⚗️ Selenium in drinking water, groundwater and surface water — hydride generation atomic absorption spectrometry (HGAAS): digestion with sulfuric acid and hydrogen peroxide, reduction to Se(IV) with hydrochloric acid, hydride with sodium tetrahydroborate, absorbance at 196.0 nm, range 1–10 µg/l; PN-ISO 9965:2001
In short
Polish standard of 2001, identical to ISO 9965:1993, with a method for determining selenium — including organically bound selenium — in drinking water, groundwater and surface water by hydride generation atomic absorption spectrometry in the range from 1 to 10 µg/l. The test is performed according to PN-ISO 9965:2001; Concentration range (clause 1): 1–10 µg/l; higher concentrations after dilution of the sample. The procedure comprises 6 steps; it is used for: Drinking water, groundwater and surface water — selenium, including organically bound selenium, 1–10 µg/l (ISO 9965:1993, clause 1; PKN card), Water intended for human consumption — comparison with the parametric value of 20 µg/l (30 µg/l in regions with high selenium in groundwater), Dz.U. 2026 item 748.
At a glance
- Standard: PN-ISO 9965:2001
- Category: Physicochemistry
- Procedure steps: 6
- STATUS (as of 18.11.2026): PN-ISO 9965:2001 without a withdrawal note in PKN; ISO 9965:1993 withdrawn by ISO on 15.07.2014 ("Revised by: ISO/TS 17379-2:2013", Standard Norge); SIST ISO 9965:1998 replaced by SIST-TS ISO/TS 17379-2:2015 (iTeh); no Polish edition of ISO/TS 17379-2
- Edition (from the PKN card): PN-ISO 9965:2001, Polish version; published 13-12-2001, 8 pages, KT 121, ICS 13.060.50; introduces ISO 9965:1993 [IDT]
- Concentration range (clause 1): 1–10 µg/l; higher concentrations after dilution of the sample
Overview
STATUS (as of 18.11.2026). The card of PN-ISO 9965:2001 (Polish version) has no "Withdrawn" header and no "Withdrawal date" row: published 13-12-2001, 8 pages, KT 121 Water Quality — Chemical Tests — Inorganic Substances, "Introduces: ISO 9965:1993 [IDT]", ICS 13.060.50; there are no "Replaces" or "Replaced by" fields. The introduced document is no longer current with its publisher: according to the Standard Norge catalogue ISO 9965:1993 has the status "Withdrawn" dated 15 July 2014 and the note "Revised by: ISO/TS 17379-2:2013", and in the iTeh catalogue the Slovenian edition SIST ISO 9965:1998 is withdrawn (29.01.2015) and replaced by SIST-TS ISO/TS 17379-2:2015 "Water quality — Determination of selenium — Part 2: Method using hydride generation atomic absorption spectrometry (HG-AAS)". The PKN search for "17379" returns no standard — there is no Polish edition of this specification. The description below follows the text of ISO 9965:1993, which PN-ISO 9965:2001 introduces unchanged (the publisher's sample in the iTeh catalogue covers the whole standard); how ISO/TS 17379-2 differs we have not established — we did not read its text.
WHAT THE STANDARD COVERS. Clause 1 of ISO 9965 and the "Scope" field of the PKN card: determination of selenium, including organically bound selenium, in drinking water, groundwater and surface water in the concentration range from 1 µg/l to 10 µg/l; higher concentrations are determined after suitable dilution of the sample. Principle (clause 3): selenium formed by thermal decomposition of selenium hydride is measured by atomic absorption. Under the conditions of the method only Se(IV) is converted quantitatively to the hydride, so selenium in other oxidation states is converted to Se(IV) before the determination. Se(IV) is reduced to gaseous selenium dihydride (SeH₂) with sodium tetrahydroborate in hydrochloric acid medium; the absorbance is measured at a wavelength of 196.0 nm.
