📏 Tension/compression testing machines — calibration and verification of the force-measuring system with a force-proving instrument according to ISO 376: inspection, resolution of the indicator, three series with increasing force, relative indication and repeatability errors, class of the range; PN-EN ISO 7500-1

Metrology PN-EN ISO 7500-1

In short

Calibration and verification of tension and compression testing machines: the force indicated by the machine is compared with the force measured by a force-proving instrument complying with ISO 376 or generated by known masses. The test is performed according to PN-EN ISO 7500-1:2018-05 (Polish version; implements EN ISO 7500-1:2018 and ISO 7500-1:2018 [IDT]); Lower limit of the range (6.4.5): not less than r × 400 (class 0,5), × 200 (class 1), × 100 (class 2), × 67 (class 3). The procedure comprises 6 steps; it is used for: Tension and compression testing machines for static tests (ISO 7500-1:2018, clause 1), Each force range used and each force indicator, including accessories (slave pointer, recorder) affecting the measurement (6.1, 6.4.6), Hydraulic machines, including double-piston machines — effect of the piston position (6.1, 6.4.7).

At a glance

  • Standard: PN-EN ISO 7500-1:2018-05 (Polish version; implements EN ISO 7500-1:2018 and ISO 7500-1:2018 [IDT])
  • Category: Metrology
  • Procedure steps: 6
  • STATUS (PKN card, 14.11.2026): PN-EN ISO 7500-1:2018-05 (Polish version, published 23-11-2022); implements EN ISO 7500-1:2018 and ISO 7500-1:2018; replaces PN-EN ISO 7500-1:2016-02
  • Reference instrument (6.1): force-proving instrument according to ISO 376, class ≤ class of the machine; masses — force error within ±0,1 %
  • Temperature (6.4.2): 10 °C to 35 °C, recorded in the report; change of the instrument temperature in a series ≤ 2 °C

Overview

WHAT THE STANDARD COVERS. According to clause 1 of ISO 7500-1:2018 the document specifies the calibration and verification of tension/compression testing machines. The verification consists of: a general inspection of the testing machine, including its accessories for the force application; a calibration of the force-measuring system of the testing machine; and a confirmation that the performance properties of the testing machine achieve the limits given for a specified class. Note to clause 1: the document addresses the static calibration and verification of force-measuring systems — the calibration values are not necessarily valid for high-speed or dynamic testing applications. Caution: some of the tests specified in the document can lead to a hazardous situation.

