to 4,2 GHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 9,5 % 1 GHz to 2 GHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 8,0 % CMC for the field of measured quantity: Electrical quantities |
Measurement by a standard HF wattmeter | KP 15/E | |
| 82 | HF level (power) / HF wattmeters, selective HF meters, spectral and circuit analyzers, testers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) | Calibration – electrical quantities 10 MHz to 20 MHz — 800 pW (−61 dBm) to 2 nW (−57 dBm) · CMC 8,5 % 20 MHz to 50 MHz — 800 pW (−61 dBm) to 2 nW (−57 dBm) · CMC 6,2 % 50 MHz to 7 GHz — 800 pW (−61 dBm) to 2 nW (−57 dBm) · CMC 5,5 % 7 GHz to 15 GHz — 800 pW (−61 dBm) to 2 nW (−57 dBm) · CMC 6,0 % 15 GHz to 18 GHz — 800 pW (−61 dBm) to 2 nW (−57 dBm) · CMC 6,6 % 10 MHz to 20 MHz — 2 nW (−57 dBm) to 10 µW (−20 dBm) · CMC 7,5 % 20 MHz to 50 MHz — 2 nW (−57 dBm) to 10 µW (−20 dBm) · CMC 4,8 % 50 MHz to 7 GHz — 2 nW (−57 dBm) to 10 µW (−20 dBm) · CMC 3,9 % 7 GHz to 15 GHz — 2 nW (−57 dBm) to 10 µW (−20 dBm) · CMC 4,5 % 15 GHz to 18 GHz — 2 nW (−57 dBm) to 10 µW (−20 dBm) · CMC 5,4 % 100 kHz to 300 kHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 7,3 % 300 kHz to 1 MHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 4,0 % 1 MHz to 50 MHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 3,6 % 50 MHz to 1 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 3,0 % 1 GHz to 2 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 3,7 % 2 GHz to 15 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 4,8 % 15 GHz to 18 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 5,1 % 18 GHz to 20 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 5,5 % 20 GHz to 26,5 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 4,6 % 26,5 GHz to 40,8 GHz — 10 µW (−20 dBm) to 100 µW (−10 dBm) · CMC 6,0 % 100 kHz to 300 kHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 7,7 % 300 kHz to 1 MHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 4,1 % 1 MHz to 50 MHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 3,4 % 50 MHz to 1 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 2,6 % 1 GHz to 2 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 3,7 % 2 GHz to 15 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 4,6 % 15 GHz to 18 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 4,9 % 18 GHz to 20 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 5,3 % 20 GHz to 26,5 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 4,3 % 26,5 GHz to 40,8 GHz — 100 µW (−10 dBm) to 10 mW (+10 dBm) · CMC 5,6 % 100 kHz to 300 kHz — 10 mW (+10 dBm) to 100 mW (+20 dBm) · CMC 6,2 % 300 kHz to 1 MHz — 10 mW (+10 dBm) to 100 mW (+20 dBm) · CMC 6,2 % 1 MHz to 10 MHz — 10 mW (+10 dBm) to 100 mW (+20 dBm) · CMC 5,4 % 10 MHz to 50 MHz — 10 mW (+10 dBm) to 100 mW (+20 dBm) · CMC 4,6 % 50 MHz to 1 GHz — 10 mW (+10 dBm) to 100 mW (+20 dBm) · CMC 4,4 % 1 GHz to 15 GHz — 10 mW (+10 dBm) to 100 mW (+20 dBm) · CMC 5,8 % 15 GHz to 18 GHz — 10 mW (+10 dBm) to 100 mW (+20 dBm) · CMC 6,0 % 18 GHz to 20 GHz — 10 mW (+10 dBm) to 100 mW (+20 dBm) · CMC 6,4 % 20 GHz to 26,5 GHz — 10 mW (+10 dBm) to 100 mW (+20 dBm) · CMC 5,5 % 26,5 GHz to 40,8 GHz — 10 mW (+10 dBm) to 100 mW (+20 dBm) · CMC 6,5 % 100 kHz to 300 kHz — 100 mW (+20 dBm) to 2 W (+33 dBm) · CMC 6,0 % 300 kHz to 1 MHz — 100 mW (+20 dBm) to 2 W (+33 dBm) · CMC 6,0 % 1 MHz to 10 MHz — 100 mW (+20 dBm) to 2 W (+33 dBm) · CMC 5,2 % 10 MHz to 500 MHz — 100 mW (+20 dBm) to 2 W (+33 dBm) · CMC 4,8 % 500 MHz to 1 GHz — 100 mW (+20 dBm) to 2 W (+33 dBm) · CMC 5,9 % 1 GHz to 4,2 GHz — 100 mW (+20 dBm) to 2 W (+33 dBm) · CMC 6,7 % 100 kHz to 300 kHz — 2 mW (+33 dBm) to 10 W (+40 dBm) · CMC 7,3 % 300 kHz to 1 MHz — 2 mW (+33 dBm) to 10 W (+40 dBm) · CMC 7,3 % 1 MHz to 10 MHz — 2 mW (+33 dBm) to 10 W (+40 dBm) · CMC 6,6 % 10 MHz to 500 MHz — 2 mW (+33 dBm) to 10 W (+40 dBm) · CMC 6,3 % 500 MHz to 1 GHz — 2 mW (+33 dBm) to 10 W (+40 dBm) · CMC 7,1 % 1 GHz to 4,2 GHz — 2 mW (+33 dBm) to 10 W (+40 dBm) · CMC 7,6 % 100 kHz to 300 kHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 7,4 % 300 kHz to 1 MHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 7,4 % 1 MHz to 10 MHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 6,7 % 10 MHz to 500 MHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 6,2 % 500 MHz to 1 GHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 6,9 % 1 GHz to 2 GHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 8,0 % 2 GHz to 4,2 GHz — 10 W (+40 dBm) to 25 W (+44 dBm) · CMC 9,5 % CMC for the field of measured quantity: Electrical quantities |
Comparison with a standard HF wattmeter | KP 14/E | |
| 83 | HF attenuation / attenuators, attenuation pads, selective HF meters, spectral and circuit analyzers, testers, radiocommunication and radio navigation testers, simulators, imitators (ATC, DME, TACAN, TCAS, VOR-ILS) | Calibration – electrical quantities 100 kHz to 400 kHz — 0 dBm to 10 dBm · CMC 0,048 dB 400 kHz to 1,3 GHz — 0 dBm to 10 dBm · CMC 0,048 dB 1,3 GHz to 3 GHz — 0 dBm to 10 dBm · CMC 0,048 dB 3 GHz to 6 GHz — 0 dBm to 10 dBm · CMC 0,059 dB 100 kHz to 400 kHz — 10 dBm to 20 dBm · CMC 0,064 dB 400 kHz to 1,3 GHz — 10 dBm to 20 dBm · CMC 0,067 dB 1,3 GHz to 3 GHz — 10 dBm to 20 dBm · CMC 0,062 dB 3 GHz to 6 GHz — 10 dBm to 20 dBm · CMC 0,076 dB 100 kHz to 400 kHz — 20 dBm to 30 dBm · CMC 0,11 dB 400 kHz to 1,3 GHz — 20 dBm to 30 dBm · CMC 0,11 dB 1,3 GHz to 3 GHz — 20 dBm to 30 dBm · CMC 0,11 dB 3 GHz to 6 GHz — 20 dBm to 30 dBm · CMC 0,11 dB 100 kHz to 400 kHz — 30 dBm to 40 dBm · CMC 0,15 dB 400 kHz to 3 GHz — 30 dBm to 40 dBm · CMC 0,15 dB 3 GHz to 6 GHz — 30 dBm to 40 dBm · CMC 0,17 dB 100 kHz to 3 GHz — 40 dBm to 60 dBm · CMC 0,18 dB 3 GHz to 6 GHz — 40 dBm to 60 dBm · CMC 0,19 dB 100 kHz to 6 GHz — 60 dBm to 65 dBm · CMC 0,25 dB CMC for the field of measured quantity: Electrical quantities |
Indirect measurement power method | KP 16/E | |
| 84 | Modulation swing of frequency modulation / generators, modulation meters, selective HF meters, spectral analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) | Calibration – electrical quantities Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 0,1 kHz to 0,2 kHz · CMC 3,4 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 0,1 kHz to 0,2 kHz · CMC 3,4 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 0,1 kHz to 0,2 kHz · CMC 3,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 0,1 kHz to 0,2 kHz · CMC 3,4 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 0,1 kHz to 0,2 kHz · CMC 6,8 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 0,1 kHz to 0,2 kHz · CMC 13 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 0,2 kHz to 0,3 kHz · CMC 2,4 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 0,2 kHz to 0,3 kHz · CMC 2,4 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 0,2 kHz to 0,3 kHz · CMC 3,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 0,2 kHz to 0,3 kHz · CMC 2,4 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 0,2 kHz to 0,3 kHz · CMC 4 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 0,2 kHz to 0,3 kHz · CMC 6,8 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 0,3 kHz to 0,4 kHz · CMC 2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 0,3 kHz to 0,4 kHz · CMC 2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 0,3 kHz to 0,4 kHz · CMC 4,9 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 0,3 kHz to 0,4 kHz · CMC 2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 0,3 kHz to 0,4 kHz · CMC 3,1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 0,3 kHz to 0,4 kHz · CMC 4,9 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 0,4 kHz to 0,5 kHz · CMC 1,9 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 0,4 kHz to 0,5 kHz · CMC 1,9 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 0,4 kHz to 0,5 kHz · CMC 4 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 0,4 kHz to 0,5 kHz · CMC 1,9 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 0,4 kHz to 0,5 kHz · CMC 2,6 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 0,4 kHz to 0,5 kHz · CMC 4 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 0,5 kHz to 0,6 kHz · CMC 1,8 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 0,5 kHz to 0,6 kHz · CMC 1,8 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 0,5 kHz to 0,6 kHz · CMC 3,4 