🍎 pH of all kinds of meat and meat products, including poultry — reference method: combined electrode in the extract of a sample homogenized with ten times its mass of KCl solution 0.1 mol/l, or non-destructive measurement in carcass meat, quarters and muscles, pH meter calibrated with two buffers, result to the nearest 0.05 pH unit; PN-ISO 2917:2001 + Ap1:2002
In short
Reference method for measuring the pH of all kinds of meat and meat products, including poultry: the potential difference is measured between a glass electrode and a reference electrode placed in a sample or a sample extract. The test is performed according to PN-ISO 2917:2001 + Ap1:2002; STATUS (as of 18.11.2026): PN-ISO 2917:2001 current; amendment Ap1:2002 — editorial (adds PN-77/A-82058 in the "instead of" field); ISO 2917:1999 — "Published", stage 90.93 (iTeh). The procedure comprises 6 steps; it is used for: All kinds of meat and meat products, including poultry — pH by the reference method (ISO 2917:1999, clause 1; PKN card), Non-destructive measurements on carcass meat, quarters and muscles, including fresh meat at 0–5 °C with temperature correction (9.4), Prerigor muscle meat — KCl solution with iodoacetic acid to stop glycolysis (5.4).
At a glance
- Standard: PN-ISO 2917:2001 + Ap1:2002
- Category: Food chemistry
- Procedure steps: 6
- STATUS (as of 18.11.2026): PN-ISO 2917:2001 current; amendment Ap1:2002 — editorial (adds PN-77/A-82058 in the "instead of" field); ISO 2917:1999 — "Published", stage 90.93 (iTeh)
- Edition (from the PKN card): PN-ISO 2917:2001, Polish version; published 13-06-2001, 9 pages, KT 93, ICS 67.120.10; introduces ISO 2917:1999 [IDT]; replaces PN-A-82058:1977
- Buffers (5.2): pH 4.00 (potassium hydrogen phthalate), 6.88 (KH₂PO₄ + Na₂HPO₄), 5.45 (citric acid + NaOH) at 20 °C; calibration with two buffers spanning the sample pH
Overview
STATUS (as of 18.11.2026). The card of PN-ISO 2917:2001 (Polish version) has no "Withdrawn" header: published 13-06-2001, 9 pages, KT 93 Meat, Eggs and their Products, "Introduces: ISO 2917:1999 [IDT]", "Replaces: PN-A-82058:1977 - Polish version", ICS 67.120.10, "Additional elements: PN-ISO 2917:2001/Ap1:2002P". The card provides amendment Ap1 (September 2002) free of charge; its entire content is the sentence: "On the title page, in the field 'instead of', PN-77/A-82058 is added" — the amendment changes nothing in the method. At the publisher: ISO 2917:1999 (second edition, 15.12.1999) has in the iTeh catalogue the status "Published" and stage 90.93 ("International Standard confirmed"), and in the Standard Norge catalogue the status "Published" and the note "Revises: ISO 2917:1974". The description below follows the text of ISO 2917:1999, which PN-ISO 2917:2001 introduces unchanged (the publisher's sample in the iTeh catalogue covers clauses 1–12).
WHAT THE STANDARD COVERS. Clause 1: reference method for measuring the pH of all kinds of meat and meat products, including poultry; the method is applicable to products which may be homogenized and to non-destructive measurements on carcass meat, quarters and muscles. By the definition of the standard, the pH of meat and meat products is the result of measurements performed according to this procedure (3.1). Principle (clause 4): the potential difference is measured between a glass electrode and a reference electrode placed in a sample or a sample extract.
REAGENTS AND APPARATUS (clauses 5–6). Water of at least grade 3 according to ISO 3696; for buffers — freshly boiled or scrubbed with carbon-dioxide-free nitrogen. Buffer solutions: commercial ready-to-use with a guaranteed pH accurate to at least 0.01, from commercial dry mixtures, or self-prepared — pH 4.00 at 20 °C (10.21 g of potassium hydrogen phthalate dried at 110–130 °C, to 1000 ml), pH 6.88 at 20 °C (3.40 g KH₂PO₄ and 3.55 g Na₂HPO₄, to 1000 ml; in a refrigerator up to 3 months), pH 5.45 at 20 °C (7.01 g of citric acid monohydrate and 375 ml of NaOH 1.0 mol/l, to 1000 ml). Potassium chloride solution 0.1 mol/l (7.5 g KCl to 1000 ml); for muscle meat in prerigor condition glycolysis is stopped by adding 925 mg of iodoacetic acid per litre and adjusting the pH to 7.0 with NaOH. For cleaning the electrodes — diethyl ether saturated with water and ethanol 95 %. Apparatus: homogenizing equipment (high-speed rotational cutter or mincer with a plate with apertures not exceeding 4.0 mm), pH meter reading to 0.01 pH (without temperature correction — scale for 20 °C), combined electrode with glass and Ag/AgCl or Hg/HgCl₂ reference electrode (the glass electrode may be spherical, conical, cylindrical or needle-shaped; for greasy samples separate electrodes are acceptable), shaft homogenizer at 20 000 min⁻¹, magnetic stirrer.
