🔥 Aviation and motor gasolines (finished) — oxidation stability by the induction period method: 50 ml of sample in a pressure bomb with oxygen at 690–705 kPa, boiling water bath 98–102 °C, time to the breakpoint of the pressure curve (drop of exactly 14 kPa within 15 min), conversion to 100 °C, result in minutes; PN-EN ISO 7536
In short
Determination of the stability of aviation and motor gasolines (finished products only) under accelerated oxidation conditions. The test is performed according to PN-EN ISO 7536:2011 (Polish version; implements EN ISO 7536:1996 and ISO 7536:1994 [IDT]); STATUS (PKN card as of 11.11.2026): PN-EN ISO 7536:2011 current (Polish version, published 23-04-2013); replaced PN-ISO 7536:1997; ISO 7536:1994 confirmed (iTeh, stage 9093). The procedure comprises 5 steps; it is used for: Aviation and motor gasolines — finished products only (ISO 7536:1994, 1.1), Indication of the tendency of gasoline to form gum in storage — with the reservation that the correlation varies (1.3), Not for individual gasoline components, particularly with a high percentage of low-boiling unsaturated compounds — risk of explosion (1.2).
At a glance
- Standard: PN-EN ISO 7536:2011 (Polish version; implements EN ISO 7536:1996 and ISO 7536:1994 [IDT])
- Category: Petrochemistry
- Procedure steps: 5
- STATUS (PKN card as of 11.11.2026): PN-EN ISO 7536:2011 current (Polish version, published 23-04-2013); replaced PN-ISO 7536:1997; ISO 7536:1994 confirmed (iTeh, stage 9093)
- Sample — condition (7.1): 50 ml ± 1 ml; bomb and gasoline at 15–25 °C
- Oxygen — condition (4.4, 7.1): purity ≥ 99 %, extra-dry; filling to 690–705 kPa, flushing with release ≤ 350 kPa/min
Overview
WHAT THE STANDARD COVERS. According to clause 1 of ISO 7536:1994 the standard specifies a method for determining the stability of aviation and motor gasolines in their finished form only, under accelerated oxidation conditions, by measuring the induction period to breakpoint in a pressure bomb apparatus (1.1). The method is not intended for determining the stability of gasoline components individually, particularly those with a high percentage of low-boiling unsaturated compounds, as they may cause explosive conditions within the apparatus; the bomb assembly therefore includes a safety burst-disc (1.2). The induction period may be used as an indication of the tendency of gasoline to form gum in storage, but this correlation may vary markedly under different storage conditions and with different gasolines (1.3). The "Scope" field of the PKN card for PN-EN ISO 7536:2011 quotes the same text in Polish ("… by measuring the induction period from the start of the test to the breakpoint, carried out in an apparatus equipped with a pressure bomb. The method is intended solely for testing finished products …").
ACCORDING TO THE TEXT OF ISO 7536:1994 — DEFINITIONS, REAGENTS, APPARATUS, PROCEDURE, CALCULATION, PRECISION (foreword, clauses 1–11). The standard was prepared by ISO/TC 28 (Petroleum products and lubricants); first edition 1994-08-01; Annexes A and B are normative. Definitions (clause 2): breakpoint — the point in the pressure–time curve preceded by a pressure drop of exactly 14 kPa within 15 min and followed by a drop of not less than 14 kPa in 15 min; induction period — the time elapsed between placing the bomb in the bath and the breakpoint at 100 °C. Principle (clause 3): the sample is oxidized in a pressure bomb initially filled at 15–25 °C with oxygen at 690 kPa and heated at 98–102 °C; the pressure is read at stated intervals or recorded continuously until the breakpoint; from the period at the test temperature the induction period at 100 °C is calculated. Reagents (clause 4): toluene and acetone of at least 99 % purity; gum solvent — a mixture of equal volumes of toluene and acetone; commercially available extra-dry oxygen of not less than 99 % purity; a detergent cleaning solution that cleans containers as well as 6 h immersion in fresh chromic acid cleaning solution (the type of detergent and conditions