📏 Extensometers — static calibration of extensometer systems (axial and diametral, contacting and non-contacting) with a calibration apparatus: relative gauge length error, resolution, bias error, classes 0,2, 0,5, 1 and 2; PN-EN ISO 9513
In short
Static calibration of extensometer systems used in uniaxial testing: axial and diametral, contacting and non-contacting. The test is performed according to PN-EN ISO 9513:2013-06 (English version; implements EN ISO 9513:2012 and ISO 9513:2012 [IDT]) with corrigendum PN-EN ISO 9513:2013-06/AC:2013-09; Calibration ranges (8.3.3): l_max/l_min ≤ 10 — 1 range; ≤ 100 — 2 ranges; > 100 — 3 ranges; each ≥ 5 increments, increment ≤ 1/3 of the range. The procedure comprises 6 steps; it is used for: Extensometer systems for uniaxial testing — axial and diametral, contacting and non-contacting (ISO 9513:2012, clause 1), Extensometers with fixed and variable gauge length, with automatic gauge length setting or gauge length defined by ridges on the test piece (clause 7), Extensometers for tests at temperatures outside 10 °C to 35 °C — calibration at or near the test temperature (8.1.2).
At a glance
- Standard: PN-EN ISO 9513:2013-06 (English version; implements EN ISO 9513:2012 and ISO 9513:2012 [IDT]) with corrigendum PN-EN ISO 9513:2013-06/AC:2013-09
- Category: Metrology
- Procedure steps: 6
- STATUS (PKN card, 14.11.2026): PN-EN ISO 9513:2013-06 (English version, published 30-06-2013) with corrigendum AC:2013-09; implements EN ISO 9513:2012 and ISO 9513:2012; replaces PN-EN ISO 9513:2005 and Ap1:2005
- Class 0,2 (Table 2): q_Le ±0,2 %; resolution 0,1 % of reading or 0,2 µm; error ±0,2 % or ±0,6 µm (the greater value)
- Class 0,5 (Table 2): q_Le ±0,5 %; resolution 0,25 % or 0,5 µm; error ±0,5 % or ±1,5 µm
Overview
WHAT THE STANDARD COVERS. According to clause 1 of ISO 9513:2012 the standard specifies a method for the static calibration of extensometer systems used in uniaxial testing, including axial and diametral, contacting and non-contacting systems. Definitions (clause 2): extensometer system — equipment measuring displacement or strain on the surface of the test piece; for the purposes of the standard it includes the indicator; some extensometers indicate strain directly (e.g. laser extensometers, image correlation), others indicate the change of gauge length, converted to strain by dividing by the gauge length; gauge length — the part of the test piece over which the extension is measured. Introduction: the standard gives criteria for calibration, the equipment, the check before calibration, the measurement of the gauge length of various types of extensometer, the evaluation of results, uncertainty, calibration intervals and the certificate; dynamic calibration is not covered (work was in progress at the time of publication).
