🔥 Crude petroleum — water content by the distillation method: heating the sample under reflux with xylene, separation of water in a calibrated 5 ml trap (0.05 ml graduations), reading to 0.025 ml, solvent blank, result in % (V/V) or % (m/m) to 0.025 %; PN-EN ISO 9029

Petrochemistry PN-EN ISO 9029

In short

Determination of the water content of crude petroleum by the distillation method. The test is performed according to PN-EN ISO 9029:2005 (Polish version; implements EN ISO 9029:1995 and ISO 9029:1990 [IDT]); STATUS (PKN card as of 11.11.2026): PN-EN ISO 9029:2005 current (Polish version); implements EN ISO 9029:1995 and ISO 9029:1990; ISO 9029:1990 confirmed (iTeh, stage 9093). The procedure comprises 5 steps; it is used for: Crude petroleum — water content, among other things to correct the volume in custody transfer of oil (ISO 9029:1990, clauses 1 and 3), Precision data only up to 1 % (V/V) water (clause 1, PKN card), Replaced in PN the standards PN-C-04523:1983 (with amendment Ap2:2004) and the English edition PN-EN ISO 9029:2002 (PKN card).

At a glance

  • Standard: PN-EN ISO 9029:2005 (Polish version; implements EN ISO 9029:1995 and ISO 9029:1990 [IDT])
  • Category: Petrochemistry
  • Procedure steps: 5
  • STATUS (PKN card as of 11.11.2026): PN-EN ISO 9029:2005 current (Polish version); implements EN ISO 9029:1995 and ISO 9029:1990; ISO 9029:1990 confirmed (iTeh, stage 9093)
  • Apparatus — condition (5.2–5.4): 1000 ml flask (24/39 joint), 5 ml trap with 0.05 ml graduations, 400 mm Liebig condenser
  • Solvent — condition (clause 6, 9.2): xylene according to ISO 5280, 400 ml in total in the flask; water content of the xylene — blank test

Overview

WHAT THE STANDARD COVERS. According to clause 1 of ISO 9029:1990 the standard specifies a method for determining water in crude oil by distillation; precision data have only been determined for water contents up to 1 % (V/V). Clause 3: knowledge of the water content of crude oil is important in the refining, purchase, sale and transfer of products, and the amount of water determined by this method is used to correct the volume involved in the custody transfer of oil. The "Scope" field of the PKN card for PN-EN ISO 9029:2005: "The method of determining the water content of crude petroleum by distillation, used for testing products with a water content of up to 1 percent (V/V), is given. The principle of the method, the apparatus and the type of solvent, calibration, sampling and preparation of samples, performance of the determination, the method of calculating the results and the precision of the method are given".