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 0,5 kHz to 0,6 kHz · CMC 1,8 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 0,5 kHz to 0,6 kHz · CMC 2,4 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 0,5 kHz to 0,6 kHz · CMC 3,4 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 0,6 kHz to 0,7 kHz · CMC 1,7 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 0,6 kHz to 0,7 kHz · CMC 1,7 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 0,6 kHz to 0,7 kHz · CMC 3,1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 0,6 kHz to 0,7 kHz · CMC 1,7 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 0,6 kHz to 0,7 kHz · CMC 2,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 0,6 kHz to 0,7 kHz · CMC 3,1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 0,7 kHz to 0,8 kHz · CMC 1,7 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 0,7 kHz to 0,8 kHz · CMC 1,7 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 0,7 kHz to 0,8 kHz · CMC 2,8 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 0,7 kHz to 0,8 kHz · CMC 1,7 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 0,7 kHz to 0,8 kHz · CMC 2,1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 0,7 kHz to 0,8 kHz · CMC 2,8 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 0,8 kHz to 0,9 kHz · CMC 1,6 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 0,8 kHz to 0,9 kHz · CMC 1,6 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 0,8 kHz to 0,9 kHz · CMC 2,6 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 0,8 kHz to 0,9 kHz · CMC 1,6 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 0,8 kHz to 0,9 kHz · CMC 2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 0,8 kHz to 0,9 kHz · CMC 2,6 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 0,9 kHz to 1 kHz · CMC 1,6 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 0,9 kHz to 1 kHz · CMC 1,6 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 0,9 kHz to 1 kHz · CMC 2,5 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 0,9 kHz to 1 kHz · CMC 1,6 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 0,9 kHz to 1 kHz · CMC 1,9 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 0,9 kHz to 1 kHz · CMC 2,5 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 1 kHz to 2 kHz · CMC 1,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 1 kHz to 2 kHz · CMC 1,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 1 kHz to 2 kHz · CMC 2,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 1 kHz to 2 kHz · CMC 1,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 1 kHz to 2 kHz · CMC 1,6 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 1 kHz to 2 kHz · CMC 2,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 2 kHz to 3 kHz · CMC 1,1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 2 kHz to 3 kHz · CMC 1,1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 2 kHz to 3 kHz · CMC 1,6 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 2 kHz to 3 kHz · CMC 1,1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 2 kHz to 3 kHz · CMC 1,3 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 2 kHz to 3 kHz · CMC 1,6 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 3 kHz to 4 kHz · CMC 1,1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 3 kHz to 4 kHz · CMC 1,1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 3 kHz to 4 kHz · CMC 1,4 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 3 kHz to 4 kHz · CMC 1,1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 3 kHz to 4 kHz · CMC 1,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 3 kHz to 4 kHz · CMC 1,4 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 4 kHz to 5 kHz · CMC 1,1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 4 kHz to 5 kHz · CMC 1,1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 4 kHz to 5 kHz · CMC 1,3 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 4 kHz to 5 kHz · CMC 1,1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 4 kHz to 5 kHz · CMC 1,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 4 kHz to 5 kHz · CMC 1,3 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 5 kHz to 8 kHz · CMC 1,1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 5 kHz to 8 kHz · CMC 1,1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 5 kHz to 8 kHz · CMC 1,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 5 kHz to 8 kHz · CMC 1,1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 5 kHz to 8 kHz · CMC 1,1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 5 kHz to 8 kHz · CMC 1,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 8 kHz to 10 kHz · CMC 1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 8 kHz to 10 kHz · CMC 1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 8 kHz to 10 kHz · CMC 1,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 8 kHz to 10 kHz · CMC 1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 8 kHz to 10 kHz · CMC 1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 8 kHz to 10 kHz · CMC 1,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 10 kHz to 100 kHz · CMC 1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 10 kHz to 100 kHz · CMC 1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 10 kHz to 100 kHz · CMC 1,1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 10 kHz to 100 kHz · CMC 1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 10 kHz to 100 kHz · CMC 1 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 10 kHz to 100 kHz · CMC 1 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 100 kHz to 300 kHz · CMC 2,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 100 kHz to 300 kHz · CMC 2,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 100 kHz to 300 kHz · CMC 2,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 100 kHz to 300 kHz · CMC 2,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 100 kHz to 300 kHz · CMC 2,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 100 kHz to 300 kHz · CMC 2,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,03 to 120) MHz — 300 kHz to 500 kHz · CMC 3,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (120 to 550) MHz — 300 kHz to 500 kHz · CMC 3,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: (0,55 to 1) GHz — 300 kHz to 500 kHz · CMC 3,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (4,25 to 120) MHz — 300 kHz to 500 kHz · CMC 3,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (120 to 550) MHz — 300 kHz to 500 kHz · CMC 3,2 % Modulation frequency: (60 to 100) kHz; carrier frequency: (0,55 to 1) GHz — 300 kHz to 500 kHz · CMC 3,2 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement with a modulation analyzer | KP 21/E | |
| 85 | Amplitude modulation depth / generators, modulation meters, selective HF meters, spectral analyzers, radio navigation and radiocommunication testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) | Calibration – electrical quantities Modulation frequency: 30 Hz — 20 % · CMC 0,015 % Modulation frequency: 30 Hz — 30 % · CMC 0,015 % Modulation frequency: 30 Hz — 40 % · CMC 0,015 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: 55 kHz to 10 MHz — 2 % to 5 % · CMC 3,6 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: 10 MHz to 1 GHz — 2 % to 5 % · CMC 3,2 % Modulation frequency: 60 kHz to 100 kHz; carrier frequency: 55 kHz to 10 MHz — 2 % to 5 % · CMC 4,3 % Modulation frequency: 60 kHz to 100 kHz; carrier frequency: 10 MHz to 1 GHz — 2 % to 5 % · CMC 4,3 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: 55 kHz to 10 MHz — 5 % to 10 % · CMC 2,7 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: 10 MHz to 1 GHz — 5 % to 10 % · CMC 2,2 % Modulation frequency: 60 kHz to 100 kHz; carrier frequency: 55 kHz to 10 MHz — 5 % to 10 % · CMC 3,6 % Modulation frequency: 60 kHz to 100 kHz; carrier frequency: 10 MHz to 1 GHz — 5 % to 10 % · CMC 3,6 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: 55 kHz to 10 MHz — 10 % to 80 % · CMC 2,2 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: 10 MHz to 1 GHz — 10 % to 80 % · CMC 1,8 % Modulation frequency: 60 kHz to 100 kHz; carrier frequency: 55 kHz to 10 MHz — 10 % to 80 % · CMC 3,4 % Modulation frequency: 60 kHz to 100 kHz; carrier frequency: 10 MHz to 1 GHz — 10 % to 80 % · CMC 3,4 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: 55 kHz to 10 MHz — 80 % to 100 % · CMC 2,4 % Modulation frequency: 30 Hz to 60 kHz; carrier frequency: 10 MHz to 1 GHz — 80 % to 100 % · CMC 1,7 % Modulation frequency: 60 kHz to 100 kHz; carrier frequency: 55 kHz to 10 MHz — 80 % to 100 % · CMC 3,3 % Modulation frequency: 60 kHz to 100 kHz; carrier frequency: 10 MHz to 1 GHz — 80 % to 100 % · CMC 3,3 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement with a standard modulation analyzer | KP 21/E | |
