{"lab":{"_id":"6ade8b79ff06824761708175","accr_number":"AP 126","country":"PL","imported_at":"2026-10-15T12:00:00.000Z","accr_body":"PCA","activity_conducted":"w stałej lokalizacji (S)","activity_conducted_en":"at permanent location (S)","address":"ul. Podleśna 61, 01-673 Warszawa","approver":"Katarzyna Wiśniewska","categories":["metrology"],"certificate_date":"28.06.2019","city":"Warszawa","cmc_note":"Niepewność pomiaru dla CMC stanowi niepewność rozszerzoną przy prawdopodobieństwie rozszerzenia ok. 95 % i jest wyrażona w jednostkach wielkości mierzonej.","cmc_note_en":"The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units.","cycle_from":"17.06.2025","cycle_to":"13.07.2029","email_source":"PCA","en_official":false,"fields_of_testing":[{"codes":"14.02","pl":"wilgotność względna","en":"relative humidity"},{"codes":"17.01","pl":"ciśnienie","en":"pressure"},{"codes":"19.01","pl":"temperatura (termometria elektryczna)","en":"temperature (electrical thermometry)"}],"has_en":true,"has_zh":false,"issue":{"no":"14","date":"17.06.2025"},"issue_date":"17.06.2025","issue_number":"14","lab_division":"Centralne Laboratorium Aparatury Pomiarowej","lab_kind":"calibration","lab_type":"calibration","location":{"type":"Point","coordinates":[20.9614259,52.2816154]},"map_coord":[52.2816154,20.9614259],"map_geo":"nominatim-street","methods_count":3,"name":"Instytut Meteorologii i Gospodarki Wodnej - Państwowy Instytut Badawczy","parse_date":"2026-10-15T12:00:00.000Z","parse_source":"reczna-analiza-labcoda-3","parse_status":"parsed","parse_version":1,"parsed_at":"2026-10-15T12:00:00.000Z","pdf_files":["AP_126.pdf"],"scope_notes":[],"scope_summary":"Laboratorium wzorcujące. Wielkości mierzone: wilgotność względna, ciśnienie, temperatura (termometria elektryczna). W stałej lokalizacji (S).","search_text":"instytut meteorologii i gospodarki wodnej - państwowy instytut badawczy centralne laboratorium aparatury pomiarowej ul. podleśna 61, 01-673 warszawa laboratorium wzorcujące. wielkości mierzone: wilgotność względna, ciśnienie, temperatura (termometria elektryczna). w stałej lokalizacji (s). ap 126 ap126 laboratorium wzorcujące wzorcowanie kalibracja relative humidity pressure temperature (electrical thermometry) instytut meteorologii i gospodarki wodnej - panstwowy instytut badawczy centralne laboratorium aparatury pomiarowej ul. podlesna 61, 01-673 warszawa laboratorium wzorcujace. wielkosci mierzone: wilgotnosc wzgledna, cisnienie, temperatura (termometria elektryczna). w stalej lokalizacji (s). ap 126 ap126 laboratorium wzorcujace wzorcowanie kalibracja relative humidity pressure temperature (electrical thermometry)","source_lang":"pl","source_page_url":"https://bip.pca.gov.pl/wykazy-i-archiwa/akredytacje-aktywne/laboratoria-wzorcujace/AP%20126,podmiot.html","source_pdf_sha256":"16bc65246e010ad54238e913543e144c08438643e6b5c074b920b70416241b7e","source_pdf_url":"https://bip.pca.gov.pl/storage/file/e-accreditations/2026/10/15/4cb73ec175676f02edf5d52597895051/AP%20126.pdf","status":"active","updated_at":"2026-10-15T12:00:00.000Z","valid_from":"14-07-2009","valid_until":"13-07-2029","website":"www.imgw.pl"},"methods":[{"_id":"09c8a82c71d7d22ca1352ebf","lab_id":"6ade8b79ff06824761708175","country":"PL","accr_number":"AP 126","accr_body":"PCA","test_object":"Higrometry\nPrzetworniki","test_object_en":"Hygrometers\nTransducers","methodology":"Wzorcowanie — Wilgotność względna","methodology_en":"Calibration — Relative humidity","parameters":[{"name":"w temp. 20 °C","range":"10 %rh do 40 %rh","cmc":"1,5 %rh"},{"name":"w