🧪 Volatile Organic Compounds (VOC) in Water

Physicochemistry PN-EN ISO 11423

Determination of volatile organic compounds (VOC) in drinking water and wastewater by purge-and-trap technique with GC-MS detection. The method allows identification and determination of dozens of chlorinated, aromatic, and aliphatic compounds.

Overview

Volatile Organic Compounds (VOCs) are a group of organic substances with boiling point 50–250°C and vapor pressure ≥0.01 kPa at 20°C. In waters they occur as anthropogenic contaminants (solvents, fuels, degreasers) and chlorine disinfection by-products (trihalomethanes — THM). Many VOCs exhibit carcinogenic, mutagenic, or hepatotoxic properties.

The purge-and-trap (P&T) technique involves purging water sample with inert gas (He), which extracts volatile analytes from aqueous phase and transfers them to sorbent trap (Tenax/activated carbon/silica gel). Then the trap is rapidly heated, desorbing concentrated analytes to gas chromatograph (GC) column coupled with mass detector (MS). The method allows simultaneous identification and quantitative determination of over 60 different VOCs.

The P&T/GC-MS technique is the reference method EPA 524.2/524.4 for drinking water and EPA 8260 for groundwater and wastewater. Standard PN-EN ISO 11423 covers analogous procedure compliant with European requirements. Typical LODs are 0.01–0.5 µg/L, allowing control of limits e.g., 1 µg/L for benzene or 100 µg/L for total THM.

The method requires specialized equipment (P&T concentrator, GC-MS), but is indispensable in trace VOC analysis in waters — offering simultaneous identification by mass spectrum and sensitivity unattainable by other techniques.

Method principle

Water sample (5 mL) is placed in purge vessel and purged with inert gas (helium, 40 mL/min, 11 min). Volatile organic compounds are purged from aqueous phase and transferred to sorbent trap (Tenax TA + activated carbon + silica gel). After purging, trap is rapidly heated (250°C, 4 min) in helium stream, desorbing analytes to gas chromatograph equipped with capillary column (e.g., DB-624, 30 m × 0.25 mm × 1.4 µm). GC temperature program separates analytes, which are identified based on mass spectra (MS in full scan mode or SIM). Quantification is performed from calibration curve with internal standard (e.g., fluorobenzene, 1,2-dichloroethane-d4).

Applications

Key parameters

ParameterValue
Measurement range0.01–200 µg/L (depending on compound)
Limit of detection (LOD)0.01–0.5 µg/L
Number of analytes60+ VOCs in one analysis
Sample volume5 mL (typical)
Analysis time (P&T + GC)30–50 minutes
IdentificationMass spectrum + retention time (dual confirmation)

Standard

Standard number
PN-EN ISO 11423:2002
Title (PL)
Jakość wody — Oznaczanie lotnych związków organicznych — Metoda chromatografii gazowej z techniką purge-and-trap i detekcją MS
Title (EN)
Water quality — Determination of volatile organic compounds — Purge-and-trap gas chromatography with mass spectrometric detection

Step-by-step procedure

1. Sample collection

Collect sample in 40 mL VOC vial with no headspace (no air bubbles). Add 25 mg ascorbic acid (chlorinated water). Store at 4°C, analyze within 14 days.

🌡 Temperature: 4°C

2. Standard preparation

From VOC concentrate in methanol prepare calibration series (5 points): 0.5; 1.0; 5.0; 10; 50 µg/L. Add internal standards (IS) to final concentration 10 µg/L.

⏱ Time: 20 min

3. P&T/GC-MS system calibration

Measure calibration series. Calculate response factors (RF) for each analyte. Check RSD RF < 20%. Average RF or linear regression.

⏱ Time: 3–4 h

4. Sample transfer to purge vessel

Transfer 5 mL sample to purge vessel. Add internal and surrogate standards (each 10 µg/L). Close vessel.

⏱ Time: 3 min

5. Purge phase

Purge sample with helium (40 mL/min) for 11 minutes. VOCs transfer from aqueous phase to sorbent trap (Tenax/carbon/silica).

⏱ Time: 11 min

6. Dry purge

Purge trap with dry helium for 1 min to remove water vapor before desorption.

⏱ Time: 1 min

7. Thermal desorption

Heat trap to 250°C for 4 minutes, desorbing analytes with helium stream onto GC column (DB-624). Transfer line 150°C.

⏱ Time: 4 min • 🌡 Temperature: 250°C

8. GC separation

Temperature program: 35°C (2 min) → 8°C/min → 200°C (2 min). Carrier gas: He 1.2 mL/min. Run time ~25 min.

⏱ Time: 25 min

9. MS detection

MS in Full Scan mode (m/z 35–300) or SIM. Identification: retention time + mass spectrum (NIST library match >70%).

⏱ Time: continuous

10. Calculations and quality control

Calculate concentrations based on IS. Check surrogate recovery (70–130%). Blank < LOD. BFB tune check every 12 h.

⏱ Time: 15 min

Required equipment and apparatus

EquipmentExampleIndicative price
Gas chromatograph GC-MSAgilent 8890 GC + 5977C MSD, Shimadzu GCMS-QP2020 NX, Thermo TRACE 1310 + ISQ 7000400,000–800,000 PLN
Purge-and-trap concentratorTeledyne Tekmar Atomx, EST Encon, OI Analytical Eclipse100,000–200,000 PLN
GC capillary columnAgilent DB-624 (30 m × 0.25 mm × 1.4 µm), Restek Rtx-VMS2,000–4,000 PLN
Autosampler for VOC vialsTeledyne Tekmar Versa, EST Centurion60,000–120,000 PLN
VOC vials 40 mL with PTFE septumGlass vials with screw cap and PTFE/silicone septum, no headspace3–8 PLN/pc
Helium 5.0 cylinderHelium 99.999%, carrier and purge gas, 50 L cylinder600–1,200 PLN/cylinder

Reagents, media and consumables

ReagentCASDetails
VOC standard mix (EPA 524.2)mixtureCertified standard 60+ VOCs in methanol, 200–2000 µg/mL, e.g., Supelco/Restek
Internal standards (IS)Fluorobenzene (CAS 462-06-6), 1,2-dichloroethane-d4 (CAS 17060-07-0), chlorobenzene-d5 (CAS 3114-55-4)
Surrogate standards4-Bromofluorobenzene (CAS 460-00-4), dibromofluoromethane (CAS 1868-53-7), toluene-d8 (CAS 2037-26-5)
Anhydrous methanol (standard solvent)67-56-1Pesticide grade quality, free from VOCs
Ascorbic acid50-81-7Dechlorination of chlorinated drinking water samples, 25 mg per 40 mL vial
Helium 5.07440-59-7GC carrier gas and purge gas, purity 99.999%

Health and safety (OHS)

🔍 Find a laboratory performing this test