📏 Coordinate measuring machines with imaging (vision) probing systems: acceptance and reverification tests — length measurement errors EB, EU, EBZ, EBXY, EBV, squareness error ESQ, probing errors PF2D and PFV2D; PN-EN ISO 10360-7

Metrology PN-EN ISO 10360-7

In short

Acceptance tests verifying the performance of a Cartesian coordinate measuring machine (CMM) with an imaging probing system as stated by the manufacturer, and reverification tests by which the user periodically reverifies it. The test is performed according to PN-EN ISO 10360-7:2011 (English and Polish version; implements EN ISO 10360-7:2011 and ISO 10360-7:2011 [IDT]) with corrigendum PN-EN ISO 10360-7:2011/Ap1:2014-09; STATUS (as of 19.11.2026): PKN cards: PN-EN ISO 10360-7:2011 “Specyfikacje geometrii wyrobów (GPS) -- Badania odbiorcze i okresowe współrzędnościowych maszyn pomiarowych (CMM) -- Część 7: CMM wyposażone w zespoły głowicy pomiarowej rejestrującej obraz” — English version (published 28-06-2011, 42 pages; corrigendum PN-EN ISO 10360-7:2011/Ap1:2014-09E) and Polish version (published 02-10-2014, 48 pages), KT 48 Fundamentals of Machine Design, implements EN ISO 10360-7:2011 and ISO 10360-7:2011 [IDT]; the cards give no withdrawal date. The procedure comprises 4 steps; it is used for: Acceptance tests of Cartesian CMMs with imaging (video, vision) probing systems in the discrete-point mode — confirmation of the performance stated by the manufacturer (clause 1), Reverification tests of such machines by the user (clause 1), Determination of length measurement errors EB, EU, EBZ, EUZ, EBXY, EUXY, EBV, EUV and of the squareness error ESQ (clause 3).

At a glance

  • Standard: PN-EN ISO 10360-7:2011 (English and Polish version; implements EN ISO 10360-7:2011 and ISO 10360-7:2011 [IDT]) with corrigendum PN-EN ISO 10360-7:2011/Ap1:2014-09
  • Category: Metrology
  • Procedure steps: 4
  • STATUS (as of 19.11.2026): PKN cards: PN-EN ISO 10360-7:2011 “Specyfikacje geometrii wyrobów (GPS) -- Badania odbiorcze i okresowe współrzędnościowych maszyn pomiarowych (CMM) -- Część 7: CMM wyposażone w zespoły głowicy pomiarowej rejestrującej obraz” — English version (published 28-06-2011, 42 pages; corrigendum PN-EN ISO 10360-7:2011/Ap1:2014-09E) and Polish version (published 02-10-2014, 48 pages), KT 48 Fundamentals of Machine Design, implements EN ISO 10360-7:2011 and ISO 10360-7:2011 [IDT]; the cards give no withdrawal date.
  • Scope (clause 1): Cartesian CMMs with imaging probing systems, discrete-point mode; contact CMMs — ISO 10360-2
  • Length measurement errors (clause 3): EB, EU (with RB, RU — range of three repetitions); EBZ, EUZ (perpendicular to the measuring plane); EBXY, EUXY (parallel); EBV, EUV (in the field of view without CMM motion)

Overview

WHAT THE STANDARD COVERS. According to clause 1 of ISO 10360-7:2011 the standard specifies the acceptance tests for verifying the performance of a CMM used for measuring linear dimensions as stated by the manufacturer, and the reverification tests that enable the user to periodically reverify it. The tests apply only to Cartesian CMMs using imaging probing systems (in the title of the Polish edition: “zespoły głowicy pomiarowej rejestrującej obraz”) of any type operating in the discrete-point probing mode. The standard does not explicitly apply to non-Cartesian CMMs or CMMs using other types of optical probing (parties may apply it by mutual agreement), nor to CMMs with contact probing systems (these — ISO 10360-2). It specifies performance requirements that can be assigned by the manufacturer or the user, the manner of execution of the tests, rules for proving conformance and applications of the tests.