REAGENTS, APPARATUS AND SAMPLES (clauses 4–6). Reagents of recognized analytical grade; the selenium content of the water and reagents shall be negligibly low compared with the lowest concentration to be determined. Sulfuric acid ρ = 1.84 g/ml, hydrochloric acid ρ = 1.16 g/ml, hydrogen peroxide 30 % (m/m), sodium hydroxide. Sodium tetrahydroborate solution: 1 g of sodium hydroxide is dissolved in about 20 ml of water, 3 g of NaBH₄ are added and the solution is diluted with water to 100 ml — prepared daily. Selenium stock solution 1000 mg/l: 1.4053 g of selenium dioxide and 2 g of sodium hydroxide in a 1000 ml flask (commercial solutions are acceptable); standard solution 1 — 10 mg/l and standard solution 2 — 0.1 mg/l (each from 10 ml of the previous solution and 20 ml of hydrochloric acid in a 1000 ml flask, stable for at least one week). Apparatus: atomic absorption spectrometer with a hydride system and a radiation source for the determination of selenium (for example an electrodeless discharge lamp or a hollow cathode lamp), with background correction if appropriate; argon or nitrogen; glassware cleaned immediately before use with warm dilute nitric acid (for example 2 mol/l) and rinsed with water. Samples are taken according to ISO 5667-1 and ISO 5667-2 into polyethylene or borosilicate glass containers cleaned with nitric acid; 20 ml of hydrochloric acid are added to each 1000 ml of sample and, if the pH is above 2, more acid until the pH is 2 or less; preservation according to ISO 5667-3.
PROCEDURE (clauses 7–9). Blank: 2 ml of hydrochloric acid in a 100 ml flask, made up with water and treated exactly like the sample. At least five calibration solutions from standard solution 2 covering the expected working range — for example for 1–10 µg/l: 1, 3, 5, 8 and 10 ml of this solution and 2 ml of hydrochloric acid in 100 ml flasks (1, 3, 5, 8 and 10 µg/l), prepared daily. For total selenium the sample is digested to decompose organic selenium compounds; digestion may be omitted if experience has shown that selenium is recovered quantitatively without it. 50 ml of sample in a round-bottomed flask with 5 ml of sulfuric acid and 5 ml of hydrogen peroxide are heated to boiling in an apparatus with a reflux condenser and condensate reservoir (example — Figure 1 of the standard) until turbid fumes of sulfuric acid appear; if the solution is turbid and almost colourless, after cooling another 5 ml of hydrogen peroxide are added and heating continues; the sample must never be evaporated to dryness; after cooling the condensate is returned to the flask. Reduction of Se(VI) to Se(IV): 20 ml of hydrochloric acid and gentle boiling under reflux for 15 min with the cock open (without digestion, if the sample contains free chlorine, the solution is purged with nitrogen at about 1 l/min for the same time); after cooling the solution is transferred quantitatively to a 100 ml flask and made up with water; the blank and calibration solutions are treated in the same way. The spectrometer is set up according to the manufacturer's instructions (196.0 nm) and the absorption cell positioned for maximum transmission. Order of measurement: blank, calibration solutions, samples. For example, 20 ml of the reduced solution in the reaction vessel are purged with argon or nitrogen until the signal returns to zero, about 5 ml of sodium tetrahydroborate solution are added and the signal is recorded; depending on the hydride system the volumes may be larger or smaller, but the defined proportions of quantities must be maintained. The measurement is repeated on separate portions of each solution. Notes of the standard: it is good practice to check the blank and calibration points from time to time; with unknown samples it is good practice to add a known amount of selenium to at least one sample and, if recovery is unsatisfactory, to use the standard addition method. The selenium concentration in µg/l is obtained from the absorbance and the calibration function, taking all dilutions into account; the result is given to two significant figures (examples of the standard: 8 µg/l, 32 µg/l).
