🔊 Class 1 and 2 sound level meters — periodic tests: conditions for admission to testing, conformance criterion with uncertainty (95 %), inspection and power supply, environmental conditions (80–105 kPa, 20–26 °C, 25–70 %), sound calibrator according to IEC 60942, indication at the calibration check frequency, self-generated noise, frequency weightings; PN-EN 61672-3
In short
Procedures for periodic testing of time-weighting, integrating-averaging and integrating sound level meters designed to class 1 or class 2 of the second edition of IEC 61672-1. The test is performed according to PN-EN 61672-3:2014-03 (Polish version; implements EN 61672-3:2013 and IEC 61672-3:2013 [IDT]); STATUS (read 17.11.2026): PN-EN 61672-3:2014-03 (Polish version, published 21-12-2016); implements EN 61672-3:2013 and IEC 61672-3:2013; replaces PN-EN 61672-3:2007. The procedure comprises 4 steps; it is used for: Periodic testing of class 1 and 2 sound level meters according to the second edition of IEC 61672-1 — time-weighting, integrating-averaging and integrating (IEC 61672-3:2013, clause 1), Calibration of sound level meters in an acoustic laboratory with an uncertainty budget according to the GUM (clause 4), Checking the sound calibrator used with the meter against the periodic-test requirements of IEC 60942 (clause 9).
At a glance
- Standard: PN-EN 61672-3:2014-03 (Polish version; implements EN 61672-3:2013 and IEC 61672-3:2013 [IDT])
- Category: Metrology
- Procedure steps: 4
- STATUS (read 17.11.2026): PN-EN 61672-3:2014-03 (Polish version, published 21-12-2016); implements EN 61672-3:2013 and IEC 61672-3:2013; replaces PN-EN 61672-3:2007
- Scope (clause 1): class 1 and 2 meters according to IEC 61672-1 (2nd ed.); for IEC 61672-1:2002 — IEC 61672-3:2006
- Conformance criterion (4.1): deviation ≤ acceptance limit and uncertainty ≤ maximum-permitted uncertainty (95 %)
Overview
WHAT THE STANDARD COVERS. According to clause 1 of IEC 61672-3:2013 the second edition of the standard describes procedures for periodic testing of time-weighting, integrating-averaging and integrating sound level meters designed to conform to the class 1 or class 2 specifications of the second edition of IEC 61672-1; the aim is consistent performance of periodic tests by all laboratories. Periodic tests are to assure the user that the meter conforms to the applicable requirements of IEC 61672-1 for a limited set of key tests and under the conditions in which they were performed; their extent is deliberately restricted to the minimum. They apply to meters whose manufacturer claims conformance to the second edition of IEC 61672-1, whether or not the model has been pattern approved according to IEC 61672-2; if evidence of pattern approval is not publicly available, no general conclusion about conformance to IEC 61672-1 can be drawn, even when the results conform to all requirements. Procedures for meters according to IEC 61672-1:2002 were given in IEC 61672-3:2006.
ACCORDING TO THE TEXT OF IEC 61672-3:2013 (publisher's sample in English: table of contents, foreword, clauses 1–11 and the beginning of clause 12). The standard was prepared by IEC TC 29 (Electroacoustics); the second edition replaces the first of 2006 and is a technical revision — conformance to a specification is demonstrated when the measured deviations from design goals do not exceed the acceptance limits and the measurement uncertainty does not exceed the maximum-permitted uncertainty, both for a coverage probability of 95 %. Normative references include: IEC 60942 (sound calibrators), IEC 61094-5 and 61094-6 (measurement microphones, electrostatic actuators), IEC 61183, IEC 61672-1 and 61672-2, IEC 62585 (corrections to the free-field response), ISO/IEC Guide 98-3 (GUM) and ISO/IEC Guide 99 (VIM).
ADMISSION TO TESTING (clauses 3, 5, 6). Without an instruction manual for the model and version of the meter (submitted, available at the laboratory or on the manufacturer's website) no periodic tests are performed; the source of the manual is described in the documentation. All accessories necessary for the tests accompany the meter and, if requested by the laboratory, a device for inserting electrical signals equivalent to those from the microphone. Tests are not performed if the marking of the meter does not comply with IEC 61672-1 (the serial number and model designation shall be visible). A sound calibrator is required — supplied by the user (the laboratory then uses it to establish the acoustical sensitivity) or by the laboratory. Before measurements — visual inspection of the meter and accessories (especially the protection grid and the microphone diaphragm) and check of the controls; the power supply is checked by the method in the manual before and after the sets of tests with acoustical and electrical signals; malfunction or a supply outside the limits (for a reason other than discharged batteries or a wrong mains voltage selection) excludes testing.