ACCORDING TO THE TEXT OF ISO 7500-1:2018 — DEFINITIONS, SYMBOLS, INSPECTION, CALIBRATION PROCEDURE (foreword, clauses 1–6.4.8). The standard was prepared by ISO/TC 164 (Mechanical testing of metals), subcommittee SC 1 (Uniaxial testing); the fifth edition replaced the fourth (ISO 7500-1:2015) as a minor revision — the definitions of ΔF, Δm, Δg, E, E′, U and U′ in Table 1 were editorially changed. Normative reference: ISO 376 (calibration of force-proving instruments). Definitions (clause 3): calibration — establishing the relation between the force values indicated by the testing machine and those measured by one or more force-proving instruments (with uncertainties); verification — confirmation, on the basis of an analysis of the measurements, that the performance properties lie within the limits of a specified class. Symbols (Table 1) include: a — relative resolution of the force indicator, b — relative repeatability error, f₀ — relative zero error, q — mean relative indication error, v — relative reversibility error, r — resolution of the indicator, U — expanded uncertainty, k — coverage factor. Inspection (clause 5): calibration is carried out only if the machine is in good working order, after a general inspection according to Annex A; note: good metrological practice is to calibrate before maintenance or adjustment to establish the "as found" condition; information on compression platens — Annex B, uncertainty of the results — Annex C. General (6.1): calibration for each force range used and with all force indicators used; a machine with several force-measuring systems — each treated as a separate machine, likewise double-piston hydraulic machines; calibration with force-proving instruments, and below the lower limit of the smallest instrument — with known masses; with several instruments the largest force for the smaller one = the smallest force for the next larger one; a set of masses is treated as one instrument; calibration may be carried out at constant indicated forces Fᵢ or at constant reference forces F; instruments with documented traceability to the SI; force-proving instrument complying with ISO 376, of a class not worse than the class of the machine being calibrated; for masses the relative error of the generated force within ±0,1 %. Force from a mass: F = m·g·(1 − ρ_air/ρ_m), approximately F = m·g; ΔF/F = Δm/m + Δg/g. Resolution (6.2): analogue scale — uniform thickness of the graduation marks, pointer width approximately equal to the width of a mark, r = (pointer width / distance between marks) × scale interval, recommended ratios 1:2, 1:5 or 1:10, reading 1/10 of a scale division requires a mark spacing of at least 2,5 mm; digital scale — one increment of the numerical indicator; when the readings fluctuate by more than the calculated resolution (unloaded instrument, with motor, drive and control running), r = half the range of fluctuation plus one increment. Relative resolution a = r/Fᵢ × 100 determined at each calibration point (6.3). Procedure (6.4): alignment of the instrument to minimize bending (in compression — a platen with a ball nut if the machine has no ball cup); when attached with threaded studs, the instrument shall have been calibrated in the same way and is rotated through 120° between the series; ambient temperature between 10 °C and 35 °C, recorded in the report; the temperature of the instrument shall not change by more than 2 °C during a series; before calibration the force is applied at least three times from zero to the maximum force. Forces (6.4.5): three series with increasing force; for machines with more than five force levels — at least five levels in each series, approximately evenly spaced between 20 % and 100 % of the range; below 20 % of the range — at least five forces per full decade, with a nominal ratio between adjacent forces not greater than 2 (e.g. 10 %, 7 %, 4 %, 2 %, 1 %, 0,7 %, 0,4 %, 0,2 %, 0,1 %); the lower limit of the range not less than r × 400 (class 0,5), × 200 (class 1), × 100 (class 2), × 67 (class 3); for auto-ranging indicators at least two force steps on each part of the range where the resolution does not change. The indicator is set to zero before each series, zero is read about 30 s after the force is removed; relative zero error f₀ = Fᵢ₀/F_N × 100. Accessories (6.4.6): slave pointer, recorder — three series with the accessory (or without) and one supplementary series in the other condition for the lowest range; condition |Fᵢc − Fc|/Fc × 100 ≤ 1,5·q_al. Hydraulic machines (6.4.7): the effect of the piston position is checked in the lowest range — a different position in each series. Reversibility (6.4.8): on request — calibration with increasing and decreasing force, with an instrument calibrated for decreasing forces according to ISO 376; one decreasing series is enough.

STATUS (PKN cards and iTeh catalogue as of 14.11.2026). Card PN-EN ISO 7500-1:2018-05 (Polish version): title "Metale -- Wzorcowanie i sprawdzanie statycznych jednoosiowych maszyn wytrzymałościowych -- Część 1: Maszyny wytrzymałościowe rozciągające/ściskające -- Wzorcowanie i sprawdzanie układu pomiarowego siły", published 23-11-2022, 24 pages, KT 123 for Testing the Properties of Metals, "Wprowadza: EN ISO 7500-1:2018 [IDT], ISO 7500-1:2018 [IDT]", "Zastępuje: PN-EN ISO 7500-1:2016-02 - wersja angielska", ICS 19.060, 77.040.10. ISO 7500-1:2018 in the iTeh catalogue — published, stage 90.93 (standard confirmed); EN ISO 7500-1:2018 superseded EN ISO 7500-1:2015 (European foreword).

HOW MANY LABORATORIES AND IN WHAT FORM (accreditation scope database of the service, read 14.11.2026). ISO 7500-1 appears in 22 records at 7 Polish calibration laboratories: "PN-EN ISO 7500-1:2018-05" — AP 045, AP 130, AP 133, AP 207, AP 225; "PN-EN ISO 7500-1:2018" — AP 072, AP 181. The standard is accompanied by ASTM E4-24 (AP 133, AP 181, AP 207, AP 225), "PN-EN 12390-4:2020-03" (AP 130) and "PN-EN ISO 12390-4:2020-03 pkt. 4.2.2" (AP 225 — literal notation). Outside Poland — 4 Czech laboratories (7 records).

