🍎 pH of fruit and vegetable juices and related products — potentiometric measurement at (20 ± 2) °C after two-point calibration of the pH meter with buffers pH 6.88 and 3.57, sample containing carbon dioxide degassed, result to two decimal places; PN-EN 1132:1999 (withdrawn standard)
In short
Method for determining the pH of fruit and vegetable juices and related products: the pH of the sample is measured potentiometrically, normally at (20 ± 2) °C, with a pH meter calibrated at two points with buffer solutions. The test is performed according to PN-EN 1132:1999; STATUS (as of 18.11.2026): PN-EN 1132:1999 withdrawn 09-04-2024, card without a "Replaced by" field; NS-EN 1132:1994 "Withdrawn" 23.02.2024 (Standard Norge); SIST EN 1132:1996 withdrawn 09.03.2026 (iTeh). The procedure comprises 4 steps; it is used for: Fruit and vegetable juices and related products — pH (EN 1132:1994, clause 1; PKN card, Standard Norge catalogue), Juices containing carbon dioxide — after degassing by shaking, vacuum or ultrasound (7.2).
At a glance
- Standard: PN-EN 1132:1999
- Category: Food chemistry
- Procedure steps: 4
- STATUS (as of 18.11.2026): PN-EN 1132:1999 withdrawn 09-04-2024, card without a "Replaced by" field; NS-EN 1132:1994 "Withdrawn" 23.02.2024 (Standard Norge); SIST EN 1132:1996 withdrawn 09.03.2026 (iTeh)
- Edition (from the PKN card): PN-EN 1132:1999, Polish version; published 23-09-1999, 8 pages, KT 38, ICS 67.160.20; introduces EN 1132:1994 [IDT]; replaced PN-A-75101-06:1990
- Buffers (5.2–5.4): pH 6.88 at 20 °C (3.40 g KH₂PO₄ + 3.55 g Na₂HPO₄ per 1000 ml, stable about one month); pH 3.57 at 20 °C (saturated potassium hydrogen tartrate, with thymol about 2 months); or commercial
Overview
STATUS (as of 18.11.2026). The card of PN-EN 1132:1999 (Polish version) has the header "Withdrawn": published 23-09-1999, "Withdrawal date: 09-04-2024", 8 pages, KT 38 Fruit and Vegetable Products, "Introduces: EN 1132:1994 [IDT]", "Replaces: PN-A-75101-06:1990 - Polish version", ICS 67.160.20; there is no "Replaced by" field. The predecessor, PN-A-75101-06:1990 "Fruit and vegetable products — Preparation of samples and methods of physicochemical testing — Determination of pH by the potentiometric method" (published 25-05-1990, 2 pages, ICS 67.080.01), has the header "Withdrawn and replaced by PN-EN 1132:1999 - Polish version" and the withdrawal date 02-12-2011, and its amendment PN-A-75101-06:1990/Az1:2002 is listed on the card as an "Additional element". The PKN search for "PN-EN 1132" returns this standard and two editions of the unrelated PN-EN ISO 11132 (sensory analysis). With other publishers: Standard Norge — NS-EN 1132:1994 "Withdrawn" on 23 February 2024, with no replacing standard named; iTeh — the Slovenian edition SIST EN 1132:1996 withdrawn on 09.03.2026. We do not name a standard replacing PN-EN 1132. The description below follows the text of EN 1132:1994, which PN-EN 1132:1999 introduces unchanged — the sample of SIST EN 1132:1996 in the iTeh catalogue contains pages 1–7 of the standard (clauses 1–10 and Annex A); Annex B with the interlaboratory data is not included.
WHAT THE STANDARD COVERS. Clause 1: method for the determination of the pH of fruit and vegetable juices and related products. Definition (clause 3): pH is the negative common logarithm of the concentration of hydrogen ions in moles per litre of solution. Principle (clause 4): the pH of the sample is measured potentiometrically. According to the card of the predecessor, PN-A-75101-06:1990, the principle of the method is the measurement of the potential difference between two electrodes immersed in the test solution.
