🌿 Cobalt and its compounds in workplace air — atomic absorption spectrometry with a graphite furnace; lowest determinable concentration per the standard about 0.002 mg/m³ (720 l of air); NDS 0.02 mg/m³ as Co; PN-Z-04454:2014-08

Environment PN-Z-04454

In short

Determination of cobalt and its compounds in workplace air by atomic absorption spectrometry with a graphite furnace according to the 2014 Polish national standard. The test is performed according to PN-Z-04454:2014-08 (Polish version); STATUS (as of 18.11.2026): current, no withdrawal note; the earlier PN-Z-04291:2003 (flame AAS) withdrawn 11-08-2014; according to CIOP-PIB 2022 the standard is due for amendment. The procedure comprises 5 steps; it is used for: Measurement of cobalt and its inorganic compounds in workplace air — assessment against the NDS of 0.02 mg/m³ as Co (Journal of Laws 2026 item 447, item 328).

At a glance

  • Standard: PN-Z-04454:2014-08 (Polish version)
  • Category: Environment
  • Procedure steps: 5
  • STATUS (as of 18.11.2026): current, no withdrawal note; the earlier PN-Z-04291:2003 (flame AAS) withdrawn 11-08-2014; according to CIOP-PIB 2022 the standard is due for amendment
  • Edition (from the PKN card): PN-Z-04454:2014-08, Polish version; published 01-08-2014, 8 pages, KT 159, ICS 13.040.30
  • Lowest determinable concentration (from the card): about 0.002 mg/m³ for 720 l of air — 1/10 of the NDS

Overview

STATUS (as of 18.11.2026). The PKN card of PN-Z-04454:2014-08 (Polish version) has no withdrawal note — the standard is CURRENT: published 01-08-2014, 8 pages, KT 159 Chemical and Dust Hazards in the Working Environment, ICS 13.040.30; the card has no “Replaces” field. The earlier standard for this substance, PN-Z-04291:2003 (flame atomic absorption spectrometry, membrane filter, aqua regia and nitric acid, 0.01 mg/m³), carries the label “Withdrawn” and the withdrawal date 11-08-2014, without a “Replaced by” field — the cards do not link the two standards. According to the 2022 CIOP-PIB article, PN-Z-04454:2014 “requires amendment”, because the Interdepartmental Commission for NDS and NDN proposed lowering the NDS for the respirable fraction to 0.001 mg/m³; there is no new standard in the PKN catalogue.

WHAT THIS STANDARD COVERS. Scope from the PKN catalogue card, in full (our translation): “This standard gives a method for the determination of cobalt (CAS No. 7440-48-4) and its compounds in workplace air using atomic absorption spectrometry with a graphite furnace. The method is used when checking sanitary and hygienic conditions. The lowest concentration of cobalt and its compounds that can be determined under the air sampling and determination conditions described in the standard is about 0.002 mg/m3 (for an air volume of 720 l)”.

LIMIT VALUES AND THE RULE ON METHODS. List of NDS values (Journal of Laws 2026 item 447), item 328 “Kobalt [7440-48-4] i jego związki nieorganiczne − w przeliczeniu na Co” (cobalt and its inorganic compounds, as Co): NDS 0.02 mg/m³, no NDSCh and no note on the fraction; the transitional provisions of the amending acts do not mention cobalt. The lowest determinable concentration according to the card (about 0.002 mg/m³) is 1/10 of the NDS. The result is reported as cobalt, not as the mass of the compound. § 16 of the regulation on tests and measurements (consolidated text Journal of Laws 2025 item 949): tests and measurements are performed by methods from Polish Standards and, where there are none, by methods recommended and validated by the laboratories referred to in § 15; this standard is a current Polish Standard, the regulation does not name it.

PROCEDURE ACCORDING TO THE CIOP-PIB PROCEDURE (second source; figures of the procedure, not of the standard). E. Gawęda, “Cobalt and its compounds — method of determination” (Podstawy i Metody Oceny Środowiska Pracy 2010, no. 1(63)) — a procedure with the same limit of quantification as the card of the 2014 standard, about 0.002 mg/m³ for 720 l of air. Up to 720 l of air is drawn through a membrane filter at up to 2 l/min (personal dosimetry) or up to 20 l/min (static measurement); filters with samples can be kept in a refrigerator for 2 weeks. The sample is first treated with aqua regia at room temperature, then digested on a hot plate successively in aqua regia and in concentrated nitric acid; the solution for analysis is prepared in dilute nitric acid and cobalt is determined by atomic absorption spectrometry with a graphite furnace. The 2022 amendment (P. Wasilewski, no. 1(111)) uses a cellulose ester filter, digestion in nitric(V) and hydrochloric acid at elevated temperature, the wavelength 242.5 nm and a range of 0.0001–0.002 mg/m³ for 2880 l of air.

WHAT WE DO NOT GIVE HERE. We did not have the text of PN-Z-04454:2014-08 — PKN sells it (8 pages). We do not know whether the standard adopted the filter, digestion and conditions of the 2010 procedure unchanged, and we do not give the furnace temperature programme, the modifier or the validation data of the standard.