ACCORDING TO THE TEXT OF ISO 9513:2012 — PRINCIPLE, EQUIPMENT, GAUGE LENGTH, PROCEDURE, CLASSIFICATION, UNCERTAINTY, INTERVALS, CERTIFICATE (foreword, introduction, clauses 1–12.2). The standard was prepared by ISO/TC 164 (Mechanical testing of metals), subcommittee SC 1 (Uniaxial testing); the third edition replaced the second (ISO 9513:1999) after a technical revision and incorporated the corrigendum ISO 9513:1999/Cor.1:2000. Symbols (Table 1): Le — nominal gauge length, L′e — measured gauge length, l_max and l_min — limits of the calibration range, lᵢ — displacement indicated by the extensometer, lₜ — displacement from the calibration apparatus, q_Le — relative gauge length error, q_rb — relative bias error, q_b — absolute bias error, r — resolution. Principle (clause 4): comparison of the readings of the extensometer with known changes in length from a calibration apparatus; the user may choose the displacement range(s) (e.g. in strain-controlled low-cycle fatigue — the middle part of the range); the output of the system includes curve fitting by the electronics and the recorder. Equipment (clause 5): calibration apparatus — e.g. a rigid frame with coaxial spindles, with a displacement mechanism and a device for accurately measuring the change in length (interferometer, incremental linear transducer, gauge blocks with a comparator, micrometer); calibrated according to Annex B, meeting Table B.1; the apparatus and auxiliary equipment (micrometers, callipers, projection microscopes) calibrated with traceability to the SI; uncertainty of measurements with auxiliary equipment ≤ 1/3 of the permissible error of the extensometer; thermometer with a resolution of 0,1 °C. Check before calibration (clause 6): mechanical parts (free movement, damage, worn knife edges and gauge length setting pins), cables and connectors; calibration in the "as found" condition, and after an adjustment — a new calibration, with both data sets reported; record of the check; unambiguous identification of the extensometer and interchangeable parts affecting the calibration. Gauge length (clause 7): fixed gauge length — direct measurement (callipers, projection microscope) or indirect (impression of the knife edges on a test piece of soft metal); q_Le = (L′e − Le)/Le × 100; variable gauge length — similarly; in creep and elevated-temperature tensile testing the gauge length is defined by ridges on the test piece; with automatic gauge length setting the largest, the smallest and three intermediate gauge lengths are measured (all of them when there are fewer than five); non-contacting extensometers — according to the manufacturer's instructions; uncertainty of the gauge length measurement three times smaller than the permissible gauge length error. Procedure (clause 8): ambient temperature recorded; recommended constant temperature within ±2 °C, in the range 18 °C to 28 °C; extensometers for tests outside 10 °C to 35 °C — calibration at or near the test temperature where possible; extensometer positioned and fixed as in the test; displacement range set by the user; the number of ranges depends on l_max/l_min: up to 10 — one range of at least 5 increments; over 10 up to 100 — two ranges; over 100 — three ranges (each at least 5 increments); increment between adjacent points ≤ 1/3 of the range; in fatigue — one range of at least 5 increments; thermal expansion and variation of mounting are taken into account when setting the limits. Calibration in the "as found" condition, without cleaning; recommended exercising of the extensometer twice over the full range (where possible to a slightly negative value and back to zero); two measurement series with removal and remounting of the extensometer between them, for extension, compression or both. Resolution: analogue scale — as for analogue indicators, not smaller than 1/5 of a division (1/10 with mark spacing ≥ 2,5 mm); electronic display — observation for 10 s, r = half the difference between the largest and smallest reading (one digit when they are equal). Bias error: q_rb = (lᵢ − lₜ)/lₜ × 100, q_b = lᵢ − lₜ. Classification (clause 9, Table 2; for resolution and bias error the greater of the relative and absolute value applies): class 0,2 — q_Le ±0,2 %, resolution 0,1 % of reading or 0,2 µm, error ±0,2 % or ±0,6 µm; class 0,5 — ±0,5 %, 0,25 % or 0,5 µm, ±0,5 % or ±1,5 µm; class 1 — ±1,0 %, 0,5 % or 1,0 µm, ±1,0 % or ±3,0 µm; class 2 — ±2,0 %, 1,0 % or 2,0 µm, ±2,0 % or ±6,0 µm. Class of the system = the largest of the classes of: gauge length error, resolution at each point, bias error at each point and calibration apparatus; adjustments to meet a class — with the agreement of the laboratory, keeping the preliminary results. Uncertainty (clause 10): the budget includes at least the uncertainty of the calibration apparatus, temperature variations, variation between operators, positioning and gauge length measurement (example — Annex A); note: reducing the permissible error by the uncertainty counts it twice, and class 0,5 does not mean an error including uncertainty below 0,5 %. Intervals (clause 11): recommended about 12 months, not more than 18 months (unless the test lasts longer — then before and after the test); for long-term creep according to ISO 204 — three years; after every repair or adjustment affecting accuracy. Certificate (clause 12): among others reference to ISO 9513, owner, identification of the extensometer (type, gauge length, designation, serial number, mounting position), calibration apparatus, temperature, direction of length changes, date, person and organization, all results (as found and after adjustment), uncertainty, class for each range; table of results with the error at each point.