UNCERTAINTY (clause 4). Uncertainties are determined according to ISO/IEC Guide 98-3 for a coverage probability of 95 %, taking into account at least the calibration of instruments (including the calibrator and the free-field facility), environmental effects, errors in the applied signals, repeatability (with a single measurement — estimated from earlier results for a similar meter), display resolution (for indications in steps of 0,1 dB — rectangular distribution with a semi-range of 0,05 dB), the mounting of the meter and deviations of the field from a free field, and each applied correction. A result whose uncertainty exceeds the maximum-permitted uncertainty is not used to assess conformance; if the excess is due only to the uncertainty of the manufacturer's free-field or random-incidence correction data, testing may continue with an explanation in the documentation.
CONDITIONS AND PROCEDURE (clauses 7–12). Environmental conditions: static pressure from 80 kPa to 105 kPa, temperature from 20 °C to 26 °C, relative humidity from 25 % to 70 %, measured at least at the start and end of testing; for tests of frequency weightings below 97 kPa, without data on the influence of the conditions, standard uncertainties of 0,09 dB up to 3 kHz and 0,14 dB above 3 kHz are assigned. All functions required by IEC 61672-1 are tested, in the configuration requested by the user; the frequency of an electrical signal shall not deviate from the specified value by more than ±0,25 %, and the laboratory's instruments shall be calibrated with traceability to national standards. The sound calibrator (clause 9) — preferably the model from the meter's manual — shall meet the periodic-test requirements of IEC 60942 for its class (level, frequency, distortion) and be calibrated with a microphone of the same model as in the meter (or an equivalent one, with corrections). The indication at the calibration check frequency (clause 10) is checked with the calibrator, adjusted if necessary and recorded before and after adjustment, with the calibrator correction for ambient conditions. Self-generated noise (clause 11) with the microphone — on the most sensitive range with frequency weighting A, preferably as a time-averaged level over at least 30 s (or the average of ten readings within 60 s, with time weighting S if available) — and with the electrical input device for all frequency weightings; the noise level is reported for information, without uncertainty. Frequency weighting C (or A) is tested with a class 1 multi-frequency calibrator according to IEC 60942, a comparison coupler with a microphone according to IEC 61094-4, an electrostatic actuator (IEC 61094-6) or in a free field, with corrections according to IEC 62585 or IEC 61183 from the meter's manual. Further clauses according to the table of contents: electrical signal tests, frequency and time weightings at 1 kHz, long-term stability, linearity, toneburst response, C-weighted peak level, overload indication and high-level stability (clauses 13–21) — the sample does not contain their text.
STATUS (read 17.11.2026). PKN card: PN-EN 61672-3:2014-03 — Polish version (published 21-12-2016), title “Elektroakustyka -- Mierniki poziomu dźwięku -- Część 3: Badania okresowe”, 22 pages, KT 105 for Electroacoustics and Sound and Image Recording, implements EN 61672-3:2013 and IEC 61672-3:2013 [IDT], replaces PN-EN 61672-3:2007; the card gives no withdrawal date. iTeh catalogue: IEC 61672-3:2013 — published 30-09-2013, stage PPUB (publication issued); EN 61672-3:2013 replaced EN 61672-3:2006.
Method principle
Periodic tests check a limited set of key properties of a sound level meter (indication with the calibrator, self-generated noise, frequency weightings and further tests with electrical signals) under laboratory conditions. A result conforms when the measured deviation is within the acceptance limit of IEC 61672-1 and the laboratory's measurement uncertainty (95 %) does not exceed the maximum-permitted uncertainty.