| 86 | Capacity / capacitors and capacity standards, capacity calibrators, capacity decades, testing and inspection instruments and their parts | Calibration – electrical quantities 1 kHz to 2 kHz — 2 pF to 5 pF · CMC 9 % 2 kHz to 5 kHz — 2 pF to 5 pF · CMC 3,5 % 5 kHz to 10 kHz — 2 pF to 5 pF · CMC 1,5 % 10 kHz to 20 kHz — 2 pF to 5 pF · CMC 0,8 % 20 kHz to 1 MHz — 2 pF to 5 pF · CMC 0,5 % 1 kHz to 2 kHz — 5 pF to 10 pF · CMC 3,5 % 2 kHz to 5 kHz — 5 pF to 10 pF · CMC 1,6 % 5 kHz to 10 kHz — 5 pF to 10 pF · CMC 0,6 % 10 kHz to 20 kHz — 5 pF to 10 pF · CMC 0,4 % 20 kHz to 100 kHz — 5 pF to 10 pF · CMC 0,3 % 100 kHz to 1 MHz — 5 pF to 10 pF · CMC 0,4 % 1 kHz to 2 kHz — 10 pF to 100 pF · CMC 1,9 % 2 kHz to 5 kHz — 10 pF to 100 pF · CMC 0,4 % 5 kHz to 10 kHz — 10 pF to 100 pF · CMC 0,35 % 10 kHz to 20 kHz — 10 pF to 100 pF · CMC 0,2 % 20 kHz to 100 kHz — 10 pF to 100 pF · CMC 0,15 % 100 kHz to 400 kHz — 10 pF to 100 pF · CMC 0,2 % 400 kHz to 1 MHz — 10 pF to 100 pF · CMC 0,4 % 0,5 kHz to 2 kHz — 100 pF to 1 000 pF · CMC 0,55 % 2 kHz to 5 kHz — 100 pF to 1 000 pF · CMC 0,15 % 5 kHz to 100 kHz — 100 pF to 1 000 pF · CMC 0,08 % 100 kHz to 400 kHz — 100 pF to 1 000 pF · CMC 0,17 % 400 kHz to 1 MHz — 100 pF to 1 000 pF · CMC 0,37 % 100 Hz to 500 Hz — 1 nF to 10 nF · CMC 0,8 % 500 Hz to 2 kHz — 1 nF to 10 nF · CMC 0,12 % 2 kHz to 100 kHz — 1 nF to 10 nF · CMC 0,08 % 100 kHz to 400 kHz — 1 nF to 10 nF · CMC 0,17 % 400 kHz to 1 MHz — 1 nF to 10 nF · CMC 0,37 % 100 Hz to 500 Hz — 10 nF to 100 nF · CMC 0,2 % 500 Hz to 100 kHz — 10 nF to 100 nF · CMC 0,1 % 100 kHz to 400 kHz — 10 nF to 100 nF · CMC 0,25 % 400 kHz to 1 MHz — 10 nF to 100 nF · CMC 0,73 % 100 Hz to 500 Hz — 100 nF to 1 000 nF · CMC 0,15 % 500 Hz to 20 kHz — 100 nF to 1 000 nF · CMC 0,08 % 20 kHz to 100 kHz — 100 nF to 1 000 nF · CMC 0,15 % 100 kHz to 400 kHz — 100 nF to 1 000 nF · CMC 0,85 % 100 Hz to 500 Hz — 1 µF to 10 µF · CMC 0,13 % 500 Hz to 10 kHz — 1 µF to 10 µF · CMC 0,08 % 10 kHz to 20 kHz — 1 µF to 10 µF · CMC 0,13 % 20 kHz to 50 kHz — 1 µF to 10 µF · CMC 0,3 % 50 kHz to 100 kHz — 1 µF to 10 µF · CMC 0,7 % 50 Hz to 100 Hz — 10 µF to 100 µF · CMC 0,25 % 100 Hz to 500 Hz — 10 µF to 100 µF · CMC 0,15 % 500 Hz to 1 kHz — 10 µF to 100 µF · CMC 0,1 % 1 kHz to 5 kHz — 10 µF to 100 µF · CMC 0,2 % 5 kHz to 10 kHz — 10 µF to 100 µF · CMC 0,4 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement using a standard RLC bridge | KP 32/E | |
| 87 | Capacity / capacity standards, capacitors, testing and inspection instruments and their parts | Calibration – electrical quantities 1 kHz — 1 pF · CMC 1,5 % 1 kHz to 100 kHz — 1 pF · CMC 1,6 % 100 kHz to 1 MHz — 1 pF · CMC 2 % 1 kHz — 10 pF · CMC 0,05 % 1 kHz to 100 kHz — 10 pF · CMC 0,2 % 100 kHz to 1 MHz — 10 pF · CMC 0,2 % 1 kHz — 100 pF · CMC 0,01 % 1 kHz to 100 kHz — 100 pF · CMC 0,1 % 100 kHz to 1 MHz — 100 pF · CMC 0,5 % 100 Hz to 1 kHz — 1 000 pF · CMC 0,05 % 1 kHz — 1 000 pF · CMC 0,02 % 1 kHz to 100 kHz — 1 000 pF · CMC 0,05 % 100 kHz to 1 MHz — 1 000 pF · CMC 0,1 % 100 Hz to 1 kHz — 10 nF · CMC 0,05 % 1 kHz — 10 nF · CMC 0,01 % 1 kHz to 100 kHz — 10 nF · CMC 0,05 % 100 kHz to 1 MHz — 10 nF · CMC 0,1 % 100 Hz to 1 kHz — 100 nF · CMC 0,05 % 1 kHz — 100 nF · CMC 0,01 % 1 kHz to 100 kHz — 100 nF · CMC 0,1 % 100 Hz to 1 kHz — 1 µF · CMC 0,15 % 1 kHz — 1 µF · CMC 0,1 % 1 kHz to 100 kHz — 1 µF · CMC 0,6 % 100 Hz to 1 kHz — 10 µF · CMC 0,15 % 1 kHz — 10 µF · CMC 0,1 % 1 kHz to 50 kHz — 10 µF · CMC 0,5 % 100 Hz to 1 kHz — 100 µF · CMC 0,1 % 1 kHz — 100 µF · CMC 0,1 % 1 kHz to 10 kHz — 100 µF · CMC 0,6 % CMC for the field of measured quantity: Electrical quantities |
Comparison with a standard capacitor | KP 32/E | |
| 88 | Capacity / RLCG bridges and meters, testing and inspection instruments and their parts | Calibration – electrical quantities 1 kHz — 1 pF · CMC 1,5 % 1 kHz to 100 kHz — 1 pF · CMC 1,6 % 100 kHz to 1 MHz — 1 pF · CMC 2 % 1 kHz — 10 pF · CMC 0,05 % 1 kHz to 100 kHz — 10 pF · CMC 0,2 % 100 kHz to 1 MHz — 10 pF · CMC 0,2 % 1 kHz — 100 pF · CMC 0,01 % 1 kHz to 100 kHz — 100 pF · CMC 0,1 % 100 kHz to 1 MHz — 100 pF · CMC 0,5 % 100 Hz to 1 kHz — 1 000 pF · CMC 0,05 % 1 kHz — 1 000 pF · CMC 0,02 % 1 kHz to 100 kHz — 1 000 pF · CMC 0,05 % 100 kHz to 1 MHz — 1 000 pF · CMC 0,1 % 100 Hz to 1 kHz — 10 nF · CMC 0,05 % 1 kHz — 10 nF · CMC 0,01 % 1 kHz to 100 kHz — 10 nF · CMC 0,05 % 100 kHz to 1 MHz — 10 nF · CMC 0,1 % 100 Hz to 1 kHz — 100 nF · CMC 0,05 % 1 kHz — 100 nF · CMC 0,01 % 1 kHz to 100 kHz — 100 nF · CMC 0,1 % 100 Hz to 1 kHz — 1 µF · CMC 0,15 % 1 kHz — 1 µF · CMC 0,1 % 1 kHz to 100 kHz — 1 µF · CMC 0,6 % 100 Hz to 1 kHz — 10 µF · CMC 0,15 % 1 kHz — 10 µF · CMC 0,1 % 1 kHz to 50 kHz — 10 µF · CMC 0,5 % 100 Hz to 1 kHz — 100 µF · CMC 0,1 % 1 kHz — 100 µF · CMC 0,1 % 1 kHz to 10 kHz — 100 µF · CMC 0,6 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement on capacity standards | KP 31/E | |
| 89 | Inductance / inductors and inductance standards, inductance calibrators, inductance decades, testing and inspection instruments and their parts | Calibration – electrical quantities 1 kHz to 2 kHz — 1 µH to 2 µH · CMC 10 % 2 kHz to 4 kHz — 1 µH to 2 µH · CMC 4,1 % 4 kHz to 8 kHz — 1 µH to 2 µH · CMC 2 % 8 kHz to 10 kHz — 1 µH to 2 µH · CMC 1 % 10 kHz to 50 kHz — 1 µH to 2 µH · CMC 0,83 % 50 kHz to 100 kHz — 1 µH to 2 µH · CMC 0,3 % 1 kHz to 2 kHz — 2 µH to 3 µH · CMC 5 % 2 kHz to 4 kHz — 2 µH to 3 µH · CMC 2,1 % 4 kHz to 8 kHz — 2 µH to 3 µH · CMC 1 % 8 kHz to 10 kHz — 2 µH to 3 µH · CMC 0,6 % 10 kHz to 50 kHz — 2 µH to 3 µH · CMC 0,5 % 50 kHz to 100 kHz — 2 µH to 3 µH · CMC 0,2 % 1 kHz to 2 kHz — 3 µH to 5 µH · CMC 3,2 % 2 kHz to 4 kHz — 3 µH to 5 µH · CMC 1,4 % 4 kHz to 8 kHz — 3 µH to 5 µH · CMC 0,69 % 8 kHz to 10 kHz — 3 µH to 5 µH · CMC 0,38 % 10 kHz to 50 kHz — 3 µH to 5 µH · CMC 0,33 % 50 kHz to 100 kHz — 3 µH to 5 µH · CMC 0,18 % 1 kHz to 2 kHz — 5 µH to 10 µH · CMC 2 % 2 kHz to 4 kHz — 5 µH to 10 µH · CMC 0,86 % 4 kHz to 8 kHz — 5 µH to 10 µH · CMC 0,44 % 8 kHz to 10 kHz — 5 µH to 10 µH · CMC 0,27 % 10 kHz to 50 kHz — 5 µH to 10 µH · CMC 0,19 % 50 kHz to 100 kHz — 5 µH to 10 µH · CMC 0,15 % 1 kHz to 2 kHz — 10 µH to 20 µH · CMC 1 % 2 kHz to 4 kHz — 10 µH to 20 µH · CMC 0,46 % 4 kHz to 10 kHz — 10 µH to 20 µH · CMC 0,25 % 10 kHz to 100 kHz — 10 µH to 20 µH · CMC 0,14 % 500 Hz to 1 kHz — 20 µH to 50 µH · CMC 1,3 % 1 kHz to 2 kHz — 20 µH to 50 µH · CMC 0,53 % 2 kHz to 4 kHz — 20 µH to 50 µH · CMC 0,26 % 4 kHz to 10 kHz — 20 µH to 50 µH · CMC 0,16 % 10 kHz to 100 kHz — 20 µH to 50 µH · CMC 0,14 % 500 Hz to 1 kHz — 50 µH to 200 µH · CMC 0,55 % 1 kHz to 4 kHz — 50 µH to 200 µH · CMC 0,25 % 4 kHz to 10 kHz — 50 µH to 200 µH · CMC 0,11 % 10 kHz to 100 kHz — 50 µH to 200 µH · CMC 0,14 % 100 Hz to 200 Hz — 200 µH to 1 mH · CMC 1,9 % 200 Hz to 500 Hz — 200 µH to 1 mH · CMC 0,6 % 500 Hz to 1 kHz — 200 µH to 1 mH · CMC 0,2 % 1 kHz to 4 kHz — 200 µH to 1 mH · CMC 0,11 % 4 kHz to 10 kHz — 200 µH to 1 mH · CMC 0,07 % 10 kHz to 100 kHz — 200 µH to 1 mH · CMC 0,13 % 100 Hz to 200 Hz — 1 mH to 1 H · CMC 0,48 % 200 Hz to 500 Hz — 1 mH to 1 H · CMC 0,2 % 500 Hz to 1 kHz — 1 mH to 1 H · CMC 0,13 % 1 kHz to 20 kHz — 1 mH to 1 H · CMC 0,09 % 100 Hz to 200 Hz — 1 H to 20 H · CMC 0,18 % 200 Hz to 1 kHz — 1 H to 20 H · CMC 0,14 % 100 Hz to 200 Hz — 20 H to 100 H · CMC 0,2 % 200 Hz to 1 kHz — 20 H to 100 H · CMC 0,16 % CMC for the field of measured quantity: Electrical quantities |
Measurement by standard RLCG bridge, two-terminal or three-terminal connection | KP 32/E | |
| 90 | Inductance / RLCG bridges, meters, testing and inspection instruments and their parts | Calibration – electrical quantities 1 kHz — 1 µH · CMC 4 % 1 kHz — 2 µH · CMC 1 % 10 kHz — 2 µH · CMC 0,5 % 100 kHz — 2 µH · CMC 0,3 % 1 kHz — 20 µH · CMC 0,2 % 1 kHz — 100 µH · CMC 0,25 % 1 kHz — 200 µH · CMC 1 % 1 kHz — 500 µH · CMC 0,1 % 1 kHz — 1 mH · CMC 0,1 % 1 kHz — 2 mH · CMC 0,2 % 1 kHz — 5 mH · CMC 0,2 % 1 kHz — 10 mH · CMC 025 % 1 kHz — 20 mH · CMC 0,5 % 1 kHz — 50 mH · CMC 0,1 % 1 kHz — 100 mH · CMC 0,6 % 1 kHz — 200 mH · CMC 0,1 % 1 kHz — 1 H · CMC 0,05 % 1 kHz — 10 H · CMC 0,1 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement on inductance standards, two- or three-terminal connection | KP 31/E | |