PRECISION, INTERFERENCES AND TEST REPORT (clauses 10–12). Interlaboratory trial of 1992 with an almost identical method (19 laboratories): drinking water — true value 3.0 µg/l, mean 2.92 µg/l, recovery 97.4 %; waste water — true value 9.0 µg/l, mean 7.76 µg/l, recovery 86.2 % (Table 1). Table 2 gives the effect of 22 other substances (among them nitrate, iodide, chromium(III), iron(III), copper(II), cadmium, mercury(II), tin(II), lead(II), antimony and bismuth) on the selenium result. The test report contains a reference to the standard, identification of the sample, pretreatment of the sample, the result according to clause 9 and operations not covered by the standard.
VALUES IN THE REGULATION. Regulation of the Minister of Health of 22 May 2026 on the quality of water intended for human consumption (Dz.U. 2026 item 748; status "in force" in the ELI database of the Sejm, 18.11.2026). Annex 1, part B, item 28: selenium — parametric value 20 μg/l, with the remark: "The parametric value of 30 μg/l applies to regions where geological conditions may lead to high levels of selenium in groundwater". Annex 4, part B: the method has to be able to measure concentrations equal to the parametric value with a limit of quantification of 30 % or less of the parametric value (for 20 μg/l — 6 μg/l), and the measurement uncertainty for selenium (item 41) is 40 % of the parametric value. The regulation does not name any standard for selenium.
WHERE A SEEMINGLY CORRECT RESULT IS EASY TO GET. First — oxidation state: only Se(IV) forms the hydride quantitatively; without reduction with hydrochloric acid selenium(VI) is not included in the result, and without digestion organically bound selenium is not, so the result is too low although the measurement runs correctly; digestion may be omitted only where quantitative recovery has been shown. Second — range: the standard covers 1–10 µg/l, while the parametric value of 20 µg/l (30 µg/l in regions with high selenium in groundwater) lies above it — such samples are determined after dilution. Third — subject: the scope of the standard is drinking water, groundwater and surface water; waste water appears only in the precision trial, with a recovery of 86.2 %. Fourth — requirements of the regulation: a limit of quantification of no more than 6 µg/l and an uncertainty of up to 40 % of the parametric value apply to water intended for human consumption assessed under the regulation. Fifth — status: the PKN card has no withdrawal note, while ISO withdrew the introduced document in 2014.
WHAT WE DO NOT GIVE HERE. We do not copy the figures of Table 2 (effect of other substances) or the standard deviations of Table 1. We do not give the differences between ISO 9965 and ISO/TS 17379-2 — we did not read the text of the specification.
Method principle
Organically bound selenium is decomposed by digestion with sulfuric acid and hydrogen peroxide, selenium(VI) is reduced to Se(IV) by boiling with hydrochloric acid, because under the conditions of the method only Se(IV) forms the hydride quantitatively; Se(IV) is reduced with sodium tetrahydroborate in hydrochloric acid medium to gaseous selenium hydride, which is decomposed thermally, and selenium is measured by atomic absorption at 196.0 nm against a calibration curve from at least five solutions (ISO 9965:1993, clauses 3 and 7). Range 1–10 µg/l, higher concentrations after dilution of the sample.