CORRIGENDUM. PN-EN ISO 9513:2013-06/AC:2013-09 (implements EN ISO 9513:2012/AC:2013 and ISO 9513:2012/Cor 1:2013, free file on the PKN card) changes only the example in Annex C: in the tables of Figures C.1 and C.2 the value "1,30" becomes "1,32" in row "0,33" (column Class 1), in Figure C.2 the headings "Fit / µm" → "v (eff)" and "Fit Error / µm" → "k", the classification in row "0,33" "Class 0,2" → "Class 0,5", and in the key item 6 "fit" → "mean reading". It does not change the content of clauses 1–12.
STATUS (PKN card and iTeh catalogue as of 14.11.2026). Card PN-EN ISO 9513:2013-06 (English version): title "Metale -- Wzorcowanie ekstensometrów stosowanych w próbie jednoosiowego rozciągania", published 30-06-2013, 58 pages, KT 123 for Testing the Properties of Metals, "Wprowadza: EN ISO 9513:2012 [IDT], ISO 9513:2012 [IDT]", "Zastępuje: PN-EN ISO 9513:2005 - wersja polska, PN-EN ISO 9513:2005/Ap1:2005 - wersja polska", corrigendum PN-EN ISO 9513:2013-06/AC:2013-09E, ICS 77.040.10. ISO 9513:2012 in the iTeh catalogue — published, stage 90.93 (standard confirmed).
HOW MANY LABORATORIES AND IN WHAT FORM (accreditation scope database of the service, read 14.11.2026). ISO 9513 appears in 12 records at 8 Polish calibration laboratories: "PN-EN ISO 9513:2013-06" — AP 045, AP 072, AP 133, AP 181, AP 194, AP 207, AP 225; "PN-EN 9513:2013-06" (without "ISO") — AP 130; "PN-EN ISO 9513:2013-06 zał. H" — AP 194, AP 207, AP 225 (crosshead displacement). Next to the standard are ASTM E83 (AP 133, AP 181, AP 194, AP 207, AP 225) and ASTM E2309/E2309M-20 (AP 133, AP 181, AP 194, AP 207, AP 225). Outside Poland — 2 Czech laboratories (3 records).
WHAT THE LABORATORIES CALIBRATE (scope database, complete items). Extensometers, also mounted in static testing machines (AP 045, AP 130); crosshead displacement (travel) of testing machines and process equipment (AP 072, AP 194, AP 207, AP 225). Ranges from 0 µm to 2500 mm (AP 207, AP 225) or 5000 mm for the crosshead; CMC from 0,08 µm (AP 207, 0 mm to 10 mm) to 0,8 % (AP 072, 0,4 mm to 25 mm).
WHERE A SEEMINGLY CORRECT RESULT IS EASY TO GET. Calibration after adjustment without "as found" results (6.1, 9.4.2). Extensometer positioned differently than in the test (8.2) or without temperature stabilization (8.1.3). One range with l_max/l_min over 10 — requirement 8.3.3 not met. Resolution from a display without the 10-second observation of fluctuations (8.5.1.3). Class assigned from the bias error alone, without the gauge length error and the class of the calibration apparatus (9.4.1). Error reduced by the uncertainty — double counting of uncertainty (note to 10.2).
WHAT WE DO NOT GIVE. Parts read: iTeh sample of ISO 9513:2012 in English (foreword, introduction, clauses 1–12.2) and the corrigendum ISO 9513:2012/Cor 1:2013 (PKN file). We do not give the content of Annexes A–H (uncertainty budget, calibration of the calibration apparatus with Table B.1, example certificate, examples of configurations, laser, video and full-field extensometry, calibration of crosshead displacement measurement — Annex H) — they are outside the sample. PKN has not issued a Polish translation of PN-EN ISO 9513:2013-06 (English version).