| 91 | AC resistance, impedance / RLCG bridges, meters, testing and inspection instruments and their parts | Calibration – electrical quantities 0 Hz — 0,1 Ω · CMC 0,017 % 0 Hz to 100 Hz — 0,1 Ω · CMC 0,21 % 100 Hz to 1 kHz — 0,1 Ω · CMC 0,32 % 1 kHz to 20 kHz — 0,1 Ω · CMC 1,8 % 20 kHz to 100 kHz — 0,1 Ω · CMC 5,7 % 100 kHz — 0,1 Ω · CMC 3,2 % 0 Hz — 1 Ω · CMC 0,005 % 0 Hz to 1 kHz — 1 Ω · CMC 0,06 % 1 kHz to 100 kHz — 1 Ω · CMC 0,26 % 100 kHz to 1 MHz — 1 Ω · CMC 0,92 % 1 MHz — 1 Ω · CMC 0,53 % 0 Hz — 10 Ω · CMC 0,005 % 0 Hz to 1 kHz — 10 Ω · CMC 0,03 % 1 kHz to 100 kHz — 10 Ω · CMC 0,04 % 100 kHz — 10 Ω · CMC 0,06 % 100 kHz to 1 MHz — 10 Ω · CMC 0,15 % 1 MHz — 10 Ω · CMC 0,1 % 0 Hz — 100 Ω · CMC 0,005 % 0 Hz to 100 kHz — 100 Ω · CMC 0,03 % 100 kHz — 100 Ω · CMC 0,06 % 100 kHz to 1 MHz — 100 Ω · CMC 0,58 % 1 MHz to 10 MHz — 100 Ω · CMC 0,5 % 10 MHz to 50 MHz — 100 Ω · CMC 12 % 0 Hz — 1 kΩ · CMC 0,005 % 0 Hz to 1 kHz — 1 kΩ · CMC 0,02 % 1 kHz to 100 kHz — 1 kΩ · CMC 0,05 % 100 kHz to 2 MHz — 1 kΩ · CMC 0,11 % 0 Hz — 10 kΩ · CMC 0,17 % 0 Hz to 1 kHz — 10 kΩ · CMC 0,02 % 1 kHz to 100 kHz — 10 kΩ · CMC 0,06 % 100 kHz to 2 MHz — 10 kΩ · CMC 1,4 % 2 MHz — 10 kΩ · CMC 0,52 % 0 Hz — 100 kΩ · CMC 0,017 % 0 Hz to 1 kHz — 100 kΩ · CMC 0,02 % 1 kHz to 100 kHz — 100 kΩ · CMC 0,25 % 100 kHz — 100 kΩ · CMC 0,09 % 100 kHz to 1 MHz — 100 kΩ · CMC 16 % 1 MHz — 100 kΩ · CMC 0,31 % 0 Hz — 1 MΩ · CMC 0,01 % 0 Hz to 1 kHz — 1 MΩ · CMC 0,06 % 1 kHz to 20 kHz — 1 MΩ · CMC 0,26 % 20 kHz to 200 kHz — 1 MΩ · CMC 2,5 % 200 kHz — 1 MΩ · CMC 1,9 % 0 Hz — 10 MΩ · CMC 0,06 % 0 Hz to 100 Hz — 10 MΩ · CMC 0,44 % 100 Hz to 400 Hz — 10 MΩ · CMC 0,5 % 400 Hz to 1 kHz — 10 MΩ · CMC 0,9 % 100 Hz — 100 MΩ · CMC 0,15 % 100 Hz to 1 kHz — 100 MΩ · CMC 0,32 % 1 kHz to 20 kHz — 100 MΩ · CMC 0,9 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement on impedance standards, four-terminal connection | KP 31/E | |
| 92 | AC resistance, impedance / impedance standards and calibrators, testing and inspection instruments and their parts | Calibration – electrical quantities 20 Hz to 50 Hz — 0,1 mΩ to 1 Ω · CMC 0,015 Ω 50 Hz to 100 Hz — 0,1 mΩ to 1 Ω · CMC 0,006 Ω 100 Hz to 500 Hz — 0,1 mΩ to 1 Ω · CMC 0,003 Ω 500 Hz to 1 kHz — 0,1 mΩ to 1 Ω · CMC 0,002 Ω 1 kHz to 10 kHz — 0,1 mΩ to 1 Ω · CMC 0,002 Ω 10 kHz to 100 kHz — 0,1 mΩ to 1 Ω · CMC 0,002 Ω 100 kHz to 500 kHz — 0,1 mΩ to 1 Ω · CMC 0,006 Ω 500 kHz to 1 MHz — 0,1 mΩ to 1 Ω · CMC 0,01 Ω 20 Hz to 50 Hz — 1 Ω to 10 Ω · CMC 1,4 % 50 Hz to 100 Hz — 1 Ω to 10 Ω · CMC 0,61 % 100 Hz to 500 Hz — 1 Ω to 10 Ω · CMC 0,35 % 500 Hz to 1 kHz — 1 Ω to 10 Ω · CMC 0,15 % 1 kHz to 10 kHz — 1 Ω to 10 Ω · CMC 0,12 % 10 kHz to 100 kHz — 1 Ω to 10 Ω · CMC 0,19 % 100 kHz to 500 kHz — 1 Ω to 10 Ω · CMC 0,6 % 500 kHz to 1 MHz — 1 Ω to 10 Ω · CMC 1 % 20 Hz to 50 Hz — 10 Ω to 100 Ω · CMC 0,58 % 50 Hz to 100 Hz — 10 Ω to 100 Ω · CMC 0,27 % 100 Hz to 500 Hz — 10 Ω to 100 Ω · CMC 0,16 % 500 Hz to 10 kHz — 10 Ω to 100 Ω · CMC 0,08 % 10 kHz to 100 kHz — 10 Ω to 100 Ω · CMC 0,11 % 100 kHz to 500 kHz — 10 Ω to 100 Ω · CMC 0,25 % 500 kHz to 1 MHz — 10 Ω to 100 Ω · CMC 0,41 % 2 MHz — 100 Ω · CMC 0,3 % 20 Hz to 50 Hz — 100 Ω to 1 kΩ · CMC 0,5 % 50 Hz to 100 Hz — 100 Ω to 1 kΩ · CMC 0,25 % 100 Hz to 500 Hz — 100 Ω to 1 kΩ · CMC 0,15 % 500 Hz to 10 kHz — 100 Ω to 1 kΩ · CMC 0,07 % 10 kHz to 100 kHz — 100 Ω to 1 kΩ · CMC 0,08 % 100 kHz to 500 kHz — 100 Ω to 1 kΩ · CMC 0,21 % 500 kHz to 1 MHz — 100 Ω to 1 kΩ · CMC 0,4 % 2 MHz — 1 kΩ · CMC 0,2 % 20 Hz to 50 Hz — 1 kΩ to 10 kΩ · CMC 0,51 % 50 Hz to 100 Hz — 1 kΩ to 10 kΩ · CMC 0,25 % 100 Hz to 500 Hz — 1 kΩ to 10 kΩ · CMC 0,18 % 500 Hz to 10 kHz — 1 kΩ to 10 kΩ · CMC 0,13 % 10 kHz to 100 kHz — 1 kΩ to 10 kΩ · CMC 0,18 % 100 kHz to 500 kHz — 1 kΩ to 10 kΩ · CMC 0,22 % 500 kHz to 1 MHz — 1 kΩ to 10 kΩ · CMC 0,4 % 2 MHz — 10 kΩ · CMC 0,3 % 20 Hz to 50 Hz — 10 kΩ to 100 kΩ · CMC 0,51 % 50 Hz to 100 Hz — 10 kΩ to 100 kΩ · CMC 0,25 % 100 Hz to 500 Hz — 10 kΩ to 100 kΩ · CMC 0,18 % 500 Hz to 10 kHz — 10 kΩ to 100 kΩ · CMC 0,13 % 10 kHz to 100 kHz — 10 kΩ to 100 kΩ · CMC 0,18 % 100 kHz to 500 kHz — 10 kΩ to 100 kΩ · CMC 0,22 % 500 kHz to 1 MHz — 10 kΩ to 100 kΩ · CMC 0,4 % 20 Hz to 50 Hz — 0,1 MΩ to 1 MΩ · CMC 2,3 % 50 Hz to 100 Hz — 0,1 MΩ to 1 MΩ · CMC 1 % 100 Hz to 500 Hz — 0,1 MΩ to 1 MΩ · CMC 0,6 % 500 Hz to 1 kHz — 0,1 MΩ to 1 MΩ · CMC 0,24 % 1 kHz to 10 kHz — 0,1 MΩ to 1 MΩ · CMC 0,18 % 10 kHz to 100 kHz — 0,1 MΩ to 1 MΩ · CMC 0,31 % 100 kHz to 500 kHz — 0,1 MΩ to 1 MΩ · CMC 0,9 % 500 kHz to 1 MHz — 0,1 MΩ to 1 MΩ · CMC 1,6 % 20 Hz to 50 Hz — 1 MΩ to 2 MΩ · CMC 4,2 % 50 Hz to 100 Hz — 1 MΩ to 2 MΩ · CMC 1,8 % 100 Hz to 500 Hz — 1 MΩ to 2 MΩ · CMC 1 % 500 Hz to 1 kHz — 1 MΩ to 2 MΩ · CMC 0,4 % 1 kHz to 10 kHz — 1 MΩ to 2 MΩ · CMC 0,3 % 10 kHz to 100 kHz — 1 MΩ to 2 MΩ · CMC 0,48 % 100 kHz to 500 kHz — 1 MΩ to 2 MΩ · CMC 1,6 % 500 kHz to 1 MHz — 1 MΩ to 2 MΩ · CMC 2,9 % 20 Hz to 50 Hz — 2 MΩ to 5 MΩ · CMC 9,6 % 50 Hz to 100 Hz — 2 MΩ to 5 MΩ · CMC 4,1 % 100 Hz to 500 Hz — 2 MΩ to 5 MΩ · CMC 2,3 % 500 Hz to 1 kHz — 2 MΩ to 5 MΩ · CMC 0,83 % 1 kHz to 10 kHz — 2 MΩ to 5 MΩ · CMC 0,65 % 10 kHz to 100 kHz — 2 MΩ to 5 MΩ · CMC 1,1 % 100 kHz to 500 kHz — 2 MΩ to 5 MΩ · CMC 3,5 % 500 kHz to 1 MHz — 2 MΩ to 5 MΩ · CMC 6,6 % 100 Hz to 500 Hz — 5 MΩ to 10 MΩ · CMC 4,5 % 500 Hz to 1 kHz — 5 MΩ to 10 MΩ · CMC 1,6 % 1 kHz to 10 kHz — 5 MΩ to 10 MΩ · CMC 1,3 % 10 kHz to 100 kHz — 5 MΩ to 10 MΩ · CMC 2,1 % 100 Hz to 500 Hz — 10 MΩ to 20 MΩ · CMC 8,7 % 500 Hz to 1 kHz — 10 MΩ to 20 MΩ · CMC 3,1 % 1 kHz to 10 kHz — 10 MΩ to 20 MΩ · CMC 2,4 % 10 kHz to 100 kHz — 10 MΩ to 20 MΩ · CMC 4 % 500 Hz to 1 kHz — 20 MΩ to 50 MΩ · CMC 7,7 % 1 kHz to 10 kHz — 20 MΩ to 50 MΩ · CMC 5,9 % 10 kHz to 50 kHz — 20 MΩ to 50 MΩ · CMC 7,1 % 1 kHz to 2 kHz — 50 MΩ to 100 MΩ · CMC 12 % 2 kHz to 10 kHz — 50 MΩ to 100 MΩ · CMC 10 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement using a standard RLC bridge | KP 32/E | |
| 93 | Insulation resistance / inspection instruments, electrical strength testers | Calibration – electrical quantities from 100 kΩ to 1 MΩ · CMC 0,14 % from 1 MΩ to 10 MΩ · CMC 0,21 % od 10 MΩ do 1 GΩ · CMC 0,36 % od 1 GΩ do 100 GΩ · CMC 0,83 % from 100 GΩ to 1 TΩ · CMC 1,5 % from 1 TΩ to 10 TΩ · CMC 2,4 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement on a calibrator of inspection instruments | KP 5/E | |
| 94 | Transition resistance / inspection instruments, multifunctional electrical safety testers | Calibration – electrical quantities od 5 mΩ do 15 mΩ · CMC 0,08 mΩ from 0,1 Ω to 0,2 Ω · CMC 0,0074 Ω from 0,2 Ω to 0,5 Ω · CMC 0,008 Ω from 0,5 Ω to 1 Ω · CMC 0,0096 Ω od 1 Ω do 3 Ω · CMC 0,013 Ω from 3 Ω to 5 Ω · CMC 0,44 % from 5 Ω to 10 Ω · CMC 0,27 % od 10 Ω do 50 Ω · CMC 0,21 % od 50 Ω do 200 Ω · CMC 0,16 % od 200 Ω do 10 kΩ · CMC 0,14 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement on a calibrator of inspection instruments | KP 5/E | |
| 95 | Earth resistance, protective conductor resistance / inspection instruments, resistance meters | Calibration – electrical quantities od 0,5 mΩ do 5 mΩ · CMC 0,14 mΩ od 5 mΩ do 30 mΩ · CMC 0,77 mΩ od 30 mΩ do 60 mΩ · CMC 1,9 mΩ od 60 mΩ do 100 mΩ · CMC 2 mΩ od 100 mΩ do 400 mΩ · CMC 3,6 mΩ od 400 mΩ do 600 mΩ · CMC 3,5 mΩ od 600 mΩ do 1 Ω · CMC 8,4 mΩ od 1 Ω do 3 Ω · CMC 0,018 Ω od 3 Ω do 6 Ω · CMC 0,019 Ω od 6 Ω do 10 Ω · CMC 0,031 Ω od 10 Ω do 30 Ω · CMC 0,034 Ω od 30 Ω do 60 Ω · CMC 0,21 Ω od 60 Ω do 100 Ω · CMC 0,35 Ω from 100 Ω to 300 Ω · CMC 0,69 Ω od 300 Ω do 600 Ω · CMC 1,8 Ω od 600 Ω do 1 000 Ω · CMC 3,5 Ω od 1 000 Ω do 2 000 Ω · CMC 6,7 Ω CMC for the field of measured quantity: Electrical quantities |
Direct measurement on a calibrator of inspection instruments | KP 5/E | |
