📏 Coordinate measuring machines (CMM) used for measuring linear dimensions — acceptance and reverification tests: length measurement error E₀ and E₁₅₀ (stylus tip offset 0 mm and 150 mm), repeatability range R₀, five calibrated lengths measured three times; PN-EN ISO 10360-2
In short
Acceptance and reverification tests of Cartesian coordinate measuring machines with contacting probing systems operating in the discrete-point probing mode. The test is performed according to PN-EN ISO 10360-2:2010 (Polish version; implements EN ISO 10360-2:2009 and ISO 10360-2:2009 [IDT]); STATUS (PKN card, 14.11.2026): PN-EN ISO 10360-2:2010 (Polish version, published 06-12-2013); implements EN ISO 10360-2:2009 and ISO 10360-2:2009; replaces PN-EN ISO 10360-2:2003. The procedure comprises 5 steps; it is used for: Cartesian CMMs with contacting probing systems of any type in the discrete-point probing mode — acceptance and reverification tests (ISO 10360-2:2009, clause 1), Dual ram CMMs — simplex and duplex mode (3.8–3.10, 6.1.1), By agreement of the parties — non-Cartesian CMMs and CMMs with optical probing (clause 1, 6.1.2).
At a glance
- Standard: PN-EN ISO 10360-2:2010 (Polish version; implements EN ISO 10360-2:2009 and ISO 10360-2:2009 [IDT])
- Category: Metrology
- Procedure steps: 5
- STATUS (PKN card, 14.11.2026): PN-EN ISO 10360-2:2010 (Polish version, published 06-12-2013); implements EN ISO 10360-2:2009 and ISO 10360-2:2009; replaces PN-EN ISO 10360-2:2003
- Characteristics (clause 4): E_L — length measurement error; R₀ — repeatability range; E_L,MPE — maximum permissible error; R₀,MPL — limit of the range
- Stylus tip offset (3.1, 3.4): by default L = 0 mm and L = 150 mm — E₀ and E₁₅₀
Overview
WHAT THE STANDARD COVERS. According to clause 1 of ISO 10360-2:2009 the standard specifies the acceptance tests for verifying that a coordinate measuring machine (CMM) used for measuring linear dimensions performs as stated by the manufacturer, and the reverification tests that enable the user to verify the machine periodically. The tests apply only to Cartesian CMMs using contacting probing systems of any type operating in the discrete-point probing mode; they do not explicitly apply to non-Cartesian machines or machines with optical probing (the parties may apply them by agreement). The standard specifies requirements that can be assigned by the manufacturer or user, the manner of executing the tests, the rules for proving conformance and applications of the tests.
ACCORDING TO THE TEXT OF ISO 10360-2:2009 — PURPOSES, DEFINITIONS, CONDITIONS, PRINCIPLE OF THE TEST (foreword, introduction, clauses 1–6.2). The standard was prepared by ISO/TC 213 (Dimensional and geometrical product specifications and verification); the third edition replaced the second (ISO 10360-2:2001) after a technical revision. Introduction: the standard is a general GPS standard (ISO/TR 14638), it influences link 5 of the chains of standards on size, distance, radius, angle, form, orientation, location, run-out and datums; the tests have three purposes — the error of indication of a calibrated length without stylus tip offset from the ram axis, with a specified offset, and the repeatability of measuring a calibrated length; the result has direct traceability to the metre. Normative references: ISO 10360-1:2000, ISO 14253-1:1998, ISO 14660-1:1999, ISO/TS 23165:2006, ISO/IEC Guide 99 (VIM). Definitions (clause 3): ram axis stylus tip offset L — distance (perpendicular to the ram axis) of the stylus tip from a reference point (specified by the manufacturer, and if not — by the user, close to the probe mounting); coefficient of thermal expansion (CTE) α at 20 °C; normal CTE material — 8 × 10⁻⁶/°C to 13 × 10⁻⁶/°C; length measurement error E_L — error of indication of a calibrated length at offset L, one probing point (or equivalent) at each end (by default L = 0 mm and L = 150 mm, i.e. E₀ and E₁₅₀); repeatability