ACCORDING TO THE TEXT OF ISO 9029:1990 — APPARATUS, CALIBRATION, SAMPLES, PROCEDURE, CALCULATION, PRECISION (foreword, clauses 1–12). The standard was prepared by ISO/TC 28 (Petroleum products and lubricants); first edition 1990-12-15; Annex A is normative. References: ISO 383 (conical ground joints), ISO 3170 and ISO 3171 (manual and automatic sampling), ISO 4259 (precision), ISO 5280 (xylene for industrial use). Principle (clause 4): the test portion is heated under reflux with a water-immiscible solvent that co-distils with the water in the sample; condensed solvent and water are continuously separated in a trap — the water settles in the graduated section of the trap and the solvent returns to the flask. Apparatus (clause 5): 1000 ml round-bottom glass distillation flask with a 24/39 ground-glass socket; 5 ml glass trap with 0.05 ml graduations; 400 mm Liebig condenser; drying tube filled with a self-indicating desiccant at the top of the condenser (prevents entry of atmospheric moisture); gas or electric heater distributing heat uniformly over the entire lower half of the flask (an electric heating mantle is preferred for safety reasons); jet-spray tube for washing down the condenser; pick of brass or bronze or steel scraper with a PTFE tip. Other apparatus may be used if it can be shown to operate within the established precision (ISO 4259). Solvent (clause 6): xylene according to ISO 5280; its water content is determined by a blank test (9.6). Calibration and recovery test (clause 7): before initial use the trap is calibrated by adding 0.05 ml increments of distilled water from a 5 ml microburette or a micro-pipette readable to 0.01 ml — a deviation of more than 0.05 ml means rejecting or recalibrating the trap; before a series of tests the whole apparatus is checked: 400 ml of dry xylene (0.02 % water maximum) — procedure as in clause 9, then 1.00 ml ± 0.01 ml of distilled water added and the procedure repeated, then 4.50 ml ± 0.01 ml; permissible readings (Table 1, 20 °C): 1.00 ± 0.025 ml and 4.50 ± 0.025 ml. A reading outside the limits indicates vapour leaks, too rapid boiling, inaccurate graduations or ingress of moisture. Samples (clause 8): only representative samples taken according to ISO 3170 or ISO 3171; size of the test portion according to Table 2 depending on the expected water content: 50.1–100.0 % — 5 g or ml; 25.1–50.0 % — 10; 10.1–25.0 % — 20; 5.1–10.0 % — 50; 1.1–5.0 % — 100; 0.5–1.0 % — 200; less than 0.5 % — 200. If the homogeneity is doubtful, the whole sample or at least three test portions are tested and all results and their average reported. A volume-based test portion is measured with a cylinder, which is rinsed with five portions of xylene; a mass-based portion is weighed directly into the flask. Procedure (clause 9): the apparatus is chemically cleaned at least daily (water droplets adhering to the walls impair precision); xylene is added to the flask to a total volume of 400 ml; a magnetic stirrer is the most effective device to reduce bumping; water at 20–25 °C is circulated through the condenser jacket; greasing of glass joints is not recommended. Heating is slow at first (about 0.5–1 h), the condensate rises no higher than three-quarters of the condenser inner tube (point A in Figure 1); the distillate is collected at 2–5 drops per second; distillation continues until no water is visible anywhere except in the trap and the volume of water in the trap remains constant for at least 5 min. Droplets in the condenser are flushed with xylene (jet-spray tube) after the heat has been shut off for at least 15 min before wash-down; adding an oil-soluble emulsion breaker to the xylene at 1000 ppm helps; after flushing, redistillation for at least 5 min. When the trap has cooled to room temperature, droplets on the walls are transferred to the water layer with the pick or PTFE scraper and the volume read — graduated in 0.05 ml, estimated to 0.025 ml. Blank (9.6): 400 ml of solvent in the flask, procedure as in 9.1–9.5. Calculation (clause 10): % (V/V) = (V2 − V0)/V1 × 100; % (V/V) = (V2 − V0)/(m/ρ) × 100; % (m/m) = (V2 − V0)/m × 100, where V0 — volume of water in the trap in the blank (ml, to 0.025 ml), V1 — volume of the test portion (ml), V2 — volume of water in the trap (ml, to 0.025 ml), m — mass of the test portion (g), ρ — density of the sample (g/ml); the density of water is assumed to be 1 g/ml. Volatile water-soluble material may be measured as water. The result is rounded to 0.025 %. Precision (clause 11, interlaboratory tests in the range 0.01–1.0 %): repeatability — from 0.0 to 0.1 % (V/V) see Figure 3, from 0.1 to 1.0 % (V/V) — 0.08 % (V/V); reproducibility — from 0.0 to 0.1 % (V/V) see Figure 3, from 0.1 to 1.0 % (V/V) — 0.11 % (V/V). Test report (clause 12): among other things identification of the product, result, individual results of the test portions, deviations, date of the test.

STATUS (PKN card as of 11.11.2026). The card for PN-EN ISO 9029:2005 (Polish version) has no withdrawal label: published 12-07-2005, 16 pages, KT 222 for Petroleum Products and Similar Products of Biological and Synthetic Origin, "Implements: EN ISO 9029:1995 [IDT], ISO 9029:1990 [IDT]", "Replaces: PN-EN ISO 9029:2002 - English version, PN-C-04523:1983 - Polish version, PN-C-04523:1983/Ap2:2004 - Polish version", ICS 75.040. The PKN search (term "PN-EN ISO 9029", 11.11.2026) returns only the cards of the 2005 (Polish) and 2002 (English) editions. iTeh catalogue (read 11.11.2026): ISO 9029:1990 — stage "9093 – International Standard confirmed".

HOW MANY LABORATORIES AND IN WHAT FORM (accreditation scope database of the service, read 11.11.2026). The notation ISO 9029 appears in 6 records at 5 Polish laboratories: AB 474 and AB 484 (2 records) — crude petroleum, in both alongside ASTM D 4006-22; AB 079 and AB 1091 — heavy fuel oil; AB 008 — waterproofing products, range (1.0–70.0) % m/m, "Gravimetric method (distillation)". Outside Poland — 2 records at 2 Czech laboratories (1328 and 1152.1: SOP procedures citing ČSN EN ISO 9029). We have not compared the current PCA scope documents of these laboratories with the database.