Applications
- Drinking water, groundwater and surface water — selenium, including organically bound selenium, 1–10 µg/l (ISO 9965:1993, clause 1; PKN card)
- Water intended for human consumption — comparison with the parametric value of 20 µg/l (30 µg/l in regions with high selenium in groundwater), Dz.U. 2026 item 748
Key parameters
| Parameter | Value |
|---|---|
| STATUS (as of 18.11.2026) | PN-ISO 9965:2001 without a withdrawal note in PKN; ISO 9965:1993 withdrawn by ISO on 15.07.2014 ("Revised by: ISO/TS 17379-2:2013", Standard Norge); SIST ISO 9965:1998 replaced by SIST-TS ISO/TS 17379-2:2015 (iTeh); no Polish edition of ISO/TS 17379-2 |
| Edition (from the PKN card) | PN-ISO 9965:2001, Polish version; published 13-12-2001, 8 pages, KT 121, ICS 13.060.50; introduces ISO 9965:1993 [IDT] |
| Concentration range (clause 1) | 1–10 µg/l; higher concentrations after dilution of the sample |
| Sample treatment (7.3) | 50 ml of sample; digestion with 5 ml H₂SO₄ and 5 ml H₂O₂ 30 % to sulfuric acid fumes; reduction with 20 ml HCl, gentle boiling 15 min under reflux; to 100 ml |
| Measurement (7.4) | 196.0 nm; e.g. 20 ml of solution, about 5 ml of NaBH₄ solution (3 g NaBH₄ and 1 g NaOH in 100 ml); argon or nitrogen; calibration ≥ 5 solutions, e.g. 1, 3, 5, 8, 10 µg/l |
| Precision (Table 1) | trial of 1992, 19 laboratories: drinking water 3.0 µg/l → mean 2.92 µg/l (recovery 97.4 %); waste water 9.0 µg/l → 7.76 µg/l (86.2 %) |
| Parametric value (Dz.U. 2026 item 748, Annex 1, part B, item 28) | 20 µg/l; 30 µg/l in regions where geological conditions may lead to high levels of selenium in groundwater |
| Requirements for the method (Dz.U. 2026 item 748, Annex 4, part B) | limit of quantification ≤ 30 % of the parametric value (6 µg/l); measurement uncertainty 40 % of the parametric value |
Standard
- Standard number
- PN-ISO 9965:2001
- Title (PL)
- Jakość wody — Oznaczanie selenu — Metoda atomowej spektrometrii absorpcyjnej (technika wodorkowa)
- Title (EN)
- Water quality — Determination of selenium — Atomic absorption spectrometric method (hydride technique)
Step-by-step procedure
-
Sampling and preservation
Polyethylene or borosilicate glass containers cleaned with nitric acid; 20 ml of hydrochloric acid per 1000 ml of sample, to pH ≤ 2 (ISO 9965, clause 6). PN-ISO 9965:2001 without a withdrawal note in PKN, ISO 9965:1993 withdrawn (status as of 18.11.2026).
-
Blank and calibration solutions
2 ml of hydrochloric acid in a 100 ml flask; at least five solutions from standard solution 2, for example 1, 3, 5, 8 and 10 µg/l, prepared daily (7.1–7.2).
-
Digestion
50 ml of sample, 5 ml of sulfuric acid and 5 ml of hydrogen peroxide; heating under reflux to turbid sulfuric acid fumes, another 5 ml of peroxide if necessary; never evaporate to dryness; condensate back into the flask. May be omitted where quantitative recovery has been shown (7.3, 7.3.1).
-
Reduction to Se(IV)
20 ml of hydrochloric acid, gentle boiling 15 min under reflux; without digestion and with free chlorine — purging with nitrogen at about 1 l/min; solution to a 100 ml flask; blank and calibration solutions in the same way (7.3.2).
-
Measurement
196.0 nm; blank, calibration solutions, samples in this order; for example 20 ml of solution, purging with gas to zero, about 5 ml of sodium tetrahydroborate solution, recording of the signal; repetition on a separate portion (7.4).
