🛡️ Electromagnetic compatibility — immunity of electrical and electronic equipment at industrial locations: generic standard, continuous and transient, conducted and radiated disturbances, electrostatic discharges, performance criteria A, B, C, range 0 Hz – 400 GHz; PN-EN IEC 61000-6-2
In short
Requirements for immunity to electromagnetic disturbances of electrical and electronic equipment intended for use at industrial locations, indoors and outdoors, including equipment supplied from a DC network or battery operated. The test is performed according to PN-EN IEC 61000-6-2:2019-04 (English and Polish version; implements EN IEC 61000-6-2:2019 and IEC 61000-6-2:2016 [IDT]); Frequency range (clause 1): from 0 Hz to 400 GHz; no tests where no requirements are specified. The procedure comprises 4 steps; it is used for: Electrical and electronic equipment intended for use at industrial locations, indoors and outdoors, where no product or product-family EMC immunity standard exists (clause 1), Equipment connected to a DC distribution network or battery operated, used at industrial locations (clause 1), Immunity tests of the enclosure port, signal/control ports and DC and AC power ports (clauses 7, 9).
At a glance
- Standard: PN-EN IEC 61000-6-2:2019-04 (English and Polish version; implements EN IEC 61000-6-2:2019 and IEC 61000-6-2:2016 [IDT])
- Category: Electronics and EMC
- Procedure steps: 4
- STATUS (as of 20.11.2026): PKN cards: PN-EN IEC 61000-6-2:2019-04 “Kompatybilność elektromagnetyczna (EMC) -- Część 6-2: Normy ogólne -- Norma dotycząca odporności w środowiskach przemysłowych” — English version (published 08-04-2019, 28 pages) and Polish version (published 19-02-2025, 24 pages), KT 104 Electromagnetic Compatibility, implements EN IEC 61000-6-2:2019 and IEC 61000-6-2:2016 [IDT], replaces PN-EN 61000-6-2:2008 with corrigenda Ap1:2009 and Ap2:2009 (Polish versions); the cards give no withdrawal date. At the publisher, IEC 61000-6-2:2016 has the status “Published” (iTeh catalogue card).
- Frequency range (clause 1): from 0 Hz to 400 GHz; no tests where no requirements are specified
- Industrial location (3.7): separate power network supplied from a high- or medium-voltage transformer, dedicated for the supply of the installation
Overview
WHAT THE STANDARD COVERS. According to clause 1 of IEC 61000-6-2:2016 the standard covers EMC immunity requirements for electrical and electronic equipment intended for use at industrial locations (definition 3.7), indoors and outdoors, including equipment directly connected to a DC distribution network or battery operated. It covers requirements in the frequency range from 0 Hz to 400 GHz; no tests are performed at frequencies where no requirements are specified. It defines immunity test requirements for continuous and transient, conducted and radiated disturbances, including electrostatic discharges, separately for each port. It is applied when no relevant product or product-family EMC immunity standard exists. The levels do not cover extreme cases with an extremely low probability of occurrence; safety considerations are not covered. Where special mitigation measures produce an environment equivalent to the residential, commercial or light-industrial environment, the generic standard for that environment or the product standard may be applied.
ACCORDING TO THE TEXT OF IEC 61000-6-2:2016 (third edition; publisher's sample: foreword, introduction, clauses 1–9 without Tables 1–4). The standard was prepared by IEC technical committee TC 77. The third edition replaces the second edition of 2005; changes: improved environmental description, extended frequency range of the radio-frequency electromagnetic field test according to IEC 61000-4-3 and amended levels at particular frequencies, changed repetition frequency in the fast transient test according to IEC 61000-4-4, requirements according to IEC 61000-4-34, revised test levels, consideration of measurement uncertainty and a new Annex A. An industrial location (3.7) has a separate power network supplied from a high- or medium-voltage transformer dedicated for the supply of the installation (examples: metalworking, pulp and paper, chemical plants, car production, farm buildings, high-voltage areas of airports). A long distance line (3.5) is a line connected to a signal/control port which inside a building is longer than 30 m or which leaves the building.
PERFORMANCE CRITERIA (clause 4). The functional description and the performance criteria are provided by the manufacturer and noted in the test report. Criterion A: the equipment continues to operate as intended during and after the test, with no degradation below the performance level specified by the manufacturer. Criterion B: the equipment continues to operate as intended after the test; during the test degradation of performance is allowed, but no change of actual operating state or stored data. Criterion C: temporary loss of function is allowed, provided the function is self-recoverable or can be restored by the operation of the controls. Equipment that becomes dangerous or unsafe as a result of the tests fails the test.
