📏 Vernier, dial and digital callipers — design and metrological characteristics: partial surface contact error E and shift error S with MPE values, conditions at 20 °C, reference point, testing with gauge blocks; PN-EN ISO 13385-1
In short
The most important design and metrological characteristics of callipers with analogue indication (vernier scale or dial) and with digital indication. The test is performed according to PN-EN ISO 13385-1:2019-12 (English version; implements EN ISO 13385-1:2019 and ISO 13385-1:2019 [IDT]); Typical jaws, range up to 150 mm (table 1): fig. 1/2: l_a 40 mm, l_b 8–20 mm; fig. 3: l_a 50 mm, l_b 8–10 mm, l_c 25–35 mm; l_d 5 mm. The procedure comprises 5 steps; it is used for: Callipers with vernier scale, with dial (circular scale) and with digital display — for external, internal, depth and step measurements (ISO 13385-1:2019, clauses 1 and 3), Acceptance testing and verification of conformity with specifications — choice of test points, standards of appropriate uncertainty (5.4, clause 6), Calibration of callipers — guidelines in annex A, default MPE in annex B.
At a glance
- Standard: PN-EN ISO 13385-1:2019-12 (English version; implements EN ISO 13385-1:2019 and ISO 13385-1:2019 [IDT])
- Category: Metrology
- Procedure steps: 5
- STATUS (PKN card, 16.11.2026): PN-EN ISO 13385-1:2019-12 (English version, published 19-12-2019); implements EN ISO 13385-1:2019 and ISO 13385-1:2019; replaces PN-EN ISO 13385-1:2011
- Types of indication (clause 1): analogue — vernier or circular scale (dial); digital — display
- Scale interval of the main scale (4.1): vernier — 1 mm; circular scale — 1 mm or 2 mm
Overview
WHAT THE STANDARD COVERS. According to clause 1 of ISO 13385-1:2019 the document provides the most important design and metrological characteristics of callipers with analogue indication (vernier scale or circular scale — dial) and with digital indication (digital display). The PKN card of PN-EN ISO 13385-1:2019-12 gives the same scope.
ACCORDING TO THE TEXT OF ISO 13385-1:2019 (publisher's sample in English: foreword, introduction, clauses 1–5.4 and the beginning of 5.5). The standard was prepared by ISO/TC 213 (Dimensional and geometrical product specifications and verification); the second edition replaced the first (ISO 13385-1:2011) after technical revision: figures were updated, general design characteristics were removed in favour of a reference to ISO 14978:2018, metrological characteristics were clarified and modified, requirements for test methods and default values of maximum permissible errors (MPE) were added. It is a general GPS standard influencing chain links F and G of the chain of standards on size and distance; the rules of ISO 8015 and the default decision rules of ISO 14253-1 apply (alternatives — ISO/TR 14253-6). Normative references: ISO 14253-1, ISO 14253-5, ISO/TR 14253-6, ISO 14978:2018, ISO/IEC Guide 98-3 (GUM) and ISO/IEC Guide 99 (VIM). Definitions (clause 3): calliper — measuring instrument which evaluates an internal or external dimension by opposing contact of the measuring jaw of a slider, moving relative to a scale on a rigid beam, and a fixed jaw (it may additionally have a depth measuring rod and faces for step measurement; analogue or digital indication); contact of the measuring face with an integral feature of a workpiece — full, partial and line contact.
DESIGN (clause 4). Design according to the general guidelines of ISO 14978, including its annex C; examples — a calliper with digital display and a vernier calliper for external, internal, depth and step measurement, and a vernier calliper with a fine adjustment device and knife edges for external measurement. Scale interval of the main scale on the beam: with a vernier — 1 mm, with a circular scale — 1 mm or 2 mm. The manufacturer states important design dimensions, e.g. those of table 1 (typical values, not requirements); with a depth rod — its cross-section dimensions, e.g. ∅ 1,5 mm or 1,2 mm × 3 mm. Typical jaw lengths (table 1): range up to 150 mm — l_a 40 mm, l_b 8–20 mm (fig. 1 and 2), l_a 50 mm, l_b 8–10 mm, l_c 25–35 mm (fig. 3), width of the faces for internal measurement l_d 5 mm; up to 300 mm — l_a 60–65 mm, l_b 10–30 mm; up to 1 000 mm — l_a 100–130 mm, l_b 20–60 mm, l_d 10–20 mm; the measuring range refers to external measurements.
