🔊 Sound calibrators — requirements for classes LS (laboratory standard), 1 and 2, pistonphones LS/M and 1/M, reference conditions (23 °C, 101,325 kPa, 50 %), nominal level and frequency, stabilization time up to 30 s, conformance criterion with measurement uncertainty; PN-EN IEC 60942
In short
Performance requirements for three classes of sound calibrators: LS (laboratory standard), 1 and 2, with the designations LS/M and 1/M for pistonphones requiring a correction for static pressure. The test is performed according to PN-EN IEC 60942:2018-06 (Polish version; implements EN IEC 60942:2018 and IEC 60942:2017 [IDT]); STATUS (read 17.11.2026): PN-EN IEC 60942:2018-06 (Polish version, published 06-12-2019); implements EN IEC 60942:2018 and IEC 60942:2017; replaces PN-EN 60942:2005. The procedure comprises 4 steps; it is used for: Sound calibrators of classes LS, 1 and 2 and pistonphones LS/M and 1/M (IEC 60942:2017, clause 1, table 1), Determination of the pressure sensitivity of measurement microphones and checking or adjusting the sensitivity of sound level meters and acoustical systems (introduction), Class 1 calibrators for class 1 sound level meters, class 2 for class 2 meters according to IEC 61672-1.
At a glance
- Standard: PN-EN IEC 60942:2018-06 (Polish version; implements EN IEC 60942:2018 and IEC 60942:2017 [IDT])
- Category: Metrology
- Procedure steps: 4
- STATUS (read 17.11.2026): PN-EN IEC 60942:2018-06 (Polish version, published 06-12-2019); implements EN IEC 60942:2018 and IEC 60942:2017; replaces PN-EN 60942:2005
- Classes and designations (clause 1, table 1): LS, LS/M, 1, 1/M, 2; M — pistonphone with a correction for static pressure
- Reference conditions (clause 4): 23 °C; 101,325 kPa; 50 % relative humidity
Overview
WHAT THE STANDARD COVERS. According to clause 1 of IEC 60942:2017 the standard specifies the performance requirements for three classes of sound calibrators: class LS (laboratory standard), class 1 and class 2. Acceptance limits are smallest for class LS and greatest for class 2; class LS calibrators are normally used only in the laboratory, and classes 1 and 2 are considered calibrators for field use. A class 1 calibrator is primarily intended for class 1 sound level meters and a class 2 calibrator for class 2 meters according to IEC 61672-1. The acceptance limits for class LS are based on the use of a laboratory standard microphone (IEC 61094-1), those for classes 1 and 2 on a working standard microphone (IEC 61094-4). The standard does not cover requirements for equivalent free-field or random-incidence sound pressure levels nor other functions of a calibrator (e.g. tonebursts).
ACCORDING TO THE TEXT OF IEC 60942:2017 (publisher's sample in English: table of contents, foreword, introduction, clauses 1–4 and 5.1 up to 5.1.10). The standard was prepared by IEC TC 29 (Electroacoustics); the fourth edition replaces the third of 2003 and is a technical revision. Changes according to the foreword: the class designations LS/C, 1/C and 2/C were deleted; the designations LS/M and 1/M intended for pistonphones were added; a conformance criterion was introduced under which conformance is demonstrated when the measured deviations from design goals do not exceed the acceptance limits and the measurement uncertainty does not exceed the corresponding maximum-permitted uncertainty; the short-term level fluctuation test and some environmental test conditions (to avoid icing) were changed, and an alternative test of immunity to radio-frequency fields in TEM waveguides was added. The introduction names two principal applications of calibrators: determination of the pressure sensitivity of specified models of microphone in specified configurations, and checking or adjusting the overall sensitivity of acoustical measuring devices and systems; the level generated by some calibrators depends on static pressure.