range R₀ — range of three repeated length measurement errors at zero offset; E_L,MPE — maximum permissible length measurement error (MPE — requires calibrated test artefacts); R₀,MPL — maximum permissible limit of the range (MPL — does not require calibrated artefacts); dual ram CMM and the simplex (separate systems) and duplex (one coordinate system) operating modes. Conditions (clause 5): the limits of the environmental conditions (temperature, humidity, vibration) are specified by the manufacturer for the acceptance test and by the user for the reverification test; the user chooses the conditions within these limits and provides the environment of the machine; outside them conformance to E₀,MPE, E_L,MPE and R₀,MPL cannot be required. The machine is operated according to the manufacturer's instructions (warm-up, stylus configuration, cleaning of the tip, probe qualification, thermal stability, mass of the stylus system, temperature sensors). The values of E_L and R₀ in micrometres shall not exceed E_L,MPE and R₀,MPL. Table load (5.5): E₀ shall not exceed E₀,MPE with a load up to the maximum workpiece mass for which the machine is specified; the load is chosen by the user (from zero to rated), fitting in the measuring volume and free-standing, normally approximately centred and symmetrical; the manufacturer may limit the load per unit area (kg/m²) and point load (kg/cm²), and the load at no support point may exceed twice that at another. Tests (6.1): acceptance according to the manufacturer's specifications and procedures — when the user does not supply an artefact, the manufacturer chooses the calibrated test length artefact from Annexes B and D; reverification according to the user's specifications and the manufacturer's procedures; with optical probing by agreement — among others the 2D index (e.g. E₀-2D), magnification and lighting, material and finish of the artefact. Principle (6.2): comparison of the indications of five different calibrated lengths, each measured three times, with their calibrated values; indications from a point-to-point measurement projected onto the alignment direction (Annex C); order of the points at ends A and B: A1 B1, A2 B2, A3 B3 or A1 B1, B2 A2, A3 B3 (not allowed e.g. A1 A2 A3, B1 B2 B3); each of the three measurements has its own, distinct points.
STATUS (PKN card and iTeh catalogue as of 14.11.2026). Card PN-EN ISO 10360-2:2010 (Polish version): title "Specyfikacje geometrii wyrobów (GPS) -- Badania odbiorcze i okresowe współrzędnościowych maszyn pomiarowych (CMM) -- Część 2: CMM stosowane do pomiaru wymiarów liniowych", published 06-12-2013, 36 pages, KT 48 for Fundamentals of Machine Design, "Wprowadza: EN ISO 10360-2:2009 [IDT], ISO 10360-2:2009 [IDT]", "Zastępuje: PN-EN ISO 10360-2:2003 - wersja polska", ICS 17.040.30. ISO 10360-2:2009 in the iTeh catalogue — published, stage 90.93 (standard confirmed); SIST EN ISO 10360-2:2011 replaced SIST EN ISO 10360-2:2002.
HOW MANY LABORATORIES AND IN WHAT FORM (accreditation scope database of the service, read 14.11.2026). ISO 10360-2 appears in 13 records at 10 Polish calibration laboratories, always as "PN-EN ISO 10360-2:2010": AP 099, AP 132, AP 141, AP 186, AP 197, AP 206, AP 214, AP 220, AP 236, AP 239; in AP 186 together with PN-EN ISO 10360-3:2002. Outside Poland — 5 Czech laboratories (6 records).
WHAT THE LABORATORIES CALIBRATE (scope database, complete items). Coordinate measuring machines with a contacting probe — numerically controlled CNC and manual, with or without temperature compensation (AP 099); coordinate measuring systems with a contacting probe, including measuring arms and tracking systems (Lasertracker) (AP 132). Ranges up to 5000 mm (AP 197 — with a laser interferometer), 6000 mm and 24 m (AP 132); CMC given as a formula depending on the length L, e.g. (0,1 + 0,5 · L) μm (AP 186), (0,1 + 0,3 · L) µm (AP 197), Q[0,08; 1,47 · L] μm (AP 099).
