🌿 1,4-Dioxane in workplace air — adsorption on activated carbon, desorption with a solution of propan-2-ol in carbon disulphide and gas chromatography, PN-Z-04431
In short
Determination of 1,4-dioxane in workplace air: adsorption on activated carbon, desorption with a solution of propan-2-ol in carbon disulphide and gas chromatography. The test is performed according to PN-Z-04431:2011 (Polish version); STATUS (as of 18.11.2026): current; card without "Replaces" and "Replaced by" fields. The procedure comprises 5 steps; it is used for: Measurements of 1,4-dioxane in workplace air — assessment against an NDS of 50 mg/m³ for two years from the entry into force of the regulation (§ 2(4)(2)), then against NDS 7.3 mg/m³ and NDSCh 22 mg/m³ (Journal of Laws 2026 item 447, item 195).
At a glance
- Standard: PN-Z-04431:2011 (Polish version)
- Category: Environment
- Procedure steps: 5
- STATUS (as of 18.11.2026): current; card without "Replaces" and "Replaced by" fields
- Edition (from the PKN card): PN-Z-04431:2011, Polish version; published 09-09-2011, 7 pages, KT 159 Chemical and Dust Hazards in the Working Environment, ICS 13.040.30
- Principle and determinability (from the card): adsorption on activated carbon, desorption with a solution of propan-2-ol in carbon disulphide, gas chromatography; lowest determinable concentration 5 mg/m³
Overview
STATUS (as of 18.11.2026). The card of PN-Z-04431:2011 (Polish version) does not have the heading "Withdrawn": published 09-09-2011, 7 pages, KT 159 Chemical and Dust Hazards in the Working Environment, ICS 13.040.30; there are no "Replaces" or "Replaced by" fields. The PKN search for "dioksanu" (18.11.2026) returns four items — the only current one is this standard.
WHAT THE STANDARD COVERS. The scope from the PKN catalogue card, quoted in full: "A method is given consisting in the adsorption of 1,4-dioxane on activated carbon, desorption of the compound with a solution of propan-2-ol in carbon disulphide and analysis of the solution obtained by gas chromatography. The lowest concentration of 1,4-dioxane that can be determined by this method is 5 mg/m3" (our translation).
LIMIT VALUES AND THE REGULATION ON THE METHOD. The NDS list — Annex 1 to the regulation of 12 June 2018 as worded by the regulation of 26 March 2026 (Journal of Laws 2026 item 447), item 195 "1,4-Dioxane" [123-91-1]: NDS (8-hour value) 7.3 mg/m³ (2 ppm), NDSCh (short-term value) 22 mg/m³ (6 ppm), "skin" notation. Under § 2(4)(2) of that regulation, "within two years from the date of entry into force" (it entered into force on 2 April 2026) the NDS for 1,4-dioxane is 50 mg/m³, and under § 2(5)(1) the NDS or NDSCh values from the annex apply "after two years from the date of entry into force"; the act gives no NDSCh for the transitional period. The lowest determinable concentration according to the card (5 mg/m³) is one tenth of the transitional NDS and about 68 % of the NDS in the list. According to the explanatory note to the list, the "skin" notation means that "absorption of the substance through the skin may be as important as exposure by inhalation" (our translation). The method is addressed by § 16 of the regulation on tests and measurements of agents harmful to health in the working environment (consolidated text — Journal of Laws 2025 item 949): "Tests and measurements are performed in accordance with the methods specified in Polish Standards and, where there are no such standards, with methods recommended and validated by the laboratories referred to in § 15(1) and (2)(1)" (our translation). The provision does not name this standard, and WE HAVE NOT IDENTIFIED any other provision that cites it.