SAMPLE AND PROCEDURE (clauses 7–9). Sampling is not part of the method (ISO 3100-1 recommended); a representative sample of at least 200 g. Non-destructive measurement: a representative point of the sample is selected. Destructive measurement: the sample is homogenized without its temperature rising above 25 °C (with a mincer — at least twice), kept in an airtight container and analysed as soon as practicable, always within 24 h after homogenization. The pH meter is calibrated with two buffer solutions spanning the expected sample pH, at the temperature of measurement, while stirring with the magnetic stirrer; without temperature correction the buffer shall be at (20 ± 2) °C. Homogenized products: a portion of the sample is homogenized in ten times as much KCl solution with the shaft homogenizer, the electrodes are introduced into the extract, temperature correction is set (without it the extract shall be at (20 ± 2) °C) and, while stirring, the pH is read to 0.01 once constant. Non-destructive measurement: a hole is pierced in the sample with a knife or a sharp pin and the electrode inserted; fresh meat, generally kept at 0–5 °C, requires a pH meter with temperature correction; the measurement is repeated at the same incision and, if useful, at several points whose number depends on the nature and size of the sample. The electrodes are cleaned with cotton wool wetted successively with diethyl ether and ethanol and finally washed with water. If the repeatability limit is to be checked, two single determinations are carried out.
RESULT, PRECISION AND TEST REPORT (clauses 10–12). Non-destructive measurement: the result is the mean of the two values at the same point, reported for each point to the nearest 0.05 pH; with several points — a sketch with the points and their means. Homogenized products: result to the nearest 0.05 pH. Precision from interlaboratory tests according to ISO 5725: repeatability 0.04 pH (standard deviation about 0.014), reproducibility 0.12 pH (about 0.042) — exceeded in not more than 5 % of cases. The test report gives identification of the sample, the sampling method (if known), the test method indicating whether the measurement was destructive or non-destructive, operating details not specified in the standard, and the result (or two results if repeatability was checked).
WHERE A SEEMINGLY CORRECT RESULT IS EASY TO GET. First — subject: the scope of the standard is all kinds of meat and meat products, including poultry; it does not list fish or other products. Second — mode of measurement: measurement in the KCl extract (1 : 10) and directly in the muscle are different conditions; the test report has to state whether the measurement was destructive or non-destructive. Third — temperature: without temperature correction the buffer and extract have to be at (20 ± 2) °C, and fresh meat at 0–5 °C requires a pH meter with correction. Fourth — prerigor meat: without iodoacetic acid in the KCl solution glycolysis continues and the pH changes. Fifth — citing the amendment: Ap1 is editorial, so citing the standard with or without it does not change the method.
WHAT WE DO NOT GIVE HERE. Interlaboratory data beyond the repeatability and reproducibility limits — the standard does not give them.
Method principle
The potential difference is measured between a glass electrode and a reference electrode placed in the extract (sample homogenized with ten times as much KCl solution 0.1 mol/l) or directly in carcass meat, quarters or muscle; the pH meter reading to 0.01 pH is calibrated with two buffers spanning the expected pH at the temperature of measurement, and the result is reported to the nearest 0.05 pH (ISO 2917:1999, clauses 4 and 9–10).
Applications
- All kinds of meat and meat products, including poultry — pH by the reference method (ISO 2917:1999, clause 1; PKN card)
- Non-destructive measurements on carcass meat, quarters and muscles, including fresh meat at 0–5 °C with temperature correction (9.4)
- Prerigor muscle meat — KCl solution with iodoacetic acid to stop glycolysis (5.4)
Key parameters
| Parameter | Value |
|---|---|
| STATUS (as of 18.11.2026) | PN-ISO 2917:2001 current; amendment Ap1:2002 — editorial (adds PN-77/A-82058 in the "instead of" field); ISO 2917:1999 — "Published", stage 90.93 (iTeh) |
| Edition (from the PKN card) | PN-ISO 2917:2001, Polish version; published 13-06-2001, 9 pages, KT 93, ICS 67.120.10; introduces ISO 2917:1999 [IDT]; replaces PN-A-82058:1977 |
| Buffers (5.2) | pH 4.00 (potassium hydrogen phthalate), 6.88 (KH₂PO₄ + Na₂HPO₄), 5.45 (citric acid + NaOH) at 20 °C; calibration with two buffers spanning the sample pH |
| Extract (9.2) | sample homogenized with ten times as much KCl 0.1 mol/l; shaft homogenizer 20 000 min⁻¹ |
| Temperature (9.1, 9.3, 9.4) | without temperature correction buffer, extract and sample at (20 ± 2) °C; fresh meat 0–5 °C — only with correction |
| Reading and result (9–10) | reading 0.01 pH; result to the nearest 0.05 pH; non-destructive — mean of two readings at the same point |
| Precision (clause 11) | repeatability 0.04 pH (sr ≈ 0.014); reproducibility 0.12 pH (sR ≈ 0.042) |
Standard
- Standard number
- PN-ISO 2917:2001 + Ap1:2002
- Title (PL)
- Mięso i przetwory mięsne — Pomiar pH — Metoda odwoławcza
- Title (EN)
- Meat and meat products — Measurement of pH — Reference method
Step-by-step procedure
-
Sample
Representative, at least 200 g; homogenization without exceeding 25 °C (with a mincer at least twice), analysis within 24 h; for non-destructive measurement — a representative point (clauses 7–8). PN-ISO 2917:2001 current (status as of 18.11.2026).