| 96 | Tripping loop and network impedance / inspection instruments, impedance meters | Calibration – electrical quantities 40 Hz to 60 Hz — 20 mΩ to 40 mΩ · CMC 5,4 mΩ 40 Hz to 60 Hz — 40 mΩ to 80 mΩ · CMC 5,7 mΩ 40 Hz to 60 Hz — 80 mΩ to 200 mΩ · CMC 5,6 mΩ 40 Hz to 60 Hz — 200 mΩ to 400 mΩ · CMC 6,3 mΩ 40 Hz to 60 Hz — 400 mΩ to 600 mΩ · CMC 6,2 mΩ 40 Hz to 60 Hz — 600 mΩ to 1 Ω · CMC 8,9 mΩ 40 Hz to 60 Hz — 1 Ω to 3 Ω · CMC 0,019 Ω 40 Hz to 60 Hz — 3 Ω to 6 Ω · CMC 0,032 Ω 40 Hz to 60 Hz — 6 Ω to 10 Ω · CMC 0,056 Ω 40 Hz to 60 Hz — 10 Ω to 30 Ω · CMC 0,1 Ω 40 Hz to 60 Hz — 30 Ω to 60 Ω · CMC 0,3 Ω 40 Hz to 60 Hz — 60 Ω to 100 Ω · CMC 0,55 Ω 40 Hz to 60 Hz — 100 Ω to 300 Ω · CMC 1,1 Ω 40 Hz to 60 Hz — 300 Ω to 600 Ω · CMC 2,9 Ω 40 Hz to 60 Hz — 600 Ω to 1 000 Ω · CMC 5,6 Ω 40 Hz to 60 Hz — 1 000 Ω to 2 000 Ω · CMC 11 Ω CMC for the field of measured quantity: Electrical quantities |
Direct measurement on a calibrator of inspection instruments | KP 5/E | |
| 97 | Tripping current / inspection instruments | Calibration – electrical quantities 40 Hz to 60 Hz — 0 mA to 0,2 mA · CMC 0,0015 mA 40 Hz to 60 Hz — 0,2 mA to 0,5 mA · CMC 1,2 % 40 Hz to 60 Hz — 0,5 mA to 1 mA · CMC 1,1 % 40 Hz to 60 Hz — 1 mA to 3 000 mA · CMC 1 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement on a calibrator of inspection instruments | KP 5/E | |
| 98 | Leakage current / inspection instruments, leakage current meters | Calibration – electrical quantities 40 Hz to 60 Hz — 0 mA to 0,1 mA · CMC 0,10 % + 0,00004 mA 40 Hz to 60 Hz — 0,1 mA to 1 mA · CMC 0,10 % + 0,00040 mA 40 Hz to 60 Hz — 1 mA to 10 mA · CMC 0,10 % + 0,0040 mA 40 Hz to 60 Hz — 10 mA to 30 mA · CMC 0,10 % + 0,040 mA CMC for the field of measured quantity: Electrical quantities |
Comparison with a reference multimeter | KP 5/E | |
| 99 | Non-linear distortion THD, THF / distortion meters and analyzers, signal analyzers, radiocommunication testers, testing and inspection instruments and their parts for fundamental harmonic level UP-P = (7 to 20) V | Calibration – electrical quantities Fundamental harmonic: 20 Hz to 50 kHz — 0,1 % to 0,3 % · CMC 2,0 % + 0,039 % (abs.) Fundamental harmonic: 50 kHz to 100 kHz — 0,1 % to 0,3 % · CMC 2,3 % + 0,039 % (abs.) Fundamental harmonic: 100 kHz to 200 kHz — 0,1 % to 0,3 % · CMC 2,9 % + 0,039 % (abs.) Fundamental harmonic: 20 Hz to 1 kHz — 0,3 % to 1 % · CMC 7,0 % + 0,033 % (abs.) Fundamental harmonic: 1 kHz to 50 kHz — 0,3 % to 1 % · CMC 4,1 % + 0,033 % (abs.) Fundamental harmonic: 50 kHz to 100 kHz — 0,3 % to 1 % · CMC 5,5 % + 0,033 % (abs.) Fundamental harmonic: 100 kHz to 200 kHz — 0,3 % to 1 % · CMC 7,0 % + 0,030 % (abs.) Fundamental harmonic: 20 Hz to 50 kHz — 1 % to 10 % · CMC 18 % + 0,010 % (abs.) Fundamental harmonic: 50 kHz to 100 kHz — 1 % to 10 % · CMC 14 % + 0,012 % (abs.) Fundamental harmonic: 100 kHz to 200 kHz — 1 % to 10 % · CMC 20 % + 0,010 % (abs.) Fundamental harmonic: 20 Hz to 1 kHz — 10 % to 40 % · CMC 20 % + 0,10 % (abs.) Fundamental harmonic: 1 kHz to 50 kHz — 10 % to 40 % · CMC 20 % + 0,15 % (abs.) Fundamental harmonic: 50 kHz to 100 kHz — 10 % to 40 % · CMC 20 % + 0,020 % (abs.) Fundamental harmonic: 20 Hz to 50 kHz — 0 % · CMC 0,010 % (abs.) Fundamental harmonic: 50 kHz — 0 % · CMC 0,031 % (abs.) Fundamental harmonic: 100 kHz — 0 % · CMC 0,054 % (abs.) Fundamental harmonic: 20 Hz — 0,1 % · CMC 0,012 % (abs.) Fundamental harmonic: 50 Hz — 0,1 % · CMC 0,012 % (abs.) Fundamental harmonic: 100 Hz — 0,1 % · CMC 0,012 % (abs.) Fundamental harmonic: 1 kHz — 0,1 % · CMC 0,012 % (abs.) Fundamental harmonic: 5 kHz — 0,1 % · CMC 0,012 % (abs.) Fundamental harmonic: 10 kHz — 0,1 % · CMC 0,012 % (abs.) Fundamental harmonic: 50 kHz — 0,1 % · CMC 0,026 % (abs.) Fundamental harmonic: 100 kHz — 0,1 % · CMC 0,026 % (abs.) Fundamental harmonic: 20 Hz — 1 % · CMC 0,10 % (abs.) Fundamental harmonic: 50 Hz — 1 % · CMC 0,10 % (abs.) Fundamental harmonic: 100 Hz — 1 % · CMC 0,10 % (abs.) Fundamental harmonic: 1 kHz — 1 % · CMC 0,10 % (abs.) Fundamental harmonic: 5 kHz — 1 % · CMC 0,10 % (abs.) Fundamental harmonic: 10 kHz — 1 % · CMC 0,10 % (abs.) Fundamental harmonic: 50 kHz — 1 % · CMC 0,025 % (abs.) Fundamental harmonic: 100 kHz — 1 % · CMC 0,025 % (abs.) Fundamental harmonic: 20 Hz — 10 % · CMC 1,0 % (abs.) Fundamental harmonic: 50 Hz — 10 % · CMC 1,0 % (abs.) Fundamental harmonic: 100 Hz — 10 % · CMC 1,0 % (abs.) Fundamental harmonic: 1 kHz — 10 % · CMC 1,0 % (abs.) Fundamental harmonic: 5 kHz — 10 % · CMC 1,0 % (abs.) Fundamental harmonic: 10 kHz — 10 % · CMC 1,0 % (abs.) Fundamental harmonic: 50 kHz — 10 % · CMC 2,5 % (abs.) Fundamental harmonic: 100 kHz — 10 % · CMC 2,5 % (abs.) CMC for the field of measured quantity: Electrical quantities |
Direct measurement on a standard distortion generator | KP 22/E | |
| 100 | Non-linear distortion of THF signals with levels from 150 mV to 300 V / signal generators, sources, oscillators, radiocommunication testers, testing and inspection instruments and their parts | Calibration – electrical quantities Fundamental harmonic: 10 Hz to 1 kHz; Measurement bandwidth: 80 kHz — 0 % to 0,1 % · CMC 29,0 % + 0,01 % (abs.) Fundamental harmonic: 1 kHz to 10 kHz; Measurement bandwidth: 80 kHz — 0 % to 0,1 % · CMC 30,0 % + 0,01 % (abs.) Fundamental harmonic: 10 kHz to 20 kHz; Measurement bandwidth: 80 kHz — 0 % to 0,1 % · CMC 36,0 % + 0,01 % (abs.) Fundamental harmonic: 10 Hz to 10 kHz; Measurement bandwidth: 80 kHz — 0,1 % to 0,3 % · CMC 18,0 % Fundamental harmonic: 10 kHz to 20 kHz; Measurement bandwidth: 80 kHz — 0,1 % to 0,3 % · CMC 19,0 % Fundamental harmonic: 20 kHz to 50 kHz; Measurement bandwidth: 80 kHz — 0,1 % to 0,3 % · CMC 39,0 % Fundamental harmonic: 10 Hz to 20 kHz; Measurement bandwidth: 80 kHz — 0,3 % to 1 % · CMC 16,0 % Fundamental harmonic: 20 kHz to 50 kHz; Measurement bandwidth: 500 kHz — 0,3 % to 1 % · CMC 27,0 % Fundamental harmonic: 50 kHz to 100 kHz; Measurement bandwidth: 500 kHz — 0,3 % to 1 % · CMC 30,0 % Fundamental harmonic: 10 Hz to 20 kHz; Measurement bandwidth: 80 kHz — 1 % to 3 % · CMC 15,0 % Fundamental harmonic: 20 kHz to 50 kHz; Measurement bandwidth: 500 kHz — 1 % to 3 % · CMC 22,0 % Fundamental harmonic: 50 kHz to 100 kHz; Measurement bandwidth: 500 kHz — 1 % to 3 % · CMC 23,0 % Fundamental harmonic: 10 Hz to 10 kHz; Measurement bandwidth: 80 kHz — 3 % to 10 % · CMC 14,0 % Fundamental harmonic: 10 kHz to 20 kHz; Measurement bandwidth: 500 kHz — 3 % to 10 % · CMC 15,0 % Fundamental harmonic: 20 kHz to 100 kHz; Measurement bandwidth: 500 kHz — 3 % to 10 % · CMC 21,0 % Fundamental harmonic: 20 Hz to 20 kHz; Measurement bandwidth: 80 kHz — 10 % to 40 % · CMC 14,0 % Fundamental harmonic: 20 kHz to 100 kHz; Measurement bandwidth: 500 kHz — 10 % to 40 % · CMC 20,0 % CMC for the field of measured quantity: Electrical quantities |
Direct measurement using a standard distortion analyzer | KP 22/E | |
| 101 | Impedance module (complex) / standards, impedance elements, coupling networks, artificial networks, filters | Calibration – electrical quantities 9 kHz to 500 MHz — 0 Ω to 10 Ω · CMC 0,5 Ω 9 kHz to 500 MHz — 10 Ω to 13 Ω · CMC 5,0 % 9 kHz to 500 MHz — 13 Ω to 16 Ω · CMC 4,0 % 9 kHz to 500 MHz — 16 Ω to 20 Ω · CMC 3,5 % 9 kHz to 500 MHz — 20 Ω to 30 Ω · CMC 3,0 % 9 kHz to 500 MHz — 30 Ω to 40 Ω · CMC 2,0 % 9 kHz to 500 MHz — 40 Ω to 70 Ω · CMC 1,9 % 9 kHz to 500 MHz — 70 Ω to 85 Ω · CMC 2,0 % 9 kHz to 500 MHz — 85 Ω to 100 Ω · CMC 2,5 % 9 kHz to 500 MHz — 100 Ω to 120 Ω · CMC 3,0 % 9 kHz to 500 MHz — 120 Ω to 170 Ω · CMC 4,0 % 9 kHz to 500 MHz — 170 Ω to 220 Ω · CMC 5,0 % 9 kHz to 500 MHz — 220 Ω to 270 Ω · CMC 6,0 % 9 kHz to 500 MHz — 270 Ω to 300 Ω · CMC 7,0 % 9 kHz to 500 MHz — 300 Ω to 350 Ω · CMC 8,0 % 9 kHz to 500 MHz — 350 Ω to 400 Ω · CMC 9,0 % CMC for the field of measured quantity: Electrical quantities |
Measurement by a vector network analyzer | KP 33/E | |
| 102 | Impedance phase (complex) / standards, impedance elements, coupling networks, artificial networks, filters | Calibration – electrical quantities −360° to +360° · CMC 1,7° CMC for the field of measured quantity: Electrical quantities |
Measurement by a vector network analyzer | KP 33/E | |
| 103 | Reflection coefficient modulus / HF instruments, HF circuit elements, scalar and vector meters | Calibration – electrical quantities 9 kHz to 2 GHz — 0,000 to 0,178 · CMC 0,009 (abs.) 2 GHz to 6 GHz — 0,000 to 0,178 · CMC 0,015 (abs.) 9 kHz to 2 GHz — 0,178 to 0,500 · CMC 3,5 % 2 GHz to 6 GHz — 0,178 to 0,500 · CMC 5,5 % 9 kHz to 2 GHz — 0,500 to 0,708 · CMC 3,0 % 2 GHz to 6 GHz — 0,500 to 0,708 · CMC 4,0 % 9 kHz to 2 GHz — 0,708 to 1,000 · CMC 2,5 % 2 GHz to 6 GHz — 0,708 to 1,000 · CMC 4,0 % CMC for the field of measured quantity: Electrical quantities |
Measurement by a vector network analyzer | KP 33/E | |