Method principle
The extensometer is mounted in the calibration apparatus as in the test, after temperature stabilization, and its readings lᵢ are compared with known displacements lₜ in two series (with removal and remounting of the extensometer between them). The number of ranges depends on the ratio l_max/l_min (1, 2 or 3 ranges, each with at least 5 increments). For each point the bias error q_rb = (lᵢ − lₜ)/lₜ × 100 or q_b = lᵢ − lₜ is calculated, from the resolution and the gauge length measurement — q_Le; the class of the system (0,2, 0,5, 1 or 2 according to Table 2) is the worst of the classes of these results and of the calibration apparatus.
Applications
- Extensometer systems for uniaxial testing — axial and diametral, contacting and non-contacting (ISO 9513:2012, clause 1)
- Extensometers with fixed and variable gauge length, with automatic gauge length setting or gauge length defined by ridges on the test piece (clause 7)
- Extensometers for tests at temperatures outside 10 °C to 35 °C — calibration at or near the test temperature (8.1.2)
- In accreditation scopes: 8 Polish calibration laboratories — extensometers and crosshead displacement of testing machines; 2 Czech
Key parameters
| Parameter | Value |
|---|---|
| STATUS (PKN card, 14.11.2026) | PN-EN ISO 9513:2013-06 (English version, published 30-06-2013) with corrigendum AC:2013-09; implements EN ISO 9513:2012 and ISO 9513:2012; replaces PN-EN ISO 9513:2005 and Ap1:2005 |
| Class 0,2 (Table 2) | q_Le ±0,2 %; resolution 0,1 % of reading or 0,2 µm; error ±0,2 % or ±0,6 µm (the greater value) |
| Class 0,5 (Table 2) | q_Le ±0,5 %; resolution 0,25 % or 0,5 µm; error ±0,5 % or ±1,5 µm |
| Classes 1 and 2 (Table 2) | class 1: ±1,0 %; 0,5 % or 1,0 µm; ±1,0 % or ±3,0 µm. Class 2: ±2,0 %; 1,0 % or 2,0 µm; ±2,0 % or ±6,0 µm |
| Calibration ranges (8.3.3) | l_max/l_min ≤ 10 — 1 range; ≤ 100 — 2 ranges; > 100 — 3 ranges; each ≥ 5 increments, increment ≤ 1/3 of the range |
| Temperature (8.1) | recorded; recommended constant within ±2 °C, in the range 18 °C to 28 °C; thermometer with 0,1 °C resolution (5.2) |
| Resolution of a display (8.5.1.3) | observation for 10 s; r = ½ (max − min); without fluctuation — 1 digit |
| Calibration intervals (clause 11) | recommended about 12 months, at most 18; long-term creep according to ISO 204 — 3 years; after repair or adjustment |
| Coverage in accreditation scopes (read 14.11.2026) | 8 Polish calibration laboratories, 12 records (3 with "zał. H"); 2 Czech |
Standard
- Standard number
- PN-EN ISO 9513:2013-06 (English version; implements EN ISO 9513:2012 and ISO 9513:2012 [IDT]) with corrigendum PN-EN ISO 9513:2013-06/AC:2013-09
- Title (PL)
- Metale — Wzorcowanie ekstensometrów stosowanych w próbie jednoosiowego rozciągania
- Title (EN)
- Metallic materials — Calibration of extensometer systems used in uniaxial testing
Step-by-step procedure
-
Check before calibration
Mechanical parts, knife edges, gauge length setting pins, cables and connectors; record of the as-found condition; identification of the extensometer and interchangeable parts (clause 6).