WHAT THE LABORATORIES TEST (scope database, complete items — 6 Polish). Crude petroleum (AB 474, AB 484 — 2), heavy fuel oil (AB 079, AB 1091) and waterproofing products (AB 008) — water content by the distillation method. The text of the standard names only crude petroleum, and the precision data reach 1 % (V/V) — for a range up to 70 % m/m (AB 008) the standard gives no precision.

WHERE AN APPARENTLY CORRECT RESULT IS EASY TO GET. Apparatus: droplets on dirty walls do not drain into the trap — chemical cleaning at least daily; leaking joints, too rapid boiling and atmospheric moisture (no drying tube) are typical causes of a reading outside the recovery tolerance. Test portion: too small for a low water content (Table 2 — 200 g or ml below 1 %); for a non-homogeneous sample three portions and the average. Procedure: too fast initial heating — bumping and loss of water; condensate above three-quarters of the condenser; flushing with the heat on risks bumping. Calculation: omitting the xylene blank (V0); a % (V/V) result from mass requires the sample density; volatile water-soluble material overstates the result. Result to 0.025 %.

WHAT WE DO NOT GIVE. Parts read: the iTeh sample of SIST ISO 9029:1996 (ISO 9029:1990 in English: foreword, clauses 1–12) and the iTeh sample of SIST EN ISO 9029:1998 (clauses 1–9.3); figures of the tables and formulae read from the page images. We do not give the content of the normative Annex A (sampling and procedure) or of Figures 1–3 (assembly, accessories, precision chart for 0.0–0.1 %) — we have not read them. We have not read the requirements of ISO 3170, ISO 3171, ISO 4259 and ISO 5280. We have not compared the text of ASTM D 4006 used alongside the standard by two laboratories.

Method principle

A test portion (5–200 g or ml according to the expected water content) is heated in a 1000 ml flask under a 400 mm Liebig condenser with xylene to a total volume of 400 ml; the water co-distils with the solvent and separates in a 5 ml trap (0.05 ml graduations), while the xylene returns to the flask. Distillation continues until the volume of water is constant for at least 5 min, the reading is estimated to 0.025 ml, the solvent blank is subtracted and the water content is calculated in % (V/V) or % (m/m) to 0.025 % (ISO 9029:1990, clauses 4–10).

Applications

Key parameters

ParameterValue
STATUS (PKN card as of 11.11.2026)PN-EN ISO 9029:2005 current (Polish version); implements EN ISO 9029:1995 and ISO 9029:1990; ISO 9029:1990 confirmed (iTeh, stage 9093)
Apparatus — condition (5.2–5.4)1000 ml flask (24/39 joint), 5 ml trap with 0.05 ml graduations, 400 mm Liebig condenser
Solvent — condition (clause 6, 9.2)xylene according to ISO 5280, 400 ml in total in the flask; water content of the xylene — blank test
Recovery test — condition (7.3, Table 1)addition of 1.00 ml and 4.50 ml (± 0.01 ml) of water; readings 1.00 ± 0.025 ml and 4.50 ± 0.025 ml at 20 °C
Test portion — condition (8.3.1, Table 2)5 (50.1–100 %), 10, 20, 50, 100 g or ml … 200 g or ml (0.5–1.0 % and below 0.5 %)
Distillation — condition (9.3–9.4)condenser water 20–25 °C; 2–5 drops/s; condensate up to 3/4 of the condenser; end — constant volume of water for ≥ 5 min
Result — calculation (clause 10)% (V/V) = (V2 − V0)/V1 × 100 or (V2 − V0)/(m/ρ) × 100; % (m/m) = (V2 − V0)/m × 100; rounded to 0.025 %
Precision — 0.1–1.0 % (V/V) (clause 11)repeatability 0.08 % (V/V), reproducibility 0.11 % (V/V); for 0.0–0.1 % — Figure 3
Coverage in accreditation scopes (database, read 11.11.2026)5 Polish laboratories, 6 records; 2 Czech laboratories

Standard

Standard number
PN-EN ISO 9029:2005 (Polish version; implements EN ISO 9029:1995 and ISO 9029:1990 [IDT])
Title (PL)
Ropa naftowa — Oznaczanie wody — Metoda destylacyjna
Title (EN)
Crude petroleum — Determination of water — Distillation method

Step-by-step procedure

  1. Calibration and recovery test

    Trap — 0.05 ml increments of water (deviation ≤ 0.05 ml); before a series: 400 ml of dry xylene, then 1.00 ml and 4.50 ml of water — readings within the tolerances of Table 1 (clause 7).

