🌿 2-Methoxyethyl acetate in workplace air — adsorption on activated carbon, extraction with chloroform or carbon disulphide and gas chromatography with sample enrichment, PN-Z-04184-02 (withdrawn standard)

Environment PN-Z-04184-02

In short

Determination of 2-methoxyethyl acetate vapour in workplace air: adsorption on activated carbon, extraction with chloroform or carbon disulphide and gas chromatography. The test is performed according to PN-Z-04184-02:1988 (Polish version); STATUS (as of 18.11.2026): withdrawn 15-08-2014; card without "Replaced by" and "Replaces" fields; both standards of the PN-Z-04184 series withdrawn. The procedure comprises 5 steps; it is used for: Measurements of 2-methoxyethyl acetate vapour in workplace air — assessment against the NDS of 5 mg/m³ (Journal of Laws 2026 item 447, item 434); the determinability of the standard (6 mg/m³) is higher than the NDS.

At a glance

  • Standard: PN-Z-04184-02:1988 (Polish version)
  • Category: Environment
  • Procedure steps: 5
  • STATUS (as of 18.11.2026): withdrawn 15-08-2014; card without "Replaced by" and "Replaces" fields; both standards of the PN-Z-04184 series withdrawn
  • Edition (from the PKN card): PN-Z-04184-02:1988, Polish version; published 03-02-1988, 3 pages, KT 159 Chemical and Dust Hazards in the Working Environment, ICS 13.040.30
  • Principle and determinability (from the card): adsorption of the vapour on activated carbon, extraction with chloroform or carbon disulphide, gas chromatography; determinability 6 mg/m³

Overview

STATUS (as of 18.11.2026). The card of PN-Z-04184-02:1988 (Polish version) has the heading "Withdrawn": published 03-02-1988, withdrawal date 15-08-2014, 3 pages; there are no "Replaced by" or "Replaces" fields. The PKN search for "octanu 2-metoksyetylu" (18.11.2026) returns for this substance only two sheets of the series (01:1988 and 02:1988), both withdrawn; the other current results concern other acetates.

WHAT THE STANDARD COVERS. The scope from the PKN catalogue card, quoted in full: "The test method consists in the adsorption of 2-methoxyethyl acetate vapour on activated carbon, its extraction with chloroform or carbon disulphide and chromatographic analysis of the solution obtained. The determinability of the method is 6 mg in a cubic metre of air" (our translation; the card uses the old Polish name "dwusiarczek węgla" for carbon disulphide).

LIMIT VALUES AND THE REGULATION ON THE METHOD. The NDS list — Annex 1 to the regulation of 12 June 2018 as worded by the regulation of 26 March 2026 (Journal of Laws 2026 item 447), item 434 "2-Methoxyethyl acetate" [110-49-6]: NDS (8-hour value) 5 mg/m³ (1 ppm), no NDSCh, "skin" notation. According to the explanatory note to the list, the "skin" notation means that "absorption of the substance through the skin may be as important as exposure by inhalation" (our translation). The determinability according to the card (6 mg/m³) is higher than the NDS — a result "below the determinability" does not decide whether the NDS is met. The method is addressed by § 16 of the regulation on tests and measurements of agents harmful to health in the working environment (consolidated text — Journal of Laws 2025 item 949): "Tests and measurements are performed in accordance with the methods specified in Polish Standards and, where there are no such standards, with methods recommended and validated by the laboratories referred to in § 15(1) and (2)(1)" (our translation). The provision does not name this standard; whether a standard withdrawn by PKN falls under the words "where there are no such standards" — WE DO NOT DECIDE.

PROCEDURE ACCORDING TO THE IMP PROCEDURE (second source; numbers of the procedure, not of the standard, different extractant and detector). Kucharska, Wesołowski, Gromiec, "Octan 2-metoksyetylu – metoda oznaczania" (2-Methoxyethyl acetate – determination method), Podstawy i Metody Oceny Środowiska Pracy 2010, no. 1(63), pp. 185–192: a tube with two sections of activated carbon (100 mg and 50 mg); up to 10 l of air at a flow rate of up to 5 l/h, according to PN-Z-04008-7; samples kept in a refrigerator are stable for 7 days; each section is desorbed with 1 ml of dichloromethane for 30 min; gas chromatograph with a mass detector (SIM mode) and a column separating 2-methoxyethyl acetate from glycol ethers; calibration curve 0.5–100 µg/ml; the amount in the shorter section should not exceed 10 % of the amount in the longer one; range 0.05–10 mg/m³ for 10 l. According to the card, the standard extracts with chloroform or carbon disulphide — we do not transfer the numbers of the procedure to the standard.

WHAT WE DO NOT GIVE HERE. We did not have the text of PN-Z-04184-02:1988 — PKN sells it (3 pages). We do not give the amount of activated carbon, air volume and flow rate, volume of extractant, column, detector, chromatographic conditions or calculations according to the standard.

Method principle

According to the card of the standard, 2-methoxyethyl acetate vapour from workplace air is adsorbed on activated carbon, extracted with chloroform or carbon disulphide, and the solution is analysed chromatographically; the determinability is 6 mg/m³ — above the NDS. The IMP procedure of 2010 desorbs the carbon with dichloromethane and determines the compound by gas chromatography–mass spectrometry. The result is compared with the NDS of 5 mg/m³ (NDS list of 2026, item 434).

