📏 Coordinate measuring machines with contacting probing systems — acceptance and reverification tests on a test sphere: form and size error with a single stylus (25 points), with multiple styli (5 × 25), location error, scanning mode; PN-EN ISO 10360-5

Metrology PN-EN ISO 10360-5

In short

Acceptance and reverification tests of coordinate measuring machines with any contacting probing system and spherical or hemispherical stylus tips. The test is performed according to PN-EN ISO 10360-5:2020-10 (Polish version; implements EN ISO 10360-5:2020 and ISO 10360-5:2020 [IDT]) with amendment PN-EN ISO 10360-5:2020-10/Ap1:2025-03; STATUS (PKN card, 14.11.2026): PN-EN ISO 10360-5:2020-10 (Polish version, published 11-10-2023) with amendment Ap1:2025-03; implements EN ISO 10360-5:2020 and ISO 10360-5:2020; replaces among others PN-EN ISO 10360-4:2002 and PN-EN ISO 10360-5:2010. The procedure comprises 5 steps; it is used for: CMMs with any contacting probing system and spherical or hemispherical tips — acceptance and reverification tests (ISO 10360-5:2020, clause 1), Single-stylus, multi-stylus (star), multiple probing systems, articulating probing systems, stylus and probe changing (clause 1), Manual and automatic machines; scanning probing systems in scanning mode (clause 1).

At a glance

  • Standard: PN-EN ISO 10360-5:2020-10 (Polish version; implements EN ISO 10360-5:2020 and ISO 10360-5:2020 [IDT]) with amendment PN-EN ISO 10360-5:2020-10/Ap1:2025-03
  • Category: Metrology
  • Procedure steps: 5
  • STATUS (PKN card, 14.11.2026): PN-EN ISO 10360-5:2020-10 (Polish version, published 11-10-2023) with amendment Ap1:2025-03; implements EN ISO 10360-5:2020 and ISO 10360-5:2020; replaces among others PN-EN ISO 10360-4:2002 and PN-EN ISO 10360-5:2010
  • Single stylus (3.11–3.12): P_Form.Sph.1×25:SS:Tact and P_Size.Sph.1×25:SS:Tact — 25 points on the test sphere
  • Multiple styli (3.7–3.9): P_Form.Sph.5×25:j:Tact, P_Size.Sph.5×25:j:Tact, L_Dia.5×25:j:Tact — 5 styli of 25 points; j = MS, MP, Emp, Inf

Overview

WHAT THE STANDARD COVERS. According to clause 1 of ISO 10360-5:2020 the document specifies the acceptance and reverification tests of the performance of a coordinate measuring machine (CMM) with contacting probing systems and applies only to CMMs using any type of contacting probing system and spherical or hemispherical stylus tips. Note: the characteristics are specified by maximum permissible errors (MPE) because in practice it is not possible to separate the performance of the probing system from that of the machine, even with a small test sphere. The document applies to machines with: a single-stylus probing system, a multi-stylus probing system (e.g. a star), multiple probing systems (a stylus per probing system), an articulating probing system, stylus and probe changing systems, manual and automatic machines and scanning probing systems; it does not apply to non-contacting probing systems. Where it is necessary to separate the performance of the probing system as far as possible, the influence of the machine can be reduced but not eliminated (Annex C).