DEFINITIONS AND REFERENCE CONDITIONS (clauses 3 and 4). Sound calibrator — a device that generates a sinusoidal sound pressure of specified level and frequency when coupled to specified models of microphone in specified configurations; pistonphone — a calibrator in which the pressure is generated in a fixed air volume by the motion of one or more pistons. The standard distinguishes the specified sound pressure level (for class LS — for the individual item, for classes 1 and 2 — for all items of the model), the nominal level (rounded to 1 dB, for marking) and the principal level, and correspondingly the specified, nominal (often rounded according to ISO 266) and principal frequency; the reference value of the level is 20 µPa. It also defines replication (re-coupling of the microphone after complete removal), total distortion + noise (in %), the reference plane, the effective load volume of a microphone (mm³) and the acceptance limit. Reference conditions: air temperature 23 °C, static pressure 101,325 kPa, relative humidity 50 %.
GENERAL REQUIREMENTS (5.1). A multi-level and multi-frequency calibrator shall meet the requirements of the same class for all combinations for which the manual declares conformance; adaptors are an integral part of the calibrator. A class LS calibrator has an individual calibration chart (6.2), and for classes 1 and 2 the specified levels and frequencies are given in the manual, each level as an absolute value. Class LS and class 1 pistonphones requiring a correction for the influence of static pressure are designated with the letter “M” (table 1: LS, LS/M, 1, 1/M, 2); the corrections and their uncertainties for a coverage probability of 95 % are given in the manual, and an LS/M or 1/M calibrator is supplied with a barometer or the manual specifies the barometer requirements. Other calibrators shall not require corrections for any environmental condition. The stabilization time of level and frequency from switching on with the microphone coupled is given in the manual; it shall not exceed 30 s, and where it exceeds 10 s the calibrator shall have a stabilization indicator.
FURTHER CONTENT (according to the table of contents, outside the sample): adaptors, sound pressure level and its short-term fluctuation, influence of supply voltage, frequency, influence of environmental conditions, total distortion + noise, microphones and their sensitivity, electromagnetic compatibility, marking, calibration chart and instruction manual (clauses 5–6), pattern evaluation tests (Annex A), periodic tests (Annex B), pattern evaluation report (Annex C) and informative Annexes D and E on the relationship between tolerance interval, acceptance interval and maximum-permitted uncertainty — the sample does not contain their text, so acceptance limits are not given.
STATUS (read 17.11.2026). PKN card: PN-EN IEC 60942:2018-06 — Polish version (published 06-12-2019), title “Elektroakustyka -- Kalibratory akustyczne”, 58 pages, KT 105 for Electroacoustics and Sound and Image Recording, implements EN IEC 60942:2018 and IEC 60942:2017 [IDT], replaces PN-EN 60942:2005; the card gives no withdrawal date. iTeh catalogue: IEC 60942:2017 — published 29-11-2017, stage PPUB (publication issued).
Method principle
A sound calibrator coupled to a microphone of a specified model generates a sinusoidal sound pressure of known level (reference value 20 µPa) and frequency; in a pistonphone the pressure is generated by the motion of a piston in a fixed air volume. Conformity with a class (LS, 1, 2) is demonstrated when the deviations from the design goals are within the acceptance limits and the measurement uncertainty does not exceed the maximum-permitted uncertainty.
Applications
- Sound calibrators of classes LS, 1 and 2 and pistonphones LS/M and 1/M (IEC 60942:2017, clause 1, table 1)
- Determination of the pressure sensitivity of measurement microphones and checking or adjusting the sensitivity of sound level meters and acoustical systems (introduction)
- Class 1 calibrators for class 1 sound level meters, class 2 for class 2 meters according to IEC 61672-1
- Pattern evaluation tests and periodic tests of calibrators (Annexes A and B of the standard)
Key parameters
| Parameter | Value |
|---|---|
| STATUS (read 17.11.2026) | PN-EN IEC 60942:2018-06 (Polish version, published 06-12-2019); implements EN IEC 60942:2018 and IEC 60942:2017; replaces PN-EN 60942:2005 |
| Classes and designations (clause 1, table 1) | LS, LS/M, 1, 1/M, 2; M — pistonphone with a correction for static pressure |
| Reference conditions (clause 4) | 23 °C; 101,325 kPa; 50 % relative humidity |
| Reference value of level (3.3, 3.4) | 20 µPa; nominal level rounded to 1 dB |
| Stabilization time (5.1.10) | ≤ 30 s; above 10 s — stabilization indicator required |
| Reference microphones (clause 1) | class LS — laboratory standard microphone (IEC 61094-1); classes 1 and 2 — working standard microphone (IEC 61094-4) |
| Conformance criterion (foreword) | deviations ≤ acceptance limits and measurement uncertainty ≤ maximum-permitted uncertainty |
| Corrections for LS/M and 1/M (5.1.5, 5.1.8) | for static pressure only, with uncertainty for 95 %; barometer supplied or described in the manual |
Standard
- Standard number
- PN-EN IEC 60942:2018-06 (Polish version; implements EN IEC 60942:2018 and IEC 60942:2017 [IDT])
- Title (PL)
- Elektroakustyka — Kalibratory akustyczne
- Title (EN)
- Electroacoustics — Sound calibrators
Step-by-step procedure
-
Class and documentation
Determination of the class and designation of the calibrator (table 1); for class LS — individual calibration chart, for classes 1 and 2 — specified levels and frequencies from the manual (5.1.4).