CONDITIONS AND ASSESSMENT (clauses 5–9). The equipment is tested in the expected most susceptible operating mode (e.g. identified by limited pre-tests), consistent with normal applications, in the configuration giving maximum susceptibility; with the minimum representative configuration of auxiliary equipment (which may be simulated) and with the external protection devices required by the user's manual. With many similar ports a sufficient number is selected, and the selection is justified in the test report. Tests are carried out at one single set of environmental parameters and at the rated supply voltage, individually, in any order; identical units may be tested in parallel (recorded in the report). Tests are applied only to existing ports according to Tables 1–4; a decision not to test shall be justified. The instrumentation uncertainty is assessed according to IEC TR 61000-1-6 or the corresponding basic standard. Test descriptions, generators and set-ups are in the basic standards: IEC 61000-4-2 (electrostatic discharge), 61000-4-3 (radiated radio-frequency electromagnetic field), 61000-4-4 (electrical fast transient/burst), 61000-4-5 (surge), 61000-4-6 (conducted disturbances induced by radio-frequency fields), 61000-4-8 (power frequency magnetic field), 61000-4-11 and 61000-4-34 (voltage dips, short interruptions and voltage variations), 61000-4-20, 61000-4-21 and 61000-4-22 (TEM waveguides, reverberation chambers, fully anechoic rooms).
WHAT WE DO NOT GIVE HERE. The sample ends at clause 9 before Tables 1–4, so we do not reproduce the test levels for the enclosure port, signal/control ports or DC and AC power ports, the performance criteria assigned to them, or the content of Annex A.
STATUS (as of 20.11.2026). PKN cards: PN-EN IEC 61000-6-2:2019-04 “Kompatybilność elektromagnetyczna (EMC) -- Część 6-2: Normy ogólne -- Norma dotycząca odporności w środowiskach przemysłowych” — English version (published 08-04-2019, 28 pages) and Polish version (published 19-02-2025, 24 pages), KT 104 Electromagnetic Compatibility, implements EN IEC 61000-6-2:2019 and IEC 61000-6-2:2016 [IDT], replaces PN-EN 61000-6-2:2008 with corrigenda Ap1:2009 and Ap2:2009 (Polish versions); the cards give no withdrawal date. At the publisher, IEC 61000-6-2:2016 has the status “Published” (iTeh catalogue card).
Method principle
The equipment, in its most susceptible operating mode, is subjected port by port to continuous and transient, conducted and radiated disturbances, including electrostatic discharges, from the generators and in the set-ups of the IEC 61000-4-x basic standards, individually and in any order, and its behaviour during and after the test is assessed against performance criteria A, B or C defined by the manufacturer.
Applications
- Electrical and electronic equipment intended for use at industrial locations, indoors and outdoors, where no product or product-family EMC immunity standard exists (clause 1)
- Equipment connected to a DC distribution network or battery operated, used at industrial locations (clause 1)
- Immunity tests of the enclosure port, signal/control ports and DC and AC power ports (clauses 7, 9)
- Definition and recording of performance criteria A, B, C in the test report (clause 4)
Key parameters
| Parameter | Value |
|---|---|
| STATUS (as of 20.11.2026) | PKN cards: PN-EN IEC 61000-6-2:2019-04 “Kompatybilność elektromagnetyczna (EMC) -- Część 6-2: Normy ogólne -- Norma dotycząca odporności w środowiskach przemysłowych” — English version (published 08-04-2019, 28 pages) and Polish version (published 19-02-2025, 24 pages), KT 104 Electromagnetic Compatibility, implements EN IEC 61000-6-2:2019 and IEC 61000-6-2:2016 [IDT], replaces PN-EN 61000-6-2:2008 with corrigenda Ap1:2009 and Ap2:2009 (Polish versions); the cards give no withdrawal date. At the publisher, IEC 61000-6-2:2016 has the status “Published” (iTeh catalogue card). |
| Frequency range (clause 1) | from 0 Hz to 400 GHz; no tests where no requirements are specified |
| Industrial location (3.7) | separate power network supplied from a high- or medium-voltage transformer, dedicated for the supply of the installation |
| Long distance line (3.5) | signal/control port line longer than 30 m inside a building or leaving the building |
| Criterion A (clause 4) | operation as intended during and after the test, no degradation below the manufacturer's level |
| Criterion B (clause 4) | operation as intended after the test; degradation allowed during the test without change of operating state and data |
| Criterion C (clause 4) | temporary loss of function, self-recoverable or restored by the controls |
| Uncertainty (clause 8) | according to IEC TR 61000-1-6 or the corresponding basic standard |
Standard
- Standard number
- PN-EN IEC 61000-6-2:2019-04 (English and Polish version; implements EN IEC 61000-6-2:2019 and IEC 61000-6-2:2016 [IDT])
- Title (PL)
- Kompatybilność elektromagnetyczna (EMC) — Część 6-2: Normy ogólne — Norma dotycząca odporności w środowiskach przemysłowych
- Title (EN)
- Electromagnetic compatibility (EMC) — Part 6-2: Generic standards — Immunity standard for industrial environments
Step-by-step procedure
-
Scope and criteria
Check that no product EMC immunity standard exists; identify ports and the manufacturer's performance criteria (clauses 1, 4, 7).
