⚗️ Mercury in drinking water, groundwater, surface water and waste water — cold vapour atomic absorption spectrometry after reduction with tin(II) chloride or sodium borohydride, measurement at 253.7 nm, range 0.1–10 µg/l, higher concentrations after dilution, digestion with potassium permanganate and peroxodisulfate; PN-EN 1483:2007 (withdrawn standard, replaced by PN-EN ISO 12846:2012)

Physicochemistry PN-EN 1483

In short

European standard (in PKN the English version PN-EN 1483:2007) with two methods for determining mercury by cold vapour atomic absorption spectrometry: after reduction with tin(II) chloride (clause 4) or sodium borohydride (clause 5), for drinking water, groundwater, surface water and waste water. The test is performed according to PN-EN 1483:2007; Methods and range (clause 1): tin(II) chloride — clause 4; sodium borohydride — clause 5; 0.1–10 µg/l; lower (0.001–5 µg/l) — mercury analysers or AFS (EN 13506, ISO 17852). The procedure comprises 5 steps; it is used for: Drinking water, groundwater, surface water and waste water — mercury in the range 0.1–10 µg/l, higher concentrations after dilution (EN 1483:2007, clause 1; PKN card), Control of water intended for human consumption — parametric value for mercury 1.0 µg/l, limit of quantification ≤ 0.3 µg/l, uncertainty 30 % of the parametric value (Dz.U. 2026 item 748).

At a glance

  • Standard: PN-EN 1483:2007
  • Category: Physicochemistry
  • Procedure steps: 5
  • STATUS (as of 18.11.2026): PN-EN 1483:2007 (English version) withdrawn 21-07-2016, replaced by PN-EN ISO 12846:2012; NS-EN 1483:2007 "Withdrawn" 01.07.2012 (Standard Norge); EN 1483:2007 withdrawn 15.04.2012 (iTeh)
  • Edition (from the PKN card): PN-EN 1483:2007, English version; published 22-05-2007, 21 pages, KT 121, ICS 13.060.50; introduces EN 1483:2007 [IDT]; replaced PN-EN 1483:2000
  • Methods and range (clause 1): tin(II) chloride — clause 4; sodium borohydride — clause 5; 0.1–10 µg/l; lower (0.001–5 µg/l) — mercury analysers or AFS (EN 13506, ISO 17852)

Overview

STATUS (as of 18.11.2026). The card of PN-EN 1483:2007 (English version) has the header "Withdrawn and replaced by PN-EN ISO 12846:2012 - English version": published 22-05-2007, "Withdrawal date: 21-07-2016", 21 pages, KT 121 Water Quality — Chemical Tests — Inorganic Substances, "Introduces: EN 1483:2007 [IDT]", "Replaces: PN-EN 1483:2000 - Polish version", ICS 13.060.50. The successor, PN-EN ISO 12846:2012 (English version) "Water quality — Determination of mercury — Method using atomic absorption spectrometry (AAS) with and without enrichment" (published 15-06-2012, 18 pages, KT 121, "Introduces: EN ISO 12846:2012 [IDT], ISO 12846:2012 [IDT]", "Replaces: PN-EN 12338:2001 - Polish version, PN-EN 1483:2007 - English version, PN-EN 1483:2007 - German version", additional element Ap1:2016-07 — "The entries in the 'Replaces' field were changed, information on replacements was added to the national foreword"), carries no withdrawal note. With the publishers: Standard Norge — NS-EN 1483:2007 "Withdrawn" on 1 July 2012, "Replaced by: NS-EN ISO 12846:2012"; iTeh — EN 1483:2007 withdrawn on 15.04.2012 and replaced by EN ISO 12846:2012. The description below is based on the sample of the Slovenian edition SIST EN 1483:2007 in the iTeh catalogue: table of contents, foreword, introduction and clauses 1–4.6 (to the start of digestion); clause 5 (sodium borohydride), precision data and annexes are known only from the table of contents and the description in the Standard Norge catalogue.