WHERE A SEEMINGLY CORRECT RESULT IS EASY TO GET. Test under conditions outside the limits of the specification — then conformance to E_L,MPE and R₀,MPL cannot be required (5.1). Measuring all points of one end and then of the other (A1 A2 A3, B1 B2 B3) or repeating the same points — not allowed (6.2). Test only at zero stylus tip offset when the specification states E₁₅₀ (3.4). Table load not complying with the manufacturer's limitations (5.5).
WHAT WE DO NOT GIVE. Parts read: iTeh sample of ISO 10360-2:2009 in English and French (foreword, introduction, clauses 1–6.2) and the sample of SIST EN ISO 10360-2:2011. We do not give the rest of clause 6 (measuring positions and directions, test procedure, derivation of results), clauses 7–9 (conformance with specification, applications, indication in product documentation) or Annexes A–F — they are outside the sample. We have not read the Polish text of PN-EN ISO 10360-2:2010.
Method principle
The machine measures five different calibrated lengths (artefacts traceable to the metre), each three times, by a point-to-point measurement projected onto the alignment direction of the artefact, with alternating order of the ends (A1 B1, A2 B2, A3 B3) and new points in each repetition. From the differences between the indications and the calibrated values the length measurement errors E₀ and E₁₅₀ (stylus tip offset 0 mm and 150 mm) and the repeatability range R₀ are determined and compared with E_L,MPE and R₀,MPL of the manufacturer or the user.
Applications
- Cartesian CMMs with contacting probing systems of any type in the discrete-point probing mode — acceptance and reverification tests (ISO 10360-2:2009, clause 1)
- Dual ram CMMs — simplex and duplex mode (3.8–3.10, 6.1.1)
- By agreement of the parties — non-Cartesian CMMs and CMMs with optical probing (clause 1, 6.1.2)
- In accreditation scopes: 10 Polish calibration laboratories — CNC and manual CMMs, measuring systems and arms; 5 Czech
Key parameters
| Parameter | Value |
|---|---|
| STATUS (PKN card, 14.11.2026) | PN-EN ISO 10360-2:2010 (Polish version, published 06-12-2013); implements EN ISO 10360-2:2009 and ISO 10360-2:2009; replaces PN-EN ISO 10360-2:2003 |
| Characteristics (clause 4) | E_L — length measurement error; R₀ — repeatability range; E_L,MPE — maximum permissible error; R₀,MPL — limit of the range |
| Stylus tip offset (3.1, 3.4) | by default L = 0 mm and L = 150 mm — E₀ and E₁₅₀ |
| Principle of the test (6.2) | 5 different calibrated lengths × 3 measurements; order A1 B1, A2 B2, A3 B3 or A1 B1, B2 A2, A3 B3; new points in each measurement |
| Normal CTE material (3.3) | α from 8 × 10⁻⁶/°C to 13 × 10⁻⁶/°C (at 20 °C) |
| Who states the limits (5.1, 5.3, 5.4) | manufacturer — acceptance test; user — reverification test; values in µm |
| Table load (5.5) | E₀ ≤ E₀,MPE with a load up to the maximum workpiece mass; load at a point ≤ 2× the load at another point |
| Coverage in accreditation scopes (read 14.11.2026) | 10 Polish calibration laboratories, 13 records; 5 Czech |
Standard
- Standard number
- PN-EN ISO 10360-2:2010 (Polish version; implements EN ISO 10360-2:2009 and ISO 10360-2:2009 [IDT])
- Title (PL)
- Specyfikacje geometrii wyrobów (GPS) — Badania odbiorcze i okresowe współrzędnościowych maszyn pomiarowych (CMM) — Część 2: CMM stosowane do pomiaru wymiarów liniowych
- Title (EN)
- Geometrical product specifications (GPS) — Acceptance and reverification tests for coordinate measuring machines (CMM) — Part 2: CMMs used for measuring linear dimensions
Step-by-step procedure
-
Conditions and specification
Establishing E_L,MPE and R₀,MPL (manufacturer or user) and the limits of the environmental conditions; checking that the environment meets them (5.1–5.4).