Applications
- Periodic testing of class 1 and 2 sound level meters according to the second edition of IEC 61672-1 — time-weighting, integrating-averaging and integrating (IEC 61672-3:2013, clause 1)
- Calibration of sound level meters in an acoustic laboratory with an uncertainty budget according to the GUM (clause 4)
- Checking the sound calibrator used with the meter against the periodic-test requirements of IEC 60942 (clause 9)
- Multi-channel meters — check of the indication for each tested channel (10.1)
Key parameters
| Parameter | Value |
|---|---|
| STATUS (read 17.11.2026) | PN-EN 61672-3:2014-03 (Polish version, published 21-12-2016); implements EN 61672-3:2013 and IEC 61672-3:2013; replaces PN-EN 61672-3:2007 |
| Scope (clause 1) | class 1 and 2 meters according to IEC 61672-1 (2nd ed.); for IEC 61672-1:2002 — IEC 61672-3:2006 |
| Conformance criterion (4.1) | deviation ≤ acceptance limit and uncertainty ≤ maximum-permitted uncertainty (95 %) |
| Environmental conditions (7.1) | 80–105 kPa; 20–26 °C; 25–70 % relative humidity |
| Uncertainty below 97 kPa (7.3) | 0,09 dB (≤ 3 kHz) and 0,14 dB (> 3 kHz) — when no data on the influence of conditions are available |
| Display resolution (4.2) | 0,1 dB → rectangular distribution, semi-range 0,05 dB |
| Frequency of electrical signal (8.3) | deviation ≤ ±0,25 % of the specified frequency |
| Self-generated noise (11.1.2) | most sensitive range, weighting A; averaging ≥ 30 s or 10 readings within 60 s (time weighting S, if available) |
| Calibrator (9.3) | conforming to the periodic-test requirements of IEC 60942 for its class (level, frequency, distortion) |
Standard
- Standard number
- PN-EN 61672-3:2014-03 (Polish version; implements EN 61672-3:2013 and IEC 61672-3:2013 [IDT])
- Title (PL)
- Elektroakustyka — Mierniki poziomu dźwięku — Część 3: Badania okresowe
- Title (EN)
- Electroacoustics — Sound level meters — Part 3: Periodic tests
Step-by-step procedure
-
Admission to testing
Instruction manual for the model and version, accessories, marking conforming to IEC 61672-1, sound calibrator; visual inspection, operation of controls and power supply status (clauses 3, 5, 6).
-
Conditions and instruments
Check of environmental conditions (80–105 kPa, 20–26 °C, 25–70 %) and recording at the start and end; instruments calibrated with traceability; calibrator conforming to IEC 60942 (clauses 7–9).
-
Indication and self-generated noise
Check and, if necessary, adjustment of the indication with the calibrator at the calibration check frequency (recorded before and after); measurement of self-generated noise with the microphone and with the electrical input device (clauses 10–11).
-
Tests of weightings and assessment
Acoustical test of the frequency weighting with free-field or random-incidence corrections, then tests with electrical signals (clauses 12–21); assessment of conformity according to 4.1 with the laboratory uncertainty for 95 %.
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Sound calibrator (clause 9) | Model from the meter manual or one with the same nominal level and frequency; conforming to IEC 60942, calibrated with a microphone of the model fitted to the meter | — |
| Multi-frequency calibrator, comparison coupler or electrostatic actuator (12.1) | For testing the frequency weighting: class 1 calibrator according to IEC 60942, working standard microphone according to IEC 61094-4, actuator according to IEC 61094-6 | — |
| Electrical input-signal device (3.3, 11.2) | Inserts signals equivalent to signals from the microphone — for tests with electrical signals and measurement of self-generated noise | — |
| Instruments for ambient conditions (7.2) | Static pressure, temperature and relative humidity — recorded at the start and end of testing | — |
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN 61672-3:2014-03 (Polish version; implements EN 61672-3:2013 and IEC 61672-3:2013 [IDT]) — “Electroacoustics — Sound level meters — Part 3: Periodic tests”.
How does this method work?
Periodic tests check a limited set of key properties of a sound level meter (indication with the calibrator, self-generated noise, frequency weightings and further tests with electrical signals) under laboratory conditions. A result conforms when the measured deviation is within the acceptance limit of IEC 61672-1 and the laboratory's measurement uncertainty (95 %) does not exceed the maximum-permitted uncertainty.
What is the measuring range and accuracy?
STATUS (read 17.11.2026): PN-EN 61672-3:2014-03 (Polish version, published 21-12-2016); implements EN 61672-3:2013 and IEC 61672-3:2013; replaces PN-EN 61672-3:2007; Scope (clause 1): class 1 and 2 meters according to IEC 61672-1 (2nd ed.); for IEC 61672-1:2002 — IEC 61672-3:2006; Conformance criterion (4.1): deviation ≤ acceptance limit and uncertainty ≤ maximum-permitted uncertainty (95 %); Environmental conditions (7.1): 80–105 kPa; 20–26 °C; 25–70 % relative humidity.
How long does the test take?
The procedure comprises 4 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Sound calibrator (clause 9), Multi-frequency calibrator, comparison coupler or electrostatic actuator (12.1), Electrical input-signal device (3.3, 11.2), Instruments for ambient conditions (7.2).
Where is this test used?
Periodic testing of class 1 and 2 sound level meters according to the second edition of IEC 61672-1 — time-weighting, integrating-averaging and integrating (IEC 61672-3:2013, clause 1); Calibration of sound level meters in an acoustic laboratory with an uncertainty budget according to the GUM (clause 4); Checking the sound calibrator used with the meter against the periodic-test requirements of IEC 60942 (clause 9); Multi-channel meters — check of the indication for each tested channel (10.1).
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 61672-3.