💧 Water Turbidity (Turbidimetry)

Physicochemistry PN-EN ISO 7027-1

Nephelometric determination of water turbidity using infrared scattered radiation at 90° angle. A fundamental drinking water quality parameter.

Overview

Water turbidity is a measure of reduction in water transparency caused by the presence of suspended and colloidal particles: clays, silts, microorganisms, organic and inorganic substances. It is expressed in NTU (Nephelometric Turbidity Units) or FNU (Formazin Nephelometric Units) — both units are numerically equivalent.

Turbidity is one of the key drinking water quality parameters — the Polish standard requires ≤1 NTU (recommended ≤0.3 NTU). High turbidity indicates potential microbiological contamination (particles shield pathogens from UV and chlorine disinfection), aesthetic problems, and technological issues.

The ISO 7027 method uses infrared light source (860 nm LED) and 90° detection (nephelometry) — this eliminates the effect of water color on results. For low turbidity waters (drinking water) nephelometry is used, for highly turbid waters (wastewater) — turbidimetry (beam attenuation measurement).

Method principle

The water sample is illuminated with monochromatic infrared radiation source (860 nm LED, according to ISO 7027). A photodiode positioned at 90°±2.5° to the incident beam measures the intensity of light scattered by suspended particles.

Nephelometry (90° angle): scattered radiation intensity is proportional to particle concentration. Used for low turbidity waters (<40 NTU, typically drinking water).

Turbidimetry (attenuation measurement): measures beam weakening passing through sample. Used for high turbidity waters (wastewater, very turbid waters >40 NTU).

Calibration using formazin suspension — primary turbidity standard (1 FNU = 1 NTU).

Applications

Key parameters

ParameterValue
Measurement range (nephelometry)0.05–400 NTU (FNU)
Measurement range (turbidimetry)40–4000 NTU
Resolution0.01 NTU (low range)
Wavelength860 nm ± 30 nm (infrared)
Detection angle90° ± 2.5° (ISO nephelometry)
Calibration standardFormazin (hydrazine + hexamethylenetetramine)

Standard

Standard number
PN-EN ISO 7027-1:2016-09
Title (PL)
Jakość wody — Oznaczanie mętności — Część 1: Metody ilościowe
Title (EN)
Water quality — Determination of turbidity — Part 1: Quantitative methods

Step-by-step procedure

1. Turbidimeter preparation

Turn on turbidimeter min. 15 minutes before measurement (light source stabilization). Check measurement chamber cleanliness.

⏱ Time: 15 min

2. Calibration

Perform calibration with standard set: 0.1 NTU (or <0.1), 20 NTU, 100 NTU, 800 NTU (or per manufacturer). Use StablCal standards or fresh formazin. Wipe cuvettes with silicone oil.

⏱ Time: 10 min

3. Sample preparation

Bring sample to room temperature. Gently mix (invert 3×) — do not shake vigorously (air bubbles increase result). Analyze immediately after sampling.

🌡 Temperature: 20–25°C

4. Filling cuvette

Fill clean cuvette with sample to indicated level (to manufacturer-specified mark). Close with cap. Wipe outer surface with cloth, apply thin layer of silicone oil.

5. Measurement

Insert cuvette into measurement chamber (marker orientation!). Close cover. Wait for reading stabilization (AutoRead function). Record result in NTU/FNU.

⏱ Time: 30–60 s

6. Measurement repetition

Remove cuvette, gently invert 3× (eliminate sedimentation), reinsert. Repeat measurement. Average of 2–3 measurements = final result.

7. Quality control

Every 10 samples measure control standard (e.g., 20 NTU). Acceptable deviation: ±10% or ±0.5 NTU (whichever is smaller). Blank: water <0.05 NTU.

8. Post-measurement cleaning

Rinse cuvettes with deionized water, leave to dry. Do not use detergents (may leave films). Clean measurement chamber periodically.

Required equipment and apparatus

EquipmentExampleIndicative price
Laboratory turbidimeter (nephelometer)Hach TU5200, Hach TL2300 ISO, WTW Turb 4308 000–25 000 PLN
Measurement cuvettesDedicated round glass cuvettes for turbidimeter (scratch-free, perfectly clean)50–150 PLN/pc
Silicone oil for cuvettesFor eliminating scratches and glass surface defects50–100 PLN
Cuvette cleaning clothsLint-free, microfiber20–50 PLN

Reagents, media and consumables

ReagentCASDetails
Formazin standard 4000 NTUPrimary turbidity standard: mixture of hydrazine sulfate and hexamethylenetetramine. Prepared fresh or purchased ready-made.
Hydrazine sulfate (NH₂)₂·H₂SO₄10034-93-2Formazin standard component. 10 g/L — solution I. WARNING: suspected carcinogen!
Hexamethylenetetramine (CH₂)₆N₄100-97-0Formazin standard component. 100 g/L — solution II. Mixing I+II produces formazin suspension.
Ready calibration standardsStablCal (Hach) or AMCO Clear sets — stable polymer standards replacing formazin. Shelf life up to 5 years.
Turbidity-free waterDeionized water filtered through 0.1 μm filter. Turbidity <0.02 NTU.

Health and safety (OHS)

🔍 Find a laboratory performing this test