WHAT THE STANDARD COVERS. Clause 1: two methods for the determination of mercury — with tin(II) chloride as reducing agent (clause 4) and with sodium borohydride (clause 5); the choice depends on the equipment and the matrix (clause 3). Both are suitable for water, for example drinking, ground, surface and waste water, in the range from 0.1 to 10 µg/l; higher concentrations after dilution of the sample, lower ones, from 0.001 to 5 µg/l, with special mercury analysers or atomic fluorescence spectrometry (EN 13506, ISO 17852). Foreword: the 2007 edition replaced EN 1483:1997, takes into account newer methods, describes continuous flow methods instead of the deleted batch methods, and in Annex A.1 adds preservation and digestion with potassium bromate and bromide. Introduction: in natural waters mercury generally occurs below 0.1 µg/l, higher for example in waste water; full decomposition of mercury compounds requires digestion, which may be omitted only if it is certain that the concentration can be measured without it. Principle of the tin(II) chloride method (4.1): mono- or divalent mercury is reduced to the elemental form by tin(II) chloride in an acid medium, stripped from the solution with a stream of inert gas or mercury-free air and transported as an atomic gas into a cuvette; absorbances are measured at 253.7 nm and concentrations calculated from a calibration curve.

INTERFERENCES (clause 3, 4.2). Mercury undergoes adsorption and desorption on vessel walls; its vapour diffuses through some plastics — glass or special plastic tubing (for example FEP, polypropylene-based pump tubing) is suitable, silicone tubing is not. Volatile organic substances absorbing in the UV are removed before reduction: potassium permanganate until a permanent purple colour and 10 min purging with inert gas (background correction often suffices). All solutions are brought to the same temperature (below 25 °C) before reduction; the cuvette is heated slightly above 100 °C to prevent condensation. Table 1 gives acceptable concentrations of matrix elements — for tin(II) chloride in HCl 0.5 mol/l: Cu(II) 500, Ni(II) 500, Ag(I) 1, iodide 0.1, As(V), Bi(III), Sb(III) 0.5 each, Se(IV) 0.05 mg/l; tin(II) chloride is less prone to heavy-metal interferences than borohydride but contaminates the apparatus with tin — separate systems are used for the two reducing agents. Iodide above 0.1 mg/l interferes with the tin(II) chloride method — then reduction with sodium borohydride is used. Without digestion tin(II) chloride does not fully reduce some inorganic (for example mercury sulfide) and organic mercury compounds.

REAGENTS, APPARATUS AND SAMPLES (4.3–4.5). Reagents at least of analytical grade or with particularly low mercury content; water grade 1 according to EN ISO 3696. Nitric acid ρ = 1.40 g/ml, sulfuric acid ρ = 1.84 g/ml, hydrochloric acid ρ = 1.16 g/ml; potassium permanganate solution 50 g/l; stabilizer solution — 5 g of potassium dichromate in 500 ml of nitric acid, to 1000 ml; potassium peroxodisulfate 40 g/l; hydroxylammonium chloride 10 g/100 ml; tin(II) chloride dihydrate 50 g/l (5 g in 30 ml HCl, to 100 ml; in flow systems a lower concentration, for example 20 g/l, according to the manufacturer; with a high blank — 30 min purging with nitrogen). Mercury solutions: stock 1000 mg/l (commercial), 10 mg/l (stable one week), standard 100 and 50 µg/l (prepared on the day of use), each with 1 ml of stabilizer solution per 100 ml; calibration for example 0.5; 1; 2; 3; 4 and 5 µg/l (1–10 ml of the 50 µg/l solution in 100 ml flasks with 1 ml of stabilizer), before each series; reagent blank with 10 ml of stabilizer per 1000 ml of water, digested like the sample. Glassware is rinsed with dilute nitric acid (150 ml HNO₃ per 1000 ml) and then with water. Apparatus: atomic absorption spectrometer with a cold vapour generation system and a quartz or glass tube atomizer (background correction recommended) — also stand-alone flow mercury analysers (range about one order of magnitude lower) and batch systems; radiation source for mercury (hollow cathode or electrodeless discharge lamp; in continuum source AAS for example a xenon short-arc lamp); accessory with peristaltic pump, valves, reaction coil or cell, gas–liquid separator and cuvette. Samples are taken according to EN ISO 5667-1 and EN ISO 5667-3 into vessels of borosilicate glass, quartz, polysulfone or FEP; 10 ml of stabilizer solution are added per 1000 ml of sample; the sample shall have pH 1 and a yellow-orange colour indicating excess dichromate.