ACCORDING TO THE TEXT OF ISO 10360-7:2011 (publisher's sample: foreword, introduction, clauses 1–3). The standard was prepared by ISO/TC 213. The tests have two objectives: to test the error of indication of a calibrated test length measured with an imaging probing system and to test the errors of the system itself; the result has direct traceability to the unit of length — the metre. The structure parallels ISO 10360-2 (CMMs with contact probing systems), and the test methodology is intentionally similar. References: ISO 10360-1:2000, ISO 10360-2:2009, ISO 14253-1:1998, ISO 14660-1:1999, ISO/TS 23165:2006 and ISO/IEC Guide 99.

TERMS (clause 3). An imaging probing system creates measurement points through an imaging system (a video or vision probing system is an imaging probing system); field of view (FOV), measuring window (region of interest in the FOV), measuring plane. Normal CTE material: coefficient of thermal expansion from 8 × 10⁻⁶/°C to 13 × 10⁻⁶/°C. Standard: test circle (circular material standard). Errors: EB and EU — bidirectional and unidirectional length measurement error (only CMMs capable of 3D spatial measurements), RB and RU — range of three repeated errors; EBZ and EUZ — length perpendicular to the measuring plane, EBXY and EUXY — parallel to it; ESQ — squareness error (combined influence of straightness and squareness of motion of the axis perpendicular to the measuring plane relative to the plane of motion); EBV and EUV — error of a length measured anywhere in the field of view without CMM motion; PF2D — probing error (range of radii of a least-squares fit of points on the test circle, with CMM motion between points, points across the usable field of view), PFV2D — the same without CMM motion.

CONTENT (table of contents). Clause 4 — symbols; clause 5 — environmental and metrological requirements (environmental conditions, operating conditions, requirements for various CMM configurations, length measurement errors). Further clauses are outside the sample, so the test procedure and the number of measurements are not given.

STATUS (as of 19.11.2026). PKN cards: PN-EN ISO 10360-7:2011 “Specyfikacje geometrii wyrobów (GPS) -- Badania odbiorcze i okresowe współrzędnościowych maszyn pomiarowych (CMM) -- Część 7: CMM wyposażone w zespoły głowicy pomiarowej rejestrującej obraz” — English version (published 28-06-2011, 42 pages; corrigendum PN-EN ISO 10360-7:2011/Ap1:2014-09E) and Polish version (published 02-10-2014, 48 pages), KT 48 Fundamentals of Machine Design, implements EN ISO 10360-7:2011 and ISO 10360-7:2011 [IDT]; the cards give no withdrawal date.

Method principle

A calibrated length (bidirectional or unidirectional) is measured with the imaging probing system of the CMM in various positions and directions — also within the field of view without machine motion — and the indication is compared with the calibrated value; the errors of the system itself are determined from the range of radii of a test circle fitted by least squares. Conformance with the requirements is assessed according to the rules of ISO 14253-1.

Applications

Key parameters

ParameterValue
STATUS (as of 19.11.2026)PKN cards: PN-EN ISO 10360-7:2011 “Specyfikacje geometrii wyrobów (GPS) -- Badania odbiorcze i okresowe współrzędnościowych maszyn pomiarowych (CMM) -- Część 7: CMM wyposażone w zespoły głowicy pomiarowej rejestrującej obraz” — English version (published 28-06-2011, 42 pages; corrigendum PN-EN ISO 10360-7:2011/Ap1:2014-09E) and Polish version (published 02-10-2014, 48 pages), KT 48 Fundamentals of Machine Design, implements EN ISO 10360-7:2011 and ISO 10360-7:2011 [IDT]; the cards give no withdrawal date.
Scope (clause 1)Cartesian CMMs with imaging probing systems, discrete-point mode; contact CMMs — ISO 10360-2
Length measurement errors (clause 3)EB, EU (with RB, RU — range of three repetitions); EBZ, EUZ (perpendicular to the measuring plane); EBXY, EUXY (parallel); EBV, EUV (in the field of view without CMM motion)
Squareness error (3.17)ESQ — combined influence of straightness and squareness of motion
Probing errors (3.20, 3.21)PF2D — with CMM motion; PFV2D — without CMM motion; test circle, least-squares fit
Normal CTE material (3.7)8 × 10⁻⁶/°C to 13 × 10⁻⁶/°C