-
Coupling to the microphone
Fitting a microphone of the correct model (with adaptor) in the required orientation of the calibrator (5.1.9); a replication means re-coupling after complete removal of the microphone (3.9).
-
Stabilization
Waiting for the stabilization time stated in the manual (not longer than 30 s; above 10 s according to the indicator) before measuring (5.1.10).
-
Measurement and assessment of conformity
Measurement of level, frequency and total distortion + noise with corrections to reference conditions (static pressure only for LS/M and 1/M); assessment against acceptance limits and maximum-permitted uncertainty — periodic tests in Annex B of the standard.
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Sound calibrator (3.1) | Source of sinusoidal sound pressure of specified level and frequency for specified models of microphone; adaptors are part of it (5.1.1) | — |
| Pistonphone (3.2) | Calibrator with a piston (pistons) in a fixed air volume; classes LS/M and 1/M — with a correction for static pressure | — |
| Standard microphone | Laboratory standard according to IEC 61094-1 (class LS) or working standard according to IEC 61094-4 (classes 1 and 2) | — |
| Barometer (5.1.8) | For the corrections of LS/M and 1/M pistonphones — supplied with the calibrator or with specifications given in the manual | — |
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN IEC 60942:2018-06 (Polish version; implements EN IEC 60942:2018 and IEC 60942:2017 [IDT]) — “Electroacoustics — Sound calibrators”.
How does this method work?
A sound calibrator coupled to a microphone of a specified model generates a sinusoidal sound pressure of known level (reference value 20 µPa) and frequency; in a pistonphone the pressure is generated by the motion of a piston in a fixed air volume. Conformity with a class (LS, 1, 2) is demonstrated when the deviations from the design goals are within the acceptance limits and the measurement uncertainty does not exceed the maximum-permitted uncertainty.
What is the measuring range and accuracy?
STATUS (read 17.11.2026): PN-EN IEC 60942:2018-06 (Polish version, published 06-12-2019); implements EN IEC 60942:2018 and IEC 60942:2017; replaces PN-EN 60942:2005; Classes and designations (clause 1, table 1): LS, LS/M, 1, 1/M, 2; M — pistonphone with a correction for static pressure; Reference conditions (clause 4): 23 °C; 101,325 kPa; 50 % relative humidity; Reference value of level (3.3, 3.4): 20 µPa; nominal level rounded to 1 dB.
How long does the test take?
The procedure comprises 4 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Sound calibrator (3.1), Pistonphone (3.2), Standard microphone, Barometer (5.1.8).
Where is this test used?
Sound calibrators of classes LS, 1 and 2 and pistonphones LS/M and 1/M (IEC 60942:2017, clause 1, table 1); Determination of the pressure sensitivity of measurement microphones and checking or adjusting the sensitivity of sound level meters and acoustical systems (introduction); Class 1 calibrators for class 1 sound level meters, class 2 for class 2 meters according to IEC 61672-1; Pattern evaluation tests and periodic tests of calibrators (Annexes A and B of the standard).
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN IEC 60942.