DIGESTION (4.6, in the sample up to the heating step). 100 ml of stabilized sample (or an appropriate volume, at most 1000 ml) in a flask of one of the materials of 4.5; carefully 15 ml of potassium permanganate solution, 1 ml of nitric acid and 1 ml of sulfuric acid, shaking after each addition; after 15 min 10 ml of potassium peroxodisulfate solution; digestion in a loosely stoppered flask at (95 ± 3) °C for 2 h. Instead, the digestions of Annexes A (bromate–bromide), B (autoclave) or C (microwave) may be used after demonstrating equivalent efficiency.

IN THE REGULATIONS. Regulation of the Minister of Health of 22 May 2026 on the quality of water intended for human consumption (Dz.U. 2026 item 748; status "in force" in the ELI database of the Sejm, 18.11.2026): Annex 1, part B, item 27 — mercury, parametric value 1.0 µg/l; Annex 4, part B — the method has to be able to measure concentrations equal to the parametric value with a limit of quantification of 30 % or less of the parametric value (for mercury — 0.3 µg/l), and the measurement uncertainty for mercury (item 40) is 30 % of the parametric value; the regulation names no standard for mercury. Regulation on the discharge of waste water (Dz.U. 2019 item 1311, status "in force"), Annex 12, item 50 "Mercury": atomic absorption spectrophotometry — PN-EN ISO 12846 (detectability 0.1 of the permissible concentration at the sampling point, accuracy and precision ± 30 % at a concentration equal to the detectability) and fluorescence spectroscopy — PN-EN ISO 17852; this standard is not listed there; according to explanatory note 2 to the annex, where a named standard has been replaced and withdrawn, the new Polish Standard is considered the recommended one.

WHERE A SEEMINGLY CORRECT RESULT IS EASY TO GET. First — status: the standard has been withdrawn since 21.07.2016 and replaced by PN-EN ISO 12846:2012, which the waste water regulation also names. Second — digestion: without it tin(II) chloride does not fully reduce mercury sulfide and organic mercury compounds, and the result is too low; it is omitted only if it is certain that mercury can be measured without it. Third — choice of reducing agent: with iodide above 0.1 mg/l the tin(II) chloride method is disturbed; with high concentrations of copper, nickel or silver the acceptable concentrations in Table 1 depend on the reducing agent and acidity; systems for the two reducing agents are not combined. Fourth — range: 0.1–10 µg/l, higher after dilution; for water intended for human consumption the limit of quantification must be no more than 0.3 µg/l. Fifth — losses and contamination: mercury adsorbs on walls and diffuses through some plastics (silicone is unsuitable), and a sample without dichromate (pH 1, yellow-orange colour) loses mercury.

WHAT WE DO NOT GIVE HERE. The rest of the digestion and determination of clause 4, the sodium borohydride method (clause 5), precision data (clause 6) and the procedures of Annexes A–C — the sample ends at 4.6.

Method principle

Mono- or divalent mercury is reduced with tin(II) chloride in an acid medium (clause 4) or with sodium borohydride (clause 5) to the elemental form, stripped with inert gas or mercury-free air and transported as an atomic gas into the cuvette of an atomic absorption spectrometer, the absorbance being measured at 253.7 nm against a calibration curve; the sample is preserved with potassium dichromate in nitric acid and digested with potassium permanganate and peroxodisulfate at (95 ± 3) °C (EN 1483:2007, clauses 1, 4.1, 4.5–4.6). Range 0.1–10 µg/l.

