📏 Rockwell hardness testing machines — direct verification (preliminary and total force, depth measuring system, cycle, hysteresis) and indirect verification on reference blocks according to ISO 6508-3: at least five indentations, repeatability and bias for each scale; PN-EN ISO 6508-2

Metrology PN-EN ISO 6508-2

In short

Verification of testing machines for Rockwell hardness testing according to ISO 6508-1, stationary and portable, by two separate methods, together with verification of the indenters. The test is performed according to PN-EN ISO 6508-2:2024-06 (English version; implements EN ISO 6508-2:2023 and ISO 6508-2:2023 [IDT]); STATUS (PKN card, 15.11.2026): PN-EN ISO 6508-2:2024-06 (English version, published 13-06-2024); implements EN ISO 6508-2:2023 and ISO 6508-2:2023; replaces PN-EN ISO 6508-2:2015-04. The procedure comprises 5 steps; it is used for: Rockwell hardness testing machines, stationary and portable — each scale used in testing (ISO 6508-2:2023, clause 1 and 6.2), Regular scales (A to K) and superficial scales (N, T) — different limits for depth measurement and repeatability (5.3, table 2), Verification of Rockwell indenters — part of the scope of the standard (clause 1).

At a glance

  • Standard: PN-EN ISO 6508-2:2024-06 (English version; implements EN ISO 6508-2:2023 and ISO 6508-2:2023 [IDT])
  • Category: Metrology
  • Procedure steps: 5
  • STATUS (PKN card, 15.11.2026): PN-EN ISO 6508-2:2024-06 (English version, published 13-06-2024); implements EN ISO 6508-2:2023 and ISO 6508-2:2023; replaces PN-EN ISO 6508-2:2015-04
  • Temperature (5.1.2, 6.1.2): (23 ± 5) °C; outside this range — stated in the report
  • Preliminary force F₀ (5.2.5): ±2,0 %; range of readings ≤ 1,5 % of F₀; held ≥ 2 s

Overview

WHAT THE STANDARD COVERS. According to clause 1 of ISO 6508-2:2023 the document specifies two separate methods of verification of testing machines for determining Rockwell hardness in accordance with ISO 6508-1 — direct and indirect — together with a method for verifying Rockwell indenters. The direct method determines whether the main parameters of the machine function (applied force, depth measurement, testing cycle timing) are within tolerances, and the indirect method — on reference blocks — how well the machine measures a material of known hardness. The document applies to stationary and portable machines; the tungsten carbide composite ball is considered the standard type of indenter ball. The PKN card of PN-EN ISO 6508-2:2024-06 gives the same scope.

ACCORDING TO THE TEXT OF ISO 6508-2:2023 (publisher's sample in English: foreword, clauses 1–5 and the beginning of clause 6 up to table 2). The standard was prepared by ISO/TC 164 (Mechanical testing of metals), subcommittee SC 3 (Hardness testing), in collaboration with CEN/TC 459 (ECISS) under the Vienna Agreement; the fourth edition replaced the third (ISO 6508-2:2015) after technical revision — requirements were removed from the scope, clause 3 was added, the requirements for calibration and verification of the force and depth measuring systems were modified, a requirement for the hardness of the indenter ball holder was added, the layout of table 8 and the requirements for direct and indirect verification were changed, as well as annex B on uncertainty. Normative references: ISO 376, ISO 6507-1, ISO 6508-1 and ISO 6508-3. General conditions (clause 4): before verification the machine is set up and operating according to the manufacturer's instructions, and the force is applied and removed without shock, vibration or overload. Direct verification (clause 5) covers the test forces, the depth measuring system, the testing cycle and the machine hysteresis test; it is recommended at (23 ± 5) °C (outside this range — stated in the report), with instruments traceable to national standards, and indirect verification follows a successful direct verification. Forces (5.2): each preliminary force F₀ and each total force F is measured, where applicable at not less than three plunger positions (with a closed-loop system — at one), the preliminary force is held for at least 2 s; three readings of each force at each position; an instrument according to ISO 376 class 1 or better, calibrated for reversibility, or balancing against a force accurate to ±0,2 %; the output of the instrument should not vary by more than 0,2 % within 1 s to 30 s after a stepped change in force. Permissible relative error: preliminary force F₀ — ±2,0 % (before application and after removal of the additional force F₁), range of readings at a position ≤ 1,5 % of F₀; total force F — ±1,0 %, range ≤ 0,75 % of F. Depth measuring system (5.3): calibrated by known incremental movements of the indenter or holder, with an instrument or gauge blocks of maximum expanded uncertainty 0,000 3 mm (95 % confidence level), at not less than four evenly spaced intervals covering the full working depth, in three cycles of readings (the maximum depth ranges from 0,25 mm for scale B to 0,04 mm for scale 15N); long-stroke systems are verified at the top, middle and bottom of the stroke (at least four intervals of about 0,05 mm each) and continuity of indication over the full travel is confirmed. The indication shall be correct within ±0,001 mm for scales A to K and ±0,000 5 mm for scales N and T (±0,5 of a scale unit). The cycle (5.4) is calibrated by the manufacturer at manufacture and after a repair that may have affected it; it shall conform to ISO 6508-1; with automatic control the uncertainty (k = 2) of the timing instrument shall not exceed 0,2 s, and with manual control it is checked that the machine can achieve the required cycle. Hysteresis (5.5): repeated tests with the scale of the highest force used in testing, with a spherical indenter of at least 10 mm diameter bearing directly on the specimen holder or through a spacer (or with a parallel block instead of a blunt indenter), material of hardness at least 60 HRC; at most 10 measurements, the average of the last three shall be (130 ± 1,0) units for the ball scales B, E, F, G, H and K and (100 ± 1,0) units for the other scales.