REAGENTS AND APPARATUS (clauses 5–6). Reagents of recognized analytical grade and only degassed water of at least grade 3 according to ISO 3696:1987. Buffer solution pH 6.88 at 20 °C: anhydrous KH₂PO₄ and Na₂HPO₄ are dried at (120 ± 10) °C to constant mass, after cooling in a desiccator 3.40 g KH₂PO₄ and 3.55 g Na₂HPO₄ are weighed, dissolved in about 800 ml of water and made up to 1000 ml at 20 °C — stable for about one month. Buffer solution pH 3.57 at 20 °C: 30 g of potassium hydrogen tartrate in about 1 l of water, shaken vigorously to saturation, the undissolved residue filtered off; the solution contains at least 5.7 g of the tartrate; preferably prepared on the day of use, since it deteriorates — with 0.05 g of thymol per 100 ml it is stable for about 2 months at 20 °C. Commercial buffer solutions of similar pH may be used instead, following the manufacturer's instructions. Apparatus: pH meter accurate to at least 0.05 pH units, glass electrode and reference electrode (for example calomel) or a combination glass electrode, desiccator with drying agent.
PROCEDURE (clause 7). The pH meter and electrodes are calibrated and serviced according to the manufacturer's instructions. The pH of the sample is normally measured at (20 ± 2) °C; with a pH meter with temperature compensation the measurement may be made at another temperature, but always between 10 and 30 °C and according to the manufacturer's instructions. Before each measurement and each calibration step the electrode is rinsed well with water and adhering water removed with filter paper; calibration and measurement are carried out while gently stirring the solution or sample. A sample containing appreciable quantities of carbon dioxide is degassed by shaking it in a stoppered conical flask (removing the stopper from time to time) or by vacuum or ultrasound until no more gas evolves. Two-point calibration with buffers pH 6.88 and 3.57 (or commercial ones) according to the manufacturer's instructions, then measurement of the sample pH.
RESULT, PRECISION AND TEST REPORT (clauses 8–10). The pH is reported to two decimal places. Repeatability r = 0.03, reproducibility R = 0.12 (pH units) — the difference between two single results exceeds them in not more than 5 % of cases; the standard notes that values from the interlaboratory test (Annex B) may not apply to other ranges and matrices. The test report contains the information identifying the sample (kind, origin, designation), a reference to the standard, the date and type of sampling (if possible), the dates of receipt and of test, the results and units, whether repeatability has been verified, particular points observed and operations not specified or optional. Annex A (informative) lists method No 11 (1989) of the International Federation of Fruit Juice Producers (IFU) and method L 31.00-2 (1980) of the official collection of food testing methods under § 35 of the German Food Act (LMBG).
WHERE A SEEMINGLY CORRECT RESULT IS EASY TO GET. First — subject: the scope of the standard is fruit and vegetable juices and related products; the standard does not describe preparation of solid samples (fruit, vegetables, products). Second — carbon dioxide: the standard requires a sample with appreciable CO₂ to be degassed before measurement until no more gas evolves; a measurement without this step is performed outside the standard. Third — temperature: without temperature compensation the measurement is to be made at (20 ± 2) °C, and with compensation only within 10–30 °C. Fourth — calibration: buffers 6.88 and 3.57; the tartrate buffer has to be prepared fresh or stabilized with thymol. Fifth — status: the standard has been withdrawn since 09.04.2024 without a named successor, and its predecessor since 02.12.2011.
WHAT WE DO NOT GIVE HERE. The interlaboratory data of Annex B — they are not in the available sample. We did not read IFU method No 11 or method L 31.00-2.
Method principle
The pH of the juice sample is measured potentiometrically with a glass electrode and a reference electrode (or a combination electrode), normally at (20 ± 2) °C, after two-point calibration of the pH meter with buffers pH 6.88 (phosphate) and 3.57 (hydrogen tartrate) or commercial ones of similar pH; a sample with much carbon dioxide is degassed first, and the result is reported to two decimal places (EN 1132:1994, clauses 4–8).
Applications
- Fruit and vegetable juices and related products — pH (EN 1132:1994, clause 1; PKN card, Standard Norge catalogue)
- Juices containing carbon dioxide — after degassing by shaking, vacuum or ultrasound (7.2)
Key parameters
| Parameter | Value |
|---|---|
| STATUS (as of 18.11.2026) | PN-EN 1132:1999 withdrawn 09-04-2024, card without a "Replaced by" field; NS-EN 1132:1994 "Withdrawn" 23.02.2024 (Standard Norge); SIST EN 1132:1996 withdrawn 09.03.2026 (iTeh) |
| Edition (from the PKN card) | PN-EN 1132:1999, Polish version; published 23-09-1999, 8 pages, KT 38, ICS 67.160.20; introduces EN 1132:1994 [IDT]; replaced PN-A-75101-06:1990 |
| Buffers (5.2–5.4) | pH 6.88 at 20 °C (3.40 g KH₂PO₄ + 3.55 g Na₂HPO₄ per 1000 ml, stable about one month); pH 3.57 at 20 °C (saturated potassium hydrogen tartrate, with thymol about 2 months); or commercial |
| Measuring temperature (7.1) | normally (20 ± 2) °C; with temperature compensation 10–30 °C |
| pH meter (6.1) | accuracy at least 0.05 pH units; glass and reference electrode or combination electrode |
| Result and precision (clauses 8–9) | two decimal places; r = 0.03, R = 0.12 |
| Predecessor (PKN cards) | PN-A-75101-06:1990 (pH of fruit and vegetable products) — withdrawn 02-12-2011, replaced by PN-EN 1132:1999 |
Standard
- Standard number
- PN-EN 1132:1999
- Title (PL)
- Soki owocowe i warzywne — Oznaczanie pH
- Title (EN)
- Fruit and vegetable juices — Determination of the pH-value
Step-by-step procedure
-
Sample
A sample with appreciable carbon dioxide is degassed by shaking in a stoppered conical flask (removing the stopper), by vacuum or ultrasound (7.2). PN-EN 1132:1999 withdrawn 09-04-2024 (status as of 18.11.2026).