Applications

Key parameters

ParameterValue
STATUS (as of 18.11.2026)PN-EN 1483:2007 (English version) withdrawn 21-07-2016, replaced by PN-EN ISO 12846:2012; NS-EN 1483:2007 "Withdrawn" 01.07.2012 (Standard Norge); EN 1483:2007 withdrawn 15.04.2012 (iTeh)
Edition (from the PKN card)PN-EN 1483:2007, English version; published 22-05-2007, 21 pages, KT 121, ICS 13.060.50; introduces EN 1483:2007 [IDT]; replaced PN-EN 1483:2000
Methods and range (clause 1)tin(II) chloride — clause 4; sodium borohydride — clause 5; 0.1–10 µg/l; lower (0.001–5 µg/l) — mercury analysers or AFS (EN 13506, ISO 17852)
Measurement (4.1, 4.4)cold vapour, quartz or glass tube atomizer, 253.7 nm; background correction recommended; solution temperature < 25 °C
Preservation (4.5)10 ml of stabilizer solution (5 g K₂Cr₂O₇ in 500 ml HNO₃, to 1000 ml) per 1000 ml of sample; pH 1, yellow-orange colour
Digestion (4.6)100 ml of sample, 15 ml KMnO₄ 50 g/l, 1 ml HNO₃, 1 ml H₂SO₄, after 15 min 10 ml K₂S₂O₈ 40 g/l; (95 ± 3) °C, 2 h; alternatively Annexes A–C after demonstrating equivalence
Regulationswater for human consumption: 1.0 µg/l, limit of quantification ≤ 0.3 µg/l, uncertainty 30 % (Dz.U. 2026 item 748); waste water: reference methods PN-EN ISO 12846 and PN-EN ISO 17852 (Dz.U. 2019 item 1311, Annex 12, item 50)

Standard

Standard number
PN-EN 1483:2007
Title (PL)
Jakość wody — Oznaczanie rtęci — Metoda z zastosowaniem atomowej spektrometrii absorpcyjnej
Title (EN)
Water quality — Determination of mercury — Method using atomic absorption spectrometry

Step-by-step procedure

  1. Sampling and preservation

    Vessels of borosilicate glass, quartz, PSU or FEP; 10 ml of stabilizer solution per 1000 ml; pH 1, yellow-orange colour (4.5). PN-EN 1483:2007 withdrawn 21-07-2016 and replaced by PN-EN ISO 12846:2012 (status as of 18.11.2026).

  2. Digestion

    100 ml of sample, 15 ml KMnO₄, 1 ml HNO₃, 1 ml H₂SO₄, after 15 min 10 ml K₂S₂O₈, (95 ± 3) °C for 2 h; reagent blank in the same way (4.3.15, 4.6).

  3. Calibration

    Solutions prepared before each series, for example 0.5; 1; 2; 3; 4 and 5 µg/l with stabilizer solution (4.3.14).

  4. Reduction and measurement

    Tin(II) chloride (clause 4) or sodium borohydride (clause 5), separate systems; stripping of mercury into the cuvette, 253.7 nm; solutions at the same temperature below 25 °C (clause 3, 4.1).

  5. Result

    Mercury concentration in µg/l from the calibration curve, taking into account dilutions and the stabilizer volume; the test report states the reducing agent (clause 4 or 5) (4.1, 4.5).