| 104 | Reflection coefficient phase / HF instruments, HF circuit elements, vector meters | Calibration – electrical quantities Reflection coefficient: 0,009 to 0,018; Frequency: 9 kHz to 2 GHz — −360° to +360° · CMC 42,0° Reflection coefficient: 0,009 to 0,018; Frequency: 2 GHz to 6 GHz — −360° to +360° · CMC 55,0° Reflection coefficient: 0,018 to 0,056; Frequency: 9 kHz to 2 GHz — −360° to +360° · CMC 20,0° Reflection coefficient: 0,018 to 0,056; Frequency: 2 GHz to 6 GHz — −360° to +360° · CMC 42,0° Reflection coefficient: 0,056 to 0,178; Frequency: 9 kHz to 2 GHz — −360° to +360° · CMC 6,0° Reflection coefficient: 0,056 to 0,178; Frequency: 2 GHz to 6 GHz — −360° to +360° · CMC 13,0° Reflection coefficient: 0,178 to 0,5; Frequency: 9 kHz to 2 GHz — −360° to +360° · CMC 2,0° Reflection coefficient: 0,178 to 0,5; Frequency: 2 GHz to 6 GHz — −360° to +360° · CMC 4,0° Reflection coefficient: 0,5 to 1; Frequency: 9 kHz to 1 GHz — −360° to +360° · CMC 1,0° Reflection coefficient: 0,5 to 1; Frequency: 1 GHz to 2 GHz — −360° to +360° · CMC 1,3° Reflection coefficient: 0,5 to 1; Frequency: 2 GHz to 3 GHz — −360° to +360° · CMC 2,0° Reflection coefficient: 0,5 to 1; Frequency: 3 GHz to 5 GHz — −360° to +360° · CMC 3,0° Reflection coefficient: 0,5 to 1; Frequency: 5 GHz to 6 GHz — −360° to +360° · CMC 3,5° CMC for the field of measured quantity: Electrical quantities |
Measurement by a vector network analyzer | KP 33/E | |
| 105 | DC power / DC wattmeters and DC power meters | Calibration – electrical quantities 0,2 V to 280 V; 0,2 mA to 20 A — 40 µW to 5,6 kW · CMC 0,021 % CMC for the field of measured quantity: Electrical quantities |
Comparison with indirectly measured reference performance | KP 9/E | |
| 106 | DC power / DC power generators and sources | Calibration – electrical quantities 0,1 V to 1 kV; 0,1 A to 100 A — 10 µW to 100 kW · CMC 0,0049 % CMC for the field of measured quantity: Electrical quantities |
Indirect power measurement by measuring voltage and current with two reference multimeters | KP 9/E | |
| 107 | AC active power / AC wattmeters and AC power meters in the range 0,2 V to 280 V; 0,2 mA to 20 A; 15 Hz to 1 kHz; 0° to 360° | Calibration – electrical quantities 0 W to 5,6 kW · CMC 0,73 mW/VA CMC for the field of measured quantity: Electrical quantities |
Generation by a reference power calibrator | KP 9/E | |
| 108 | AC apparent power / AC wattmeters and AC power meters in the range 0,2 V to 280 V; 0,2 mA to 20 A; 15 Hz to 1 kHz | Calibration – electrical quantities 40 µVA to 5,6 kVA · CMC 0,75 mVA/VA CMC for the field of measured quantity: Electrical quantities |
Generation by a reference power calibrator | KP 9/E | |
| 109 | AC reactive power / AC wattmeters and AC power meters in the range 0,2 V to 280 V; 0,2 mA to 20 A; 15 Hz to 1 kHz; 0° to 360° | Calibration – electrical quantities 0 var to 5,6 kvar · CMC 0,73 mvar/VA CMC for the field of measured quantity: Electrical quantities |
Generation by a reference power calibrator | KP 9/E | |
| 110 | AC active power / AC power generators and sources in the range 0.1 V to 700 V; 0° to 360° | Calibration – electrical quantities 0,01 A to 0,1 A; 40 Hz to 1 kHz — 0 W to 70 W · CMC 0,97 mW/VA 0,1 A to 16 A; 15 Hz to 1 kHz — 0 W to 11,2 kW · CMC 0,97 mW/VA 16 A to 100 A; 40 Hz to 1 kHz — 0 W to 70 kW · CMC 0,97 mW/VA CMC for the field of measured quantity: Electrical quantities |
Measurement by a digital sampling wattmeter | KP 9/E | |
| 111 | AC apparent power / AC power generators and sources in the range 0.1 V to 700 V | Calibration – electrical quantities 0,01 A to 0,1 A; 40 Hz to 1 kHz — 1 mVA to 70 VA · CMC 0,97 mVA/VA 0,1 A to 16 A; 15 Hz to 1 kHz — 10 mVA to 11,2 kVA · CMC 0,97 mVA/VA 16 A to 100 A; 40 Hz to 1 kHz — 1,6 VA to 700 kVA · CMC 0,97 mVA/VA CMC for the field of measured quantity: Electrical quantities |
Measurement by a reference digital sampling wattmeter | KP 9/E | |
| 112 | AC reactive power / AC power generators and sources in the range 0.1 V to 700 V; 0° to 360° | Calibration – electrical quantities 0,01 A to 0,1 A; 40 Hz to 1 kHz — 0 var to 70 var · CMC 0,97 mvar/VA 0,1 A to 16 A; 15 Hz to 1 kHz — 0 var to 11,2 kvar · CMC 0,97 mvar/VA 16 A to 100 A; 40 Hz to 1 kHz — 0 var to 70 kvar · CMC 0,97 mvar/VA CMC for the field of measured quantity: Electrical quantities |
Measurement by a reference digital sampling wattmeter | KP 9/E | |
| 113 | Phase angle / wattmeters and power meters in the range 0.2 V to 280 V; 0.2 mA to 20 A | Calibration – electrical quantities 15 Hz to 200 Hz — 0° to 360° · CMC 0,15° 200 Hz to 1 kHz — 0° to 360° · CMC 0,25° CMC for the field of measured quantity: Electrical quantities |
Generation by a reference power calibrator | KP 9/E | |
| 114 | Reflection coefficient modulus / scalar and vector reflection and impedance meters | Calibration – electrical quantities APC7 Connector; 100 kHz to 3 GHz — 0,00 to 0,03 · CMC 0,004 abs. N Connector; 100 kHz to 3 GHz — 0,00 to 0,03 · CMC 0,004 abs. N Connector; 3 GHz to 6 GHz — 0,00 to 0,03 · CMC 0,006 abs. APC7 Connector; 100 kHz to 1,5 GHz — 0,95 to 1 · CMC 0,005 abs. APC7 Connector; 1,5 GHz to 2,8 GHz — 0,95 to 1 · CMC 0,007 abs. APC7 Connector; 2,8 GHz to 3,0 GHz — 0,95 to 1 · CMC 0,008 abs. N Connector; 100 kHz to 2 GHz — 0,95 to 1 · CMC 0,014 abs. N Connector; 2 GHz to 6 GHz — 0,95 to 1 · CMC 0,026 abs. CMC for the field of measured quantity: Electrical quantities |
Direct measurement of fixed reference elements | KP 34/E | |
| 115 | Reflection coefficient phase / vector reflection and impedance meters | Calibration – electrical quantities Module 0,95 to 1; APC7 Connector; 100 kHz to 1,8 GHz — −180° to +180° · CMC 0,3° Module 0,95 to 1; APC7 Connector; 1,8 GHz to 3,0 GHz — −180° to +180° · CMC 0,35° Module 0,95 to 1; N Connector; 100 kHz to 3,0 GHz — −180° to +180° · CMC 0,4° Module 0,95 to 1; N Connector; 3,0 GHz to 6,0 GHz — −180° to +180° · CMC 0,8° CMC for the field of measured quantity: Electrical quantities |
Direct measurement of fixed reference elements | KP 34/E | |
| 116 | Absorbance (transmittance) in the UV-VIS-NIR spectrum / photometers and spectrometers, “Elisa readers”, solid and liquid filters for checking absorbance, stray light and spectral resolution | Calibration – optical quantities wavelength: 235 nm — 0,1 to 0,3 · CMC 0,0030 abs. wavelength: 257 nm — 0,1 to 0,3 · CMC 0,0030 abs. wavelength: 302 nm — 0,1 to 0,3 · CMC 0,0025 abs. wavelength: 313 nm — 0,1 to 0,3 · CMC 0,0025 abs. wavelength: 345 nm — 0,1 to 0,3 · CMC 0,0025 abs. wavelength: 350 nm — 0,1 to 0,3 · CMC 0,0025 abs. wavelength: 395 nm — 0,1 to 0,3 · CMC 0,0025 abs. wavelength: 512 nm — 0,1 to 0,3 · CMC 0,0040 abs. wavelength: 678 nm — 0,1 to 0,3 · CMC 0,0030 abs. wavelength: 235 nm — 0,3 to 0,6 · CMC 0,0040 abs. wavelength: 257 nm — 0,3 to 0,6 · CMC 0,0040 abs. wavelength: 302 nm — 0,3 to 0,6 · CMC 0,0090 abs. wavelength: 313 nm — 0,3 to 0,6 · CMC 0,0040 abs. wavelength: 345 nm — 0,3 to 0,6 · CMC 0,0040 abs. wavelength: 350 nm — 0,3 to 0,6 · CMC 0,0040 abs. wavelength: 395 nm — 0,3 to 0,6 · CMC 0,0080 abs. wavelength: 512 nm — 0,3 to 0,6 · CMC 0,0055 abs. wavelength: 678 nm — 0,3 to 0,6 · CMC 0,0040 abs. wavelength: 235 nm — 0,6 to 1,6 · CMC 0,0050 abs. wavelength: 257 nm — 0,6 to 1,6 · CMC 0,0050 abs. wavelength: 302 nm — 0,6 to 1,6 · CMC 0,0098 abs. wavelength: 345 nm — 0,6 to 1,6 · CMC 0,0050 abs. wavelength: 350 nm — 0,6 to 1,6 · CMC 0,0050 abs. wavelength: 395 nm — 0,6 to 1,6 · CMC 0,0090 abs. wavelength: 512 nm — 0,6 to 1,6 · CMC 0,0060 abs. wavelength: 678 nm — 0,6 to 1,6 · CMC 0,0040 abs. wavelength: 190 nm to 390 nm — 1,6 to 5,0 · CMC 0,010 abs. CMC for the field of measured quantity: Optical quantities |
Direct measurement and comparison using standard filters and CRM | KP 1/S | |
