🥜 Food Allergens (ELISA)

Food Chemistry AOAC 2012.01 (seria)

Quantitative determination of food allergens (gluten, milk, egg, peanuts, soy etc.) in food by enzyme-linked immunosorbent assay (ELISA) using certified test kits.

Overview

Determination of food allergens is a key element of food safety — Regulation (EU) 1169/2011 requires declaration of 14 major allergens on food product labels. The ELISA (Enzyme-Linked Immunosorbent Assay) method is the most widely used laboratory technique for quantitative determination of allergenic proteins in food and contact surfaces.

The method principle is based on specific antigen–antibody binding. Two main formats are available: sandwich ELISA — detects intact proteins and large fragments with two epitopes, and competitive ELISA — also detects small protein fragments, which is crucial for highly thermally processed food.

Most commonly used commercial kits are: RIDASCREEN (R-Biopharm), Veratox and Reveal (Neogen), AgraQuant (Romer Labs) and ELISA Systems. These kits have AOAC-OMA (Official Methods of Analysis) validation and are recommended by Codex Alimentarius. Typical determined allergens are: gluten (wheat, rye, barley prolamins), milk proteins (casein, β-lactoglobulin), egg proteins (ovalbumin, ovomucoid), peanut proteins (Ara h1, Ara h2), soy, nuts, sesame and mustard proteins.

ELISA detection limits are typically 1–5 mg/kg (ppm) of allergen in the product, allowing assessment of compliance with safety thresholds. Laboratories performing these tests participate in PT programs (FAPAS, LGC) and are accredited according to ISO 17025.

Method principle

In sandwich ELISA: microplate wells are coated with antibodies specific for a given allergen. Food sample extract is incubated in the well — allergen binds to antibodies. After washing, second enzyme-labeled antibody (HRP or AP) is added, which binds to another epitope of the allergen (forming a "sandwich"). After re-washing, enzyme substrate is added — a colored product is formed, whose intensity (absorbance) is proportional to allergen concentration. Reading on microplate reader at 450 nm.

Applications

Key parameters

ParameterValue
Limit of detection (LOD)1–5 mg/kg (ppm) of allergen
Determination range5–200 mg/kg (typical calibration curve range)
Analysis time30–90 min (depending on kit — FAST vs standard)
Format96-well microplate (12 strips × 8 wells)
Reading wavelength450 nm (with reference 620–650 nm)
Gluten limit according to Codex< 20 mg/kg gliadin (= < 20 ppm gluten)

Standard

Standard number
AOAC 2012.01 / AOAC 2014.03 / AOAC 991.19 (seria metod ELISA)
Title (PL)
Oznaczanie alergenów pokarmowych metodą immunoenzymatyczną ELISA
Title (EN)
Determination of food allergens by enzyme-linked immunosorbent assay (ELISA)

Step-by-step procedure

1. Sample preparation

Homogenize food sample to obtain uniform mass. Weigh 1 g into extraction tube.

⏱ Time: 10 min

2. Extraction

Add 10 mL extraction buffer from kit. Shake/mix for 10 min at room temperature (or 60°C for gluten with Mendez cocktail).

⏱ Time: 10–40 min

3. Centrifugation and dilution

Centrifuge 10 min at 2500 × g. Collect supernatant. Dilute according to kit instructions (typically 1:10 or 1:100).

⏱ Time: 15 min

4. Application on microplate

Apply 100 µL of calibration standards (5 concentrations + zero), samples and controls to appropriate wells. Each in duplicate.

⏱ Time: 5 min

5. Incubation I

Incubate microplate 30 min at room temperature (or 10 min for FAST kits). Gentle shaking.

⏱ Time: 10–30 min

6. Washing

Wash 3× with 300 µL washing buffer. Remove residual liquid by inverting and blotting on absorbent paper.

7. Incubation with conjugate

Add 100 µL of antibody-enzyme solution (conjugate). Incubate 30 min RT.

⏱ Time: 30 min

8. Washing II

Wash 3× as in step 6.

9. Color detection

Add 100 µL enzyme substrate (TMB). Incubate 15 min in darkness. Add 100 µL stop solution.

⏱ Time: 15 min

10. Absorbance reading

Measure absorbance on microplate reader at 450 nm (reference 620 nm). Perform within 30 min of stopping.

⏱ Time: 5 min

11. Results calculation

Plot calibration curve (absorbance vs concentration). Read allergen concentration in samples. Convert to mg/kg in product.

Required equipment and apparatus

EquipmentExampleIndicative price
Microplate readerTecan Sunrise, BioTek ELx800, Thermo Multiskan FC15 000–50 000 PLN
Incubator (37°C or RT)Microplate shaking incubator, Heidolph Titramax 10003 000–8 000 PLN
Microplate washerBioTek ELx50, Tecan HydroFlex, Thermo Wellwash15 000–40 000 PLN
HomogenizerIKA Ultra-Turrax T25, Stomacher 4005 000–15 000 PLN
Multichannel pipettesEppendorf Research Plus 8-channel (10–100 µL, 100–1200 µL)2 000–5 000 PLN
Analytical balanceMettler Toledo MS105DU, RADWAG AS 220.R28 000–20 000 PLN

Reagents, media and consumables

ReagentCASDetails
ELISA gluten kit (RIDASCREEN Gliadin)R-Biopharm R7001 — sandwich ELISA for gliadin determination, AOAC-OMA certificate, 96 tests, ~2 500 PLN
ELISA milk kit (RIDASCREEN FAST Milk)R-Biopharm R4652 — FAST 30 min, LOD 0.7 mg/kg, 48 tests
ELISA egg kit (Neogen Veratox Egg)Neogen 8470 — sandwich ELISA, LOD 2.5 mg/kg, 48 tests
ELISA peanuts kit (RIDASCREEN FAST Peanut)R-Biopharm R6802 — LOD 0.1 mg/kg, AOAC-OMA validation
ELISA soy kit (AgraQuant Soy)Romer Labs — competitive ELISA for soy proteins, 96 tests
Extraction bufferSupplied in kit or according to manufacturer's instructions (e.g. Mendez cocktail for gluten)

Health and safety (OHS)