⚡ Immunity of electrical and electronic equipment in residential, commercial, public and light-industrial environments — generic standard: requirements from 0 Hz to 400 GHz for the enclosure port, signal/control ports and DC and AC power ports, performance criteria A, B and C; PN-EN IEC 61000-6-1

Electronics and EMC PN-EN IEC 61000-6-1

In short

Generic EMC standard specifying immunity requirements for electrical and electronic equipment intended for residential, commercial, public and light-industrial locations where no relevant dedicated product or product-family standard exists. The test is performed according to PN-EN IEC 61000-6-1:2019-03 (English version, published 27-03-2019; implements EN IEC 61000-6-1:2019 and IEC 61000-6-1:2016 [IDT]); Frequency range (Clause 1): from 0 Hz to 400 GHz; no tests where no requirements are specified. The procedure comprises 4 steps; it is used for: Equipment for houses and apartments connected to the low-voltage public mains network (3.8), Equipment for shops, offices, hotels, data centres, hospitals, schools, railway stations, workshops and laboratories (3.9), Battery operated equipment or equipment powered by a non-public, non-industrial low voltage network in these locations (Clause 1).

At a glance

  • Standard: PN-EN IEC 61000-6-1:2019-03 (English version, published 27-03-2019; implements EN IEC 61000-6-1:2019 and IEC 61000-6-1:2016 [IDT])
  • Category: Electronics and EMC
  • Procedure steps: 4
  • STATUS (as of 27.11.2026): PKN card: PN-EN IEC 61000-6-1:2019-03 — English version (27-03-2019, 28 pages), KT 104, implements EN IEC 61000-6-1:2019 [IDT] and IEC 61000-6-1:2016 [IDT], replaces PN-EN 61000-6-1:2008; the card gives no withdrawal date.
  • Frequency range (Clause 1): from 0 Hz to 400 GHz; no tests where no requirements are specified
  • Ports (3.1–3.4, Tables 1–4): enclosure; signal/control; DC power; AC power

Overview

WHAT THE STANDARD COVERS. According to Clause 1 of IEC 61000-6-1:2016 the standard specifies EMC immunity requirements for electrical and electronic equipment intended for use in residential, commercial, public and light-industrial locations, in the range from 0 Hz to 400 GHz (no tests at frequencies where no requirements are specified). It applies if no relevant dedicated product or product-family EMC immunity standard exists. It covers equipment indoors and outdoors, also battery operated or powered by a non-public, but non-industrial, low voltage distribution system in these locations; it concerns continuous and transient, conducted and radiated disturbances, including electrostatic discharges. The levels do not cover extreme cases with an extremely low probability of occurrence; safety is not covered.

ACCORDING TO THE TEXT OF IEC 61000-6-1:2016 (third edition; the publisher's sample covers the contents, foreword, introduction and Clauses 1–9 without Tables 1–4). The standard was prepared by IEC TC 77. The third edition replaces the second edition of 2005; changes according to the foreword: improved environmental description, extended frequency range and amended levels at particular frequencies for the radiated field test according to IEC 61000-4-3, changed repetition frequency for the fast transients test according to IEC 61000-4-4, requirements according to IEC 61000-4-34, revised test levels, consideration of measurement uncertainty and new Annex A. Normative references: IEC 61000-4-2, -4-3, -4-4, -4-5, -4-6, -4-8, -4-11, -4-20, -4-21, -4-22 and -4-34 in the cited editions.

LOCATIONS (3.8, 3.9). Residential — an area of land designated for domestic dwellings (houses, apartments, farm buildings used for living), where equipment is connected directly to the low-voltage public mains network or to a dedicated DC source interfacing between the equipment and the mains. Commercial, public and light-industrial — e.g. shops, offices, banks, hotels, data centres, cinemas, places of worship, petrol stations, car parks, hospitals, schools, railway stations, public areas of airports and workshops, laboratories and service centres. Long distance line (3.6) — connected to a signal/control port and longer than 30 m inside a building or leaving the building. Ports (3.1–3.4): enclosure, signal/control, power (AC, DC).

PERFORMANCE CRITERIA (Clause 4). The manufacturer provides a functional description and EUT-specific performance criteria, noted in the test report, consistent with the general criteria: A — the EUT continues to operate as intended during and after the test, without degradation below the level specified by the manufacturer; B — it continues to operate as intended after the test; degradation is allowed during the test, but no change of actual operating state or stored data; C — temporary loss of function is allowed if self-recoverable or restorable by operating the controls. An EUT that becomes dangerous or unsafe has failed the test.

CONDITIONS AND APPLICABILITY (Clauses 5–9). The EUT is tested in the most susceptible operating mode consistent with normal applications (e.g. identified by pre-tests), with the minimum representative configuration of auxiliary equipment (which may be simulated), with external protection devices required in the user's manual; with many similar ports a sufficient number is selected and the selection justified in the report; at one set of climatic parameters and the rated supply voltage. A manufacturer's own specification of acceptable degradation is stated in the user documentation. Tests are applied to the existing ports according to Tables 1–4; a decision not to test is justified in the report. Uncertainty — according to IEC TR 61000-1-6 or the basic standard. Tests are carried out individually, in any order, possibly in parallel on identical units; the test description, generators and set-ups are given in the basic standards referred to in Tables 1–4.

WHAT WE DO NOT GIVE HERE. Tables 1–4 (tests and levels for the enclosure port, signal/control ports, DC and AC power ports) and Annex A — the sample ends before them.

