📏 Gauge blocks of rectangular cross section (0,5 mm to 1 000 mm) — design and metrological characteristics, calibration grade K and grades 0, 1, 2: length and central length, reference temperature 20 °C, reference orientation, cross section, coupling holes; PN-EN ISO 3650

Metrology PN-EN ISO 3650

In short

The most important design and metrological characteristics of gauge blocks with a rectangular cross section and a nominal length from 0,5 mm to 1 000 mm, with limit deviations and tolerances for calibration grade K and grades 0, 1 and 2. The test is performed according to PN-EN ISO 3650:2000 (Polish version; implements EN ISO 3650:1998 and ISO 3650:1998 [IDT]); Range of nominal lengths (clause 1): 0,5 mm ≤ lₙ ≤ 1 000 mm; calibration grade K, grades 0, 1, 2. The procedure comprises 4 steps; it is used for: Gauge blocks with a rectangular cross section and a nominal length from 0,5 mm to 1 000 mm — calibration grade K and grades 0, 1, 2 (ISO 3650:1998, clause 1), Length standards in the calibration of measuring instruments — among others in testing callipers according to ISO 13385-1 (5.4 of that standard), Composite assemblies of blocks joined by wringing and blocks over 100 mm with coupling holes (3.1, 6.1).

At a glance

  • Standard: PN-EN ISO 3650:2000 (Polish version; implements EN ISO 3650:1998 and ISO 3650:1998 [IDT])
  • Category: Metrology
  • Procedure steps: 4
  • STATUS (PKN card, 16.11.2026): PN-EN ISO 3650:2000 (Polish version, published 11-08-2000); implements EN ISO 3650:1998 and ISO 3650:1998; replaces PN-M-53101:1983
  • Range of nominal lengths (clause 1): 0,5 mm ≤ lₙ ≤ 1 000 mm; calibration grade K, grades 0, 1, 2
  • Reference conditions (5.3): 20 °C (ISO 1); 101 325 Pa = 1,013 25 bar

Overview

WHAT THE STANDARD COVERS. According to clause 1 of ISO 3650:1998 the standard specifies the most important design and metrological characteristics of gauge blocks with a rectangular cross section and a nominal length lₙ from 0,5 mm to 1 000 mm; limit deviations and tolerances are stated for calibration grade K and for grades 0, 1 and 2 intended for various measuring purposes. The introduction explains that gauge blocks are length standards representing specified fractions of the metre, offered in several grades depending on the application, and that the basis for their use as standards is calibration — measurement of the length at a specified point of the measuring face and evaluation of the measurement uncertainty. The PKN card of PN-EN ISO 3650:2000 gives the same scope.

ACCORDING TO THE TEXT OF ISO 3650:1998 (publisher's sample in English: foreword, introduction, clauses 1–5 and the beginning of clause 6; in French — up to 6.2.1). The standard was prepared by ISO/TC 213 (Dimensional and geometrical product specifications and verification); the second edition replaced the first (ISO 3650:1978) after technical revision. It is a general GPS standard influencing chain link 6 (calibration requirements — calibration standards) of the chain of standards on size and distance. Normative references: ISO 1:1975 (reference temperature), ISO 1101, ISO 6507-1:1997 (Vickers hardness), ISO 14253-1:1998 and the VIM. Definitions (clause 3): gauge block — material measure of rectangular section made of wear-resistant material, with one pair of planar, mutually parallel measuring faces which can be wrung to the measuring faces of other gauge blocks (composite assemblies) or to similarly finished auxiliary plates; length of a gauge block l — perpendicular distance from a particular point of one measuring face to the planar surface of an auxiliary plate of the same material and surface texture to which the other measuring face is wrung (it includes the effect of one-face wringing); central length l_c — the length at the centre point of the free measuring face; deviation of the length from nominal e = l − lₙ; deviation from flatness f_d; variation in length v = l_max − l_min (the sum of the deviations f_o and f_u from the central length); wringing — the property of measuring faces to adhere under the action of molecular forces. Clause 4 names the faces (left and right measuring face, marked measuring face, side faces, marked side face — separately for lₙ < 6 mm and lₙ ≥ 6 mm).