ACCORDING TO THE TEXT OF ISO 10360-5:2020 — INTRODUCTION, SCOPE, DEFINITIONS (foreword, introduction, clauses 1–3.23). The document was prepared by ISO/TC 213 in collaboration with CEN/TC 290 (Vienna Agreement); the third edition replaced the second (ISO 10360-5:2010) after a technical revision and incorporated with a technical revision the tests of ISO 10360-4:2000, which it also replaced. Main changes: new symbols, an additional (optional) ring gauge test, changes to the permissible test parameters (e.g. the test sphere diameter), changes to the evaluation of location results, including the evaluation of "opposing styli". Introduction: a general GPS standard (ISO 14638), link F of the chains of size, distance, form, orientation, location and run-out; the principles of ISO 8015 and the default decision rules of ISO 14253-1 apply; the tests apply to CMMs with contacting probes — with multiple styli or articulating probe positions — in discrete-point and/or scanning mode; multi-stylus errors may be significant and sometimes dominate the machine error; the tests cover a practical subset of configurations; with them the user can reduce errors (e.g. by removing long extensions and styli and testing the new configuration); the tests supplement the length tests of ISO 10360-2 — with multiple styli the machine travel can be quite different from the measured length (Annex C); they also supplement ISO 10360-7, -8 and -9. References: ISO 10360-1, ISO 10360-2, ISO 14253-1, ISO/IEC Guide 99:2007. Definitions (clause 3; they replace 18 definitions of ISO 10360-1:2000, clause 9; zero form and location are assumed for a single sphere — limitations on the form of the sphere in 6.2.1): rated operating conditions; empirical qualification (measurement of the reference sphere at each position of the articulating probe) and inferred qualification (parameters for other positions by interpolation, extrapolation or a model); effective stylus diameter; multi-stylus probing system and multiple probing system. Characteristics: multi-stylus form error P_Form.Sph.5×25:j:Tact — form of the test sphere measured with five styli of 25 points each (5 × 25) in discrete-point mode, where j = MS (multi-stylus), MP (multiple probes), Emp (articulating, empirical qualification), Inf (articulating, inferred qualification); size error P_Size.Sph.5×25:j:Tact — error of indication of the sphere diameter; location error L_Dia.5×25:j:Tact — error of indication of the location of the sphere measured from five directions; opposing-styli projected location error L_Dia.Proj.Sph.2×25:j:Tact (e.g. for co-axiality of crank shaft journals measured from opposite sides); single-stylus form and size error P_Form.Sph.1×25:SS:Tact and P_Size.Sph.1×25:SS:Tact (25 points); in scanning mode — form error P_Form.Sph.Scan:k:Tact and size error P_Size.Sph.Scan:k:Tact on the sphere and scanning time τ_Sph.Scan:k:Tact (k = PP — pre-defined path, NPP — not pre-defined path); on a ring gauge (optional test, Annex A) — P_Form.Cir.Scan:k.lo:Tact and P_Size.Cir.Scan:k.lo:Tact (lo = 0 mm — stylus aligned to the ram axis, by default 150 mm — orthogonal) and L_Dia.Proj.Cir.Scan:j:Tact; each has a corresponding MPE (how to express it — Annex D).

AMENDMENT. PN-EN ISO 10360-5:2020-10/Ap1:2025-03 (national amendment of PKN/KT 48, approved on 5 March 2025, free file on the PKN card) changes two wordings of the Polish version: in 6.1, NOTE 1, "błąd próbkowania" (probing error) becomes "błąd głowicy pomiarowej" (probing system error), and in 6.3.4, second paragraph, "błąd wymiaru próbkowania" becomes "błąd wymiaru głowicy pomiarowej".

STATUS (PKN card and iTeh catalogue as of 14.11.2026). Card PN-EN ISO 10360-5:2020-10 (Polish version): title "Specyfikacje geometrii wyrobów (GPS) -- Badania odbiorcze i okresowe współrzędnościowych systemów pomiarowych (CMS) -- Część 5: Współrzędnościowe maszyny pomiarowe (CMM) z zespołami głowic pomiarowych stykowych z pojedynczym trzpieniem lub wielotrzpieniowych działające w trybie próbkowania punktowego i/lub skanowania", published 11-10-2023, 56 pages, KT 48 for Fundamentals of Machine Design, "Wprowadza: EN ISO 10360-5:2020 [IDT], ISO 10360-5:2020 [IDT]", "Zastępuje: PN-EN ISO 10360-4:2002 - wersja polska, PN-EN ISO 10360-4:2002/AC:2005 - wersja polska, PN-EN ISO 10360-5:2010/Ap1:2013-11 - wersja angielska, PN-EN ISO 10360-5:2010 - wersja polska, PN-EN ISO 10360-5:2010 - wersja angielska, PN-EN ISO 10360-4:2002/Ap1:2017-01 - wersja polska", amendment PN-EN ISO 10360-5:2020-10/Ap1:2025-03P, ICS 17.040.30. ISO 10360-5:2020 in the iTeh catalogue — published, stage 90.93 (standard confirmed).

HOW MANY LABORATORIES AND IN WHAT FORM (accreditation scope database of the service, read 14.11.2026). ISO 10360-5 appears in 14 records at 10 Polish calibration laboratories: "PN-EN ISO 10360-5:2020" — AP 099, AP 132, AP 141, AP 186, AP 206, AP 214; "PN-EN ISO 10360-5:2020-10" — AP 197, AP 220, AP 236, AP 239. In AP 220 it is accompanied by the withdrawn PN-EN ISO 10360-4:2002 (replaced by PN-EN ISO 10360-5:2020-10 — PKN card). Outside Poland — 2 Czech laboratories (3 records).