PROCEDURE ACCORDING TO THE CIOP-PIB PROCEDURE (second source; numbers of the procedure, not of the standard). Jeżewska, "1,4-Dioksan – metoda oznaczania" (1,4-Dioxane – determination method), Podstawy i Metody Oceny Środowiska Pracy 2009, no. 1(59), pp. 141–146 — the same principle and the same determinability (5 mg/m³) as on the card of the standard. A tube with two sections of activated carbon (100 mg and 50 mg); up to 12 l of air is drawn through the tube at a flow rate not exceeding 12 l/h, sampling according to PN-Z-04008-7; samples kept in a freezer are stable for 6 days. Each section is desorbed separately with 1 ml of a mixture of carbon disulphide and propan-2-ol (98 + 2) for 30 min. Chromatograph with a flame ionisation detector; a column separating 1,4-dioxane from carbon disulphide, propan-2-ol and co-occurring substances (example in the procedure: capillary column with polyethylene glycol 2-nitroterephthalate, 50 m). Calibration curve from six solutions of 0.06–1.2 mg/ml. The desorption coefficient is determined for each batch of carbon; the amount in the shorter section should not exceed 10 % of the amount in the longer one, otherwise the result is indicative. Concentration: X = (m₁ + m₂) · 1000 / (V · d). Measuring range of the procedure 5–100 mg/m³ for a 12 l sample.
WHAT WE DO NOT GIVE HERE. We did not have the text of PN-Z-04431:2011 — PKN sells it (7 pages). We do not claim that the numbers of the CIOP-PIB procedure are those of the standard; we do not describe the newer CIOP-PIB procedure of 2024 (different determinability).
Method principle
1,4-Dioxane vapour from workplace air is adsorbed on activated carbon, desorbed with a solution of propan-2-ol in carbon disulphide, and the solution is analysed by gas chromatography; the lowest determinable concentration according to the card is 5 mg/m³. The CIOP-PIB procedure of 2009 with the same principle uses a flame ionisation detector. The result is compared with the NDS in force on the day of measurement: 50 mg/m³ in the transitional period, then NDS 7.3 mg/m³ and NDSCh 22 mg/m³ (NDS list of 2026).
Applications
- Measurements of 1,4-dioxane in workplace air — assessment against an NDS of 50 mg/m³ for two years from the entry into force of the regulation (§ 2(4)(2)), then against NDS 7.3 mg/m³ and NDSCh 22 mg/m³ (Journal of Laws 2026 item 447, item 195)
Key parameters
| Parameter | Value |
|---|---|
| STATUS (as of 18.11.2026) | current; card without "Replaces" and "Replaced by" fields |
| Edition (from the PKN card) | PN-Z-04431:2011, Polish version; published 09-09-2011, 7 pages, KT 159 Chemical and Dust Hazards in the Working Environment, ICS 13.040.30 |
| Principle and determinability (from the card) | adsorption on activated carbon, desorption with a solution of propan-2-ol in carbon disulphide, gas chromatography; lowest determinable concentration 5 mg/m³ |
| Limit values against which the result is assessed | for two years from the entry into force of the regulation (2.04.2026) NDS 50 mg/m³, no NDSCh (§ 2(4)(2)); then NDS 7.3 mg/m³ (2 ppm) and NDSCh 22 mg/m³ (6 ppm) from item 195 (§ 2(5)(1)); "skin" notation |
| CIOP-PIB procedure 2009 (numbers of the procedure) | activated carbon 100 mg/50 mg; up to 12 l of air, ≤ 12 l/h; desorption with 1 ml of CS₂ with propan-2-ol (98 + 2), 30 min; GC-FID; 5–100 mg/m³ for 12 l |
Standard
- Standard number
- PN-Z-04431:2011 (Polish version)
- Title (PL)
- Ochrona czystości powietrza — Oznaczanie 1,4-dioksanu na stanowiskach pracy metodą chromatografii gazowej
- Title (EN)
- Air purity protection — Determination of 1,4-dioxane at workplaces by gas chromatography
Step-by-step procedure
-
Sampling
Up to 12 l of air through the activated carbon tube, flow rate ≤ 12 l/h, according to PN-Z-04008-7 (CIOP-PIB procedure).