are established by the laboratory). Apparatus (clause 5): pressure bomb of corrosion-resistant steel with inside dimensions according to Figure 1 and a highly polished interior, constructed to withstand a working pressure of 1 240 kPa at 100 °C, with an ultimate strength at least equal to that of a bomb of 18 % chromium and 8 % nickel alloy steel (material according to Annex A); closure tight when filled with oxygen at 690 kPa at 15–25 °C and plunged into a bath at 100 °C; gasket — passes the test if the bomb without gasoline, with oxygen at 690 kPa, in a bath at 100 °C ± 1.0 °C does not lose more than 14 kPa from the maximum in 24 h; glass sample container with cover (Figure 2); bomb stem with filler rod and a circular plate 89 mm in diameter closing the bath; stainless-steel burst-disc assembly rupturing at a pressure greater than 1 530 kPa ± 10 %; needle valve; indicating or recording pressure gauge reading to at least 1 380 kPa, with intervals of division of 35 kPa or less and an accuracy of 1 % of the total scale interval or better — any half of the scale interval between 690 kPa and 1 380 kPa shall be at least 25 mm along the arc; total volume of tubing, connections and stem with filler rod not exceeding 30 ml; water bath of at least 18 l for one bomb plus 8 l for each additional bomb, water depth not less than 290 mm, top of the bomb lid at least 50 mm below the surface, boiling vigorously, with a condenser; thermometer with a range of 95–103 °C (Annex B); spade-ended forceps of corrosion-resistant steel. Preparation (clause 6): the container is washed with gum solvent, rinsed with water, immersed in hot detergent solution, then handled only with forceps, washed with tap and distilled water and dried at 100–150 °C for at least 1 h; the bomb, stem, rod and valve are cleaned with gum solvent; the bomb and lines must be dry. Procedure (clause 7): the bomb and the gasoline are brought to 15–25 °C; 50 ml ± 1 ml of sample is added to the container in the bomb, the container covered, the bomb closed and oxygen introduced to 690–705 kPa; the gas is released slowly (not faster than 350 kPa per min) to flush out the air and the bomb refilled to 690–705 kPa; an initial rapid drop (generally not greater than 40 kPa) results from dissolution of oxygen in the sample; if the pressure drop does not exceed 7.0 kPa in 10 min, absence of leaks is assumed. The bomb is placed in the vigorously boiling bath and the time of immersion recorded; bath temperature 98–102 °C, observed to 0.1 °C, the average being the test temperature; pressure recorded continuously or read at intervals of 15 min or less. If a leak develops during the initial 30 min (a steady drop considerably in excess of 14 kPa in 15 min), the test is discarded. The test continues to the breakpoint; the number of minutes from immersion to the breakpoint is the observed induction period at the test temperature. Where the atmospheric pressure is consistently below normal, a higher-boiling liquid such as ethylene glycol may be added to the water to keep about 100 °C (Note 4). After the test the bomb is cooled in cool tap water and the pressure released slowly (not faster than 350 kPa/min). Calculation (clause 8): induction period at 100 °C, in minutes — when the test temperature is above 100 °C: IP100 = (IPt)[1 + 0.101 (Ta − 100)]; when below: IP100 = (IPt)/[1 + 0.101 (100 − Tb)], where IPt — observed induction period at the test temperature (min), Ta and Tb — test temperature above and below 100 °C. Result (clause 9): IP100 to the nearest whole minute. Precision (clause 10): repeatability — the difference between successive results of the same operator differs from the mean by more than 5 % only in one case in twenty; reproducibility — by more than 10 % only in one case in twenty; precision was determined using equipment without a burst-disc assembly (Note 5 — the effect of the assembly is likely to be minimal). Test report (clause 11): identification of the product, reference to the standard, result, deviations, date.
