🍎 Preservatives (sorbates, benzoates) — HPLC

Food Chemistry PN-EN 12856

Determination of sorbic acid and benzoic acid (preservatives E200, E210) in food by HPLC-UV. Control of permitted food additives.

Overview

Sorbic acid (E200) and benzoic acid (E210) and their salts are the most commonly used preservatives in food. Their permitted concentrations are regulated by EU Regulation 1333/2008: sorbic acid 150–2000 mg/kg (depending on product category), benzoic acid 150–2000 mg/kg.

The HPLC-UV method involves extraction of preservatives from food sample, purification, separation on a C18 column, and UV detection: sorbic acid 255 nm, benzoic acid 228 nm. The method is simple, fast, and selective.

Method principle

The sample is homogenized with water, ultrasonically extracted, and deproteinated with Carrez reagents. After filtration (0.22 µm), the solution is analyzed by HPLC on a C18 column with buffered mobile phase (ammonium acetate pH 4.2 + methanol). UV detection at 228 nm (benzoic acid) and 255 nm (sorbic acid). Quantitative determination using a calibration curve.

Applications

Key parameters

ParameterValue
Substances determinedSorbic acid (E200), benzoic acid (E210)
Measurement range5–5000 mg/kg
Limit of detection2 mg/kg
Wavelength228 nm (benzoic), 255 nm (sorbic)
Precision (CV)2–5%

Standard

Standard number
PN-EN 12856:2001
Title (PL)
Artykuły żywnościowe — Oznaczanie kwasu sorbowego i benzoesowego metodą HPLC
Title (EN)
Foodstuffs — Determination of acesulfame-K, aspartame and saccharin — HPLC method

Step-by-step procedure

1. Extraction

Weigh 5 g sample, add 40 mL water, ultrasonic 15 min. Deproteinate with Carrez I+II. Fill to 50 mL.

⏱ Time: 25 min

2. Filtration

Filter through 0.22 µm PVDF into vial.

⏱ Time: 5 min

3. Calibration

Standards: 5, 10, 50, 100, 500 mg/L mixture of sorbic and benzoic acid.

⏱ Time: 15 min

4. HPLC analysis

C18, 20 mM ammonium acetate pH 4.2 / MeOH gradient, 1 mL/min, UV 228+255 nm.

⏱ Time: 15 min

5. Calculations

Concentration from calibration curve, conversion to mg/kg product.

⏱ Time: 5 min

Required equipment and apparatus

EquipmentExampleIndicative price
HPLC system with UV/DAD detectorAgilent 1260, Shimadzu Nexera, Waters Alliance200 000–400 000 PLN
C18 columnAgilent Zorbax Eclipse Plus C18, 150×4.6 mm, 5 µm1 500–3 500 PLN
Ultrasonic bathBandelin Sonorex RK 5103 000–8 000 PLN
Syringe filters 0.22 µmPVDF or nylon, ø 13 mm1–3 PLN/pc

Reagents, media and consumables

ReagentCASDetails
Sorbic acid (standard)110-44-1Purity ≥99%, CRM
Benzoic acid (standard)65-85-0Purity ≥99.5%, CRM
Ammonium acetate631-61-8Mobile phase: 20 mmol/L, pH 4.2
Methanol HPLC grade67-56-1Mobile phase component (gradient 10–60%)

Health and safety (OHS)