temp. 20 °C","range":"40 %rh do 60 %rh","cmc":"1,8 %rh"},{"name":"w temp. 20 °C","range":"60 %rh do 80 %rh","cmc":"2,2 %rh"},{"name":"w temp. 20 °C","range":"80 %rh do 95 %rh","cmc":"2,5 %rh"}],"parameters_en":[{"name":"at 20 °C","range":"10 %rh to 40 %rh","cmc":"1,5 %rh"},{"name":"at 20 °C","range":"40 %rh to 60 %rh","cmc":"1,8 %rh"},{"name":"at 20 °C","range":"60 %rh to 80 %rh","cmc":"2,2 %rh"},{"name":"at 20 °C","range":"80 %rh to 95 %rh","cmc":"2,5 %rh"}],"technique":"Procedura wewnętrzna; W stałej lokalizacji (S)","technique_en":"Internal procedure; At permanent location (S)","techniques":["Procedura wewnętrzna","W stałej lokalizacji (S)"],"techniques_en":["Internal procedure","At permanent location (S)"],"computed":[],"computed_en":[],"activity":"Wzorcowanie — Wilgotność względna\nZakres:\nw temp. 20 °C 10 %rh do 40 %rh CMC 1,5 %rh\nw temp. 20 °C 40 %rh do 60 %rh CMC 1,8 %rh\nw temp. 20 °C 60 %rh do 80 %rh CMC 2,2 %rh\nw temp. 20 °C 80 %rh do 95 %rh CMC 2,5 %rh\nProcedura wewnętrzna; W stałej lokalizacji (S)","reference":"P.W.-HCB-02/CLAP","norm_bases":["P.W.-HCB-02/CLAP"],"internal_procedure":true,"lab_section":null,"lab_section_en":null,"lab_address":null,"regulatory_area":"Wilgotność względna","regulatory_area_en":"Relative humidity","outside_premises":false,"location_of_activity":"S","note":null,"note_en":null,"search_text":"Higrometry\nPrzetworniki Wzorcowanie — Wilgotność względna\nZakres:\nw temp. 20 °C 10 %rh do 40 %rh CMC 1,5 %rh\nw temp. 20 °C 40 %rh do 60 %rh CMC 1,8 %rh\nw temp. 20 °C 60 %rh do 80 %rh CMC 2,2 %rh\nw temp. 20 °C 80 %rh do 95 %rh CMC 2,5 %rh\nProcedura wewnętrzna; W stałej lokalizacji (S) P.W.-HCB-02/CLAP Higrometry\nPrzetworniki Wzorcowanie — Wilgotnosc wzgledna\nZakres:\nw temp. 20 °C 10 %rh do 40 %rh CMC 1,5 %rh\nw temp. 20 °C 40 %rh do 60 %rh CMC 1,8 %rh\nw temp. 20 °C 60 %rh do 80 %rh CMC 2,2 %rh\nw temp. 20 °C 80 %rh do 95 %rh CMC 2,5 %rh\nProcedura wewnetrzna; W stalej lokalizacji (S) P.W.-HCB-02/CLAP","search_en":"Hygrometers\nTransducers Calibration — Relative humidity\nScope:\nat 20 °C 10 %rh to 40 %rh CMC 1,5 %rh\nat 20 °C 40 %rh to 60 %rh CMC 1,8 %rh\nat 20 °C 60 %rh to 80 %rh CMC 2,2 %rh\nat 20 °C 80 %rh to 95 %rh CMC 2,5 %rh\nInternal procedure; At permanent location (S) P.W.-HCB-02/CLAP","categories":["metrology"],"categories_source":"reczne","source_lang":"pl","translation":{"en":"reczna-analiza-labcoda-3"},"row_index":0,"parse_version":1,"parse_source":"reczna-analiza-labcoda-3"},{"_id":"9f1125ea97eff17d1f616831","lab_id":"6ade8b79ff06824761708175","country":"PL","accr_number":"AP 126","accr_body":"PCA","test_object":"Ciśnieniomierze elektroniczne","test_object_en":"Electronic pressure gauges","methodology":"Wzorcowanie — Ciśnienie","methodology_en":"Calibration — Pressure","parameters":[{"name":"ciśnienie absolutne (czynnik gaz)","range":"600 hPa do 1100 hPa","cmc":"0,17 hPa"}],"parameters_en":[{"name":"absolute pressure (gas medium)","range":"600 hPa to 1100 hPa","cmc":"0,17 hPa"}],"technique":"Procedura wewnętrzna; W stałej lokalizacji (S)","technique_en":"Internal procedure; At permanent location (S)","techniques":["Procedura wewnętrzna","W stałej lokalizacji (S)"],"techniques_en":["Internal procedure","At permanent location (S)"],"computed":[],"computed_en":[],"activity":"Wzorcowanie — Ciśnienie\nciśnienie absolutne (czynnik gaz) 600 hPa do 1100 hPa CMC 0,17 hPa\nProcedura wewnętrzna; W stałej lokalizacji (S)","reference":"P.W.-PCB-1/CLAP w oparciu o Euramet/cg-17/v.4.1","norm_bases":["P.W.