-
Gauge length measurement
Directly or indirectly (impression on soft metal); q_Le = (L′e − Le)/Le × 100; with automatic setting — 5 gauge lengths (clause 7).
-
Conditions and positioning
Temperature recorded, stabilization of the extensometer and apparatus; position and fixing as in the test (8.1–8.2).
-
Ranges and increments
Setting l_min and l_max, the number of ranges (1–3) and increments (≥ 5, each ≤ 1/3 of the range) (8.3).
-
Two measurement series
Calibration without cleaning; recommended exercising twice; series, removal and remounting of the extensometer, second series — for extension, compression or both (8.4).
-
Results and class
Resolution, q_rb or q_b at each point, comparison with Table 2; class = the worst of four assessments; uncertainty and certificate (clauses 9–12).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Calibration apparatus (5.1) | Rigid frame with coaxial spindles and measurement of the change in length: interferometer, incremental linear transducer, gauge blocks with a comparator or micrometer; calibrated according to Annex B | — |
| Auxiliary equipment (5.2, 7.1) | Micrometers, callipers, projection microscopes — uncertainty ≤ 1/3 of the permissible error of the extensometer | — |
| Thermometer (5.2) | Resolution 0,1 °C | — |
| Soft metal test piece (7.1.1.2) | For indirect gauge length measurement — impression of the knife edges or tips of the extensometer | — |
Health and safety (OHS)
- The standard contains no safety warning in the text read
- TECHNICAL NOTE: a class assigned after calibration does not guarantee that the error including uncertainty is smaller than the class value (note to 10.2)
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN ISO 9513:2013-06 (English version; implements EN ISO 9513:2012 and ISO 9513:2012 [IDT]) with corrigendum PN-EN ISO 9513:2013-06/AC:2013-09 — “Metallic materials — Calibration of extensometer systems used in uniaxial testing”.
How does this method work?
The extensometer is mounted in the calibration apparatus as in the test, after temperature stabilization, and its readings lᵢ are compared with known displacements lₜ in two series (with removal and remounting of the extensometer between them). The number of ranges depends on the ratio l_max/l_min (1, 2 or 3 ranges, each with at least 5 increments).
What is the measuring range and accuracy?
STATUS (PKN card, 14.11.2026): PN-EN ISO 9513:2013-06 (English version, published 30-06-2013) with corrigendum AC:2013-09; implements EN ISO 9513:2012 and ISO 9513:2012; replaces PN-EN ISO 9513:2005 and Ap1:2005; Class 0,2 (Table 2): q_Le ±0,2 %; resolution 0,1 % of reading or 0,2 µm; error ±0,2 % or ±0,6 µm (the greater value); Class 0,5 (Table 2): q_Le ±0,5 %; resolution 0,25 % or 0,5 µm; error ±0,5 % or ±1,5 µm; Classes 1 and 2 (Table 2): class 1: ±1,0 %; 0,5 % or 1,0 µm; ±1,0 % or ±3,0 µm. Class 2: ±2,0 %; 1,0 % or 2,0 µm; ±2,0 % or ±6,0 µm.
How long does the test take?
The procedure comprises 6 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Calibration apparatus (5.1), Auxiliary equipment (5.2, 7.1), Thermometer (5.2), Soft metal test piece (7.1.1.2).
Where is this test used?
Extensometer systems for uniaxial testing — axial and diametral, contacting and non-contacting (ISO 9513:2012, clause 1); Extensometers with fixed and variable gauge length, with automatic gauge length setting or gauge length defined by ridges on the test piece (clause 7); Extensometers for tests at temperatures outside 10 °C to 35 °C — calibration at or near the test temperature (8.1.2); In accreditation scopes: 8 Polish calibration laboratories — extensometers and crosshead displacement of testing machines; 2 Czech.
What safety precautions apply?
The standard contains no safety warning in the text read; TECHNICAL NOTE: a class assigned after calibration does not guarantee that the error including uncertainty is smaller than the class value (note to 10.2).
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 9513.