STATUS (PKN cards as of 11.11.2026). The card for PN-EN ISO 7536:2011 (Polish version) has no withdrawal label: published 23-04-2013, 16 pages, KT 222 for Petroleum Products and Similar Products of Biological and Synthetic Origin, "Implements: EN ISO 7536:1996 [IDT], ISO 7536:1994 [IDT]", "Replaces: PN-ISO 7536:1997 - Polish version", ICS 75.160.20. The PKN search (term "PN-EN ISO 7536", 11.11.2026) returns the cards of PN-EN ISO 7536:2011 in the Polish and English versions and PN-ISO 7536:1997 (withdrawn 05-08-2011, replaced by PN-EN ISO 7536:2011). iTeh catalogue (read 11.11.2026): ISO 7536:1994 — stage "9093 – International Standard confirmed".
HOW MANY LABORATORIES AND IN WHAT FORM (accreditation scope database of the service, read 11.11.2026). The notation ISO 7536 appears in 5 records at 5 Polish laboratories: AB 079, AB 297 and AB 484 ("PN-EN ISO 7536:2011"), AB 009 and AB 387 ("PN-EN ISO 7536"); all with the technique "Induction period method". Outside Poland — 1 record of the Czech laboratory 1152.1 (procedure SOP 114 citing ČSN EN ISO 7536; E-85). We have not compared the current PCA scope documents of these laboratories with the database.
WHAT THE LABORATORIES TEST (scope database, complete items — 5 Polish). Unleaded gasoline (AB 079, AB 484 — also "other petroleum products: fractions and/or petroleum products with a boiling range"), unleaded gasoline, aviation gasoline and unleaded gasoline with bioethanol (AB 297), unleaded gasoline and liquid biohydrocarbons for spark-ignition engines (AB 009), liquid fuels (AB 387) — oxidation resistance / oxidation stability. The standard excludes individual gasoline components (1.2) — the text of the standard does not directly cover the petroleum fractions tested by AB 484.
WHERE AN APPARENTLY CORRECT RESULT IS EASY TO GET. Sample and filling: bomb and gasoline outside 15–25 °C; flushing with oxygen (removal of air) omitted; tightness not checked — threshold 7.0 kPa in 10 min, and a steady drop well over 14 kPa in 15 min during the first 30 min means discarding the test. Bath: not boiling vigorously or temperature outside 98–102 °C; the test temperature is the average of the readings, not the set value. Breakpoint: requires a drop of exactly 14 kPa in 15 min before it and not less than 14 kPa in 15 min after it — readings less frequent than every 15 min may shift it. Calculation: the formula depends on whether the test temperature was above or below 100 °C; without conversion the result is the period at the test temperature, not at 100 °C. Cleanliness: residues of gum and gasoline in the bomb and container; after washing the container is handled only with forceps.
WHAT WE DO NOT GIVE. Parts read: the iTeh sample of ISO 7536:1994 (English version: foreword, clauses 1–11) and the sample of SIST ISO 7536:1996; formulae and figures of clauses 4–10 checked on the page images. We do not give the content of the normative Annexes A (bomb material) and B (thermometer) or the dimensions from Figures 1 and 2 — we have not read them. We have not examined differences between PN-ISO 7536:1997 and PN-EN ISO 7536:2011 (both implement ISO 7536:1994).
Method principle
50 ml ± 1 ml of finished gasoline in a glass container is placed in a pressure bomb of corrosion-resistant steel, flushed and filled with oxygen to 690–705 kPa at 15–25 °C and immersed in a vigorously boiling water bath (98–102 °C); the pressure is recorded or read every 15 min up to the breakpoint (a drop of exactly 14 kPa in 15 min before it, not less than 14 kPa in 15 min after it). The time from immersion to the breakpoint is the induction period at the test temperature, converted to 100 °C: IP100 = (IPt)[1 + 0.101 (Ta − 100)] or IP100 = (IPt)/[1 + 0.101 (100 − Tb)], result to the nearest whole minute (ISO 7536:1994, clauses 2–9).