Method principle

The aerosol of cobalt and its compounds is retained on a membrane filter, the sample is digested in aqua regia and concentrated nitric acid, and cobalt in the solution in dilute nitric acid is determined by atomic absorption spectrometry with a graphite furnace (CIOP-PIB procedure 2010; card of the standard: AAS with a graphite furnace, about 0.002 mg/m³ for 720 l). The result expressed as Co is compared with the NDS of 0.02 mg/m³ (Journal of Laws 2026 item 447, item 328).

Applications

Key parameters

ParameterValue
STATUS (as of 18.11.2026)current, no withdrawal note; the earlier PN-Z-04291:2003 (flame AAS) withdrawn 11-08-2014; according to CIOP-PIB 2022 the standard is due for amendment
Edition (from the PKN card)PN-Z-04454:2014-08, Polish version; published 01-08-2014, 8 pages, KT 159, ICS 13.040.30
Lowest determinable concentration (from the card)about 0.002 mg/m³ for 720 l of air — 1/10 of the NDS
Limit value against which the result is assessedNDS 0.02 mg/m³ as Co, no NDSCh — Journal of Laws 2026 item 447, item 328
Sampling and digestion (CIOP-PIB procedure 2010)membrane filter; up to 720 l, ≤ 2 l/min (personal) or ≤ 20 l/min (static); aqua regia and concentrated nitric acid; solution in dilute nitric acid

Standard

Standard number
PN-Z-04454:2014-08 (Polish version)
Title (PL)
Ochrona czystości powietrza — Oznaczanie kobaltu i jego związków na stanowiskach pracy metodą absorpcyjnej spektrometrii atomowej z kuwetą grafitową
Title (EN)
Air purity protection — Determination of cobalt and its compounds at workplaces by graphite furnace atomic absorption spectrometry

Step-by-step procedure

  1. Status of the document

    PKN card: PN-Z-04454:2014-08 current (as of 18.11.2026).

  2. Sampling

    CIOP-PIB procedure 2010: membrane filter in a holder, up to 720 l of air, up to 2 l/min (personal) or up to 20 l/min (static); filters in a refrigerator for up to 2 weeks.

  3. Digestion

    Aqua regia at room temperature, then on a hot plate aqua regia and concentrated nitric acid; solution in dilute nitric acid (CIOP-PIB procedure 2010).

  4. Determination

    Atomic absorption spectrometry with a graphite furnace (PKN card); lowest determinable concentration about 0.002 mg/m³ for 720 l.

  5. Result and assessment

    Cobalt concentration in mg/m³ as Co; comparison with the NDS of 0.02 mg/m³ (Journal of Laws 2026 item 447).

Required equipment and apparatus

EquipmentExampleIndicative price
Atomic absorption spectrometer with a graphite furnaceTechnique of the standard according to the PKN card; the temperature programme of the standard is not known to us (in the CIOP-PIB 2022 amendment — 242.5 nm)—
Membrane filters in holders and suction pumpCIOP-PIB procedure 2010: up to 720 l of air, up to 2 l/min or up to 20 l/min—
Hot plate and PTFE beakersDigestion of the filters (CIOP-PIB procedure 2010)—

Reagents, media and consumables

ReagentCASDetails
Aqua regia—For pre-treatment and digestion (CIOP-PIB procedure 2010)
Nitric(V) acid, concentrated and dilute7697-37-2Digestion and solution for analysis (CIOP-PIB procedure 2010)
Cobalt standard solution—For the calibration curve (CIOP-PIB procedure 2010); the concentrations of the standard are not known to us

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-Z-04454:2014-08 (Polish version) — “Air purity protection — Determination of cobalt and its compounds at workplaces by graphite furnace atomic absorption spectrometry”.

How does this method work?

The aerosol of cobalt and its compounds is retained on a membrane filter, the sample is digested in aqua regia and concentrated nitric acid, and cobalt in the solution in dilute nitric acid is determined by atomic absorption spectrometry with a graphite furnace (CIOP-PIB procedure 2010; card of the standard: AAS with a graphite furnace, about 0.002 mg/m³ for 720 l).

What is the measuring range and accuracy?

STATUS (as of 18.11.2026): current, no withdrawal note; the earlier PN-Z-04291:2003 (flame AAS) withdrawn 11-08-2014; according to CIOP-PIB 2022 the standard is due for amendment; Edition (from the PKN card): PN-Z-04454:2014-08, Polish version; published 01-08-2014, 8 pages, KT 159, ICS 13.040.30; Lowest determinable concentration (from the card): about 0.002 mg/m³ for 720 l of air — 1/10 of the NDS; Limit value against which the result is assessed: NDS 0.02 mg/m³ as Co, no NDSCh — Journal of Laws 2026 item 447, item 328.

How long does the test take?

The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Atomic absorption spectrometer with a graphite furnace, Membrane filters in holders and suction pump, Hot plate and PTFE beakers.

Where is this test used?

Measurement of cobalt and its inorganic compounds in workplace air — assessment against the NDS of 0.02 mg/m³ as Co (Journal of Laws 2026 item 447, item 328).

What safety precautions apply?

Work with concentrated acids is carried out wearing personal protective equipment in an efficient laboratory fume hood (CIOP-PIB procedure 2022); TECHNICAL NOTE: the NDS concerns cobalt and its inorganic compounds as Co — the result is reported as cobalt, not as the mass of the compound.

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-Z-04454.

🔍 Find a laboratory performing this test