📏 Hand torque tools — calibration (measurement method according to ISO 6789-1) and determination of uncertainty: relative measurement error a_s, uncertainty interval of the device W′md ≤ 1/4 W′, intervals of 12 months or 5 000 cycles, torque measurement device every 24 months; PN-EN ISO 6789-2
In short
The method for the calibration of hand torque tools (indicating Type I and setting Type II) and the method of calculating the measurement uncertainty of the calibration, together with minimum requirements for calibrating the torque measurement device where its uncertainty interval is not provided by a calibration certificate. The test is performed according to PN-EN ISO 6789-2:2017-03 (English version; implements EN ISO 6789-2:2017 and ISO 6789-2:2017 [IDT]); STATUS (PKN card, 16.11.2026): PN-EN ISO 6789-2:2017-03 (English version, published 30-03-2017); implements EN ISO 6789-2:2017 and ISO 6789-2:2017; replaces PN-EN ISO 6789:2009. The procedure comprises 5 steps; it is used for: Calibration of hand torque wrenches and screwdrivers of Type I and II with determination of the measurement uncertainty (ISO 6789-2:2017, clause 1), Calibration of the torque measurement device where W′md is not provided by a traceable calibration certificate (clause 1, annex C), Calibration during use: by default every 12 months or 5 000 cycles and after overload, repair and improper handling (4.1).
At a glance
- Standard: PN-EN ISO 6789-2:2017-03 (English version; implements EN ISO 6789-2:2017 and ISO 6789-2:2017 [IDT])
- Category: Metrology
- Procedure steps: 5
- STATUS (PKN card, 16.11.2026): PN-EN ISO 6789-2:2017-03 (English version, published 30-03-2017); implements EN ISO 6789-2:2017 and ISO 6789-2:2017; replaces PN-EN ISO 6789:2009
- Interval between calibrations of the tool (4.1): according to the control procedure; without it by default 12 months or 5 000 cycles (whichever occurs first), from first use
- Torque measurement device (4.3): W′md ≤ 1/4 W′ at each target value; certificate from a laboratory according to ISO/IEC 17025; without a procedure — at most every 24 months; after overload > 20 % of T_E — recalibration
Overview
WHAT THE STANDARD COVERS. According to clause 1 of ISO 6789-2:2017 the document specifies the method for the calibration of hand torque tools and describes the method of calculation of measurement uncertainties for the calibration; it also specifies the minimum requirements for the calibration of the torque measurement device where the relative measurement uncertainty interval W′md is not already provided by a traceable calibration certificate. ISO 6789 is applicable to the step-by-step (static) and continuous (quasi-static) calibration of torque measurement devices in which the torque is determined from the elastic deformation of a body or a quantity proportional to it. The standard applies to indicating (Type I) and setting (Type II) tools identified in ISO 1703:2005 by reference numbers 6 1 00 11 0, 6 1 00 11 1, 6 1 00 12 0, 6 1 00 12 1, 6 1 00 14 0 and 6 1 00 15 0. The scope on the PKN card of PN-EN ISO 6789-2:2017-03 gives the calibration method and the uncertainty calculation.
ACCORDING TO THE TEXT OF ISO 6789-2:2017 (publisher's sample in English: foreword, introduction, clauses 1–6.1). The standard was prepared by ISO/TC 29 (Small tools), subcommittee SC 10; together with ISO 6789-1 it replaced ISO 6789:2003 after technical revision: part 2 specifies the requirements for traceable calibration certificates, contains detailed methods for calculating the uncertainty budget, which shall be performed for each tool, example calculations for different types of tools and annex C with requirements for calibrating the torque measurement device where the laboratory does not use a national standard giving such requirements. The introduction explains that the aim is a calibration in which the sources of uncertainty are evaluated and define the range within which the readings lie; with sufficient data the Type B components may be determined statistically for a given model of tool, provided they are reviewed periodically. References: ISO 6789-1:2017 and ISO/IEC 17025. Definitions (3.1): Type A evaluation of uncertainty (statistical analysis of a series of observations — data from the calibration of each tool, not prepared in advance) and Type B (by other means; ISO/IEC Guide 98-3), calibration system (measurement device and loading system — the standard for the tool), measurement device (working standard, mechanical or with an electronic transducer and display), reference measurement standard and measurement error (measured value minus reference value; ISO/IEC Guide 99). Table 1 lists the symbols: among others a_s — calculated relative measurement error of the tool, b_re, b_rep, b_od, b_int, b_l — variations due to repeatability, reproducibility, geometry of the output drive, the interface between the output drive and the calibration system and the position of the force loading point, b_z — hysteresis error of the zero signal, I₀ and I_z — zero indications 30 s after preload and after unloading, k = 2 (confidence level approx. 95 %), components of the relative standard uncertainty w (w_re, w_rep, w_od, w_int, w_l, w_r) and of the device w_md, relative expanded uncertainty W, relative uncertainty interval W′, W′md and W′ref.