-
Calibration
Two-point with buffers pH 6.88 and 3.57 or commercial ones, according to the manufacturer's instructions; the electrode is rinsed with water and dried with filter paper before each step; gentle stirring (7.1, 7.3).
-
Measurement
pH of the sample at (20 ± 2) °C, with temperature compensation at 10–30 °C, with gentle stirring (7.1, 7.4).
-
Result
pH to two decimal places; r = 0.03, R = 0.12; the test report states whether repeatability was verified (clauses 8–10).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| pH meter | Accuracy at least 0.05 pH units; calibration and servicing according to the manufacturer's instructions (EN 1132, 6.1, 7.1) | — |
| Electrodes | Glass electrode and reference electrode, for example calomel, or a combination glass electrode (6.2–6.4) | — |
| Desiccator | With drying agent — for cooling the buffer salts after drying (6.5) | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Buffer solution pH 6.88 (20 °C) | — | Anhydrous KH₂PO₄ and Na₂HPO₄ dried at (120 ± 10) °C; 3.40 g and 3.55 g to 1000 ml at 20 °C; stable about one month (EN 1132, 5.2) |
| Buffer solution pH 3.57 (20 °C) | — | 30 g of potassium hydrogen tartrate in about 1 l of water, saturated, filtered; preferably fresh, with 0.05 g thymol per 100 ml stable about 2 months (5.3) |
| Degassed water | — | At least grade 3 according to ISO 3696:1987 (5.1) |
Health and safety (OHS)
- Chemical laboratory safety rules and safety data sheets of the buffer salts
- TECHNICAL NOTE: the scope of the standard is fruit and vegetable juices and related products; a sample containing carbon dioxide has to be degassed before measurement
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN 1132:1999 — “Fruit and vegetable juices — Determination of the pH-value”.
How does this method work?
The pH of the juice sample is measured potentiometrically with a glass electrode and a reference electrode (or a combination electrode), normally at (20 ± 2) °C, after two-point calibration of the pH meter with buffers pH 6.88 (phosphate) and 3.
What is the measuring range and accuracy?
STATUS (as of 18.11.2026): PN-EN 1132:1999 withdrawn 09-04-2024, card without a "Replaced by" field; NS-EN 1132:1994 "Withdrawn" 23.02.2024 (Standard Norge); SIST EN 1132:1996 withdrawn 09.03.2026 (iTeh); Edition (from the PKN card): PN-EN 1132:1999, Polish version; published 23-09-1999, 8 pages, KT 38, ICS 67.160.20; introduces EN 1132:1994 [IDT]; replaced PN-A-75101-06:1990; Buffers (5.2–5.4): pH 6.88 at 20 °C (3.40 g KH₂PO₄ + 3.55 g Na₂HPO₄ per 1000 ml, stable about one month); pH 3.57 at 20 °C (saturated potassium hydrogen tartrate, with thymol about 2 months); or commercial; Measuring temperature (7.1): normally (20 ± 2) °C; with temperature compensation 10–30 °C.
How long does the test take?
The procedure comprises 4 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
pH meter, Electrodes, Desiccator.
Where is this test used?
Fruit and vegetable juices and related products — pH (EN 1132:1994, clause 1; PKN card, Standard Norge catalogue); Juices containing carbon dioxide — after degassing by shaking, vacuum or ultrasound (7.2).
What safety precautions apply?
Chemical laboratory safety rules and safety data sheets of the buffer salts; TECHNICAL NOTE: the scope of the standard is fruit and vegetable juices and related products; a sample containing carbon dioxide has to be degassed before measurement.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 1132.