BASIS OF MEASUREMENT (clause 5). The unit of length is the metre — the path travelled by light in vacuum in 1/299 792 458 s (17th CGPM, 1983), realized by wavelength standards recommended by the CIPM. The length of a gauge block is traceable to national or international standards if the result can be related by an unbroken chain of comparisons with stated uncertainties to a gauge block calibrated by interferometry using appropriate wavelength standards. The nominal length and the measured lengths apply at 20 °C (ISO 1) and the standard pressure 101 325 Pa = 1,013 25 bar; the effect of pressure deviations may be ignored under normal atmospheric conditions. Reference orientation: blocks up to and including 100 mm — vertical, with the measuring faces horizontal; blocks over 100 mm — horizontal, supported on one of the narrow side faces without additional stress, on supports at 0,211 times the nominal length from the ends; when measured by interferometry in horizontal orientation, the weight of the wrung auxiliary plate shall be compensated.

DIMENSIONS AND MATERIAL (clause 6). Cross section (table 1): dimension a = 30 mm for 0,5 mm ≤ lₙ ≤ 10 mm and 35 mm for 10 mm < lₙ ≤ 1 000 mm, limit deviations 0/−0,3 mm; dimension b = 9 mm, limit deviations −0,05/−0,20 mm. Blocks over 100 mm may have coupling holes dimensioned and located as in figure 5 (hole ∅ 10 mm with deviation +0,2/0 mm, axis 25 mm from the measuring face); grade K blocks shall not be combined with coupling devices. Material (6.2.1): high-grade steel or another similarly wear-resistant material giving surfaces that wring readily and stable for length within table 2. Further clauses according to the table of contents: coefficient of thermal expansion, hardness and dimensional stability (6.2.2–6.2.4), marking (6.3), metrological requirements — flatness tolerance, measuring faces, side faces (flatness, parallelism, perpendicularity, edges) (clause 7) and calibration of gauge blocks — wringing test, measurement by interferometry and by comparison, calibration certificates (clause 8); the sample does not contain their text, so tolerance values are not given.

CORRIGENDUM. ISO 3650:1998/Cor.1:2008 (technical corrigendum in the French version) corrects in 7.4.1 and 7.4.2 the unit in the formulas for the flatness and parallelism tolerances of the side faces: “40 mm” to “40 µm” and “80 mm” to “80 µm”. EN ISO 3650:1998/AC:2009 takes over this corrigendum for the official French version of the EN only (effective from 25 March 2009).

STATUS (PKN card, read 16.11.2026). PN-EN ISO 3650:2000 — Polish version, title “Specyfikacje geometrii wyrobów (GPS) -- Wzorce długości -- Płytki wzorcowe”, published 11-08-2000, 24 pages, KT 48 for Fundamentals of Machine Design, implements EN ISO 3650:1998 and ISO 3650:1998 [IDT], replaces PN-M-53101:1983; the card gives no withdrawal date. iTeh catalogue (16.11.2026): ISO 3650:1998 — published 20-12-1998, stage “International Standard confirmed” (90.93).

IN ACCREDITATION SCOPES (scope database of the service, read 16.11.2026, after loading the scopes of this batch). PN-EN ISO 3650 appears at 13 Polish calibration laboratories (20 records) — always in the 2000 edition (two records as “2000P”), most often as the document on which an internal procedure for calibrating gauge blocks is based; it does not appear in Czech scopes.

Method principle

A gauge block represents length as the distance between a point of one measuring face and the planar surface of an auxiliary plate to which the other face is wrung; at the centre of the free face this is the central length l_c. The length refers to 20 °C, 101 325 Pa and a specified orientation (up to 100 mm vertical, above that horizontal on supports at 0,211 lₙ from the ends), and its traceability to the metre is ensured by an unbroken chain of comparisons with a block calibrated by interferometry. Calibration grade K and grades 0, 1, 2 differ in limit deviations and tolerances.