| 117 | Absorbance (transmittance) in the UV-VIS-NIR spectrum / photometers and spectrometers, “Elisa readers”, solid and liquid filters for checking absorbance, stray light and spectral resolution | Calibration – optical quantities wavelength: 250 nm to 400 nm — 0,03 (93,3 %) to 0,6 (25,1 %) · CMC 0,0045 abs. (1,1 % abs.) wavelength: 250 nm to 400 nm — 0,6 (25,1 %) to 1,1 (8 %) · CMC 0,0050 abs. (1,2 % abs.) wavelength: 396 nm to 401 nm — 0 (100 %) to 0,4 (40 %) · CMC 0,0060 abs. (1,4 % abs.) wavelength: 401 nm to 406 nm — 0 (100 %) to 0,4 (40 %) · CMC 0,0030 abs. (0,7 % abs.) wavelength: 406 nm to 420 nm — 0 (100 %) to 0,4 (40 %) · CMC 0,0030 abs. (0,7 % abs.) wavelength: 396 nm to 401 nm — 0,4 (40 %) to 0,8 (16 %) · CMC 0,0030 abs. (0,7 % abs.) wavelength: 401 nm to 406 nm — 0,4 (40 %) to 0,8 (16 %) · CMC 0,0030 abs. (0,7 % abs.) wavelength: 406 nm to 420 nm — 0,4 (40 %) to 0,8 (16 %) · CMC 0,0030 abs. (0,7 % abs.) wavelength: 396 nm to 401 nm — 0,8 (16 %) to 1,2 (6,3 %) · CMC 0,013 abs. (3,0 % abs.) wavelength: 401 nm to 406 nm — 0,8 (16 %) to 1,2 (6,3 %) · CMC 0,0050 abs. (1,2 % abs.) wavelength: 406 nm to 420 nm — 0,8 (16 %) to 1,2 (6,3 %) · CMC 0,0050 abs. (1,2 % abs.) wavelength: 396 nm to 401 nm — 1,2 (6,3 %) to 1,6 (2,5 %) · CMC 0,0060 abs. (2,1 % abs.) wavelength: 401 nm to 406 nm — 1,2 (6,3 %) to 1,6 (2,5 %) · CMC 0,026 abs. (5,9 % abs.) wavelength: 406 nm to 420 nm — 1,2 (6,3 %) to 1,6 (2,5 %) · CMC 0,0090 abs. (1,4 % abs.) wavelength: 396 nm to 401 nm — 1,6 (2,5 %) to 2,8 (0,16 %) · CMC 0,030 abs. (6,7 % abs.) wavelength: 401 nm to 406 nm — 1,6 (2,5 %) to 2,8 (0,16 %) · CMC 0,010 abs. (1,9 % abs.) wavelength: 406 nm to 420 nm — 1,6 (2,5 %) to 2,8 (0,16 %) · CMC 0,0080 abs. (2,3 % abs.) wavelength: 420 nm to 460 nm — 0 (100 %) to 0,35 (45 %) · CMC 0,0035 abs. (0,9 % abs.) wavelength: 460 nm to 600 nm — 0 (100 %) to 0,35 (45 %) · CMC 0,0030 abs. (0,7 % abs.) wavelength: 600 nm to 860 nm — 0 (100 %) to 0,35 (45 %) · CMC 0,0035 abs. (0,9 % abs.) wavelength: 420 nm to 460 nm — 0,35 (45 %) to 0,7 (20 %) · CMC 0,0040 abs. (1,0 % abs.) wavelength: 460 nm to 600 nm — 0,35 (45 %) to 0,7 (20 %) · CMC 0,0030 abs. (0,7 % abs.) wavelength: 600 nm to 860 nm — 0,35 (45 %) to 0,7 (20 %) · CMC 0,0030 abs. (0,7 % abs.) wavelength: 420 nm to 460 nm — 0,7 (20 %) to 1,0 (10 %) · CMC 0,0050 abs. (1,2 % abs.) wavelength: 460 nm to 600 nm — 0,7 (20 %) to 1,0 (10 %) · CMC 0,0040 abs. (1,0 % abs.) wavelength: 600 nm to 860 nm — 0,7 (20 %) to 1,0 (10 %) · CMC 0,0040 abs. (1,0 % abs.) wavelength: 420 nm to 460 nm — 1,0 (10 %) to 1,2 (6,3 %) · CMC 0,0065 abs. (1,5 % abs.) wavelength: 460 nm to 600 nm — 1,0 (10 %) to 1,2 (6,3 %) · CMC 0,0050 abs. (1,2 % abs.) wavelength: 600 nm to 860 nm — 1,0 (10 %) to 1,2 (6,3 %) · CMC 0,0050 abs. (1,2 % abs.) wavelength: 420 nm to 460 nm — 1,2 (6,3 %) to 2,2 (0,63 %) · CMC 0,0080 abs. (1,9 % abs.) wavelength: 460 nm to 600 nm — 1,2 (6,3 %) to 2,2 (0,63 %) · CMC 0,006 abs. (1,5 % abs.) wavelength: 600 nm to 860 nm — 1,2 (6,3 %) to 2,2 (0,63 %) · CMC 0,0065 abs. (1,5 % abs.) wavelength: 860 nm to 950 nm — 0 (100 %) to 0,3 (50,1 %) · CMC 0,0035 abs. (0,9 % abs.) wavelength: 950 nm to 1000 nm — 0 (100 %) to 0,3 (50,1 %) · CMC 0,0035 abs. (0,9 % abs.) wavelength: 1000 nm to 1500 nm — 0 (100 %) to 0,3 (50,1 %) · CMC 0,0030 abs. (0,7 % abs.) wavelength: 860 nm to 950 nm — 0,3 (50,1 %) to 0,65 (22,4 %) · CMC 0,0030 abs. (0,7 % abs.) wavelength: 950 nm to 1000 nm — 0,3 (50,1 %) to 0,65 (22,4 %) · CMC 0,0055 abs. (1,3 % abs.) wavelength: 1000 nm to 1500 nm — 0,3 (50,1 %) to 0,65 (22,4 %) · CMC 0,0060 abs. (1,4 % abs.) wavelength: 860 nm to 950 nm — 0,65 (22,4 %) to 0,8 (16 %) · CMC 0,0040 abs. (1,0 % abs.) wavelength: 950 nm to 1000 nm — 0,65 (22,4 %) to 0,8 (16 %) · CMC 0,0055 abs. (1,3 % abs.) wavelength: 1000 nm to 1500 nm — 0,65 (22,4 %) to 0,8 (16 %) · CMC 0,0060 abs. (1,4 % abs.) wavelength: 860 nm to 950 nm — 0,8 (16 %) to 1,0 (10 %) · CMC 0,0040 abs. (1,0 % abs.) wavelength: 950 nm to 1000 nm — 0,8 (16 %) to 1,0 (10 %) · CMC 0,0040 abs. (1,0 % abs.) wavelength: 860 nm to 950 nm — 1,0 (10 %) to 1,4 (4 %) · CMC 0,0050 abs. (1,2 % abs.) wavelength: 950 nm to 1000 nm — 1,0 (10 %) to 1,4 (4 %) · CMC 0,0050 abs. (1,2 % abs.) CMC for the field of measured quantity: Optical quantities |
Direct measurement using standard filters, comparison with standard filters | KP 1/S | |
| 118 | Absorbance ratio / spectral resolution of UV spectrometers, liquid filters | Calibration – optical quantities wavelength: 265 nm to 270 nm — 1,0 to 2,2 · CMC 0,0020 abs. CMC for the field of measured quantity: Optical quantities |
Direct measurement of CRM and calculation | KP 1/S | |
| 119 | UV-VIS-NIR wavelength / photometers and spectrometers, filters for controlling wavelengths and spectral resolution | Calibration – optical quantities 200 nm to 650 nm · CMC 0,035 nm 650 nm to 900 nm · CMC 0,060 nm 900 nm to 2 600 nm · CMC 0,5 nm CMC for the field of measured quantity: Optical quantities |
Direct measurement of CRM or standard filters, comparison with CRM | KP 1/S | |
| 120 | IR wavenumber / IR and FTIR spectrometers | Calibration – optical quantities 10 000 cm⁻¹ to 4 000 cm⁻¹ · CMC 0,60 cm⁻¹ 4 000 cm⁻¹ to 400 cm⁻¹ · CMC 0,30 cm⁻¹ CMC for the field of measured quantity: Optical quantities |
Direct measurement of standard filters | KP 3/S | |
| 121 | Integral transmittance / fade and transmittance meters | Calibration – optical quantities 18 % to 100 % · CMC 0,7 % CMC for the field of measured quantity: Optical quantities |
Direct measurement of standard filters | KP 1/S | |
| 122 | Colorimetric coordinates / spectrophotometers for colour measurement | Calibration – optical quantities Lighting type: D65, viewing angle: 2°, 10°; L* — 1 to 5 · CMC 0,5 abs. Lighting type: D65, viewing angle: 2°, 10°; L* — 5 to 100 · CMC 0,1 abs. Lighting type: D65, viewing angle: 2°, 10°; a* — -20 to 110 · CMC 0,2 abs. Lighting type: D65, viewing angle: 2°, 10°; a* — -40 to -20 · CMC 0,5 abs Lighting type: D65, viewing angle: 2°, 10°; b* — -20 to 110 · CMC 0,1 abs Lighting type: D65, viewing angle: 2°, 10°; b* — -40 to -20 · CMC 0,25 abs. CMC for the field of measured quantity: Optical quantities |
Direct measurement of colorimetric standards | KP 5/S | |
| 123 | Colorimetric coordinates / colorimetric standards and samples | Calibration – optical quantities Lighting type: D65, viewing angle: 2°, 10°; L* — 0 to 80 · CMC 0,5 abs. Lighting type: D65, viewing angle: 2°, 10°; L* — 80 to 100 · CMC 0,4 abs. Lighting type: D65, viewing angle: 2°, 10°; a* — -90 to -5 · CMC 0,7 abs. Lighting type: D65, viewing angle: 2°, 10°; a* — -5 to +5 · CMC 0,3 abs. Lighting type: D65, viewing angle: 2°, 10°; a* — +5 to 110 · CMC 0,6 abs. Lighting type: D65, viewing angle: 2°, 10°; b* — -90 to -5 · CMC 0,7 abs. Lighting type: D65, viewing angle: 2°, 10°; b* — -5 to +5 · CMC 0,5 abs. Lighting type: D65, viewing angle: 2°, 10°; b* — +5 to 110 · CMC 0,6 abs. CMC for the field of measured quantity: Optical quantities |
Comparison with colorimetric standards | KP 5/S | |
| 124 | Frequency / frequency counters, meters, measuring receivers, oscilloscopes, spectral, modulation and frequency analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) | Calibration – time and frequency quantities from 0,01 Hz to 0,1 Hz · CMC 3,2·10⁻⁹ from 0,1 Hz to 1 Hz · CMC 4,9·10⁻¹⁰ from 1 Hz to 100 Hz · CMC 2,2·10⁻¹⁰ od 100 Hz do 40 GHz · CMC 1,9·10⁻¹⁰ 0,1 MHz · CMC 9,5·10⁻¹¹ 1 MHz · CMC 9,5·10⁻¹¹ 5 MHz · CMC 9,5·10⁻¹¹ 10 MHz · CMC 9,5·10⁻¹¹ CMC for the field of measured quantity: Time and frequency quantities |
Direct measurement on a standard generator | KP 1/F, KP 18/E | |
| 125 | Frequency / reference oscillators, HF, LF, function and pulse generators, frequency converters, spectral, circuit, modulation and frequency analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) | Calibration – time and frequency quantities Sinusoidal signal U = 3,57 V — 0,1 Hz to 1 Hz · CMC 6,5·10⁻⁶ Sinusoidal signal U = 3,57 V — 1 Hz to 10 Hz · CMC 2,1·10⁻⁷ Sinusoidal signal U = 3,57 V — 10 Hz to 100 Hz · CMC 1,4·10⁻⁸ Sinusoidal signal U = 3,57 V — 100 Hz to 1 kHz · CMC 3,1·10⁻⁹ Sinusoidal signal U = 3,57 V — 1 kHz to 10 kHz · CMC 4,1·10⁻¹⁰ Sinusoidal signal U = 3,57 V — 10 kHz to 500 MHz · CMC 1,4·10⁻¹⁰ Sinusoidal signal U = 3,57 V — 500 MHz to 6 GHz · CMC 1,2·10⁻¹⁰ Sinusoidal signal U = 1,0 V — 6 GHz to 15 GHz · CMC 1,0·10⁻¹⁰ + 2,8 Hz Sinusoidal signal U = 1,0 V — 15 GHz to 46 GHz · CMC 1,0·10⁻¹⁰ + 3,7 Hz Rectangular signal U = 5.0 V — 0,01 Hz to 0,1 Hz · CMC 1,3·10⁻¹⁰ Rectangular signal U = 5.0 V — 0,1 Hz to 10 Hz · CMC 1,2·10⁻¹⁰ Rectangular signal U = 5.0 V — 10 Hz to 350 MHz · CMC 1,1·10⁻¹⁰ CMC for the field of measured quantity: Time and frequency quantities |
Direct frequency measurement by a standard counter | KP 2/F, KP 3/F, KP 4/F | |
| 126 | Bandwidth / oscilloscopes | Calibration – time and frequency quantities od 100 kHz do 1 GHz · CMC 13 % od 1 GHz do 3 GHz · CMC 17 % from 3 GHz to 6 GHz · CMC 20 % CMC for the field of measured quantity: Time and frequency quantities |
