🌡️ Industrial platinum resistance thermometers and platinum temperature sensors (−200 °C to +850 °C) — resistance versus temperature relationship (A, B, C), tolerance classes AA, A, B, C and W/F, marking, measuring current, connecting wire configuration, production and type tests; PN-EN IEC 60751

Metrology PN-EN IEC 60751

In short

Requirements for industrial platinum resistance thermometers and platinum temperature sensors whose resistance is a defined function of temperature in the range from −200 °C to +850 °C. The test is performed according to PN-EN IEC 60751:2022-11 (English version; implements EN IEC 60751:2022 and IEC 60751:2022 [IDT]); Temperature range (clause 1): −200 °C to +850 °C; ITS-90; α = 3,851 · 10⁻³ °C⁻¹. The procedure comprises 4 steps; it is used for: Industrial platinum resistance thermometers and platinum resistors for the range from −200 °C to +850 °C, also for a part of this range (IEC 60751:2022, clause 1), Conversion of measured resistance to temperature (Pt100, Pt500, Pt1000 and other R₀) according to the equations of 4.2 and the tables of Annex A, Checking conformity with the tolerance class and the marking of a thermometer (tables 1 and 2, 5.2.3.3).

At a glance

  • Standard: PN-EN IEC 60751:2022-11 (English version; implements EN IEC 60751:2022 and IEC 60751:2022 [IDT])
  • Category: Metrology
  • Procedure steps: 4
  • STATUS (read 17.11.2026): PN-EN IEC 60751:2022-11 (English version, published 09-11-2022); implements EN IEC 60751:2022 and IEC 60751:2022; replaces PN-EN 60751:2009
  • Temperature range (clause 1): −200 °C to +850 °C; ITS-90; α = 3,851 · 10⁻³ °C⁻¹
  • Equation −200 °C to 0 °C (4.2): R_t = R₀ [1 + At + Bt² + C(t − 100 °C) t³]

Overview

WHAT THE STANDARD COVERS. According to clause 1 of IEC 60751:2022 the standard specifies the requirements — in addition to the resistance versus temperature relationship — for industrial platinum resistance thermometers (“thermometers”) and industrial platinum resistance temperature sensors (“platinum resistors”) whose electrical resistance is a defined function of temperature. Temperatures are given in the International Temperature Scale of 1990 (ITS-90). It applies to resistors with the temperature coefficient α = (R₁₀₀ − R₀)/(R₀ · 100 °C), conventionally written as α = 3,851 · 10⁻³ °C⁻¹, and to resistors and thermometers for the range from −200 °C to +850 °C in different tolerance classes (also for a part of this range). Where the resistance versus temperature relationship is to have an uncertainty below 0,1 °C — possible only with exceptionally stable, individually calibrated items — a more complex interpolation equation is needed, which is outside the scope of the standard.

ACCORDING TO THE TEXT OF IEC 60751:2022 (publisher's sample in English: foreword, clauses 1–5 and 6.1 up to 6.1.3). The standard was prepared by IEC subcommittee SC 65B (Measurement and control devices) of TC 65; the third edition replaces the second edition of 2008 and is a technical revision. Main changes according to the foreword: the formula of the resistance versus temperature relationship becomes the specification and the numerical table ceases to be one (it is now the informative Annex A), a clause “Compliance and requirements” is added, the tolerance acceptance test is modified, an expanded marking system for special ranges of validity is introduced, the vibration test method is revised and a “cold seal” type test is added. Normative references: IEC 60068-2-6 (sinusoidal vibration), IEC 61152 (dimensions of metal-sheathed thermometer elements) and IEC 61515:2016. The definitions (clause 3) cover, among others, the platinum resistor, the thermometer (PRT), the nominal resistance R₀ (a resistor with R₀ = 100 Ω is called a Pt100), the minimum immersion depth (change from full immersion not more than 0,1 °C), the tolerance Δt(t), self-heating and its coefficient (°C/mW), the thermal response time, hysteresis and the expanded uncertainty with coverage factor k = 2 (confidence level 95 %).

RESISTANCE VERSUS TEMPERATURE (4.2–4.3). For the range from −200 °C to 0 °C: R_t = R₀ [1 + At + Bt² + C(t − 100 °C) t³]; for the range from 0 °C to 850 °C: R_t = R₀ (1 + At + Bt²); constants: A = 3,908 3 × 10⁻³ °C⁻¹, B = −5,775 × 10⁻⁷ °C⁻², C = −4,183 × 10⁻¹² °C⁻⁴. Tables A.1 and A.2 (for R₀ = 100 Ω) follow from these equations and are applied to any R₀ after multiplication by R₀/100 Ω; R₀ = 100 Ω is used most often, 10 Ω, 500 Ω and 1000 Ω frequently as well. With a two-wire connection the resistance of the leads between the connecting point of the resistor and the terminals shall be subtracted from the measured value (4.1).