CALIBRATION (clause 4). If the user has procedures for the control of test devices, they include torque tools, and the interval between calibrations is chosen considering the required maximum permissible error, frequency of use, typical load and conditions of operation and storage, and adapted on the basis of successive results; without such procedures the default is 12 months or 5 000 cycles, whichever occurs first, starting from the first use; a shorter interval may be required by the user, their customer or legislation. The tool is also calibrated after an overload greater than the values of ISO 6789-1:2017, 5.1.6, after repair and after improper handling. The calibration method corresponds to the measurement method of ISO 6789-1:2017, clause 6, with the requirement for the torque measurement device in 6.1 of part 1 replaced by 4.3: at each target value the relative uncertainty interval of the device W′md shall not exceed 1/4 of the expected maximum relative uncertainty interval of the tool W′; the device has a valid calibration certificate from a laboratory meeting ISO/IEC 17025 or is calibrated by a laboratory maintaining the national standard; without a control procedure — at most every 24 months; recalibration after an overload above 20 % of T_E, after repair or after improper use.
MEASUREMENT ERROR (clause 5). The calibration values are measured and recorded according to ISO 6789-1:2017, 6.5; relative measurement error: a_s = (X_a − X_r) × 100 / X_r (formula 1), and the mean of n measurements at each calibration torque — formula 2. Example 1 (indicating and graduated setting tools, X_a = 100 N·m, five reference values 104,0; 96,5; 102,6; 99,0; 101,0 N·m): a_s = −3,85 %, +3,63 %, −2,53 %, +1,01 %, −0,99 %. Example 2 (Type II classes C and F, X_a = 100 N·m, ten reference values from 101,0 N·m to 104,0 N·m): a_s from −0,99 % to −3,85 %. Sources of uncertainty (6.1): the components are determined according to 6.2 — with sufficient data statistically for at least 10 specimens of a model of tool, without repetition at each calibration of that model, with systematic review of validity — or taken from manufacturer or third-party data that can be validated and reproduced in the laboratory; example calculations for Type I wrenches — annex A, Type II — annex B. The sample ends at 6.1, so the uncertainty budget and the certificate requirements are not summarized.
STATUS (PKN card, read 16.11.2026). PN-EN ISO 6789-2:2017-03 — English version, Polish title “Narzędzia montażowe do wkrętów i nakrętek -- Ręczne narzędzia dynamometryczne -- Część 2: Wymagania dotyczące kalibracji i wyznaczania niepewności pomiarów”, published 30-03-2017, 54 pages, KT 236 for Fasteners and Assembly Tools, implements EN ISO 6789-2:2017 and ISO 6789-2:2017 [IDT], replaces PN-EN ISO 6789:2009 (Polish version); the card gives no withdrawal date. iTeh catalogue (16.11.2026): ISO 6789-2:2017 — stage “International Standard to be revised” (90.92); European draft prEN ISO 6789-2 — stage “closure of enquiry” (40.60), planned publication 14-12-2026; its abstract adds minimum requirements for a calibration certificate to this standard.
IN ACCREDITATION SCOPES (scope database of the service, read 16.11.2026, after loading the scopes of this batch). PN-EN ISO 6789-2 appears at 10 Polish calibration laboratories (16 records) — in the 2017 or 2017-03 edition, once with the note “annex C” alongside BS 7882:2017 — and at 1 Czech laboratory.
Method principle
The tool is calibrated by the measurement method of ISO 6789-1 in a calibration system whose torque measurement device has, at each target value, a relative uncertainty interval W′md not greater than 1/4 of the expected interval of the tool W′. For each target value X_a and reference value X_r the relative measurement error a_s = (X_a − X_r) × 100 / X_r and its mean are calculated, and the uncertainty is determined from the budget of components (repeatability, reproducibility, geometry of the output drive and interface, force loading point, resolution, device) with k = 2.
