In short
Basic EMC publication describing the immunity test of electrical and electronic equipment to conducted common mode disturbances in the range from direct current (DC) to 150 kHz, applied to power supply, control, signal and communication ports. The test is performed according to PN-EN 61000-4-16:2016-05 (English version, published 23-05-2016; implements EN 61000-4-16:2016 and IEC 61000-4-16:2015 [IDT]); STATUS (as of 27.11.2026): PKN card: PN-EN 61000-4-16:2016-05 — English version (23-05-2016, 34 pages), KT 104, implements EN 61000-4-16:2016 [IDT] and IEC 61000-4-16:2015 [IDT], replaces PN-EN 61000-4-16:2001 with A1:2006 and A2:2011; the card gives no withdrawal date. The procedure comprises 4 steps; it is used for: Equipment in industrial plants and power stations, near power electronic convertors (Clause 1, Clause 4), Power supply, control, signal and communication ports connected to long cables (Clause 1), Equipment used in the proximity of railway systems supplied at 16⅔ Hz (5.2).
At a glance
- Standard: PN-EN 61000-4-16:2016-05 (English version, published 23-05-2016; implements EN 61000-4-16:2016 and IEC 61000-4-16:2015 [IDT])
- Category: Electronics and EMC
- Procedure steps: 4
- STATUS (as of 27.11.2026): PKN card: PN-EN 61000-4-16:2016-05 — English version (23-05-2016, 34 pages), KT 104, implements EN 61000-4-16:2016 [IDT] and IEC 61000-4-16:2015 [IDT], replaces PN-EN 61000-4-16:2001 with A1:2006 and A2:2011; the card gives no withdrawal date.
- D.c. and mains frequency — continuous (Table 1): levels 1–4: 1, 3, 10, 30 V (r.m.s., open circuit); x — open
- D.c. and mains frequency — short duration (Table 2): levels 1–4: 10, 30, 100, 300 V; normally 1 s
Overview
WHAT THE STANDARD COVERS. According to Clause 1 of IEC 61000-4-16:2015 the standard relates to the immunity requirements and test methods for electrical and electronic equipment to conducted, common mode disturbances in the range d.c. to 150 kHz applied to power supply, control, signal and communication ports; it defines the test voltage and current waveform, the range of test levels, test equipment, test set-up and test procedures. The test is intended to demonstrate immunity to disturbances such as those originating from power line currents and return leakage currents in the earthing/grounding system. Outside the scope: disturbances produced by 400 Hz mains systems, harmonics and interharmonics in differential mode on a.c. power ports (IEC 61000-4-13 and IEC 61000-4-19) and disturbances from intentional radio-frequency transmitters (IEC 61000-4-6). According to the standard, actual interference due to these phenomena is relatively rare, except in industrial plants, and the test is not relevant for ports intended to be connected to short cables — in the text of the standard: "having a length less than 20 m or less".
ACCORDING TO THE TEXT OF IEC 61000-4-16:2015 (second edition; the publisher's sample covers the contents, foreword, introduction and Clauses 1–6.3.2.1). The standard was prepared by subcommittee IEC SC 77A of IEC TC 77. The second edition replaces the first edition of 1998 and Amendments 1:2001 and 2:2009; change according to the foreword: clarification and complement of the test generators' specifications and performances.
SOURCES OF DISTURBANCES (Clause 4). Disturbances are generated by the power distribution system (fundamental frequency, harmonics, interharmonics; fault and leakage currents in the ground and earthing systems) and by power electronic equipment (e.g. power convertors), which injects disturbances into the ground conductors and earthing system or induces them in signal and control lines. Two categories are distinguished: voltage/current at power frequency (d.c., 16⅔ Hz, 50 Hz and 60 Hz) and in the range 15 Hz to 150 kHz (including mains harmonics). The applicability of the tests is defined in the product standard.
TEST LEVELS (Clause 5). The test voltage is applied in common mode to power supply, control, signal and communication ports. Mains frequency (Tables 1 and 2, open circuit r.m.s. voltage): continuous disturbance — level 1: 1 V, 2: 3 V, 3: 10 V, 4: 30 V; short duration — 10, 30, 100 and 300 V; level x — open. The normal duration of a short duration disturbance is 1 s. The test is carried out at d.c., 16⅔ Hz, 50 Hz or 60 Hz according to the mains frequency at the equipment location; 16⅔ Hz only for equipment intended to be used in the proximity of railway systems at this frequency. Range 15 Hz–150 kHz (Table 3): level 1 — 1 V at 15 Hz decreasing to 0.1 V at 150 Hz, 0.1 V up to 1.5 kHz, increasing to 1 V at 15 kHz and 1 V up to 150 kHz; levels 2, 3 and 4 — 3/0.3, 10/1 and 30/3 V respectively; profile: decrease of 20 dB/decade from 15 to 150 Hz, constant from 150 Hz to 1.5 kHz, increase of 20 dB/decade up to 15 kHz, constant up to 150 kHz. No test level is defined below 15 Hz (except d.c.).