Applications

Key parameters

ParameterValue
STATUS (PKN card, 16.11.2026)PN-EN ISO 3650:2000 (Polish version, published 11-08-2000); implements EN ISO 3650:1998 and ISO 3650:1998; replaces PN-M-53101:1983
Range of nominal lengths (clause 1)0,5 mm ≤ lₙ ≤ 1 000 mm; calibration grade K, grades 0, 1, 2
Reference conditions (5.3)20 °C (ISO 1); 101 325 Pa = 1,013 25 bar
Reference orientation (5.4)up to 100 mm — vertical; over 100 mm — horizontal on a narrow side face, supports at 0,211 lₙ from the ends
Cross section a (table 1)30 mm (0,5 ≤ lₙ ≤ 10 mm), 35 mm (10 < lₙ ≤ 1 000 mm); deviations 0/−0,3 mm
Cross section b (table 1)9 mm; deviations −0,05/−0,20 mm
Coupling holes (6.1, fig. 5)blocks > 100 mm: ∅ 10 mm (+0,2/0), axis 25 mm from the measuring face; grade K — no coupling devices
Quantities (clause 3)l, l_c, e = l − lₙ, f_d, v = l_max − l_min
Corrigendum (Cor.1:2008, French version)7.4.1 and 7.4.2: “40 mm” → “40 µm”, “80 mm” → “80 µm”
Coverage in accreditation scopes (read 16.11.2026)13 Polish calibration laboratories (20 records)

Standard

Standard number
PN-EN ISO 3650:2000 (Polish version; implements EN ISO 3650:1998 and ISO 3650:1998 [IDT])
Title (PL)
Specyfikacje geometrii wyrobów (GPS) — Wzorce długości — Płytki wzorcowe
Title (EN)
Geometrical Product Specifications (GPS) — Length standards — Gauge blocks

Step-by-step procedure

  1. Grade and dimensions

    Determination of the grade (K, 0, 1, 2) and check of the cross-section dimensions according to table 1; blocks over 100 mm — coupling holes according to figure 5 (6.1).

  2. Reference conditions

    Length referred to 20 °C and 101 325 Pa (5.3); orientation of the block according to 5.4 — vertical up to 100 mm, horizontal on supports at 0,211 lₙ above 100 mm.

  3. Definition of length

    Length of the block measured relative to a wrung auxiliary plate; central length l_c at the centre of the free measuring face; deviation e = l − lₙ and variation in length v (clause 3).

  4. Traceability

    Result related by an unbroken chain of comparisons to a block calibrated by interferometry (5.2); interferometric and comparison calibration — clause 8 of the standard.

Required equipment and apparatus

EquipmentExampleIndicative price
Auxiliary plate (3.2)Of the same material and surface texture as the block — for wringing the other measuring face in the definition of length—
Interferometer with wavelength standards (5.2)Calibration of blocks at the top of the traceability chain — wavelength standards recommended by the CIPM—
Supports for blocks over 100 mm (5.4)Two supports at 0,211 times the nominal length from the ends of a block lying on a narrow side face—

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN ISO 3650:2000 (Polish version; implements EN ISO 3650:1998 and ISO 3650:1998 [IDT]) — “Geometrical Product Specifications (GPS) — Length standards — Gauge blocks”.

How does this method work?

A gauge block represents length as the distance between a point of one measuring face and the planar surface of an auxiliary plate to which the other face is wrung; at the centre of the free face this is the central length l_c. The length refers to 20 °C, 101 325 Pa and a specified orientation (up to 100 mm vertical, above that horizontal on supports at 0,211 lₙ from the ends), and its traceability to the metre is ensured by an unbroken chain of comparisons with a block calibrated by interferometry.

What is the measuring range and accuracy?

STATUS (PKN card, 16.11.2026): PN-EN ISO 3650:2000 (Polish version, published 11-08-2000); implements EN ISO 3650:1998 and ISO 3650:1998; replaces PN-M-53101:1983; Range of nominal lengths (clause 1): 0,5 mm ≤ lₙ ≤ 1 000 mm; calibration grade K, grades 0, 1, 2; Reference conditions (5.3): 20 °C (ISO 1); 101 325 Pa = 1,013 25 bar; Reference orientation (5.4): up to 100 mm — vertical; over 100 mm — horizontal on a narrow side face, supports at 0,211 lₙ from the ends.

How long does the test take?

The procedure comprises 4 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Auxiliary plate (3.2), Interferometer with wavelength standards (5.2), Supports for blocks over 100 mm (5.4).

Where is this test used?

Gauge blocks with a rectangular cross section and a nominal length from 0,5 mm to 1 000 mm — calibration grade K and grades 0, 1, 2 (ISO 3650:1998, clause 1); Length standards in the calibration of measuring instruments — among others in testing callipers according to ISO 13385-1 (5.4 of that standard); Composite assemblies of blocks joined by wringing and blocks over 100 mm with coupling holes (3.1, 6.1); In accreditation scopes: 13 Polish calibration laboratories.

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 3650.

🔍 Find a laboratory performing this test