WHAT THE LABORATORIES CALIBRATE (scope database, complete items). Contacting probing systems — touch-trigger and scanning, on CNC and manual machines (AP 099). Quantity 25 mm (AP 206 — 25 mm to 30 mm; AP 220, AP 236, AP 239 — 0 mm to 25 mm); CMC from 0,1 μm (AP 186, AP 220) to 0,7 μm (AP 099 — manual machines).

WHERE A SEEMINGLY CORRECT RESULT IS EASY TO GET. Treating the result as the error of the probing system alone — the result contains the influence of the machine, which cannot be eliminated (note to clause 1, Annex C). Testing only with a single stylus when the machine works with a stylus star or an articulating probe — multi-stylus errors may dominate (introduction). Confusing empirical and inferred qualification (j = Emp and Inf are different characteristics, 3.7). Test sphere with form outside the limitations of 6.2.1. Stating the characteristics with old symbols of ISO 10360-4 or ISO 10360-5:2010 instead of the symbols of clause 3.

WHAT WE DO NOT GIVE. Parts read: iTeh sample of ISO 10360-5:2020 in English (foreword, introduction, clauses 1–3.23) and the amendment PN-EN ISO 10360-5:2020-10/Ap1:2025-03 (PKN file). We do not give clauses 4–8 (symbols, rated operating conditions, test procedure: test sphere, styli, distribution of points, data analysis, conformance with specification, interim checks) or Annexes A–G — they are outside the sample; the numbers of points 25 and 5 × 25 come from the definitions in clause 3. We have not read the Polish text of PN-EN ISO 10360-5:2020-10.

Method principle

On a test sphere (and optionally a ring gauge) the machine with the tested probing configuration measures in discrete-point mode 25 points with one stylus, 5 × 25 points with five styli, articulating probe positions or probes, and with scanning probes — also in scanning mode. From these measurements the form error (P_Form), the diameter size error (P_Size) and the location error (L_Dia, also for opposing styli) are determined and compared with the maximum permissible errors MPE; the result covers the combined performance of the machine and the probing system.

Applications

Key parameters

ParameterValue
STATUS (PKN card, 14.11.2026)PN-EN ISO 10360-5:2020-10 (Polish version, published 11-10-2023) with amendment Ap1:2025-03; implements EN ISO 10360-5:2020 and ISO 10360-5:2020; replaces among others PN-EN ISO 10360-4:2002 and PN-EN ISO 10360-5:2010
Single stylus (3.11–3.12)P_Form.Sph.1×25:SS:Tact and P_Size.Sph.1×25:SS:Tact — 25 points on the test sphere
Multiple styli (3.7–3.9)P_Form.Sph.5×25:j:Tact, P_Size.Sph.5×25:j:Tact, L_Dia.5×25:j:Tact — 5 styli of 25 points; j = MS, MP, Emp, Inf
Opposing styli (3.10, 3.18)L_Dia.Proj.Sph.2×25:j:Tact on the sphere; L_Dia.Proj.Cir.Scan:j:Tact on the ring
Scanning (3.13–3.17)P_Form.Sph.Scan:k:Tact, P_Size.Sph.Scan:k:Tact, time τ_Sph.Scan:k:Tact; k = PP or NPP; ring gauge (optional) — lo = 0 mm or 150 mm
Assessment (note to clause 1)maximum permissible errors MPE — probing system and machine combined
National amendment Ap1:2025-036.1 note 1: "błąd próbkowania" → "błąd głowicy pomiarowej"; 6.3.4: "błąd wymiaru próbkowania" → "błąd wymiaru głowicy pomiarowej"
Coverage in accreditation scopes (read 14.11.2026)10 Polish calibration laboratories, 14 records; quantity 25 mm; 2 Czech

Standard

Standard number
PN-EN ISO 10360-5:2020-10 (Polish version; implements EN ISO 10360-5:2020 and ISO 10360-5:2020 [IDT]) with amendment PN-EN ISO 10360-5:2020-10/Ap1:2025-03
Title (PL)
Specyfikacje geometrii wyrobów (GPS) — Badania odbiorcze i okresowe współrzędnościowych systemów pomiarowych (CMS) — Część 5: Współrzędnościowe maszyny pomiarowe (CMM) z zespołami głowic pomiarowych stykowych z pojedynczym trzpieniem lub wielotrzpieniowych działające w trybie próbkowania punktowego i/lub skanowania
Title (EN)
Geometrical product specifications (GPS) — Acceptance and reverification tests for coordinate measuring systems (CMS) — Part 5: Coordinate measuring machines (CMMs) using single and multiple stylus contacting probing systems using discrete point and/or scanning measuring mode

Step-by-step procedure

  1. Specification

    Establishing the characteristics tested and their MPE for the machine configuration (single stylus, MS, MP, Emp, Inf, scanning) (clause 3).