INDIRECT VERIFICATION (clause 6) is recommended at (23 ± 5) °C on reference blocks calibrated according to ISO 6508-3, for each scale the machine will use — with blocks from each hardness range of table 1 (e.g. scale C: 10 to 30 HRC, 35 to 55 HRC, 60 to 70 HRC; scale A: 20 to 40, 45 to 75 and 80 to 95 HRA; scale B: 10 to 50, 60 to 80 and 85 to 100 HRBW), chosen to approximate the limits of intended use; the same cycle as in the calibration of the blocks is recommended, and only the calibrated surfaces are used. On each block at least five indentations uniformly distributed, read to a resolution not worse than 0,2 HR units; before them at least two preliminary indentations whose results are ignored. Repeatability range r = H_n − H₁ and mean H̄; permissible bias b and repeatability r (table 2), e.g.: scale C (10 to 70 HRC) — b ±1,5 HRC, r ≤ 0,02 (100 − H̄) or 0,8 HRC (whichever is greater); scale A — b ±2 HRA (20 to 75 HRA) and ±1,5 HRA (> 75 to 95 HRA), r as for C; scale B — b ±4, ±3 and ±2 HRBW (10 to 45, > 45 to 80, > 80 to 100 HRBW), r ≤ 0,04 (130 − H̄); scales 15N, 30N, 45N — b ±2 HRN, r ≤ 0,04 (100 − H̄) or 1,2 HRN; scales 15T, 30T, 45T — b ±3 HRTW, r ≤ 0,06 (100 − H̄) or 2,4 HRTW. According to the note to table 2 the requirements of ASTM E18 may be different. The sample does not include the further parts of the document (verification of indenters, intervals between verifications, report, annexes) — we do not give them here.

STATUS (PKN card, read 15.11.2026). PN-EN ISO 6508-2:2024-06 — English version, Polish title “Metale -- Pomiar twardości sposobem Rockwella -- Część 2: Sprawdzanie i wzorcowanie twardościomierzy i wgłębników”, published 13-06-2024, 34 pages, KT 123 for Testing of Properties of Metals, implements EN ISO 6508-2:2023 and ISO 6508-2:2023 [IDT], replaces PN-EN ISO 6508-2:2015-04 (English version); the card gives no withdrawal date. iTeh catalogue (15.11.2026): ISO 6508-2:2023 — published 13.12.2023, stage “International Standard published” (60.60).

IN ACCREDITATION SCOPES (scope database of the service, read 15.11.2026, after loading the scopes of this batch). PN-EN ISO 6508-2 appears at 8 Polish calibration laboratories (1 record each) — six in the 2024-06 edition, one as “2024”, and one cites the 2015-04 edition as the basis of an internal procedure; at four alongside ASTM E18-25 — and at 4 Czech laboratories (5 records).

Method principle

The machine is verified by two separate methods. Directly — by measuring the preliminary (±2,0 %) and total (±1,0 %) force with an instrument according to ISO 376 class 1 or better, the depth measuring system by known movements (±0,001 mm for scales A to K, ±0,000 5 mm for scales N and T), the cycle and the hysteresis (average of the last three measurements 130 ± 1,0 or 100 ± 1,0 units). Indirectly — by making at least five indentations on blocks from each hardness range of the scale and comparing the repeatability r and the bias b with table 2.