-
Result
Selenium concentration in µg/l from the calibration function taking dilutions into account, to two significant figures; for water intended for human consumption comparison with the parametric value of 20 µg/l (clauses 8–9; Dz.U. 2026 item 748).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Atomic absorption spectrometer with hydride system | Radiation source for selenium, for example an electrodeless discharge lamp or a hollow cathode lamp; background correction if appropriate; 196.0 nm (ISO 9965, 5.1) | — |
| Gas supply | Argon or nitrogen (ISO 9965, 5.2) | — |
| Digestion apparatus | 250 ml round-bottomed flask with riser tube, reflux condenser, condensate reservoir and polytetrafluoroethene cock — example in Figure 1 of the standard | — |
| Glassware | Cleaned immediately before use with warm dilute nitric acid (for example 2 mol/l) and rinsed with water (ISO 9965, 5.3) | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Sulfuric acid, ρ = 1.84 g/ml | — | Digestion: 5 ml per 50 ml of sample (ISO 9965, 4.1 and 7.3.1) |
| Hydrochloric acid, ρ = 1.16 g/ml | — | Sample preservation (20 ml per 1000 ml), reduction of Se(VI) to Se(IV) (20 ml), blank and calibration solutions (ISO 9965, 4.2) |
| Hydrogen peroxide, 30 % (m/m) | — | Digestion: 5 ml, another 5 ml if necessary (ISO 9965, 4.3) |
| Sodium tetrahydroborate solution | — | 1 g NaOH in about 20 ml of water, 3 g NaBH₄, to 100 ml; prepared daily (ISO 9965, 4.5) |
| Selenium solutions | — | Stock 1000 mg/l from 1.4053 g SeO₂ and 2 g NaOH in 1000 ml; standard 1 — 10 mg/l, standard 2 — 0.1 mg/l, stable for at least one week (ISO 9965, 4.6–4.8) |
Health and safety (OHS)
- Digestion with concentrated sulfuric acid and hydrogen peroxide until sulfuric acid fumes appear — the sample must never be evaporated to dryness (ISO 9965, note 2 in 7.3.1)
- TECHNICAL NOTE: the result covers only selenium converted to Se(IV); without digestion and reduction with hydrochloric acid the result is too low
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-ISO 9965:2001 — “Water quality — Determination of selenium — Atomic absorption spectrometric method (hydride technique)”.
How does this method work?
Organically bound selenium is decomposed by digestion with sulfuric acid and hydrogen peroxide, selenium(VI) is reduced to Se(IV) by boiling with hydrochloric acid, because under the conditions of the method only Se(IV) forms the hydride quantitatively; Se(IV) is reduced with sodium tetrahydroborate in hydrochloric acid medium to gaseous selenium hydride, which is decomposed thermally, and selenium is measured by atomic absorption at 196.0 nm against a calibration curve from at least five solutions (ISO 9965:1993, clauses 3 and 7).
What is the measuring range and accuracy?
STATUS (as of 18.11.2026): PN-ISO 9965:2001 without a withdrawal note in PKN; ISO 9965:1993 withdrawn by ISO on 15.07.2014 ("Revised by: ISO/TS 17379-2:2013", Standard Norge); SIST ISO 9965:1998 replaced by SIST-TS ISO/TS 17379-2:2015 (iTeh); no Polish edition of ISO/TS 17379-2; Edition (from the PKN card): PN-ISO 9965:2001, Polish version; published 13-12-2001, 8 pages, KT 121, ICS 13.060.50; introduces ISO 9965:1993 [IDT]; Concentration range (clause 1): 1–10 µg/l; higher concentrations after dilution of the sample; Sample treatment (7.3): 50 ml of sample; digestion with 5 ml H₂SO₄ and 5 ml H₂O₂ 30 % to sulfuric acid fumes; reduction with 20 ml HCl, gentle boiling 15 min under reflux; to 100 ml.
How long does the test take?
The procedure comprises 6 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Atomic absorption spectrometer with hydride system, Gas supply, Digestion apparatus, Glassware.
Where is this test used?
Drinking water, groundwater and surface water — selenium, including organically bound selenium, 1–10 µg/l (ISO 9965:1993, clause 1; PKN card); Water intended for human consumption — comparison with the parametric value of 20 µg/l (30 µg/l in regions with high selenium in groundwater), Dz.U. 2026 item 748.
What safety precautions apply?
Digestion with concentrated sulfuric acid and hydrogen peroxide until sulfuric acid fumes appear — the sample must never be evaporated to dryness (ISO 9965, note 2 in 7.3.1); TECHNICAL NOTE: the result covers only selenium converted to Se(IV); without digestion and reduction with hydrochloric acid the result is too low.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-ISO 9965.