GENERATORS (6.1). All have a source impedance Voc/Isc = 50 Ω ±10 %. D.c.: ripple less than 5 %, 1–30 V (continuous) and 10–300 V (short duration), rise and fall time at switching between 1 µs and 5 µs. Mains frequency: variable transformer, isolation transformer and time controlled switch synchronized at 0° (0° ± 5 %), total harmonic distortion less than 10 %. Range 15 Hz–150 kHz: sweep at 1 × 10⁻² decade/s or slower or steps of 10 % of the preceding frequency, distortion less than 1 %, 0.1–30 V. Verification (6.2): waveform, source impedance (at the highest and lowest level; for the sweep also at 15 Hz, 1.5 kHz, 15 kHz and 150 kHz; the first 50 ms of a short duration disturbance may be disregarded), frequency, output voltage, switching times — with probes and an oscilloscope of at least 1 MHz bandwidth and accuracy better than ±5 %. Coupling network for power supply and input/output ports (6.3.2.1): for each conductor a resistor and a capacitor in series, C = 1.0 µF, R = 100 × n Ω (n ≥ 2 — number of conductors), components matched within 1 %; for the d.c. test the capacitors are short-circuited; for screened cables the test signal is injected directly onto the shield.
WHAT WE DO NOT GIVE HERE. The sample ends at 6.3.2.1 — we do not reproduce coupling for communication ports (T network), decoupling devices, the test set-up (Clause 7), procedure (8), evaluation (9) or report (10).
STATUS (as of 27.11.2026). PKN card: PN-EN 61000-4-16:2016-05 "Kompatybilność elektromagnetyczna (EMC) -- Część 4-16: Metody badań i pomiarów -- Badanie odporności na asymetryczne zaburzenia przewodzone w zakresie częstotliwości od 0 Hz do 150 kHz" — English version (published 23-05-2016, 34 pages), KT 104 (electromagnetic compatibility), implements EN 61000-4-16:2016 [IDT] and IEC 61000-4-16:2015 [IDT], replaces PN-EN 61000-4-16:2001 (Polish version) with A1:2006 (Polish version) and A2:2011 (English version), ICS 33.100.20; the card gives no withdrawal date and no additional elements.
Method principle
A voltage from a generator with a 50 Ω source impedance is applied in common mode through an RC coupling network (for screened cables — onto the shield) to the power supply, control, signal and communication ports of the equipment under test: d.c. or mains frequency (continuous and short duration, 1 s) and swept from 15 Hz to 150 kHz according to the profile of Table 3, and the operation of the equipment is observed.
Applications
- Equipment in industrial plants and power stations, near power electronic convertors (Clause 1, Clause 4)
- Power supply, control, signal and communication ports connected to long cables (Clause 1)
- Equipment used in the proximity of railway systems supplied at 16⅔ Hz (5.2)
- Immunity tests required by product standards that refer to IEC 61000-4-16 (IEC Guide 107)
Key parameters
| Parameter | Value |
|---|---|
| STATUS (as of 27.11.2026) | PKN card: PN-EN 61000-4-16:2016-05 — English version (23-05-2016, 34 pages), KT 104, implements EN 61000-4-16:2016 [IDT] and IEC 61000-4-16:2015 [IDT], replaces PN-EN 61000-4-16:2001 with A1:2006 and A2:2011; the card gives no withdrawal date. |
| D.c. and mains frequency — continuous (Table 1) | levels 1–4: 1, 3, 10, 30 V (r.m.s., open circuit); x — open |
| D.c. and mains frequency — short duration (Table 2) | levels 1–4: 10, 30, 100, 300 V; normally 1 s |
| Frequencies (5.2) | d.c., 16⅔ Hz, 50 Hz, 60 Hz — according to the mains at the equipment location |
| 15 Hz–150 kHz (Table 3) | level 1: 1 → 0.1 V (15–150 Hz), 0.1 V (150 Hz–1.5 kHz), 0.1 → 1 V (1.5–15 kHz), 1 V (15–150 kHz); levels 2–4: ×3, ×10, ×30 |
| Source impedance (6.1) | Voc/Isc = 50 Ω ±10 % for all generators |
| Sweep generator (6.1.4) | 1 × 10⁻² decade/s or slower or 10 % steps; THD < 1 %; 0.1–30 V |
| Coupling network (6.3.2.1) | C = 1.0 µF and R = 100 × n Ω per conductor (n ≥ 2); matched within 1 %; capacitors short-circuited for d.c. |
Standard
- Standard number
- PN-EN 61000-4-16:2016-05 (English version, published 23-05-2016; implements EN 61000-4-16:2016 and IEC 61000-4-16:2015 [IDT])
- Title (PL)
- Kompatybilność elektromagnetyczna (EMC) — Część 4-16: Metody badań i pomiarów — Badanie odporności na asymetryczne zaburzenia przewodzone w zakresie częstotliwości od 0 Hz do 150 kHz
- Title (EN)
- Electromagnetic compatibility (EMC) — Part 4-16: Testing and measurement techniques — Test for immunity to conducted, common mode disturbances in the frequency range 0 Hz to 150 kHz
Step-by-step procedure
-
Selection of tests and levels
Applicability of the tests and levels from the product standard (5.1); mains frequency according to the location (5.2); selection of the ports exposed to the disturbances (Clause 4).