Applications

Key parameters

ParameterValue
STATUS (PKN card, 15.11.2026)PN-EN ISO 6508-2:2024-06 (English version, published 13-06-2024); implements EN ISO 6508-2:2023 and ISO 6508-2:2023; replaces PN-EN ISO 6508-2:2015-04
Temperature (5.1.2, 6.1.2)(23 ± 5) °C; outside this range — stated in the report
Preliminary force F₀ (5.2.5)±2,0 %; range of readings ≤ 1,5 % of F₀; held ≥ 2 s
Total force F (5.2.6)±1,0 %; range ≤ 0,75 % of F; 3 readings at each position
Depth measuring system (5.3)standard ≤ 0,000 3 mm (95 %); ≥ 4 intervals × 3 cycles; ±0,001 mm (A–K), ±0,000 5 mm (N, T)
Cycle (5.4)timing uncertainty (k = 2) ≤ 0,2 s; limits of ISO 6508-1
Hysteresis (5.5)ball ≥ 10 mm, ≥ 60 HRC; ≤ 10 measurements, average of the last 3: (130 ± 1,0) for B, E, F, G, H, K; (100 ± 1,0) for the other scales
Indirect verification (6.2)blocks from each range of table 1; ≥ 5 indentations; reading ≤ 0,2 HR; ≥ 2 preliminary indentations ignored
Bias and repeatability — scale C (table 2)b ±1,5 HRC; r ≤ 0,02 (100 − H̄) or 0,8 HRC
Scales N and T (table 2)N: b ±2 HRN, r ≤ 0,04 (100 − H̄) or 1,2 HRN; T: b ±3 HRTW, r ≤ 0,06 (100 − H̄) or 2,4 HRTW
Coverage in accreditation scopes (read 15.11.2026)8 Polish calibration laboratories (8 records); 4 Czech (5 records)

Standard

Standard number
PN-EN ISO 6508-2:2024-06 (English version; implements EN ISO 6508-2:2023 and ISO 6508-2:2023 [IDT])
Title (PL)
Metale — Pomiar twardości sposobem Rockwella — Część 2: Sprawdzanie i wzorcowanie twardościomierzy i wgłębników
Title (EN)
Metallic materials — Rockwell hardness test — Part 2: Verification and calibration of testing machines and indenters

Step-by-step procedure

  1. General conditions

    Set-up and operation of the machine according to the manufacturer's instructions; application of the force without shock, vibration or overload (clause 4).

  2. Test forces

    Each preliminary and total force, three readings at each plunger position; error and range against 5.2.5–5.2.6.

  3. Depth measuring system

    Known movements at not less than four intervals, three cycles; long-stroke systems — three locations and continuity; limits ±0,001 mm and ±0,000 5 mm (5.3).

  4. Cycle and hysteresis

    Cycle in accordance with ISO 6508-1; hysteresis test with a spherical indenter — average of the last three measurements (5.4–5.5).

  5. Indirect verification

    For each scale blocks from the ranges of table 1; two preliminary indentations, then at least five; r and b against table 2 (clause 6).

Required equipment and apparatus

EquipmentExampleIndicative price
Force-measuring instrument (5.2.3)According to ISO 376, class 1 or better, calibrated for reversibility; or balancing against a force accurate to ±0,2 %—
Instrument or gauge blocks for the depth measuring system (5.3.2)Expanded uncertainty ≤ 0,000 3 mm (95 %)—
Spherical indenter and spacer for the hysteresis test (5.5.1)Ball of diameter ≥ 10 mm or a parallel block; material of hardness ≥ 60 HRC—
Reference blocks (6.1.2)Calibrated according to ISO 6508-3, from each hardness range of table 1 for the scale—

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN ISO 6508-2:2024-06 (English version; implements EN ISO 6508-2:2023 and ISO 6508-2:2023 [IDT]) — “Metallic materials — Rockwell hardness test — Part 2: Verification and calibration of testing machines and indenters”.

How does this method work?

The machine is verified by two separate methods. Directly — by measuring the preliminary (±2,0 %) and total (±1,0 %) force with an instrument according to ISO 376 class 1 or better, the depth measuring system by known movements (±0,001 mm for scales A to K, ±0,000 5 mm for scales N and T), the cycle and the hysteresis (average of the last three measurements 130 ± 1,0 or 100 ± 1,0 units).

What is the measuring range and accuracy?

STATUS (PKN card, 15.11.2026): PN-EN ISO 6508-2:2024-06 (English version, published 13-06-2024); implements EN ISO 6508-2:2023 and ISO 6508-2:2023; replaces PN-EN ISO 6508-2:2015-04; Temperature (5.1.2, 6.1.2): (23 ± 5) °C; outside this range — stated in the report; Preliminary force F₀ (5.2.5): ±2,0 %; range of readings ≤ 1,5 % of F₀; held ≥ 2 s; Total force F (5.2.6): ±1,0 %; range ≤ 0,75 % of F; 3 readings at each position.

How long does the test take?

The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Force-measuring instrument (5.2.3), Instrument or gauge blocks for the depth measuring system (5.3.2), Spherical indenter and spacer for the hysteresis test (5.5.1), Reference blocks (6.1.2).

Where is this test used?

Rockwell hardness testing machines, stationary and portable — each scale used in testing (ISO 6508-2:2023, clause 1 and 6.2); Regular scales (A to K) and superficial scales (N, T) — different limits for depth measurement and repeatability (5.3, table 2); Verification of Rockwell indenters — part of the scope of the standard (clause 1); In accreditation scopes: 8 Polish calibration laboratories and 4 Czech.

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 6508-2.

🔍 Find a laboratory performing this test