  2. Test portion

    Size according to Table 2; by volume — cylinder and rinsing five times with xylene; by mass — weighing into the flask; xylene to 400 ml in total (8.3, 9.2).

  3. Distillation

    Tight assembly, condenser 20–25 °C; heating slow at first (0.5–1 h), then 2–5 drops/s, condensate up to 3/4 of the condenser; until the volume of water is constant for ≥ 5 min (9.3–9.4).

    🌡 Temperature: cooling water 20–25 °C

  4. Wash-down and reading

    Droplets in the condenser — flushing with xylene after the heat has been off for ≥ 15 min, redistillation ≥ 5 min; after cooling, droplets on the walls into the water layer, reading to 0.025 ml (9.4–9.5).

  5. Blank and result

    400 ml of solvent treated the same way (9.6); % (V/V) or % (m/m) according to the formulae of clause 10, rounded to 0.025 %; report according to clause 12. PN-EN ISO 9029:2005 — current (PKN card as of 11.11.2026).

Required equipment and apparatus

EquipmentExampleIndicative price
Distillation flask (5.2)Round-bottom, glass, 1000 ml, 24/39 conical ground-glass socket—
Trap (5.3)Glass, 5 ml, 0.05 ml graduations, 24/39 joints; calibrated before initial use (7.2)—
Condenser and drying tube (5.4–5.5)400 mm Liebig condenser; at the top a tube with a self-indicating desiccant—
Heater (5.6)Gas or electric, heating the lower half of the flask uniformly; electric heating mantle preferred—
Accessories (5.7–5.8)Jet-spray tube for washing down the condenser; pick of brass or bronze or steel scraper with a PTFE tip; magnetic stirrer (9.2.2)—

Reagents, media and consumables

ReagentCASDetails
Xylene—According to ISO 5280 (clause 6); for the recovery test dry — 0.02 % water maximum (7.3); 400 ml in total in the flask
Distilled water—For calibrating the trap (0.05 ml increments) and for the recovery test (1.00 ml and 4.50 ml ± 0.01 ml)
Oil-soluble emulsion breaker—Added to the wash xylene at a concentration of 1000 ppm — helps dislodge water droplets (9.4)

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN ISO 9029:2005 (Polish version; implements EN ISO 9029:1995 and ISO 9029:1990 [IDT]) — “Crude petroleum — Determination of water — Distillation method”.

How does this method work?

A test portion (5–200 g or ml according to the expected water content) is heated in a 1000 ml flask under a 400 mm Liebig condenser with xylene to a total volume of 400 ml; the water co-distils with the solvent and separates in a 5 ml trap (0.05 ml graduations), while the xylene returns to the flask.

What is the measuring range and accuracy?

STATUS (PKN card as of 11.11.2026): PN-EN ISO 9029:2005 current (Polish version); implements EN ISO 9029:1995 and ISO 9029:1990; ISO 9029:1990 confirmed (iTeh, stage 9093); Apparatus — condition (5.2–5.4): 1000 ml flask (24/39 joint), 5 ml trap with 0.05 ml graduations, 400 mm Liebig condenser; Solvent — condition (clause 6, 9.2): xylene according to ISO 5280, 400 ml in total in the flask; water content of the xylene — blank test; Recovery test — condition (7.3, Table 1): addition of 1.00 ml and 4.50 ml (± 0.01 ml) of water; readings 1.00 ± 0.025 ml and 4.50 ± 0.025 ml at 20 °C.

How long does the test take?

The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Distillation flask (5.2), Trap (5.3), Condenser and drying tube (5.4–5.5), Heater (5.6), Accessories (5.7–5.8).

Where is this test used?

Crude petroleum — water content, among other things to correct the volume in custody transfer of oil (ISO 9029:1990, clauses 1 and 3); Precision data only up to 1 % (V/V) water (clause 1, PKN card); Replaced in PN the standards PN-C-04523:1983 (with amendment Ap2:2004) and the English edition PN-EN ISO 9029:2002 (PKN card); In accreditation scopes: 5 Polish laboratories — crude petroleum, heavy fuel oil, waterproofing products.

What safety precautions apply?

The standard warns that its use may involve hazardous materials, operations and equipment and does not address all safety problems — health and safety practices are set by the user; An electric heating mantle is preferred for safety reasons (5.6); wash-down of the condenser only after the heat has been shut off at least 15 min before (9.4); TECHNICAL NOTE: precision was determined only up to 1 % (V/V) water; volatile water-soluble material is measured as water.

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 9029.

🔍 Find a laboratory performing this test