-
Preparation of the machine
Warm-up, stylus configuration, cleaning of the tip, probe qualification, thermal stability according to the manufacturer's instructions; table load when it is part of the test (5.2, 5.5).
-
Length measurements
5 calibrated lengths × 3 measurements, point-to-point measurement projected onto the alignment direction; alternating order of the ends, new points in each measurement (6.2).
-
Offset 150 mm
Repetition with the specified stylus tip offset from the ram axis when the specification states E₁₅₀ (3.1, 3.4).
-
Assessment
E_L from the differences between indications and calibrated values, R₀ from the range of three repetitions at L = 0; comparison with MPE and MPL — conformance rules in clause 7 (outside the sample read).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Calibrated test length artefacts (6.1.1, 6.2) | Calibrated, traceable to the metre; choice of artefacts — Annexes B and D (outside the sample read) | — |
| Table load (5.5) | Free-standing, within the measuring volume, normally centred and symmetrical; within the load limits stated by the manufacturer | — |
| Machine temperature sensors (5.2) | Location, type and number according to the manufacturer's instructions | — |
Health and safety (OHS)
- The standard contains no safety warning in the text read
- TECHNICAL NOTE: under environmental conditions outside the specification conformance to E₀,MPE, E_L,MPE and R₀,MPL cannot be required (5.1)
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN ISO 10360-2:2010 (Polish version; implements EN ISO 10360-2:2009 and ISO 10360-2:2009 [IDT]) — “Geometrical product specifications (GPS) — Acceptance and reverification tests for coordinate measuring machines (CMM) — Part 2: CMMs used for measuring linear dimensions”.
How does this method work?
The machine measures five different calibrated lengths (artefacts traceable to the metre), each three times, by a point-to-point measurement projected onto the alignment direction of the artefact, with alternating order of the ends (A1 B1, A2 B2, A3 B3) and new points in each repetition. From the differences between the indications and the calibrated values the length measurement errors E₀ and E₁₅₀ (stylus tip offset 0 mm and 150 mm) and the repeatability range R₀ are determined and compared with E_L,MPE and R₀,MPL of the manufacturer or the user.
What is the measuring range and accuracy?
STATUS (PKN card, 14.11.2026): PN-EN ISO 10360-2:2010 (Polish version, published 06-12-2013); implements EN ISO 10360-2:2009 and ISO 10360-2:2009; replaces PN-EN ISO 10360-2:2003; Characteristics (clause 4): E_L — length measurement error; R₀ — repeatability range; E_L,MPE — maximum permissible error; R₀,MPL — limit of the range; Stylus tip offset (3.1, 3.4): by default L = 0 mm and L = 150 mm — E₀ and E₁₅₀; Principle of the test (6.2): 5 different calibrated lengths × 3 measurements; order A1 B1, A2 B2, A3 B3 or A1 B1, B2 A2, A3 B3; new points in each measurement.
How long does the test take?
The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Calibrated test length artefacts (6.1.1, 6.2), Table load (5.5), Machine temperature sensors (5.2).
Where is this test used?
Cartesian CMMs with contacting probing systems of any type in the discrete-point probing mode — acceptance and reverification tests (ISO 10360-2:2009, clause 1); Dual ram CMMs — simplex and duplex mode (3.8–3.10, 6.1.1); By agreement of the parties — non-Cartesian CMMs and CMMs with optical probing (clause 1, 6.1.2); In accreditation scopes: 10 Polish calibration laboratories — CNC and manual CMMs, measuring systems and arms; 5 Czech.
What safety precautions apply?
The standard contains no safety warning in the text read; TECHNICAL NOTE: under environmental conditions outside the specification conformance to E₀,MPE, E_L,MPE and R₀,MPL cannot be required (5.1).
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 10360-2.