STATUS (as of 27.11.2026). PKN card: PN-EN IEC 61000-6-1:2019-03 "Kompatybilność elektromagnetyczna (EMC) -- Część 6-1: Normy ogólne -- Norma dotycząca odporności w środowiskach: mieszkalnym, handlowym i lekko uprzemysłowionym" — English version (published 27-03-2019, 28 pages), KT 104 (electromagnetic compatibility), implements EN IEC 61000-6-1:2019 [IDT] and IEC 61000-6-1:2016 [IDT], replaces PN-EN 61000-6-1:2008 (Polish version), ICS 33.100.20; the card gives no withdrawal date and no additional elements.

Method principle

The equipment is tested in its most susceptible typical operating mode with successive immunity tests according to the basic standards of the IEC 61000-4 series listed in Tables 1–4 for each existing port (enclosure, signal/control ports, DC and AC power ports), and the result is evaluated according to criterion A, B or C assigned to the test, with the specific criteria provided by the manufacturer.

Applications

Key parameters

ParameterValue
STATUS (as of 27.11.2026)PKN card: PN-EN IEC 61000-6-1:2019-03 — English version (27-03-2019, 28 pages), KT 104, implements EN IEC 61000-6-1:2019 [IDT] and IEC 61000-6-1:2016 [IDT], replaces PN-EN 61000-6-1:2008; the card gives no withdrawal date.
Frequency range (Clause 1)from 0 Hz to 400 GHz; no tests where no requirements are specified
Ports (3.1–3.4, Tables 1–4)enclosure; signal/control; DC power; AC power
Long distance line (3.6)longer than 30 m inside a building or leaving the building
Performance criteria (Clause 4)A — operation during and after the test; B — after the test, no change of operating state or data during the test; C — temporary loss of function, self-recoverable or restored by the controls
Test conditions (Clause 5)most susceptible typical operating mode; minimum representative configuration; one set of climatic parameters; rated voltage
Basic standards (Clause 2)IEC 61000-4-2, -4-3, -4-4, -4-5, -4-6, -4-8, -4-11, -4-20, -4-21, -4-22, -4-34

Standard

Standard number
PN-EN IEC 61000-6-1:2019-03 (English version, published 27-03-2019; implements EN IEC 61000-6-1:2019 and IEC 61000-6-1:2016 [IDT])
Title (PL)
Kompatybilność elektromagnetyczna (EMC) — Część 6-1: Normy ogólne — Norma dotycząca odporności w środowiskach: mieszkalnym, handlowym i lekko uprzemysłowionym
Title (EN)
Electromagnetic compatibility (EMC) — Part 6-1: Generic standards — Immunity standard for residential, commercial and light-industrial environments

Step-by-step procedure

  1. Determination of applicability

    Check that no relevant product standard exists; location according to 3.8/3.9; list of existing ports and tests from Tables 1–4; justification of omitted tests (Clause 7).

  2. Criteria and configuration

    Functional description and specific criteria from the manufacturer (Clause 4); most susceptible operating mode, minimum representative configuration, external protection from the manual, selection of ports (Clause 5).

  3. Tests

    Each test individually, in any order, according to the basic standard; possibly in parallel on identical units (Clause 9).

  4. Evaluation and report

    Result according to criterion A, B or C; in the report: criteria, configuration and operating mode, selection of ports, omitted tests, units used in parallel.

Required equipment and apparatus

EquipmentExampleIndicative price
Test equipment according to the basic standards of the IEC 61000-4 series (Clause 9)Generators, set-ups and methods are described in the basic standards referred to in Tables 1–4—

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN IEC 61000-6-1:2019-03 (English version, published 27-03-2019; implements EN IEC 61000-6-1:2019 and IEC 61000-6-1:2016 [IDT]) — “Electromagnetic compatibility (EMC) — Part 6-1: Generic standards — Immunity standard for residential, commercial and light-industrial environments”.

How does this method work?

The equipment is tested in its most susceptible typical operating mode with successive immunity tests according to the basic standards of the IEC 61000-4 series listed in Tables 1–4 for each existing port (enclosure, signal/control ports, DC and AC power ports), and the result is evaluated according to criterion A, B or C assigned to the test, with the specific criteria provided by the manufacturer.

What is the measuring range and accuracy?

STATUS (as of 27.11.2026): PKN card: PN-EN IEC 61000-6-1:2019-03 — English version (27-03-2019, 28 pages), KT 104, implements EN IEC 61000-6-1:2019 [IDT] and IEC 61000-6-1:2016 [IDT], replaces PN-EN 61000-6-1:2008; the card gives no withdrawal date.; Frequency range (Clause 1): from 0 Hz to 400 GHz; no tests where no requirements are specified; Ports (3.1–3.4, Tables 1–4): enclosure; signal/control; DC power; AC power; Long distance line (3.6): longer than 30 m inside a building or leaving the building.

How long does the test take?

The procedure comprises 4 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Test equipment according to the basic standards of the IEC 61000-4 series (Clause 9).

Where is this test used?

Equipment for houses and apartments connected to the low-voltage public mains network (3.8); Equipment for shops, offices, hotels, data centres, hospitals, schools, railway stations, workshops and laboratories (3.9); Battery operated equipment or equipment powered by a non-public, non-industrial low voltage network in these locations (Clause 1); Products without their own product or product-family immunity standard (Clause 1).

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN IEC 61000-6-1.

🔍 Find a laboratory performing this test