METROLOGICAL CHARACTERISTICS (clause 5). The characteristics and MPE values apply to any indication permitted by the manufacturer when used according to the manufacturer's recommendations; the MPE cannot be smaller than the digital step or the scale interval of the circular or vernier scale. The rated operating conditions are stated by the manufacturer; MPE values apply at exactly 20 °C unless otherwise stated — test values are corrected to 20 °C, and without correction its consequences are included in the measurement uncertainty (6.2); a calliper is a manually operated instrument, so the user — reasonably skilled — is part of the measuring system. Reference point (5.3): callipers with a digital display or a circular scale have an adjustable zero point (digital — in any position within the range, dial — within the range of the circular scale), and the characteristics apply with zero set on the external measuring faces in contact; in callipers without an adjustable zero, the error when these faces are brought into contact is included in the characteristics without correction. Test methods (5.4): errors of indication are tested with suitable instruments or standards of appropriate uncertainty, e.g. gauge blocks according to ISO 3650, step gauges or setting ring gauges, sufficient to establish confidence in the results; for acceptance testing the customer chooses the test points, and the points are chosen with regard to cyclic and local errors — with vernier and circular scales they shall cover the range of that scale (with a dial — various pointer angles). The partial surface contact error E (limited by E_MPE, 5.5) is the error of indication when partial contact of the external measuring faces is used to measure a standard. Further subclauses according to the table of contents: shift error S (limited by S_MPE) for internal faces, crossed knife-edge faces and depth or step faces (5.6), MPE values (5.7), special cases (5.8), conformity — uncertainty and decision rule (clause 6), marking (clause 7), calibration guidelines (annex A) and default MPE values (annex B, normative); the sample does not contain their text, so MPE values are not given.
STATUS (PKN card, read 16.11.2026). PN-EN ISO 13385-1:2019-12 — English version, Polish title “Specyfikacje geometrii wyrobów (GPS) -- Sprzęt pomiarowy do wymiarów -- Część 1: Charakterystyki konstrukcyjne i metrologiczne suwmiarek”, published 19-12-2019, 28 pages, KT 48 for Fundamentals of Machine Design, implements EN ISO 13385-1:2019 and ISO 13385-1:2019 [IDT], replaces PN-EN ISO 13385-1:2011 (English and German versions); the card gives no withdrawal date. iTeh catalogue (16.11.2026): ISO 13385-1:2019 — published 21-08-2019, stage “International Standard confirmed” (90.93); EN ISO 13385-1:2019 — published 18-09-2019, replaced EN ISO 13385-1:2011.
IN ACCREDITATION SCOPES (scope database of the service, read 16.11.2026, after loading the scopes of this batch). PN-EN ISO 13385-1 appears at 8 Polish calibration laboratories (1 record each) — in the 2019 or 2019-12 edition; it does not appear in Czech scopes.
Method principle
A calliper is assessed by its errors of indication when measuring standards of known length (e.g. gauge blocks according to ISO 3650): the partial surface contact error E of the external measuring faces and the shift error S for internal faces, crossed knife edges and depth or step faces, compared with MPE values. The errors are referred to 20 °C and to the reference point set by zeroing on the external measuring faces in contact; the test points cover the range of the vernier or circular scale.