TOLERANCE CLASSES (5.2). Platinum resistors (table 1): wire-wound W 0,1 (−100 °C to +350 °C), W 0,15 (−100 to +450), W 0,3 (−196 to +660), W 0,6 (−196 to +660); film F 0,1 (0 to +150), F 0,15 (−30 to +300), F 0,3 (−50 to +500), F 0,6 (−50 to +600); tolerances ±(0,1 + 0,001 7 |t|), ±(0,15 + 0,002 |t|), ±(0,3 + 0,005 |t|), ±(0,6 + 0,01 |t|) °C respectively. Thermometers (table 2): class AA — from −50 °C to +250 °C for wire-wound resistors and from 0 °C to +150 °C for film resistors, ±(0,1 + 0,0017 |t|); A — from −100 to +450 and from −30 to +300, ±(0,15 + 0,002 |t|); B — from −196 to +600 and from −50 to +500, ±(0,3 + 0,005 |t|); C — from −196 to +600 and from −50 to +600, ±(0,6 + 0,01 |t|). A class without a specified range of validity is not permitted; tolerances and ranges other than in the tables are agreed between supplier and user, and the thermometer is then marked with the suffix “-sp” (5.2.3.2). The marking (5.2.3.3) gives the number of resistors, the nominal resistance, the class, the resistor type, the wire configuration and the range of validity, e.g. “2 × Pt100 / (2/3B) –F-sp / 3 / −50 / +250” or “1 × Pt100 / AA-W-sp / 4 / −50 / +300”.

CURRENT, SUPPLY, WIRES (5.3–5.5). The measuring current is limited so that self-heating under the conditions of 6.4.3 does not exceed 25 % of the tolerance of the declared class; for a 100 Ω wire-wound resistor usually not more than 1 mA. Resistors and thermometers shall be suitable for systems supplied with direct current or alternating current at frequencies up to 100 Hz. Thermometers of a class better than B shall have a three- or four-wire configuration (four-wire recommended).

TESTS (6.1). Categories: routine production tests (on every item; they may be replaced by a sampling test with documented statistical control, informing the user), type tests (on a resistor or thermometer of each design and temperature range, together with the routine tests) and additional type tests. Further clauses according to the table of contents: tolerance acceptance test, insulation resistance, sheath integrity, dimensions, stability at the upper temperature limit, self-heating, response time, thermoelectric effect, temperature cycling, hysteresis, minimum immersion depth, capacitance, inductance, dielectric strength, vibration, drop and “cold seal”, information from the supplier (clause 7) and resistance tables (Annex A) — the sample does not contain their text, so the conditions of these tests are not given.

STATUS (read 17.11.2026). PKN card: PN-EN IEC 60751:2022-11 — English version, title “Platynowe czujniki przemysłowych termometrów rezystancyjnych i platynowe czujniki temperatury”, published 09-11-2022, 34 pages, KT 51 for Industrial Measurements of Non-electrical Quantities, implements EN IEC 60751:2022 and IEC 60751:2022 [IDT], replaces PN-EN 60751:2009; the card gives no withdrawal date. According to the European foreword EN IEC 60751:2022 was approved by CENELEC on 2022-03-03 (date of publication of the national standard — dop 2022-12-03, date of withdrawal of conflicting standards — dow 2025-03-03); it supersedes EN 60751:2008. iTeh catalogue: EN IEC 60751:2022 — published 04-03-2022, stage 90.93 (decision to confirm).

Method principle

The resistance of platinum increases with temperature according to the equation R_t = R₀ (1 + At + Bt²) for 0 °C to 850 °C (below 0 °C with the additional term C(t − 100 °C) t³), with the constants A, B, C from the standard. The thermometer indicates temperature by measuring the resistance of the platinum resistor (e.g. Pt100, R₀ = 100 Ω) and converting it with this formula; conformity with a tolerance class (AA, A, B, C for thermometers, W/F for resistors) applies only within the range of validity assigned to the class.