WHAT THE LABORATORIES CALIBRATE (scope database, complete items). Static testing machines and process equipment — for tensile and compressive forces; concrete compression testing machines (AP 072, AP 130). Ranges from 0,01 N (AP 133, AP 181, AP 207, AP 225) to 6 MN (AP 072, AP 133); CMC from 0,03 % (AP 207 — 1 N to 200 N, AP 225 — 1 N to 250 N) to 0,25 %, for larger forces 0,24 % (AP 045, AP 133, AP 207, AP 225) or 0,35 % (AP 130 — 3000 kN to 5000 kN). Some laboratories distinguish the CMC with masses and with control force-proving instruments of class 00 and 05 (AP 181) or work in the laboratory and at the customer's site (AP 045).

WHERE A SEEMINGLY CORRECT RESULT IS EASY TO GET. Adjustment before calibration — without an "as found" measurement it is not known in what condition the machine worked since the previous verification (note to clause 5). A force-proving instrument of a worse class than the declared class of the machine (6.1). Too few points below 20 % of the range or a lower limit of the range below r × 400/200/100/67 (6.4.5). No triple preloading and unstable temperature of the instrument (change > 2 °C in a series, 6.4.2–6.4.3). A machine with several indicators or force systems calibrated on only one of them (6.1). Static results transferred to high-speed or dynamic tests (note to clause 1).

WHAT WE DO NOT GIVE. Parts read: iTeh sample SIST EN ISO 7500-1:2018 (SIST title page, European foreword, contents, ISO foreword, clauses 1–3.1) and iTeh samples of ISO 7500-1:2018 in English and French (clauses 1–6.4.8). We do not give the content of 6.5 (assessment of the indicator: formulae for the indication and repeatability errors, agreement of two instruments), Table 2 (permissible values for the classes), clauses 7–9 (class of the range, report, intervals between verifications) or Annexes A–C — they are outside the sample. We have not read the Polish text of PN-EN ISO 7500-1:2018-05.

Method principle

The force indicated by the machine is compared with the force measured by a force-proving instrument complying with ISO 376 (of a class not worse than the class of the machine) or generated by known masses. After the inspection and triple preloading, three series with increasing force are performed — at constant indicated forces Fᵢ or constant reference forces F — at 10 °C to 35 °C, with at least five levels from 20 % to 100 % of the range and additional points in each decade below 20 %. From the results the relative indication and repeatability errors are calculated for each point, and from the resolution — the lower limit of the range for the given class.

Applications

Key parameters

ParameterValue
STATUS (PKN card, 14.11.2026)PN-EN ISO 7500-1:2018-05 (Polish version, published 23-11-2022); implements EN ISO 7500-1:2018 and ISO 7500-1:2018; replaces PN-EN ISO 7500-1:2016-02
Reference instrument (6.1)force-proving instrument according to ISO 376, class ≤ class of the machine; masses — force error within ±0,1 %
Temperature (6.4.2)10 °C to 35 °C, recorded in the report; change of the instrument temperature in a series ≤ 2 °C
Preloading (6.4.3)at least 3 times from zero to the maximum force
Series and points (6.4.5)3 series with increasing force; at least 5 levels from 20 % to 100 % of the range; below 20 % — at least 5 forces per decade, ratio of adjacent forces ≤ 2
Lower limit of the range (6.4.5)not less than r × 400 (class 0,5), × 200 (class 1), × 100 (class 2), × 67 (class 3)
Zero (6.4.5)zero setting before each series, reading about 30 s after removing the force; f₀ = Fᵢ₀/F_N × 100
Resolution (6.2–6.3)analogue: (pointer width / mark spacing) × scale interval, 1/10 of a division with spacing ≥ 2,5 mm; digital: 1 increment; a = r/Fᵢ × 100
Coverage in accreditation scopes (read 14.11.2026)7 Polish calibration laboratories, 22 records; 4 Czech