-
Desorption
Each carbon section separately: 1 ml of a mixture of carbon disulphide and propan-2-ol (98 + 2), 30 min with shaking (CIOP-PIB procedure).
-
Chromatographic analysis
GC-FID, two measurements of each solution; calibration curve from six solutions (CIOP-PIB procedure).
-
Breakthrough check
The amount in the shorter section should not exceed 10 % of the amount in the longer one — otherwise the result is indicative (CIOP-PIB procedure).
-
Result and assessment
X = (m₁ + m₂) · 1000 / (V · d) in mg/m³; comparison with the value in force on the day of measurement (NDS 50 mg/m³ in the transitional period, then NDS 7.3 and NDSCh 22 mg/m³). "Skin" notation: the measurement in air does not cover absorption through the skin.
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Activated carbon tubes (100 mg/50 mg) | CIOP-PIB procedure 2009; desorption coefficient for each batch | — |
| Suction pump | Constant flow rate not exceeding 12 l/h, up to 12 l of air (CIOP-PIB procedure) | — |
| Gas chromatograph with flame ionisation detector | Column separating 1,4-dioxane from carbon disulphide and propan-2-ol (CIOP-PIB procedure) | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Mixture of carbon disulphide and propan-2-ol (98 + 2) | — | Desorption solution, 1 ml per sorbent section (CIOP-PIB procedure; card of the standard: "solution of propan-2-ol in carbon disulphide") |
| 1,4-Dioxane — reference substance | 123-91-1 | Working standard solutions of 0.06–1.2 mg/ml (CIOP-PIB procedure) |
Health and safety (OHS)
- Work with organic solvents is carried out in an efficient laboratory fume hood, and used solutions are collected in containers intended for that purpose (CIOP-PIB procedure 2009, point 4.3)
- The "skin" notation in the NDS list: the measurement in air does not cover absorption through the skin (1,4-dioxane, item 195)
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-Z-04431:2011 (Polish version) — “Air purity protection — Determination of 1,4-dioxane at workplaces by gas chromatography”.
How does this method work?
1,4-Dioxane vapour from workplace air is adsorbed on activated carbon, desorbed with a solution of propan-2-ol in carbon disulphide, and the solution is analysed by gas chromatography; the lowest determinable concentration according to the card is 5 mg/m³. The CIOP-PIB procedure of 2009 with the same principle uses a flame ionisation detector.
What is the measuring range and accuracy?
STATUS (as of 18.11.2026): current; card without "Replaces" and "Replaced by" fields; Edition (from the PKN card): PN-Z-04431:2011, Polish version; published 09-09-2011, 7 pages, KT 159 Chemical and Dust Hazards in the Working Environment, ICS 13.040.30; Principle and determinability (from the card): adsorption on activated carbon, desorption with a solution of propan-2-ol in carbon disulphide, gas chromatography; lowest determinable concentration 5 mg/m³; Limit values against which the result is assessed: for two years from the entry into force of the regulation (2.04.2026) NDS 50 mg/m³, no NDSCh (§ 2(4)(2)); then NDS 7.3 mg/m³ (2 ppm) and NDSCh 22 mg/m³ (6 ppm) from item 195 (§ 2(5)(1)); "skin" notation.
How long does the test take?
The procedure comprises 5 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Activated carbon tubes (100 mg/50 mg), Suction pump, Gas chromatograph with flame ionisation detector.
Where is this test used?
Measurements of 1,4-dioxane in workplace air — assessment against an NDS of 50 mg/m³ for two years from the entry into force of the regulation (§ 2(4)(2)), then against NDS 7.3 mg/m³ and NDSCh 22 mg/m³ (Journal of Laws 2026 item 447, item 195).
What safety precautions apply?
Work with organic solvents is carried out in an efficient laboratory fume hood, and used solutions are collected in containers intended for that purpose (CIOP-PIB procedure 2009, point 4.3); The "skin" notation in the NDS list: the measurement in air does not cover absorption through the skin (1,4-dioxane, item 195).
Which laboratory can perform this test?