Applications
- Callipers with vernier scale, with dial (circular scale) and with digital display — for external, internal, depth and step measurements (ISO 13385-1:2019, clauses 1 and 3)
- Acceptance testing and verification of conformity with specifications — choice of test points, standards of appropriate uncertainty (5.4, clause 6)
- Calibration of callipers — guidelines in annex A, default MPE in annex B
- In accreditation scopes: 8 Polish calibration laboratories
Key parameters
| Parameter | Value |
|---|---|
| STATUS (PKN card, 16.11.2026) | PN-EN ISO 13385-1:2019-12 (English version, published 19-12-2019); implements EN ISO 13385-1:2019 and ISO 13385-1:2019; replaces PN-EN ISO 13385-1:2011 |
| Types of indication (clause 1) | analogue — vernier or circular scale (dial); digital — display |
| Scale interval of the main scale (4.1) | vernier — 1 mm; circular scale — 1 mm or 2 mm |
| Typical jaws, range up to 150 mm (table 1) | fig. 1/2: l_a 40 mm, l_b 8–20 mm; fig. 3: l_a 50 mm, l_b 8–10 mm, l_c 25–35 mm; l_d 5 mm |
| Depth rod (4.2) | manufacturer states the cross section, e.g. ∅ 1,5 mm or 1,2 mm × 3 mm |
| Rated operating conditions (5.2) | MPE at exactly 20 °C; without correction to 20 °C — consequences in the uncertainty (6.2) |
| MPE and scale interval (5.1) | MPE not smaller than the digital step or the scale interval of the circular or vernier scale |
| Reference point (5.3) | digital — zero setting in any position; dial — within the circular scale; without adjustable zero — contact error included in the characteristics |
| Testing (5.4) | gauge blocks according to ISO 3650, step gauges, setting ring gauges; points covering the range of the vernier / circular scale |
| Errors (5.5–5.6) | E — partial contact of external faces (E_MPE); S — shift: internal, crossed knife edges, depth or steps (S_MPE) |
| Coverage in accreditation scopes (read 16.11.2026) | 8 Polish calibration laboratories (8 records) |
Standard
- Standard number
- PN-EN ISO 13385-1:2019-12 (English version; implements EN ISO 13385-1:2019 and ISO 13385-1:2019 [IDT])
- Title (PL)
- Specyfikacje geometrii wyrobów (GPS) — Sprzęt pomiarowy do wymiarów — Część 1: Charakterystyki konstrukcyjne i metrologiczne suwmiarek
- Title (EN)
- Geometrical product specifications (GPS) — Dimensional measuring equipment — Part 1: Design and metrological characteristics of callipers
Step-by-step procedure
-
Manufacturer data
Determination of the type of indication, scale interval, rated operating conditions and important design dimensions (4.1–4.2, 5.2).
-
Reference point
Zero setting on the external measuring faces in contact; in callipers without adjustable zero — the contact error included without correction (5.3).
-
Choice of test points
Points chosen with regard to cyclic and local errors, covering the range of the vernier or circular scale (various pointer angles) (5.4).
-
Errors E and S
Measurement of standards (gauge blocks according to ISO 3650, step gauges, rings) — partial surface contact error E and shift error S; correction to 20 °C or inclusion in the uncertainty (5.2, 5.5–5.6).
-
Conformity
Comparison with MPE values taking into account the uncertainty and the decision rule (5.7, clause 6).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Gauge blocks (5.4) | According to ISO 3650 — standards for testing the errors of indication | — |
| Step gauges and setting ring gauges (5.4) | Standards of appropriate measurement uncertainty, among others for internal faces and steps | — |
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN ISO 13385-1:2019-12 (English version; implements EN ISO 13385-1:2019 and ISO 13385-1:2019 [IDT]) — “Geometrical product specifications (GPS) — Dimensional measuring equipment — Part 1: Design and metrological characteristics of callipers”.
How does this method work?
A calliper is assessed by its errors of indication when measuring standards of known length (e.g.
What is the measuring range and accuracy?
STATUS (PKN card, 16.11.2026): PN-EN ISO 13385-1:2019-12 (English version, published 19-12-2019); implements EN ISO 13385-1:2019 and ISO 13385-1:2019; replaces PN-EN ISO 13385-1:2011; Types of indication (clause 1): analogue — vernier or circular scale (dial); digital — display; Scale interval of the main scale (4.1): vernier — 1 mm; circular scale — 1 mm or 2 mm; Typical jaws, range up to 150 mm (table 1): fig. 1/2: l_a 40 mm, l_b 8–20 mm; fig. 3: l_a 50 mm, l_b 8–10 mm, l_c 25–35 mm; l_d 5 mm.
How long does the test take?
The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Gauge blocks (5.4), Step gauges and setting ring gauges (5.4).
Where is this test used?
Callipers with vernier scale, with dial (circular scale) and with digital display — for external, internal, depth and step measurements (ISO 13385-1:2019, clauses 1 and 3); Acceptance testing and verification of conformity with specifications — choice of test points, standards of appropriate uncertainty (5.4, clause 6); Calibration of callipers — guidelines in annex A, default MPE in annex B; In accreditation scopes: 8 Polish calibration laboratories.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 13385-1.