-
Verification of the generators
Waveform, 50 Ω source impedance (at the highest and lowest level, for the sweep also at 15 Hz, 1.5 kHz, 15 kHz, 150 kHz), frequency, voltage, switching times (6.2).
-
Coupling
RC network on each conductor of the port (C = 1.0 µF, R = 100 × n Ω), capacitors short-circuited for d.c.; screened cables — voltage onto the shield (6.3.2.1).
-
Test
Continuous and short duration (1 s) disturbance at d.c. and mains frequency; sweep 15 Hz–150 kHz according to the profile of Table 3 (Clauses 5 and 8).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| D.c. generator (6.1.2) | Ripple < 5 %; 1–30 V continuous, 10–300 V short duration; 50 Ω ±10 %; switching 1–5 µs | — |
| Mains frequency generator (6.1.3) | Variable transformer, isolation transformer, switch synchronized at 0°; THD < 10 % | — |
| Sweep generator 15 Hz–150 kHz (6.1.4) | THD < 1 %; 0.1–30 V; also manual setting | — |
| Voltage and current probes with an oscilloscope (6.2) | Bandwidth at least 1 MHz, accuracy better than ±5 % | — |
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN 61000-4-16:2016-05 (English version, published 23-05-2016; implements EN 61000-4-16:2016 and IEC 61000-4-16:2015 [IDT]) — “Electromagnetic compatibility (EMC) — Part 4-16: Testing and measurement techniques — Test for immunity to conducted, common mode disturbances in the frequency range 0 Hz to 150 kHz”.
How does this method work?
A voltage from a generator with a 50 Ω source impedance is applied in common mode through an RC coupling network (for screened cables — onto the shield) to the power supply, control, signal and communication ports of the equipment under test: d.c.
What is the measuring range and accuracy?
STATUS (as of 27.11.2026): PKN card: PN-EN 61000-4-16:2016-05 — English version (23-05-2016, 34 pages), KT 104, implements EN 61000-4-16:2016 [IDT] and IEC 61000-4-16:2015 [IDT], replaces PN-EN 61000-4-16:2001 with A1:2006 and A2:2011; the card gives no withdrawal date.; D.c. and mains frequency — continuous (Table 1): levels 1–4: 1, 3, 10, 30 V (r.m.s., open circuit); x — open; D.c. and mains frequency — short duration (Table 2): levels 1–4: 10, 30, 100, 300 V; normally 1 s; Frequencies (5.2): d.c., 16⅔ Hz, 50 Hz, 60 Hz — according to the mains at the equipment location.
How long does the test take?
The procedure comprises 4 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
D.c. generator (6.1.2), Mains frequency generator (6.1.3), Sweep generator 15 Hz–150 kHz (6.1.4), Voltage and current probes with an oscilloscope (6.2).
Where is this test used?
Equipment in industrial plants and power stations, near power electronic convertors (Clause 1, Clause 4); Power supply, control, signal and communication ports connected to long cables (Clause 1); Equipment used in the proximity of railway systems supplied at 16⅔ Hz (5.2); Immunity tests required by product standards that refer to IEC 61000-4-16 (IEC Guide 107).
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 61000-4-16.