-PCB-1/CLAP","Euramet/cg-17/v.4.1"],"internal_procedure":true,"lab_section":null,"lab_section_en":null,"lab_address":null,"regulatory_area":"Ciśnienie","regulatory_area_en":"Pressure","outside_premises":false,"location_of_activity":"S","note":null,"note_en":null,"search_text":"Ciśnieniomierze elektroniczne Wzorcowanie — Ciśnienie\nciśnienie absolutne (czynnik gaz) 600 hPa do 1100 hPa CMC 0,17 hPa\nProcedura wewnętrzna; W stałej lokalizacji (S) P.W.-PCB-1/CLAP w oparciu o Euramet/cg-17/v.4.1 Cisnieniomierze elektroniczne Wzorcowanie — Cisnienie\ncisnienie absolutne (czynnik gaz) 600 hPa do 1100 hPa CMC 0,17 hPa\nProcedura wewnetrzna; W stalej lokalizacji (S) P.W.-PCB-1/CLAP w oparciu o Euramet/cg-17/v.4.1","search_en":"Electronic pressure gauges Calibration — Pressure\nabsolute pressure (gas medium) 600 hPa to 1100 hPa CMC 0,17 hPa\nInternal procedure; At permanent location (S) P.W.-PCB-1/CLAP w oparciu o Euramet/cg-17/v.4.1","categories":["metrology"],"categories_source":"reczne","source_lang":"pl","translation":{"en":"reczna-analiza-labcoda-3"},"row_index":1,"parse_version":1,"parse_source":"reczna-analiza-labcoda-3"},{"_id":"00a285213526bc5c505e2e21","lab_id":"6ade8b79ff06824761708175","country":"PL","accr_number":"AP 126","accr_body":"PCA","test_object":"Czujniki termometrów rezystancyjnych\nPrzetworniki temperatury\nTermometry elektryczne","test_object_en":"Resistance thermometer sensors\nTemperature transducers\nElectrical thermometers","methodology":"Wzorcowanie — Temperatura (termometria elektryczna)","methodology_en":"Calibration — Temperature (electrical thermometry)","parameters":[{"name":"","range":"-30 °C do 50 °C","cmc":"0,05 °C"}],"parameters_en":[{"name":"","range":"-30 °C to 50 °C","cmc":"0,05 °C"}],"technique":"Procedura wewnętrzna; W stałej lokalizacji (S)","technique_en":"Internal procedure; At permanent location (S)","techniques":["Procedura wewnętrzna","W stałej lokalizacji (S)"],"techniques_en":["Internal procedure","At permanent location (S)"],"computed":[],"computed_en":[],"activity":"Wzorcowanie — Temperatura (termometria elektryczna)\n-30 °C do 50 °C CMC 0,05 °C\nProcedura wewnętrzna; W stałej lokalizacji (S)","reference":"P.W.-TCB-2/CLAP P.W.-TCB-3/CLAP P.W.-TCB-4/CLAP","norm_bases":["P.W.-TCB-2/CLAP","P.W.-TCB-3/CLAP","P.W.-TCB-4/CLAP"],"internal_procedure":true,"lab_section":null,"lab_section_en":null,"lab_address":null,"regulatory_area":"Temperatura (termometria elektryczna)","regulatory_area_en":"Temperature (electrical thermometry)","outside_premises":false,"location_of_activity":"S","note":null,"note_en":null,"search_text":"Czujniki termometrów rezystancyjnych\nPrzetworniki temperatury\nTermometry elektryczne Wzorcowanie — Temperatura (termometria elektryczna)\n-30 °C do 50 °C CMC 0,05 °C\nProcedura wewnętrzna; W stałej lokalizacji (S) P.W.-TCB-2/CLAP P.W.-TCB-3/CLAP P.W.-TCB-4/CLAP Czujniki termometrow rezystancyjnych\nPrzetworniki temperatury\nTermometry elektryczne Wzorcowanie — Temperatura (termometria elektryczna)\n-30 °C do 50 °C CMC 0,05 °C\nProcedura wewnetrzna; W stalej lokalizacji (S) P.W.-TCB-2/CLAP P.W.-TCB-3/CLAP P.W.-TCB-4/CLAP","search_en":"Resistance thermometer sensors\nTemperature transducers\nElectrical thermometers Calibration — Temperature (electrical thermometry)\n-30 °C to 50 °C CMC 0,05 °C\nInternal procedure; At permanent location (S) P.W.-TCB-2/CLAP P.W.-TCB-3/CLAP P.W.-TCB-4/CLAP","categories":["metrology"],"categories_source":"reczne","source_lang":"pl","translation":{"en":"reczna-analiza-labcoda-3"},"row_index":2,"parse_version":1,"parse_source":"reczna-analiza-labcoda-3"}],"has_pdf":true}