Applications

Key parameters

ParameterValue
STATUS (as of 18.11.2026)withdrawn 15-08-2014; card without "Replaced by" and "Replaces" fields; both standards of the PN-Z-04184 series withdrawn
Edition (from the PKN card)PN-Z-04184-02:1988, Polish version; published 03-02-1988, 3 pages, KT 159 Chemical and Dust Hazards in the Working Environment, ICS 13.040.30
Principle and determinability (from the card)adsorption of the vapour on activated carbon, extraction with chloroform or carbon disulphide, gas chromatography; determinability 6 mg/m³
Limit values against which the result is assessedNDS 5 mg/m³ (1 ppm), no NDSCh, "skin" notation — Journal of Laws 2026 item 447, item 434 "2-Methoxyethyl acetate" [110-49-6]
IMP procedure 2010 (numbers of the procedure)activated carbon 100 mg/50 mg; up to 10 l, up to 5 l/h; desorption with 1 ml of dichloromethane, 30 min; GC-MS (SIM); 0.05–10 mg/m³ for 10 l

Standard

Standard number
PN-Z-04184-02:1988 (Polish version)
Title (PL)
Ochrona czystości powietrza — Badania zawartości octanu 2-metoksyetylu — Oznaczanie octanu 2-metoksyetylu na stanowiskach pracy metodą chromatografii gazowej z wzbogacaniem próbki
Title (EN)
Air purity protection — Tests for 2-methoxyethyl acetate content — Determination of 2-methoxyethyl acetate at workplaces by gas chromatography with sample enrichment

Step-by-step procedure

  1. Status of the document

    PKN card: standard withdrawn on 15-08-2014, no successor indicated (as of 18.11.2026).

  2. Sampling

    The vapour is adsorbed on activated carbon (from the card); in the IMP procedure 2010 — up to 10 l of air, up to 5 l/h.

  3. Extraction

    Card: with chloroform or carbon disulphide; IMP procedure 2010 — 1 ml of dichloromethane, 30 min.

  4. Chromatographic analysis

    Gas chromatography (from the card); in the IMP procedure 2010 — mass detector in SIM mode.

  5. Result and assessment

    The concentration in mg/m³ is compared with the NDS of 5 mg/m³. With the determinability of the standard of 6 mg/m³, a result "below the determinability" does not decide whether the NDS is met. "Skin" notation: the measurement in air does not cover absorption through the skin.

Required equipment and apparatus

EquipmentExampleIndicative price
Activated carbon tubes (100 mg/50 mg)IMP procedure 2010; WE DO NOT GIVE the values according to the standard—
Suction pumpIMP procedure 2010: up to 10 l of air, up to 5 l/h—
Gas chromatographThe card of the standard says only "chromatographic analysis"; IMP procedure 2010 — mass detector—

Reagents, media and consumables

ReagentCASDetails
Chloroform or carbon disulphide—Extractants according to the card of the standard (card: "dwusiarczkiem węgla"); WE DO NOT GIVE the volume
Dichloromethane75-09-2Desorbent in the IMP procedure 2010 (1 ml per section) — not according to the standard
2-Methoxyethyl acetate — reference substance110-49-6Working standard solutions of 0.5–100 µg/ml (IMP procedure 2010)

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-Z-04184-02:1988 (Polish version) — “Air purity protection — Tests for 2-methoxyethyl acetate content — Determination of 2-methoxyethyl acetate at workplaces by gas chromatography with sample enrichment”.

How does this method work?

According to the card of the standard, 2-methoxyethyl acetate vapour from workplace air is adsorbed on activated carbon, extracted with chloroform or carbon disulphide, and the solution is analysed chromatographically; the determinability is 6 mg/m³ — above the NDS. The IMP procedure of 2010 desorbs the carbon with dichloromethane and determines the compound by gas chromatography–mass spectrometry.

What is the measuring range and accuracy?

STATUS (as of 18.11.2026): withdrawn 15-08-2014; card without "Replaced by" and "Replaces" fields; both standards of the PN-Z-04184 series withdrawn; Edition (from the PKN card): PN-Z-04184-02:1988, Polish version; published 03-02-1988, 3 pages, KT 159 Chemical and Dust Hazards in the Working Environment, ICS 13.040.30; Principle and determinability (from the card): adsorption of the vapour on activated carbon, extraction with chloroform or carbon disulphide, gas chromatography; determinability 6 mg/m³; Limit values against which the result is assessed: NDS 5 mg/m³ (1 ppm), no NDSCh, "skin" notation — Journal of Laws 2026 item 447, item 434 "2-Methoxyethyl acetate" [110-49-6].

How long does the test take?

The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Activated carbon tubes (100 mg/50 mg), Suction pump, Gas chromatograph.

Where is this test used?

Measurements of 2-methoxyethyl acetate vapour in workplace air — assessment against the NDS of 5 mg/m³ (Journal of Laws 2026 item 447, item 434); the determinability of the standard (6 mg/m³) is higher than the NDS.

What safety precautions apply?

Work with the reference substances is carried out in rubber gloves and protective clothing, in an efficient laboratory fume hood (IMP procedure 2010, point 4.3); The "skin" notation in the NDS list: the measurement in air does not cover absorption through the skin (2-methoxyethyl acetate, item 434).

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-Z-04184-02.

🔍 Find a laboratory performing this test