Applications

Key parameters

ParameterValue
STATUS (read 17.11.2026)PN-EN IEC 60751:2022-11 (English version, published 09-11-2022); implements EN IEC 60751:2022 and IEC 60751:2022; replaces PN-EN 60751:2009
Temperature range (clause 1)−200 °C to +850 °C; ITS-90; α = 3,851 · 10⁻³ °C⁻¹
Equation −200 °C to 0 °C (4.2)R_t = R₀ [1 + At + Bt² + C(t − 100 °C) t³]
Equation 0 °C to 850 °C (4.2)R_t = R₀ (1 + At + Bt²)
Constants (4.2)A = 3,908 3 × 10⁻³ °C⁻¹; B = −5,775 × 10⁻⁷ °C⁻²; C = −4,183 × 10⁻¹² °C⁻⁴
Thermometer classes (table 2)AA ±(0,1 + 0,0017 |t|); A ±(0,15 + 0,002 |t|); B ±(0,3 + 0,005 |t|); C ±(0,6 + 0,01 |t|) °C
Resistor classes (table 1)W 0,1 / F 0,1 … W 0,6 / F 0,6 — the same tolerance formulas, different ranges of validity for wire-wound (W) and film (F) resistors
Measuring current (5.3)self-heating ≤ 25 % of the class tolerance; usually ≤ 1 mA for a 100 Ω wire-wound resistor
Supply and wires (5.4, 5.5)DC or AC up to 100 Hz; class better than B — 3- or 4-wire configuration (4-wire recommended)
Minimum immersion depth (3.6)change from full immersion ≤ 0,1 °C

Standard

Standard number
PN-EN IEC 60751:2022-11 (English version; implements EN IEC 60751:2022 and IEC 60751:2022 [IDT])
Title (PL)
Platynowe czujniki przemysłowych termometrów rezystancyjnych i platynowe czujniki temperatury
Title (EN)
Industrial platinum resistance thermometers and platinum temperature sensors

Step-by-step procedure

  1. Identification of the item

    Reading the thermometer marking: number of resistors, R₀, tolerance class, resistor type (W/F, possibly “-sp”), wire configuration and range of validity (5.2.3.3).

  2. Resistance measurement

    Measurement with a current not greater than permitted with regard to self-heating (5.3); with a two-wire configuration the lead resistance is taken into account (4.1).

  3. Conversion to temperature

    Application of the equation of 4.2 (for t < 0 °C with the C term) or of tables A.1/A.2 multiplied by R₀/100 Ω.

  4. Assessment of conformity with the class

    Comparison of the deviation Δt with the class tolerance (table 1 or 2) only within the range of validity of that class; production and type tests — clause 6 of the standard.

Required equipment and apparatus

EquipmentExampleIndicative price
Platinum resistor (3.1)Platinum wire or film embedded in an insulator (usually glass or ceramic) — wire-wound (W) or film (F), e.g. Pt100—
Platinum resistance thermometer PRT (3.2)One or more platinum resistors in a protective sheath with internal connecting wires and terminals; without a separable protection tube (thermowell)—
Resistance measuring instrumentMeasurement in a 2-, 3- or 4-wire configuration (fig. 1); with 2 wires the lead resistance is subtracted (4.1)—

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN IEC 60751:2022-11 (English version; implements EN IEC 60751:2022 and IEC 60751:2022 [IDT]) — “Industrial platinum resistance thermometers and platinum temperature sensors”.

How does this method work?

The resistance of platinum increases with temperature according to the equation R_t = R₀ (1 + At + Bt²) for 0 °C to 850 °C (below 0 °C with the additional term C(t − 100 °C) t³), with the constants A, B, C from the standard. The thermometer indicates temperature by measuring the resistance of the platinum resistor (e.

What is the measuring range and accuracy?

STATUS (read 17.11.2026): PN-EN IEC 60751:2022-11 (English version, published 09-11-2022); implements EN IEC 60751:2022 and IEC 60751:2022; replaces PN-EN 60751:2009; Temperature range (clause 1): −200 °C to +850 °C; ITS-90; α = 3,851 · 10⁻³ °C⁻¹; Equation −200 °C to 0 °C (4.2): R_t = R₀ [1 + At + Bt² + C(t − 100 °C) t³]; Equation 0 °C to 850 °C (4.2): R_t = R₀ (1 + At + Bt²).

How long does the test take?

The procedure comprises 4 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Platinum resistor (3.1), Platinum resistance thermometer PRT (3.2), Resistance measuring instrument.

Where is this test used?

Industrial platinum resistance thermometers and platinum resistors for the range from −200 °C to +850 °C, also for a part of this range (IEC 60751:2022, clause 1); Conversion of measured resistance to temperature (Pt100, Pt500, Pt1000 and other R₀) according to the equations of 4.2 and the tables of Annex A; Checking conformity with the tolerance class and the marking of a thermometer (tables 1 and 2, 5.2.3.3); Selection of the measuring current and of the wire configuration (2-, 3-, 4-wire) when measuring the sensor resistance (5.3, 5.5).

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN IEC 60751.