Applications
- Aviation and motor gasolines — finished products only (ISO 7536:1994, 1.1)
- Indication of the tendency of gasoline to form gum in storage — with the reservation that the correlation varies (1.3)
- Not for individual gasoline components, particularly with a high percentage of low-boiling unsaturated compounds — risk of explosion (1.2)
- Replaced in PN the standard PN-ISO 7536:1997 (PKN card)
- In accreditation scopes: 5 Polish laboratories — unleaded, aviation gasoline, gasoline with bioethanol, liquid biohydrocarbons
Key parameters
| Parameter | Value |
|---|---|
| STATUS (PKN card as of 11.11.2026) | PN-EN ISO 7536:2011 current (Polish version, published 23-04-2013); replaced PN-ISO 7536:1997; ISO 7536:1994 confirmed (iTeh, stage 9093) |
| Sample — condition (7.1) | 50 ml ± 1 ml; bomb and gasoline at 15–25 °C |
| Oxygen — condition (4.4, 7.1) | purity ≥ 99 %, extra-dry; filling to 690–705 kPa, flushing with release ≤ 350 kPa/min |
| Tightness — condition (7.1, 7.2) | drop ≤ 7.0 kPa in 10 min; leak in the first 30 min (steady drop well > 14 kPa in 15 min) — test discarded |
| Bath — condition (5.9, 7.2) | water boiling vigorously, 98–102 °C (read to 0.1 °C); ≥ 18 l per bomb, depth ≥ 290 mm, bomb lid ≥ 50 mm below the surface |
| Breakpoint — definition (2.1) | pressure drop of exactly 14 kPa in 15 min before it and not less than 14 kPa in 15 min after it |
| Result — calculation (clauses 8–9) | IP100 = (IPt)[1 + 0.101 (Ta − 100)] or (IPt)/[1 + 0.101 (100 − Tb)]; to the nearest whole minute |
| Precision (clause 10) | repeatability — deviation from the mean > 5 % in 1 case in 20; reproducibility — > 10 % in 1 case in 20 |
| Coverage in accreditation scopes (database, read 11.11.2026) | 5 Polish laboratories, 5 records; 1 Czech laboratory |
Standard
- Standard number
- PN-EN ISO 7536:2011 (Polish version; implements EN ISO 7536:1996 and ISO 7536:1994 [IDT])
- Title (PL)
- Przetwory naftowe — Oznaczanie odporności benzyn na utlenianie — Metoda okresu indukcyjnego
- Title (EN)
- Petroleum products — Determination of oxidation stability of gasoline — Induction period method
Step-by-step procedure
-
Preparation
Container: gum solvent, water, hot detergent, then only with forceps, tap and distilled water, drying at 100–150 °C ≥ 1 h; bomb, stem, rod, valve — gum solvent; everything dry (clause 6).
-
Filling
Bomb and gasoline at 15–25 °C; 50 ml ± 1 ml of sample, cover, oxygen to 690–705 kPa, release ≤ 350 kPa/min, refilling; tightness — ≤ 7.0 kPa in 10 min (7.1).
-
Oxidation
Bomb into the vigorously boiling bath (98–102 °C), time of immersion = start; temperature to 0.1 °C, average = test temperature; pressure recorded continuously or every ≤ 15 min (7.2).
-
Breakpoint
Until a drop of exactly 14 kPa in 15 min followed by a drop of ≥ 14 kPa in 15 min; leak in the first 30 min — discard; time in minutes = period at the test temperature; cooling of the bomb and release of gas ≤ 350 kPa/min (7.2–7.3).