Standard

Standard number
PN-EN ISO 7500-1:2018-05 (Polish version; implements EN ISO 7500-1:2018 and ISO 7500-1:2018 [IDT])
Title (PL)
Metale — Wzorcowanie i sprawdzanie statycznych jednoosiowych maszyn wytrzymałościowych — Część 1: Maszyny wytrzymałościowe rozciągające/ściskające — Wzorcowanie i sprawdzanie układu pomiarowego siły
Title (EN)
Metallic materials — Calibration and verification of static uniaxial testing machines — Part 1: Tension/compression testing machines — Calibration and verification of the force-measuring system

Step-by-step procedure

  1. Inspection of the machine

    General inspection according to Annex A; calibration only if the machine is in good working order, preferably first in the "as found" condition (clause 5).

  2. Resolution of the indicator

    Determination of r for an analogue or digital scale, taking fluctuations of the readings into account; a = r/Fᵢ × 100 at each point (6.2–6.3).

  3. Alignment and conditions

    Instrument aligned without bending; temperature 10 °C to 35 °C, stable (≤ 2 °C in a series); triple preloading (6.4.1–6.4.3).

  4. Three measurement series

    Constant Fᵢ or constant F; at least 5 levels from 20 % to 100 % of the range and points in the decades below 20 %; zero setting and zero reading about 30 s after unloading (6.4.4–6.4.5).

  5. Accessories, piston, reversibility

    Supplementary series with or without the accessory; different piston positions in hydraulic machines; on request a decreasing series (6.4.6–6.4.8).

  6. Assessment and class

    Relative indication and repeatability errors and comparison with the permissible values of the class — clauses 6.5 and 7 (outside the sample read).

Required equipment and apparatus

EquipmentExampleIndicative price
Force-proving instrument (6.1)Complying with ISO 376, of a class not worse than the class of the machine being calibrated; traceability to the SI—
Known masses (6.1)For forces below the lower limit of the smallest instrument; a set of masses = one instrument; force error within ±0,1 %—
Platen with a ball nut, threaded studs (6.4.1)Alignment of the instrument in compression; with threaded studs — instrument calibrated in the same way, rotation through 120° between the series—

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN ISO 7500-1:2018-05 (Polish version; implements EN ISO 7500-1:2018 and ISO 7500-1:2018 [IDT]) — “Metallic materials — Calibration and verification of static uniaxial testing machines — Part 1: Tension/compression testing machines — Calibration and verification of the force-measuring system”.

How does this method work?

The force indicated by the machine is compared with the force measured by a force-proving instrument complying with ISO 376 (of a class not worse than the class of the machine) or generated by known masses. After the inspection and triple preloading, three series with increasing force are performed — at constant indicated forces Fᵢ or constant reference forces F — at 10 °C to 35 °C, with at least five levels from 20 % to 100 % of the range and additional points in each decade below 20 %.

What is the measuring range and accuracy?

STATUS (PKN card, 14.11.2026): PN-EN ISO 7500-1:2018-05 (Polish version, published 23-11-2022); implements EN ISO 7500-1:2018 and ISO 7500-1:2018; replaces PN-EN ISO 7500-1:2016-02; Reference instrument (6.1): force-proving instrument according to ISO 376, class ≤ class of the machine; masses — force error within ±0,1 %; Temperature (6.4.2): 10 °C to 35 °C, recorded in the report; change of the instrument temperature in a series ≤ 2 °C; Preloading (6.4.3): at least 3 times from zero to the maximum force.

How long does the test take?

The procedure comprises 6 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Force-proving instrument (6.1), Known masses (6.1), Platen with a ball nut, threaded studs (6.4.1).

Where is this test used?

Tension and compression testing machines for static tests (ISO 7500-1:2018, clause 1); Each force range used and each force indicator, including accessories (slave pointer, recorder) affecting the measurement (6.1, 6.4.6); Hydraulic machines, including double-piston machines — effect of the piston position (6.1, 6.4.7); In accreditation scopes: 7 Polish calibration laboratories — static testing machines, process equipment, concrete compression testing machines; 4 Czech.