-
Operating mode and configuration
Most susceptible operating mode, configuration giving maximum susceptibility, minimum configuration of auxiliary equipment, protection devices from the manual, justified selection of similar ports (clause 5).
-
Tests
Tests port by port according to Tables 1–4 and the basic standards, individually, in any order; identical units may be tested in parallel (clause 9).
-
Assessment and report
Assessment against criteria A, B, C; equipment that becomes unsafe fails; the report gives configuration, operating mode, criteria and the justification of tests not performed (clauses 4, 5, 7).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Test generators and set-ups according to basic standards (clauses 2, 9) | Electrostatic discharge (IEC 61000-4-2), electrical fast transient/burst (IEC 61000-4-4), surge (IEC 61000-4-5), voltage dips and interruptions (IEC 61000-4-11, IEC 61000-4-34) | — |
| Radiated field test facilities (clause 2) | Test according to IEC 61000-4-3, TEM waveguides (IEC 61000-4-20), reverberation chambers (IEC 61000-4-21), fully anechoic rooms (IEC 61000-4-22) | — |
| Conducted disturbance and magnetic field test set-ups (clause 2) | Conducted disturbances induced by radio-frequency fields (IEC 61000-4-6), power frequency magnetic field (IEC 61000-4-8) | — |
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN IEC 61000-6-2:2019-04 (English and Polish version; implements EN IEC 61000-6-2:2019 and IEC 61000-6-2:2016 [IDT]) — “Electromagnetic compatibility (EMC) — Part 6-2: Generic standards — Immunity standard for industrial environments”.
How does this method work?
The equipment, in its most susceptible operating mode, is subjected port by port to continuous and transient, conducted and radiated disturbances, including electrostatic discharges, from the generators and in the set-ups of the IEC 61000-4-x basic standards, individually and in any order, and its behaviour during and after the test is assessed against performance criteria A, B or C defined by the manufacturer.
What is the measuring range and accuracy?
STATUS (as of 20.11.2026): PKN cards: PN-EN IEC 61000-6-2:2019-04 “Kompatybilność elektromagnetyczna (EMC) -- Część 6-2: Normy ogólne -- Norma dotycząca odporności w środowiskach przemysłowych” — English version (published 08-04-2019, 28 pages) and Polish version (published 19-02-2025, 24 pages), KT 104 Electromagnetic Compatibility, implements EN IEC 61000-6-2:2019 and IEC 61000-6-2:2016 [IDT], replaces PN-EN 61000-6-2:2008 with corrigenda Ap1:2009 and Ap2:2009 (Polish versions); the cards give no withdrawal date. At the publisher, IEC 61000-6-2:2016 has the status “Published” (iTeh catalogue card).; Frequency range (clause 1): from 0 Hz to 400 GHz; no tests where no requirements are specified; Industrial location (3.7): separate power network supplied from a high- or medium-voltage transformer, dedicated for the supply of the installation; Long distance line (3.5): signal/control port line longer than 30 m inside a building or leaving the building.
How long does the test take?
The procedure comprises 4 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Test generators and set-ups according to basic standards (clauses 2, 9), Radiated field test facilities (clause 2), Conducted disturbance and magnetic field test set-ups (clause 2).
Where is this test used?
Electrical and electronic equipment intended for use at industrial locations, indoors and outdoors, where no product or product-family EMC immunity standard exists (clause 1); Equipment connected to a DC distribution network or battery operated, used at industrial locations (clause 1); Immunity tests of the enclosure port, signal/control ports and DC and AC power ports (clauses 7, 9); Definition and recording of performance criteria A, B, C in the test report (clause 4).
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN IEC 61000-6-2.