Indirect measurement by a standard generator and HF voltmeter | KP 18/E | |
| 127 | Time, time interval / counters, stopwatches, generators, timers, synthesizers, converters | Calibration – time and frequency quantities 1 ns to 10 ns · CMC 0,75 ns 10 ns to 10 ms · CMC 0,40 ns 10 ms to 1 s · CMC 0,51 ns CMC for the field of measured quantity: Time and frequency quantities |
Direct measurement using a standard oscilloscope and counter | KP 2/F | |
| 128 | Time, time interval / counters, stopwatches, generators, timers, synthesizers, converters | Calibration – time and frequency quantities 1 s to 10 s · CMC 1,5 ns 10 s to 100 s · CMC 1,5·10⁻¹⁰ 100 s to 100000 s · CMC 1,1·10⁻¹⁰ CMC for the field of measured quantity: Time and frequency quantities |
Direct comparison, electrical triggering | KP 18/E | |
| 129 | Time, time interval / counters, stopwatches, generators, timers, synthesizers, converters | Calibration – time and frequency quantities 10 s to 12 h · CMC 0,06 s CMC for the field of measured quantity: Time and frequency quantities |
Comparison with a standard, manual triggering | KP 5/F | |
| 130 | Tripping time of residual current circuit breakers / inspection instruments | Calibration – time and frequency quantities 10 ms to 40 ms · CMC 0,26 ms 40 ms to 80 ms · CMC 0,27 ms 80 ms to 200 ms · CMC 0,30 ms 200 ms to 500 ms · CMC 0,35 ms 500 ms to 1000 ms · CMC 0,07 % 1000 ms to 2000 ms · CMC 0,045 % 2000 ms to 5000 ms · CMC 0,033 % CMC for the field of measured quantity: Time and frequency quantities |
Direct measurement on a calibrator of inspection instruments | KP 5/E | |
| Zlín | |||||
| 131 | Hardness tester Shore A | Calibration – hardness Spring force 31 mN — 1 Shore A to 100 Shore A · CMC 0,3 Shore A CMC for the field of measured quantity: Hardness |
Direct measurement of force on a calibrator and measurement of tip geometry using an image analyzer | KP 1/H | |
| 132 | Hardness tester Shore C | Calibration – hardness Spring force 71 mN — 1 Shore C to 100 Shore C · CMC 0,3 Shore C CMC for the field of measured quantity: Hardness |
Direct measurement of force on a calibrator and measurement of tip geometry using an image analyzer | KP 1/H | |
| 133 | Hardness tester Shore D | Calibration – hardness Spring force 71 mN — 1 Shore D to 100 Shore D · CMC 0,3 Shore D CMC for the field of measured quantity: Hardness |
Direct measurement of force on a calibrator and measurement of tip geometry using an image analyzer | KP 1/H | |
| 134 | Hardness tester IRHD/N | Calibration – hardness 10 IRHD to 100 IRHD · CMC 0,6 IRHD CMC for the field of measured quantity: Hardness |
Direct measurement of force on a calibrator and measurement of tip geometry using an image analyzer | KP 2/H | |
| 135 | Hardness tester IRHD/m | Calibration – hardness 10 IRHD to 100 IRHD · CMC 1,1 IRHD CMC for the field of measured quantity: Hardness |
Direct measurement of force on a calibrator and measurement of tip geometry using an image analyzer | KP 2/H | |
| 136 | Volume concentration of ozone / ozone generators | Calibration – physicochemical quantities from 0 nmol/mol to 100 nmol/mol · CMC 2,4 nmol/mol od 100 nmol/mol do 5 000 nmol/mol · CMC 17 nmol/mol CMC for the field of measured quantity: Physicochemical quantities |
Spectrophotometric measurement of ozone concentration | KP 1/O | |
| Brno | |||||
| 137 | Thermocouples | Calibration – temperature from −40 °C to 0 °C · CMC 0,6 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in liquid bath | KP 1/T | |
| 138 | Thermocouples | Calibration – temperature 0 °C · CMC 0,6 °C from 0 °C to 100 °C · CMC 0,6 °C from 100 °C to 250 °C · CMC 0,7 °C od 250 °C do 600 °C · CMC 0,8 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in dry block | KP 1/T | |
| 139 | Thermocouples | Calibration – temperature from 600 °C to 800 °C · CMC 1,5 °C od 800 °C do 1 200 °C · CMC 2,2 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard thermoelectric sensor in a horizontal oven | KP 1/T | |
| 140 | Resistance temperature sensors, resistance temperature sensors with a transducer | Calibration – temperature from −40 °C to 0 °C · CMC 0,07 °C 0 °C · CMC 0,05 °C from 0 °C to 100 °C · CMC 0,08 °C from 100 °C to 250 °C · CMC 0,10 °C CMC for the field of measured quantity: Temperature |
Comparison with a reference resistance sensor in a liquid bath or dry block | KP 2/T | |
| 141 | Resistance temperature sensors, resistance temperature sensors with a transducer | Calibration – temperature from 250 °C to 350 °C · CMC 0,43 °C od 350 °C do 600 °C · CMC 0,62 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in dry block | KP 2/T | |
| 142 | Glass thermometers | Calibration – temperature from −40 °C to 0 °C · CMC 0,20 °C 0 °C · CMC 0,05 °C from 0 °C to 100 °C · CMC 0,14 °C from 100 °C to 250 °C · CMC 0,15 °C from 250 °C to 300 °C · CMC 0,30 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in liquid bath | KP 3/T | |
| 143 | Direct-indicating thermometers | Calibration – temperature −80 °C to −40 °C · CMC 0,3 °C CMC for the field of measured quantity: Temperature |
Comparison with a reference direct indicating temperature sensor in an electrical vertical furnace | KP 4/T | |
| 144 | Direct-indicating thermometers | Calibration – temperature from −40 °C to 0 °C · CMC 0,07 °C 0 °C · CMC 0,06 °C from 0 °C to 100 °C · CMC 0,09 °C from 100 °C to 250 °C · CMC 0,10 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in liquid bath | KP 4/T | |
| 145 | Direct-indicating thermometers | Calibration – temperature from 250 °C to 350 °C · CMC 0,44 °C od 350 °C do 600 °C · CMC 0,62 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard resistance sensor in dry block | KP 4/T | |
| 146 | Direct-indicating thermometers | Calibration – temperature from 600 °C to 800 °C · CMC 1,6 °C od 800 °C do 1 200 °C · CMC 2,2 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard thermoelectric sensor in a horizontal oven | KP 4/T | |
| 147 | Controllers and indicators | Calibration – temperature Type K — −270 °C to 0 °C · CMC 0,3 °C Type K — 0 °C to 1370 °C · CMC 0,3 °C Type S — −50 °C to 0 °C · CMC 0,9 °C Type S — 0 °C to 1 780 °C · CMC 0,6 °C Type B — 200 °C to 500 °C · CMC 1,5 °C Type B — 500 °C to 1 820 °C · CMC 0,6 °C Type R — 0 °C to 1 768 °C · CMC 0,6 °C Type J — −210 °C to 0 °C · CMC 0,2 °C Type J — 0 °C to 1 200 °C · CMC 0,2 °C Type N — −270 °C to 0 °C · CMC 0,2 °C Type N — 0 °C to 1 300 °C · CMC 0,2 °C CMC for the field of measured quantity: Temperature |
Simulation of thermoelectric temperature sensors using DC voltage | KP 5/T | |
| 148 | Controllers and indicators | Calibration – temperature RTD — −200 °C to −50 °C · CMC 0,05 °C RTD — −50 °C to 850 °C · CMC 0,05 °C CMC for the field of measured quantity: Temperature |
Simulation of resistance temperature sensors using DC electrical resistors | KP 5/T | |
| 149 | Non-contact thermometers | Calibration – temperature od −30 °C do 23 °C · CMC 2,2 °C od 23 °C do 150 °C · CMC 2,1 °C from 150 °C to 250 °C · CMC 2,2 °C from 250 °C to 350 °C · CMC 2,5 °C from 350 °C to 500 °C · CMC 3,0 °C CMC for the field of measured quantity: Temperature |
Direct measurement on a black body, comparison with a standard contactless thermometer | KP 6/T | |
| 150 | Direct-indicating thermometers integrated in conditioning and thermal chambers and baths | Calibration – temperature od −80 °C do 250 °C · CMC 1,0 °C od 250 °C do 600 °C · CMC 1,1 °C from 600 °C to 700 °C · CMC 1,7 °C od 700 °C do 850 °C · CMC 1,9 °C od 850 °C do 1 200 °C · CMC 2,9 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard digital thermometer in a chamber (characterisation) | KP 22/T | |
| Prague | |||||
| 151 | IR wavenumber / IR and FTIR spectrometers | Calibration – optical quantities 10 000 cm⁻¹ to 4 000 cm⁻¹ · CMC 0,60 cm⁻¹ 4 000 cm⁻¹ to 400 cm⁻¹ · CMC 0,30 cm⁻¹ CMC for the field of measured quantity: Optical quantities |
Direct measurement of standard filters | KP 3/S | |
Reference standards (49)
KP 1/D
KP 1/E
KP 1/F
KP 1/H
KP 1/O
KP 1/P
KP 1/S
KP 1/T
KP 1/V
KP 11/E
KP 12/E
KP 13/E
KP 14/E
KP 15/E
KP 16/E
KP 18/E
KP 2/D
KP 2/E
KP 2/F
KP 2/H
KP 2/P
KP 2/T
KP 21/D
KP 21/E
KP 21/T
KP 22/E
KP 22/T
KP 3/D
KP 3/E
KP 3/F
KP 3/P
KP 3/S
KP 3/T
KP 31/E
KP 32/E
KP 33/E
KP 34/E
KP 4/D
KP 4/F
KP 4/T
KP 5/E
KP 5/F
KP 5/S
KP 5/T
KP 6/E
KP 6/F
KP 6/T
KP 8/E
KP 9/E