-
Calculation and test report
IP100 according to the formula for a temperature above or below 100 °C, to the nearest whole minute; report according to clause 11. PN-EN ISO 7536:2011 — current (PKN card as of 11.11.2026).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Pressure bomb (5.1) | Corrosion-resistant steel, polished interior; working pressure 1 240 kPa at 100 °C; inside dimensions according to Figure 1, material according to Annex A | — |
| Burst-disc assembly and valve (5.5–5.7) | Stainless steel, ruptures above 1 530 kPa ± 10 %; needle valve and quick-release coupling for oxygen | — |
| Pressure gauge (5.8) | Indicating or recording, to ≥ 1 380 kPa, division ≤ 35 kPa, accuracy ≤ 1 % of the scale interval; volume of tubing with stem ≤ 30 ml | — |
| Oxidation bath and thermometer (5.9–5.10) | Boiling water, ≥ 18 l for one bomb and 8 l for each additional bomb, with a condenser; thermometer 95–103 °C (Annex B) | — |
| Sample container with cover and forceps (5.3, 5.11) | Glass container and cover according to Figure 2; forceps of corrosion-resistant steel | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Toluene | — | Purity ≥ 99 % (4.1) |
| Acetone | — | Purity ≥ 99 % (4.2) |
| Gum solvent | — | Mixture of equal volumes of toluene and acetone (4.3) |
| Oxygen | — | Commercially available, extra-dry, purity ≥ 99 % (4.4) |
| Detergent cleaning solution | — | Cleans containers as well as 6 h in fresh chromic acid cleaning solution; type and conditions established by the laboratory (4.5) |
Health and safety (OHS)
- The standard warns: the bomb should be operated behind an appropriate safety shield to protect against possible explosive rupture (clause 3); the assembly has a burst-disc, and expelled gas is directed away from the operator (5.5)
- Chromic acid — potentially hazardous in contact with organic materials, toxic and highly corrosive; if used, full-face shield, full-length protective clothing and gloves (warning in 4.5); spent cleaning solutions are disposed of as toxic waste (6.2)
- Gasoline components with a high percentage of low-boiling unsaturated compounds may cause explosive conditions in the apparatus — the method does not cover them (1.2)
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN ISO 7536:2011 (Polish version; implements EN ISO 7536:1996 and ISO 7536:1994 [IDT]) — “Petroleum products — Determination of oxidation stability of gasoline — Induction period method”.
How does this method work?
50 ml ± 1 ml of finished gasoline in a glass container is placed in a pressure bomb of corrosion-resistant steel, flushed and filled with oxygen to 690–705 kPa at 15–25 °C and immersed in a vigorously boiling water bath (98–102 °C); the pressure is recorded or read every 15 min up to the breakpoint (a drop of exactly 14 kPa in 15 min before it, not less than 14 kPa in 15 min after it). The time from immersion to the breakpoint is the induction period at the test temperature, converted to 100 °C: IP100 = (IPt)[1 + 0.
What is the measuring range and accuracy?
STATUS (PKN card as of 11.11.2026): PN-EN ISO 7536:2011 current (Polish version, published 23-04-2013); replaced PN-ISO 7536:1997; ISO 7536:1994 confirmed (iTeh, stage 9093); Sample — condition (7.1): 50 ml ± 1 ml; bomb and gasoline at 15–25 °C; Oxygen — condition (4.4, 7.1): purity ≥ 99 %, extra-dry; filling to 690–705 kPa, flushing with release ≤ 350 kPa/min; Tightness — condition (7.1, 7.2): drop ≤ 7.0 kPa in 10 min; leak in the first 30 min (steady drop well > 14 kPa in 15 min) — test discarded.
How long does the test take?
The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Pressure bomb (5.1), Burst-disc assembly and valve (5.5–5.7), Pressure gauge (5.8), Oxidation bath and thermometer (5.9–5.10), Sample container with cover and forceps (5.3, 5.11).
Where is this test used?
Aviation and motor gasolines — finished products only (ISO 7536:1994, 1.1); Indication of the tendency of gasoline to form gum in storage — with the reservation that the correlation varies (1.3); Not for individual gasoline components, particularly with a high percentage of low-boiling unsaturated compounds — risk of explosion (1.2); Replaced in PN the standard PN-ISO 7536:1997 (PKN card); In accreditation scopes: 5 Polish laboratories — unleaded, aviation gasoline, gasoline with bioethanol, liquid biohydrocarbons.
What safety precautions apply?
The standard warns: the bomb should be operated behind an appropriate safety shield to protect against possible explosive rupture (clause 3); the assembly has a burst-disc, and expelled gas is directed away from the operator (5.5); Chromic acid — potentially hazardous in contact with organic materials, toxic and highly corrosive; if used, full-face shield, full-length protective clothing and gloves (warning in 4.5); spent cleaning solutions are disposed of as toxic waste (6.2); Gasoline components with a high percentage of low-boiling unsaturated compounds may cause explosive conditions in the apparatus — the method does not cover them (1.2).
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 7536.