🌿 SO₂ in air (UV fluorescence)

Environmental PN-EN 14212

Continuous automatic measurement of sulfur dioxide (SO₂) concentration in ambient air by ultraviolet fluorescence method. Reference air quality monitoring method.

Overview

Sulfur dioxide (SO₂) is a gaseous air pollutant formed mainly from combustion of sulfur-containing fossil fuels (coal, fuel oil). SO₂ is toxic to the respiratory system, causes acid rain and smog.

Standard PN-EN 14212 defines the reference method for continuous SO₂ measurement in ambient air, based on UV fluorescence. The method is used in the National Environmental Monitoring (PMŚ) network managed by GIOŚ/WIOŚ. Permissible SO₂ concentrations: 350 µg/m³ (1h), 125 µg/m³ (24h) according to Directive 2008/50/EC and Polish regulation.

Method principle

The air sample is drawn into the analyzer measurement chamber, where it is irradiated with UV radiation at ~214 nm wavelength. SO₂ molecules absorb UV radiation and transition to an excited state, from which they return to the ground state emitting fluorescence at longer wavelength (~330 nm). Fluorescence intensity is proportional to SO₂ concentration and measured by photomultiplier. Interferences (e.g., H₂O, aromatic hydrocarbons) are eliminated by selective optical filter and Teflon membrane.

Applications

Key parameters

ParameterValue
Measurement range0–500 µg/m³ (typical), expandable to 2000 µg/m³
Detection limit0.5–2 µg/m³
Response time (T90)<120 s
1h limit350 µg/m³ (not to exceed >24 times/year)
24h limit125 µg/m³ (not to exceed >3 times/year)

Standard

Standard number
PN-EN 14212:2013
Title (PL)
Powietrze atmosferyczne — Standardowa metoda pomiaru stężenia dwutlenku siarki metodą fluorescencji w ultrafiolecie
Title (EN)
Ambient air — Standard method for the measurement of the concentration of sulphur dioxide by ultraviolet fluorescence

Step-by-step procedure

1. Analyzer installation

Place analyzer in monitoring station with air conditioning (20±5°C). Connect inlet probe and Teflon line.

⏱ Time: 2 h

2. Startup and stabilization

Turn on analyzer. Heat UV lamp (30–60 min). Check sample flow (0.5–1 L/min).

⏱ Time: 1 h

3. Zero calibration

Supply zero air. Set analyzer zero point. Deviation <2 µg/m³.

⏱ Time: 15 min

4. Span calibration

Supply SO₂ calibration gas of known concentration (e.g., 200 µg/m³). Set span point. Deviation <5%.

⏱ Time: 15 min

5. Continuous measurement

Start automatic data recording (1-minute and hourly averages). Data transmission to GIOŚ database.

⏱ Time: continuous

6. Quality control

Zero and span calibration every 2 weeks. Service every 6 months. UV lamp replacement every 12–18 months.

⏱ Time: periodic

Required equipment and apparatus

EquipmentExampleIndicative price
SO₂ analyzer (UV fluorescence)Thermo 43i, Horiba APSA-370, Teledyne T10030,000–60,000 PLN
Gas calibrator (permeation or cylinders)Thermo 146i, Environics 610025,000–60,000 PLN
Zero air generator (synthetic air)Thermo 111, carbon filter + dryer + HEPA filter10,000–25,000 PLN
Sample intake system with probePM₁₀ probe with filter and Teflon connection2,000–5,000 PLN

Reagents, media and consumables

ReagentCASDetails
Calibration gas SO₂7446-09-5Cylinder SO₂/N₂, 50–500 ppb, certified ±2%
Zero airSynthetic air free of SO₂ (<0.5 ppb), from generator or cylinder

Health and safety (OHS)