-
Calibration
Two buffers spanning the expected pH, at the temperature of measurement, with stirring; without temperature correction buffer at (20 ± 2) °C (9.1).
-
Measurement on extract
Portion of sample with ten times as much KCl 0.1 mol/l, shaft homogenizer; electrodes in the extract, temperature correction, stirring, reading 0.01 pH once constant (9.2–9.3).
-
Non-destructive measurement
Hole with a knife or pin, electrode in the incision, temperature correction (meat at 0–5 °C — mandatory), two readings at the same point, several points if useful (9.4).
-
Cleaning the electrodes
Cotton wool with diethyl ether, then ethanol, finally washing with water (9.5).
-
Result
pH to the nearest 0.05; non-destructive — mean for each point and sketch of points; repeatability 0.04, reproducibility 0.12 pH; test report states destructive or non-destructive measurement (clauses 10–12).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| pH meter | Digital or analogue display accurate to 0.01 pH; without temperature correction the scale applies at 20 °C; protected from induction currents (ISO 2917, 6.2) | — |
| Combined electrode | Glass electrode with Ag/AgCl or Hg/HgCl₂ reference electrode in one shaft; glass electrode spherical, conical, cylindrical or needle-shaped; for greasy samples separate electrodes (6.3) | — |
| Equipment for homogenizing the sample | High-speed rotational cutter or mincer with a plate with apertures not exceeding 4.0 mm (6.1) | — |
| Shaft homogenizer and magnetic stirrer | Homogenizer at a rotational frequency of 20 000 min⁻¹; stirrer for calibration and extract measurement (6.4–6.5) | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Buffer solutions pH 4.00, 6.88 and 5.45 (20 °C) | — | Ready-to-use with pH guaranteed to 0.01, from commercial mixtures or prepared according to 5.2.1–5.2.3 (ISO 2917, 5.2) |
| Potassium chloride solution, c(KCl) = 0.1 mol/l | — | 7.5 g KCl to 1000 ml; for prerigor meat 925 mg iodoacetic acid per litre, pH 7.0 (5.4) |
| Sodium hydroxide solution, c(NaOH) = 1.0 mol/l | — | 40 g NaOH to 1000 ml — for the pH 5.45 buffer and adjusting the KCl solution (5.3) |
| Diethyl ether saturated with water and ethanol 95 % | — | Cleaning the electrodes (5.5, 9.5) |
Health and safety (OHS)
- Diethyl ether and ethanol for cleaning electrodes — chemical laboratory safety rules and safety data sheets of the substances
- TECHNICAL NOTE: the scope of the standard is meat and meat products, including poultry; it does not list fish or other products
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-ISO 2917:2001 + Ap1:2002 — “Meat and meat products — Measurement of pH — Reference method”.
How does this method work?
The potential difference is measured between a glass electrode and a reference electrode placed in the extract (sample homogenized with ten times as much KCl solution 0.1 mol/l) or directly in carcass meat, quarters or muscle; the pH meter reading to 0.
What is the measuring range and accuracy?
STATUS (as of 18.11.2026): PN-ISO 2917:2001 current; amendment Ap1:2002 — editorial (adds PN-77/A-82058 in the "instead of" field); ISO 2917:1999 — "Published", stage 90.93 (iTeh); Edition (from the PKN card): PN-ISO 2917:2001, Polish version; published 13-06-2001, 9 pages, KT 93, ICS 67.120.10; introduces ISO 2917:1999 [IDT]; replaces PN-A-82058:1977; Buffers (5.2): pH 4.00 (potassium hydrogen phthalate), 6.88 (KH₂PO₄ + Na₂HPO₄), 5.45 (citric acid + NaOH) at 20 °C; calibration with two buffers spanning the sample pH; Extract (9.2): sample homogenized with ten times as much KCl 0.1 mol/l; shaft homogenizer 20 000 min⁻¹.
How long does the test take?
The procedure comprises 6 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
pH meter, Combined electrode, Equipment for homogenizing the sample, Shaft homogenizer and magnetic stirrer.
Where is this test used?
All kinds of meat and meat products, including poultry — pH by the reference method (ISO 2917:1999, clause 1; PKN card); Non-destructive measurements on carcass meat, quarters and muscles, including fresh meat at 0–5 °C with temperature correction (9.4); Prerigor muscle meat — KCl solution with iodoacetic acid to stop glycolysis (5.4).
What safety precautions apply?
Diethyl ether and ethanol for cleaning electrodes — chemical laboratory safety rules and safety data sheets of the substances; TECHNICAL NOTE: the scope of the standard is meat and meat products, including poultry; it does not list fish or other products.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-ISO 2917.