Standard

Standard number
PN-EN ISO 10360-7:2011 (English and Polish version; implements EN ISO 10360-7:2011 and ISO 10360-7:2011 [IDT]) with corrigendum PN-EN ISO 10360-7:2011/Ap1:2014-09
Title (PL)
Specyfikacje geometrii wyrobów (GPS) — Badania odbiorcze i okresowe współrzędnościowych maszyn pomiarowych (CMM) — Część 7: CMM wyposażone w zespoły głowicy pomiarowej rejestrującej obraz
Title (EN)
Geometrical product specifications (GPS) — Acceptance and reverification tests for coordinate measuring machines (CMM) — Part 7: CMMs equipped with imaging probing systems

Step-by-step procedure

  1. Requirements and conditions

    Establishing the requirements (EB, EU, PF2D etc.) assigned by the manufacturer or user and the environmental and operating conditions (clauses 1, 5).

  2. Length measurement errors

    Measurement of calibrated lengths in the positions and directions relevant to the given error — also within the field of view without CMM motion (EBV, EUV) (clause 3).

  3. Probing errors

    Measurement of points on the test circle distributed across the usable field of view, with CMM motion (PF2D) or without motion (PFV2D), least-squares circle fit (3.20, 3.21).

  4. Conformance assessment

    Comparison of the results with the requirements according to the rules of ISO 14253-1 (clauses 1, 2).

Required equipment and apparatus

EquipmentExampleIndicative price
Length standard (clause 3)Calibrated bidirectional or unidirectional test length—
Test circle (3.8)Circular material standard for acceptance and reverification tests of the probing error—

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN ISO 10360-7:2011 (English and Polish version; implements EN ISO 10360-7:2011 and ISO 10360-7:2011 [IDT]) with corrigendum PN-EN ISO 10360-7:2011/Ap1:2014-09 — “Geometrical product specifications (GPS) — Acceptance and reverification tests for coordinate measuring machines (CMM) — Part 7: CMMs equipped with imaging probing systems”.

How does this method work?

A calibrated length (bidirectional or unidirectional) is measured with the imaging probing system of the CMM in various positions and directions — also within the field of view without machine motion — and the indication is compared with the calibrated value; the errors of the system itself are determined from the range of radii of a test circle fitted by least squares. Conformance with the requirements is assessed according to the rules of ISO 14253-1.

What is the measuring range and accuracy?

STATUS (as of 19.11.2026): PKN cards: PN-EN ISO 10360-7:2011 “Specyfikacje geometrii wyrobów (GPS) -- Badania odbiorcze i okresowe współrzędnościowych maszyn pomiarowych (CMM) -- Część 7: CMM wyposażone w zespoły głowicy pomiarowej rejestrującej obraz” — English version (published 28-06-2011, 42 pages; corrigendum PN-EN ISO 10360-7:2011/Ap1:2014-09E) and Polish version (published 02-10-2014, 48 pages), KT 48 Fundamentals of Machine Design, implements EN ISO 10360-7:2011 and ISO 10360-7:2011 [IDT]; the cards give no withdrawal date.; Scope (clause 1): Cartesian CMMs with imaging probing systems, discrete-point mode; contact CMMs — ISO 10360-2; Length measurement errors (clause 3): EB, EU (with RB, RU — range of three repetitions); EBZ, EUZ (perpendicular to the measuring plane); EBXY, EUXY (parallel); EBV, EUV (in the field of view without CMM motion); Squareness error (3.17): ESQ — combined influence of straightness and squareness of motion.

How long does the test take?

The procedure comprises 4 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Length standard (clause 3), Test circle (3.8).

Where is this test used?

Acceptance tests of Cartesian CMMs with imaging (video, vision) probing systems in the discrete-point mode — confirmation of the performance stated by the manufacturer (clause 1); Reverification tests of such machines by the user (clause 1); Determination of length measurement errors EB, EU, EBZ, EUZ, EBXY, EUXY, EBV, EUV and of the squareness error ESQ (clause 3); Determination of probing errors PF2D and PFV2D on a test circle (clause 3).

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 10360-7.