Required equipment and apparatus

EquipmentExampleIndicative price
Atomic absorption spectrometer with cold vapour systemQuartz or glass tube atomizer, background correction recommended; also stand-alone flow mercury analysers and batch systems (EN 1483, 4.4.1)—
Radiation source for mercuryHollow cathode or electrodeless discharge lamp; in continuum source AAS for example a xenon short-arc lamp (4.4.2)—
Mercury accessoryPeristaltic pump, carrier gas valves, reaction coil or cell, gas–liquid separator, borosilicate glass or quartz cuvette; glass or FEP tubing, not silicone (4.4.3, clause 3)—

Reagents, media and consumables

ReagentCASDetails
Tin(II) chloride solution, 50 g/l SnCl₂·2H₂O—5 g in 30 ml HCl ρ 1.16, to 100 ml; in flow systems less, for example 20 g/l (EN 1483, 4.3.9)
Stabilizer solution—5 g potassium dichromate in 500 ml HNO₃ ρ 1.40, to 1000 ml; 10 ml per 1000 ml of sample (4.3.6, 4.5)
Potassium permanganate 50 g/l and potassium peroxodisulfate 40 g/l—Digestion: 15 ml and 10 ml per 100 ml of sample (4.3.5, 4.3.7, 4.6)
Hydroxylammonium chloride 10 g/100 ml—Reagent of the tin(II) chloride method (4.3.8)
Mercury solutions—Stock 1000 mg/l (commercial), 10 mg/l, standard 100 and 50 µg/l, calibration for example 0.5–5 µg/l — each with stabilizer solution (4.3.10–4.3.14)

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN 1483:2007 — “Water quality — Determination of mercury — Method using atomic absorption spectrometry”.

How does this method work?

Mono- or divalent mercury is reduced with tin(II) chloride in an acid medium (clause 4) or with sodium borohydride (clause 5) to the elemental form, stripped with inert gas or mercury-free air and transported as an atomic gas into the cuvette of an atomic absorption spectrometer, the absorbance being measured at 253.7 nm against a calibration curve; the sample is preserved with potassium dichromate in nitric acid and digested with potassium permanganate and peroxodisulfate at (95 ± 3) °C (EN 1483:2007, clauses 1, 4.

What is the measuring range and accuracy?

STATUS (as of 18.11.2026): PN-EN 1483:2007 (English version) withdrawn 21-07-2016, replaced by PN-EN ISO 12846:2012; NS-EN 1483:2007 "Withdrawn" 01.07.2012 (Standard Norge); EN 1483:2007 withdrawn 15.04.2012 (iTeh); Edition (from the PKN card): PN-EN 1483:2007, English version; published 22-05-2007, 21 pages, KT 121, ICS 13.060.50; introduces EN 1483:2007 [IDT]; replaced PN-EN 1483:2000; Methods and range (clause 1): tin(II) chloride — clause 4; sodium borohydride — clause 5; 0.1–10 µg/l; lower (0.001–5 µg/l) — mercury analysers or AFS (EN 13506, ISO 17852); Measurement (4.1, 4.4): cold vapour, quartz or glass tube atomizer, 253.7 nm; background correction recommended; solution temperature < 25 °C.

How long does the test take?

The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Atomic absorption spectrometer with cold vapour system, Radiation source for mercury, Mercury accessory.

Where is this test used?

Drinking water, groundwater, surface water and waste water — mercury in the range 0.1–10 µg/l, higher concentrations after dilution (EN 1483:2007, clause 1; PKN card); Control of water intended for human consumption — parametric value for mercury 1.0 µg/l, limit of quantification ≤ 0.3 µg/l, uncertainty 30 % of the parametric value (Dz.U. 2026 item 748).

What safety precautions apply?

Mercury and its compounds are very toxic — samples and solutions that may contain mercury require extreme caution; potassium dichromate is toxic (EN 1483:2007, introduction); TECHNICAL NOTE: the standard is withdrawn and replaced by PN-EN ISO 12846:2012; for mercury in waste water the regulation (Dz.U. 2019 item 1311, Annex 12) names PN-EN ISO 12846 and PN-EN ISO 17852.

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 1483.

🔍 Find a laboratory performing this test