  2. Single stylus

    25 points on the test sphere in discrete-point mode — form and size error (3.11–3.12).

  3. Multiple styli

    5 styli (or probe positions, or probes) of 25 points each on one sphere — form, size and location error; evaluation of opposing styli (3.7–3.10).

  4. Scanning

    With a scanning probe — form and size of the sphere in scanning mode and scanning time; optionally a ring gauge (3.13–3.18, Annex A).

  5. Assessment

    Comparison with MPE; procedure, distribution of points and data analysis — clauses 6–7 (outside the sample read).

Required equipment and apparatus

EquipmentExampleIndicative price
Test sphere (3.7, 6.2.1)Limitations on the form of the sphere — 6.2.1 (outside the sample read)—
Ring gauge (3.16–3.18)For the optional test in scanning mode (Annex A)—
Stylus and probe configuration (clause 1, 3.5–3.6)Single stylus, star, multiple probes, articulating probe with empirical or inferred qualification—

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN ISO 10360-5:2020-10 (Polish version; implements EN ISO 10360-5:2020 and ISO 10360-5:2020 [IDT]) with amendment PN-EN ISO 10360-5:2020-10/Ap1:2025-03 — “Geometrical product specifications (GPS) — Acceptance and reverification tests for coordinate measuring systems (CMS) — Part 5: Coordinate measuring machines (CMMs) using single and multiple stylus contacting probing systems using discrete point and/or scanning measuring mode”.

How does this method work?

On a test sphere (and optionally a ring gauge) the machine with the tested probing configuration measures in discrete-point mode 25 points with one stylus, 5 × 25 points with five styli, articulating probe positions or probes, and with scanning probes — also in scanning mode. From these measurements the form error (P_Form), the diameter size error (P_Size) and the location error (L_Dia, also for opposing styli) are determined and compared with the maximum permissible errors MPE; the result covers the combined performance of the machine and the probing system.

What is the measuring range and accuracy?

STATUS (PKN card, 14.11.2026): PN-EN ISO 10360-5:2020-10 (Polish version, published 11-10-2023) with amendment Ap1:2025-03; implements EN ISO 10360-5:2020 and ISO 10360-5:2020; replaces among others PN-EN ISO 10360-4:2002 and PN-EN ISO 10360-5:2010; Single stylus (3.11–3.12): P_Form.Sph.1×25:SS:Tact and P_Size.Sph.1×25:SS:Tact — 25 points on the test sphere; Multiple styli (3.7–3.9): P_Form.Sph.5×25:j:Tact, P_Size.Sph.5×25:j:Tact, L_Dia.5×25:j:Tact — 5 styli of 25 points; j = MS, MP, Emp, Inf; Opposing styli (3.10, 3.18): L_Dia.Proj.Sph.2×25:j:Tact on the sphere; L_Dia.Proj.Cir.Scan:j:Tact on the ring.

How long does the test take?

The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Test sphere (3.7, 6.2.1), Ring gauge (3.16–3.18), Stylus and probe configuration (clause 1, 3.5–3.6).

Where is this test used?

CMMs with any contacting probing system and spherical or hemispherical tips — acceptance and reverification tests (ISO 10360-5:2020, clause 1); Single-stylus, multi-stylus (star), multiple probing systems, articulating probing systems, stylus and probe changing (clause 1); Manual and automatic machines; scanning probing systems in scanning mode (clause 1); In accreditation scopes: 10 Polish calibration laboratories — touch-trigger and scanning contacting probes; 2 Czech.

What safety precautions apply?

The standard contains no safety warning in the text read; TECHNICAL NOTE: the test result covers the performance of the machine and the probing system together — the influence of the machine cannot be completely eliminated (note to clause 1).

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 10360-5.

🔍 Find a laboratory performing this test