Applications
- Calibration of hand torque wrenches and screwdrivers of Type I and II with determination of the measurement uncertainty (ISO 6789-2:2017, clause 1)
- Calibration of the torque measurement device where W′md is not provided by a traceable calibration certificate (clause 1, annex C)
- Calibration during use: by default every 12 months or 5 000 cycles and after overload, repair and improper handling (4.1)
- In accreditation scopes: 10 Polish calibration laboratories and 1 Czech
Key parameters
| Parameter | Value |
|---|---|
| STATUS (PKN card, 16.11.2026) | PN-EN ISO 6789-2:2017-03 (English version, published 30-03-2017); implements EN ISO 6789-2:2017 and ISO 6789-2:2017; replaces PN-EN ISO 6789:2009 |
| Interval between calibrations of the tool (4.1) | according to the control procedure; without it by default 12 months or 5 000 cycles (whichever occurs first), from first use |
| Torque measurement device (4.3) | W′md ≤ 1/4 W′ at each target value; certificate from a laboratory according to ISO/IEC 17025; without a procedure — at most every 24 months; after overload > 20 % of T_E — recalibration |
| Relative measurement error (5.1, formula 1) | a_s = (X_a − X_r) × 100 / X_r; mean of n measurements — formula 2 |
| Example 1 (5.2.1) | X_a = 100 N·m; X_r = 104,0 / 96,5 / 102,6 / 99,0 / 101,0 N·m → a_s = −3,85 / +3,63 / −2,53 / +1,01 / −0,99 % |
| Coverage factor (table 1) | k = 2 — confidence level approx. 95 % |
| Zero indications (table 1) | I₀ — 30 s after preload; I_z — 30 s after unloading |
| Statistical data of a model (6.1) | at least 10 specimens of a model of tool; systematic review of validity |
| Coverage in accreditation scopes (read 16.11.2026) | 10 Polish calibration laboratories (16 records); 1 Czech |
Standard
- Standard number
- PN-EN ISO 6789-2:2017-03 (English version; implements EN ISO 6789-2:2017 and ISO 6789-2:2017 [IDT])
- Title (PL)
- Narzędzia montażowe do wkrętów i nakrętek — Ręczne narzędzia dynamometryczne — Część 2: Wymagania dotyczące kalibracji i wyznaczania niepewności pomiarów
- Title (EN)
- Assembly tools for screws and nuts — Hand torque tools — Part 2: Requirements for calibration and determination of measurement uncertainty
Step-by-step procedure
-
Need for calibration
Interval elapsed (control procedure or 12 months / 5 000 cycles), overload, repair or improper handling (4.1).
-
Calibration system
Choice of a system suited to the range of the tool; check of W′md ≤ 1/4 W′ at each target value and of the validity of the device certificate (4.3).
-
Measurements
Measurement and recording of the calibration values according to ISO 6789-1:2017, 6.5 — measurement method of clause 6 of part 1 (4.2, 5.1).
-
Measurement error
Calculation of a_s for each measurement (formula 1) and of the mean at each calibration torque (formula 2) (5.1–5.2).
-
Uncertainty
Determination of the uncertainty components according to 6.2 or from validated manufacturer or third-party data; examples — annexes A and B (6.1).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Calibration system (3.1.3) | Measurement device with a loading system — the standard for the tool; suited to the range of the tool (4.3) | — |
| Torque measurement device (3.1.4, 4.3) | Mechanical or with an electronic transducer and display; W′md ≤ 1/4 W′, valid calibration certificate | — |
| Reference measurement standard (3.1.5) | For calibrating the torque measurement device according to annex C when there is no certificate giving W′md | — |
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN ISO 6789-2:2017-03 (English version; implements EN ISO 6789-2:2017 and ISO 6789-2:2017 [IDT]) — “Assembly tools for screws and nuts — Hand torque tools — Part 2: Requirements for calibration and determination of measurement uncertainty”.
How does this method work?
The tool is calibrated by the measurement method of ISO 6789-1 in a calibration system whose torque measurement device has, at each target value, a relative uncertainty interval W′md not greater than 1/4 of the expected interval of the tool W′. For each target value X_a and reference value X_r the relative measurement error a_s = (X_a − X_r) × 100 / X_r and its mean are calculated, and the uncertainty is determined from the budget of components (repeatability, reproducibility, geometry of the output drive and interface, force loading point, resolution, device) with k = 2.
What is the measuring range and accuracy?
STATUS (PKN card, 16.11.2026): PN-EN ISO 6789-2:2017-03 (English version, published 30-03-2017); implements EN ISO 6789-2:2017 and ISO 6789-2:2017; replaces PN-EN ISO 6789:2009; Interval between calibrations of the tool (4.1): according to the control procedure; without it by default 12 months or 5 000 cycles (whichever occurs first), from first use; Torque measurement device (4.3): W′md ≤ 1/4 W′ at each target value; certificate from a laboratory according to ISO/IEC 17025; without a procedure — at most every 24 months; after overload > 20 % of T_E — recalibration; Relative measurement error (5.1, formula 1): a_s = (X_a − X_r) × 100 / X_r; mean of n measurements — formula 2.
How long does the test take?
The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Calibration system (3.1.3), Torque measurement device (3.1.4, 4.3), Reference measurement standard (3.1.5).
Where is this test used?
Calibration of hand torque wrenches and screwdrivers of Type I and II with determination of the measurement uncertainty (ISO 6789-2:2017, clause 1); Calibration of the torque measurement device where W′md is not provided by a traceable calibration certificate (clause 1, annex C); Calibration during use: by default every 12 months or 5 000 cycles and after overload, repair and improper handling (4.1); In accreditation scopes: 10 Polish calibration laboratories and 1 Czech.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 6789-2.